Proline Promass 40
Coriolis Mass Flow Measuring System
BA061D/06/en/03.10
71111274
Valid as of version:
V 3.01.XX (Device software)
Proline Promass 40
2 Endress+Hauser
Proline Promass 40 Table of Contents
Table of Contents
Endress+Hauser 3
Table of Contents Proline Promass 40
4 Endress+Hauser
Proline Promass 40 1 Safety instructions
1 Safety instructions
Endress+Hauser 5
1 Safety instructions Proline Promass 40
• External surface temperature of the transmitter can increase by 10 K due to power consumption
of internal electronical components. Hot process fluids passing through the measuring device will
further increase the surface temperature of the measuring device. Especially the surface of the
sensor can reach temperatures which are close to process temperature. Additionally safety
precautions are required when increased process temperatures are present.
• The manufacturer reserves the right to modify technical data without prior notice. Your
Endress+Hauser distributor will supply you with current information and updates to this
Operating Instruction.
1.4 Return
The following procedures must be carried out before a flowmeter requiring repair or calibration,
for example, is returned to Endress+Hauser:
• Always enclose a duly completed "Declaration of contamination" form.
Only then can Endress+Hauser transport, examine and repair a returned device.
• Enclose special handling instructions if necessary, for example a safety data sheet as per
Regulation (EC) No 1907/2006 REACH.
• Remove all residues. Pay special attention to the grooves for seals and crevices which could
contain residues. This is particularly important if the substance is hazardous to health,
e.g. flammable, toxic, caustic, carcinogenic, etc.
! Note!
You will find a preprinted blank of the "Declaration of contamination" form at the back of this
manual.
# Warning!
• Do not return a measuring device if you are not absolutely certain that all traces of hazardous sub-
stances have been removed, e.g. substances which have penetrated crevices or diffused through
plastic.
• Costs incurred for waste disposal and injury (burns, etc.) due to inadequate cleaning will be
charged to the owner operator.
# Warning!
"Warning" indicates an action or procedure which, if not performed correctly, can result in injury
or a safety hazard. Comply strictly with the instructions and proceed with care.
" Caution!
"Caution" indicates an action or procedure which, if not performed correctly, can result in incorrect
operation or destruction of the device. Comply strictly with the instructions.
! Note!
"Note" indicates an action or procedure which, if not performed correctly, can have an indirect
effect on operation or trigger an unexpected response on the part of the device.
6 Endress+Hauser
Proline Promass 40 2 Identification
2 Identification
5 6
ENDRESS+HAUSER
PROMASS 40
Order Code: 40E25-XXXXXXXXXXXX IP67/NEMA/Type4X
1
Ser.No.: 12345678901
TAG No.: ABCDEFGHJKLMNPQRST
16-62VDC/20-55VAC 15VA/W
2 50-60Hz
A0009770
Endress+Hauser 7
2 Identification Proline Promass 40
ENDRESS+HAUSER
PROMASS E
Order Code: 40E25-XXXXXXXXXXXX
1 Ser. No.: 12345678901
TAG No.: ABCDEFGHJKLMNPQRST
2 K-factor: 2.510 / -11
i
3 Size: DN 25 / 1”
4 DN25 / 1" DIN EN PN 100 R
FEK0716-0002
3P-CAL, 0.8um/150grit
9 IP 67 NEMA/Type4X
Pat. US 5,610,342
8 -40°C (-40°F)< Tamb <+60°C (140°F)
10
A0009769
8 Endress+Hauser
Proline Promass 40 2 Identification
A: active
See operating manual 2
P: passive
Betriebsanleitung beachten
NO: normally open contact
Observer manuel d'instruction 3
NC: normally closed contact
1 2
20(+) / 21(-)
22(+) / 23(-)
24(+) / 25(-)
26(+) / 27(-)
1 Ser.No.: 12345678912
Supply / L1/L+
4 Versorgung / N/L-
Tension d'alimentation
PE
Active: 0/4...20mA, RL max. = 700 Ohm
A
Passive: 4...20mA, max. 30VDC
I-OUT (HART)
(HART: RL.min. = 250 OHM)
fmax = 1kHz
Active: 24VDC/25mA (max. 250mA/20ms) f-OUT P
Passive: 30VDC, 250mA
5
Passive: 30VDC, 250mA
STATUS-OUT X
3...30VDC, Ri = 5kOhm
STATUS-IN X
10
A0000963
Endress+Hauser 9
2 Identification Proline Promass 40
The devices comply with the applicable standards and regulations in accordance with EN 61010-1
"Protection Measures for Electrical Equipment for Measurement, Control, Regulation and Labora-
tory Procedures" and with the EMC requirements of IEC/EN 61326.
The measuring system described in this Operating Instruction thus complies with the statutory
requirements of the EC Directives. Endress+Hauser confirms successful testing of the device by
affixing to it the CE mark.
The measuring system is in conformity with the EMC requirements of the "Australian Communica-
tions and Media Authority (ACMA)".
SWAGELOK ®
is a registered trademark of Swagelok & Co., Solon, USA
HART ®
is a registered trademark of HART Communication Foundation, Austin, USA
10 Endress+Hauser
Proline Promass 40 3 Installation
3 Installation
3.1.2 Transport
The following instructions apply to unpacking and to transporting the device to its final location:
• Transport the devices in the containers in which they are delivered.
• The covers or caps fitted to the process connections prevent mechanical damage to the sealing
faces and the ingress of foreign matter to the measuring tube during transportation and storage.
Consequently, do not remove these covers or caps until immediately before installation.
• Do not lift measuring devices of nominal diameters ≥ DN 40 (≥ DN 1½") by the transmitter
housing or the connection housing in the case of the remote version (Fig. 4). Use webbing slings
slung round the two process connections (Fig. 4). Do not use chains, as they could damage the
housing.
# Warning!
Risk of injury if the measuring device slips. The center of gravity of the assembled measuring device
might be higher than the points around which the slings are slung. At all times, therefore, make sure
that the device does not unexpectedly turn around its axis or slip.
A0004294
3.1.3 Storage
Note the following points:
• Pack the measuring device in such a way as to protect it reliably against impact for storage
(and transportation). The original packaging provides optimum protection.
• The permissible storage temperature is −40 to +80 °C (–40 to +176 °F),
preferably +20 °C (68 °F).
• Do not remove the protective covers or caps on the process connections until you are ready
to install the device.
• The measuring device must be protected against direct sunlight during storage in order to
avoid unacceptably high surface temperatures.
Endress+Hauser 11
3 Installation Proline Promass 40
3.2.1 Dimensions
Dimensions and fitting lengths of the transmitter and the sensor are provided in the separate
documentation "Technical Information" (→ Page 66).
A0003605
12 Endress+Hauser
Proline Promass 40 3 Installation
A0003597
8 3/8" 6 0.25
15 ½" 10 0.40
25 1" 14 0.55
40 1½" 22 0.87
50 2" 28 1.10
80 3" 50 2.00
System pressure
It is important to ensure that cavitation does not occur, because it would influence the oscillation
of the measuring tube. No special measures need to be taken for fluids which have properties similar
to water under normal conditions.
In the case of liquids with a low boiling point (hydrocarbons, solvents, liquefied gases) or in suction
lines, it is important to ensure that pressure does not drop below the vapor pressure and that the
liquid does not start to boil. It is also important to ensure that the gases that occur naturally in many
liquids do not outgas. Such effects can be prevented when system pressure is sufficiently high.
Endress+Hauser 13
3 Installation Proline Promass 40
3.2.3 Orientation
Make sure that the direction of the arrow on the nameplate of the sensor matches the direction of
flow (direction of fluid flow through the pipe).
Vertical (Fig. V)
Recommended orientation with upward direction of flow. When fluid is not flowing, entrained
solids will sink down and gases will rise away from the measuring tube. Thus the measuring tubes
can be completely drained and protected against solids buildup.
Standard,
✔✔ ✔✔ ✔✔
Compact version
In order to ensure that the permissible ambient temperature range for the transmitter (→ Page 62)
is not exceeded, we recommend the following orientations:
• For fluids with very high temperatures we recommend the horizontal orientation with the
transmitter head pointing downwards (Fig. H2) or the vertical orientation (Fig. V).
• For fluids with very low temperatures, we recommend the horizontal orientation with the
transmitter head pointing upwards (Fig. H1) or the vertical orientation (Fig. V).
" Caution!
When using a bent measuring tube and horizontal installation, the position of the sensor has to be
matched to the fluid properties.
1 2
A0004581
14 Endress+Hauser
Proline Promass 40 3 Installation
! Note!
• Do not use any heating elements with thyristor controlled voltage sources.
• When using electrical heat tracing whose heat is regulated using phase control or by pulse packs,
it cannot be ruled out that the measured values are influenced by magnetic fields which may
occur, i.e. at values greater than those permitted by the EC standard (Sinus 30 A/m). In such
cases, the sensor must be magnetically screened.
The secondary containment can be shielded with tin plates or electric sheets without privileged
direction (e.g. V330-35A) with the following properties:
– Relative magnetic permeability μr ≥ 300
– Plate thickness d ≥ 0.35 mm (d ≥ 0,014")
" Caution!
Risk of electronics overheating!
• Make sure that the connector between sensor and transmitter always remains free of insulating
material.
• Bear in mind that a certain orientation might be required, depending on the temperature of the
fluid → Page 12
• Information on permissible temperature ranges → Page 62
3.2.6 Vibrations
The high oscillation frequency of the measuring tubes ensures that the correct operation of the
measuring system is not influenced by pipe vibrations. Consequently, the sensors require no special
measures for attachment.
Endress+Hauser 15
3 Installation Proline Promass 40
# Warning!
The turning mechanism in devices with EEx d/de or FM/CSA Cl. I Div. 1 classification is not the
same as that described here. The procedure for turning these housings is described in the Ex-specific
documentation.
2 4
1 3
6
A0004302
16 Endress+Hauser
Proline Promass 40 3 Installation
4 x 45°
A0003236
Does the device correspond to specifications at the measuring point, including process tempera- see Page 56 ff.
ture and pressure, ambient temperature, measuring range, etc.?
Installation Notes
Does the arrow on the sensor nameplate match the direction of flow through the pipe? −
Are the measuring point number and labeling correct (visual inspection)? –
Is the orientation chosen for the sensor correct, in other words suitable for sensor type, fluid see Page 12 ff.
properties (outgassing, with entrained solids) and fluid temperature?
Endress+Hauser 17
3 Installation Proline Promass 40
18 Endress+Hauser
Proline Promass 40 4 Wiring
4 Wiring
# Warning!
When connecting Ex-certified devices, see the notes and diagrams in the Ex-specific supplement to
this Operating Instruction. Please do not hesitate to contact your Endress+Hauser representative if
you have any questions.
! Note!
The device does not have an internal power switch. For this reason, assign the device a switch or
power-circuit breaker which can be used to disconnect the power supply line from the power grid.
# Warning!
• Risk of electric shock. Switch off the power supply before opening the device. Do not install or
wire the device while it is connected to the power supply. Failure to comply with this precaution
can result in irreparable damage to the electronics.
• Risk of electric shock. Connect the protective conductor to the ground terminal on the housing
before the power supply is applied (not necessary if the power supply is galvanically isolated).
• Compare the specifications on the nameplate with the local voltage supply and frequency.
The national regulations governing the installation of electrical equipment also apply.
1. Remove the cover of the connection compartment (f) from the transmitter housing.
2. Feed the power supply cable (a) and signal cables (b) through the appropriate cable entries.
3. Connect the cables:
– Wiring diagramm → Fig. 10
– Terminal assignment → Page 20
4. Screw the cover of the connection compartment (f) firmly onto the transmitter housing.
f
– 27
+ 26
– 25 b
b e + 24
– 23 d
+ 22
– 21
+ 20
a
N (L-) 2 c
L1 (L+) 1
g a
A0004582
Fig. 10: Connecting the transmitter (aluminium field housing); Cable cross-section: max. 2.5 mm2
a Cable for power supply: 85…260 V AC, 20…55 V AC, 16…62 V DC
Terminal No. 1: L1 for AC, L+ for DC
Terminal No. 2: N for AC, L− for DC
b Signal cable: Terminals Nos. 20–27 → Page 19
c Ground terminal for protective conductor
d Ground terminal for signal cable shield
e Service adapter for connecting service interface FXA 193 (Fieldcheck, FieldCare)
f Cover of the connection compartment
g Securing clamp
Endress+Hauser 19
4 Wiring Proline Promass 40
Order variant 20 (+) / 21 (–) 22 (+) / 23 (–) 24 (+) / 25 (–) 26 (+) / 27 (–)
! Note!
• The measuring loop’s minimum load must be at least 250 Ω.
• The CURRENT SPAN function must be set to "4–20 mA" (individual options see device function).
• See also the documentation issued by the HART Communication Foundation, and in particular
HCF LIT 20: "HART, a technical summary".
³ 250 W
-27
+26
2
4 3
A0004586
20 Endress+Hauser
Proline Promass 40 4 Wiring
³ 250 W
–27
+26
4 2
3
5
A0004592
Compliance with the following points is mandatory following installation in the field or servicing,
in order to ensure that IP 67 protection is maintained:
• The housing seals must be clean and undamaged when inserted into their grooves.
The seals must be dried, cleaned or replaced if necessary.
• The threaded fasteners and screw covers must be firmly tightened.
• The cables used for connection must be of the specified outside diameter
→ Page 58, cable entries.
• The cable entries must be firmly tighten (point a → Fig. 13).
• The cable must loop down in front of the cable entry ("water trap") (point b → Fig. 13).
This arrangement prevents moisture penetrating the entry.
! Note!
The cable entries may not be point up.
a b
a0001914
" Caution!
Do not loosen the screws of the sensor housing, as otherwise the degree of protection guaranteed
by Endress+Hauser no longer applies.
Endress+Hauser 21
4 Wiring Proline Promass 40
Does the supply voltage match the specifications on the nameplate? 85…260 V AC (45…65 Hz)
20…55 V AC (45…65 Hz)
16…62 V DC
Are the power supply and signal cables correctly connected? See the wiring diagram inside the
cover of the terminal compartment
Are all cable entries installed, firmly tightened and correctly sealed? see Page 21
Cables looped as "water traps"?
22 Endress+Hauser
Proline Promass 40 5 Operation
5 Operation
3
+48.25 m /h
3
+3702.6 m
A0003802
You can change the assignment of display lines to different variables to suit your needs and
preferences by means of the HART interface or using the "FieldCare" program (→ see "Description
of Device Functions" manual).
! Note!
• The transmitter continues to measure while data entry is in progress, i.e. the current measured
values are output via the signal outputs in the normal way.
• If the power supply fails all preset and parameterized values remain safely stored in the EEPROM.
" Caution!
• All functions are described in detail, as is the function matrix itself, in the "Description of
Device Functions" manual, which is a separate part of this Operating Instruction.
• Changing certain parameters such as all sensor characteristics, for example, influences numerous
functions of the entire measuring system, particularly measuring accuracy.There is no need to
change these parameters under normal circumstances and consequently, they are protected by a
special code known only to the Endress+Hauser service organization. Please contact
Endress+Hauser if you have any questions.
Endress+Hauser 23
5 Operation Proline Promass 40
1 P XXXXXXXXXX
#000 00:00:05
2 4 5 3
A0000991
Fault message ( $)
• Displayed as → Lightning flash ( $), error type (S: system error, P: process error).
• The error in question has a direct effect on the inputs or outputs.
The response of the inputs or outputs (failsafe mode) can be defined by means of functions in the
function matrix (see Page 49).
! Note!
• For security reasons, error messages should be output via the status output.
• When an error message is present, an upper or lower failure signal level can be output in
accordance with NAMUR NE 43 via the current output.
24 Endress+Hauser
Proline Promass 40 5 Operation
Universal commands:
All HART device support and use universal commands.
The following functionalities are linked to them:
• Recognizing HART devices
• Reading digital measured values (mass flow, totalizer, etc.)
Device-specific commands:
These commands allow access to device-specific functions which are not HART standard. Such
commands access individual field device information, amongst other things, such as calibration
values, creepage settings, etc.
! Note!
Promass 40 has access to all three command classes. On Page 27, you will find a list with all the
supported "Universal Commands" and "Common Practice Commands".
! Note!
The HART protocol requires the "4–20 mA setting in the CURRENT SPAN function (individual
options see "Description of Device functions" Manual).
Endress+Hauser 25
5 Operation Proline Promass 40
HART protocol:
Fieldcheck • Update by means of FieldCare via flow device FXA 193/291 DTM in Fieldflash
Module
0 OFF (unassigned)
2 Mass flow
5 Volume flow
250 Totalizer 1
Process variables:
At the factory, the process variables are assigned to the following device variables:
• Primary process variable (PV) → Mass flow
• Second process variable (SV) → Totalizer 1
• Third process variable (TV) → Volume flow
• Fourth process variable (FV) → Corrected volume flow
! Note!
You can set or change the assignment of device variables to process variables
using Command 51 → Page 31
26 Endress+Hauser
Proline Promass 40 5 Operation
Universal Commands
0 Read unique device identifier none Device identification delivers information on the device and
the manufacturer. It cannot be changed.
Access type = read
The response consists of a 12-byte-device ID:
– Byte 0: fixed value 254
– Byte 1: Manufacturer ID, 17 = Endress+Hauser
– Byte 2: Device type ID, e.g. 83 = Promass 40
– Byte 3: Number of preambles
– Byte 4: Universal commands rev. no.
– Byte 5: Device-specific commands rev. no.
– Byte 6: Software revision
– Byte 7: Hardware revision
– Byte 8: Additional device information
– Byte 9-11: Device identification
1 Read primary process variable none – Byte 0: HART unit code of the primary process variable
– Bytes 1-4: Primary process variable
Access type = read
Factory setting:
Primary process variable = Mass flow
! Note!
• You can set the assignment of device variables to process
variables using Command 51.
• Manufacturer specific units are represented using the HART
unit code "240".
2 Read the primary process variable as none – Bytes 0-3: Actual current of the primary process variable
current in mA and percentage of the set in mA
measuring range – Bytes 4-7: Percentage of the set measuring range
! Note!
You can set the assignment of device variables to process
variables using Command 51.
3 Read the primary process variable as cur- none 24 bytes are sent as a response:
rent in mA and four (preset using Com- – Bytes 0-3: Primary process variable current in mA
mand 51) dynamic process variables – Byte 4: HART unit code of the primary process variable
– Bytes 5-8: Primary process variable
Access type = read – Byte 9: HART unit code of the second process variable
– Bytes 10-13: Second process variable
– Byte 14: HART unit code of the third process variable
– Bytes 15-18: Third process variable
– Byte 19: HART unit code of the fourth process variable
– Bytes 20-23: Fourth process variable
Factory setting:
• Primary process variable = Mass flow
• Second process variable = Totalizer 1
• Third process variable = Volume flow
• Fourth process variable = Corrected volume flow
! Note!
• You can set the assignment of device variables to process
variables using Command 51.
• Manufacturer specific units are represented using the HART
unit code "240".
Endress+Hauser 27
5 Operation Proline Promass 40
Universal Commands
6 Set HART short form address Byte 0: desired address (0…15) Byte 0: active address
Factory setting:
Access type = write 0
! Note!
With an address >0 (multidrop mode), the
current output of the primary process
variable is set to 4 mA.
11 Read unique device identification using the Bytes 0-5: TAG Device identification delivers information on the device and
TAG (measuring point designation) the manufacturer. It cannot be changed.
The response consists of a 12-byte-device ID if the given TAG
Access type = read agrees with the one saved in the device:
! Note!
You can write the TAG, descriptor and date using
Command 18.
14 Read sensor information on primary none – Bytes 0-2: Sensor serial number
process variable – Byte 3: HART unit code of sensor limits and measuring
range of the primary process variable
– Bytes 4-7: Upper sensor limit
– Bytes 8-11: Lower sensor limit
– Bytes 12-15: Minimum span
! Note!
• The data relate to the primary process variable
(= Mass flow).
• Manufacturer specific units are represented using the HART
unit code "240".
28 Endress+Hauser
Proline Promass 40 5 Operation
Universal Commands
Factory setting:
Primary process variable = Mass flow
! Note!
• You can set the assignment of device variables to process
variables using Command 51.
• Manufacturer specific units are represented using the HART
unit code "240".
16 Read the device production number none Bytes 0-2: Production number
17 Write user message You can save any 32-character long text in Displays the current user message in the device:
the device under this parameter:
Access = write Bytes 0-23: Current user message in the device
Bytes 0-23: Desired user message
18 Write TAG, descriptor and date With this parameter, you can store an 8 Displays the current information in the device:
character TAG, a 16 character descriptor
Access = write and a date: – Bytes 0-5: TAG
– Bytes 6-17: Descriptor
– Bytes 0-5: TAG – Byte 18-20: Date
– Bytes 6-17: Descriptor
– Byte 18-20: Date
Endress+Hauser 29
5 Operation Proline Promass 40
The following table contains all the common practice commands supported by the device.
34 Write damping value for primary process Bytes 0-3: Damping value of the primary Displays the current damping value in the device:
variable process variable in seconds
Bytes 0-3: Damping value in seconds
Access = write Factory setting:
Primary process variable = Mass flow
35 Write measuring range of primary process Write the desired measuring range: The currently set measuring range is displayed as a response:
variable – Byte 0: HART unit code of the primary
process variable – Byte 0: HART unit code for the set measuring range of the
Access = write – Bytes 1-4: Upper range, value for 20 mA primary process variable
– Bytes 5-8: Start of measuring range, – Bytes 1-4: Upper range, value for 20 mA
value for 4 mA – Bytes 5-8: Start of measuring range, value for 4 mA
Factory setting:
Primary process variable = Mass flow
! Note!
Manufacturer-specific units are represented using the HART
unit code "240".
! Note!
• You can set the assignment of device
variables to process variables using
Command 51.
• If the HART unit code is not the correct
one for the process variable, the device
will continue with the last valid unit.
Access = write
40 Simulate output current of primary process Simulation of the desired output current of The momentary output current of the primary process variable
variable the primary process variable. is displayed as a response:
Access = write An entry value of 0 exits the simulation Byte 0-3: Output current in mA
mode:
Byte 0-3: Output current in mA
Factory setting:
Primary process variable = Mass flow
! Note!
You can set the assignment of device
variables to process variables with
Command 51.
Access = write
30 Endress+Hauser
Proline Promass 40 5 Operation
44 Write unit of primary process variable Set unit of primary process variable. Only The current unit code of the primary process variable is
unit which are suitable for the process vari- displayed as a response:
Access = write able are transferred to the device:
Byte 0: HART unit code
Byte 0: HART unit code
Factory setting:
! Note!
Manufacturer-specific units are represented using the HART
Primary process variable = Mass flow
unit code "240".
! Note!
• If the written HART unit code is not the
correct one for the process variable, the
device will continue with the last valid
unit.
• If you change the unit of the primary
process variable, this has no impact on
the system units.
48 Read additional device status none The device status is displayed in extended form as the
response:
Access = read
Coding: see table on Page 33
50 Read assignment of the device variables to none Display of the current variable assignment of the process
the four process variables variables:
– Byte 0: Device variable code to the primary process variable
Access = read – Byte 1: Device variable code to the second process variable
– Byte 2: Device variable code to the third process variable
– Byte 3: Device variable code to the fourth process variable
Factory setting:
• Primary process variable: Code 1 for mass flow
• Second process variable: Code 250 for totalizer 1
• Third process variable: Code 5 for volume flow
• Fourth process variable: Code 6 for corrected volume flow
! Note!
You can set the assignment of device variables to process
variables with Command 51.
51 Write assignments of the device variables Setting of the device variables to the four The variable assignment of the process variables is displayed as
to the four process variables process variables: a response:
– Byte 0: Device variable code to the – Byte 0: Device variable code to the primary process variable
Access = write primary process variable – Byte 1: Device variable code to the second process variable
– Byte 1: Device variable code to the – Byte 2: Device variable code to the third process variable
second process variable – Byte 3: Device variable code to the fourth process variable
– Byte 2: Device variable code to the third
process variable
– Byte 3: Device variable code to the
fourth process variable
Factory setting:
• Primary process variable = Mass flow
• Second process variable = Totalizer 1
• Third process variable = Volume flow
• Fourth process variable = Corrected
volume flow
! Note!
The totalizer cannot be assigned as the
primary process variable.
Endress+Hauser 31
5 Operation Proline Promass 40
53 Write device variable unit This command sets the unit of the given The current unit of the device variables is displayed in the
device variables. Only those units which device as a response:
Access = write suit the device variable are transferred:
– Byte 0: Device variable code – Byte 0: Device variable code
– Byte 1: HART unit code – Byte 1: HART unit code
59 Write number of preambles in response This parameter sets the number of pream- As a response, the current number of the preambles is dis-
message bles which are inserted in the response played in the response message:
messages:
Access = write Byte 0: Number of preambles
Byte 0: Number of preambles (2…20)
32 Endress+Hauser
Proline Promass 40 5 Operation
! Note!
You can find a detailed explanation of the device status and error messages and their elimination in
the "System error messages" section → Page 44 ff.
Endress+Hauser 33
5 Operation Proline Promass 40
5 705 The electronics' measuring range will be exceeded. The mass flow is too
high.
6…7 not assigned –
0…4 not assigned –
20 5 731 The zero point adjustment is not possible or has been cancelled.
6…7 not assigned –
34 Endress+Hauser
Proline Promass 40 6 Commissioning
6 Commissioning
PROMASS 40
Start-up message
START-UP
DEVICE SOFTWARE
Current software version (example)
XX.XX.XX
SYSTEM OK
Beginning of normal measuring mode
→ OPERATION
! Note!
If start-up fails, an error message indicating the cause is displayed.
Endress+Hauser 35
6 Commissioning Proline Promass 40
6.3 Configuration
# Warning!
Risk of electric shock!
Exposed components carry dangerous voltages. Make sure that the power supply is switched off
before you remove the cover of the electronics compartment.
1. Switch off power supply.
2. Remove the I/O board → see Page 52 ff.
3. Set the jumpers in accordance with Fig. 16.
" Caution!
Risk of destroying the measuring device. Set the jumpers exactly as shown in Fig. 16.
Incorrectly set jumpers can cause overcurrents that would destroy either the measuring device
or external devices connected to it.
A0001044
36 Endress+Hauser
Proline Promass 40 6 Commissioning
6.4 Adjustment
Experience shows that the zero point adjustment is advisable only in special cases:
• to achieve higest measuring accuracy at very low flow rates
• under extreme process or operating conditions (e.g. very high process pressures or very
high viscosity fluids).
" Caution!
• If the fluid is very difficult to measure (e.g. containing entrained solids or gas) it may prove impos-
sible to obtain a stable zero point despite repeated zero point adjustments. In instances of this
nature, please contact your Endress+Hauser service center.
• You can view the currently valid zero point value using the "ZERO POINT" function (see the
"Description of Device Functions" manual).
A0003601
Endress+Hauser 37
6 Commissioning Proline Promass 40
Procedure
Select "START"
Zero adjustment now starts. While zero adjustment is in progress, the "Zero adjustment running" message is visible for
30…60 seconds.
If the flow of fluid in the pipe exceeds 0.1 m/s, an error message appears on the display: "A: ZERO ADJUST NOT
POSSIBLE"
38 Endress+Hauser
Proline Promass 40 6 Commissioning
" Caution!
• On-site density adjustment can be performed only if the user has detailed knowledge of the fluid
density, obtained for example from detailed laboratory analyses.
• The target density value specified in this way must not deviate from the fluid density measured
in the device by more than ±10%.
• An error in defining the target density affects all calculated volume functions.
• Density adjustment changes the factory density calibration values or the calibration values set by
the service technician.
The functions outlined in the instructions below are described in detail in the "Description of Device
Functions" manual.
1. Fill the sensor with fluid. Make sure that the measuring tubes are completely filled and that
liquids are free of gas bubbles.
2. Wait until the temperature difference between fluid and measuring tube has equalized. The
time you have to wait for equalization depends on the fluid and the temperature level.
3. Select the density adjustment function within the function matrix:
PROCESSPARAMETER → DENSITY SET VALUE
Enter the fluid's target density and save this value. Input range = actual density value ±10%
4. Select the "MEASURE FLUID" function.
Select the "START" setting. A "MEASURING DENSITY" message appears for approx. 10 sec-
onds on the display. During this time the current density of the fluid (measured density value)
is measured.
5. Select the "DENSITY ADJUST" function.
Select the "DENSITY ADJUST" setting. Promass compares the measured density value with the
target density value and calculates the new density coefficient.
" Caution!
If density adjustment does not complete correctly, you can select the "RESTORE ORIGINAL"
function to reactivate the default density coefficient.
Endress+Hauser 39
7 Maintenance Proline Promass 40
# Warning!
• Make sure that the function and operation of the rupture disk is not impeded through the
installation. Triggering overpressure in the housing as stated on the indication label. Take
adequate precautions to ensure that no damage occurs, and risk to human life is ruled out, if
the rupture disk is triggered. Rupture disk: Burst pressure 10 to 15 bar (145 to 217 psi).
• Please note that the housing can no longer assume a secondary containment function if a rupture
disk is used.
• It is not permitted to open the connections or remove the rupture disk.
" Caution!
• Rupture disks can not be combined with separately available heating jacket.
• The existing connection nozzles are not designed for a rinse or pressure monitoring function.
! Note!
• Before commissioning, please remove the transport protection of the rupture disk.
• Please note the indication labels.
7 Maintenance
40 Endress+Hauser
Proline Promass 40 7 Maintenance
8 Accessories
Various accessories, which can be ordered separately from Endress+Hauser, are available for the
transmitter and the sensor. Detailed information on the order code in question can be obtained from
the Endress+Hauser service organization.
Transmitter Transmitter for replacement or for stock. Use the 40XXX − XXXXX * * * * * *
Promass 40 order code to define the following specifications:
– Approvals
– Degree of protection / version
– Cable entries
– Display / power supply / operation
– Software
– Outputs / inputs
Memograph M graphic The Memograph M graphic display recorder provides RSG40 - ************
display recorder information on all the relevant process variables.
Measured values are recorded correctly, limit values
are monitored and measuring points analyzed. The
data are stored in the 256 MB internal memory and
also on a DSD card or USB stick.
Memograph M boasts a modular design, intuitive
operation and a comprehensive security concept.
The ReadWin® 2000 PC software is part of the
standard package and is used for configuring, visual-
izing and archiving the data captured.
The mathematics channels which are optionally
available enable continuous monitoring of specific
power consumption, boiler efficiency and other
parameters which are important for efficient energy
management.
HART Communicator Handheld terminal for remote parameterization and SFX100 - *******
Field Xpert handheld for obtaining measured values via the current output
terminal HART (4 to 20 mA).
Endress+Hauser 41
7 Maintenance Proline Promass 40
42 Endress+Hauser
Proline Promass 40 9 Trouble-shooting
9 Trouble-shooting
No display visible, but output 1. Check whether the ribbon cable connector of the display module is correctly
signals are present. plugged into the amplifier board → Page 53
2. Display module defective → order spare parts → Page 51
3. Measuring electronics defective → order spare parts → Page 51
Measured value indicated, but Measuring electronics defective → order spare parts → Page 51
no signal at the current or pulse
output
Errors which occur during commissioning or measuring operation are displayed immediately.
Error messages consist of a variety of icons. The meanings of these icons are as follows:
" Caution!
Also observe the information → Page 24 ff.
Endress+Hauser 43
9 Trouble-shooting Proline Promass 40
" Caution!
In the event of a serious fault, a flowmeter might have to be returned to the manufacturer for repair.
The procedures on Page 6 must be carried out before you return a flowmeter to Endress+Hauser.
Always enclose a duly completed "Declaration of contamination" form. You will find a preprinted
form at the back of this manual.
! Note!
• The listed error message types below correspond to the factory setting.
• Also observe the informations on → Page 24 ff. and 49.
S = System error
$ = Fault message (with an effect on the outputs)
! = Notice message (without an effect on the outputs)
001 S: CRITICAL FAILURE Serious device error Replace the amplifier board.
$: # 001 Spare parts → Page 51
013 S: AMP SW EEPROM Amplifier: The EEPROM data blocks in which an error has occurred
$: # 012 Error accessing EEPROM data are displayed in the "RESTORE DATA FAILURE" function.
Press Enter to acknowledge the errors in question; default
values are automatically inserted instead of the errored
parameter values.
! Note!
The measuring device has to be restarted if an error has
occurred.
031 S: SENSOR HW DAT Sensor: 1. Check whether the S-DAT is correctly plugged into
$: # 031 1. S-DAT is not plugged into the amplifier board correctly the amplifier board→ Page 53
(or is missing). 2. Replace the S-DAT if it is defective.
032 S: SENSOR SW DAT 2. S-DAT is defective. Spare parts→ Page 51
$: # 032 Before replacing the S-DAT, check that the new,
replacement S-DAT is compatible with the measuring
electronics. Check the:
– Spare part set number
– Hardware revision code
3. Replace measuring electronics boards if necessary.
Spare parts→ Page 51
4. Plug the S-DAT into the amplifier board.
051 S: A / C COMPATIB. The I/O board and the amplifier board are not compatible. Use only compatible modules and boards.
$: # 051 Check the compatibility of the modules used.
Check the:
– Spare part set number
– Hardware revision code
44 Endress+Hauser
Proline Promass 40 9 Trouble-shooting
111 S: CHECKSUM TOTAL Totalizer checksum error 1. Restart the measuring device
$: # 111 2. Replace the amplifier board if necessary.
Spare parts→ Page 51
121 S: A / C COMPATIB. Due to different software versions, I/O board and ampli- Module with lower software version has either to be actu-
!: # 121 fier board are only partially compatible (possibly restricted alized by FieldCare with the required software version or
functionality). the module has to be replaced. Spare parts → Page 51
! Note!
– This message is only listed in the error history.
– Nothing is displayed on the display.
251 S: COMMUNICATION I/O Internal communication fault on the amplifier board. Remove the amplifier board.
$: # 251 Spare parts → Page 51
261 S: COMMUNICATION I/O No data reception between amplifier and I/O board or Check the BUS contacts
$: # 261 faulty internal data transfer.
351 S: CURRENT RANGE Current output: 1. Change the upper or lower limit setting, as applicable.
!: # 351 Flow is out of range. 2. Increase or reduce flow, as applicable.
355 S: FREQUENCY RANGE Frequency output: 1. Change the upper or lower limit setting, as applicable.
!: # 355 Flow is out of range. 2. Increase or reduce flow, as applicable.
359 S: PULSE RANGE Pulse output: 1. Increase the setting for pulse weighting
!: # 359 Pulse output frequency is out of range. 2. Increase the max. pulse frequency, if the totalizer can
handle a higher number of pulses.
3. Reduce flow.
379 S: FREQ. LIM The measuring tube oscillation frequency is outside the Contact your Endress+Hauser service organization.
380 $: # 379 / 380 permitted range.
Causes:
– Damaged measuring tube
– Sensor defective or damaged
381 S: FLUIDTEMP. LIM The temperature sensor on the measuring tube is likely Check the following electrical connections before you
382 $: # 381 / 382 defective. contact your Endress+Hauser service organization:
– Verify that the sensor signal cable connector is correctly
plugged into the amplifier board → Page 53.
385 S: INL.SENS.DEF. One of the measuring tube exciter coils (inlet) is likely Check the following electrical connections before you
$: # 385 defective. contact your Endress+Hauser service organization:
– Verify that the sensor signal cable connector is correctly
plugged into the amplifier board → Page 53.
386 S: OUTL.SENS.DEF. One of the measuring tube exciter coils (outlet) is likely Check the following electrical connections before you
$: # 386 defective. contact your Endress+Hauser service organization:
– Verify that the sensor signal cable connector is correctly
plugged into the amplifier board→ Page 53.
Endress+Hauser 45
9 Trouble-shooting Proline Promass 40
387 S: SEN.ASY.EXCEED Measuring pipe excitation coil is probably faulty. Check the following electrical connections before you
$: # 387 contact your Endress+Hauser service organization:
– Verify that the sensor signal cable connector is correctly
plugged into the amplifier board → Page 53.
388 S: AMP. FAULT Amplifier error Contact your Endress+Hauser service organization.
389 $: # 388 / 389 / 390
390
501 S: SW.-UPDATE ACT. New amplifier or communication (I/O module) software Wait until process is finished. The device will restart
!: # 501 version is loaded. Currently no other functions are automatically.
possible.
502 S: UP-/DOWNLOAD ACT. Up- or downloading the device data via configuration pro- Wait until process is finished.
!: # 502 gram. Currently no other functions are possible.
601 S: POS. ZERO-RET. Positive zero return active. Switch off positive zero return
!: # 601
" Caution!
This message has the highest display priority.
611 S: SIM. CURR. OUT. Simulation current output active Switch off simulation
!: # 611
621 S: SIM. FREQ. OUT Simulation frequency output active Switch off simulation
!: # 621
631 S: SIM. PULSE Simulation pulse output active Switch off simulation
!: # 631
641 S: SIM. STAT. OUT. Simulation status output active Switch off simulation
!: # 641
671 S: SIM. STAT. IN Simulation status input active Switch off simulation
!: # 671
691 S: SIM. FAILSAFE Simulation of response to error (outputs) active Switch off simulation
$: # 691
692 S: SIM. MEASURAND Simulation of measuring variables (e.g. mass flow) Switch off simulation
!: # 692
46 Endress+Hauser
Proline Promass 40 9 Trouble-shooting
! Note!
• The error types listed in the following correspond to the factory settings.
• Also observe the information on → Page 24 ff. and 49
P = Process error
$ = Fault message (with an effect on the outputs)
! = Notice message (without an effect on the outputs)
586 P: OSC. AMP. LIMIT The fluid properties do not allow a continuation of the Change or improve process conditions.
$: # 586 measurement.
Causes:
– Extremely high viscosity
– Process fluid is very inhomogeneous
(gas or solid content)
587 P: TUBE NOT OSC Extreme process conditions exist. The measuring system Change or improve process conditions.
$: # 587 can therefore not be started.
588 P: NOISE LIMIT Overdriving of the internal analog to digital converter. Change or improve process conditions, e.g. by reducing
$: # 588 the flow velocity.
Causes:
– Cavitation
– Extreme pressure pulses
– High gas flow velocity
700 P: EMPTY PIPE The process fluid density is below the lower limit value set 1. Ensure that there is no gas content in the process
!: # 700 in the "EPD" function. liquid.
2. Adapt the values in the "EPD RESPONSE TIME"
Causes: function to the current process conditions.
– Air in the measuring tube
– Partly filled measuring tube
701 P: EXC. CURR. LIM The maximum current value for the measuring tube In particular with outgassing fluids and/or increased gas
!: 701 exciter coils has been reached, since certain process fluid content, the following measures are recommended to
characteristics are extreme, e.g. high gas or solid content. increase system pressure:
The instrument continues to work correctly. 1. Install the device at the outlet side of a pump.
2. Install the device at the lowest point of an ascending
pipeline.
3. Install a flow restriction, e.g. reducer or orifice,
downstream from the instrument.
702 P: FLUID INHOM. Frequency control is not stable, due to inhomogeneous In particular with outgassing fluids and/or entrained gas
!: 702 process fluid, e.g. entrained gas or solid content. content, the following measures are recommended to
increase system pressure:
Endress+Hauser 47
9 Trouble-shooting Proline Promass 40
703 P: NOISE LIMIT Overdriving of the internal analog to digital converter. Change or improve process conditions, e.g. by reducing
704 !: 703 / 704 the flow velocity.
Causes:
– Cavitation
– Extreme pressure pulses
– High gas flow velocity
705 P: FLOW LIMIT The mass flow is too high. The electronics' measuring Reduce flow
$: # 705 range will be exceeded.
731 P: ADJ. ZERO FAIL The zero point adjustment is not possible or has been Make sure that zero point adjustment is carried out at
!: 731 cancelled "zero flow" only (v = 0 m/s) → Page 37
Symptoms Rectification
Remark:
You may have to change or correct certain settings of the function matrix in order to rectify faults.
The functions outlined below, such as DISPLAY DAMPING, for example, are described in detail in the "Description of Device Functions" manual.
Measured value reading fluctuates 1. Check liquids for the presence of gas bubbles.
even though flow is steady.
2. In the "TIME CONSTANT" function (CURRENT OUTPUT) → increase value
Measured value reading shown on 1. Check liquids for the presence of gas bubbles.
display, even though the fluid is at a
standstill and the measuring tube is 2. Activate the "LOW FLOW CUTOFF" function (PROCESS PARAMETER), i.e. enter or increase the value for the
full. switching point.
The fault cannot be rectified or some The following options are available for tackling problems of this nature:
other fault not described above has
occurred. Request the services of an Endress+Hauser service technician
In these instances, please contact your If you contact our service organization to have a service technician sent out, please be ready with the following
Endress+Hauser service organization. information:
– Brief description of the fault
– Nameplate specifications (Page 7 ff.): Order code and serial number
48 Endress+Hauser
Proline Promass 40 9 Trouble-shooting
! Note!
The failsafe mode of totalizers, current, pulse and frequency outputs can be customized by means
of various functions in the function matrix. You will find detailed information on these procedures
in the "Description of Device Functions" manual.
" Caution!
System or process errors defined as "Notice messages" have no effect whatsoever on the inputs and outputs.
See the information on → Page 24 ff.
MAXIMUM CURRENT
The current output will be set to the higher value of
the signal on alarm level depending on the setting
selected in the CURRENT SPAN function (see the
"Description of Device Functions" manual).
HOLD VALUE
Measured value display on the basis of the last saved
value preceding occurrence of the fault.
ACTUAL VALUE
Measured value display on the basis of the current
flow measurement. The fault is ignored.
HOLD VALUE
Last valid value (preceding occurrence of the fault) is
output.
ACTUAL VALUE
Fault is ignored, i.e. normal measured value output
on the basis of ongoing flow measurement.
Endress+Hauser 49
9 Trouble-shooting Proline Promass 40
FAILSAFE LEVEL
Output of the frequency specified in the FAILSAFE
VALUE function.
HOLD VALUE
Last valid value (preceding occurrence of the fault) is
output.
ACTUAL VALUE
Fault is ignored, i.e. normal measured value output
on the basis of ongoing flow measurement.
ACTUAL VALUE
The fault is ignored. The totalizers continue to count
in accordance with the current flow value.
HOLD VALUE
The totalizers continue to count the flow in accord-
ance with the last valid flow value (before the error
occurred).
Status output In the event of a fault or power supply failure: No effect on the status output
Status → non-conductive
50 Endress+Hauser
Proline Promass 40 9 Trouble-shooting
! Note!
You can order spare parts directly from your Endress+Hauser service organization by providing the
serial number printed on the transmitter's nameplate (see Page 7).
A0009764
Endress+Hauser 51
9 Trouble-shooting Proline Promass 40
# Warning!
• Risk of electric shock. Exposed components carry dangerous voltages. Make sure that the power
supply is switched off before you remove the cover of the electronics compartment.
• Risk of damaging electronic components (ESD protection). Static electricity can damage electronic
components or impair their operability. Use a workplace with a grounded working surface pur-
posely built for electrostatically sensitive devices!
• If you cannot guarantee that the dielectric strength of the device is maintained in the following
steps, then an appropriate inspection must be carried out in accordance with the manufacturer’s
specifications.
1. Unscrew cover of the electronics compartment from the transmitter housing (see Fig. 19).
2. Remove the local display (1) as follows:
– Press in the latches (1.1) at the side and remove the display module.
– Disconnect the ribbon cable (1.2) of the display module from the amplifier board.
3. Remove the screws and remove the cover (3) from the electronics compartment.
4. Remove power unit board and I/O board (4, 6):
Insert a thin pin into the hole (3) provided for the purpose and pull the board clear of its holder.
5. Remove amplifier board (5):
– Disconnect the plug of the sensor signal cable (5.1) including S-DAT (5.3) from the board.
– Loosen the plug of the excitation current cable (5.2) and gently disconnect the plug from the
board, i.e. without moving it back and forth.
– Insert a thin pin into the hole (3) provided for the purpose and pull the board clear of its
holder.
6. Installation is the reverse of the removal procedure.
" Caution!
Use only original Endress+Hauser parts.
52 Endress+Hauser
Proline Promass 40 9 Trouble-shooting
6
5.1
3
5.3
3
5.2
2
1.2
1.1
A0009765
Endress+Hauser 53
9 Trouble-shooting Proline Promass 40
# Warning!
Risk of electric shock. Exposed components carry dangerous voltages. Make sure that the power
supply is switched off before you remove the cover of the electronics compartment.
" Caution!
Use only original Endress+Hauser parts.
A0001148
Fig. 20: Replacing the device fuse on the power unit board
1 Protective cap
2 Device fuse
54 Endress+Hauser
Proline Promass 40 9 Trouble-shooting
9.9 Return
→ Page 6
9.10 Disposal
Observe the regulations applicable in your country!
New functionalities:
– Function PULSE WIDTH
– Function CURRENT SPAN
– Function FAILSAFE MODE
! Note!
Usually, an upload or download between the different software versions is only possible with a
special service software.
Endress+Hauser 55
10 Technical data Proline Promass 40
10 Technical data
10.1.1 Applications
The measuring device is for mass flow and volume flow measurement of liquids and gases in sealed
piping systems. Application examples:
• additives
• oils, fats
• acids, alkalis
• lacquers, paints
• suspensions
• gases
10.1.3 Input
Measured variable • Mass flow (proportional to the phase difference between two sensors mounted on the measuring
tube to register a phase shift in the oscillation)
• Volume flow (calculated from mass flow and fluid density, which is proportional to the resonance
frequency of the measuring tubes)
• Fluid temperature (registered by temperature sensors) for compensating temperature effects.
56 Endress+Hauser
Proline Promass 40 10 Technical data
DN
x
[mm] [inch]
8 3/8" 85
15 ½" 110
25 1" 125
40 1½" 125
50 2" 125
80 3" 155
Operable flow range Flowrates above the preset full scale value do not overload the amplifier, i.e. the totalizer values are
registered correctly.
Endress+Hauser 57
10 Technical data Proline Promass 40
10.1.4 Output
Signal on alarm • Current output → failsafe mode selectable (e.g. acc. to NAMUR recommendation NE 43)
• Pulse/frequency output → failsafe mode selectable
• Status output → "non-conductive" in the event of fault or power supply failure
Low flow cutoff Switch points for low flow cut off are selectable
Galvanic isolation All circuits for inputs, outputs, and power supply are galvanically isolated from each other.
Switch-on current:
• max. 13.5 A (< 50 ms) at 24 V DC
• max. 3 A (< 5 ms) at 260 V AC
58 Endress+Hauser
Proline Promass 40 10 Technical data
Maximum measured error The following values refer to the pulse/frequency output.
Measured error at the current output is typically ±5 μA.
o.r. = of reading
Density (liquid)
• ±0.0005 g/cc (under reference conditions)
• ±0.0005 g/cc (after field density calibration under process conditions)
• ±0.02 g/cc (over the entire measuring range of the sensor)
Temperature
±0.5 °C ± 0.005 · T °C
(±1 °F ± 0.003 · (T - 32) °F)
T = medium temperature
Endress+Hauser 59
10 Technical data Proline Promass 40
[%]
±1.0
±0.5
±0.2
0
0 1 2 4 6 8 10 12 14 16 18 t/h
A0012900
Fig. 21: Max. measured error in % of measured value (example: Promass 40E / DN 25)
o.r. = of reading
o.r. = of reading
Density (liquids)
±0.00025 g/cc
1 g/cc = 1 kg/l
Temperature
±0.25 °C ± 0.0025 · T °C
(±0.5 °F ± 0.0015 · (T - 32) °F)
T = medium temperature
60 Endress+Hauser
Proline Promass 40 10 Technical data
Influence of fluid temperature When there is a difference between the temperature for zero point adjustment and the process tem-
perature, the typical measured error of the Promass sensor is ±0.0003% of the full scale value / °C
(±0.0001% of the full scale value / °F).
Influence of fluid pressure The table below shows the effect on accuracy of mass flow due to a difference between calibration
pressure and process pressure.
DN Promass E
8 3/8" no influence
15 ½" no influence
25 1" no influence
40 1½" no influence
50 2" –0.009
80 3" –0.020
o.r. = of reading
o.r. = of reading
Inlet and outlet runs There are no installation requirements regarding inlet and outlet runs.
Endress+Hauser 61
10 Technical data Proline Promass 40
! Note!
• Install the device in a shady location. Avoid direct sunlight, particularly in warm climatic regions.
• At ambient temperatures below –20 °C (-4°F) the readability of the display may be impaired.
Storage temperature –40…+80 °C (–40 to +175 °F), preferably +20 °C (+68 °F)
Fluid pressure limit • Flanges: according to EN (DIN) PN 40…100 / acording to ANSI Cl 150, Cl 300, Cl 600 /
(nominal pressure) JIS 10K, 20K, 40K, 63K
• The Promass E sensor does not have a secondary containment
Rupture disk in the sensor The sensor housing protects the inner electronics and mechanics and is filled with dry nitrogen.
housing (optional) The housing of this sensor does not fulfill any additional secondary containment function.
However, 15 bar (217.5 psi) can be specified as a reference value for the pressure loading capacity.
For increased safety, a version with rupture disk (triggering pressure 10 to 15 bar (145 to 217.5 psi))
can be used, which is available for order as a separate option.
Limiting flow Select nominal diameter by optimizing between required flow range and permissible pressure loss.
See chapter "Measuring range" for a list of maximum possible full scale values → Page 56 ff.
• The minimum recommended full scale value is approx. 1/20 of the maximum full scale value.
• In most applications, 20…50% of the maximum full scale value can be considered ideal.
• Select a lower full scale value for abrasive substances such as fluids with entrained solids
(flow velocity < 1 m/s (< 3 ft/s)).
• For gas measurement the following rules apply:
– Flow velocity in the measuring tubes should not be more than half the sonic velocity
(0.5 Mach).
– The maximum mass flow depends on the density of the gas (formula → Page 57).
62 Endress+Hauser
Proline Promass 40 10 Technical data
Pressure loss (SI units) Pressure loss depends on the properties of the fluid and on the flow. The following formulas can be
used to approximately calculate the pressure loss:
Reynolds number 2 ⋅ m·
Re = --------------------------
π⋅d⋅υ⋅ρ
Re ≥ 2300 1) Δp = K ⋅ υ
0.25
⋅ m·
1.85
⋅ρ
– 0.86
0.25 · 2
Re < 2300 K2 ⋅ υ ⋅m
Δp = K1 ⋅ υ ⋅ m· + ------------------------------------
ρ
1)
To compute the pressure loss for gases, always use the formula for Re ≥ 2300.
DN d [m] K K1 K2
[mbar]
10000
DN 40
DN 25
DN 15 DN 50
DN 8
DN 80
1000
100
10
0.1
0.001 0.01 0.1 1 10 100 1000 [t/h]
A0004606
Endress+Hauser 63
10 Technical data Proline Promass 40
Pressure loss (US units) Pressure loss is dependent on fluid properties nominal diameter. Consult Endress+Hauser for Appli-
cator PC software to determine pressure loss in US units. All important instrument data is contained
in the Applicator software program in order to optimize the design of measuring system.
Design / dimensions The dimensions and lengths of the sensor and transmitter are provided in the separate "Technical
Information" document on the measuring instrument in question. This can be downloaded as a PDF
file from www.endress.com. A list of the "Technical Information" documents available is provided
in the "Documentation" section → Page 66.
Compact version 8 8 10 15 22 31
Weight in US units
DN [inch] 3/8" ½" 1" 1½" 2" 3"
Compact version 18 18 22 33 49 69
64 Endress+Hauser
Proline Promass 40 10 Technical data
Sensor housing
• Acid and alkali-resistant outer surface
• Stainless steel 1.4301/ASTM 304
Process connections
• Stainless steel 1.4404/316L
– Flanges according to EN 1092-1 (DIN 2501) and according to ASME B16.5
– DIN 11864-2 Form A (flat flange with groove)
– Threaded hygienic connection: DIN 11851, SMS 1145, ISO 2853, DIN 11864-1 Form A
– VCO connections
• Stainless steel SUS 316L
– Flanges to JIS B2220
Measuring tubes
Stainless steel EN 1.4539 / ASTM 904L
Seals
Welded process connections without internal seals
Material load diagram The material load diagrams (pressure-temperature diagrams) for the process connections
are provided in the separate "Technical Information" document on the device in question.
This can be downloaded as a PDF file from www.endress.com.
List of the available "Technical Information" documents → Page 66
Process connections • Flanges according to EN 1092-1 (DIN 2501), according to ASME B16.5, JIS B2220,
VCO connections
• Sanitary connections: Tri-Clamp, threaded hygienic connections (DIN 11851, SMS 1145, ISO
2853, DIN 11864-1), DIN 11864-2 Form A (flat flange with groove)
Display element • Liquid crystal display (optional): illuminated, two lines with 16 characters per line
• Selectable display of different measured values and status variables
• At ambient temperatures below –20 °C (–4 °F) the readability of the display may be impaired
Endress+Hauser 65
10 Technical data Proline Promass 40
Ex Approvals Information about currently available Ex versions (ATEX, FM, CSA, IECEx, NEPSI) can be supplied
by your Endress+Hauser Sales Center on request. All explosion protection data are given in a sepa-
rate documentation which is available upon request.
Pressure device approval Flowmeters with a nominal diameter smaller or equal DN 25 are covered by Art. 3(3) of the
European directive 97/23/EC (Pressure Equipment Directive) and are designed according to sound
engineer practice. For larger nominal diameters, optional approvals according to Cat. II/III are
available when required (depends on fluid and process pressure).
CE mark The measuring system complies with the EMC requirements of the "Australian Communications
and Media Authority (ACMA)".
C-Tick mark The measuring system is in conformity with the EMC requirements of the Australian
Communications Authority (ACMA).
EN 61010-1:
Protection Measures for Electrical Equipment for Measurement, Control, Regulation and Laboratory
Procedures.
IEC/EN 61326
"Emission in accordance with requirements for Class A". Electromagnetic compatibility (EMC-
requirements)
NAMUR NE 21:
Electromagnetic compatibility (EMC) of industrial process and laboratory control equipment.
NAMUR NE 43:
Standardization of the signal level for the breakdown information of digital transmitters with analog
output signal.
NAMUR NE 53:
Software of field devices and signal-processing devices with digital electronics
The Endress+Hauser service organization can provide detailed ordering information and informa-
tion on the order codes upon request.
10.1.14 Accessories
10.1.15 Documentation
66 Endress+Hauser
Proline Promass 40 Index
Index
A Error messages
Ambient temperature . . . . . . . . . . . . . . . . . . . . . . . . . . . . 62 display mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
Applications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56 process errors (application errors) . . . . . . . . . . . . . . . . 47
Applicator (selection and configuration software) . . . . . . . 42 system errors (device errors) . . . . . . . . . . . . . . . . . . . . 44
Error response (inputs/outputs) . . . . . . . . . . . . . . . . . . . . 49
B Error types (system and process errors) . . . . . . . . . . . . . . . 24
Boards Ex certification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 66
see Spare parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51 Ex documentation, supplementary . . . . . . . . . . . . . . . . . . . 5
Exchange
C device fuse . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54
Cable entries printed circuit boards . . . . . . . . . . . . . . . . . . . . . . . . . 52
Degree of protection . . . . . . . . . . . . . . . . . . . . . . . . . . 21 Exterior cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
technical data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58
Calibration factor (default) . . . . . . . . . . . . . . . . . . . . . . . . . 8 F
Cleaning Field Xpert . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
exterior cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40 FieldCare . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
Commissioning Fieldcheck (tester and simulator) . . . . . . . . . . . . . . . . . . . . 42
configuring current output (active/passive) . . . . . . . . . 36 Fitting length
density adjustment . . . . . . . . . . . . . . . . . . . . . . . . . . . 39 see Dimensions
zero point adjustment . . . . . . . . . . . . . . . . . . . . . . . . . 37 Fluid temperature ranges . . . . . . . . . . . . . . . . . . . . . . . . . 62
Communication . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25 Frequency output
Connection technical data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58
see electrical connection Function check . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
Current output Function descriptions
configuring for active/passive . . . . . . . . . . . . . . . . . . . 36 see “Description of Device Functions” manual
technical data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58 Fuse, replacing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54
D G
DAT-Speicherbaustein (S-DAT, Messaufnehmer) . . . . . . . 58 Galvanic isolation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58
Declaration of Contamination form . . . . . . . . . . . . . . . . . . 71
Degree of protection . . . . . . . . . . . . . . . . . . . . . . . . . 21, 62 H
Density adjustment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39 HART
Designated use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 command No. . . . . . . . . . . . . . . . . . . . . . . . . . . . 27, 30
Device Description files (DD) . . . . . . . . . . . . . . . . . . . . . . 26 commands (universal or common) . . . . . . . . . . . . . . . . 25
Device designation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 Device Description files (DD) . . . . . . . . . . . . . . . . . . . 26
Device functions device status, error messages . . . . . . . . . . . . . . . . . . . . 33
see “Description of Device Functions” Manual electrical connection . . . . . . . . . . . . . . . . . . . . . . . . . . 20
Dimensions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12 error messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23 operating possibilities . . . . . . . . . . . . . . . . . . . . . . . . . 25
Display, turning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17 Hazardous substances . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Disposal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55 Heating (general notes, insulation, etc.) . . . . . . . . . . . . . . . 15
Documentation, additional . . . . . . . . . . . . . . . . . . . . . . . . 66 HOME position (operating mode) . . . . . . . . . . . . . . . . . . . 23
Down pipes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
I
E Incoming acceptance . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Electrical connection Inlet runs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Commubox FXA 195 . . . . . . . . . . . . . . . . . . . . . . . . . 21 Input signal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57
Degree of protection . . . . . . . . . . . . . . . . . . . . . . . . . . 21 Input variables . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56
HART handheld communicator . . . . . . . . . . . . . . . . . . 20 Inputs/outputs, response to errors . . . . . . . . . . . . . . . . . . 49
post-connection check (checklist) . . . . . . . . . . . . . . . . 22 Installation conditions
transmitter, terminal assignment . . . . . . . . . . . . . . . . . 19 dimensions, fitting lengths . . . . . . . . . . . . . . . . . . . . . . 12
Electronics boards (replacing, installing) . . . . . . . . . . . . . . 52 down pipes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Environment conditions . . . . . . . . . . . . . . . . . . . . . . . . . . 62 inlet and outlet runs . . . . . . . . . . . . . . . . . . . . . . . . . . 15
location . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
system pressure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
vibrations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Insulation for sensor, heating . . . . . . . . . . . . . . . . . . . . . . 15
Endress+Hauser 67
Index Proline Promass 40
L R
Limiting flow Registered trademarks . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
see Measuring range . . . . . . . . . . . . . . . . . . . . . . . . . 62 Remote operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 65
Limiting medium pressure range . . . . . . . . . . . . . . . . . . . 62 Repair . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Load (outputs) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58 Returning devices . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Local display
see Display S
Low flow cutoff . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58 Safety icons . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Safety instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
M Safety of operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40 Sanitary compatibility . . . . . . . . . . . . . . . . . . . . . . . . . . . . 66
Material load curves . . . . . . . . . . . . . . . . . . . . . . . . . . . . 65 Seals
Materials . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 65 fluid temperature ranges . . . . . . . . . . . . . . . . . . . . . . . 62
Measured variables . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56 materials . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 65
Measuring principle . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56 Sensor installation
Measuring range . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56 see Installation conditions
Measuring system . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56 Serial number . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7, 8, 9
Mounting the sensor Shock resistance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 62
see Installation conditions Signal on alarm . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58
Software
N amplifier . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
Nameplate versions (history) . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55
connections (electrical) . . . . . . . . . . . . . . . . . . . . . . . . 9 Spare parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51
sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8 Standards, guidelines . . . . . . . . . . . . . . . . . . . . . . . . . . . . 66
transmitter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 Status input
Nominal pressure technical data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57
see Limiting medium pressure range Status output
Notice message . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24 technical data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58
Storage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
O Supply voltage (power supply) . . . . . . . . . . . . . . . . . . . . . 58
Operable flow range . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57 Switching output (status output) . . . . . . . . . . . . . . . . . . . 58
Operating conditions . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61 System error (definition) . . . . . . . . . . . . . . . . . . . . . . . . . 24
Operation System error messages . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
Device Description files (DD) . . . . . . . . . . . . . . . . . . . 26 System pressure, requirements . . . . . . . . . . . . . . . . . . . . . 61
HART handheld . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
operating possibilities . . . . . . . . . . . . . . . . . . . . . . . . . 25 T
Operational safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 Technical data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56
Ordering code Temperature ranges
accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41 ambient temperature . . . . . . . . . . . . . . . . . . . . . . . . . 62
sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8 fluid temperature . . . . . . . . . . . . . . . . . . . . . . . . . . . . 62
transmitter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 storage temperature . . . . . . . . . . . . . . . . . . . . . . . . . . 62
Ordering information . . . . . . . . . . . . . . . . . . . . . . . . . . . . 66 Transmitter
Outlet runs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15 electrical connection . . . . . . . . . . . . . . . . . . . . . . . . . 19
Output signal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58 turning the transmitter housing . . . . . . . . . . . . . . . . . 16
Output variables . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58 Transporting the sensor . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Trouble-shooting and remedy . . . . . . . . . . . . . . . . . . . . . 43
P
Post-installation check (checklist) . . . . . . . . . . . . . . . . . . 17 U
Power consumption . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58 User interface, display elements . . . . . . . . . . . . . . . . . . . . 23
Power supply (voltage supply) . . . . . . . . . . . . . . . . . . . . . 58
Power supply failure . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58 V
Pressure loss (formulas, pressure loss diagrams) . . . . . . . . 63 Vibration resistance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 62
Printed circuit boards (removing/installing) . . . . . . . . . . . 52 Vibrations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15, 62
Process connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . 65
Process error (definition) . . . . . . . . . . . . . . . . . . . . . . . . . 24
Process error messages . . . . . . . . . . . . . . . . . . . . . . . . . . 47
Process errors without messages . . . . . . . . . . . . . . . . . . . 48
Pulse output
see frequency output
Pumps, mounting location, system pressure . . . . . . . . . . . 13
68 Endress+Hauser
Proline Promass 40 Index
W
Weight . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 65
Wiring
see electrical connection
Z
Zero point adjustment . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
Endress+Hauser 69
Index Proline Promass 40
70 Endress+Hauser
Declaration of Contamination
Erklärung zur Kontamination
Because of legal regulations and for the safety of our employees and operating equipment, we need the "declaration of
contamination", with your signature, before your order can be handled. Please make absolutely sure to include it with
the shipping documents, or - even better - attach it to the outside of the packaging.
Aufgrund der gesetzlichen Vorschriften und zum Schutz unserer Mitarbeiter und Betriebseinrichtungen, benötigen wir die
unterschriebene "Erklärung zur Kontamination", bevor Ihr Auftrag bearbeitet werden kann. Legen Sie diese unbedingt den
Versandpapieren bei oder bringen Sie sie idealerweise außen an der Verpackung an.
harmful/
Medium /concentration Identification flammable toxic corrosive irritant other * harmless
Medium /Konzentration CAS No. entzündlich giftig ätzend gesundheits- sonstiges* unbedenklich
schädlich/
reizend
Process
medium
Medium im
Prozess
Medium for
process cleaning
Medium zur
Prozessreinigung
Returned part
cleaned with
Medium zur
Endreinigung
* explosive; oxidising; dangerous for the environment; biological risk; radioactive
* explosiv; brandfördernd; umweltgefährlich; biogefährlich; radioaktiv
Please tick should one of the above be applicable, include security sheet and, if necessary, special handling instructions.
Zutreffendes ankreuzen; trifft einer der Warnhinweise zu, Sicherheitsdatenblatt und ggf. spezielle Handhabungsvorschriften beilegen.
_____________________________________________ ______________________________________________
(place, date / Ort, Datum) (Company stamp and legally binding signature)
(Firmenstempel und rechtsverbindliche Unterschrift)