Вы находитесь на странице: 1из 12

Fuji Medical Dry Laser Imager

DRYPIX Smart
Product Specifications

1st Edition
December 2012

Z30N937
Fuji Medical Dry Laser Imager DRYPIX Smart Product Specifications

Contents
1. Overview ......................................................................................................................................2
2. Features .......................................................................................................................................2
3. System Components ....................................................................................................................3
3.1 Standard Components.........................................................................................................3
3.2 Options ...............................................................................................................................3
3.3 Supplies ..............................................................................................................................3
4. Specifications ...............................................................................................................................4
4.1 Image Recording System ....................................................................................................4
4.2 Processing Capacity............................................................................................................4
4.3 Time Required for Output of 1st Film ................................................................................4
4.4 Time Required for Startup ...................................................................................................4
4.5 Pixel Size ............................................................................................................................4
4.6 Maximum Number of Recording Pixels ...............................................................................4
4.7 Gray Scale ..........................................................................................................................4
4.8 Maximum Density................................................................................................................5
4.9 Number of Input I/F .............................................................................................................5
4.10 Film Supply .........................................................................................................................5
4.11 Density Calibration Function ...............................................................................................5
4.12.1 Connectable Devices (standard) ....................................................................................5
4.12.2 Format ...........................................................................................................................5
4.12.3 Gradation Conversion ....................................................................................................6
4.12.4 Interpolation Mode .........................................................................................................6
4.12.5 Image Frame .................................................................................................................6
4.12.6 Background....................................................................................................................6
4.12.7 DICOM Connection Specifications .................................................................................6
5. Operation Overview......................................................................................................................6
6. Power Supply Conditions .............................................................................................................7
7. Environmental Conditions.............................................................................................................8
8. Footprint for Installation / Space Required for Maintenance .........................................................9
9. External Dimensions and Mass ..................................................................................................10

1
Z30N0937
Fuji Medical Dry Laser Imager DRYPIX Smart Product Specifications

1. Overview The Fuji Medical Dry Laser Imager DRYPIX Smart is the equipment
that outputs digital images received from FCR/FDR image readers or
CT, MRI and other imaging modalities onto Fuji Medical Dry Imaging
Film.

2. Features (1) Fast process at approx. in 80 films/hour (14”x17”) and approx. in


100 films/hour in 10” x 14”.

(2) High image quality is attained


When DI-HL is used, Dmax 3.0.
When DI-ML is used, Dmax 4.0.

(3) It is equipped with maximum of 2ch, and 2 types from


14”x17”, ,14”x14”, B4, 8”x10” and 10”x12” sizes of the film can be
selected and set.

(4) Compact and space saving design, and full daylight processing

(5) Network communication protocol: DICOM compliant

(6) Power consumption of the device can be saved by setting the


power saving mode

(7) Management of drug solution, restock or effluent is not required so


that it is fuss-free and it can prevent environmental contamination.

(8) Liquid such as water, developing solution, fixing solution or drug


solution is not required at all, so that there is no need to secure the
space for plumbing system or exhaust system and construction as
well as the space for stocking drug solution.

Z30N0937
2
Fuji Medical Dry Laser Imager DRYPIX Smart Product Specifications

3. System Components
3.1 Standard Components (1) DRYPIX Smart main unit 100V/200V model
(Model: DRYPIX 6000)

(2) Film tray 2 Trays


Tray 1: 14” x 17” (35×43cm)
Tray 2: 8” x 10” (20 x 25cm)

(3) Three CDs:

z Individual data CD
z Main unit software CD
z Reference Guide & QC Reference Guide
& QC for mammography CD

(4) Two documents:


z Operation Manual
z Quick Guide

3.2 Options (1) Additional option tray

(2) Cleaning roller

(3) Additional protective earth kit

(4) Fall prevention fixture kit

3.3 Supplies Fuji Medical Dry Imaging Film


• DI-HL (blue base)
35¯43cm (14”×17”) 100 films/pack
35¯35cm (14”×14”) 100 films/pack
26¯36cm (10”×14”) 150 films/pack
25¯30cm (10”×12”) 150 films/pack
20¯25cm (8”×10”) 150 films/pack
• DI-ML (blue base)
26¯36cm (10”×14”) 150 films/pack
25¯30cm (10”×12”) 150 films/pack
20¯25cm (8”×10”) 150 films/pack

3 3
Z30N0937
Fuji Medical Dry Laser Imager DRYPIX Smart Product Specifications

4. Specifications
4.1 Image Recording Laser exposure thermal development system
System

4.2 Processing Capacity Film size Processing capability (films/hour)


35 × 43cm (14”x17”) Approx. 80
35 × 35cm (14”x14”) Approx. 85
26 × 36cm (10”x14”) Approx. 100
25 × 30cm (10”x12”) Approx. 100
20 × 25cm (8”x10”) Approx. 100

* Processing capacity achieved when this equipment is in


actual use will depend on the model of the connected imaging
modality and how it is used.

4.3 Time Required for Approx. 100 sec. in 14”x17” (35x43cm, while test printing)
Output of 1st Film

4.4 Time Required for ¾ Approx. 15 minutes under 25°C room temperature
Startup conditions
¾ Approx. 10 minutes under 25°C room temperature
conditions in Power Saving Mode

4.5 Pixel Size 50μm(508dpi) / 100μm(254dpi)

4.6 Maximum Number of When recorded in 100μ


Recording Pixels 35 × 43cm (14"×17"): 3520 x 4280
35 × 35cm (14”×14”): 3520 x 3520
26 × 36cm (10”×14”): 2540 x 3600
25 × 30cm (10"×12"): 2506 x 3016
20 × 25cm (8"×10"): 2000 x 2510

When recorded in 50μ


35 × 43cm (14"×17"): 7040 x 8560
35 × 35cm (14”×14”): 7040 x 7040
26 × 36cm (10”×14”): 5080 x 7200
25 × 30cm (10"×12"): 5012 x 6032
20 × 25cm (8"×10"): 4000 x 5020

4.7 Gray Scale 14 bits

Z30N0937
4
Fuji Medical Dry Laser Imager DRYPIX Smart Product Specifications

4.8 Maximum Density DI-HL :3.0


DI-ML :4.0

4.9 Number of Input I/F 1ch of DICOM Network input

4.10 Film Supply Two supplies (two trays)


Film loading tray system under daylight conditions
・Up to five film trays can be used (Changed by operator)

4.11 Density Calibration Automatic density calibration function embedded


Function
4.12 Connectable Devices Major FujiFilm devices connectable to DRYPIX Smart include:

(1) Connection with each type of the DICOM printing compliant


image diagnostic device

(2) In case of connection with DICOM non-network compliant


image diagnostic device, DrypixLink is used

(3) In case of connection with E-DMS/Toshiba protocol


compliant image diagnostic device, FN-PS551 and
DrypixLink are used

(4) Major FujiFilm devices connectable to the DRYPIX Smart


include:
z V Station T (CR-IR392V CL)
z C@RNACORE (CC-WS674)
z CR Console (CR-IR346CL/CR-IR348CL)
z Console Advance (DR-ID300CL)
z AWS (CR-IR363AWS)

4.12.1 Format 1) Standard format (for all film sizes)


1, 2, 3, 4, 6, 8, 9, 12, 15, 16, 18, 20, 24, 25, 28, 30, 32, 35,
36, 40, 42, 48, 49, 54, 56, 63, 64, 70, 72, 80

(2) Mixed format


• 14"×17" Portrait mixed format: 6 formats
• 14"×14" mixed format: 5 formats
• 10"×14" (26×36cm) Portrait mixed format: 1 format
• 10"×14" (26×36cm) Landscape mixed format: 4 formats
• 10"×12" Portrait mixed format: 1 format
• 10"×12" Landscape mixed format: 4 formats
• 8"×10" Portrait mixed format: 1 format
• 8"×10" Landscape mixed format: 4 formats

5 5
Z30N0937
Fuji Medical Dry Laser Imager DRYPIX Smart Product Specifications

4.12.2 Gradation Eight Look-up tables (LUTs) are supported per a client (AE Title)
Conversion

4.12.3 Interpolation Three modes (Sharp, Medium and Smooth) per a client (AE
Mode Title) are available for selection

4.12.4 Image Frame Possible to attach an image frame around each image

4.12.5 Background Black and clear border are available for the background, and
one of them can be selected

4.12.6 DICOM (1) Network connection I/F


Connection ・ 10 / 100 / 1000BASE-T (Auto-Negotiation)
Specifications ・ Half duplex / Full duplex (Auto-Negotiation)

(2) DICOM network connection I/F specifications


z Verification SOP Class
z Basic Gray Scale Print Management Meta SOP Class
z Print Job SOP Class
z Basic Annotation Box SOP Class
z Print Queue Management SOP Class
z Presentation LUT SOP Class

(3) Number of clients can be registered: Maximum of 64


(4) Number of sessions that can be established at the same
time: Maximum of 10

(5) Maximum input image size Row: 8800


Column: 8800

Mammogram image : Four images of B4

4.12.7 Number of spools

5.Operation Overview (1) Image data sent from the imaging modalities are received by
the input interface and stored in image memory.

(2) Following magnification/reduction processing, image data is


sent to the recording unit.

(3) Film is taken out from the film supply tray one sheet at a
time, image is recorded, and it is output to the tray.

Z30N0937
6
Fuji Medical Dry Laser Imager DRYPIX Smart Product Specifications

6. Power Supply Conditions North America Other countries


Line voltage AC100/110/120V AC200/220/230/240V
±10% ±10%
No. of phase Single phase
Frequency 50-60Hz
Rated current 10.6/11.2/12A 5.3/5.6/5.9/6A
Grounding Class D
• Power consumption: Maximum 1.5kW

• Electric energy: Printing: Approx. 350Wh


Ready: Approx. 180Wh
Power Saving: Approx. 60Wh
• Maximum heat generation:
Printing: 1,300kJ
Ready: 700kJ

7 7
Z30N0937
Fuji Medical Dry Laser Imager DRYPIX Smart Product Specifications

7. Environmental Conditions (1) Operating conditions


Temperature: 15 ℃ ( 40%RH ) -30 ℃
(70%RH)
Humidity: 15%RH (30 ℃ ) -70%RH
(30℃)
(No dew condensation)
Atmospheric pressure: 750-1060hPa
Equipment noise
Printing: 55dB or less (except singly sound)
Ready: 45dB or less

(2) Non-operating conditions (except film)


Temperature: 0 - 45°C
Humidity: 10 - 90%RH
(No dew condensation)
Atmospheric pressure: 500 - 1,060hPa

Z30N0937
8
Fuji Medical Dry Laser Imager DRYPIX Smart Product Specifications

8. Footprint for Installation / (1) Installation space


Space Required for When Not Fixed with Fixtures
Maintenance
Front Back Left side Right side
800mm 100mm or 50mm or 200mm or
or more more more more

When Fixed with Fixtures

Front Back Left side Right side


800mm 100mm or 250mmor 250mm or
or more more more more

(2) Space required for maintenance work

Front Back Left side Right side


1000mm 600mm or 800mm or 800mm or
or more more more more

Back Top
600mm or more 400mm or more

Left side Right side


800mm or more 800mm or more

Left side Right side


800mm or more 800mm or more

Front
1000mm or more

9 9
Z30N0937
Fuji Medical Dry Laser Imager DRYPIX Smart Product Specifications

9. External Dimensions
and Mass
Width (mm) Depth (mm) Height (mm) Mass (kg)
Approx. 610 Approx. 630 Approx. 893 Approx. 104
Dimensions and Mass are approximate and are subject to change
without prior to notice. (Values in the table above do not include those
of protrusions.)

Approx.


Approx.630

Approx. Approx.


Approx.893

Z30N0937
10
Fuji Medical Dry Laser Imager DRYPIX Smart Product Specifications

11 11
Z30N0937

Вам также может понравиться