Вы находитесь на странице: 1из 3

Radiation Protection Dosimetry (2010), Vol. 139, No. 1–3, pp. 310–312 doi:10.

1093/rpd/ncq011

QUALITY CONTROL AND PATIENT DOSIMETRY IN DENTAL


CONE BEAM CT
J. Vassileva* and D. Stoyanov
Department for Radiation Protection at Medical Exposure, National Centre of Radiobiology and Radiation
Protection, 132 Kliment Ohridsky blv., 1756 Sofia, Bulgaria

*Corresponding author: j.vassileva@ncrrp.org

This paper presents the initial experience in performing quality control and patient dose measurements in a cone beam

Downloaded from http://rpd.oxfordjournals.org/ at University of Birmingham on March 22, 2015


computed tomography (CT) scanner (ILUMATM Ultra, IMTEC Imaging, USA) for oral and maxillofacial radiology. The
X-ray tube and the generator were tested first, including the kVp accuracy and precision, and the half-value layer (HVL). The
following tests specific for panoramic dental systems were also performed: tube output, beam size and beam alignment to the
detector. The tests specific for CT included measurements of noise and CT numbers in water and in air, as well as the homo-
geneity of CT numbers. The most appropriate dose quantity was found to be the air kerma-area product (KAP) measured
with a KAP-metre installed at the tube exit. KAP values were found to vary from 110 to 185 mGy m2 for available adult
protocols and to be 54 mGy m2 for the paediatric protocol. The effective dose calculated with the software PCXMC (STUK,
Finland) was 0.05 mSv for children and 0.09 –0.16 mSv for adults.

INTRODUCTION USA). This system has a fixed FOV and a pyrami-


dal-shaped X-ray beam, directed towards a flat panel
Cone beam computed tomography (CBCT) is
detector on the other side of the patient’s head. The
increasingly being used in the oral and maxillofacial
C-arm with the X-ray assembly and detector per-
radiology practice since 2001 for implant planning,
forms a single 3608 rotation around the head,
assessment of bony and dental pathological con-
acquiring multiple 2D projection images. This varies
ditions, orthodontic treatment planning, temporo-
from a traditional medical CT which uses a fan-
mandibular joint imaging, etc.(1,2,3).
shaped X-ray beam acquiring axial image slices of
Published reports showed that the effective doses
the FOV. Reconstruction software generates a 3D
from CBCT are 5– 80 times higher than doses from
volumetric data set used to provide secondary recon-
single panoramic radiograph and 1 –23 % of a com-
structed images in axial, sagittal and coronal planes,
parable conventional CT(1 – 5). The CBCT dose
as well as multi-planar (oblique, curved, cross-sec-
varies substantially depending on the device, field of
tional) reformation and 3D visualisation. The
view (FOV) and selected technique factors.
system characteristics are summarised in Table 1.
The ‘basic principles’ on the use of dental CBCT,
The system operates in a continuous mode with a
recently established by the European Academy of
fixed tube voltage of 120 kVp; two steps of the tube
DentoMaxilloFacial Radiology and the project
current are used – 3.8 mA for adult patients and
‘SEDENTEXCT’ (safety and efficacy of a new and
1 mA for children. Scanning time is selectable
emerging dental X-ray modality) require a quality
between 20 and 40 ms.
assurance program to be implemented for each
CBCT facility, including acceptance tests and regular
routine tests to ensure optimal radiation protection
for staff and patients(6,7). No standard quality control
(QC) and dosimetry protocols are available in the lit- TESTING PROCEDURE
erature. Very few publications deal with QC for Because the CBCT system has a rotational geometry
CBCT(8,9); a few more present studies of patient like panoramic dental systems, tomographic recon-
doses(1 – 5). struction like a conventional ‘fan-beam’ CT and a
This paper presents the initial experience with large area detector like digital radiography/fluoro-
acceptance testing and patient dosimetry for one scopy systems, the elaborated QC program included
type of CBCT. relevant parts of the tests performed for these three
systems. A similar approach was reported by other
authors(8,9). For most of the parameters were
CBCT SYSTEM
adopted remedial levels (RLs) and suspension levels
Measurements were carried out on the latest full- (SLs), similar to the existing national requirements
view CBCT for craniofacial imaging (ILUMAw for conventional radiography, CT and dental
Ultra Cone Beam CT Scanner, IMTEC Imaging, systems(10).

# The Author 2010. Published by Oxford University Press. All rights reserved. For Permissions, please email: journals.permissions@oxfordjournals.org
QC AND DOSIMETRY IN DENTAL CBCT
Table 1. Technical specification of the cone beam CT scanner (ILUMATM Ultra, IMTEC Imaging, USA).

Tube voltage Fixed 120 kVp HF continuous mode

Tube current Selectable 1 mA or 3.8 mA


Scan time One rotation 360; selectable 20 s or 40 s
Reconstructed voxel size Isotropic; 0.4; 0.3; 0.2 or 0.1 mm
Grey scale 16 bits
X-ray detector Amorphous Si flat panel with active pixel area: size 193 mm  242 mm; matrix 1516  1900
FOV Fixed
Light field Two lasers for centring and a third for the midline

Tube voltage accuracy and precision central and four peripheral regions of interests (ROI)

Downloaded from http://rpd.oxfordjournals.org/ at University of Birmingham on March 22, 2015


with an area of 500 mm2.
Tube voltage (kVp) was measured with an X-ray
multimetre (Barracuda, RTI Electronics, Sweden).
The multipurpose detector (MPD) was centred at Patient dose
the image detector surface. Measurements were per-
formed in two possible modes—in the ‘scout’ mode Taking into account the rotational geometry and
(without rotation), using the MPD calibration for the large detector size, the most appropriate dose
radiographic systems, and in CT mode, using the quantity for patient dose measurements in CBCT
corresponding calibration of the MPD. Tube voltage was decided to be the air kerma-area product
precision was tested with three consecutive (KAP). Measurements were performed with a trans-
measurements. mission ion chamber (Diamentor E, PTW, Freiburg,
Germany) positioned at the tube exit. KAP was
measured for all available examination protocols.
Specific tube output Organ doses and the effective dose E were calcu-
Tube output was measured with a flat 30 cm3 ionis- lated with a commercial software (PCXMC v.2.0,
ation chamber (type 233612, PTW Freiburg, STUK, Finland). The examination for a typical
Germany) positioned in air at 60 cm from the tube patient was simulated with 12 views in 308 intervals.
focus. Measurements were performed in a ‘scout’ All available protocols for patient examinations were
mode. Consistency of the output was tested with simulated. The effective dose was calculated with
three repeated measurements. tissue weighting factors from Publication 60 and
Publication 103 of the ICRP.
Half-value layer
RESULTS AND DISCUSSION
The same detector and geometry as for the output
measurements were used also for half-value layer The system operates at a fixed tube voltage of
(HVL) measurements. Aluminium filters of 99.9 % 120 kVp. The measured tube voltage was 126 +
purity were used. 3 kV. The results were similar when measured in a
‘scout’ mode and in a rotational mode. For kVp
accuracy +6 kVp was applied as an RL and 15 %
Size of radiation beam and beam alignment
as an SL. The measured value was within the
to the detector
accepted limits but it was expected that for a new
Measurement was performed with a ready pack system the tube voltage accuracy should be better.
X-ray film with a size larger than the detector area, The possible reason for the found inaccuracy of the
positioned directly on the flat panel detector surface. tube voltage could be the calibration of the kVp-
The film was exposed during the full rotation at metre. The CBCT system operates at high kVp but
120 kV, 3.8 mA and 40 ms. with a low tube current, and corresponding cali-
bration of kVp-metres has to be introduced by the
manufacturers of measuring instruments.
CT number uniformity and image noise
The tube voltage precision was better than 1 %.
CT number uniformity and image noise were A maximum deviation of 5 % of each measured
measured with a cylindrical water phantom with a value from the mean is proposed as an RL for this
diameter of 16 cm, positioned at the patient’s head parameter.
support. Using the software incorporated into the Specific tube output consistency was tested with
system, the mean CT number and the standard devi- three consecutive measurements at fixed exposure
ation of mean CT numbers were calculated in a parameters and was found to be better than 1 %.

311
J. VASSILEVA AND D. STOYANOV
Table 2. Measured KAP and calculated effective doses.

Exposure parameters (protocol) Measured KAP, mGy m2 E, mSv (ICRP 60) E, mSv (ICRP 103)

3.8 mA; 40 s (standard adult) 184 126 157


3.8 mA; 20 s (low-dose adult) 110 74 94
1 mA; 20 s (paediatric) 54 37 46

The proposed RL for this parameter is a deviation CONCLUSION


of 5 % of each measured value from the mean.
The present study demonstrated the applicability of
The HVL measured at 120 kVp was 7.3 mm Al.
some of the QC tests to the CBCT for conventional
The same RL as for conventional CT was proposed
fan-beam CT, for panoramic dental systems and for
for an HVL of 3.8 mm Al at 120 kVp. The test of the

Downloaded from http://rpd.oxfordjournals.org/ at University of Birmingham on March 22, 2015


conventional radiography. The following additional
size of radiation beam and beam alignment to the
tests, not performed here, could also be included for
detector is important because of the strong influence
testing the flat panel detector quality: dark noise,
on the patient dose. It is expected that the radiation
image retention and resolution. In addition, image
field will not exceed the size of the active detector
quality tests are recommended, but for that purpose
area. The measured irradiated field was found to be
a dedicated phantom should be available.
185 mm  234 mm, which is lower than the declared
active detector area of 193 mm  242 mm.
CT number uniformity was tested by the calcu- REFERENCES
lation of mean CT numbers in five regions of the
image of the water phantom. Applicability of the cri- 1. Kau, C. H., Richmond, S., Palomo, J. M. and Hans,
teria for fan-beam CT was tested: maximum differ- M. G. Three-dimensional cone beam computerized tom-
ography in orthodontics. J. Orthod. 32, 282–93 (2005).
ence of +8 HU between CT numbers in different
2. Scarfe, W., Farman, A. and Sukovic, P. Clinical appli-
areas of the phantom. The results for the mean values cations of cone beam computed tomography in dental
were as follows: for the central ROI: 7.7 HU and for practice. J. Can. Dent. Assoc. 72(1), 72– 80 (2006).
the peripheral ROI: 16.6; 12.0; 22.0; 14.5. The found 3. Scarfe, W. and Farman, A. Cone beam computed tom-
deviations between mean CT numbers are higher than ography: a paradigm shift for clinical dentistry. Aus.
expected. The question under investigation is whether Dent. Pract. 102–110 (2007).
CT numbers in CBCT are the same as in axial CT. 4. Ludlow, J. B. and Ivanovic, M. Comparative dosimetry
The measured value of the standard deviation of of dental CBCT devices and 64-slice CT for oral and
the mean CT numbers in the central ROI, s ¼ 0.35, maxillofacial radiology. Oral Surg. Oral Med. Oral
was recorded as a baseline value for the image noise. Pathol. Oral Radiol. Endod. 106, 106–114 (2008).
5. Roberts, J. A., Drage, N. A., Davies, J. and Thomas,
Similar to CT, remedial actions at future QC tests D. W. Effective dose from cone beam CT examinations
will be needed when the deviation of s from the in dentistry. Br. J. Radiol. 82, 35–40 (2009).
baseline value exceeds 20 %. 6. Horner, K., Islam, M., Flygare, L., Tsiklakis, T. and
The measured air KAP and calculated effective Whaites, E. Basic principles for use of dental cone beam
doses for three available clinical protocols are pre- CT: consensus guidelines of the European Academy of
sented in Table 2. KAP values were found to vary Dental and Maxillofacial Radiology. Dentomaxillofac.
from 110 to 185 mGy m2 for available adult proto- Radiol. 38, 187– 195 (2009).
cols and to be 54 mGy m2 for the paediatric proto- 7. Radiation Protection: Cone beam CT for dental and
maxillofacial radiology. Provisional guideliness. Vol. 1.
col. The effective dose calculated with tissue
A report prepared by the SEDENTEX CT project.
weighing factors from the ICRP Publication 103 was (2009) Available on www.sedentexct.eu.
0.05 mSv for children and 0.09 –0.16 mSv for adults. 8. Pryor, M. and Hollaway, P. Dental cone beam CT–a
These values are 24–27 % higher than those calcu- review of our experience. Ninth CT users group
lated with the factors from the ICRP Publication 60. meeting. Birmingham (2007) Available on http://www.
The reason is the inclusion of the salivary glands ctug.org.uk/meet07-10-23/
and brain as individually weighted tissue with a 9. Betancourt Benı́tez, R., Ning, R., Conover, D., Zhang,
factor 0.01 and the addition of the oral mucosa to Y. and Cai, W. Development of a quality control
the remaining organs with a factor of 0.12. program for a cone beam CT imaging system. Proc.
The calculated effective doses are lower than those SPIE. 6913, DOI:10.1117/12.770427 (2008).
10. Ministry of Health of Republic of Bulgaria. Ordinance
calculated by Ludlow for the same CBCT system on No 30/31.10.2005 for protection of individuals at
the basis of phantom measurements of organ medical exposure. Promulgated in the State Gazette
doses(4). Further verification of the results is needed (official organ of the National Assembly of the
with a comparison of calculations and measure- Republic of Bulgaria) No 91, November 15 (2005).
ments for the same system. (In Bulgarian)

312

Вам также может понравиться