Академический Документы
Профессиональный Документы
Культура Документы
ULTRACENTRIFUGE
SOP # 01-REP
This SOP covers the proper and safe use of the Sorvall WX+
Scope: Ultracentrifuge. It is applicable to all staff members/students
working with this machine.
1
- SOP Outline
Procedures
Summary of Significant Health and Physical
3
Hazards
Personal Protective Equipment Required 3
Risk Assessment 3
Emergency Contacts 3
Preparatory Steps 3
Warnings 3
Laboratory Procedures
(1) Preparation 4
(2) Centrifuge Operation 5
(3) Emergency Procedures 9
(4) Maintenance 9
(5) Service 9
(6) Cleaning Procedures 10
(7) Training 10
References 11
Procedures Quality Assurance 11
Appendix
I. Equipment Details 12
II. List of Compatible Rotors 13
III. Ultracentrifuge and Rotor Log Sheet 14
IV. Acceleration and Deceleration Modes 15
V. Operation Flowchart 16
VI. Troubleshooting Table 17
VII. Inspection, Maintenance and Service
20
Record
VIII. Documentation of Training 21
2
- Summary of Significant Health and Physical Hazards
- Risk Assessment
-
The overall health and safety risk of this procedure is described
in Ref: PL-ES-006-RA-xxxx - Using a Centrifuge
- Emergency Contacts
-
List Of QU Emergency Contacts Here
- Preparatory Steps
- Warnings
(1) Make sure that the rotor is appropriate for the instrument and
use the lid that matches the rotor. Rotor can fail due to
excessive speed, causing catastrophic failure of centrifuge.
3
Know the speed limits of the rotor and do not exceed them
(See Appendix II).
(2) Rotor imbalance can cause catastrophic failure of centrifuge
explosion), injuring user.
(3) Swing bucket rotor are designed to be used with all the buckets
in place, even if some of them are empty. Do not use this type
of rotors without all the buckets in place.
(4) Rotor can fail due to extensive wear and tear, causing
catastrophic failure of centrifuge or even injuring the user.
Inspect all the rotors regularly and replace if damaged.
(5) Centrifuge tubes can break inside the rotor due to excessive
wear, tear or inappropriate use causing catastrophic failure.
Ensure that you use the correct tubes for your samples, and do
not use cracked or broken tubes. Phenol and chloroform will
rapidly destroy polycarbonate tubes and must be used in
polypropylene tubes.
(6) Plastic centrifuge tubes not certified as reusable from the
manufacturer, should be discarded after one cycle of
ultracentrifugation. The failure rate for used tubes is a hazard
that justifies using new tubes for each high G run.
(7) Do not lean or place items on the instrument while it is
operating.
(8) Do not leave the centrifuge until full operating speed is
attained, and the instrument appears to be running normally
without vibration.
(9) If vibration occurs, stop the run immediately; wait until the
rotor stops, and check the load balances.
(10) In event of a power failure, do not try to open the lid to retrieve
samples for at least one hour.
- Laboratory Procedures
(1) Preparation
- Make sure that the sample tubes are appropriate for your
samples and the speed required. All tubes used should be of the
same type. Check all tubes for signs of stress and/or crack.
Discard any tubes if necessary.
4
(2) Centrifuge Operation
5
Figure 3:
Symmetrical
Placement of Tubes
on a 12-Position
Rotor
6
remaining run time. For example, if the centrifuge has been
running for 5 hours and you want operation to stop 1 hour and
30 minutes later, touch the TIME field and enter [6][ : / . ] [3]
[0].
Acceptable Values
Speed: Set a value between 1,000 rpm to maximum speed in
100 rpm increment
Time: Set a value up to 999 hours and 59 minutes in 1-minute
increments.
Temp: Temperatures in the 0 to 40°C range can be set in 0.1 ℃
increments.
- When the Rotor management screen does not display the desired
rotor, press the [Rotor Catalog] button. In the ROTOR
CATALOG screen, select the desired rotor type. In the screen
displaying rotor types, press the field of the desired rotor. Check
that the pressed field turns blue and press the [Enter] button.
- To confirm rotor details, press the rotor field to turn it blue and
press the [Details] button. A window that allows you to check
rotor information appears. If this is the selected rotor, press the
[Enter] button. To select a different rotor, press the [X] button
to return to the Select rotor screen.
7
Figure 6: Rotor Selection
Screen
8
- Press the [VACUUM] button. The vacuum pump stops, the air
leak valve starts operating to return the rotor chamber to
normal atmospheric pressure. The door lock is released and the
door can be opened. When the rotor chamber returns to normal
atmospheric pressure, the "Now open the door" message
appears above the [VACUUM] button.
9
lab.
10
Users of this device must have attended an instrument-specific
training session with the equipment custodian or the facility
approved lab manager prior to any use, covering general use
and safe practices of the instrument in question.
References
Quality Assurance Improper centrifugation speed, temperature and time can have
significant effects on sample quality. To ensure the validity
and reproducibility of the centrifugation protocols used, the
machine’s speed of rotation, chronometer and thermometer
must be verified once a year, and if necessary, calibrated by
authorized technicians. All calibration certificates should be
gathered and kept in the lab.
11
APPENDIX
I. Equipment Details:
12
II. Compatible Rotors (Non-exhaustive list)
Rotor
Model Maximum RPM Supported Tubes
Type
13
III. Sorvall WX 80+ Ultracentrifuge and Rotor Log Sheet
From: To:
14
IV. Acceleration and Deceleration Modes
The acceleration and deceleration rates can be adjusted between 0 to 500 rpm. By setting an
ideal acceleration and deceleration time samples will not be disturbed, which improves
efficiency and reduces centrifugation time. The table below shows the relationship between
ACCEL and DECEL code numbers and acceleration and deceleration time.
*1 Minimum time: The time it takes the drive motor to accelerate and decelerate the rotor using
maximum torque. This time will vary with rotor type and the mechanical
resistance of the drive motor.
*2 DECEL code "F": A deceleration method that does not involve brake control using the drive
motor. The deceleration time will then vary widely depending on
differences in the small mechanical resistance of the drive motor and the
vacuum level of the rotor chamber and other factors.
15
V. Operation Flowchart
16
VI. Troubleshooting Table
Rotor Problems
17
Tube Problems
The tubes become fragile with Discard expired tubes, use new
age and use. ones.
18
Various Problems
19
VII. Inspection, Maintenance and Service Record
Description of
Date Service Agent Signature
Action
20
VIII. Documentation of Training (Signature of all users is required)
I, the undersigned, verify that I have read and understand the content and
responsibilities of this SOP and have undergone training by an approved
facility personnel.
21