O2/M-100
Oxygen Gas Sensor in Mini Housing
MEASUREMENT
Operation Principle 3-Electrode Electrochemical
Miniature-Size Outline Dimensions
Nominal Range 0 – 30 %
Maximum Overload 100 %
BOTTOM VIEW SIDE VIEW
Inboard Filter ‒
Output Signal 90 ± 24 µA/%
Sensing 20 mm
Resolution
< 0.05 %
(Electronics dependent) Reference Counter
16.4 mm
15.7 mm
T90 Response Time < 8 sec 90°
Typical Baseline Range (in
0.1 %-equivalent
pure nitrogen) 13.5 mm 5 mm
Maximum Zero Shift 18 mm PCD Pin Projection 1.5 mm Pins
N.D.
(+20°C to +40°C) on 13.5 mm PCD
Repeatability < 2 % of signal
1
Output Linearity Linear
Gain ‒ ± 0.10 mm
ELECTRICAL MECHANICAL
Rec. Load Resistor 10 Ohm Weight 5.5 g
Bias (V_Sens-V_Ref) -600 mV Position Sensitivity None
Conformity to RoHS directive RoHS Compliance
APPLICATIONS
ENVIRONMENTAL Oxygen Monitoring in Diffusion Mode
Relative Humidity Range 15 % to 90 % R.H. non- Continuous Air Quality Monitoring
condensing Safety and Environmental Control
Temperature Range -40 °C to 50 °C
Pressure Range Atmospheric ± 20% CROSS-SENSITIVITY DATA
Pressure Coefficient N.D.
The table below does not claim to be complete. Interfering
Humidity Effect none gases should not be used for calibration.
Interfering Gas Conc. Reading
LIFETIME
2
ppm %
Expected Operation Life 3 years in air
H₂S 500 -0.02
Expected Long Term Output
< ±4% signal / 3 years SO₂ 200 -0.01
Drift in air
Filter Life ‒
Storage Life 6 months in container
Rec. Storage Temperature 5 °C – 20 °C
Warranty Period 12 months from date of
dispatch
O2/M-100
Oxygen Gas Sensor in Mini Housing
TEMPERATURE DEPENDENCE
The output of an electrochemical sensor varies with temperature. The graphs below show the variation in output with
temperature for this type of sensor. The results are shown in the graphs as a mean for a batch of sensors. The sensitivity
dependence is expressed as a percentage of the signal at 20 °C. The shift in baseline is shown in ppm referenced to 20
°C and a relative humidity of 50%.
Please note:
It is highly recommended to acquire the temperature dependence curves with the whole instrument. The sampling
system, the humidity, the electronics, the interaction between the electronics and the sensor, all have a significant impact
on the temperature dependence of the final measurement reading.
105%
Output (referenced to 20 °C)
100%
Mean
95%
Max
Min
90%
85%
80%
-40 -30 -20 -10 0 10 20 30 40 50
Temperature [°C]
Figure 1: Sensitivity dependence expressed as a percentage of the signal at 20 °C. The result is shown along with
confidence intervals corresponding to ±3 times the standard deviation.
O2/M-100
Oxygen Gas Sensor in Mini Housing
LINEARITY AND RESOLUTION
28
24
Signal (% O2 indication)
20
16
12
0
0 5 10 15 20 25 30
Oxygen Concentration [%]
Figure 2: Linearity of O₂-Sensor
Resolution
150
140
130
120
110
100
90
Signal [uA]
80
70
60
50
40
30
20
10
0
0.0 0.2 0.4 0.6 0.8 1.0 1.2 1.4 1.6 1.8 2.0
Oxygen [%]