Академический Документы
Профессиональный Документы
Культура Документы
Document History
Table 1 – Document History
Date Revision Description of Revision Author
18/2/2011 1d0 Adaptation from F. Evans Notes C. Sinadinos
9/3/2011 1d1 Minor spelling corrections and screen shot update S. Brede
4/7/2012 1d2 Additional photos and section updates S. King
22/5/2014 1V2 Company Details Updated S. Brede
© Copyright
This document is protected by copyright and the information contained herein is confidential. The
document may not be copied and the information herein may not be disclosed except by written
permission of and in a manner permitted by the proprietors of IPC Global Pty Ltd.
Table of Contents
Contents
1 INTRODUCTION. .......................................................................................................................................................1
1.1 Scope. ...........................................................................................................................................................1
1.2 Document Overview. ...................................................................................................................................1
1.3 Assumptions.................................................................................................................................................1
1.4 Recommendations. ......................................................................................................................................1
1.5 Referenced Documents. ..............................................................................................................................1
1.6 Document Location .....................................................................................................................................1
2 WARNINGS. ...............................................................................................................................................................2
3 THE CYCLIC TRIAXIAL SYSTEM COMPONENTS ............................................................................................3
3.1 0002-2620 UTM-9P Basic System Assembly ...............................................................................................3
3.2 0002-2056 IMACS data acquisition and controls system ..........................................................................3
3.3 0002-2084 Auxiliary air receiver..................................................................................................................3
3.4 Transducers ..................................................................................................................................................3
3.5 Water distribution panel .............................................................................................................................3
3.6 Air water interface........................................................................................................................................3
4 THE BASE SYSTEM .................................................................................................................................................4
4.1 Triaxial Load frame......................................................................................................................................4
4.2 Pneumatic actuator ......................................................................................................................................4
4.3 Integrated Multi-Axis Control System (IMACS) ........................................................................................4
4.4 Auxiliary Air Reservoir ................................................................................................................................4
5 TRIAXIAL CELL & ACCESSORIES .......................................................................................................................5
5.1 Triaxial Cell..................................................................................................................................................5
5.2 Transducers ..................................................................................................................................................5
6 THE PRESSURE SYSTEM ......................................................................................................................................6
6.1 Air water interface(s). ...................................................................................................................................6
6.2 Water distribution panel .............................................................................................................................6
7 CONNECTING ACTUATOR TO AIR SUPPLY.....................................................................................................7
8 CELL, BACK AND PORE PRESSURE TRANSDUCERS..................................................................................8
8.1 Location .......................................................................................................................................................8
8.2 Cell Port Allocations....................................................................................................................................8
9 LOAD AND DISPLACEMENT TRANSDUCERS .................................................................................................9
10 CONNECTING THE IMACS TO THE CELL AND FRAME...............................................................................10
11 TRIAXIAL CELL .......................................................................................................................................................11
12 FILLING THE DE-AIRED SYSTEM WITH WATER............................................................................................13
13 TO FILL THE AIR WATER INTERFACES...........................................................................................................13
14 VOLUME CHANGE / BACK PRESSURE ASSEMBLY ....................................................................................14
14.1 Filling and Bleeding of Volume Change / Back Pressure Assembly .......................................................14
15 SETTING UP FOR A TEST ....................................................................................................................................16
15.1 Platen selection ..........................................................................................................................................16
15.2 Filling the pore water lines and De-airing of Pore pressure transducer ...................................................16
16 SOFTWARE..............................................................................................................................................................18
16.1 Checking system control............................................................................................................................19
List of Figures
List of Tables
Table 1 – Document History...................................................................................................................... i
Table 2 Referenced Documents. ............................................................................................................... 1
Table 3 Acronyms and Abbreviations ..................................................................................................... 40
Table 4 Troubleshooting ......................................................................................................................... 40
1 INTRODUCTION.
1.1 Scope.
This document is the Hardware Reference for the IPC Global Universal Cyclic Tri-axial
System.
1.2 Document
Document Overview.
This document contains the necessary information for the person(s) operating the
Universal Cyclic Tri-axial System to successfully setup the hardware and in conjunction
with the IPC Global Software Manuals (referenced below) operate the machine.
1.3 Assumptions.
The operator should have an understanding of the potential hazards associated with the
IPC Global apparatus as mentioned in 2 below.
It is also assumed the operator has a general working knowledge of materials testing
methodologies as applied to this apparatus (not included in this document).
1.4 Recommendations.
It is recommended the user thoroughly familiarise themselves with the UTS012 test
software prior to using the system.
1.5 Referenced Documents.
Other documents referred to within this document.
Table 2 Referenced Documents.
Document No. Document Title
UTS Soft.pdf UTS General Software Reference
UTS012.pdf UTS012 Test Software referemce
2 WARNINGS.
WARNINGS.
IPC Global testing apparatus are powered by mains voltages from 110-240VAC
depending on the model. No covers or panels should be removed as live terminals may
! be exposed.
If any cover is removed whilst mains power is connected any exposed terminals should
be considered live regardless of the status of the power switch.
IPC Global apparatus also operate on pneumatic pressures up to1000kPa (150 psi) and
can crush a trapped body part.
Where a confining cell (e.g. Triaxial Cell) is part of the apparatus particular care must be
paid to checking the integrity of the cell walls as due to the confining pressures utilised
there is the risk of a damaged cell exploding. Any cracks or severe scratches should be
inspected and the affected piece replaced or repaired before continued use.
Care should be taken to ensure no body parts are placed within the reaction frame such
that the part might be pinched by movement of the actuator.
4 THE BASE
BASE SYSTEM
5.2 Transducers
De-Aired
Water In
Confining
Supply
To To To To Back To Pore
Air/Water Confining Recharge Pressure Pressure
Interface Cell Port 3 Supply on Supply. Supply
Volume Cell Cell
Change Port 1 Port 2
From
Compressed
Air Source
The five ports across the cell base are configured Left to Right as follows.
Connection Connects to
Back Pressure The back pressure line connects to the back of the volume change
transducer
Pore Pressure Connects to the Pore Supply port of the Water Panel
Confining Pressure Connects to the Confining Cell Port of the Water Panel
Pore Drain Connects to the Pore Pressure Transducer and Specimen Vacuum
Generator
Back Drain Connects to a drain bucket.
Axis 1
Ctrl A Actuator displacement
Ctrl B Load cell
Ctrl C & D Spare acquisition channel
Servo Output Displacement servo valve
Axis 2
Ctrl A Confining pressure
Ctrl B Pore pressure
Ctrl C Mid Height pore pressure
(if fitted)
Ctrl D Spare channel
Servo output Confining pressure servo valve
X Recharge Solenoid Valve
Axis 3
Ctrl A Back pressure
Ctrl B Volume Change
Ctrl C&D Spare channels
Servo output Back pressure servo valve
Group 1
A External LVDT (if fitted)
B,C&D On-Specimens (if fitted)
X Back pressure Solenoid Valve
11 TRIAXIAL CELL
The pictures below details the main components of the cell assembly.
Not shown are the two pore supply and drain connections and the bottom vacuum platen.
These are essentially a mirror image of the top vacuum platen and back supply and drain
connections.
Ram coupling
Air bleed
Triaxial cell
Load cell
Top half of suction top
Back Supply
Back Drain
Connects to the
vacuum connection
on cell top
Check that all valves on all panels are in the closed position.
To Cell Back
Pressure Port
Back Pressure
To Back Pressure Transducer
Servo Valve
With the de-aired water supply open and all valves on the water distribution panel closed.
Open the air bleed tap on top of the assembly and ensure the drain hose is placed in a
bucket or container to capture the extracted water.
Open the Back Supply valve on the Water Distribution Panel to allow water to enter the
volume change / back pressure chamber. The water will enter on the left hand side as you
look at the assembly.
When a continuous stream of water flows out and no air bubble are visible then close the
tap.
Open the air supply, activate the Virtual Pendant, select Axis 3 and apply 5kPa back
pressure.
To complete the bleeding the UTS012 software must be started, a Levels screen opened
and the OPEN toolbar button clicked so the back pressure solenoid operates (see section
15). Remove the tubing at the cell Back Pressure port and hold over a suitable container to
capture any water.
Reconnect when the tube is full of water.
On the PC click CLOSE to shut off the Back Pressure Supply to the Cell.
At the end of the operation the bar protruding from the side of the assembly should be
near the bottom of its stroke. This allows the maximum amount of water to be initially
available for Back Pressure Supply to the specimen.
After removing the air, it is advisable to leave the apparatus overnight, or for at least eight
hours, with an internal pressure of approximately 700 kPa. This will allow any remaining
trapped air to be absorbed into the solution. After this period, the apparatus should be
carefully flushed out using new de-aired water, and thus displacing the aerated water. It
may be necessary to repeat this procedure once more, if any signs of air pockets occur
during the first two days of operation.
Keeping an eye on the drain line and de-airing block, air bubbles should be seen pushing
through.
Once a steady stream of water is evident then close the black ball tap on the right side of
the de-airing block and open the screw tap on the side.
Rotate the de-airing block so that the pore pressure transducer is facing downwards and
remove any air trapped in the transducer cavity.
Once complete close the screw tap and the pore water pressure supply tap on the base of
the tri-axial cell.
16 SOFTWARE
Install the software as described in the software manual.
The communication with the IMACS should be USB, This can be setup via the File-
System Setup main menu item. Once communications has been established the
transducers need to be allocated (as shown in Section 10) and the provided calibration files
will need to be loaded and save into a template file.
Secondly the Hardware options need to be configured. This can be done via the Options –
Hardware Options main menu item.
Once the software has been configured for the system click the New button followed by
clicking on the Levels button to activate the transducer levels screen.
If you have all transducers and cables correctly positioned you should see all the transducer
signals in colour on the levels screen.
You can see that the virtual pendant has an active and disabled appearance. The disabled
appears when a test is running to prevent accidental changes made during the test.
Click to AXIS 2 cell pressure and by activation the up and down buttons to raise the lower
the cell pressure.
Repeat this for the back pressure AXIS 3
You are now ready to run a test. Before running a test on a real sample it is advisable to
familiar yourselves with the operation oft the software and the performance off the system.
This will allow you to conduct many tests without having to prepare anew sample every
time.
17 AIR SYSTEM
Note:- Before applying Air pressure to the system the IMACS should be switched ON.
Failure to do this can result in uncontrolled actuator movement or pressure being applied
to the system.
17.1 Adjusting
Adjusting System Pressure and switching on.
With all the valves on the auxiliary air receiver and the controls valves closed start up the
compressor and open the valve to the auxiliary air receiver. Adjust the input regulator until
the desired working pressure is indicated on the regulator pressure gauge. This should be
up to 800kPa depending on your test requirements. A pressure of 750 – 800kPa should be
available.
Suction plate
Suction ring
Top Cap
O ring
Membrane
Dummy sample
O ring
Base Pedestal
Assemble the triaxial cell as normal as if the sample was set up on the pedestal .
This is the time when the load frame may need to be adjusted to the corrected usable
height.
When the triaxial cell is in position the actuator ram can be lowered using the virtual
pendant.
The picture below shows the arrangement between the actuator piston and the triaxial cell
ram
The alignment between the actuator ram and the triaxial cell ram is critical to the ability to
produce good waveshapes applied to the sample.
Because of the difficulty of maintaining perfect angular alignment between the actuator
and cell shafts a ball coupling is included
Click on start and you will see the drop down box. The consolidation is greyed out because
you have to saturate before consolidation. Click on saturation to move to the saturation set
up page. You will be prompted to save the
data, select a file name and save
18.2 Saturation
This is the saturation page where you can select methods of saturation. Select manual
pressure increments.
the required value click the open button. This will open the back pressure solenoid valve
and allow water to enter the sample, the volume change will be seen on the live graph
when the volume change has become stable close the valve, select increment and cell
pressure. The cell pressure will increase to the next required value and the pore pressure
coefficient B will be displaced. Repeat this procedure until the required B value is
achieved.
This procedure follows the saturation stage in BS1377 part 8. A copy of this will explain
the saturation in detail.
Because the pressure is only being applied up to the valve on the triaxial cell the saturation
can be repeated as many times as required.
18.2.2 Automatic Pressure Increment Saturation Mode
The automatic pressure increments method can only be performed on a real sample
because it requires the values of pore water and volume change transducers to activate
changes from one stage to next.
18.3 Consolidation
A drop down box shows the options available for consolidation. Select Isotropic and enter
a cell pressure and time to reach that pressure.
Note: For K0 consolidation modes additional transducers are required.
Once the target confining cell pressure has been reached and has stabilised then press the
Start button to begin the consolidation stage.
After consolidation select the Cyclic shear setup parameters tab and the following page will
appear.
Enter data as shown and click the Start button followed by the Cyclic Shear sub-menu item
and the system will perform the cyclic stage as described in the set up page.
By adjusting the Proportional and Integral gains the measured load can be adjusted to
equal the requested load and a correct wave shape obtained.
Tests can be done using the dummy sample to cover all the variations offered in the
software before starting tests for real
19 RUNNING A TEST
Move the ram down towards the sample until it just touches the rubber suction ring.
Clamp in this position Zero the load cell reading by right clicking and set zero
Bring the actuator down to make contact with the triaxial cell ram and clamp together with
the knurled nut. The actuator should be at about mid travel or just above. On the Levels
screen right click on the actuator Displacement bar and set zero offset.
Free the triaxial ram clamp and slowly bring the actuator down using the virtual pendant
to make contact with the sample. Keeping the reading at zero using the virtual pendant
When contact is made, apply a vacuum to the top cap. This is done by opening the air
supply to the platen vacuum generator.
19.10 Filling the Cell with Water
Fill the triaxial cell with water, when water escapes from the air bleed close the cell valve to
stop water entering. Do not closed the bleed valve first as this will trap the supply pressure
in the triaxial cell.
Closed
Open
19.12 Select New and enter tests data and consolidation type
Click start and select Saturation
When saturation is complete close window and click start and select Consolidation
One of these consolidation methods would have
been selected in the initial set up page
19.12.4 Isotropic
Enter target cell pressure and time to reach target
cell pressure as shown
19.12.5 Anisotropic
The anisotropic consolidation stage allows for up
to five stages to be performed too reach the final
consolidation stress.
19.13 Ko Consolidation
The following shows examples of the different stress path types with test data entered
When the test has been completed. Click unload and the stresses will reduce to zero.
20 MAINTENANCE.
20.1 Daily.
Daily.
20.1.1 Frame.
The frame and surrounding area should be cleaned of specimen residue and fluid spills at
the end of each day
20.1.2 IMACS.
If the IMACS has had fluid spilled on it immediately switch off and disconnect the power
lead from the wall outlet. If necessary unscrew each module and withdraw and using low
pressure compressed air blow away any moisture evident on the circuit boards.
The Power Supply Module (rightmost position) should have the side covers removed and
any moisture should be blown off with low pressure air. Leave the modules in a warm, dry
area to dry out before re-fitting.
20.1.3 Cell.
Check the cell acrylic for any signs of damage (scratches or crazing). Any scratches should
be removed using a very mild abrasive (e.g. Brasso) on a rag rubbed over the area. Do Not
use solvents to clean the acrylic as the surface will craze. Use only soapy water to clean the
acrylic.
Check all o-rings for wear and damage. Replace as necessary
20.1.4 Air Reservoir.
Check the moisture traps for signs of moisture and empty of any fluid trapped. If fluid is
evident the compressed air supply should be checked for moisture as well.
20.2 Weekly
The weekly checks consist of the daily checks listed above plus checking the reaction frame
for leaks and evidence of corrosion on plated metal surfaces. Wipe corroded areas with a
rag sprayed with WD-40 or equivalent water dispersant.
20.3 Monthly
Monthly checks are the same as weekly.
20.4 Yearly.
Yearly checks are the same as monthly.
20.4.1
APPENDIX B Troubleshooting.
Table 4 Troubleshooting
Symptom Possible Cause Fix
No control Power switched Off Switch On IMACS Power switch(es).
IMACS fuse(s) blown Replace fuse(s)
PC Communications errors File\System setup\Comms\re-open cdas port
Air Supply OFF Switch on Air supply
For more troubleshooting assistance please contact your IPC Global Agent or
IPC Global Technical Support Directly
Email techsupport@ipcglobal.com.au
Phone: +613 9800 2200
- END -