Operation Manual
Content
V. Operation methods……………………………………………………………...5
VI. Attention………………………………………………………………………..7
1
Chongqing Gold GD-1884 Petroleum Products Density Tester
2
Chongqing Gold GD-1884 Petroleum Products Density Tester
(9)
(1) Electric controlling case (2) Control panel (3) Bath cover
(4) Spare hole and its rubber stopper (5) Motor shield (6) Stirring motor
(7) Temperature sensor (8) Thermometer (9) Sample cylinder clamps
(10) Main power switch (11) Pedestal (12) Bath
(13) Fixing ring of water bath (14) Sample cylinder (15) Foot
Fig.1
1. Electric heater
It is made of sealed stainless steel tube. It is used to heat the water bath.
3
Chongqing Gold GD-1884 Petroleum Products Density Tester
4
Chongqing Gold GD-1884 Petroleum Products Density Tester
Fig.2
V. Operation methods
(I) Preparation before test
1. Please carefully read the Operation Manual before use.
2. Please carefully read the National Standard of People’s Republic of China GB/T1884-2000
“Test Methods for Density of Petroleum Oils and Liquid Petroleum Products (Density Meter
Method)” to acquaint and familiar with preparation, test procedures, and test requirements.
3. Please get ready for all test apparatus and materials as per GB/T1884-2000.
4. Please check the working state of the instrument to make sure it is in accord with the
5
Chongqing Gold GD-1884 Petroleum Products Density Tester
6
Chongqing Gold GD-1884 Petroleum Products Density Tester
Input the modified value: Press the Set button to the menu B. Then keep pressing the Set
button until SC is shown at the SV position. Press the move button, plus button or minus button to
input the modified value. Repress the Set button to escape from the menu B after modification.
8. When the bath temperature keeps stable at the required testing temperature, please rotate the
sample cylinder clamps for several degrees and take it off from the bath cover. Then insert the
sample cylinder (containing sample) into the clamps and put the sample cylinder into the bath.
Rotate the clamps to fix it on the bath cover.
9. Insert the thermometer into the sample and stir the sample carefully. When the sample
temperature keeps stable, please vertically place a proper density meter into the sample and keep it
stable. Read the reading of density meter and record the sample temperature. Then convert it to the
density at 20 ℃ as per the methods stipulated in the National Standard of People’s Republic of
China GB/T1885.
VI. Attention
1. You can use water bath if the controlling temperature is lower than 80 ℃. Then it is not
necessary to change the stirring vane. If you use the oil bath when the controlling temperature is
higher than 80 ℃ and the temperature controlling accuracy can not meet the requirement, you can
change the stirring vane. Disassemble the protection shield of heater, and enlarge the angle of
stirring vane to increase the stirring force. Assemble the protection shield again.
2. Stir the bath during heating to ensure the bath temperature is even and uniform. If the
stirring motor does not rotate, please check the capacitance at first. Then check the electric motor.
3. If there is nothing shown at the PV position after the instrument gets power, please pull off
the temperature sensor plug on the back panel and check it. The resistance between the pin opposite
to the gap on the plug and each pin at its two sides is less than 110 Ω at the room temperature and
the resistance will increase about 0.4 Ω for each temperature increase of 1 ℃. The resistance
between pins not opposite to the gap on the plug is zero. You can check whether the temperature
sensor is damaged according to the method described above.
4. If the temperature increases slowly after you turn on the two Temp.ctrl switches, please
7
Chongqing Gold GD-1884 Petroleum Products Density Tester
check the aviation plug of eight cores. The resistance between the No.7 pin and the No.5 pin is
about 50Ω, and the resistance between the No.8 pin and the No.5 pin is about 69 Ω. If the first two
is open circuit, it indicates that the heater of 1000 W is damaged. If the later two is open circuit, it
indicates the heater of 700 W is damaged. Then you can check whether the heaters are damaged
according to the method described above.
5. If the heaters are fine but the temperature increases slowly, you can check the silicon
controlled rectifier on the bottom board of the controlling case. Please change the silicon controlled
rectifier if necessary.
Note: When the ambient temperature is lower than 0 ℃, please exhaust water in the water bath
to avoid any damage to the instrument caused by freezing.
Warning: If there is any trouble appears, please cut off the power supply immediately. Then
ask a professional to check and repair it. You can use it again only after the problem is solved to
avoid any accidents.
8
Chongqing Gold GD-1884 Petroleum Products Density Tester
2. Technical documents
(1) Operation Manual 1 piece
(2) Quality Certificate 1 piece
(3) Repair Guarantee 1 piece
Checker: Packager:
Date: Date: