OxyGen: A Low-Cost Oxygen Concentrator for Low-Resource Settings

Gaurav Begur1,3, Neha Kapate1,2, Blair Rohring1,2, Shridhar Singh1,2


Department of Biomedical Engineering1, Department of Chemical Engineering2, Department of Material Science & Engineering3
Carnegie Mellon University – Pittsburgh, PA
CLINICAL NEED PROPOSED SOLUTION MARKET ANALYSIS & DEVICE COST
Hypoxemia: Inadequate supply of blood oxygen, • Estimated market for oxygen concentrators in Nepal: ~8 million patients
can lead to brain & organ damage1,2 Needs Statement: Construct a durable and o Based on the incidence of hypoxemia for adults in Nepal
• Nepal: 20-30% of adults affected, accounts for portable oxygen concentrator that provides oxygen Parts Price Quantity Total Price
1/3 of hospital visits3 at 85% purity. Solenoid Valve 6 $125.28
$20.88
Hypoxemia in Low-Resource Settings
Polyethylene Tubing $0.25 10 $2.50
• Lack of oxygen supply for treatment
Piped Systems Goals Implementation Compressor $50 1 $50.00
• Expensive to implement and maintain4 • >85% oxygen purity • 2 cylinder Pressure Arduino Uno/Logic Board $30 1 $30.00
Oxygen Cylinders • >10 L/min output Swing Adsorption (PSA) Zeolite $30 1.3 $39.00
• Difficult to transport4
• <20 pounds, portable • Zeolite adsorbent Pressure Vessel $1.42 2 $2.84
Oxygen Concentrators
• Ideal WHO price: $450; Devilbiss 525 series: $5852 Total Cost $249.62

RESULTS
Single BedSingle-Bed
TestingSystem Testing Design of Electrical Implementation of Dual-Bed System
Table 1: Oxygen Purity vs. Time Circuit
35 Depressurization 1. Arduino controls relays
30 40 psi 2. Relays control solenoid
25 50 psi Trial 1 valves
50 psi Trial 2 3. Solenoid valves cycle to
% O2

20
50 psi Trial 3 control PSA
15
4. Valves powered through
10 transformer
5
0 Figure 1: Schematic of circuit
Figure 2a: Maximum oxygen purity achieved (left)
0 20 40 60 80
Time (s) Figure 2b: Fully wired dual-bed system with PSA (right)

CONCLUSIONS FUTURE WORKS ACKNOWLEDGEMENTS REFERENCES


• Identified LiX zeolite to be best option for nitrogen Next Priority The OxyGen BME Design team would like to thank 1. “Hypoxemia.” Mayo Clinic, Mayo Foundation for Medical Education and Research, 25
Dec. 2015, www.mayoclinic.org/symptoms/hypoxemia/basics/definition/sym-20050930.
adsorption • Increase purity of oxygen output Dr. Conrad Zapanta for his guidance and mentorship 2. Samuel, Jacob, and Cory Franklin. “Hypoxemia and Hypoxia.” Common Surgical Diseases,
pp. 391–394., doi:10.1007/978-0-387-75246-4_97.
• Designed circuit for controlling solenoid valves to >85% with this project, along with course TA Angela Lai for 3. Pendleton, Linda D. “When Humans Fly High: What Pilots Should Know About High-

• Progressed from single-bed purification system to • Use air filter to purify input air her continuous assistance. We would also like to Altitude Physiology, Hypoxia, and Rapid Decompression.” AVweb, AVweb, 9 Apr. 2003,
www.avweb.com/news/aeromed/181893-1.html.
thank the Undergraduate Resource Office for funding 4. “Elevation of Tanahun,Nepal Elevation Map, Topo, Contour.” Flood Map,
dual-bed system with pressure swing Long-term www.floodmap.net/elevation/ElevationMap/?gi=1282721.
this research. Additionally, we would like to thank
• Achieved 32.5% purity oxygen with dual-bed • Increase volumetric flow output Emily Reichert and Professor Vishwa Shrivastava for
5. Sah, Hd. “Study of Predictors of Hypoxemia in Children with Pneumonia.” Journal of
College of Medical Sciences-Nepal, vol. 9, no. 4, 2014, doi:10.3126/jcmsn.v9i4.10230.
pressure swing system of oxygen to accommodate providing context about resources in Nepal, and Matt
6. PhD, Charles Patrick Davis MD. “Hypoxia vs Hypoxemia: Click for Signs & Causes.”
MedicineNet, www.medicinenet.com/hypoxia_and_hypoxemia/article.htm.
• Optimal cycle time: 8 seconds multiple patients Cline and Tim Alpert for assisting with lab work, all of
• Optimal starting pressure: 50 psi • Develop portable casing which helped bring this project to fruition.