This monitor is suitable for the adult, the pediatric and the neonate, may monitor
the life parameters, such as the electrocardiograph (ECG), the noninvasive blood
pressure (NIBP), the oxygen saturation (SpO₂), the respiration rate (Resp), the body
temperature (Temp) and so on,, can in the same screen demonstrate maximum 8
The purpose and function of the patient monitor is to monitor: ECG; heartrate;
functional arterial oxygen saturation; respiration rate and temperature for adult and
pediatric patients in all hospital areas and hospital-type facilities. It can be used
ambulances.
2) The oxygen saturation (SpO₂), including: the oxygen saturation , pulse rate,
pulse wave;
3) The noninvasive blood pressure (NIBP), including: Systolic pressure,
diastolic pressure, mean pressure;
4) The body temperature (Temp): 2 channels body temperature data;
5) The respiration (Resp): Breathing rate, breath waveform.
Chapter 2 Interface
2.1 System Structure Diagram
sensor socket
socket Function
J1 Power
J5 Net
J6 TFT inverter
J8 TFT LCD
J11 Recorder
J17 Keyboard
J20 Speaker
ECG ECG
SPO2 SPO2
TEMP Temperature
socket Function
J1 To main board
J2 Encoder
J4 Alarm led
1) Use a screwdriver to remove the two screws fastening the rear case to the
front case.
2) Open the front case.
3) Disconnect the keyboard cable, power switch cable, power led cable, TFT
cable and inverter cable from FY0501 board and keyboard, remove front
case.
1) Disconnect the keyboard cable, power switch cable and power led cable from
keyboard. Use a screwdriver remove keyboard.
2) Remove the TFT plate.
3) Remove the inverter board.
4) Remove the alarm led board.
AC Power Cable
AC Ground Cable
DC Output Cable
Fan Cable
Power Switch
Battery Cable
LED Cable(LED1/LED2)
TFP Cable
Keyboard Cable
Net Cable
Encoder Cable
ECG Cable
SpO2 Cable
Temperature Cable
Chapter 4 Troubleshooting
4.1 Troubleshooting
No Power up With external power supply connected, push power switch, check
the green POWER LED on the front panel, if it is not lit, replace
power cable and power board.
LCD screen is totally Ensure that the backlight cable is connected to the FY0501
black or white. board. Ensure that the LCD connector is properly connected to
the FY0501 board and LCD. If problem persists, replace
backlight board or FY0501 board.
LCD screen is Ensure that the LCD connector is properly connected to the
illuminated, but no data FY0501 board and LCD. If problem persists, replace FY0501
is visible. board.
“COMM. STOP ERR” Ensure that the FY0501-FY0801 cable is properly connected to
the FY0501 board and FY00801 board. If problem persists,
replace FY0801 board.
Key or encoder do not Replace keyboard cable or keyboard. If problem persists, replace
work FY0501 board.
Recorder do not work Ensure that recorder’s power led is on. Ensure the recorder cable
is properly connected to FY0501 board. If problem persists,
replace recorder or FY0501 board.
ECG no signal Check ECG cable,If problem persists, replace FY0801 board.
NIBP can’t inflate Ensure there is pump inflation sound when NIBP start, if no
pressure up sound, replace pump cable or pump. If pump is ok, check if there
is leakage in valve, replace valve cable or valve when there is
leakage, If problem persists, replace recorder or FY0801 board.
SPO2 no signal Ensure SPO2 cable is ok, If problem persists after changing
SpO2 cable, replace FY0801 board.
Temperature no signal Check temperature cable and temperature cable type setting,If
problem persists, replace recorder or FY0801 board.
IBP no signal Ensure IBP cable is ok,If problem persists, replace FY0801
board.
CO2 no wavefrom Ensure CO2 cable is ok,If problem persists, replace CO2
module.
“CO2 COMM STOP” Ensure that the FY0501-CO2 cable is properly connected to the
FY0501 board and CO2 module. If problem persists, replace
CO2 module.
information
** ECG LEAD OFF RL or more than 2 ECG leads check the ECG lead
falls off connection
** ECG LEAD V OFF V lead fall off check the ECG lead
connection
***MODULE COMM The module and the main engine Restart the machine, if
STOP communication have the problem error still existed,
contact the factory
service
** MODULE COMM The module and the main engine Restart the machine, if
ERR communication have the problem error still existed,
contact the factory
service
** PARA ALARM The parameter of the alarm limit contact the factory
LMT ERR is modified by the accident service
** RANGE EXEED The parameter observed value has contact the factory
exceed the measurement scope service
which the system can carry on
** SpO ₂ SENSOR SpO₂ sensor does not connected Check SpO ₂ sensor
OFF connection
** SpO ₂ FINGER The finger fall off from SpO ₂ Check SpO ₂ sensor
OFF sensor connect with the finger
** SYSTEM TIME The system clock has not set Change the system time
LOST as the current time, if
error still existed, related
the factory to carry on
the service
** NIBP OVER The pressure value exceeds the check the pipe
PRESSURE measurement scope connection situation or
replace cuff, if the error
still existed, please
contact the factory
service
* NIBP PNEUMATIC The cuff has not connected good check the pipe
FAIL or the air course leaks air connection situation or
replace cuff, if the error
still existed, please
** NIBP CUFF TYPE Patient type for adult when has Check the patient type or
WRONG used the neonate cuff replace cuff, if the error
still existed, please
contact the factory
service
Chapter 5 Test
Resistor box
Ruler
Network cable
Connect the simulator, power cable and sensor cable to the patient monitor; turn on the
power. Patient monitor shall display startup picture and then enters monitor screen.
Display Test
After patient monitor enter monitor screen, the characters and waves should display clearly
without dithers and blurs.
Key Test
Press every key on the keypad, the patient monitor shall execute the corresponding
function as instructed in the operation manual. The encoder shall execute the corresponding as
well.
Sound Test
There should be key sound when rotate encoder; there should be alarm sound when
triggers alarm.
ECG Calibrate
Set ECG channel 1 and channel 2’s gain to: ×1 and open ECG calibrate switch, ECG
wave should be a 1mV, 1Hz square wave, check the wave amplitude should be 10mm±0.5mm.
ECG Gain
Set ECG channel 1’s gain to: ×1, ECG channel 1 lead to II and set ECG simulator to input
a 1mV 1Hz square wave, ECG1 wave should be a 1mV, 1Hz square wave, check the wave
amplitude should be 10mm±0.5mm.
Set ECG channel 2’s gain to: ×1, ECG channel 2 lead to II and set ECG simulator to input
a 1mV 1Hz square wave, ECG2 wave should be a 1mV, 1Hz square wave, check the wave
amplitude should be 10mm±0.5mm.
ECG Lead
Set ECG simulator to input normal ECG wave, Select all ECG lead and gain, ECG wave
should be correct according to simulator.
Lead Off
Disconnect ECG lead from ECG simulator, there should be correct ECG lead off alarm.
Heart Rate
Set ECG simulator heart rate to be 30, 80, 300, the patient monitor’s heart rate should be
30, 80 and 300 respectively and error is below 1%.
Pace
Set “PACE” switch on and set simulator to input PACE signal, there should be PACE mark
on ECG wave.
Set the base impedance to 1K and resistance respectively to 0.5Ω and 3Ω, RESP Rate
respectively to 30 and 120, the system shall measure the RESP Rate to be 29~31 and 118~
122.
1) Set NIBP simulator to adult mode and pressure to 255/195/215 mmHg, set patient
monitor to adult mode and start NIBP test, when measurement finish, systolic, mean,
diastolic pressure should be 255±8mmHg,215±8mmHg 、195±8mmHg.
2) Set NIBP simulator to neonate mode and pressure to 120/80/90 mmHg, set patient
monitor to pediatric mode and start NIBP test, when measurement finish, systolic,
mean, diastolic pressure should be 120±8mmHg,90±8mmHg 、80±8mmHg.
3) Set NIBP simulator to neonate mode and pressure to 120/80/90 mmHg, set patient
monitor to neonate mode and start NIBP test, when measurement finish, systolic,
mean, diastolic pressure should be 40±8mmHg,20±8mmHg 、10±8mmHg.
NIBP calibrations
Using a pressure gauge (or mercury sphygmomanometer) to carry on the calibration,
choose “NIBP Calibration” in the “Service” menu to start calibration.
Connect the pressure gauge, the cuff through a 3-way tube to the blood pressure trachea
jack on the monitor, setting the monitor as “calibration” mode, then charge the cuff using a air
pump, first make the pressure to 250 mmHg, then slowly deflates, when the monitor display 200,
150 and 50 mmHg, the difference between the standard pressure gauge value and the monitor
pressure value should in 3 mmHg. If the value exceeds 3 mmHg, please contact our company’s
attendant.
Put finger to SpO2 sensor and connect SpO2 sensor to the patient monitor, SpO2 value
should be above 95%.
Press “RECORD” key, the recorder should output ECG waveform accordant to display.
Take out recorder paper, there should be “RECORDER OUT OF PAPER” alarm on patient
monitor.
Connect patient monitor and central monitor station with network cable and start up patient
monitor and central monitor station, the net icon on patient monitor should be “ ”, there
should be waveform and parameter displayed on the central monitor station.
2) Check if all the outer cables, inserted modules and accessories are in good
condition;
3) Check if all the monitoring functions of the monitor can work normally so as to make
sure that the monitor is in good condition.
If you find any damage on the monitor, stop using the monitor on patient, and contact the
biomedical engineer of the hospital or Customer Service Department immediately.
Routine check
The overall check of the monitor including the functional safety check, must be performed
by qualified personnel once every 6 to 12 month or each time after fix up. All checks that need to
open the monitor enclosure must be performed by qualified service personnel.
It is recommended that you should clean the outside surface of the monitor enclosure and
the display screen regularly. Only use non-caustic detergents such as soap and water to clean
the monitor enclosure.
2) Most cleaning agents must be diluted before use. Dilute the cleaning agent as per
the manufacturer's direction.
4) Do not let the cleaning agent enter the monitor. Do not immerse any part of the
system into liquid.
4) Hydrogen Peroxide 3%
5) Alcohol
6) Isopropanol
6.4 Sterilization