CALIBRATION CERTIFICATE

Company N/A Contact(s) N/A

CERTIFICATE NUMBER: CH09D0N/A JOB NUMBER: N/A

Date 23rd June 2009 Date due June 2010

Instrument Type: Microlite 16K Serial No: 1808569

Calibration Environment: 21°C +/- 3°C Number of Points: 5

Calibrator Name Jeremy Rouse Signature At foot of page

COMMENTS

The instruments were tested inside a Hygrogen chamber.


Temperature reference was ASL F100

Results were within acceptable limits of +/- 0.5 °C and +/- 5%

Report Explanation:
Instrument Reference: The instrument being calibrated
Target value: The value the temperature or humidity source was set to.
Reference Measurement: The value the calibrated thermometer/Hygrometer measured
Instrument Measurement: The value the instrument being calibrated measured
Difference between reference and instrument: The deviation between true and actual
Off set: If the difference between true and actual is outside of acceptable tolerances, an offset is added.

Reference units used:


Hygroclip ASL ASL
Instrument Details F200 F100
  
CALIBRATION REFERENCE INSTRUMENT
2818041 008408/20 K1642A
S/N:
±0.3C ± 0.016C ± 0.02°C
ACCURACY: Temperature ±1.5%rH ---- ----
Humidity ---- ---- ----
---- ---- ----
CERTIFICATE OF REFERENCE
14647 N1173104T N1181512T
INSTRUMENT:

RECALIBRATION DATE OF REFERENCE


24.4.09 23.10.08 23.09.09
INSTRUMENT:

Signature: ………………………...................................…
Jeremy Rouse

LoggerShop www.loggershop.co.uk
59 Southwick Road, Bournemouth, Dorset, BH6 5PR: email: Sales@loggershop.co.uk
Batch 1

Difference
Instrument Reference Instrument between
Target value Off set
Reference Measurement Measurement reference and
instrument
°C °C °C °C °C

1808569 5 4.992 4.74 -0.252


N/A
1808569 10 10.013 10.07 +0.057
N/A
1808569 23 23.087 23.12 +0.033
N/A
1808569 25 25.122 25.15 +0.028
N/A
1808569 30 30.205 30.03 +0.175
N/A

LoggerShop www.loggershop.co.uk
59 Southwick Road, Bournemouth, Dorset, BH6 5PR: email: Sales@loggershop.co.uk