Академический Документы
Профессиональный Документы
Культура Документы
a
Department of Surgery, University Hospital Basel, Spitalstrasse 21, CH-4031 Basel, Switzerland
b
German Cancer Research Centre, INF 280, D-69120 Heidelberg, Germany
c
Department of Radiology, University Hospital Basel, Spitalstrasse 21, CH-4031 Basel, Switzerland
Accepted 28 April 2003
KEYWORDS Summary A prospective controlled clinical study was performed to compare fluoro-
Fluoroscopy-based scopic guidance with fluoroscopy-based surgical navigation for distal locking of intra-
surgical navigation; medullary implants. Forty-two patients with fractures of the lower extremity treated
Virtual fluoroscopy; by intramedullary nailing were divided in two groups: distal locking either with
Distal locking; fluoroscopic guidance (group I) or with surgical navigation (group II). The average
fluoroscopic time to insert one interlocking screw with fluoroscopic guidance was 108 s
Intramedullary nailing;
compared with 7.3 s in the navigation group. The average procedure time to insert one
Exposure to radiation
interlocking screw in group I was 13.7 min compared with 17.9 min in group II. The drill
bit failed to pass through the interlocking hole in one patient from group II. There
was no significant difference in the technical reliability between both groups. Fluoro-
scopic times to achieve equivalent precision are reduced with fluoroscopy-based
surgical navigation compared with fluoroscopic guidance. Fluoroscopy-based surgical
navigation is recommended for intraoperative guidance in situations where reduction
of exposure to radiation is considered advantageous over the increase of procedure
time.
ß 2003 Elsevier Ltd. All rights reserved.
Figure 1 Image acquisition with fluoroscopy-based surgical navigation for distal locking in revision surgery: patient’s
feet left margin, patient’s head right margin, patient in supine position. A Grosse Kempf femoral nail had to be
removed from the left femur for implant failure, spongious bone grafting was performed and a long PFNÕ nail inserted
into the medullary canal (see large incision). The fluoroscope is aligned in the lateral projection for image acquisition.
The fluoroscope’s reference base (A, arrow) and the nail reference base (B, arrow) must be in line of sight with the
infrared tracking unit (not visible).
nail insertion handle are tracked. A graphic display The precision of passing the drill through the
of the drill bit then is overlaid on the stored image of interlocking holes in the nail was assessed by the
the distal locking holes in real time. The drill is then surgeon as ‘‘successful without contact with the
introduced through an incision until it contacts nail’’, ‘‘successful with slight or severe contact
bone. The drill bit is moved along the surface of with the nail’’, or as ‘‘failure in distal locking’’.
the bone until its projection on the computer screen The failure rate of a guidance method was cal-
is located exactly in the centre of the distal locking culated as number of failures in distal locking per
hole. After the drill’s axis has been aligned with the number of interlocking screws inserted using the
axis of the interlocking hole the drill is driven into relevant guidance method.
the bone and through the interlocking hole in the Intraoperative problems during the interlocking
nail into the opposite cortex (Fig. 2). Precision in procedure were classified as related to lack of
passing the drill through the interlocking holes is precision or as related to the technical equipment
evaluated as described below. The interlocking that was used for guidance. The severity of intrao-
screw’s correct position is confirmed with intrao- perative problems was rated as major or minor by
perative check fluoroscopy. the surgeon as follows: technical problems were
rated as major when they could not be solved within
Documentation 10 min during the procedure and were rated as
minor in all other cases.
The data evaluated intraoperatively included A major problem related to lack of precision was
fluoroscopic time and procedure time, precision defined as intraoperative situation when measures
in passing the drill bit through the interlocking had to be taken in order to improve guidance of the
holes, type and severity of problems with the inter- drill. This included detection of failures in distal
locking procedure, and the experience of the sur- locking or severe contact to the nail. A minor
geon with the guidance method. problem related to lack of precision did not mean
Documented fluoroscopic and procedure time in any change of the procedure. This included slight
both groups included all steps beginning with c-arm contact between the drill bit and the nail.
alignment for image acquisition ending with check In patients within group II in whom a major intra-
fluoroscopy to verify correct screw position. operative problem was recognized, the scheduled
570 N. Suhm et al.
Figure 2 Continuous intraoperative guidance with fluoroscopy-based surgical navigation: the c-arm fluoroscope is
removed from the patient and the computer screen positioned opposite to the surgeon. The nail’s reference base (B,
arrow) and the compact air drive’s reference base (C, arrow) must be in line of sight for the infrared tracking unit (not
visible). Based on the actual position data, a graphic display of the drill bit is overlaid on the stored fluoroscopic
images.
performed with use of the Statistical Analysis Soft- Table 2 The surgeon’s experience with guidance
ware (SASÕ, version 8.E for Windows; The SAS Insti- methods used in the study
tute, Heidelberg, Germany). All continuous data are
Group I Group II
expressed as the mean and the standard deviation.
For all tests, probability less than 0.05 was consid- Guidance Fluoroscopic Surgical
ered significant. method guidance navigation
Experience JR/SR/C 3/13/5 5/18/0
Fracture types included in the study (fracture classification according to the ASIF), type of intramedullary nail used
for stabilization, geometry of the interlocking holes in the nail, drill bit required and number of patients included in
groups I and II of the study. PFNÕ: proximal femoral nail; UFNÕ: unreamed femoral nail; UTNÕ: unreamed tibia nail.
572 N. Suhm et al.
eight precision-related minor intraoperative pro- be detected. The surgeon’s experience with fluoro-
blems out of 39 interlocking screws. There were five scopic guidance in group I was found on a significantly
major technical-related intraoperative problems higher level than experience with fluoroscopy-based
out of 23 applications; in two patients a problem surgical navigation for distal locking in group II
concerning the tracking unit of the navigation sys- (P ¼ 0:044).
tem occurred and fluoroscopy-based surgical navi-
gation for distal locking had to be abandoned.
Fluoroscopic guidance then was used and the track- Discussion
ing unit was checked after the procedure. In two
patients fluoroscopy-based surgical navigation had Radiation exposure to the surgical team as well as to
to be abandoned as stored images could not be the patient during orthopaedic procedures using
properly distorted from the navigation system. In fluoroscopic guidance is a universal concern.12
one patient the nail’s reference base was not visible Several studies have evaluated the radiation expo-
due to unusual fixation. sure to orthopaedic surgeons,1 to the theatre
There were two minor technical-related intrao- staff,5,6,8,10 and to the patients.4 While all studies
perative problems out of 23 applications; in two showed that the radiation exposure was well within
patients the authors observed software problems. A guidelines, the exposure times reported by these
system reset bypassed the problem. authors were considered to be too optimistic.3
The surgeon’s experience with fluoroscopy-based Ionizing radiation has no safe threshold of exposure
surgical navigation for interlocking hole placement below which it ceases to have adverse effects.
is shown in Table 2. Kwong et al.4 stressed the importance of minimising
There was only one stab wound with evidence of the radiation dose related to fluoroscopic imaging.
infection. This patient had reosteosynthesis of his Surgical navigation offers the potential to reduce
femoral fracture 4 weeks postoperatively because fluoroscopic times and thus radiation doses.
of deep wound infection. The wound infection was In this study, the authors compared two guidance
located proximal at the entry point of the intrame- methods for distal locking of intramedullary
dullary nail and was diagnosed in the early rehabi- implants. The distal locking procedure was selected
litation hospital phase. as clinical model to evaluate fluoroscopy-based
surgical navigation for several reasons. The distal
Comparison of groups I and II locking procedure is common. Furthermore, Levin
et al.5 reports that 60—307 s of fluoroscopic time
An unpaired Chi-square test showed no preoperative are required to insert the distal interlocking screws
significant differences between groups I and II using a free-hand method with fluoroscopic gui-
with respect to gender (P ¼ 0:39), fractured side dance. This means 40—50% of the total fluoroscopic
(P ¼ 0:062) or type of anaesthesia (P ¼ 0:57). Fish- time for nailing are required for the distal locking
er’s exact test revealed no preoperative significant procedure.5,10
differences between groups I and II with respect to In the present study, fluoroscopic times for distal
the type of intramedullary nails used (P ¼ 0:96) and locking with fluoroscopic guidance in group I were
to the diameter of interlocking screws inserted found in good accordance with those reported in the
(P ¼ 0:42). An unpaired t-test revealed no preo- literature. With fluoroscopy-based surgical naviga-
perative significant differences between groups I tion the authors succeeded in reducing the fluoro-
and II with respect to the patient’s age (P ¼ 0:45). scopic time required to insert one interlocking
Fluoroscopic time needed per interlocking screw screw to 7.3 s.
in group II was significantly less than fluoroscopic Slomczykowski et al.9 reported a fluoroscopy
time needed in group I (P < 0:0001). Drilling an time per pair of distal locking screws of 1.67 s when
interlocking hole using guidance by fluoroscopy- the distal locking procedure was performed with an
based surgical navigation required significantly equivalent surgical navigation system. Aligning the
longer procedure times (P ¼ 0:031) than passing c-arm fluoroscope, image acquisition and final doc-
the drill bit through an interlocking hole with fluoro- umentation take more than 2 s fluoroscopy time in
scopic guidance. our experience.
No significant differences in the rate of major Even though the reduction of fluoroscopy time
(P ¼ 0:93) and in the rate of minor (P ¼ 0:99) achievable is impressive, the impact of this reduc-
precision-related intraoperative problems could tion in fluoroscopic time on the patient’s radiation
be detected. No significant differences in the dose might be limited. The locking procedures are
rate of major (P ¼ 0:10) or in the rate of minor performed in a great distance from the radiosensi-
(P ¼ 0:61) intraoperative technical problems could tive organs, so there is only a small contribution to
Fluoroscopic guidance versus surgical navigation 573
the patient’s effective radiation dose.4 In contrast, between drill bit and nail was subjective. This has
radiation exposure to the surgeon is potentially to be mentioned when interpreting the numbers
reduced with fluoroscopy-based surgical navigation, on interlocking holes passed through with minor
as the new technique allows the surgeon to stay far precision-related problems or with no precision-
from the fluoroscope during image acquisition. related problems.
Using a free-hand aiming technique with fluoro- These limitations notwithstanding, this prospec-
scopic guidance for distal locking, the surgeon is tive study on distal locking of intramedullary
forced to work within or close to the primary X-ray implants showed that fluoroscopic times to achieve
beam.5,7 equivalent precision are reduced with fluoroscopy-
Fluoroscopy-based surgical navigation increased based surgical navigation compared with fluoro-
the need for resources. An additional 40 min were scopic guidance. The authors do not advocate
required prior to skin incision and after skin closure using fluoroscopy-based surgical navigation in
as set up time for the navigation system. As these every procedure possible. Fluoroscopy-based sur-
tasks were fulfilled by a trained technician in par- gical navigation is recommended for intraopera-
allel with other duties the patient turnover was tive guidance in situations where reduction of
not delayed. While the new method prolonged pro- exposure to radiation is considered advantageous
cedure times initially, this drawback disappeared over the demand of resources and overall procedure
towards the end of our study, as the surgeon time.
became more familiar with the new technical
equipment. Procedure times of computer assisted
interventions might also decline when dedicated Acknowledgements
operation suites become available that are adapted
to the specific needs of image guided surgery. This study was supported by the Swiss National
Technical reliability was better with the standard Science Foundation (SNSF) Grant No. 3200-059
guidance method compared with fluoroscopy- 222.99. Other than clinical research money, no
based surgical navigation. The technical problems additional benefits have been received. The authors
observed were partly related to a lack of experience thank Lisa Churchill and Bernhard Suhm for their
with the complex equipment. Proper instruction assistance in the preparation of this manuscript and
and testing in the laboratory is strongly advised the Anaesthetic Department and the operating
prior to application of surgical navigation tech- room staff for continuous cooperation during this
nology in clinical practice. study.
The authors encountered one case of postopera-
tive wound infection among 42 patients, which is
clearly within the limits. The patient was assigned References
to group II. This result by no means can be inter-
preted as increased infection rate in interven- 1. Barry TP. Radiation exposure to an orthopedic surgeon. Clin
tions performed with fluoroscopy-based surgical Orthop 1984;182:160—4.
2. Hofstetter R, Slomczykowski M, Sati M, Nolte LP. Fluoro-
navigation.
scopy as an imaging means for computer-assisted surgical
The failure rate in actually doing the interlocking navigation. Comput Aided Surg 1999;4:65—76.
in the clinical setting was found higher than 3. Hudson I. Locking nailing: an aid to distal targeting. Injury
expected, as the potential accuracy of the system 1989;20:129—30.
used by the authors is much greater than this. 4. Kwong LM, Johanson PH, Zinar DM, Lenihan MR, Herman
MW. Shielding of the patient’s gonads during intramedullary
However, the navigation system used allowed track-
interlocking femoral nailing. J Bone Joint Surg (Am)
ing the compact air drive instead of a drill guide. 1990;72:1523—6.
Bending the drill bit while drilling the interlocking 5. Levin PE, Schoen RW, Browner BD. Radiation exposure to
hole might therefore be responsible for the the surgeon during closed interlocking intramedullary
inaccuracies observed. This problem has been nailing. J Bone Joint Surg (Am) 1987;69:761—6.
6. Mehlman CT, DiPasquale TG. Radiation exposure to the
addressed by the manufacturer and drill guides that
orthopaedic surgical team during fluoroscopy: ‘‘how far
can be tracked are now available. away is far enough?’’. J Orthop Trauma 1997;11:392—8.
Limitations of our study include the lack of ran- 7. Miller ME, Davis ML, MacClean CR, Davis JG, Smith BL,
domization in the assignment of patients into Humphries JR. Radiation exposure and associated risks to
groups I and II but the anaesthetic risk score clearly operating-room personnel during use of fluoroscopic gui-
dance for selected orthopaedic surgical procedures. J Bone
does not introduce a bias. There was a potential for
Joint Surg (Am) 1983;65:1—4.
user bias because the surgeon could not be blinded 8. Sanders R, Koval KJ, DiPasquale T, Schmelling G, Stenzler S,
with respect to the method used for guidance. Ross E. Exposure of the orthopaedic surgeon to radiation. J
To decide between slight contact or no contact Bone Joint Surg (Am) 1993;75:326—30.
574 N. Suhm et al.
9. Slomczykowski MA, Hofstetter R, Sati M, Krettek C, 11. Suhm N, Jacob AL, Nolte LP, Regazzoni P, Messmer P.
Nolte LP. Novel computer-assisted fluoroscopy system Surgical navigation based on fluoroscopy-clinical application
for intraoperative guidance: feasibility study for distal for computer-assisted distal locking of intramedullary
locking of femoral nails. J Orthop Trauma 2001;15: implants. Comput Aided Surg 2000;5:391—400.
122—31. 12. Tremains MR, Georgiadis GM, Dennis MJ. Radiation exposure
10. Sugarman ID, Adam I, Bunker TD. Radiation dosage during with use of the inverted-c-arm technique in upper-extre-
AO locking femoral nailing. Injury 1988;19:336—68. mity surgery. J Bone Joint Surg (Am) 2001;83:674—8.