Академический Документы
Профессиональный Документы
Культура Документы
Christopher Maurino
Attenuation Project
Objective:
The objective of this project is to understand the principle of attenuation of the radiation
beam as it passes through an object and to calculate the transmission factor appropriately, which
Purpose:
Any object in the path of a radiation beam causes the radiation beam to lose energy as it
transverses the matter in that object.1 As a result, physical wedges were developed in order to
attenuate the radiation beam in order to optimize the dose given to the target and optimize dose
distribution by tilting the isodose lines in a desired orientation.2 Physical wedges are high density
pieces of metal that have a thick portion on one side, called the heel, and tapers to a thin portion
on the other side, call the toe. The toe of the wedge attenuates less of the radiation; therefore,
more radiation passes through the toe and results with the wedge changing the orientation of the
dose distribution, depending on the thickness of the heel. With the development of multileaf
collimators (MLC’s) in modern linear accelerators, Enhanced Dynamic Wedges (EDW) were
created. EDW use the MLC’s to slowly move across the treatment field in order to alter the dose
distribution the same way a physical wedge would (P. Archer CMD, personal communication,
February 12, 2020). Each angled wedge or corresponding EDW alters the isodose lines
differently, therefore, each has a different wedge transmission factor. This paper aims to find the
2
Wedge Transmission Factor (WF) for a 25-Degree EDW by finding the dose rate with the wedge
Methods
The dose was measured on a Varian TrueBeam Edge utilizing the Exradin A12 ion chamber
connected to a Innovision Model-35040 electrometer with and without the 25-degree EDW. The
set-up for the experiment used solid water and was standard for the machines, 90 SSD, 100 SAD,
10 cm depth with 10 cm of back scatter, as shown in image 1 and image 2. The field that was
analyzed was a single antero-posterior (AP) field from a plan that I created for treatment using a
6 MV photon beam. The field size was 13.0 x 12.4 cm with a blocked field of 12.5 x 7.0 cm field
with a collimator angle at 90 degrees. The dose rate that was given was 445 monitor units (MU).
The machine was calibrated to 0.8 cGy/ monitor unit at 10 cm depth, 100 SAD on a 10x 10 field.
Image 1: Solid water set up, 90 SSD, 100 SAD with 10 cm depth and 10 cm backscatter.
3
Image 2: Solid water set up, gantry angle 0 degrees, collimator angle 90 degrees couch angle 0
degrees.
Results
Three measurements were taken with the EDW and without the EDW after delivering 445
monitor units, results are displayed in table 1. The mean charge of the EDW field was -56.53 nC
and the mean charge without the EDW was -68.546 nC. As described above, our wedge
transmission factor (WF) is Dose with wedge/Dose without wedge. Therefore, our WF is -
56.53/-68.546 = 0.825. The equivalent square for the treatment field was approximately 13 cm
4
and the equivalent square for the blocked treatment field was approximately 9 cm (see image 3
for calculations). Dose to the ion chamber was 21% higher without the EDW compared to with
the EDW, 281 cGy to 232 cGy, respectively (see image 3 for calculations). The beam data table
used for hand calculations was provided in our DOS-522 class and was produced by Varian
Clinac 21X (S/N 2081) for 6 MV Tissue Maximum Ratio Dosimetry.3 Finally, image 4 shows
the hand MU calculation of the AP field of the treatment that we simulated in this experiment.
The MU calculated is 441 MU while the treatment planning system provides 445 MU, a
difference of 0.9%.
Discussion
Wedges are invaluable to radiation therapy because of the ability to change the dose distribution
of the radiation that the patient receives. Wedges can be used to avoid high doses to unwanted
areas, or they can be used as compensators for when the patient has to receive dose to variable
depths. They are a great asset to have in a Dosimetrists toolbox when trying to create a treatment
plan. The advent of the EDW eliminates errors from happening in the clinic and has improved
the efficiency compared to physical wedges. For instance, physical wedges must be placed in the
machine, and if forgotten, the patient would be treated to a much high dose. As calculated in this
experiment, the patient would receive 21% more dose if there was no wedge; as we can see from
the two dose distributions in images 5 and 6. These images are shown from the treatment
planning system dose distribution of an AP field with a 25-degree EDW and without the wedge
after giving 445 monitor units. However, with the EDW, radiation therapists are not required to
enter the room and add a physical wedge.4 When an EDW is used, the computer system
6
automatically engages the wedge, thus, it cannot be forgotten. Therefore, in order for the patient
to receive the higher dose of forgetting the wedge, there must be a malfunction with the machine
or operating system, which would be very unlikely. In addition, EDW are more efficient than
physical wedges provided that EDW have a much lower wedge transmission factor (WF). For
example, this experiment showed that the WF of a 25-degree wedge was 0.825, while in the
beam tables that were provided to us in our class, the WF for a 15-degree wedge and 30-degree
wedge is 0.708 and 0.551, respectfully.3 That means that in order to give the same amount of
dose, you have to give approximately 50% more monitor units when using a 30- degree physical
wedge compared to the 25-degree EDW. As a result, by using a EDW, the patient is exposed to
Image 5: Dose distribution of AP field of humerus with 25-degree EDW giving 445 MU.
7
Image 6: Dose distribution of AP field of humerus without 25-degree EDW giving 445 MU.
Conclusion
Every object that is placed in the path of a radiation beam, attenuates that beam. Wedges are
used to purposefully attenuate the beam in order to achieve an optimized dose distribution in
order to efficiently treat the patient. In this experiment, we found that a 25-degree EDW had a
WF of 0.825. This experiment has shown that the WF is an important component of a dose
calculation since omitting the WF would increase the patient’s exposure to radiation by 21%.
Finally, there are two main advantages of the EDW compared to physical wedges. Firstly, an
EDW does not require a therapist to physically add the wedge, resulting in the reducing clinical
errors. Secondly, the EDW is more efficient because the EDW requires less monitor units in
Acknowledgements
Thank you, Charlie Matrosic (physics Ph.D student at the University of Michigan), for being my
mentor and taking the time to set up this experiment with me.
8
References:
and beam profiles for physical and enhanced dynamic wedges. Journal of medical
https://uwlac.instructure.com/courses/278253/pages/all-weeks-
4. Khan FM, Gibbons JP. The Physics of Radiation Therapy, 5th ed. Philadelphia, PA: