GE Medical Systems
LOGIQ™ S6 Software Option Installation
Instructions
Operating Documentation
Important Precautions
Language
- 1
GE MEDICAL SYSTEMS
DIRECTION 5173830, REVISION 1 LOGIQ™ S6 SOFTWARE OPTION INSTALLATION INSTRUCTIONS
2 -
GE MEDICAL SYSTEMS
DIRECTION 5173830, REVISION 1 LOGIQ™ S6 SOFTWARE OPTION INSTALLATION INSTRUCTIONS
- 3
GE MEDICAL SYSTEMS
DIRECTION 5173830, REVISION 1 LOGIQ™ S6 SOFTWARE OPTION INSTALLATION INSTRUCTIONS
DAMAGE IN TRANSPORTATION
All packages should be closely examined at time of delivery. If damage is apparent write “Damage In
Shipment” on ALL copies of the freight or express bill BEFORE delivery is accepted or “signed for” by
a GE representative or hospital receiving agent. Whether noted or concealed, damage MUST be
reported to the carrier immediately upon discovery, or in any event, within 14 days after receipt, and the
contents and containers held for inspection by the carrier. A transportation company will not pay a claim
for damage if an inspection is not requested within this 14 day period.
Call Traffic and Transportation, Milwaukee, WI (262) 827-3468 or 8*285-3468 immediately after
damage is found. At this time be ready to supply name of carrier, delivery date, consignee name, freight
or express bill number, item damaged and extent of damage.
Complete instructions regarding claim procedure are found in Section “S” of the Policy And Procedures
Bulletins.
The purchaser of GE equipment shall only utilize qualified personnel (i.e., GE’s field engineers,
personnel of third-party service companies with equivalent training, or licensed electricians) to perform
electrical servicing on the equipment.
GE Medical Systems employees should use the iTrak System to report all documentation errors or
omissions.
4 -
GE MEDICAL SYSTEMS
DIRECTION 5173830, REVISION 1 LOGIQ™ S6 SOFTWARE OPTION INSTALLATION INSTRUCTIONS
Revision History
- 5
GE MEDICAL SYSTEMS
DIRECTION 5173830, REVISION 1 LOGIQ™ S6 SOFTWARE OPTION INSTALLATION INSTRUCTIONS
6 -
GE MEDICAL SYSTEMS
DIRECTION 5173830, REVISION 1 LOGIQ™ S6 SOFTWARE OPTION INSTALLATION INSTRUCTIONS
Section 1-1
Introduction
This Service instructions describe the LOGIQ™ S6 software option installation as follows:
NOTICE The Stress Echo, 4D, Vocal-volume calculation, and VCI-Volume contrast imaging require 512MB
memory option.
R.6.x.x or
Option Part Number HCAT# later
Advanced 3D 5171754 H41932LF Yes
B-Flow 5169730 H41932LD Yes
Coded Contrast 5173023 H41932LG Yes
DICOM 5173166 H41932LH Yes
LOGIQ View 5166914 H41932LE Yes
CrossXBeam 5173238 H41932LJ Yes
Report Writer 5154930 H41932LK Yes
Stress Echo 5172058 H41932LL Yes
PFD 5168447 H41932LM Yes
TVI 5173189 H41932LA Yes
SRI 5172730 H41932LN Yes
Vocal-Volume
Calculation 5173079 H41932LB Yes
VCI-Volume Contrast
Imaging 5173017 H41932LC Yes
LS6 M12L Enable Option 5175356 Yes
LS6 M7C Enable Option 5175187 Yes
LS6 M3S Enable Option 5175946 Yes
NOTICE Vocal-volume calculation and VCI-Volume contrast imaging require the realtime 4D option.
- 1-1
GE MEDICAL SYSTEMS
DIRECTION 5173830, REVISION 1 LOGIQ™ S6 SOFTWARE OPTION INSTALLATION INSTRUCTIONS
Section 1-2
Tools Required
Option strings
Section 1-3
Time Required
Approximately 15 minutes plus travel time
Section 1-4
Parts Required
The software options listed in Introduction contain the same parts in the kit.
Section 1-5
Procedure
READ and UNDERSTAND these instructions thoroughly before proceeding with the installation.
Perform each step in sequence and check it off when completed. If a problem occurs after completion,
go back through the procedures and check for implementation errors before contacting your Ultrasound
Region Engineer.
1-5-1 Preparations
• Option strings (This is a option key with a sequence of letters and numbers. Consult your local GE
personnel.)
- 1-3
GE MEDICAL SYSTEMS
DIRECTION 5173830, REVISION 1 LOGIQ™ S6 SOFTWARE OPTION INSTALLATION INSTRUCTIONS
6.) Enter Option string into New Key dialogue box. Click on Add, then Save.
Add
Key String
Entry
Disable
1-5-3-1 Advanced3D
1.) Press 3D key on the keyboard.
2.) Touch 3D soft key on the touch panel.
1-5-3-2 B-Flow
NOTICE You can use 10L, M12L, 7L, 12L, 3.5C, 3.5CS, 4C, or T739 probe for this check-out.
- 1-5
GE MEDICAL SYSTEMS
DIRECTION 5173830, REVISION 1 LOGIQ™ S6 SOFTWARE OPTION INSTALLATION INSTRUCTIONS
2.) Select the related probe on the touch panel.
3.) Verify that “Option not available” message does NOT appear when the B-Flow button is pressed.
NOTICE You can use 10L, 7L, 3.5C, 3.5CS, 4C, M3S or 3S probe for this check-out.
Verify that “Option not available” message does NOT appear when the Contrast button is pressed.
1-5-3-4 DICOM
DICOM Setting
After installing DICOM option software, you must set the parameters to establish communications
between the scanner and DICOM device(s).
Item Descriptions
Select the device name to which the scanner is connected via network.
Name
When the DICOM device is not listed in this menu, add a new device using Add button.
IP address Enter IP Address of the DICOM device to which the scanner is connected via network.
Item Descriptions
- 1-7
GE MEDICAL SYSTEMS
DIRECTION 5173830, REVISION 1 LOGIQ™ S6 SOFTWARE OPTION INSTALLATION INSTRUCTIONS
Item Descriptions
Select
This is the same parameter as Service Type.
Destinations
Functional Check
Send an image via DICOM to a DICOM device (server, printer.), then verify that it should be received
with no error.
1-5-3-5 LOGIQView
Verify that “Option not available” message does NOT appear when the LOGIQView button is pressed.
- 1-9
GE MEDICAL SYSTEMS
DIRECTION 5173830, REVISION 1 LOGIQ™ S6 SOFTWARE OPTION INSTALLATION INSTRUCTIONS
1-5-3-8 Stress Echo
1.) Select Preset > Card (Cardiac).
2.) Select Scan > Protocol. Verify that the protocol screens appears on the CRT monitor and the
protocol buttons appears on the LCD screen.
1-5-3-9 PFD
1.) Enter the color mode.
PFD button
- 1 - 11
GE MEDICAL SYSTEMS
DIRECTION 5173830, REVISION 1 LOGIQ™ S6 SOFTWARE OPTION INSTALLATION INSTRUCTIONS
1-5-3-10 TVI
1.) Connect a sector probe to the scanner.
2.) Select a cardiac application.
3.) Press the PFD button.
4.) Verify that TVI tab appears on the touch panel screen.
1-5-3-11 SRI
1.) Select B mode and touch SRI HD to verify that level selection button appears.
- 1 - 13
GE MEDICAL SYSTEMS
DIRECTION 5173830, REVISION 1 LOGIQ™ S6 SOFTWARE OPTION INSTALLATION INSTRUCTIONS
5.) Touch Vocal tab. Verify that the vocal menu appears on the screen.
- 1 - 15
GE MEDICAL SYSTEMS
DIRECTION 5173830, REVISION 1 LOGIQ™ S6 SOFTWARE OPTION INSTALLATION INSTRUCTIONS
1-5-4 Final Procedure
1.) Unscrew four screws and remove the Right Side Cover. Refer to Components Replacement in
Chapter 8, LOGIQ™ 7 Service Manual, 5162630 (included in LOGIQ S6 eDoc CD (5166927-200)).
2.) Attach the Option Label onto the Nest Cover as shown in Figure 1-6.
Nest Cover
Option Labels
3.) Re-install the Right Side Cover. Refer to Components Replacement in Chapter 8, LOGIQ™ 7
Service Manual, 5162630.
4.) Fill out the proper customer information on the Product Locator Card. Mail the installation card
“Product Locator” to the address corresponding to your region.
NOTE: The Product Locator Card indicated in Figure 1-7 may be different from the actually provided on.