Вы находитесь на странице: 1из 1

2550 TEMPERATURE*

2550 A. Introduction

Temperature readings are used in the calculation of various peratures as a function of depth often are required. Elevated
forms of alkalinity, in studies of saturation and stability with temperatures resulting from discharges of heated water may
respect to calcium carbonate, in the calculation of salinity, in have significant ecological impact. The source of water sup-
a number of colorimetric tests, and in general laboratory ply, such as deep wells, often can be identified by temperature
operations. In limnological studies, knowledge of water tem- measurements alone. Industrial plants often require data on
water temperature for process use or heat-transmission calcu-
* Approved by Standard Methods Committee, 2010. lations.

2550 B. Laboratory and Field Methods

1. Laboratory and Other Non-Depth Temperature water long enough to permit complete equilibration. Report results
Measurements to the nearest 0.1 or 1.0°C, depending on need.
A reversing type thermometer is commonly used for depth
Normally, temperature can be measured using any standard liq- measurements. It often is mounted on the sample-collection
uid-in-glass or electronic thermometer with an analog or digital apparatus so a water sample can be obtained simultaneously.
readout. The device should be able to distinguish temperature Correct reversing-thermometer readings for changes due to dif-
changes of 0.1°C or less, and equilibrate rapidly (have a minimal ferences between temperature at reversal and temperature at time
thermal capacity). Abstain from using mercury-filled thermometers of reading. Calculate as follows (all units in degrees):
whenever possible to avoid the possibility of releasing mercury into

冋 册 冋 册
the environment if the thermometer breaks. To prevent breakage in (T 1⫺t)(T 1⫹V 0) (T 1⫺t)(T 1⫹V 0)
field operations, use a thermometer with a metal case. ⌬T ⫽ ⫻ 1⫹ ⫹L
K K
Periodically check the device’s bias (within the temperature
range of use) against a reference thermometer certified by the where:
National Institute of Standards and Technology (NIST, formerly
National Bureau of Standards) and using tolerances suggested in ⌬T ⫽ correction to be added algebraically to uncorrected reading,
NIST Handbook 105-6.* The certified thermometer must be used T1 ⫽ uncorrected reading at reversal,
with its certificate and correction chart. t ⫽ temperature at which thermometer is read,
A total immersion thermometer is designed to indicate tem- V0 ⫽ degree volume of thermometer, the volume of small bulb
peratures correctly when the bulb and the entire liquid column end of capillary up to 0°C graduation,
are exposed to the temperature being measured (except for a K ⫽ constant depending on relative thermal expansion of
minimal emergent length for handling). A partial-immersion mercury and glass (usual value of K ⫽ 6100), and
thermometer has a line around it at the immersion distance from L ⫽ calibration correction of thermometer depending on T1.
the bottom. It indicates correctly when the bulb and the liquid If series observations are made, it is convenient to prepare graphs
column to that line are exposed to the temperature being mea- for a thermometer to obtain ⌬T from any values of T1 and t.
sured and the emergent stem is at ambient temperature.
The quality control (QC) practices considered to be an integral 3. Bibliography
T2020 part of each method are summarized in Table 2020:II.
WARREN, H.F. & G.C. WHIPPLE. 1895. The thermophone—A new in-
2. Depth Temperature Measurements strument for determining temperatures. Mass. Inst. Technol. Quart.
8:125.
Depth temperatures required for limnological studies may be SVERDRUP, H.V., M.W. JOHNSON & R.H. FLEMING. 1942. The Oceans.
measured with a reversing thermometer, thermophone, or thermis- Prentice-Hall, Inc., Englewood Cliffs, N.J.
tor. The thermistor is most convenient and accurate; however, AMERICAN SOCIETY FOR TESTING & MATERIALS. 1949. Standard Specifi-
higher cost may preclude its use. Before field use, verify a temper- cations for ASTM Thermometers; No. E1-58. Philadelphia, Pa.
ature measurement device’s calibration according to NIST specifi- REE, W.R. 1953. Thermistors for depth thermometry. J. Amer. Water
cations. Make readings with the thermometer or device immersed in Works Assoc. 45:259.
NATIONAL INSTITUTE OF STANDARDS AND TECHNOLOGY. 1997. Specifications
and Tolerances for Reference Standards and Field Standard
* Some commercial thermometers may be as much as 3°C in error. Weights and Measures; NIST Handbook 105-6. Gaithersburg, Md.

https://doi.org/10.2105/SMWW.2882.031 1

Вам также может понравиться