Академический Документы
Профессиональный Документы
Культура Документы
Application
Gas Chromatography
March 1997
In LVI, a large volume of sample is The Agilent 6890 Series GC uses a avoiding most of the problems associ-
injected. The bulk of the solvent is standard automatic sampler with ated with hot inlets such as sample
evaporated before transfer of the syringe sizes up to 50 mL. A 50-mL discrimination, liner overload, and
sample to the analytical column and syringe can inject up to 25 mL. Multi- sample decomposition.
the start of the analytical sepa-ration. ple injections can be used with the
There are two primary techniques to PTV inlet when even larger volumes For large volume injections, the PTV
eliminate solvent: PTV and cool on- are required. With the 6890 GC is used in a "solvent vent" or "solvent
column injection with solvent vapor system, delay between injections can elimination" mode. Sample is intro-
exit (COC-SVE). COC-SVE is most be controlled as well as the number duced into the inlet with the inlet
appropriate for clean samples with of injections or the total injection temperature near the boiling point of
volatile (early-eluting) compo-nents volume. Injection parameters are set the solvent and with a relatively high
such as extracts of drinking water. through the Agilent ChemStation. split ratio. The solvent (and
The COC-SVE technique is discussed low-boiling solutes) is vented while
elsewhere.1 A problem with multiple injections is the higher boiling solutes (more than
the increased number of punctures of about 100°C above the solvent boiling
the GC inlet and vial septa. This point) remain and are concentrated
PTV reduces septum life and increases the in the inlet. After a preset time, the
possibility of contamination of split vent is closed and the inlet tem-
LVI with PTV is ideal for trace analy- sample and inlet. Using a "septumless perature increased to transfer the
sis of later eluting solutes (boiling head" for the inlet can eliminate the solutes and any residual solvent to
points approximately 100°C higher inlet problems. Figure 2A shows the column for separation.
than the solvent) and for dirty sam- septum cap extract that contami-
ples. Typical injection volumes for nated the sample after the vial was Because the sample is evaporated
solvent elimination PTV are 25 to pierced 40 times during several from the inlet, nonvolatile sample
100 mL. Injection volumes up to 1 mL multiple-injection experiments. 100% components and degradation prod-
have been demonstrated.2 The large Teflon septa minimize sample conta- ucts remain behind in the inlet, mini-
sample volumes are injected by mina-tion such as this, but once punc- mizing column contamination. There
manual injection, by multiple sample tured, Teflon septa do not reseal. is evidence that inlet contamination
injections from a standard automatic in PTVs influences subsequent injec-
sampler, or by LVI with a variable- The PTV inlet can be considered a tions less than in hot inlets. If conta-
speed injector. temperature-programmable mination becomes an issue, the inlet
split/ splitless inlet with the same liner is easily changed. Thus, PTV is a
An automatic injector is recom- basic configuration. While it can be better choice for dirty samples than
mended for maximum reproducibility. used hot for split and splitless appli- cool-on-column and split/ splitless
A standard automatic sampler making cations, this is not recommended inlet.
repeat injections is more cost effec- because the volume of vaporized
tive than purchasing a variable-speed solvent may exceed the low internal
sampler and requires less solvent for volume of the PTV inlet. PTV is ideal
syringe cleaning. for cold split or splitless applications,
2
For a PTV inlet to work well, inlet Table 2. Advantages and Disadvantages of LVI by Solvent Elimination PTV
temperature must be programmed Advantages Disadvantages
independently from the column oven. Most flexible LVI technique Loss of volatile sample components
The 6890 provides this function. For Good for late-eluting compounds such as pesticides, Possibility of sample decomposition(although less
example, the inlet can be heated with PAHs, etc than with split/splitless)
the split flow off to transfer the Inlet protects column so one can use dirty samples More difficult to use than conventional inlets like
sample to the column before the oven split/splitless
temperature program begins. After
sample transfer, the inlet can be
heated further to bake off contami-
nants with a high split flow to mini-
mize inlet contamination.
Experimental
A 6890 GC with electronic pneumat-
ics control (EPC) was used. A
G1916A automatic liquid sampler
(ALS) with a G1513A controller per-
formed sample injection. An Agilent
ChemStation (version A.04.02) con-
trolled the instruments and acquired
and processed data.
3
Results and Discussion Area
Count
Multiple Injections 12,000 Packed Liner
4000
3000
2000
C12
1000
C10
0
4
PTV
5 X 5 µL Injections
Figure 3 is a chromatogram of an LVI
1. Methamidophos
of pesticides using a PTV inlet. By 5. Chlorothalonil 9. Imazalil
2. Acephate
injecting 25 mL divided into five injec- 3. Dimethoate
6. Chlorpyrifos-Methyl 10. Ethion
7. Chlorpyrifos 11. Phosmet
tions, good response is obtained from 4. Diazinon
8. Thiabendazole 12. Azinphos-methyl
a 0.01-ppm mixture
5
Abundance 9e+05
9e+06
Initial PTV Temp = 20 ˚C
5e+05 Vent Flow = 300 mL/min
7e+06
5e+06 2e+05
500000
0
5 6 7 8 9 10 11 12
1.2e+07
6e+07 Initial PTV Temp = 35 ˚C
9e+06 Vent Flow = 100 mL/min
4e+07 5e+06
2e+06
B
2e+07 8.20 8.60 9.00
5000000
0
5 6 7 8 9 10 11 12
Minutes
Copyright© 2000
Agilent Technologies, Inc.