Вы находитесь на странице: 1из 8

See discussions, stats, and author profiles for this publication at: https://www.researchgate.

net/publication/225358264

Puncture of elastomer membranes by medical needles. Part I: Mechanisms

Article  in  International Journal of Fracture · March 2009


DOI: 10.1007/s10704-009-9326-7

CITATIONS READS
27 776

4 authors, including:

Chien Thang Nguyen Patricia I. Dolez


Duy Tan University 96 PUBLICATIONS   538 CITATIONS   
19 PUBLICATIONS   201 CITATIONS   
SEE PROFILE
SEE PROFILE

Jaime Lara
Université de Montréal
181 PUBLICATIONS   836 CITATIONS   

SEE PROFILE

Some of the authors of this publication are also working on these related projects:

Practure, Fatigue of Polymer, Steel Tube View project

Development of a method of measuring nanoparticle penetration through protective glove under conditions simulating workplace use View project

All content following this page was uploaded by Jaime Lara on 07 July 2014.

The user has requested enhancement of the downloaded file.


Int J Fract (2009) 155:75–81
DOI 10.1007/s10704-009-9326-7

ORIGINAL PAPER

Puncture of elastomer membranes by medical needles.


Part I: Mechanisms
C. Thang Nguyen · Toan Vu-Khanh ·
Patricia I. Dolez · Jaime Lara

Received: 6 November 2008 / Accepted: 12 February 2009 / Published online: 4 March 2009
© Springer Science+Business Media B.V. 2009

Abstract Resistance to puncture is a critical prop- similar way as for cutting. A lubricant was tentatively
erty for several applications, in particular for elastomer used to reduce the friction contribution for the compu-
materials used in protective clothing. To evaluate the tation of the material fracture energy.
puncture resistance of membranes, some methods have
been proposed as standard tests. However, the rounded Keywords Puncture · Elastomer · Medical needle ·
puncture probes used in these tests are very different Friction · Fracture energy
from real pointed objects like medical needles, and may
not measure the level of material resistance that cor-
responds to them. In fact, puncture by medical nee- 1 Introduction
dles is shown to proceed gradually as the needle cuts
into the membrane. This behavior is highly different Puncture resistance is among the major mechanical
from puncture by rounded probes which occurs sud- properties of elastomer membranes, especially in the
denly when the strain at the probe tip reaches the failure case of their use in protective clothing. In terms of punc-
value. In addition, maximum force values are observed ture agents, medical needles are becoming an increas-
to be much smaller with medical needles. A method has ingly encountered mechanical hazard, not only in health
been developed based on the change in strain energy care but also for law enforcement and maintenance
with the puncture depth to evaluate the fracture energy occupations, with the associated risk of blood-borne
associated to puncture. The results show that the phe- pathogen transmission. Some investigations have been
nomenon of puncture by medical needles involves con- performed on specific cases using medical needles.
tributions both from friction and fracture energy, in a In two studies, a total of seven commercially avail-
able surgical gloves and glove liners were tested for
C. T. Nguyen
Faculté de Génie, Université de Sherbrooke, Sherbrooke,
resistance to puncture with different types of medi-
QC, Canada cal needles (Leslie et al. 1996; Hewett, 1993, personal
communication). The main goal was to rank the gloves
T. Vu-Khanh (B) · P. I. Dolez and compare the protection efficiency of new mate-
Département de génie mécanique, École de technologie
supérieure, 1100 rue Notre-Dame Ouest, Montréal, QC,
rials with regular latex gloves. In another work, the
H3C 1K3, Canada influence of various needle characteristics on the resis-
e-mail: toan.vu-khanh@etsmtl.ca tance to puncture by medical needles of materials rele-
vant to protective gloves was investigated (Dolez et al.
J. Lara
Institut de recherche Robert-Sauvé en santé et en sécurité du
2008). However, these studies were only qualitative
travail, Montréal, QC, Canada for a comparison purpose and did not provide a

123
76 C. T. Nguyen et al.

that the puncture of rubber membranes is controlled


by a critical local deformation at the probe tip that is
independent of the indentor geometry.
Furthermore, the puncture probes used in the ASTM
F1342 standard and other available standard test meth-
ods are very different from real pointed objects like
medical needles. In particular, their rounded tip does
not bear the cutting edge of medical needles as illus-
trated in Fig. 2.
Results have also shown that the shape of the force-
displacement curves recorded during puncture is very
different for ASTM-type rounded probes and medi-
cal needles (Nguyen et al. 2005). While puncture by
Fig. 1 Schematic representation of the ASTM F1342 probes, rounded probes occurs suddenly when the strain at the
a probe A, b probe B, and c probe C
probe tip reaches the failure value, medical needles pen-
etrate gradually through the sample. Puncture forces
measured with medical needles have also been reported
fundamental understanding of the mechanisms con- to be much smaller (Leslie et al. 1996; Hewett, 1993,
trollling puncture. personal communication, Dolez et al. 2008).
To evaluate the puncture resistance of materials, the From a fundamental point of view, the puncture
ASTM F1342 standard test (ASTM F1342 2005) is cur- caused by sharp-pointed objects like medical needles is
rently the most commonly used method. This standard a complex phenomenon that may involve various frac-
test method considers the use of one of three alterna- ture processes. For elastomer membranes tearing and
tive puncture probes (see Fig. 1); Probe A has a 2 mm cutting are the two major modes of failure that have
diameter and a conic extremity with an angle of 26◦ been investigated. While tearing only involves the frac-
and a rounded tip with a radius of 0.25 mm; Probe B ture energy of the material, both the fracture energy and
has a 1 mm diameter and a spherical extremity with the friction caused by the contact between the blade
a rounded tip radius of 0.50 mm; Probe C is similar and the material contribute to cutting (Gent et al. 1994;
to probe A but with a rounded tip radius of 0.50 mm. Vu Thi et al. 2005). Based on the results obtained dur-
The test is designed for any type of protective cloth- ing a preliminary study carried out with neoprene, it
ing, including coated fabrics, laminates, textiles, plas- was shown that puncture by medical needles seems to
tics, elastomeric films or flexible materials. This test involve cutting, which is related to the material fracture
method determines the puncture resistance of a mate- energy (Nguyen et al. 2005). By comparison, puncture
rial by measuring the maximum force required for a by ASTM-type rounded probes has been related to the
puncture probe that moves at a speed of 500 mm/min material tension strength and failure strain. However,
to penetrate through a specimen clamped between two it must be noted that a large difference in geometrical
plates with chamfered holes not more than 10 mm in configuration exists between cutting and needle punc-
diameter. ture. Indeed, cutting involves a rectangular blade slicing
In previous studies using rounded probes (Nguyen through the whole sample thickness at the same time
and Vu-Khanh 2004; Nguyen et al. 2004), it was found while moving along its length. On the other hand, nee-
that the probe tip geometry strongly affects the results dle puncture involves an elliptical sharp tip penetrating
in puncture of elastomer membranes. The maximum through the sample surface, into its thickness.
puncture force depends on the contact surface between The aim of this work is to investigate the mate-
the membrane and the probe tip. Using the Mooney rial parameters that control the puncture resistance of
strain-energy function, the indentation force was cal- thin elastomer membranes to sharp-pointed objects like
culated for elastomer membranes with large deforma- medical needles. In particular, the contributions of intrin-
tions in the absence of friction. The puncture strengths sic material parameters to puncture by medical needles
of elastomer membranes were found to be lower than are verified. An approach is derived to compute the frac-
their tensile and biaxial strengths. It was also found ture energy associated to puncture by medical needles.

123
Puncture of elastomer membranes 77

Fig. 2 a Puncture probe A


used in ASTM F1342
standard test; b Medical
needle

Fig. 4 Schematic
representation of medical
needles

Fig. 3 Sample holder setup


Table 1 Medical needles used as puncture probes
Probe diameter: d (mm) 0.35 0.50 0.65
The contribution of the friction energy to the calculated
Probe tip angle: α 13.5◦ 11.2◦ 10.4◦
fracture energy is also discussed.

2 Experimental

Puncture tests were carried out on an Instron 1137 Two types of commercial rubbers commonly used
universal-testing machine. The puncture probes are for protective gloves, neoprene and nitrile rubber, were
held by a pin chuck mounted on a 167.7 N load cell. investigated. Neoprene sheets with three different
A setup [see Fig. 3] comprising two steel plates bear- thicknesses, 0.40, 0.78 and 1.57 mm, were obtained
ing 10 mm holes is used to secure the elastomer sam- from Fairprene Industrial Products. The nitrile rubber
ples. The edge of the hole in the lower plate is rounded samples, 0.83 mm thick, were cut from nitrile rubber
to avoid stress concentration. All the tests were per- gloves manufactured by Ansell Co (glove model
formed with a displacement rate of 50 mm/min unless Sol-Vex® 37-165).
specified otherwise. For each condition, a minimum of Force-displacement data were recorded for the var-
four replicates were produced. ious puncture tests. Puncture force was measured as
In addition to the ASTM puncture probe A (see the maximum of the force-displacement curve. Nee-
Fig. 2a), a series of stainless steel medical needles dles were reused up to five times for puncture tests.
obtained from PrecisionGlideTM Pharma Co were used Indeed, previous work has shown an increase in punc-
as puncture probes in this study. Their external diameter ture force of less than 7% after ten successive uses of
and tip angle measured as shown in Fig. 4 are provided the same needle as puncture probe (Vu-Khanh et al.
in Table 1. 2005).

123
78 C. T. Nguyen et al.

Table 2 Puncture test results measured with ATSM punc-


ture probe A and 0.5-mm diameter medical needles for three
thicknesses of neoprene (coefficient of variation in parenthesis)
Sample thickness (mm) 0.40 0.78 1.57

ASTM F1342 conical probe A


Puncture force (N) 5.1 (4%) 10.4 (3%) 19.2 (3%)
Probe displacement at 14.7 (14%) 14.9 (12%) 15.3 (13%)
puncture (mm)
Medical needle 0.5 mm
Puncture force (N) 0.5 (8%) 1.4 (7%) 2.2 (5%)
Probe displacement 2.7 (7%) 3.6 (6%) 5.3 (6%)
at puncture (mm)

Fig. 5 Typical force—displacement curve for puncture with


ASTM conical probe A (0.8- mm thick neoprene)
In addition to the difference in the shape of the force-
displacement curves, a strong reduction in the values
of the maximum force and probe displacement at max-
imum force is observed with medical needles as illus-
trated in Table 2 for three thicknesses of neoprene.
The very slow puncture speed (0.05 mm/min) used
for Fig. 6 allows a clear observation of the puncture pro-
cess with medical needles. Before label 1 on the graph,
the probe produces only a deformation of the sample.
Label 1 indicates the point where the crack starts being
initiated as illustrated in Fig. 7a. Between labels 1 and
2, the probe penetrates further into the sample thick-
ness. At label 2, the probe has reached the bottom face
of the sample, corresponding to the situation illustrated
Fig. 6 Typical force—displacement curve for puncture with
medical needles (0.8- mm thick neoprene, 0.5- mm diameter in the schematic in Fig. 7b.
medical needle and 0.05 mm/min displacement rate) When the needle tip emerges on the bottom surface
of the membrane (label 3 on the graph in Fig. 6 and
schematic in Fig. 7c), the needle tip can widen easily
3 Results and discussions the punctured hole. As a consequence, the deforma-
tion of the sample under the needle tip is released and
Puncture tests carried out with the ASTM puncture the whole sample moves upward, with a decline in the
probe A and the medical needles were compared using force. The plateau observed in Fig. 6 at large displace-
the same thickness of neoprene. As shown in Figs. 5 ments corresponds to the configuration where the cylin-
and 6 and in agreement to what has already been drical shaft of the needle is in contact with the sample.
reported in the literature (Nguyen et al. 2005; Dolez The measured force thus relates to the friction between
et al. 2008), large differences can be observed in the the probe shaft and the puncture hole circumference.
shape of the force-displacement curve. For the rounded Figure 8 displays a zoom on the interval correspond-
probes, puncture occurs instantly at maximum load ing to the needle tip first penetration into the sample
(Fig. 5); it has been associated to the point where the (i.e. between labels 1 and 2 in Fig. 6), with the probe
strain at the probe tip reaches the failure value (Nguyen displacement data converted to crack depth values (see
et al. 2005). On the contrary, needles penetrate grad- insert in Fig. 8). It can be seen that, for neoprene, the
ually through the sample thickness (Fig. 6); after the puncture force varies linearly with crack depth. In addi-
force reaches a maximum, it diminishes slightly before tion, the slope is independent of the sample thickness
reaching a plateau. on the studied range (between 0.4 and 1.5 mm).

123
Puncture of elastomer membranes 79

Fig. 7 Schematic
representation of the sample
deformation during the
puncture process by medical
needles

Fig. 8 Variation of the


puncture force with crack
depth for three thicknesses
of neoprene (0.5- mm
diameter medical needles)

2
Puncture force (N) at crack

1,5
started

0,5
Fig. 10 Fracture surface created into a neoprene sample by a
medical needle (optical microscopy, 20x)
0
0 1 2 3 4
Sample thickness (mm) Earlier studies have suggested that the puncture pro-
Fig. 9 Variation of the puncture force at crack start (F1) as a
cess associated with medical needles involves a cutting
function of the sample thickness for neoprene (0.5- mm diameter phenomenon (Nguyen et al. 2005). Other studies have
medical needles) related cutting to the fracture energy G c of the material
(Lake and Yeoh 1987; Cho and Lee 1998). According
In Fig. 9, the force at crack start, i.e. at the point cor- to the energy balance of fracture mechanics, the energy
responding to label 1, is expressed as a function of the necessary to create a new fracture surface, the fracture
sample thickness in the case of neoprene. An increase energy G s , can be expressed by (Felbeck and Atkins
in the force with the sample thickness can be observed. 1996):
 
This result is rather surprising since at initiation, only ∂U
Gs = − ≈ −U/A (1)
the sample surface is penetrated and the contact force at ∂A
the needle tip should be the same. Therefore it suggests where U is the change in strain energy correspond-
that, in order for a crack to start, the local strain defor- ing to the change in fracture surface A. For the case
mation of the sample around the needle tip must reach of needle puncture, the fracture surface created into
a critical value. For thicker samples, a larger force is the sample by medical needles is elliptical as shown in
needed to create the same level of deformation. Fig. 10.

123
80 C. T. Nguyen et al.

Fig. 11 Change in strain


energy due to different
puncture depths

Fig. 12 Measured puncture


energy as a function of
crack depth for various
thicknesses of neoprene and
a 0.65- mm diameter
medical needle

Table 3 Extrapolated
Extrapolated fracture energy Gs (kJ/m2 ) Probe diameter: d (mm)
fracture energy of puncture
for neoprene by medical 0.35 0.50 0.65
needles (coefficient of
variation in parenthesis) Sample thickness (mm) 0.40 1.7 (18%) 1.9 (14%) 1.8 (17%)
1.57 1.8 (15%) 1.7 (14%) 1.7 (12%)

U can be measured from the loading curves: when puncture energy G s as a function of the puncture depth
the puncture has reached a certain depth into the sam- for different neoprene thicknesses. The results suggest
ple, corresponding to a fracture surface Ai , the needle is a slight increase in G s with the puncture depth, i.e. with
withdrawn to get the return curve. Figure 11 illustrates puncture propagation.
how the change in strain energy U is computed by A first approximation of the fracture energy at punc-
subtracting the released energies corresponding to the ture initiation can be provided by the extrapolation at
different puncture depths. The changes in strain energy crack depth d = 0 of the measured puncture energies
−Ui j corresponding to the change of fracture surface shown in Fig. 12. This energy is displayed in Table 3 for
Ai j are thus: three diameters of the medical needles and two thick-
−Ui j = Ui − U j ; nesses (0.40 and 1.57 mm) of neoprene. The results
(2) seem to suggest a constant value of puncture energy at
Ai j = Ai − A j
initiation. However, a significant contribution of fric-
Similar tests carried out on the same sample and with tion to the fracture energy has already been reported
the same needle at different puncture depths provide for cutting (Vu Thi et al. 2005).
different values of the released energy: −U1 , −U2 . . ., In order to verify the hypothesis of the friction con-
and the corresponding fracture surfaces: A1 , A2 . . .. tribution to the puncture energy, a BPTM lubricant was
Figure 12 displays the variation of the measured sprayed on the medical needles before their use as a

123
Puncture of elastomer membranes 81

Table 4 Extrapolated
Extrapolated fracture energy (kJ/m2 ) Probe diameter: d (mm)
fracture energy for puncture
by medical needles with and 0.35 0.50 0.65
without lubricant
(coefficient of variation in Neoprene (1.57 mm) Without lubricant 1.8 (15%) 1.7 (14%) 1.7 (12%)
parenthesis) With lubricant 1.6 (13%) 1.6 (13%) 1.5 (11%)
Nitrile rubber (0.83 mm) Without lubricant 3.8 (18%) 4.1 (14%) 4.2 (17%)
With lubricant 3.6 (14%) 3.7 (13%) 4.0 (12%)

puncture probe. Table 4 displays the results obtained International Annual Book of ASTM Standards, vol 11.03.
in terms of extrapolated fracture energy both with and West Conshohocken, PA, pp 1489–1493
Cho K, Lee D (1998) Viscoelastic effects in cutting
without lubricant for three medical needle diameters of elastomers by a sharp object. J Polym Sci
and two membrane materials, 1.57- mm thick neoprene Part B Polym Phys 36(8):1283–1291. doi:10.1002/
and 0.83- mm thick nitrile rubber. The values of frac- (SICI)1099-0488(199806)36:8<1283::AID-POLB3>3.0.
ture energy seem to be slightly reduced by the use of CO;2-T
Dolez P, Vu-Khanh T, Nguyen CT, Guero G, Gauvin C, Lara
the lubricant even if the difference is not statistically J (2008) Influence of medical needle characteristics on the
significant. In the case of nitrile rubber, the effect of resistance to puncture of protective glove materials. J ASTM
friction does not appear to be totally removed by the Int 5(1):1–12
use of lubricant as shown by the increasing values of the Felbeck DK, Atkins AG (1996) Strength and Fracture of Engi-
neering Solids, 2nd edn. Prentice-Hall Inc, NJ
extrapolated fracture energy with the needle diameter. Gent AN, Lai S-M, Nah C, Wang C (1994) Viscoelastic effects
in cutting and tearing rubber. Rubber Chem Technol
67(4):610–619
4 Conclusion Lake GJ, Yeoh OH (1987) Effect of crack tip sharpness on the
strength of vulcanized rubbers. J Polym Sci 25:1157–1190
Leslie LF, Woods JA, Thacker JG, Morgan RF, McGregor W,
This work has demonstrated that the puncture by sharp- Edlich RF (1996) Needle puncture resistance of medical
pointed objects like medical needles is very different gloves, finger guards, and glove liners. J Biomed Mater
from the puncture by conical probes like those used Res 33:41–46. doi:10.1002/(SICI)1097-4636(199621)33:
in the ASTM standard test for protective clothing. For 1<41::AID-JBM7>3.0.CO;2-M
Nguyen CT, Vu-Khanh T (2004) Mechanics and mecha-
elastomer materials, the puncture by conical probes is nisms of puncture of elastomer membranes. J Mater
controlled by a local deformation or failure strain. On Sci 39(24):7361–7364. doi:10.1023/B:JMSC.0000048751.
the other hand, for medical needles, the puncture pro- 55710.44
cess involves crack growth and fracture energy dissi- Nguyen CT, Vu-Khanh T, Lara J (2004) Puncture characteriza-
tion of rubber membranes. Theor Appl Fract Mech 42:25–
pation. A method based on the change in strain energy 33. doi:10.1016/j.tafmec.2004.06.002
with the change in fracture surface is proposed for the Nguyen CT, Vu-Khanh T, Lara J (2005) A Study on the puncture
characterization of the crack initiation energy due to resistance of rubber materials used in protective clothing.
puncture. However, even with the application of a lubri- J ASTM Int 2(4):245–258. doi:10.1520/JAI12147
Vu-Khanh T, Vu Thi BN, Nguyen CT, Lara J (2005) Protective
cant on the needle surface, the effect of friction on the gloves: Study of the resistance of gloves to multiple mechan-
puncture process was not eliminated, preventing the ical aggressors. Rapport Études et Recherche R-424, Institut
determination of the material fracture energy. de recherche Robert-Sauvé en santé et en sécurité au travail,
Montréal, QC, Canada, 74 pp
Acknowledgements This work has been supported by the Insti- Vu Thi BN, Vu-Khanh T, Lara J (2005) Effect of friction on
tut de recherche Robert-Sauvé en santé et en sécurité du travail. cut resistance of polymers. J Thermoplast Compos Mater
18(1):23–36. doi:10.1177/0892705705041157

References

ASTM Standard F1342 (2005) Standard test method for pro-


tective clothing material resistance to puncture. ASTM

123

View publication stats

Вам также может понравиться