Вы находитесь на странице: 1из 6

Large Volume Injection for Gas

Chromatography Using a PTV Inlet

Application

Gas Chromatography
March 1997

Authors Introduction Detection limits can be reduced by


lowering system background and
Bill Wilson, Philip L. Wylie, The demand for lower detection interferences. This can be done
and Matthew S. Klee limits is important in environmental, through sample cleanup, such as
Agilent Technologies, Inc. pharmaceutical, food analysis, and florisil column chromatography, SPE,
2850 Centerville Road other gas chromatography (GC) and SPME, and by using selective
Wilmington, DE 19808-1610 applications. This demand has driven detectors that do not respond to the
USA instrument manufacturers to provide background. How-ever, the cleanup
more sensitive instrumentation and requires time, and the need for
Abstract procedures, which has prompted reg- reduced limits has surpassed the sen-
ulators to reduce allowable limits, sitivity of even the best detectors.
Reduced sample detection limits is a and so on in a never-ending cycle.
continuing goal in gas chromatography. Improvements in sample handling, Large volume injection (LVI) is
Large sample injection volume is one sample injection techniques, and another approach to lower detection
detectors have all contributed to the limits. The typical injection volume
possible approach. New inlets and injec-
ability to measure compounds at for capillary column analysis is 0.5 to
tion techniques supporting large volume 2 mL. Agilent 6890 Series and 5890 gas
decreasing levels.
injection (LVI) have been developed in chromatographs (GCs) allow approxi-
recent years. This paper discusses the Concentrating samples is an estab- mately two times the normal injection
uses and limitations of LVI, describes lished approach for increasing volume (up to 5 mL, depending on the
LVI with a programmable temperature method sensitivity. For many environ- solvent) using "pulsed" splitless injec-
vaporizer (PTV) inlet, and reviews the mental analysis, this involves extrac- tion. Injecting still larger volumes
results of LVI analysis of pesticides and tion followed by solvent evaporation, with standard techniques can lead to
straight-chain hydrocarbons. which generates large volumes of contamination of the system, irrepro-
waste solvent and increases sample ducible results, and loss of sample.
Key Words preparation time significantly. Recent With the new LVI technique, good
advances, such as supercritical fluid chromatography can be obtained with
Large volume injection, LVI, pro- extration (SFE), solid-phase extrac- injection volumes of 5 to 500 mL or
grammed temperature vaporizer, tion (SPE), solid-phase microextrac- more.
PTV, gas chromatography, GC, pesti- tion (SPME), and pressurized fluid
cide analysis, hydrocarbon analysis extraction, are making inroads on Table 1 summarizes several common
liquid/liquid extractions, but these ways to lower detection limits.
still involve additional sample prepa-
ration time.
Table 1. Approaches to Lowering Detection Limits
Concentrate the Sample Transfer More Sample into the Use More Sensitive or Selective Decrease System Noise
Column Detectors
Extraction followed by evaporation of Cool on-column (COC) injection Electron capture detector (ECD) Selective detectors
solvent Splitless injection Electrolytic conductivity detector Sample cleanup to reduce interferences
Solid-phase extraction (SPE) Large volume injection (LVI) (ELCD) Use headspace or purge and trap
Solid-phase micro extraction (SPME) Atomic emission detector (AED) sampling
Headspace sampling Selected-ion monitoring mass Decrease column bleed
Purge and trap (P&T) sampling spectrometry (SIM-MS)
Large volume injection (LVI) Nitrogen-phosphorus detector (NPD)
Flame photometric detector (FPD)

In LVI, a large volume of sample is The Agilent 6890 Series GC uses a avoiding most of the problems associ-
injected. The bulk of the solvent is standard automatic sampler with ated with hot inlets such as sample
evaporated before transfer of the syringe sizes up to 50 mL. A 50-mL discrimination, liner overload, and
sample to the analytical column and syringe can inject up to 25 mL. Multi- sample decomposition.
the start of the analytical sepa-ration. ple injections can be used with the
There are two primary techniques to PTV inlet when even larger volumes For large volume injections, the PTV
eliminate solvent: PTV and cool on- are required. With the 6890 GC is used in a "solvent vent" or "solvent
column injection with solvent vapor system, delay between injections can elimination" mode. Sample is intro-
exit (COC-SVE). COC-SVE is most be controlled as well as the number duced into the inlet with the inlet
appropriate for clean samples with of injections or the total injection temperature near the boiling point of
volatile (early-eluting) compo-nents volume. Injection parameters are set the solvent and with a relatively high
such as extracts of drinking water. through the Agilent ChemStation. split ratio. The solvent (and
The COC-SVE technique is discussed low-boiling solutes) is vented while
elsewhere.1 A problem with multiple injections is the higher boiling solutes (more than
the increased number of punctures of about 100°C above the solvent boiling
the GC inlet and vial septa. This point) remain and are concentrated
PTV reduces septum life and increases the in the inlet. After a preset time, the
possibility of contamination of split vent is closed and the inlet tem-
LVI with PTV is ideal for trace analy- sample and inlet. Using a "septumless perature increased to transfer the
sis of later eluting solutes (boiling head" for the inlet can eliminate the solutes and any residual solvent to
points approximately 100°C higher inlet problems. Figure 2A shows the column for separation.
than the solvent) and for dirty sam- septum cap extract that contami-
ples. Typical injection volumes for nated the sample after the vial was Because the sample is evaporated
solvent elimination PTV are 25 to pierced 40 times during several from the inlet, nonvolatile sample
100 mL. Injection volumes up to 1 mL multiple-injection experiments. 100% components and degradation prod-
have been demonstrated.2 The large Teflon septa minimize sample conta- ucts remain behind in the inlet, mini-
sample volumes are injected by mina-tion such as this, but once punc- mizing column contamination. There
manual injection, by multiple sample tured, Teflon septa do not reseal. is evidence that inlet contamination
injections from a standard automatic in PTVs influences subsequent injec-
sampler, or by LVI with a variable- The PTV inlet can be considered a tions less than in hot inlets. If conta-
speed injector. temperature-programmable mination becomes an issue, the inlet
split/ splitless inlet with the same liner is easily changed. Thus, PTV is a
An automatic injector is recom- basic configuration. While it can be better choice for dirty samples than
mended for maximum reproducibility. used hot for split and splitless appli- cool-on-column and split/ splitless
A standard automatic sampler making cations, this is not recommended inlet.
repeat injections is more cost effec- because the volume of vaporized
tive than purchasing a variable-speed solvent may exceed the low internal
sampler and requires less solvent for volume of the PTV inlet. PTV is ideal
syringe cleaning. for cold split or splitless applications,

2
For a PTV inlet to work well, inlet Table 2. Advantages and Disadvantages of LVI by Solvent Elimination PTV
temperature must be programmed Advantages Disadvantages
independently from the column oven. Most flexible LVI technique Loss of volatile sample components
The 6890 provides this function. For Good for late-eluting compounds such as pesticides, Possibility of sample decomposition(although less
example, the inlet can be heated with PAHs, etc than with split/splitless)
the split flow off to transfer the Inlet protects column so one can use dirty samples More difficult to use than conventional inlets like
sample to the column before the oven split/splitless
temperature program begins. After
sample transfer, the inlet can be
heated further to bake off contami-
nants with a high split flow to mini-
mize inlet contamination.

LVI by PTV is not a good choice when


the target compounds include highly Table 3. Software, Hardware, and Firmware Versions that Support LVI-PTV
volatile species because these low- Item Software/Firmware
boiling compounds are vented along ChemStation software A.04.02 or higher
with the solvent. The lowest boiling G1513A injector A.09.10
target compound should boil at least G1512A ALS A.01.08
100°C above the solvent to have a rea- 6890 GC A.02.01 or higher
sonable chance of success.

Table 2 lists the advantages and dis-


advantages of LVI by solvent elimina-
tion PTV.

Experimental
A 6890 GC with electronic pneumat-
ics control (EPC) was used. A
G1916A automatic liquid sampler
(ALS) with a G1513A controller per-
formed sample injection. An Agilent
ChemStation (version A.04.02) con-
trolled the instruments and acquired
and processed data.

Table 3 lists the hardware and soft-


ware revisions that support multiple
injections, postdwell time, and the
slow plunger mode required for
LVI-PTV.

Experimental conditions for the


GC methods are given with the
chromatograms.

3
Results and Discussion Area
Count
Multiple Injections 12,000 Packed Liner

Multiple injections are a 10,000


straight-forward and reliable way to
introduce large sample volumes into Response Factor RSD
8,000
the inlet. Figure 1 shows the linearity C23: 2.1%
obtained from two sets of multiple C22: 2.6%
injections using a standard 6,000
G1513 ALS.
4,000
Depending on solvent type and injec-
tion volume, liquid sample may run
2,000
down the liner and enter the column.
If this occurs, the column may over-
load with solvent causing cata- 0
0 50 100 150 200 250
strophic peak splitting and possible Total Injection Volume, µL
damage to the stationary phase. To
minimize this possibility, a packed Figure 1. Linearity of multiple injections
liner should be used with multiple
pA
injections of more than 5 mL.
Inlet = 40 ˚C at injection
2500 A Open baffle liner
In addition to preventing solvent from
50 mL/min purge for 2.5 min
flowing into the column, glass wool
30 m x 0.32 mm x 0.25 µm HP-5
or other packing provides a surface 2000
C18
to retain a film of solvent which, in
turn, helps retain early-eluting com-
1500
pounds. Figure 2A shows loss of ana-
C16
lytes eluting before C18. Figure 2B,
Septum extract
with glass wool packing in the liner, 1000
C14 (vial pierced 40 times)
shows complete recovery down to
C14.
500
C12
C10
0

0 2.5 5 7.5 10 12.5 15 17.5 20 22.5


Minutes
pA
B C14
6000
C16
C18 Glass wool packed liner
5000
10 x 25 µL injections

4000

3000

2000
C12
1000
C10
0

0 2.5 5 7.5 10 12.5 15 17.5 20 22.5


Minutes

Figure 2. Comparison of packed and open baffle liners using PTV


sample: C10-C44 in Hexane

4
PTV
5 X 5 µL Injections
Figure 3 is a chromatogram of an LVI
1. Methamidophos
of pesticides using a PTV inlet. By 5. Chlorothalonil 9. Imazalil
2. Acephate
injecting 25 mL divided into five injec- 3. Dimethoate
6. Chlorpyrifos-Methyl 10. Ethion
7. Chlorpyrifos 11. Phosmet
tions, good response is obtained from 4. Diazinon
8. Thiabendazole 12. Azinphos-methyl
a 0.01-ppm mixture

COC-SVE is usually the preferred


technique for samples with 4 5 6 7 10 11
8 12
low-boiling components, because 2
1 3 9
volatile compounds evaporate with
the solvent (as shown in Figure 2A)
with PTV. However, for volatile sam- 0
ples that are too dirty for COC-SVE 5 6 7 8 9 10 11 12
sampling, the PTV inlet can be Minutes
cryocooled below ambient tempera- Figure 3. LVI with solvent elimination PTV
ture resulting in much better recovery Pesticides: 0.01 ppm; PTV conditions: vent flow, 300 mL/min;vent pressure: 0 until 1 min;
of the low boilers. Figure 4 compares purge flow: 50 mL/min at 3.50 min; gassaver on at 4.70 min; PTV initial temperature: 20 °C;
the recovery of C10 using PTV with- PTV initial time:1.1 min; PTV rate: 700 °C/min; PTV final temperature: 300 °C; injection delay:
the inlet temperature at 40°C and 0.00 min; column: 30 m x 0.25 mm x 0.25 mm HP-5MS
-10°C during the injection step. There
is 100% recovery of C10 with
cryocooling. Much less solvent was
eliminated at the lower temperature,
pA
which helped retain the early-eluting
compounds. C23
3000 PTV = 40 ˚C
100 % Recovery
It is important to determine carefully
the best injection parameters for LVI 2000
C20
methods. Figure 5 shows the improve- C16
ment in peak shape obtained for a 1000 Hexane C14
sample containing pesticides when Solvent C10 C12
the initial PTV temperature, vent 0
flow, and injection delay are opti- 2 4 6 8 10 12
0
mized. Inlet temperature and vent pA C10
flow influence the speed and extent
C14
of solvent removal. The chro- 3000 C23
matogram in figure 5B indicates that PTV = 10 ˚C
insufficient solvent was vented, 2000 C12 C16
C20
thereby increasing the chance of
carry-over,degrading peak shape, and- 1000
increasing ghost peaks.
0
0 2 4 6 8 10 12
Conclusion Minutes
Figure 4. PTV with cryocooled inlet
GC using LVI is useful for lowering
detection limits. It has broad applica-
PTV use. LVI with solvent elimination UsingCOC-SVE," Agilent
bility for applications that require
PTV requires careful method develop- Technologies, Application Note
more sensitivity than can be obtained
ment for maximum accuracy and 228-377, Publication Number
with standard injection volumes. PTV
reproducibility. (23)5965-7923E, March 1997.
inlets are appropriate for LVI of late-
eluting samples, for dirty samples, 2. J. Staniewski and J. Rijks, HRC,
and for any cold split/splitless injec-
References
16(1993)182.
tion. Total automation of tempera- 1. B. Wilson, et al., "Large Volume
ture, pressure, and flow simplifies Injection for Gas Chromatography

5
Abundance 9e+05
9e+06
Initial PTV Temp = 20 ˚C
5e+05 Vent Flow = 300 mL/min
7e+06

5e+06 2e+05

7.60 8.00 8.40 8.70


3e+06
A

500000
0
5 6 7 8 9 10 11 12

1.2e+07
6e+07 Initial PTV Temp = 35 ˚C
9e+06 Vent Flow = 100 mL/min

4e+07 5e+06

2e+06
B
2e+07 8.20 8.60 9.00

5000000
0
5 6 7 8 9 10 11 12
Minutes

Figure 5. Effect of PTV setpoints on peak shape


sample: pesticides at 1.0 ppm (peaks identified in figure 3); column: 30 m x 0.25 mm x 0.25 mm HP-5MS

Agilent shall not be liable for errors contained herein or for


incidental or consequential damages in connection with the
furnishing, performance, or use of this material.

Information, descriptions, and specifications in this publication


are subject to change without notice.

Copyright© 2000
Agilent Technologies, Inc.

Printed in the USA 2/2000


5965-7770E

Вам также может понравиться