Technical specification

Dimensions 300W x 490D x 1,480H mm (Without IV pole)


Weight Approx. 100kg
Supply AC110V,220V,230V± 10 %, 50 Hz/60 Hz
Power
Single patient dialysis machine
2.50kVA or less (AC220V,AC230V)
Consumption
1.65kVA or less (AC110V)
Class I Type B applied part (Hydraulic part)
Class and grade
Protection Type BF applied part (Cuff for BPM)
Against water penetration IPX1
Pressure 0.05 to 0.74 MPa
Average flow 900 mL/min or more
Water supply 2 to 30°C (With the Heat Exchanger Assy)
Temperature
15 to 30°C (Without the Heat Exchanger Assy)
Water quality Online HDF treatment should be satisfied the provision of AAMI RD52: 2004.
Flow 100 to 800 mL/min
Temperrature 32.0 to 39.0 °C
Dialysate
Conductivity (BICARBONATE) 125~165 mmol/L (Bicaronate:24~70 mmol/L)
Conductivity (ACETATE) 125~165 mmol/L
Method Volumetric control with piston pump
UF Control UF rate 0.00, 0.10 to 5.00 L/h
Accuracy ± 30 g/h
Chemical / Heat / Hot citric
Disinfection Method
IHR (Integrated Hot Rinse)
10 to 400 mL/min (φ6.35×φ 9.75 tube)
Flow range
Blood pump 10 to 600 mL/min (φ8.00×φ12.00 tube)
Flow accuracy Set value ± 10 %
Substitution pump 10 to 300 mL/min (φ6.35×φ 9.75 tube)
(For Online HDF) Flow range 10 to 500 mL/min (φ8.00×φ12.00 tube)
Syringe size 10mL, 20mL, 30mL
Flow setting 0.0 to 20.0 mL/h
Heparin pump
Fast-forward injection Approx. 550/900/1,200mL/h : 10/20/30mL
Flow accuracy Flow accuracy ± 10 %
Sensor method Ultrasonic
Bubble detector Corresponding tube φ3.4×φ5.6 / φ4.4×φ6.6 / φ4.6×φ6.8 ±0.1mm
Detection capability 0.3μL bubbles for being cumulated in 5 min (Flow 200 mL/min)
Sensor method Optical
Blood leak detector
Detection value 50 to 500 ppm
Measurable range - 500 to + 500 mmHg
Venous pressure
Indicated value accuracy ±10 mmHg
Measurable range -500 to +500 mmHg
Arterial pressure
Indicated value accuracy ±10 mmHg
Measurable range -500 to +500 mmHg
TMP
Indicated value accuracy ±10 mmHg
NIBP
Remote SW for NIBP
ETRF leak check uni t (For CF1)
Bicarbonate Powder Assy Factory option
Central supply system 2 ports. 2 kind of A solution or A and B solution.
Heat Exchanger Assy
Auto priming and Auto blood return Factory option.
Option External network
Online HDF Factory option
Drain port
Nurse call
IC card Factory option
Solution tank tray for dialysate
Solution tank tray for disinfectant

Manufacturer :

NIPRO CORPORATION
3-9-3, Honjo-Nishi, Kita-ku, Osaka, 531-8510 Japan NIPRO MEDICAL CORPORATION NIPRO TRADING (SHANGHAI) CO., LTD
Tel:+81-6-6372-2331 Fax:+81-6-6371-7422 3150 N.W. 107th Avenue, Miami, FL33172, U.S.A. Room 2001, 2015, 2016, Tower B, City Center of Shanghai,
http://www.nipro.co.jp/ Tel:+1-305-599-7174 Fax:+1-305-599-8454 100 Zunyi Road, Changning District, Shanghai 200051, China
Tel:+86-21-6237-0606 Fax:+86-21-6237-0186
Overseas Office :
NIPRO EUROPE N.V NIPRO ASIA PTE.LTD NIPRO SALES (THAILAND) Co.,Ltd
Weihoek 3H, B-1930, Zaventem, Belgium 8 Temasek Boulevard, Suntec tower 3 #40-02, 2nd Floor, Juthamard Building, 89/170,Moo 3, Viphavadee Rangsit Road,
Tel:+32-2-725-5533 Fax:+32-2-725-7041 Singapore 038988 Talad Bangkhen, Laksi, Bangkok 10210 Thailand
http://www.nipro-europe.com/ Tel:+65-6588-2281 Fax:+65-6588-2481 Tel:+66-2-551-2000 Fax:+66-2-551-2006
Optimized treatment for patient & medical staff

Flexibility Automatic priming & Automatic blood return


HD double needle by saline or Online dialysate*
HD single needle HD model uses saline.
HDF double needle; pre or post dilution* HDF model uses saline or dialysate.
HDF single needle; pre or post dilution*
HF; pre or post dilution* Automatic priming
Dialysate flow; 100ml/min to 800ml/min Automatic blood return
SQUF (Sequential UF) Automatic drainage
Bicarbonate and sodium profiling
UF profile
2 central concentration connectors*
Advanced accurate Kt/V calculation "Dose finder"

User friendly
Hygiene and disinfection First aid button
Saline or Online dialysate can be injected to blood line.
Sheath-type dialysis concentration suction UFR, blood flow, substitution flow, dialysate bypass,
Arterial and venous drip chamber level control blood pressure measurement are controlled.
ETRF filter for dialysate, CF-609N* NIBP*
Leak check function for Online HDF filter, CF-609N* Automatic non-invasive blood pressure monitoring
Bicarbonate cartridge function* Systolic, diastolic, MAP and pulse are monitored.
Heat disinfection Individual control with links to UF and Sodium profiling
IHR (Integrated Hot Rinse) The turnable 10.4” inch screen
Automatic disinfection programs after treatment English, Spanish, Chinese (simplified Chinese, traditional Chinese) is available.
2 disinfection connectors 300mm width slim body
Guidance for error message

IC card & Network system*


Treatment setting can be set by IC card system or Network system. Safety
IC card system Power failure backup
Patient treatment setting is written into NCU-18 via IC card which is written to Start up test
each card from PC beforehand. Closed circuit leak test during treatment
Network system
Patient treatment setting is written into NCU-18 via network cable from PC.

Reduce the labor burden