Вы находитесь на странице: 1из 5

COLUMN WATCH

The Cleaning and


Regeneration of
Reversed-Phase HPLC Columns
Ronald E. Majors, Agilent Technologies, Wilmington Delaware, USA.

This month’s “Column Watch” looks at practical ways to return a contaminated


column to — or close to — its original state. Ron Majors also discusses cleaning
procedures for bonded-silica and other types of reversed-phase columns.

`Reversed-phase chromatography is by far phenyl–hexyl), endcapped and non- short-chain-bonded phases such as C2 or
the most widely used technique in high endcapped varieties, and polar-embedded C4 phases.
performance liquid chromatography phases — also exist within these bonded Users must be aware of the surface
(HPLC).1 It is popular because it is applicable silicas. Various other packing materials characteristics of their particular stationary
to most non-polar analytes and to many have been used in reversed-phase phases and of possible analyte–surface
ionizable and ionic compounds. Most of chromatography, including polymers, interactions, so they can take into account
the stationary phases used in reversed- polymer-coated silicas and aluminas, possible matrix interactions when they are
phase chromatography are hydrophobic in inorganic–organic hybrids, coated zirconia developing and using reversed-phase
nature. Therefore, analytes are separated and graphitized carbon. Each type of phase methods. For example, very hydrophobic
by their degree of hydrophobic interaction has its own advantages and disadvantages. sample matrices such as corn oil, highly
with the stationary phase and matrices Reversed-phase chromatography columns aromatic materials and waxes can stick to
with hydrophobic components can also be are used in various applications with a wide reversed-phase packing surfaces and change
retained in a similar manner. variety of mobile phases and additives. their characteristics. Biological fluids
Table 1 lists the most popular stationary Some of these techniques use additives containing proteinaceous materials can
phases usually bonded to silica gel.1 Phase that can change or modify the surfaces of adsorb on packing surfaces. Despite analysts’
subspecies — such as mixed phases (e.g., the packing material. Sometimes, these best attempts to protect HPLC columns
additives themselves may contaminate the from foreign substances, eventually certain
surface or bonded phase. analyte–matrix combinations can affect
Table 1: Relative use of stationary As with hydrophobic bonded phases, the stationary phases detrimentally.
phases in HPLC* surface of silica-gel packings has other After a column is contaminated, its
Relative chemical features. Residual silanols are present
Phase Usage (%) on the surface of all silica-gel bonded-phase
packings. Figure 1 depicts the different types Figure 1: Types of silanols on the
C18 (octadecylsilane) 39
of silanols that can be present.2 Being surface of silica gel (2).
C8 (octyl) 26
weakly acidic in character, these silanols can
Cyanopropyl† 14.5 interact with certain analytes and matrix OH OH OH OH OH
Phenyl 12 components, particularly with basic
C4 (butyl) 3.7 compounds. As the pKa of the silanol is Si Si Si Si
roughly 4.5, ionization can occur at Free Geminal Associated
Hydrophobic interaction 1.8
intermediate pH values, and thus the silanol silanols silanols
C2 (ethyl) 1.1 possibility of electrostatic interactions with
C1 (methyl) 0.8 Decreased acidity
cationic species exists. The older Type A
Other 0.8 silicas can contain high concentrations of OH
metal ions (sometimes 100 ppm or more)
Polymers 0.5  
that impart even greater acidity to the silica
* Adapted from reference 1 and normalized. M M Si
surface and can also interact with metal
† Includes normal-phase usage, because reversed- Surface metal Internal metal
phase versus normal-phase chromatography use
chelating or scavenging compounds.3
(most acidic)
was not queried. Residual silanols are more bothersome on
non-endcapped bonded silicas and on

2 LC•GC Europe July 2003


Column Watch

chromatographic performance can be contamination results from the accumulation very non-polar (e.g., ethyl acetate or even a
different from that of an uncontaminated of strongly retained substances from hydrocarbon), which helps to solubilize
column. Contaminated columns can exhibit repeated injections, a simple washing non-polar substances such as lipids and oils.
back-pressure problems. A contaminated process to strip these contaminants can It is important to ensure that each solvent
reversed-phase column must be cleaned often restore column performance. in the series is miscible with the next solvent.
and regenerated to return it to its original Sometimes, after isocratic operation, At the conclusion of the wash cycle, go
operating condition. This instalment of flushing a column with 20 column volumes backwards through an intermediately
“Column Watch” will discuss practical ways of 90–100% solvent B (the stronger miscible solvent before returning to the
to return a column to or nearly to its original solvent in a binary reversed-phase system) original mobile-phase system. For example,
state. Because bonded-silica columns are can remove the contaminants. (Table 2 lists isopropanol is an excellent solvent for this
the most popular, I will focus on them. At the column volumes for various sizes of intermediate step because it is miscible
the end, I will discuss cleaning procedures HPLC columns, so readers can easily with organic solvents such as hexane or
for other types of reversed-phase columns. determine a flush volume for a particular methylene chloride and also is miscible
by washing a column with non-aqueous with aqueous solutions. Because isopropanol
What Causes Contaminant Build- solvents such as methanol, acetonitrile or is quite viscous, make sure that the flow-
up in Reversed-Phase Columns? tetrahydrofuran. If you are using a buffered rate is not high enough to cause pump
Usually, sample matrices contain compounds aqueous mobile phase, do not jump overpressure. Also, if using a UV detector,
that are of no interest to analysts. Salts, immediately to the strong solvent. An avoid using solvents that absorb in the
lipids, fatty compounds, humic acids, abrupt change to high organic solvent ultraviolet region of the spectrum because
hydrophobic proteins and other biological content could result in buffer precipitation it may require a great deal of washing to
compounds are a few of the possible in the HPLC flow system, which could cause remove all of the absorbing solvent to get
substances that can come in contact with even bigger problems such as plugged frits, a stable baseline.
an HPLC column during its use. These plugged connecting tubing, pump seal A recommended column washing system
materials can have lesser or greater failure, a scratched piston or injection valve for a typical bonded-silica column and a
retention than the analytes of interest. rotor failure. Instead, flush the column with mobile phase without buffer salts is to use
Those compounds that have lesser a buffer-free mobile phase (that is, replace • 100% methanol,
retention, such as salts, will usually be the buffer with water). After flushing with • 100% acetonitrile,
eluted from the column at the void volume. 5–10 column volumes, the stronger solvent • 75% acetonitrile–
These undesired interferences might be then can be passed through the column. 25%isopropanol,
observed by a detector and appear as Occasionally, the strong solvent • 100% isopropanol,
chromatographic peaks, blobs, baseline component of a mobile phase is insufficient • 100% methylene chloride and
upsets or even negative peaks. If sample to remove the column contaminants. A • 100% hexane.
matrix components are retained strongly stronger solvent or series of solvents will be When using methylene chloride or
on the column and if the mobile-phase necessary to clean the column. If the hexane, the column must be flushed with
solvent composition itself never becomes contaminants are non-biological, then users isopropanol before returning to an
strong enough to elute them, these can pass one or more additional organic aqueous mobile phase because of solvent
adsorbed or absorbed compounds will solvents through the column to remove the immiscibility. A minimum of 10 column
accumulate, usually at the head of the undesired compounds. The solvents and volumes of each wash solvent should be
column, after many injections. This solvent combinations that can be used are passed through a column. For 250 mm 
behaviour is often observed with isocratic numerous. Visit one or more column 4.6 mm analytical columns, analysts can
conditions. Sample compounds that are of manufacturers’ websites to see various use a typical 1–2 mL/min HPLC flow-rate.
intermediate retention can be eluted slowly recommended solvent systems. To return to the original mobile phase,
and appear as wide peaks, baseline Generally, all washing approaches follow chromatographers can usually skip going
disturbances or baseline drift. a similar pattern. The wash solvents used through the entire series in reverse order.
Sometimes the sorbed sample are increased in their solvent strength, Using isopropanol as an intermediate
components build up to levels high enough often ending with a solvent that could be solvent is recommended, followed by
that they begin to act as a new stationary mobile phase without buffer, then finally
phase. Analytes can interact with these with the starting mobile-phase composition.
impurities that contribute to the separation Table 2: Column volumes of analytical Tetrahydrofuran is another popular solvent
mechanism. Retention times can shift and columns that can be used for cleaning contaminated
tailing can occur. If sufficient contamination Column Size Void Volume columns. If users suspect severe fouling,
occurs, the column backpressure can build (mm  mm) (mL) they can mix dimethyl sulphoxide (DMSO)
up to intolerably high levels, which 250  4.6 2.5 or dimethylformamide mixed 50:50 with
overpressures the pump and can cause a water and pass them at flow-rates less
150  4.6 1.5
column to settle and create a void, than 0.5 mL/min. The successful
depending upon where the blockage occurs. 150  3.0 0.64 regeneration of a reversed-phase column
150  2.1 0.28 can be a time-consuming process, and
Washing Bonded-Silica Columns 50  4.6 0.50 solvent washings can be programmed into
The keys to rejuvenating a contaminated a gradient system for overnight operation.
30  4.6 0.30
HPLC column are knowing the nature of A question arises as to whether to reverse
the contaminants and finding an appropriate 15  4.6 0.15 the HPLC column during the washing
solvent that will remove them. When procedure. As most of the strongly held

www.lcgceurope.com 3
Column Watch

contaminants are usually at the head of the before resolution loss or elution of extraneous manufacturer to ensure that these solvents
column, reversing the column can shorten compounds, users tend to wait until they are compatible with the packing material.
the migration distance that the solubilized observe some unusual behaviour. However, Silica-based columns are usually compatible
contaminants must travel to exit the an increased build-up of contaminants will but organic polymer–based columns can
column. As far as the packed-bed stability make it more difficult to clean the column. swell or shrink with certain solvent
is concerned, most modern HPLC columns For this reason, if you know that you are combinations, and the performance could
have been packed at a considerably higher subjecting your reversed-phase columns to be affected.
pressure than the normal operating dirty sample matrices, I recommend cleaning As with the previous solvent series,
pressure; therefore, their beds should not your columns on a regular basis. The more ensure that the sets of solvents in Table 3
be disturbed by the reversed flow. However, frequent the cleaning, the less rigorous used in series are miscible. Propanol is a
if a top frit is of a higher porosity than a cleaning conditions you will need. good intermediate flush solvent. A minimum
bottom frit, this type of reversal could be of 20 column volumes should be used for
detrimental. For example, if the bottom frit Cleaning Protein Residues from each solvent system. Because some of the
is of 2 µm porosity, it is usually sufficient to Bonded-Silica Reversed-Phase solvent systems are quite viscous, the
contain column packing with an average Columns flushing flow-rates should be adjusted
particle size of 5 µm (with a 2 µm particle If biological materials such as plasma or accordingly to ensure that no
size distribution). However, a manufacturer serum build up on a reversed-phase column, overpressuring occurs. After cleaning a
will sometimes put a larger-porosity frit at chromatographers must use a somewhat column with the guanidine or urea
the top of the column to prevent plugging different cleaning process. In most instances, reagents, use a minimum of 40–50 column
with sample or mobile-phase particulates. neat organic solvents such as acetonitrile or volumes of HPLC-grade water to flush the
If the porosity of this frit is larger than that methanol do not dissolve peptides and column.
of the smallest particles in the particle-size proteins and are ineffective for cleaning For reversed-phase columns, it has been
distribution curve, some of the packing reversed-phase columns. However, inadvisable to use detergents such as
material could conceivably pass through mixtures of organic solvents with buffer, sodium dodecyl sulphate (SDS) and Triton,
the frit and be removed from the column, acids and, sometimes, ion-pairing reagents because these compounds are apparently
thereby creating a void. If a column has an can be effective. Initially, flushing a column adsorbed strongly on bonded-silica packing
arrow to recommend the direction of flow, with mobile phase that has a somewhat and are difficult to remove. Using detergents
I would consult the manual or instruction higher percentage of the stronger solvent can affect the surface of the packing and
sheet, the manufacturer’s website or the (solvent B) should be attempted. change its characteristics. However, a study
technical support group before reversing Freiser and co-workers4 found that by the Separations Group found that a
the column to make sure that it is a safe repeated up-and-down gradients between column contaminated by a protecting group
practice. Whether you reverse the column aqueous trifluoroacetic acid and and scavenger products from a peptide
or not, it is a good practice to disconnect trifluoroacetic acid–propanol can regenerate synthesis could be cleaned by injecting 500
the column from the HPLC detector so that contaminated reversed-phase columns. µL of 1% SDS solution into the mobile
contaminants or particulates lodged on the Bhadwaj and Day5 suggested that a plug phase flowing at 1 mL/min.7 If followed by
frit are not swept into the detector cell, injection of 100 µL of trifluoroethanol in a a gradient from 5% to 95% acetonitrile
where they can cause contamination. 250 mm  4.6 mm column could work. If with 0.1% (v/v) trifluoroacetic acid and
The frequency of cleaning fouled these procedures fail, the strong eluents or equilibration at the starting conditions, the
reversed-phase columns depends upon solubilizing agents recommended by polypeptide separation was restored.
how much unretained material has been Cunico and colleagues6 can be used to
injected onto the column. Because strip proteins (see Table 3). Before flushing Special Techniques for Cleaning
reversed-phase columns can sometimes columns with these solvents, however, Bonded-Silica Reversed-Phase
withstand a great deal of contamination consult the column manual or the Columns
Sometimes, washing with organic solvents
can fail to remove the column
Table 3: Wash solvents for removing proteinaceous material from HPLC contaminants. This situation is particularly
reversed-phase columns* true if metallic ions are sorbed to the silica
Solvent Composition or bonded phase. A chelating reagent such
Acetic acid 1% in water as 0.05 M ethylenediaminetetraacetic acid
(EDTA) can be flushed through a column.
Trifluoroacetic acid 1% in water The EDTA complexes with many metallic
0.1% Trifluoroacetic acid–propanol 40:60 (v/v) (viscous; use reduced flow-rate) species and solubilizes them. After
treatment with an EDTA solution, analysts
TEA–propanol 40:60 (v/v) (adjust 0.25 N phosphoric acid to can wash the column thoroughly with
pH 2.5 with triethylamine before mixing) water. If the sample matrix contains
Aqueous urea or guanidine 5–8 M (adjusted to pH 6–8) ionizable compounds, a change of pH
could put them into an un-ionized form,
Aqueous sodium chloride, sodium 0.5–1.0 M (sodium phosphate pH 7.0) and they could be flushed from the column
phosphate or sodium sulphate
with water–organic solvent mixtures. For
DMSO–water or dimethylformamide–water 50:50 (v/v) example, a strongly basic matrix
*Adapted from reference 6. component can sometimes be removed by
adjusting the pH to less than 3, at which

4 LC•GC Europe July 2003


Column Watch

the protonated amine becomes more behaviours. Bidlingmeyer and co-workers10 flushing the column in a reversed direction
water soluble. Acidic matrix components demonstrated that when using a C18 with 10 column volumes each of 1.0 M
can be removed by adjusting the pH to a column with mobile-phase concentrations sodium hydroxide, water, 20% ethanol
higher value — greater than the pKa — greater than 70% methanol, SDS, which is solution and the working buffer.12 For
approximately pH 8 or 9, at which the a long-chain anion-pairing agent, is not more hydrophobic proteins, users should
acids are in their ionized form. However, be adsorbed onto the stationary phase. This insert a wash step of isopropanol (30% v/v)
cautious with silica-based columns because finding agrees with the Separations Group or 70% ethanol (v/v) after the water wash.
they can be damaged by long-term work.7 For sanitization or the inactivation of
exposure to high pH levels.8 Bonded-silica monolith columns such as microbial populations, a PS–DVB
To control bacterial growth that could be Chromolith columns (Merck KGaA, monolithic column should be thoroughly
present in a buffer system or in columns Darmstadt, Germany) should be treated as washed with 0.5–1.0 M sodium hydroxide.
left unattended in aqueous buffer, any other silica-based columns. The packed monolith column should be
chromatographers can use common exposed to sodium hydroxide for at least
household bleach diluted 1:10 or 1:20. Run Regeneration for Polymeric 1 h at room temperature.
at least 50 column volumes followed by Columns Columns with traditional polymer-based
another 50 column volumes of HPLC-grade Polymeric columns used to separate biological packing materials used to separate difficult
water. Do not run the bleach through the molecules can also become contaminated proteins such as membrane proteins,
detector, because it could attack the or require sanitization. The chemical structural proteins and viral-coat proteins
flowcell. To prevent bacterial growth in the stability of polymeric materials is generally that are very insoluble require harsh
solvent reservoir, use just enough buffer for considered one of their strengths. In fact, cleaning conditions. For example, a solvent
the day’s use and store unused buffer in many manufacturers recommend washing such as 50% isopropanol with 3 M
the refrigerator, use 0.1% sodium azide in their columns with 1.0 M nitric acid or 1.0 guanidine hydrochloride at 60 °C might be
the buffer, and don’t let the column sit in M sodium hydroxide. Certain reversed- necessary to elute such difficult proteins.13
buffer solution for long periods of time phase polymeric columns such as those The cleavage of synthetic peptides from
without any flow. packed with poly(styrene–divinylbenzene) solid-phase resins generates reactive
Chromatographers have frequently (PS–DVB) beads and polymeric monoliths carbonium ions that are scavenged by
discussed the effect of the ion-pairing such as CIM RP-SDVB discs (BIA anisole and thioanisole. The scavenger-
reagents on the stationary phase for Separations, Ljubljana, Slovenia) and Swift carbonium reactions yield large, aromatic
columns used for ion-pairing columns (Isco, Lincoln, Nebraska,USA) can molecules that can foul reversed-phase
chromatography. Apparently, ion-pairing withstand a wide range of pH values columns during peptide purification. These
reagents such as octanesulphonic acid (usually pH 1–13 or sometimes pH 0–14), contaminants are highly retained on C18
(used for cations) and tetraalkylammonium but users should exercise some care when columns and cannot be removed with
bromide (used for anions) strongly sorb on washing these columns with harsh organic 100% acetonitrile or methanol. To clean
the surfaces of bonded-silica columns at solvents. Depending upon their degree of these columns, reverse them and wash
certain concentrations of organic modifier. cross-linking, swelling or shrinking can them with three to five column volumes of
The columns become contaminated and occur when the columns are exposed to 100% isopropanol, three to five column
cannot be regenerated to their original some organic solvents. Highly cross-linked volumes of methylene chloride, three to
state, and the story goes that any column polymers with greater than 8–10% cross- five volumes of isopropanol, and then the
used for ion-pairing work should be linking usually have good mechanical original solvent system.14 The elution of
dedicated to that technique and never stability with minimal shrinking in aqueous aromatic impurities can be verified with a
used again for regular reversed-phase solvents and minimal swelling in organic UV detector at 260 nm.
chromatography. solvents. Before washing a polymer column
Bidlingmeyer9 disagrees with this with a series of solvents, however, it is a Regeneration of Zirconia-Based
generality and feels that the aggressive pH good practice to consult the column’s HPLC Columns
values used for the ion-pairing coupling manual or contact the technical support ZirChrom Separations, Inc. (Anoka,
can actually change the nature of some group of the column’s manufacturer. Minnesota, USA), has manufactured a
columns by either hydrolysis of the bonded According to BIA Separations,11 users series of zirconia-based columns. The
phase or endcapping silane under acidic can regenerate a polymer-based monolithic product line has several reversed-phase
conditions (pH 1–3) or by silica dissolution column made of PS–DVB by columns, including polybutadiene,
at higher pH values (pH 7–8). To remove • washing the column with 10 column polystyrene and graphitized carbon
sulphonic acid ion-pairing reagents, he volumes of 0.1% trifluoroacetic acid in versions. Zirconia is more pH resistant than
recommends first washing the column 2-propanol at one-half of the working silica gel, so coated zirconia should be
(minimum of 20 column volumes) with the flow-rate more rugged and able to withstand
same mobile phase without the ion-pairing • washing the column with at least 5 harsher conditions such as higher pH and
reagent and then washing with mobile column volumes of 100% mobile phase operating temperatures. However, due to
phase without the buffer (methanol might B at one-half of the working flow-rate the special surface characteristics, analysts
be a better organic solvent than • re-equilibrating the column with at least must maintain certain experimental
acetonitrile in this wash step; for very 10 column volumes of 100% mobile conditions to use these columns
long–chain ion-pairing reagents, use phase A at the working flow-rate. successfully with various analytes.
tetrahydrofuran). Apparently, sulphonic If a methacrylate-based monolith with Carboxylic acids, fluoride and phosphate
acid ion-pairing reagents and amine butyl or ethyl chemistry is cleaned, ions all adsorb strongly to zirconia-based
ion-pairing reagents exhibit different precipitated protein can be removed by columns. To remove them from a column,

www.lcgceurope.com 5
Column Watch

“Basic HPLC and CE of Biomolecules,” Bay


flush the column with 50 column volumes Bioanalytical Laboratory (Richmond, California,
of a mixture of 20% acetonitrile–0.1 M 1998), p. 254.
sodium hydroxide or 0.1 M 7. “Reversed-Phase Column Cleaning: SDS is
OK,” Vydac Advances, Grace/Vydac (The
tetramethylammonium hydroxide, 10 Separations Group) (Hesperia, California,
column volumes of water, 50 column Winter 1998).
volumes of 20% acetonitrile–0.1 M nitric 8. R.K. Iler, The Chemistry of Silica (John Wiley &
Son, New York, 1979), p. 639.
acid, 10 column volumes of water again,
9. B.A. Bidlingmeyer, J. Chromatogr. Sci. 38, 226
and 20 column volumes of 100% of (2000).
organic solvent. For the polybutadiene and 10. B.A. Bidlingmeyer, F.V. Warren Jr. and R.A.
polystyrene columns, chromatographers Grohs, Anal. Chem. 63, 384–390 (1991).
11. “Product Specific Information Sheet for CIM
can use methanol, acetonitrile, isopropanol RP-SDVB Disk,” publication number PSIS-
or tetrahydrofuran. The graphitized carbon RPSD-0502, BIA Separations (Ljubljana,
column will require the same solvents with Slovenia, 2002).
12. Specific Information Sheet for CIM C4 Disk,”
at least 20% tetrahydrofuran.15
publication number PSIS-C4D-0702, BIA
Separations (Ljubljana, Slovenia, 2002).
Conclusions 13. “A New Chemically Resistant and Heat-Stable
Reversed-phase HPLC columns can become Reversed-Phase Column for Protein and
Peptide Separations,” Vydac application note
contaminated by the repeated injection of number 9702, Grace Vydac (Hesperia,
samples that contain strongly retained California, February 1997).
substances in their matrices, especially 14. “Cleaning of Reversed-Phase Columns Used
for Synthetic Peptides,” Vydac application
compounds that are of high molecular note number 9603, Grace Vydac (Hesperia,
weight or are very hydrophobic in nature, California, March 1996).
biological fluid components such as 15. “Recommendations for Use, Cleaning and
Storage of Zirconia-Based HPLC Columns,”
proteins, and strongly basic compounds
ZirChrom Separations, Inc. (Anoka, Minnesota,
that can adsorb on silanol groups. In 2002).
addition, certain mobile-phase additives
such as ion-pairing reagents and surfactants
can sorb onto packing surfaces and change “Column Watch” editor Ronald E. Majors
their nature. Contaminated columns can is the business development manager,
generate poor peak shapes, non-reproducible consumables and accessories business unit,
retention, high back pressures and baseline Agilent Technologies, Wilmington,
artifacts. Often, washing with organic Delaware, USA and is a member of LC•GC
solvents and reagents that can break the Europe Editorial Advisory Board.
strong interactions with the bonded phase, Direct correspondance about this column
or underlying silica surface can clean these to LC•GC Europe, Advanstar House,
columns. Sealand Road, Chester CH1 4RN, UK, e-
Chromatographers should exercise some mail dhills@advanstar.com
caution because using reagents that are
too harsh can damage the bonded phase
itself. Furthermore, it is good practice to
prevent column contamination by using
sample preparation procedures that will
minimize column exposure to undesirable
substances. For all applications, I highly
recommend using guard columns to
prevent contamination and possible
damage to the analytical column.

References
1. R.E. Majors, LCGC 15(11), 1008–1015 (1997).
2. L.R. Snyder, J.J. Kirkland and J.L. Glajch,
Practical HPLC Method Development (John
Wiley & Sons, New York, 2nd ed., 1997),
p. 180.
3. L.R. Snyder, J.J. Kirkland and J.L. Glajch,
Practical HPLC Method Development (John
Wiley & Sons, New York, 2nd ed., 1997),
pp. 181–182.
4. H.H. Freiser et al., “The Use of Stabilized Silica
Support with a Short Alkyl Chain for Reversed-
Phase Preparative Chromatography of
Proteins,” product bulletin number 10, Micra
Scientific (Northbrook, Illinois, 1988).
5. S. Bhardwaj and R.A. Day, LCGC 17(4),
354–356 (1997).
6. R.L. Cunico, K.M. Gooding and T. Wehr,

6 LC•GC Europe July 2003

Вам также может понравиться