MANUAL
Laboratory Vertical Steam Sterilizers
models
2540, 3150, 3170, 3850, 3870, 5050, 5075
ELV Standard Autoclave
ELVC Including Fast Cooling System
ELVPRC Including Preparation for Fast Cooling System
10.12 Disassembling and Assembling the Door Cover, Models 5050, 5075.............. 46
10.13 Replacing the locking microswitch (Models 5050, 5075) ................................ 49
10.14 Replacing the safety-opening valve (Models 5050, 5075)................................ 50
10.15 Replacing the “DOOR CLOSED” microswitch (Models 5050, 5075) .............51
10.16 Replacing the solenoid lock (Models 5050, 5075) ...........................................52
10.17 Cleaning and Replacing the Water Level Electrodes ...................................... 53
10.18 Replacing the Printer ...................................................................................... 54
10.19 Replacing the Door Switch (models 2540, 3150, 3170, 3850, 3870)................56
10.20 Replacing the Locking Device (models 2540, 3150, 3170, 3850, 3870) ........... 57
10.21 Replacing the Plunger or Coil of the BACCARA Solenoid Valve...................58
11. IN-OUT TEST ........................................................................................................ 59
12. TROUBLESH OOTING ......................................................................................... 60
12.1 Displayed Messages and Reference to Trouble Shooting................................ 60
12.2 Pre-process malfunction ................................................................................. 61
12.3 In Process Malfunction................................................................................... 67
12.4 Mechanical malfunction ................................................................................. 86
13. LIST OF SPARE PARTS .....................................................................................87
14. PRESSURE VS TEMPERATURE FOR SATURATED STEAM ................. 97
15. XPCS Manual ....................................................................................................... 101
16. VALVES NUMBERING ..................................................................................... 114
2. SYMBOL DESCRIPTION
Stand by
1 2 3 4 5 6
FLOW DIRECTION
PLUG
20 37
SKT
20 37
1 19
JP1
1 19
JP1
P1
+ C2 U12
DIG-T2 BOARD
R3 JP7 1 21
JP5 1 JP6
1
C9 U8 JP2
R2
U16
C7
R1 U11
C5 U14 U3 C4 U18
R9 JP12 JP8
DIG T3
97-05 C25 C24 U22
D1
C15 U19 1 1
JP3 C23
8
1
C22 U1 C6
2 34
DS 1
1 33
JP 1
1 19
R8
R7 R6 R5 C2
C1 C1 R1
0
C1 U5 9 3 R1
3 U1
2
3
C1 C1 U1 R1
+C2 C1 JP 1 C2 U2
U1 4 2 0 +
0 4 1 0
2
C8
R3 C1
RP 1 U2
2 3
U7
U6
C9 U8
R2
U1
6 C7
R1 U1
C5 U1 U3 C4 U1
1 4 8
C3 + U4 R4
C1
C1 6
U2 7
Q1
1 RP Y1 BZ1
1
U1
U1 5
7
D IG C2 C2 U2 R9 D1
H2 C1 U1
T917 -.V
05
3 5 4 2
9
5
1 JP3 C2 8
3 C2 U1 C6
2 2 34
DS
1 1 33
ANL T1 BOARD
GUL P
20 37
1 19
JP1
P1
1 21 1 1
1 1 1 1
1 1 1
AC-T1 BOARD
JP5 JP1
1 RV3 RV2
JP2
R9 R5 R1
1
U3 U2 U1
R12 R10 R6 R2
C3 R11 R8 C2 R7 R4 C1 R3
JP6 L1
JP7 JP4 RV1
JP3
1
C5
Q3 Q2 Q1
ETYPEAC-97
CAT TECH.
Y
310
300 ACTUAL
PRESURE
100
90
X
V1 V2
a INPUT FROM SENSOR
The calibration method enables to input this data into the system in order to
perform automatic correction of OFFSET and GAIN.
It is necessary to define the two points, which have to be input to the system,
prior to performing calibration.
— Select CALIB CODE: 107 to calibrate the system by means of the UP/DN
keys and press the PROGRAM key again until TEMP CAL:
will be displayed.
— Set calibration code to TEMP CAL: 1
— Press the PROGRAM key again and do not change the code.
— The cursor on the lower row will move to R130.0, change to R123.0 by
means of keys UP/DN, then press the PROGRAM key
— When the cursor is set under any figure, it enables to feed this value to the
reading of temperature by pressing the PARAMETERS key.
Example:
If the cursor is set under R68.0 on the upper row and the actual temperature of
the autoclave is now 66.0ºC, by pressing the PARAMETERS key, the
temperature reading will be
corrected from R68.0 to R 66.0.
Example:
When the working temperatures are 134ºC and 121ºC but the actual
temperature are 132ºC and 121ºC
The measured temperature for 134ºC must be corrected without changing the
temperature at 121ºC,
Considering the above example, the setting of data can be done as follows:
A 121.0 R 121.0
A 132.0 R 134.0
The system will calibrate the new offset and gain and retain them into the non-
erasable memory.
As a result, while running a 134ºC program the correction is automatically
made, while at 121ºC no change is felt.
To perform the pressure calibration two points have to be defined allowing the
calculation of new OFFSET and GAIN values.
— Press the PROGRAM key.
— Skip date and the time setting by pressing the PROGRAM key six
times.
— CODE: will be displayed.
— Select the Technician Code (011) by the UP/DN keys and press the
PROGRAM key.
— Select CALIB CODE: 107 to calibrate the system by means of the
UP/DN keys and press the PROGRAM key again until the system
displays: PRES CAL:
— Press the PROGRAM key.
The following is displayed:
Point 1 – Upper Row A 100 R 100
Point 2 - Lower Row A 300 R 300
— Open the door, do not alter A 100.
— Press the PROGRAM key, the cursor will move to R 100.
— Press the PARAMETERS key. The system performs reading
pressure and displays it instead of R100. E.g. R97 (i.e. the
system reads out 97kPa, when the door is open).
— Press the PROGRAM key.
— The cursor moves to the lower row to A300.
If the system reads out 280kPa, do not change A300. Press the PROGRAM
key to move to R300 and change it to the actual
pressure value 280kPa by means of the keys UP/DN.
— Press the PROGRAM key again.
The system calculates the new values for offset and gain and stores these
values in the non-volatile memory.
This completes the calibration procedure for the pressure.
NU TP FUNCTION VALVE
TP1 GND
TP2 +5V DC
TP3 +12V DC
TP4 OUT - HEATERS ON – 5V; OFF –0V
TP5 OUT-VACUUM PUMP ON – 5V; OFF –0V
TP6 OUT-EXH-2 VALVE ON – 5V; OFF –0V
TP7 OUT-AIR VALVE ON – 5V; OFF –0V
TP8 OUT-EXH-1 VALVE ON – 5V; OFF –0V
TP9 OUT-WATER TO CHAMB. ON – 5V; OFF –0V
TP10
TP11 OUT-DOOR LOCK ON – 5V; OFF –0V
TP12 OUT-AIR RELEASE ON – 5V; OFF –0V
TP13 OUT-COMPR. AIR ON – 5V; OFF –0V
TP14 OUT-WATER COOLING ON – 5V; OFF –0V
TP15
TP16
TP17
TP18
TP19
TP20 IN –TEMPERATURE (OPTION) 134ºC-1.97v
TP21 IN –TEMPERATURE (MAIN) 134ºC-1.97v
TP22
TP23
TP24 IN – PRESSURE 2.0B – 2.031v
TP25 IN – ELECRODE LOW ‘0V’ – YES, ‘2.5V’ – NO
TP26 IN – ELECTRODE HIGH ‘0V’ – YES, ‘2.5V’ - NO
Set Cycle 1 2 3 4 5 6 7
value Value 134 121 121 121 105 121 0
Set Cycle 1 2 3 4 5 6 7
value Value 3 15 20 20 18 30 0
Set Cycle 1 2 3 4 5 6 7
value Value 0 0 0 0 0 0 0
Set Cycle 1 2 3 4 5 6 7
value Value 20 20 20 20 20 20 20
9.2.5 Heat T.O. - The maximum time allowed for the heating stage
(testing)
After this time, if the system does not enter the sterilization stage, it
will receive a ‘Low Heat’ message and the process will abort.
Entry Code – 11
Resolution – 1min
Minimum Value – 20min
Maximum Value – 300 min
Set Cycle 1 2 3 4 5 6 7
value Value 60 60 60 90 90 90 0
Set Cycle 1 2 3 4 5 6 7
value Value 120 120 100 85 85 85 134
Set Cycle 1 2 3 4 5 6 7
value Value 2 1 2 1 1 1 1
Set Cycle 1 2 3 4 5 6 7
value Value 12 20 20 20 20 20 20
Set Cycle 1 2 3 4 5 6 7
value Value 20 20 20 20 20 20 20
Set Cycle 1 2 3 4 5 6 7
value Value 30 30 30 30 30 30 30
Set Cycle 1 2 3 4 5 6 7
value Value 30 30 30 30 30 30 30
9.2.12 Ex Mode – The method for exhausting the steam at the end of the
process
Entry Code – 11
Resolution – 1
Ex Mode
Fast Ex 1
N.A. 2
Slow Ex (Waste) 3
Slow Ex (Liquids) 4
Cooling with compressed air 5
Cooling with water circulation 6
Set Cycle 1 2 3 4 5 6 7
value Value 1 1 3 4 4 6 1
Set Cycle 1 2 3 4 5 6 7
value Value 0 0 0 0 5 0 0
Set Cycle 1 2 3 4 5 6 7
value Value 100 100 100 100 100 100 100
9.2.15 Ex Time
This defines the time the exhaust valve should remain open after
reaching a pressure of 127kpa for removing the water in the chamber.
In some autoclaves 15 seconds is enough, and in the large autoclaves
it is possible that there will be a need to extend this to 40 seconds.
Entry Code – 11
Resolution – 10 sec
Minimum Value – Ø sec
Maximum Value – 240 sec
Cycle
1 2 3 4 5 6 7
Model
2540 45 45 30 30 30 30 30
Set
value 3150/70 45 45 30 30 30 30 30
Value
3850/70 45 45 30 30 30 30 30
5050/75 60 60 60 60 30 45 30
Set Cycle 1 2 3 4 5 6 7
value Value 0 0 0 0 0 0 0
Set Cycle 1 2 3 4 5 6 7
value Value 5 3 1 1 1 1 1
Set Cycle 1 2 3 4 5 6 7
value Value 0 0 0 0 0 0 0
Set Cycle 1 2 3 4 5 6 7
value Value 400 400 400 400 400 400 180
9.2.21 Ex Rate
It sets the rate of the exhaust in the slow exhaust. It is possible in
autoclaves that have a top exhaust valve. A combination of natural
exhaust (through the condense valve) and the operation of the top
exhaust valve in Shoot Mode when:
Entry Code – 11
Resolution – 1 (step)
Minimum Value – 0
Maximum Value – 5
Ex Rate Top Exhaust
0 = Closed
1 = 5% of the time Open
2 = 20% of the time Open
3 = 35% of the time Open
4 = 50% of the time Open
5 = 70% of the time Open
Set Cycle 1 2 3 4 5 6 7
value Value 0 0 0 0 0 0 0
Set Cycle 1 2 3 4 5 6 7
value Value 1.5 1.5 1.5 1.5 1.5 1.5 1.5
Set Cycle 1 2 3 4 5 6 7
value Value 0 0 0 0 0 0 0
9.2.24 Temp2+
For Hot Water systems, there is an extra sensor placed outside of the
liquid media to measure the temperature of the hot water. If this
temperature rises above the value of the Ster. Temp and
Temp2+combined, then the heating elements will shut down.
For example, if Ster Temp is 121C, and Temp2 is 3C, if the hot
water temperature outside the liquid rises above 124C (i.e. 124.1C
or above), the heaters will not activate.
Entry Code – 33
Resolution – 1C
Minimum Value – 1C
Maximum Value – 5C
Set Cycle 1 2 3 4 5 6 7
value Value 3 3 3 3 3 3 3
Set Cycle 1 2 3 4 5 6 7
value Value 0 0 0 0 0 0 0
After performing these tests, the Service Engineer should complete and sign
the Work Order.
Rear
1
2 3 3 4 6 3 5
For models 2540, 3850, 3870 For models 5050, 5075
Cat. No.
No. Description 2540, 3150,
5050, 5075
3170, 3850, 3870
1 CE-marked SVL029-0028 SVL029-0005
Safety valve
1 ASME SVL029-0004 SVL029-0007
2 Safety valve base CMV100-0004
3 Washer CMT240-0022 LOK505-0001
4 Nut CMT240-0020
5 Street elbow F*M ½*½ 90 FIT100-0065
6 Male connector ½*½ FIT100-0296
5
3
N Cat No.
Description
o. 2540 3150, 3170 3850, 3870 5050, 5075
BOL194- BOL194- BOL194- BOL191-
2 Screw 0340 0340 0340 0140
BOL194- BOL194- BOL194- BOL194-
3 Screw 0341 0341 0341 0341
BOL194- BOL194- BOL194- BOL194-
8 Screw 0340 0340 0340 0340
BOL194- BOL194- BOL194- BOL194-
9 Screw 0340 0340 0340 0340
NUT192- NUT192-
Nut
0185 0185
NUT193- NUT193-
Washer
0276 0276
Spring Washer NUT193- NUT193-
10
¼ 0317 0317
NUT193-
Star Disk
0291
BOL191-
Screw
0140
Sheet Nut NUT192- NUT192- NUT192- NUT192-
11
4.8x1.25 0230 0230 0230 0230
CMT254- CMT315- CMT240- CMT507-
12 Cover 0020 0001 0031 0008
12
2 5
2
2
1 11
5
Cut-Out Thermostat
Reset button 6
7 6 5 4 3 2 1
For models
3150, 3170,
3850, 3870
Locking solenoid
6
A
5 3 4 3 1
No. Description Model Cat. No. No. Description Model Cat. No.
2540 BOL191-0032 3 washer All models ELE036-0009
1 screw 3150, 3170 4 spring All models SPR177-0012
BOL191-0091
3850, 3870 2540 POL065-0053
2540 BOL191-0033 6 door cover 3150, 3170 COV314-0001
Warning:
Before starting, disconnect the instrument from the power source
and ensure that there is no pressure in the autoclave chamber.
1. Release the two (2) fastening screws (1).
2. Remove the locking handle (10) and the Teflon ring (9).
3. Remove the screws covers (2).
4. Unscrew the cover fastening screws (3, 5) and remove the
screws and the washer (4).
5. Remove the door cover. Be careful not to loose the springs (6)
and the washers (4a) located beneath the door cover. Washer
(7) is a disposable nut and must be replaced buy a new nut
when the door cover is re-assembled.
1 2 3 4
8 3
4
9
5
7 6
No No
Description Cat. No. Description Cat. No.
. .
Spring washer
1 Arm guide LOK507-0031 6 NUT193-0315
M5
2 Valve PNE195-0054 7 Side bracket HLD505-0001
Spring washer
3 NUT193-0317 8 Nut M6 NUT192-0193
¼”
4 Nut ¼” NUT192-0185 9 Screw M6 BOL191-0150
5 Nut M5 NUT192-0192 10 Screw ¼ x 1½ BOL191-0174
1
2 A
7
3
4
5
No No
Description Cat. No. Description Cat. No.
. .
1 Nut M5 NUT192-0192 4 Bolt ¼ x ¾ BOL191-0142
2 Washer 3/16 NUT193-0275 5 Nut ¼ with flange NUT192-0155
3 Spring washer M5 NUT193-0315 8 Microswitch ELE036-0004
3 4 5 1
10.17.2 Cleaning
1. Remove the basket from the autoclave.
2. Clean the electrode tip.
3. Test the unit.
1, 2, 3, 4, 5, 6, 7 8, 9 10
up down up down
1
3
Cat. No.
No. Description 3150, 3170,
2540
3850, 3870
6 Door locking device pin LOK240-0031 LOK387-0016
9 Closing bridge “c” clip NUT193-0302 NUT193-0303
10 Okolon disc LOK240-0017 LOK387-007
The solenoid valves may be out of order due to faulty plunger or coil. To repair
the solenoid valve – replace the faulty plunger or solenoid
Caution!
Before starting, disconnect the instrument from the power source and
ensure that there is no pressure in the chamber, coil or generator.
Allow the autoclave to cool before removing outer covers.
1. Open the relevant service door to reveal the faulty BACCARA solenoid
valve.
2. Unscrew and remove nut (1).
3. Remove the coil (2).
4. Unscrew the plunger + spring (4) with the plunger housing (3) and replace
it with a new one. The plunger the spring and the housing are replaced as a
kit.
5. Reassemble the coil (2) and the nut (1).
6. Re-install the autoclave cover.
5 4 3 2 1
DISPLAYED
ITEM ACTIVATED REMARKS
NOTICE
Exh On Exhaust valve Exhaust valve is open
Heating elements are activated.
Heat on Heating Elements Do not perform this test
without water covering the
heating elements
Water On Water inlet valve Water inlet valve is open
Exh2 On N/A N/A
Air inlet valve is open. Applicable
Air Valve Air inlet valve
only for ELVC models
Air Release On Air release valve Air release valve is open.
Door lock solenoid is open. Verify
Door Lock On Door lock solenoid
that the solenoid has retracted.
Water circulation valve is open.
Cool&Circ On Water circulation valve
Applicable only for ELVC.
Compressed air inlet open.
Comprs Air On Compressed air inlet
Applicable only for ELVC.
Thermost is in “connected”
Thermostat “0”
position
Safety thermostat
Thermost is in “disconnected”
Thermostat “1”
position
Door Sw “0” Door switch closed
Door switch
Door Sw “1” Door switch open
Electrd-high High water level electrode does
255 not sense water
High Water level electrode
High water level electrode does
Electrd-high 13
sense water
Electrd-Low Low water level electrode does not
255 sense water
High Water level electrode
Low water level electrode does
Electrd- Low 40
sense water
Low Heat 1 4 1 8, 11
4, 11,
High Temp. 1 1 1 2, 5
13, 16
4, 11,
Low Pres. 1 2, 4, 5
13,
Door Unlock 2 19
1, 14,
Low Water 1, 2, 3 3 2 2, 5
21
1. No response when the 1.1 Check the power cord and the connections 1.1 Replace damaged cord or repair faulty
system is turned on at the autoclave input. connections.
1.2 Check the cut-off thermostat. 1.2 Reset cut-off.
If operation continues after resetting the cut-off.
find out what caused the thermostat to
switch-off.
1.3 Check the circuit breaker. If it tripped,
there is a short-circuit. 1.3 Return the circuit breaker lever to the
“ON” position after fixing the cause for
1.4 Check the fuse of the power supply. If the the short circuit.
fuse is blown, check if one of the valve
coils or one of the electronic boards 1.4 Replace the blown fuse after fixing the
causes a short circuit. cause for the short circuit.
1.5 Check the AC voltages.
— The input supplies voltage at the control
1. (cont.) 1.6 Check the DC voltages supply to the — Replace faulty connector.
system.
Check main power switch.
1.6.1 Check the grounding
— Plug connector 26 pin FLAT of the
test board into connector JP5 mounted
on board ANL-T2.
— Connect the negative probe of the Replace faulty power switch
multitester to TP1 on the test board.
— Check with the positive probe of
the multitester the voltage at different
grounding points of the autoclave, on
the ANL-T2 and DIG-T2 boards, on
the enclosure of the electric box and
the main grounding screw of the
machine.
The voltage differences must be less
2. System is on and display is 2.1 If the other functions are OK, the display 2.1 Replace the DIG-T2 board
not lit. is probably damaged.
3. System on and erroneous or 3.1 DIG-T2 board is faulty. 3.1 Replace DIG-T2 board.
fragmented digits.
4. No response from one of the 4.1 Keyboard is faulty. 4.1 Replace the keyboard.
6. Pressing the START key 6.1 Incorrect reading of the door switch 6.1 See symptoms 18 & 19 in para. 13.3.5.
does not start the process, position
despite the door is closed.
12.3 In Process Malfunction
12.3.1 Water Inlet Stage
1. If there is any malfunction 1. Press door switch while the door is open
related to inlet water and simultaneously press START push-
quantity (Low Water, High button.
Temp, Low Press or Low
Message Water inlet will appear on the
Heat), check the automatic display and water enters the chamber until
water filling into the it reaches the grove adjacent to the door.
chamber. On this point the water must stop inserting
the chamber.
If the autoclave does not operate as
described above, proceed as follows:
2. The water inlet phase is 2.1 Incorrect reading of the water sensing 2.1 See symptoms 16 in para. 13.3.5.
short and the water level in electrode.
the chamber is low. 2.2 Programmed interval after the electrode
4. Water does not enter 4.1 No water supply 4.1 Resume water supply.
chamber during “FILL 4.2 Open main water valve.
4.2 Main water valve closed.
WATER” stage. 4.3 Fix disconnection.
4.3 No voltage at solenoid input.
Verify command at Test Point TP9 from 4.4 Replace plunger or complete solenoid.
electronic box. 4.5 Clean water inlet pipe
4.6 See symptoms 2 & 3 in para. 13.3.5.
4.4 Solenoid faulty. Replace water inlet valve.
4.5 Water inlet pipe is clogged. 4.7 Replace plunger or complete solenoid.
4.8 Replace air release valve.
4.6 Water inlet valve (21) is faulty.
4.7 Valve’s solenoid is faulty.
4.8 Air release valve (41) is faulty.
1. Process fails while initiating 1.1 The problem is the incorrect reading of 1.1 See symptoms 9, 11 and 12 in para.
the heating stage. message the temperature sensor. 13.3.5.
HIGH TEMP is displayed.
2. Message DOOR UNLOCK 2.1 The pressure in the chamber rises during 2.1 Close the door more tightly or set the
is displayed during the the heating stage and is pressing on the door switch more steadily.
heating stage, although the door. If the door switch is lightly pressed
door is closed. it could be turned off.
3. LOW WATER message is 3.1 The process fails if the electrode is above 3.1 If amount of water is not correct, clean
displayed during the heating the water surface and the safety the electrode and set the water level
stage and process fails. thermostat switches off.
Check if the amount of water in the
4. Message LOW HEAT is 4.1 The heating time is very long and the
displayed during the heating heating rate is normal.
stage. 4.1.1 Excess quantity of water in the 4.1.1 Clean or replace electrode.
chamber due to dirty electrode,
loose electrode connections or
faulty electrode.
4.1.2 Excess quantity of water in the 4.1.2 Level autoclave.
chamber due to unleveled
autoclave. 4.2 See symptoms 8 in para. 13.3.5.
1. The fast exhaust operation is 1.1 Exhaust valve is faulty or pipe is clogged. 1.1 Replace exhaust valve or clean pipe.
too slow.
2. (cont.) 2.3 Check the solenoid coil input voltage with 2.3 Repair or replace the valve.
the solenoid connected. remove the cover
that protects the terminals of the coil. If
the valve is open despite the voltage being
less than 2VDC, there is a mechanical
problem.
2.4 If all items checked above conform to
requirements, ANL-T2 board is faulty.
3. Air valve is always closed. The problem may be the mechanical part of the
valve, the solenoid, the driver on ANL-T2 board
or the control circuit on DIG-T2 board.
3.1 Check TP7. If voltage is constantly 0V at
the stage when the valve must be open
then DIG-T2 board is faulty.
3.1 Replace board DIG-T2.
3.2 Check the connection between connector
JP5/3 and the valve.
Fix the connections.
4. Air valve is always open. The problem could be the mechanical part of the
valve, the driver located on ANL-T2 board or the
control circuit on DIG-T2 board.
4.1 Check TP7. If the voltage is higher than
4VDC and there is a control signal to the 4.1 Replace board DIG-T2.
valve, the problem is on DIG-T2 board.
If the voltage is 0VDC and the valve is open, the
problem is not the control circuit but one of the
other elements of the system.
4.2 Check the connection between connector
JP5/3 and the valve.
4.3 Check the voltage at the solenoid coil,
while the solenoid is connected to the 4.2 Fix the connections.
valve, by removing the cover that protects
the terminals of the coil. If the valve is
open despite the voltage being lower than 4.3 Repair or replace the valve.
8. Heating elements do not 8.1 Check the rated AC voltage (115V or 8.1 If ac voltage exists – check according
work 230V) at the AC-OUTPUT on AC-T1/ to par. 8.2 ahead.
JP4. Verify that AC voltage exists and has
If ac voltage exists – check according
the required value.
to par. 8.3 ahead.
8.2 Uncover cautiously the heating elements
connections and measure the heating
elements input voltage. If no voltage
reaches the heating elements locate the
interruption between connector AC-
OUTPUT (JP4) and the heating elements. 8.2 Fix the connections.
8.3 If there is no AC voltage at the output of
the SSR the problem is the SSR.
8.4 The heating elements receive the normal
voltage, but do not work.
8.3 Replace the SSR.
10. The temperature readout 10.1 The sensor is short-circuit. Check the 10.1 Fix the connections.
permanently indicates connections of the sensor to the board.
temperature less than 30ºC, Make sure that the wires and connectors
although chamber are in good condition.
temperature is high. 10.2 The temperature sensor is defective.
10.2 Replace The temperature sensor (See
symptoms 9.2 in this paragraph).
10.3 The temperature measuring circuit on
ANL-T2 board is defective, 10.3 Replace ANL-T2 board.
10.4 The A/D component on ANL-T2 board is
defective. 10.4 Replace ANL-T2 board
11. The temperature readout is 11.1 The system must be calibrated. 11.1 Calibrate the temperature measuring
not accurate. There is a circuit, as described in para. 6.2
11.2 System cannot be calibrated.
deviation of ± 20ºC. (Temperature Calibration).
11.2 replace ANT-T2board.
14. The pressure readout 14.1 Check the sensor connections and the 14.1 Fix the connections.
indicates permanently soldering at the pins of the sensor
pressure over 350 kPa connections to ANL - T2 board.
14.2 The pressure sensor is defective.
14.2 Replace the pressure sensor.
14.3 The pressure measuring circuit on ANL-
T2 board is defective. 14.3 Replace board ANL-T2.
14.4 The A/D component on ANL-T2 board is
defective. 14.4 Replace board ANL-T2.
15. The-pressure readout 15.1 Check the sensor connections and the 15.1 Fix the connections.
indicates permanently soldering at the pins of the sensor
pressure less than 50kPa. connections to ANL-T2 board.
15.2 The pressure sensor is defective.
15.2 Replace the pressure sensor.
15.3 The pressure measuring circuit on ANL-
T2 board is defective. 15.3 Replace board ANL-T2.
15.4 The A/D component on ANL-T2 board is
defective.
18. The door switch indicates 18.1 Check if door switch is stuck or faulty. 18.1 Replace the door switch.
door is permanently closed. 18.2 Check connection of switch to ANL-T2 18.2 Fix the connections.
board. Verify that no short circuit exist on
both connections (to switch and to board)
and on the entire wire.
18.3 If the switch is in order the input circuit
on ANL-T2 board is faulty. 18.3 Replace board ANL-T2.
19. The door switch indicates 19.1 Check connection of switch to ANL-T2 19.1 Fix the connections.
door is permanently open. board. Verify that no short circuit exist on
both connections (to switch and to board)
and on the entire wire.
19.2 If the switch is in order the switch-input
circuit on ANL-T2 board is faulty. 19.2 Replace board ANL-T2.
20. The safety thermostat circuit 20.1 Disconnect the wires from the thermostat 20.1 Replace faulty thermostat.
22. Steam leaks through the 22.1 Check if the gasket is smooth, making 22.1 Replace the faulty gasket.
door, despite the fact that sure the door closes properly.
door is tightly closed.
23. On models 2540, 3150, 3170, 23.1 The locking mechanism is faulty. 23.1 Repair or replace the locking mechanism
3850 and 3870 the door
locking mechanism enables
opening the door, pressure
higher than 150kPa.
24. Cooling water valve is The problem may be the mechanical part of the
always closed. valve, the solenoid, the driver located on the
ANL-T2 board or the control circuit on the DIG-
T2 board.
24.1 Replace DIG-T2 board
24.1 Check TP14. If the voltage is constantly
0V at the cooling stage, when the valve
must be open, then the problem must be
on DIG-T2 board. 24.2 Replace board ANL-T2.
25. Compressed air valve is The problem could be the mechanical part of the
always open. valve, the driver located on the ANL –T2 board
or the control circuit on the DIG-T2 board.
25.1 Check TP13. If the voltage is higher than 4
VDC there is a control signal to the valve. 25.1 Replace DIG-T2 board
The problem is on the DIG-T2.
26. Compressed air valve is The problem could be the mechanical part of the
always closed valve, the solenoid, the driver located on the
ANL-T2 board or the control circuit on the DIG-
T2 board.
26.1 Check TP13. If voltage is constantly 0V at 26.1 Replace DIG-T2 board
the cooling stage while the valve must be
open then the problem must be on DIG-T2
board.
26.2 check the voltage on JP5/9. If the voltage 26.2 replace board ANL-T2.
is above 10VDC the driver on ANL-T2 is
faulty
26.3 Check the connection between the 26.3 Fix faulty connections.
connector JP5/9 and the valve.
26.4 Check the solenoid coil input voltage 26.4 replace the solenoid. If malfunction
persists repair or replace the valve.
while the solenoid is connected. if the
1. The safety valve does not 1.1 Check the path of safety valve ensuring 1.1 If circulation is not free, remove the block.
release pressure when blow- circulation is free. 1.2 Replace the faulty valve.
off test is performed. 1.2 Safety relief valve is faulty
2. Safety relief valve opens at 2.1 Safety relief valve is clogged. 2.1 Replace relief valve and check the water
pressure lower than quality.
specified. 2.2 Safety relief valve is faulty. 2.2 Replace relief valve
3. Safety relief valve opens at 3.1 Safety relief valve is faulty. 3.1 Replace relief valve
pressure higher (more than
10%) than specified.
4. Safety relief valve does not 4.1 Safety relief valve is faulty. 4.1 Replace relief valve
6. Locking pin is loose in the 6.1 Wear of components. 6.1 Replace pin and sleeve.
bronze sleeve.
7. Vibration during operation of 7.1 Pump’s rubber legs are loos or damaged. 7.1 Tighten or replace rubber legs.
pump.
13. LIST OF SPARE PARTS
Cat.No.
Description
2540 3150, 3170 3850, 3870 5050, 5075
DPU 20 THE002-0005
Printer
DPU 30 THE002-0022 THE002-0022 THE002-0022
DPU 20 THE002-0003
Printer paper roll
DPU 30 THE002-0025 THE002-0025 THE002-0025
Sensor, Temperature, PT-100, ELC240-0022 ELC240-0022 ELC240-0022 ELC240-0022
5*100, Assy.
O-Ring, 2.84 x 8.08, W=2.62,
GAS080-0022 GAS080-0022 GAS080-0022 GAS080-0022
EPDM (for PT100)
Thermostat, Cut-Off, TY95-H,
THE005-0016 THE005-0016 THE005-0016 THE005-0016
155C, Campini
Pressure transducer MPX2200 THE006-0003 THE006-0003 THE006-0003 THE006-0003
Heater 2400W/230V HEA012-0012 — — —
Heater, 3300W/230V — HEA012-0006 — —
Heater 3 x 2000W/230V — — HEA012-0009 —
Heater 3 x 3000W/230V — — — HEA012-0011
Switch, Rocker, 16A ELE035-0012 ELE035-0012 ELE035-0012 ELE035-0012
Microswitch, E13-00M, CHEERY
ELE036-0001 ELE036-0001 ELE036-0001 —
15A, 125/250VAC, 3/4HP
Microswitch, MP320-5MAL,
— — — ELE036-0019
Microprecision Electronics
Microswitch with Antenna Spring — — — ELE036-0004
Strainer, Water, 1/4", Y type 192,
FIL175-0028 FIL175-0028 FIL175-0028 FIL175-0028
ITAP
Gauge, Pressure, Steam, 0-60 psi, 2.5" — — — GAU029-0003
Handle, Door, Bakelite HAN071-0003 HAN071-0006 HAN071-0006 HAN071-0006
DPU20 POL062-0002 —
Base, Control Panel
DPU30 POL062-0013 POL062-0003 POL064-0013
DPU20 POL067-0002 —
Printer opening cover
DPU30 POL067-0007 POL067-0007 POL067-0007
Base, Pressure Gauge, 50xx ELV, Pol. — — — POL064-0002
Completion for front panel
POL063-0002 POL063-0005 POL063-0003 —
Polyethylene
Cover, Door With Manometer POL066-0002 — — —
Cover, Door — POL065-0009 POL065-0003 POL065-0005
Cover, Validation Port POL065-0006 POL065-0006 POL065-0006 —
Gasket, Door, Black, 2540 GAS080-0184 — — —
Gasket, Door, Gray, 31XX — GAS080-0195 — —
Gasket, Door, Green, 38XX — — GAS080-0198 —
Cat. No.
No. Description 3150, 3170,
2540
3850, 3870
1 Bushing LOK240-0003 LOK387-0003
2 Door tightening bolt assembly LOK240-0034 LOK387-0029
3 Locking screw housing LOK240-0043 LOK387-0042
4 Locking base LOK240-0030 LOK387-0031
5 Locking housing axe LOK240-0013 LOK387-0036
6 Door locking device pin LOK240-0035 LOK387-0016
7 Bakelite handle HAN071-0016 HAN071-0015
8 Closing bridge “c” clip NUT193-0302 NUT193-0303
9 Cotter pin LOK692-0039 LOK692-0039
10 Okolon disc LOK240-0017 LOK387-0037
Door tightening bolt – Door with locking
LOK240-0028 LOK387-0028
solenoid, assembly
Definitions
PC – Personal computer.
Program – a complete, self-contained set of computer instructions that you use to perform a
specific task such as word processing, accounting or data management. Program is also called
Application.
PCS Target platform – Cat Technologies LTD. controlling system based on PCS card.
PCS communication – exchange data between PCS components via RS232 or RS485, using
PCS data exchange protocol.
Calibration - adjusting analog input presentation value by changing gain and offset values.
Overview
XPCS is a Data exchange program that lets you perform downloading of PCS code,
downloading/uploading analog inputs gain and offset values. The program using PCS
communication module, to connect between target platform and PC.
XPCS Aim
Download PCS code data.
Calibration.
Minimum Requirements
Pentium processor or equivalent.
32MB of RAM, 32 MB of free hard drive space.
CD-ROM drive.
Available serial communication port.
Windows 9X/NT 4.
User Requirements
To operate this program, user needs a basic knowledge of PCS.
Using XPCS
In 'Port Number' text box type serial port number connected to the PCS Target platform, or
modem port number to remotely connection.
In 'Baudrate' text box type the PCS Target platform baud rate. Usually the value is: 19200.
'Wait loop' it is a communication time error counter. To fix communication problems with
remote targets via modem, try to insert greater value.
To connect to remote target via a modem, select 'Active' check box in 'Modem Config' frame.
'Connection Timeout' it is a waiting time in seconds to requested data form the PCS target
platform. If no data retrieved from the target after wait time, the program generates a
communication time error.
'Dial Time Out' it is a waiting time in seconds to be used by the local modem trying to
connect to the remote modem. On end of time, the program generates timeout error.
If 'Get and set Gain Offset' checked, when program performing download, all gain and offset
values will be stored before PCS code downloading, and restored after downloading are
finished.
If 'Get and Set Parameters' checked, when program performing download, all parameters will
be stored before PCS code downloading and restored after downloading are finished.
Note: Only current selected Program (For example: 1-Instruments) parameters will be
saved and restored.
Note: Because the PCS target platform parameters saving process are complex, restoring
Parameters task takes long time (few minutes)!
Note: Theirs no need to edit options every time that program start running. The program
saves setting on 'OK' button pressing and starting up with last settings.
If one of the tasks fails, the program will abort connection process and will display
relevant error message.
If 'Modem active' option not selected, by pressing on 'Connect' button, the program will open
communication port. In this case, communication establishment tasks will be:
Check if COM Port is available.
Open COM Port.
Note: In this case the program not checking if target platform connected.
After connection was established, the main XPCS window will be displayed.
When program is online, buttons 'Go offline', 'Calibration', 'Download' will be displayed.
Pressing 'Go Offline' button will close current connection and buttons 'Go Online',
'Options' and 'Database' will be displayed on the main XPCS window:
Calibration window
When program connected, double click on one of the inputs table rows, will display
calibration tool. With this tool you can calibrate selected analog input.
Min and Max – limits for gain and offset values. If ADMC compatible type file not
found, Min and Max labels will show "N/A".
Note: When Min and Max are N/A, you must be very careful, because the gain and
offset values not limited at this case and may be inserted wrong gain or offset value.
Upload button – Press this button to upload gain and offset values from the target
platform. All 16 analog inputs gain and offset values will be uploaded.
Download button – Press this button to download Gain and Offset of the current selected
analog input. The values displayed in Gain and offset text boxes will be downloaded. If
Inserted values are wrong, relevant error message will be displayed and download
operation will be aborted.
Calculation
Calculate Gain and Offset values by inserting 4 values: Actual High, Actual Low, Read
High and Read Low to the formula. For more information about Gain and offset
calculation see User Manual supplied with PCS Target platform. Pressing 'Calculate'
button will calculate new Gain and Offset values. The new values will be displayed in the
Gain and Offset text boxes.
Download window
Note: No need to select type file, if requested file name already displayed on the button
(In example: "LabTn2.DWN").
Status Bar
The status bar at the bottom of the window, displaying current downloading process status
(In Example: "Ready to process").
'Download' Button
Press this button to perform download with selected file. When download performing, the
program displaying on the window downloading process progress messages. After
download process are finished, the status bar displaying "Download Finished".
Wait for target platform startup. After downloading of the file complete, the PCS
Target platform restarts. It takes several seconds to target start running with new
code and before downloading parameters and gain/offset, we must wait to target will
start again communicate with the PC. Status message: "Waiting for machine
startup..."
If 'Get and Set Parameters' option in 'Options' window is selected, and downloaded
file type is ADMC compatible, the program will download all parameters of the
current selected program (for example: 1-instruments). Status message:
"Downloading Parameters..."
If 'Get and Set Gain & Offset' option in 'Options' window is selected, the program
will download values of all analog inputs gain and offset. Status message:
"Downloading Gain and Offset..."
Note: If error occurs during download process, relevant error massage will be displayed.
The message box with 'Retry' option will happen.
At the end of the download process, the window looks like in the picture:
Reason: Probably the communication between PCS Target and PC fail, or downloaded file
are wrong or damaged.
Reason: Probably the communication between PCS Target and PC fail, or PCS Target
platform are damaged and can not save data on flash memory.
Reason: Probably the communication between PCS Target and PC fail, or PCS Target
platform are damaged and can not save data on flash memory.
Tips: Try to download other file. If operation success, try to find what application using the
file and in the worse case, restart the PC. If the problem continues, probably the files are
damaged. In this case try to replace the file with new same file from the ADMC installation
CD under 'Support' directory.
Reason: In one of the text boxes inserted wrong number, or not numeric data.
Reason: In one of the text boxes inserted wrong number. The number must be:
1>Number>1000.
Tips: Read High value must bee great than Read Low value.
Tips: Read High value must bee great than Read Low value.
Description: This message displayed when user trying to download Gain and Offset by
pressing 'Download' button on 'Calibration' window.
Reason: Gain or Offset value displayed in Gain or Offset text boxes are not in limits.
Tips: The value in the Gain or Offset boxes must be in Min/Max limits.
Description: This message displayed when user trying to download Gain and Offset by
pressing 'Download' button on 'Calibration' window.
Reason: Probably the Gain or Offset value displayed in Gain or Offset text boxes are not
numeric data.
Description: This message displayed when user pressing on 'OK' button on 'Options'
window.
Description: This message displayed when user pressing on 'OK' button on 'Options' window.
Description: This message displayed when user pressing on 'OK' button on 'Options' window.
81. Inlet
8 82. Main inlet
GAS
. 83. Inlet through humidifier
To enable the connection of the autoclave to the 1 phase electrical net, a special switch box is
supplied with the autoclave. This switch box shall be placed on the wall and the wiring
connection shall be as described below.
FROM FROM
AUTOCLAVE NETWORK GND N
GND R
N S
L T