Академический Документы
Профессиональный Документы
Культура Документы
Anees Ahmed
70043971
Babar Kahn
70037506
Hypoxia is a condition which occurs when the human body don’t have enough
oxygen. This condition can cause the body to loss the conscious.
The crew system will last a 5-man crew approximately for 4 hours using diluted
oxygen.
Function
Oxygen regulator
On the front of the oxygen regulator are controls for oxygen dilution and
pressure
The regulated oxygen pressure gauge shows pressure entering the
oxygen regulator. Oxygen will not flow from the oxygen regulator to the
masks unless the on-off selector is in the ON position shown on the
figure of the oxygen regulator panel.
With the 100% oxygen normal oxygen selector in the 100% position,
undiluted oxygen will be supplied to the mask regardless of flight station
altitude. With this selector in the normal oxygen position, oxygen to the
mask will be diluted with flight station ambient air in proportion to flight
station air pressure. Dilution decreases with increasing altitude. At a
flight station altitude of 28,000 feet or higher, oxygen to the masks will
not be diluted at all. Normal flight station altitude (pressurized) is 8,000
feet.
With the emergency-normal-test mask selector in the normal position
flow to the mask is on demand (inhalation).
In the emergency position, flow to the mask is continuous, and at
positive pressure. The emergency position mechanically overrides the
normal position.
In the test mask position, flow to the mask is continuous, and at a
pressure higher than in emergency position. The selector is spring-
Loaded out of the test mask position to the normal position.
On the back of the oxygen regulator are oxygen inlet and outlet ports, and an
electrical connector for the 5 VAC panel light.
Oxygen masks, with associated supply tubes, are nested in a clear plastic rack and
secured in the module by a hinged door. The door is magnetically latched, and for
mask deployment, is electro magnetically unlatched by a solenoid. Oxygen
generators and the masks associated with them are divided into fifteen groups;
each group being identified numerically and composed of sectionally situated
modules.
Oxygen generators are available in three sizes to supply oxygen to one, two, or
three masks.
The case exterior contains black and white comparison stripes. The white stripe is
heat-sensitive and will darken when the generator is used. Oxygen from all masks
constitutes about 1 percent of the cabin atmosphere, so there is no fire hazard
due to oxygen from unused masks. Oxygen flow is continuous until the generator
is expended.
BOEING 747-400 OXYGEN SYSTEMS
The flight crew and passenger oxygen systems use oxygen cylinders to supply
oxygen if needed. Several passenger oxygen systems have used chemical oxygen
generators to supply the passengers in an emergency.
The 747 use nine high-pressure oxygen cylinders to store the passenger
emergency oxygen.
OFF Position
In the OFF (reset) condition, the flow control units prevent oxygen flow into
the low-pressure distribution manifold by closing the flow control units.
Safety Any small amount of oxygen leakage through the flow control units is
discharged by the automatic vent valve to prevent unscheduled mask
deployment.
On Position
When the flow control units are actuated, the resulting pressure surge
unlatches the oxygen module doors, causing the masks to deploy. The
surge also closes the automatic vent valve.
Continuous flow systems use compressed oxygen cylinders, a valve is kept closed
by an actuating pin. The valve is opened when the mask is pulled toward the
passenger's face.
• A lanyard attached to the mask hose pulls the actuating pin out of the
valve, allowing the valve to open and oxygen to flow to the mask. When
oxygen is no longer required, the flow control units are reset by the flight
crew, stopping the flow of oxygen to the distribution manifold.
• Test fittings allow connection of an external pressure source and gauge for
system testing.
• Switch at Norm
When the switch in NORM, the; system will activate one of the flow control units
automatically at 14,000 feet cabin altitude.
When the switch is at ON position it activates the two other flow control units.
When oxygen is no longer required, the system is turned off by placing the switch
move to the RESET position.