Overview
Index
897N102438A
Introduction
This Manual is applicable to the following software.
• DRYPIX Smart Main Software V. 1.0 or later
This Manual provides an explanation about the Quality Control (hereafter indicated as QC) function
of the Fuji Medical Dry Laser Imager DRYPIX 6000 (hereinafter referred to as “DRYPIX Smart”).
The QC function can only be used if the QC function is set to “Available” during installation. The QC
is a function to implement the quality control of the images printed by the DRYPIX Smart.
For the concepts behind the quality control, refer to the following guidebook.
• “FCR Quality Assurance Program Plus – For Medical Institutions – Image Recorder Guidebook”
(sold separately, hereafter indicated as “The guidebook”)
Accompanying documents were originally drafted in the English language.
CAUTIONS
1. No part or all of this manual may be reproduced in any form without prior permission.
2. The information contained in this manual may be subject to change without prior notice.
3. FUJIFILM Corporation shall not be liable for malfunctions and damages resulting from
installation, relocation, remodeling, maintenance, and repair performed by other than
dealers specified by FUJIFILM Corporation.
4. FUJIFILM Corporation shall not be liable for malfunctions and damages of FUJIFILM
Corporation products due to products of other manufacturers not supplied by FUJIFILM
Corporation.
5. FUJIFILM Corporation shall not be liable for malfunctions and damages resulting from
remodeling, maintenance, and repair using repair parts other than those specified by
FUJIFILM Corporation.
6. FUJIFILM Corporation shall not be liable for malfunctions and damages resulting from
negligence of precautions and operating methods contained in this manual.
7. FUJIFILM Corporation shall not be liable for malfunctions and damages resulting from use
under environment conditions outside the range of using conditions for this product such
as power supply, installation environment, etc. contained in this manual.
8. FUJIFILM Corporation shall not be liable for malfunctions and damages resulting from
natural disasters such as fires, earthquakes, floods, lightning, etc.
Trademark
FCR is a trademark or a registered trademark of FUJIFILM Corporation.
Overview
Chapter 1 This chapter provides an overview of the quality control of the DRYPIX Smart’s
images as well as an overview of the QC functions to implement quality control.
Appendix
Appendix A worksheet is attached for a recording of the results of the QC functions. Copy
the worksheet to use it.
Index
Chapter 1 Overview
1.1 The Quality Control (QC) Function���������������������������������������������������������������������������� 1-1
1.2 Precautions When Implementing the Uniformity Test Using the QC Function���������� 1-2
Index
Overview
In order to maintain the quality of the images printed by the DRYPIX Smart, this device is subjected
to an uniformity test. With the DRYPIX Smart keep the following points in mind when implementing
the QC function for the uniformity test.
1. Use the DRYPIX Smart to print the “test patterns” used for quality control.
2. Make sure that no error code is displayed on the display panel of the DRYPIX Smart.
The measured result and the measured results besides those of the QC test pattern are recorded
on the worksheet.The uniformity test is periodically implemented and its results are recorded
each time.Check the results measured by the uniformity test do not exceed the standard value.
Contact our official dealer should the measured results exceed the standard value. In this case, the
worksheet with a record of the results measured by the uniformity test is required.
Quality control can be implemented using other means than the uniformity test. In addition, the
uniformity test can be implemented without the use of the QC function, however the use of the
uniformity test will make quality control easier and more reliable.
1. External Densitometer
For the external densitometer, use a densitometer calibrated with the standard density step tablet
that complies with the expiration date. The use of a standard density step tablet manufactured
based on a density step tablet (NIST : National Institute of Standards and Technology (US)) is
recommended. The recommended density step tablet and densitometer are as follows.
● Recommended densitometer
Manufacturer : x-rite
Product name : 331
2. Plastic Ruler
Use a general straight ruler enabling measurement down to the millimeter. The use of a transparent
ruler with an easy to check indication is recommended.
3. Worksheet
Make a copy of the worksheet in the appendix of this Manual and use that.
2
Select the tray
Select either tray 1 or tray 2.
Implementation of the QC Function
Checking artifacts
Visually check the QC test pattern for irregularities
and disorderly patterns.
Record the visual check results in the worksheet.
you confirm the presence of artifacts.
The “Guidebook” provides details of the standard values (basic values) and the setting standard.
HINT
The utility mode starts.
Basic Operations in Utility Mode
The Display panel changes and shows “U-1”.
The table below shows the basic Utility menu
operations available when the printer has
switched to Utility Mode.
2
10 seconds have elapsed without any operation.
HINT
When a QC test pattern is output, the Utility mode
Implementation of the QC Function
Slot 1 is displayed.
To b e u s e d f o r t h e
measurement by the
2 external densitometer
Format B
of density 0, 1, 2 and 3. Measure format B
*(When set at time of using a plastic ruler.
installation, density 4
Implementation of the QC Function
Format B
Format B
Format A
Format A
Format B
Format B
Format A
Format A
2
Implementation of the QC Function
lW
hen density 4 is set using the service settings, density
4 is measured as well.
1 Using a plastic ruler, measure the length 2 Record the format measurement results in
(mm) [Format A] and [Format B] of the QC the worksheet.
test patterns. Record the measured results and the determination
Format B of the measurement (PASS/FAIL) in the worksheet.
● Determination of the measurement results
PASS........ When the format length is within the
2
standard range of the standard values
FAIL.......... When the values exceeds the range above
No distorted and
bent lines
Measurement results of the print density: Record the measurement values at density 0-4 with the judgment result. (*Density 4 is the service setting)
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21
Density 4 3.30
3.25
3.20
3.15
3.10
3.05
3.00
Appendix
2.95
2.90
2.85
2.80
2.75
2.70
Measurement value
Determination of PASS PASS PASS PASS PASS PASS PASS PASS PASS PASS PASS PASS PASS PASS PASS PASS PASS PASS PASS PASS PASS
the measurement FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21
Density 3 2.20
2.15
2.10
2.05
2.00
1.95
1.90
1.85
1.80
1.75
1.70
1.65
Measurement value
Determination of PASS PASS PASS PASS PASS PASS PASS PASS PASS PASS PASS PASS PASS PASS PASS PASS PASS PASS PASS PASS PASS
the measurement
FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21
Density 2 1.45
1.40
1.35
1.30
1.25
1.20
1.15
1.10
1.05
1.00
0.95
Measurement value
Determination of PASS PASS PASS PASS PASS PASS PASS PASS PASS PASS PASS PASS PASS PASS PASS PASS PASS PASS PASS PASS PASS
the measurement
FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL
0.55
0.50
0.45
0.40
0.35
0.30
Measurement value
Determination of PASS PASS PASS PASS PASS PASS PASS PASS PASS PASS PASS PASS PASS PASS PASS PASS PASS PASS PASS PASS PASS
the measurement
FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21
Density 0 0.30
0.25
0.20
0.15
0.10
Measurement value
Determination of PASS PASS PASS PASS PASS PASS PASS PASS PASS PASS PASS PASS PASS PASS PASS PASS PASS PASS PASS PASS PASS
the measurement
FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL
Appendix
Format Measurement Result: Record the measurement value of format A and B along with the determination of the measurement in the table below.
Format A
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21
Measured
value (mm)
Determination of PASS PASS PASS PASS PASS PASS PASS PASS PASS PASS PASS PASS PASS PASS PASS PASS PASS PASS PASS PASS PASS
the measurement FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL
Format B
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21
Measured
value (mm)
Determination of PASS PASS PASS PASS PASS PASS PASS PASS PASS PASS PASS PASS PASS PASS PASS PASS PASS PASS PASS PASS PASS
the measurement
FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL
Artifact Determination of the Measurement: Record the determination of the measurement of the artifact in the table below.
Artifact
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21
Determination of PASS PASS PASS PASS PASS PASS PASS PASS PASS PASS PASS PASS PASS PASS PASS PASS PASS PASS PASS PASS PASS
the measurement FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL
Auto Density Measurement Result: Record the result of the Auto Density Measurement in the table below.
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21
Determination of PASS PASS PASS PASS PASS PASS PASS PASS PASS PASS PASS PASS PASS PASS PASS PASS PASS PASS PASS PASS PASS
the measurement FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL