Вы находитесь на странице: 1из 21

IPM Safety

Table of Contents
Table of Contents ........................................................................................................ 1
IPM Safety ................................................................................................................... 2
Electrical and mechanical devices ........................................................................... 2
Lockout/tagout requirements ................................................................................... 3
Compressed gases .................................................................................................... 5
Lasers ....................................................................................................................... 6
Radiology equipment ............................................................................................... 7
Personnel exposure limits..................................................................................... 8
Radiation measurement units ............................................................................... 8
Precautions during IPM ........................................................................................ 8
Infection control ....................................................................................................... 9
Basic infection control practices .......................................................................... 9
Decontamination ................................................................................................ 11
Mercury.................................................................................................................. 15
Health risks......................................................................................................... 15
Exposure monitoring .......................................................................................... 16
Federal regulation............................................................................................... 16
Mercury-spill cleanup and disposal .................................................................... 16
Managing mercury sources ................................................................................ 18
Sphygmomanometers ..................................................................................... 18
Bougies ........................................................................................................... 18
Infant incubators ............................................................................................. 19
Laboratory chemicals ..................................................................................... 19
Other sources .................................................................................................. 19
Recommendations for avoiding mercury contamination ................................... 19
References .......................................................................................................... 21

Page 1
IPM Safety
Safety in the workplace is the responsibility of under test, especially if the grounding is defective.
both the employer and the employee: each You can safely test receptacle wiring (e.g., with a
employer is obligated to provide a safe working three-lamp polarity tester) and measure line
environment, and each employee should strive to voltage and low-current ground resistance on
work in as safe a manner as possible. Inspecting branch circuits or receptacles in use. (See
medical devices involves some risks; to protect Electrical Receptacles Procedure 437 for further
test personnel, staff, and patients, observe the details.)
special precautions noted in many of the
procedures. When performing hazardous Tests of isolated power systems temporarily lower
inspections, post signs to warn visitors and staff. the protective barrier normally associated with
Do not leave hazardous test setups unattended. isolated power. If the receptacle has a
Follow manufacturer or other appropriate ground-fault circuit interrupter (GFCI), power to
guidelines concerning lasers, ionizing radiation, all devices on the line could be shut off. This
and chemical hazards. And use common sense. emphasizes the need to ensure that critical or life-
support equipment is never powered from a
Electrical and mechanical GFCI-protected circuit.
devices
Confirm that the outlet used to power the safety
Inspecting medical devices for electrical safety analyzer is correctly wired. Grounding the
involves proximity to voltages and currents that analyzer case through the receptacle ground
can cause injury to test personnel, staff, and prevents a shock hazard while testing a defective
patients. Deliberately simulating faults increases piece of equipment. A GFCI trip point
the need for caution. measurement often uses the analyzer ground for a
return path of the test current; if the analyzer is
Electrical safety tests intentionally simulate faults connected to an ungrounded receptacle, there may
that may be hazardous. Not only would a patient be line voltage (115/230 VAC) on the analyzer
be exposed to unnecessary risks, but test results case. Do not touch the case of the equipment
may also be misleading. Arrange with appropriate being tested, especially when measuring
clinical personnel to disconnect the device from ungrounded leakage current.
the patient, or have them advise clinical
engineering personnel when the device is no To avoid damaging the device being tested or
longer in use. interrupting a fuse or circuit in the device or
branch circuit panel, turn off devices with motors
Do not test electrical power distribution systems and compressors and wait until they completely
on which patient equipment is operating. Some stop before reversing hot-neutral polarity; turn off
receptacle ground integrity tests may inject and wait at least 10 seconds for microprocessor-
several amperes into the ground line and can controlled devices, including computer and
cause hazardously high current to flow through a clinical laboratory analyzers. Also, turn off power
patient who is connected to a device on the branch

Page 2
IPM Safety

to the device being tested if it is necessary to as well as injury from other energy sources, such
insert or remove modules. as mechanical (both stored, as in compressed
springs, and kinetic), pneumatic, hydraulic,
Lead isolation testing involves line voltage use. thermal, and chemical sources. Thus, a lockout
We recommend performing the test only with can—and should—be applied to a pressurized
equipment properly designed to allow safe steam valve and an electrical switch box, as well.
application of the voltage to the patient leads. If
an electrical safety analyzer is not available, Lockout is accomplished when all of the
devise or buy an adapter that limits the current technicians (maintenance personnel) servicing a
flow to 1 mA or less to ground from any exposed piece of equipment ensure that all power sources
terminals of the test setup. Do not touch patient to the device are turned off and apply a lock
leads during lead isolation testing, and avoid using (usually a keyed or combination padlock) to the
exposed clips or improvised test setups that make switch(es) before beginning work. This ensures
it easy to accidentally contact the input circuit. that the unit can be energized only when the last
Touching the leads may cause a shock even with a technician has completed work and has removed
current-limiting resistor in the test circuit. the lock. When a switch provides for only one
lock, a key box is provided into which the single
Testing pneumatic and mechanical devices can key for the lock can be placed; the box, in turn,
result in injuries such as crushed fingers, has multiple lock holes so that each technician can
lacerations, and punctures and in hazards such as place a lock on the key box.
flying projectiles. When inspecting devices with
cams, gears, levers, sliding components, or other Tagout is a comparable, though less secure,
moving parts (e.g., blood pumps, electric beds, x- procedure. Under OSHA regulations, it may be
ray film processors), keep fingers and clothing used where providing lockout is not feasible and
away from moving parts. Perform parts where adequate procedures that provide the
inspection, cleaning, and lubrication with power necessary protection are in place and properly
disconnected. Do not get underneath electric beds, understood. A durable and prominent tag, with
patient lifts, radiology systems, or similar devices appropriate information clearly displayed on it, is
while they are connected to a power source or placed on the disconnecting means at the time the
loaded. unit is deenergized. The tag warns others not to
turn the power on and is removed only by the
Lockout/tagout requirements technician after work is completed. The tag is
usually accompanied by some means (e.g., a
Lockout and tagout procedures are protective sturdy nylon wire tie) that impedes operation of
measures that usually include the use of security the disconnecting device.
devices such as padlocks applied to manual circuit
breakers and cutout switches or tags prominently Clinical engineering personnel commonly work
placed on such devices to warn of work in with line cord and plug-connected devices.
progress. Locks are placed so that electrical Servicing these devices, whether they are portable
service to a device will not be activated or mobile or operate in fixed locations, is
inadvertently by someone who is unaware of the exempted from lockout/tagout requirements
work in progress. provided that the line cord and plug are under the
exclusive control of the person doing the
Occupational Safety and Health Administration servicing. In these cases, lockout is accomplished
(OSHA) regulations codify requirements for simply by unplugging the device.
lockout and tagout procedures to reduce the risk
of injury from energized sources during Clinical engineers (CEs) and biomedical
maintenance and servicing (29 CFR 1910.147). equipment technicians (BMETs) should practice
The regulations protect against electrical mishaps lockout or tagout of hard-wired devices such as

Page 3
IPM Safety

radiology equipment, CT and MRI scanners, and  Perform service or maintenance.


film processors. Lockout/tagout also applies to
major installed appliances and systems (e.g.,  After completion of service or maintenance:
steam and ethylene oxide sterilizers, cart washers,  Replace guards, and remove tools.
blanket and solution warmers, hyperbaric
chambers). Likewise, facility engineering  Check area to be sure that no personnel
personnel should use these procedures on such are exposed.
equipment as air handlers, trash compactors,  Be sure that controls are returned to
incinerators, boilers, pumps, and elevator off/neutral position.
equipment. CEs and BMETs should respect and
pay attention to these procedures as they are  Remove lockout/tagout and related
encountered. Where isolation devices for such protective devices.
equipment lack lockout provisions, tagout  Operate isolating devices to restore
procedures must be used. New installation or energy to equipment.
major replacement, repair, renovation, or
modification of equipment must include The OSHA requirements also stress the need to
energy-isolating devices with lockout provision. protect against exposure to stored energy, which
can be released suddenly and unexpectedly. Some
Employers must develop documented of the activities that must be considered include
energy-source control procedures appropriate for blocking spring-loaded components that might
the equipment in their facility and ensure that accidentally be tripped, chocking mobile devices,
employees are trained in those procedures. immobilizing counterweights so they cannot fall,
Employers must also conduct inspections at least and discharging pressurized chambers or charged
annually to ensure that these procedures are capacitors. While steps taken to provide such
carried out and that requirements of the standard protection would not strictly be viewed as locking
are being met. or tagging out, they are, nevertheless, important to
implement.
Below is a checklist of typical procedures that
should be performed diligently within an The OSHA standard generally requires employers
institution to reduce the risk of injury and ensure to document specific procedures for the
conformity with OSHA’s requirements: equipment on the property. However, one
exception relieves the need to document
 Identify all energy-isolating devices that must lockout/tagout procedures for specific machines
be locked or tagged out. or equipment when a series of eight specific
conditions are met. The essence of these
 Carry out lockout/tagout procedure: conditions can be summarized as follows: The
 Notify affected personnel. device is supplied from a single power source that
can be totally isolated by a single lockout device
 Shut off equipment by normal means. and has no potential for releasing any form of
 Operate isolators (e.g., switches, valves, stored energy after shutdown; the employee
disconnects); relieve or protect against authorized to service the device has exclusive
stored energy (see below). control over the lockout device, and the
maintenance poses no risk to other employees;
 Lock out or tag out device. and the employer has had no accidents with this
device that resulted from unexpected activation or
 Operate normal controls to be certain that
reenergization during maintenance or servicing.
equipment will not operate.
 Be sure that controls are returned to The standard does not describe responsibility for
off/neutral position. or obligations to outside service personnel. It

Page 4
IPM Safety

seems reasonable, however, to assume that  Use commercial leak-detector solutions to


facilities are responsible for installing lockout detect gas leaks. Use only oxygen-compatible
capabilities on their equipment, whereas training leak-detector solutions on oxygen and nitrous
is the responsibility of the companies employing oxide systems.
these service personnel.
 Do not deface or remove any markings (e.g.,
Even with lockout and tagout procedures labels, decals, tags, stenciled marks) used to
implemented, CEs, BMETs, or other identify the contents of a cylinder.
engineering/maintenance personnel may find an  Never attempt to repair or alter cylinders,
open disconnect or other energy-control device stem valves, or indexing pins. Replace any
that is not locked or tagged out. Under no damaged cylinder with a new one, and return
circumstances should anyone close an open switch the damaged cylinder to the supplier.
or other energy isolator (with or without lockout
or tagout devices) without being absolutely certain  Never drop cylinders or permit them to strike
that it is safe and appropriate to do so. each other violently. Ensure that all cylinders
are securely mounted or chained so that they
cannot roll or fall during use or while in
Compressed gases storage.
Compressed gas cylinders must be handled
 Never drag, roll, or slide cylinders. Never use
carefully to avoid being contaminated or knocked
a cylinder valve as a handle to move a
about. High pressures can turn loose connectors
cylinder. Move larger cylinders, even for
into projectiles. Gas-driven particles can cause
short distances, with a suitable truck, making
ignition of downstream components. The
sure that the cylinder-retaining chain or strap
Compressed Gas Association's (CGA)*
is fastened in place.
Characteristics and Safe Handling of Medical
Gases provides recommended practices for  Do not use regulators, pressure gauges, or
handling medical gases, including many of the manifolds intended for use with a particular
following: gas or group of gases with cylinders
containing other gases.
 Ensure that the workspace, into which  Never interconnect medical gases without
compressed gases may be vented, is well appropriate check valves. They may become
ventilated with fresh air. contaminated by the feedback of other gases
or foreign material.
 Before disconnecting regulating devices from
cylinders, close the cylinder valve and release  Always use pressure-reducing regulators
all pressure from the regulator. Cylinder when withdrawing the contents of gas
valves should be closed at all times except cylinders, because regulators deliver a
when the gas is actually being used. constant, safe working pressure. Do not use
needle valves or similar devices without
 Never permit oil (e.g., from oily hands or
pressure-regulating mechanisms in place;
gloves), grease, or other combustible
excessive pressures may develop downstream
substances to come in contact with cylinders,
of such devices and may result in injury or
valves, regulators, gauges, hoses, or fittings.
damage to equipment.
Oil and certain gases (e.g., oxygen, nitrous
oxide) may combine with explosive violence.  Ensure that the threads on
regulator-to-cylinder valve connections or the
* pin indexing devices on yoke-to-cylinder
Compressed Gas Association, 4221 Walney Rd,
Chantilly, VA 20151-2923; (703) 788-2700. valve connections are properly mated. Never
www.cganet.com force connections that do not fit.

Page 5
IPM Safety

 After removing the protective valve cap, and room). Therefore, exercise great care whenever an
with the opening pointed away from all unfocused laser beam is accessible.
personnel, slightly and briefly open the valve
to clear the outlet of any dust and dirt. Do not Area security and use of service personnel
do this with cylinders containing flammable protective devices and practices should be
or toxic gas. consistent with facilitywide laser safety
procedures and/or approved by the laser safety
 When opening a valve, point the outlet away committee. Windows should be covered with
from all personnel. Never use wrenches or absorbing or laser-opaque material to prevent
tools except those provided or approved by transmission of laser energy into other areas.
the gas supplier; incompatible tools may (Window covers are not necessary with carbon
damage the valve. Use only nonferrous (e.g., dioxide lasers.) Wear appropriate laser safety
brass, aluminum) tools in the presence of eyewear at all times whenever the laser is in the
flammable or combustion-supporting gases to operating mode. WARNING: Verify that laser
prevent sparks. Never hammer the valve safety eyewear is appropriate for the specific
wheel when attempting to open or close the wavelength of laser being used (e.g., glasses
valve. appropriate for a Ho:YAG 2,100 nm wavelength
 Open the cylinder valve slowly, and keep it are not appropriate for Er:YAG lasers at 2,940
fully open when the cylinder is in use. nm). The American National Standards Institute
(ANSI) standard Z136.1-2007, Safe Use of
 Use secure fittings to prevent a tube or Lasers, calls for protective eyewear to be labeled
connector from suddenly disconnecting and with the wavelength for which protection is
whipping about or becoming a projectile. Do afforded. Laser safety eyewear may not protect
not use friction fittings (e.g., hose barbs, Luer the wearer from the aiming system light. Do not
slip) for 50 psi and higher circuits—use stare directly into the aiming system beam or the
threaded or positive-locking devices. therapeutic laser, even when wearing laser safety
eyewear. Eyewear should be inspected
Lasers periodically for physical condition and light leaks
Inspecting and maintaining lasers is a dangerous (see Section 4.6.2.8 of the ANSI standard). Avoid
but necessary process that demands far greater placing the laser beam path at eye level (sitting or
care than is required with most devices. U.S. Food standing).
and Drug Administration (FDA) medical-laser
incident reports reveal that many of the accidents Do not perform inspection and preventive
that occur during laser maintenance can be maintenance (IPM) procedures when a patient is
attributed to a failure to follow basic laser safety present or clinical staff is working. However, a
precautions. Personnel who inspect or service second person should be present to summon help
lasers should receive special training from the if an accident occurs while performing the
manufacturer or from a qualified alternative procedure. Do not aim the laser across a path that
training source. a person might normally use as a thoroughfare.
Post doors to the room with an appropriate laser
Laser energy can cause serious injury. This hazard safety sign that identifies that the laser is in use
is especially great when an interlock is overridden and that it is unsafe to enter the room without
or in any other situation where the energy does authorization by the service person performing the
not diverge significantly over long distances. procedure.
Under some circumstances, the beam may not
diverge significantly even a full room length or The laser should be kept powered off when not in
more from the laser (an unfocused laser beam use. When in use, the laser should be in the
from a mirrored articulating arm or from an exit standby/disabled mode. Do not switch it to the
port can harm tissue or burn material across a operating mode until the procedure is about to

Page 6
IPM Safety

begin and the laser and its delivery system are Radiology equipment
properly positioned. If the procedure must be
interrupted, disconnect the laser from line voltage; Personnel working on equipment that emits
remove the laser operating key, and store it in a ionizing radiation (e.g., x-rays) must be
controlled location. knowledgeable in radiation safety procedures. For
personnel not already familiar with the safety and
Do not use the laser in the presence of flammable maintenance of this equipment, radiation safety
anesthetics or other volatile substances or should be included as a formal part of their
materials (e.g., alcohol, acetone) because of the training. Even if maintenance personnel spend far
serious risk of explosion and fire. Remove from less time with radiology equipment than users
the working area or cover with flame-resistant (technologists and physicians), they should be
opaque material all reflective surfaces likely to be aware of occupational radiation exposure issues
irradiated by the laser beam. Whenever possible, and take practical steps to minimize exposure. In
use a firebrick or other nonflammable material addition to radiation exposure issues faced by
(e.g., black Delrin) behind the target material typical users, maintenance personnel should also
when the laser is to be activated. A carbon dioxide take precautions against special risks such as the
fire extinguisher should be readily available in the following:
room in which the laser is used.
 Excessive or unnecessary radiation exposure
Some surgical lasers use high voltages (e.g., 20 from unintentional activation or activation
kV), which can be lethal. Capacitors may store with normal protective components of the
charges long after the device has been device removed or disabled.
disconnected from line voltage. Consult the
manufacturer’s recommended procedures for  Risk of exposure to high voltages made
servicing high-voltage laser circuits (including accessible during maintenance procedures;
using proper lockout/tagout procedures), and contact with these voltages can lead to serious
avoid contact with any portion of the high-voltage injury or death.
circuit until the charge has been drained. When  Thermal risks from contact with a hot tube or
possible, disconnect the laser from line voltage tube housing or with the hot oil that insulates
before entering the laser cabinet, and use insulated the x-ray tube. (Contact with oil might occur
gloves for those procedures in which contact with if the tube is damaged from overheating.) For
a high-voltage source is possible (and the gloves personnel safety, and to avoid damaging the
are not otherwise contraindicated). Ensure that tube, do not exceed the tube rating for housing
equipment intended to be used to measure, drain, or anode heat capacities.
or insulate high voltages carries the appropriate
insulation rating (e.g., above 20 kV).  Mechanical risks (e.g., crushing) caused by
component movement associated with
Report any laser accident immediately to the laser inadvertent activation (or brake mechanism
safety officer (or equivalent) and to the facility release) and by disconnection of
risk manager. compensating weights, brakes, or other
restraining components. Also, precautions
For a comprehensive discussion of laser safety, must be taken against movement of the
see ANSI Z136.1-2007, Safe Use of Lasers, and compensating weights when removing the
Z136.3-2005, Safe Use of Lasers in Health Care corresponding components, such as the x-ray
Facilities. tube or image intensifier.
 Chemical risks from fixer and developer
solutions in film processors and, on old x-ray
generators, the polychlorinated biphenyl
(PCB) insulating fluid used in transformers.

Page 7
IPM Safety

Personnel exposure limits organs being exposed. Each organ has a weighting
To ensure safety, personnel exposure to radiation factor and the total effective dose is the sum of the
should be kept as low as reasonably possible corrected dose absorbed by all organs. Effective
(ALARA). Personnel exposure to radiation is dose uses the same units as dose equivalent.
commonly specified in dose-equivalent units. Regulatory dose limits and overall cancer risk are
Current federal regulations limit whole-body expressed in terms effective dose.
occupational exposure to 5 rem (50 mSv) per year
(10 CFR 20.1201). Precautions during IPM
Radiation measurement units  When possible, avoid x-ray exposure by
RADIATION EXPOSURE—A measurement of standing behind a permanent shield.
the ionization created in air by radiation, Otherwise, wear a lead apron and thyroid
measured with an ionization chamber. The shield at all times during x-ray exposure, or
traditional measurement unit is the roentgen (R); use overhead or mobile lead shields.
the standard international (SI) unit is coulombs  Maintain the greatest possible reasonable
per kilogram (C/kg) (1 R = 2.58  10-4 C/kg). distance from the x-ray source and all
(Also see ABSORBED DOSE.) scattering material.

ABSORBED DOSE—A measurement of the  Do not place hands or fingers in the x-ray
amount of energy deposited in a medium. The beam. If unavoidable, wear lead gloves.
traditional measurement unit is the rad; the SI unit  Keep x-ray exposure time to a minimum.
is the gray (Gy) (1 rad = 10 mGy). (RADIATION
EXPOSURE can also be specified in Gy; when  For tests that require the use of cine or serial
using this unit, it is implicit that the medium in film changer acquisitions, staff should leave
which the energy is deposited is air.) For x-ray the room, if possible, when these images are
photons, a radiation exposure of 1 R is about being acquired because much higher exposure
equivalent to an absorbed dose of 1 rad. rates are used for these images than in
conventional fluoroscopy.
DOSE EQUIVALENT—A measurement of the  Radiation badges should be worn by all
ABSORBED DOSE that produces the same personnel servicing radiology equipment and
biological effect regardless of the type of ionizing they should be processed monthly. A radiation
radiation (e.g., alpha particles, beta particles, x- safety officer should review all badge
ray or gamma ray photons) commonly used in readings and take immediate remedial actions
protection literature. The traditional measurement if exposures exceed expected limits.
unit is the rem or mrem (thousandths of a rem);
the SI unit is the sievert (Sv) (1 rem = 10 mSv).  Take precautions against radiation exposure,
The dose equivalent is obtained by multiplying high voltage, thermal, mechanical, infection,
the absorbed dose by the quality factor (Q), which and chemical risks associated with inadvertent
varies according to the type of radiation: activation, device disassembly and removal,
or disabling of protective features.
Dose equivalent = Absorbed dose  Q  It is critical that personnel performing IPM or
servicing of radiology devices receive initial
For x-rays, Q = 1; therefore, 1 rem = 1 rad and periodic training to raise their awareness
(approximately equal to 1 R). of radiation safety and to help them use
optimum protection practices.
EFFECTIVE DOSE—An estimate of the
biological risk of a particular dose. Effective dose
takes into account the radiosensitivity of the

Page 8
IPM Safety

Infection control Wear a lab coat or other appropriate outer garment


to prevent contaminating street clothing. Do
Clinical engineering personnel face a risk of not take lab coats home. Have them washed
infection during medical equipment maintenance frequently in the facility laundry.
and should take appropriate infection control
measures. Disease can be transmitted by several Wear gloves when in contact with equipment that
modes; for clinical engineering personnel, contact may have come in contact with blood, body
transmission and, to a lesser extent, airborne fluids, or other infectious materials (e.g.,
transmission are the most likely modes. Contact surgical instruments, transducers, sensors,
transmission can occur through direct contact with breathing circuits). CAUTION: Wearing gloves
an infected patient, droplet contact (e.g., from a does not protect against cuts and puncture
cough or sneeze), or indirect contact (i.e., from wounds; exercise care when handling
handling equipment that is contaminated by contaminated medical equipment. (Also see
infectious material from a patient). Airborne “Synthetic Surgical Gloves” in Health Devices
transmission can occur through inhaling 2000 Feb-Mar;29:37-66.)
disease-causing microorganisms on dust particles Do not touch clean items (e.g., doorknobs,
or infectious residue on evaporated droplets telephones, test equipment, computer
(droplet nuclei); the airborne transmission of terminals, keyboards) with soiled gloved
tuberculosis (TB) by this process is a particular hands.
risk in healthcare facilities. To protect against
unnecessary exposure to infectious diseases, Wear face shields or masks and protective
clinical engineers should be familiar with basic eyewear during cleaning and decontamination
infection control practices, as well as disinfection, procedures that are likely to aerosolize or
sterilization, and decontamination procedures that splash droplets of blood or body fluid onto
may be necessary before servicing medical mucous membranes.
devices. Wear gowns or protective aprons during cleaning
and decontamination procedures that are likely
Basic infection control practices to aerosolize or splash droplets of patient
material onto clothing.
Good personal hygiene and common sense are
instrumental to developing an effective infection Where there is a special concern about exposure
control program. Infection control practitioners to aerosolized infectious agents, especially TB,
should provide clinical engineering personnel take appropriate respiratory precautions,
with training that includes review of general including wearing a respirator approved by the
principles of TB and bloodborne pathogen National Institute for Occupational Safety and
transmission and prevention, as well as voluntary Health (NIOSH). However, first obtain
hepatitis B virus (HBV) immunization and training on the use and fit-testing of the
postexposure follow-up procedures. Below are respirator. Follow your facility's TB control
some basic infection control practices that are plan. (See Handling and disposing of TB-
appropriate for clinical engineering personnel: contaminated HEPA filters, below.)
Do not rub eyes or other mucous membranes.
Wash hands routinely—when hands are obviously Inspect or repair devices that normally require
soiled, after handling soiled equipment, after cleaning, disinfecting, or sterilizing only after
removing protective gloves, before eating, and such procedures are completed. Clean,
before leaving the facility. disinfect, or sterilize these devices again before
Do not eat, drink, chew gum, smoke, or apply returning them to patient care areas.
cosmetics in work areas. Use standard facility procedures when entering,
leaving, and working in areas that pose special

Page 9
IPM Safety

infection control problems (e.g., surgical contamination by bloodborne pathogens,


suites, dialysis units, nurseries, neonatal ICUs, some dialysis units may be dedicated to
burn units, isolation rooms, critical care areas). patients with HBV or HIV; consider
Patients in these areas may be unusually maintaining separate, dedicated tool sets
susceptible to infection, and the inspector may for servicing dialysis equipment and for
be at considerable risk. units dedicated to patients with HBV or
HIV)
Follow the facility's TB control plan when
entering, working in, and leaving areas where  All devices or items in the OR and ER,
confirmed or suspected infectious TB patients especially footswitches of devices used
are present (e.g., a negative-pressure TB during surgery (e.g., electrosurgical units)
isolation room).
 Any handheld items or other items that
Even if the patient is not in the area, continue to can be found in beds (e.g., nurse call
follow appropriate procedures when entering buttons, remote controls for television
an isolation room or area where a TB patient sets, pillow speakers, blood-pressure
was present until the area has received final cuffs, telemetry transmitters)
cleaning and adequate time has passed. Check
your facility's policy for the appropriate time to  Clinical laboratory equipment, including
wait; the Centers for Disease Control and centrifuges, in which blood splashes and
Prevention (CDC); (in: CDC final guidelines fragments of glass from broken blood
for preventing the transmission of collection tubes may be present
Mycobacterium Tuberculosis in health-care (Vacuuming the interior may be
facilities, 1994. Fed Regist, 1994 Oct necessary.)
28;59[208]:54242-303) implies that cleaning  Breathing circuits
personnel require personal protective
equipment for at least 69 minutes after the  Blood warmers
patient has left if the room has an air exchange  In general, devices from any area where
rate of six air changes per hour (ACH) and body fluids are spilled, splashed, or
99.9% removal efficiency is to be achieved. routinely handled, as well as devices that
(This time can be increased as much as 10-fold are in contact with or contain patient
to take into account nonideal mixing. Also, the material (While device contact is usually
time will depend on the air exchange rate [6 limited to or contained within a
ACH is the minimum recommended rate for an disposable set, fluid squirts, leaks, and
isolation room] and the desired percent spills or handling by personnel whose
removal efficiency.) hands are contaminated often results in
Any device that is visibly soiled with patient device contamination [e.g., in dialysis
material—even if dried—or that has been in equipment, blood warmers,
contact with patient fluids or tissues should be autotransfusion units]).
treated in accordance with the facility's
bloodborne pathogens exposure plan. Handling and disposing of TB-
However, even if a device appears to be clean, contaminated HEPA filters
do not handle it in an unhygienic manner.
Consideration and awareness of infection Manufacturers' instructions for inspecting,
control issues may be particularly appropriate cleaning, and maintaining high-efficiency
for the following devices (even when they particulate air (HEPA) filters and respirators
appear to be clean): should be followed to ensure proper function.
 Dialysis equipment (In addition to a HEPA filter maintenance should be performed
routinely high risk of equipment only by adequately trained personnel. Appropriate

Page 10
IPM Safety

respiratory protection and gloves must be worn handle, can be accomplished by thorough cleaning
while performing maintenance or testing followed by an appropriate disinfection or
procedures or while removing or disposing of the sterilization procedure.
filter. In addition, filter housing and ducts leading
to the housing must be labeled clearly (e.g., Disinfection is the process of killing virtually all
“contaminated air” or a similar warning). recognized pathogenic organisms, but not
necessarily all forms of microbial life, on
To date, CDC has not issued any inanimate objects (Rutala 1995). Disinfectants are
recommendations or guidance regarding the classified in three levels (high, intermediate, or
infectious nature of HEPA filters. Because HEPA low) depending on their lethality. The
filters may contain infectious organisms (the effectiveness of a disinfectant depends on several
survival characteristics and re-aerosolization factors, including the nature and number of
properties of TB from HEPA filter media are contaminating microorganisms (especially the
unknown), ECRI Institute recommends that presence of spores), concentration of the
healthcare facilities dispose of these filters in disinfectant, the length of the application period,
accordance with their state laws for the disposal of the amount of organic soil present, the type and
infectious or regulated medical wastes. condition of the material to be disinfected, and the
Manufacturers of HEPA air filtration units temperature. Thus, the effectiveness of
sometimes provide recommended handling and disinfectants can range from sterility to minimal
disposal techniques in the service or operating reduction of microorganisms.
manuals that include disinfecting with
glutaraldehyde, with formaldehyde gas, or by Sterilization is the process of eliminating all
spraying a phenolic disinfectant such as Amphyl, microorganisms, including vegetative bacteria,
Lysol, or undiluted bleach into the filter. viruses, fungi, and bacterial spores (dormant
Disinfected filters can be disposed of as ordinary forms of microorganisms that are more resistant to
waste. Follow these recommendations (but first heat and chemicals than their nondormant forms).
resolve any conflicts with ECRI Institute's In healthcare settings, sterilization is generally
recommendations or other procedures that have accomplished by using physical agents (e.g.,
been set up at the facility by infection control or steam autoclaving) or chemical agents (e.g.,
biosafety engineering). ethylene oxide gas, gas plasma). These procedures
require specialized equipment and should be
Because the viability of TB on HEPA filters is performed only by personnel trained in sterile
presently unknown, transportation of a processing.
contaminated HEPA filter appears to be regulated
by the U.S. Department of Transportation (DOT) To assess the need for decontamination, medical
as a Class 6 hazard (e.g., poisonous or infections devices can be classified in three categories:
materials) (49 CFR 173.2). This is because the
DOT regulations apply to CDC's list of “certain
Devices visibly contaminated (e.g., with blood,
etiologic agents,” which specifically includes all
feces, urine)
species of Mycobacterium (42 CFR 72.3). This
means that it is subject to certain packaging and Devices used for perfusion, hemodialysis units,
label requirements. Consult with your legal autotransfusion devices, aspiration probes of
counsel, risk manager, or biohazardous material clinical laboratory instruments, and any other
shipping expert. device that routinely comes into contact with
blood (Treat these devices as though they are
contaminated even if no visible contamination
Decontamination is evident)
Decontamination, the process of reducing
microorganisms to a level considered safe to Devices not visibly contaminated and not very
likely to be contaminated (e.g., infusion

Page 11
IPM Safety

pumps, bedside monitors) (Although these the manufacturer's recommendations,


devices may not require decontamination, including the proper length of time,
handle them and all clinical equipment with concentration, and temperature.
care and an awareness of infection control
Tag and bag, if possible, all visibly contaminated
issues)
medical equipment (or equipment known or
strongly suspected to be contaminated) that
Manufacturers are responsible for outlining
must be transported to the clinical engineering
decontamination procedures in the instruction
department for service to indicate the nature of
manuals for their devices. If decontamination
the contamination and to contain the
procedures are not included in the manual, contact
contamination before transport. Some facilities
the manufacturer for explicit instructions. Before
may have policies regarding transport of
purchasing equipment, insist that manufacturers
contaminated equipment. Be consistent with
supply appropriate decontamination procedures
existing policies where applicable.
and have the infection control practitioner review
them. If a manufacturer does not include Use CDC standard precautions when handling
decontamination procedures in its manual, or if its medical equipment that is or may have been
instructions for proper decontamination for a contaminated with blood or body fluids. Treat
particular device are not clear, file a problem all blood and body fluids as though they are
report with ECRI Institute, and consult with the able to transmit infection.
facility's infection control practitioner to develop a Designate specific areas where decontamination
decontamination protocol. Consider the following and servicing of contaminated movable
guidelines for decontamination procedures. equipment can be performed without risk of
infecting patients or other employees. The
Central supply personnel should routinely perform decontamination area should not be near any
decontamination procedures; clinical patient care areas, food preparation or storage
engineering staff should perform them only if areas, medication areas, or other clean areas.
necessary. When it is necessary, use an Preferably, the designated area should be away
appropriate physical or chemical disinfection from direct employee traffic and free from
or sterilization procedure as outlined in excessive draft. (Proper ventilation is essential
Decontamination Procedures, below, and do because many chemicals used for disinfection
the following: can cause eye, nose, throat, or respiratory
irritation.) The surface of the decontamination
 Remove all visible contamination before area should itself be able to withstand
proceeding with the decontamination decontamination (i.e., be made of metal or a
process—large quantities of protein (as nonporous surface other than wood).
found in blood, body fluids, and other
patient material) may hinder disinfection. CAUTION: Do not perform electrical device
repairs on metal surfaces—this increases the
 Use plastic-backed gauze or risk of electric shock.
puncture-resistant gloves to reduce the
risk of cuts or puncture wounds when Use barrier protection (gloves, face shields or
cleaning instruments or components with masks and protective eyewear, gowns, and
sharp edges, metal probes, or other respiratory protection) as needed during
structures that present a risk of skin cleaning and decontamination procedures to
injury. Punctures may occur even with prevent exposure to potentially infectious
protective materials or gloves, so always materials. (See Basic infection control
use caution. practices, above.)
 Be sure that the medical device is exposed Use appropriate procedures to decontaminate any
to the decontamination agent according to tools or equipment that come into contact with

Page 12
IPM Safety

blood or body fluids or other potentially  When special concerns exist, use the
infectious materials during servicing. respiratory precautions described above.
Exercise care when opening a device for cleaning Decontaminate any contaminated equipment or
or repair. components that must be returned to suppliers
or sent out for service.
 The likelihood of being infected as a Dispose of all used cleaning and disinfecting
result of contamination from the materials, gloves, masks, and gowns in
interior of a device is low unless accordance with the facility's policy. Dispose
visible contamination with blood, of glass and any other sharp objects in a proper
body fluids, or other infectious sharps container. (See: Sharps safety devices.
materials is present. If contamination Health Devices 2006;35[337-53].)
is present (or suspected), contact the
In the event of a percutaneous or mucous
device manufacturer for specific
membrane exposure to blood or body fluids,
recommendations for
report the incident to the facility's infection
decontaminating the interior of the
control specialist or employee health
device. If the manufacturer cannot
administrator.
offer specific recommendations, use
detergent and water to remove visible
contamination, rinse, soak in alcohol, Decontamination procedures
and thoroughly dry the device; this
will significantly reduce the number In the absence of manufacturer recommendations
of contaminating microorganisms. If for decontaminating equipment, clinical
soaking is not practical, spraying the engineering personnel can help decide what
device with alcohol may be an decontamination procedure to use depending on
appropriate solution. (Also, see the nature of the contamination, the level of
Decontamination procedures, below.) disinfection or sterilization needed, and the type
of material to be decontaminated. Central supply
 WARNING: Even this procedure may personnel trained in sterilization and disinfection
damage some components. Obtain and procedures should process contaminated
use the manufacturer's recommendations equipment and devices. If this is not possible in a
if possible. specific instance, disinfect the device using a
chemical disinfectant. Discuss the advantages and
 Whenever decontaminating the disadvantages of various disinfectants with an
interior of a device, verify device infection control practitioner.
safety and performance before
returning it to use. Where relevant, procedures forth in the facility's
 Remove dust by brushing and vacuuming written bloodborne pathogens exposure control
(rather than blowing) to minimize plan and TB control plan must be followed.
dispersing dust particles and to avoid
blowing dust into other parts of the device Clinical engineering decontamination applications
or other devices. If filters are present, and some of the disinfectants that may be
clean them in a manner that will minimize particularly suitable for them are listed below.
dispersing the dust. Vacuuming followed Most of these agents contain chemicals that have
by cleaning with soap and water may be OSHA-recommended exposure limits. Exercise
adequate. (Although the risk of personnel care to reduce exposure—avoid inhaling vapors,
infection from dust is not well minimize skin contact, and protect eyes if splashes
established, it is likely to be minimal.) are anticipated.

Page 13
IPM Safety

CAUTION: Because of the diversity of medical ECRI Institute believes that users should
devices and the various materials used to follow all instructions of a
manufacture them, one procedure cannot disinfectant/sterilant supplier, including
adequately decontaminate all equipment. Unless its directions for precleaning. For
manufacturer approved, any decontamination additional information and guidelines
procedure poses the risk of damage. We cannot regarding clinical applications of
guarantee that the following procedures will not glutaraldehyde, see “Choosing a Low-
harm some devices, and we stress that they should Temperature Sterilization Technology,” a
be used only in the absence of a guidance article in the November 1999
manufacturer-approved decontamination protocol. issue of Health Devices. Clinical
Report any problems regarding engineering personnel can also check with
manufacturer-supplied decontamination a healthcare facility's housekeeping
procedures to ECRI Institute for further department to determine what it is using
investigation and resolution. for surface decontamination

1. Sterilization or high-level disinfection Comment: Federal OSHA currently does


(depends on concentration and exposure not have a permissible exposure limit for
time): glutaraldehyde, but some states (e.g.,
California) either do or are soon to have
• Glutaraldehyde enforceable exposure limits, so check
with your state. The American
The required exposure time when using Conference of Governmental Industrial
glutaraldehyde remains unclear. The Hygienists (ACGIH) has set a threshold
Association for Professionals in Infection limit value ceiling (TLV-C) of 0.05 ppm
Control and Epidemiology Inc. (APIC) for glutaraldehyde. Because
recommends guidelines for high-level glutaraldehyde is an eye, nose, and
disinfection calling for 20 minutes or respiratory irritant, contact the appropriate
longer at 20C. However, the time and staff member (e.g., infection control
temperature specified by Johnson & practitioner, facility safety officer,
Johnson for CIDEX-activated dialdehyde employee health specialist) to be sure that
(a 2.4% solution of glutaraldehyde-base its use is permitted in the work area, and
disinfectant) is 45 minutes at 25C to ask about any precautions that should be
support a high-level disinfection claim. taken when using any glutaraldehyde-
containing product.
CIDEX has had this label claim since
1984. However, with its 510(k) clearance, • Formaldehyde
FDA required the package insert to state
precisely that immersion for 45 minutes at Comment: Formaldehyde use may be
25C is required for high-level limited to certain areas within the facility
disinfection. Competitive 2% alkaline and is not recommended for typical
glutaraldehyde product manufacturers clinical engineering applications. It can
face the same labeling requirements. Of cause eye, nose, throat, and respiratory
course, all disinfectant and sterilant irritation and allergic reactions and is a
labeling directions assume precleaning of suspected carcinogen. OSHA has
equipment with an enzymatic detergent or established a specific standard for
a detergent that removes debris and formaldehyde exposure (the permissible
significantly reduces microbial exposure limit is 0.75 ppm and the short-
contaminant. term exposure limit is 2 ppm), which
requires exposure monitoring, medical

Page 14
IPM Safety

examination, and medical surveillance for contact time for decontamination.


employees exposed to formaldehyde in Generally, soaking in alcohol for a period
the workplace. of time is necessary. Alcohol may damage
some plastics. Alcohol is also very
2. Intermediate-level disinfection: flammable; avoid sparks and heat.

• Chlorine compounds (>500 mg/L free • Phenolic compounds, aqueous (0.5% to


available chlorine). 3%) (rated as intermediate to low-level
disinfectants)
Comment: A 1:10 dilution of household
bleach provides 5,000 mg/L of available 3. Low-level disinfection:
chlorine. This concentration exceeds the
intermediate level of 500 mg/L, which • Quaternary ammonium compounds (0.1%
inactivates HBV in 10 minutes and HIV to 0.2%)
in 2 minutes (CLSI 2004). At this
concentration, chlorine-containing Comment: Quaternary ammonium
disinfectants are a good general-purpose, compounds are eye, nose, and throat
broad-spectrum disinfectant for tabletops irritants. They are good, general-purpose
and other surfaces, but they may corrode cleaners for surfaces such as tabletops,
metals. Note: A bleach solution should be walls, and floors.
mixed daily to be effective.
Mercury
• Iodophors (30 to 50 mg/L free iodine or
70 to 150 mg/L available iodine). Although mercury use has decreased because of
concerns over toxicity, mercury (and mercury
Comment: Iodophors are serious eye, compounds) can still be found in
nose, and throat irritants and, if sold as sphygmomanometers, thermometers and
antiseptics, should not be used as thermostats, esophageal dilators, gastric
equipment disinfectants. Only iodophors decompression tubes, and histology fixatives and
registered with U.S. Environmental stains.
Protection Agency as hard-surface
disinfectants should be used, and Mercury and mercury compounds are among the
manufacturers' instructions for most toxic substances found in healthcare
concentration and exposure time should facilities, and few healthcare employees fully
be followed (CLSI 2004). understand the importance of proper handling and
spill cleanup. Exposure to high levels of mercury
• Iodine + alcohol (0.5% + 70%) and mercury compounds can lead to acute
poisoning, even death. It is essential to identify
• Isopropyl alcohol (70%) any mercury related problems in the facility.
Policies and procedures must be established to
Comment: Overexposure to isopropyl handle these problems; employees need to know
alcohol can cause eye, nose, and throat the procedures and must adhere to them.
irritation, as well as neurobehavioral Health risks
effects. Consult OSHA's exposure limits
Mercury is poisonous in all forms. Elemental or
before using. Rapid evaporation of
metallic mercury is a liquid at room temperature
isopropyl alcohol may shorten contact
that readily vaporizes. Accidental spills, improper
time unacceptably. Although an alcohol handling, and poor ventilation can result in toxic
wipedown may reduce contamination
levels of mercury vapor in room air. When
levels, swabbing does not allow sufficient
mercury vapor is inhaled, approximately 80% is

Page 15
IPM Safety

absorbed into the bloodstream; to a lesser degree, treated with mercury fungicides) or environmental
mercury can also be absorbed through the skin sources (water or air contamination). Gloves can
and digestive tract. protect personnel against contact with mercury
and mercury-containing compounds. Mercury
Mercury vapor is easily absorbed through the thermometers should be phased out as soon as
alveoli of the lungs. Short-term, high-level possible.
exposures can cause acute poisoning, leading to
Exposure monitoring
interstitial pneumonitis, bronchitis, bronchiolitis,
or, in severe cases, death. Chronic (long-term) Mercury-vapor monitoring in high-use areas is
exposure to lower levels of mercury can cause important for a safe working environment.
symptoms such as muscle tremor, irritability, and However, cleanup personnel and employees who
gingivitis (inflammation of the gums). Other work in high-use areas should undergo periodic
symptoms include personality changes, such as screening urinalysis for mercury. Employees who
outbursts of temper, increased excitability, handle mercury on a daily basis should wear a
shyness, or indecision. In addition, researchers are mercury exposure dosimeter. Screening should
finding that long-term exposures can slow worker also be conducted following any spills or if
reaction times and interfere with performance of employees exhibit signs or symptoms of mercury
intricate tasks. poisoning.
Federal regulation
Mercury salts (inorganic mercury compounds) are
OSHA lists mercury and mercuric chloride as
readily absorbed through the skin and digestive
hazardous chemicals. Facilities must maintain up-
tract and can cause localized skin reactions or
to-date material safety data sheets (MSDSs) on
generalized symptoms; mercuric chloride is
these chemicals and are required to comply with
commonly used in the histology laboratory as a
the OSHA hazard communication standard (ECRI
fixative. Organomercurials (organic mercury
Institute 2000). OSHA has also established
compounds) can produce a variety of toxic
permissible air exposure limits (PELs) for
effects; these compounds include fungicides,
mercury and mercury compounds (29 CFR
germicides, and bacteriostats such as merbromin
1910.1000, Table Z2).
(Mercurochrome), mercocresols (e.g., tincture of
Mercresin), and thimerosal (Merthiolate).
Used mercury is a Resource Conservation and
Occupational mercury exposure can cause contact
Recovery Act D-listed hazardous waste, according
dermatitis, but reports in the literature of mercury
to the EPA (40 CFR 261.24, 40 CFR 261.33).
allergy are rare.
Disposal of mercury and of the waste resulting
from spill cleanup must meet EPA requirements.
While the mercury from a broken thermometer in
As with other hazardous substances, state and
a patient's room is not likely to pose a significant
local legislation may be even more stringent than
health risk, high use areas such as
federal legislation.
sphygmomanometer calibration locations can
easily have mercury vapor levels exceeding the Mercury-spill cleanup and disposal
OSHA permissible exposure limit (0.05 mg/m3, as Accidental spills are the greatest source of
a time-weighted average) (29 CFR 1910.1000, Ide mercury vapor contamination. Whether small
1986). (thermometer) or large (sphygmomanometer,
esophageal dilator), all spills should be cleaned up
Dose depends on the concentration and duration immediately and completely. (Note that if the spill
(i.e., frequency and/or duration) of exposure to a is too large to be handled by the immediate staff,
particular toxicant like mercury. The effect of it may be considered an emergency and trigger
workplace exposure can be exacerbated by OSHA's Hazardous Waste Emergency Response
exposures to mercury and its compounds from Operation [29 CFR 1910.120]. Because this will
dietary sources (e.g., contaminated fish, seed grain necessitate use of a fully trained HAZMAT team,

Page 16
IPM Safety

declaring a mercury spill an emergency should be container. Choosing a cleanup device should
avoided whenever possible by making sure that be based on the amount of mercury to be
staff in the immediate work area know how to collected, as well as device cost and
contain and control the spill.) effectiveness.
Use appropriate cleanup equipment. Mercury
When mercury is spilled, it breaks up into tiny
vacuums are specially designed to handle
droplets that lodge in cracks and sink traps, mix
mercury liquid and vapor and can be purchased
with dust, and penetrate cracked or porous
commercially. The exhaust of the vacuum is
materials. It can also cling to gold jewelry,
equipped with special mercury-absorbing
clothing (especially knitted fabrics), and shoe
filters; these filters must be changed when their
soles, so that a staff member can easily carry the
absorbing capacity is consumed to the point
mercury to places outside the immediate work
that effective filtering is lost. Because these
site. In high-use areas, enough mercury can
units can vacuum contaminated dust and
accumulate on work surfaces to cause a serious
mercury decontamination powders, workers
health hazard.
can clean up spills more quickly with less
likelihood of exposure. Mercury-spill control
Floor surfaces can affect the extent of mercury-
kits are designed for cleanup of small spills
spill cleanup. A carpet may lessen scatter but
and should be placed in designated areas where
provides crevices in which mercury droplets can
mercury spillage is possible. These kits include
hide. In many cases, contaminated carpet must be
a method for retrieving mercury droplets and
removed. Tile and vinyl flooring allow wider
chemicals for decontaminating surfaces. Also,
dispersal and formation of smaller droplets, and
in any mercury spill cleanup, protective
mercury can pool beneath tiles under certain
equipment such as impervious gloves, shoe
conditions.
covers, and respirators should be used.
The surface area of each droplet allows mercury Dispose of mercury. All mercury droplets should
to vaporize (at a rate proportional to the be disposed of in a manner that will prevent
temperature—droplets near a heat source will further vaporization and that is in compliance
vaporize more readily). Left untouched, the with federal, state, and local hazardous waste
surfaces of these drops oxidize, slowing the regulations. All waste elemental (metallic)
vaporization rate. Disturbing the droplets with a mercury should be recycled. (All healthcare
broom or foot traffic can break them into smaller staff using devices containing mercury should
droplets, providing fresh surfaces for be aware of policies and procedures
vaporization. Mops, brooms, and vacuum cleaners established for cleanup of large and small
can also carry mercury contamination to other spills. Mercury should never be disposed of
areas of the healthcare facility. into a drain.)
Decontaminate surfaces. Because spilled mercury
To clean mercury spills, designated spill team tends to disperse into microdroplets and fall
personnel should do the following: into cracks and crevices (invisible to the
unaided eye), surface decontamination should
Clean up all visible droplets of mercury. This is be accomplished by applying one of the
accomplished by aspirating the droplets with a chemicals available to coat mercury droplets
plastic disposable syringe or vacuuming the (sulfur compounds) or to react with the
droplets using a mercury vacuum with a trap mercury (Hg Absorb), forming an amalgam;
for the mercury droplets and filters to absorb these chemicals virtually eliminate mercury
mercury vapor. In some cases, small amounts vaporization. Some surfaces (such as shag
of mercury can be picked up with adhesive carpet) cannot be decontaminated and may
tape or a sponge device designed to pick up need to be removed.
mercury and deposit it in a collection

Page 17
IPM Safety

Perform area monitoring. Area monitoring is sphygmomanometer calibration laboratory, where


necessary to ensure that decontamination is mercury vapor levels can build.
complete and that the area is safe for
occupancy. As long as small spills are cleaned Any room where sphygmomanometer calibration
up promptly and effectively (e.g., from a and repair is performed should be well ventilated.
thermometer), area monitoring is typically not It should be reserved for the exclusive task of
necessary. handling mercury, especially if calibration or
repairs are performed often. Traffic through the
Dirty metallic mercury should not be disposed of area should be limited. There should be no
as waste but should be recycled. Facilities should smoking, drinking, or eating in the room. Floors
not clean and reprocess their own mercury should not be carpeted, and benches should be
because this increases the risk to personnel from equipped with troughs to collect mercury spills.
accidental spills or poor technique. Used mercury Personnel should remove all jewelry, especially if
can be accumulated in small, airtight, unbreakable it is gold or gold plated (mercury readily
containers and sold as scrap to mercury refineries. combines with gold), and they should wear
disposable, chemically-resistant gloves. In
Managing mercury sources
high-use areas, workers should wear disposable
Sphygmomanometers gowns and shoe coverings to minimize skin and
Mercury-containing sphygmomanometers are clothing contamination, which can increase
used in patient areas to measure blood pressure; worker exposure and carry mercury to other areas
they may be mounted on a wall, on a mobile of the healthcare facility. Consider use of a
stand, or in a carrying case. mercury vapor respirator and periodic surveys of
mercury vapor levels in areas where
Planned elimination of mercury sphygmomanometer maintenance and calibration
sphygmomanometers is advisable to reduce the is performed.
risk of staff and patient mercury exposure. Bougies
Mercury sphygmomanometers are traditionally
Maloney or Hurst bougies (esophageal dilators)
considered more accurate and reliable in high-use
are, typically, red (natural) rubber tubes
facility settings, but newer, high-quality,
approximately 30 inches long that are filled with
wall-mounted aneroid gauges are available that
mercury. A variety of diameters (approximately
are more reliable than earlier designs. Although
0.2 to 0.8 in) are available to treat esophageal
costs will be associated with replacing and
constriction. They are commonly found in
maintaining new aneroid units, the personnel and
operating rooms, gastrointestinal labs, and
materials costs associated with maintaining the
endoscopy departments; each tube may contain as
mercury units and controlling mercury exposure
much as three pounds of mercury. Over time, the
will be eliminated.
red rubber becomes brittle and can rupture,
spilling its contents. Before each use, these tubes
All workers who use, clean, or maintain mercury
should be inspected for any signs of deterioration
sphygmomanometers should understand the
so that mercury can be collected in an airtight
properties of mercury and its associated hazards
container for recycling, thus avoiding employee
and should be instructed in safe handling
exposure and spill-site cleanup and
procedures. Specific policies and procedures must
decontamination.
be established for mercury spill cleanup, and
specific areas should be designated for
Gastric decompression tubes (e.g., Cantor,
maintenance activities.
Miller-Abbott), which are used to remove gas
from the alimentary tract, have a balloon to hold
When sphygmomanometers are dropped or
the tube in place. The balloon is filled with
broken, prompt, thorough cleanup is important,
mercury by the surgeon during the procedure. The
especially in high-use areas, such as a

Page 18
IPM Safety

tube is usually removed in the patient's room. (Tungsten-weighted tubes are still available, but
Personnel involved with insertion or removal of unweighted tubes are generally considered as
these tubes and with handling the mercury must effective as weighted tubes.) However, it is
be educated in safe handling of the mercury and possible that some facilities still have
what procedures to follow in the event of a spill. mercury-weighted tubes in stock; their disposal
Infant incubators
can pose a problem because the mercury in the
tube makes the tube a hazardous waste and its
Some older incubators still in use today have hood contact with the patient can make it an infectious
thermometers and/or high-temperature waste.
thermostats that contain mercury. ECRI Institute
recommends replacing all mercury-containing Various instruments in the clinical laboratory use
components in infant incubators. Should an mercury in their operating mechanisms. Some
incubator thermometer or thermostat containing hematology analyzers, for example, use mercury
mercury break, personnel should immediately in their hydraulic system to draw samples through
transfer the infant to another incubator and the counting chamber. Disposal of old instruments
remove the contaminated incubator from service. or leaks from the instrument will require
Personnel who are unfamiliar with appropriate cleanup and decontamination.
decontamination procedures and equipment
should not attempt to clean up the spill. Recommendations for avoiding
Laboratory chemicals
mercury contamination
Mercuric chloride, a major constituent of B5 and  Urge the safety committee to adopt
Zenker's solution, is an extremely toxic chemical procedures for safely storing and using toxic
used in pathology laboratories for fixing tissue substances such as mercury. The safety
specimens. Sources of contamination include committee should consult with appropriate
accidental spills on benchtops, floors, and clothing health professionals in the facility concerning
and in sink traps and plumbing (Stewart et al. the proper handling of toxic substances.
1977). To control mercury contamination in the
 Confine large-scale use of mercury to
laboratory, do the following:
properly designed areas. Floor coverings
Use chemicals without mercury when possible should be seamless, and carpets should not be
(e.g., Bovin's fixative solution [picric acid]). installed in areas where mercury might be
Cover workbenches and floors with impermeable spilled.
surfaces.  Require all personnel involved in
Clean spills immediately. decontamination of mercury spills or
maintenance in which significant mercury
Wear gloves whenever using mercury solutions. exposure is possible to undergo
Adequately ventilate the work space. preemployment testing. Also, include annual
mercury-level testing as a part of their regular
Other sources physical examinations.
Thermometers used in patient areas and the
laboratory frequently contain mercury. All spills  Survey maintenance shops and other
from broken thermometers should be cleaned up locations where mercury has been spilled and
with proper mercury disposal techniques. improperly cleaned to determine the location
and severity of contamination.
In the past, it was believed that weighting an  Draft formal procedures, and instruct all
enteral feeding tube with mercury made it easier affected staff in the safe handling of mercury
to insert the tube and anchor it in the stomach. as part of the OSHA hazard communication
Mercury-weighted enteral feeding tubes are standard program. These procedures should
reportedly no longer manufactured. include development of a mercury-spill

Page 19
IPM Safety

reporting system, which involves placing


warning placards in designated mercury use
areas (e.g., sphygmomanometer calibration
area) and designation of a spill-cleanup team.
Very minor spills can be handled locally by
using strategically located emergency spill
kits.
 Should there be a mercury spill (e.g., a
sphygmomanometer break) in a patient room,
relocate the patient (if possible) until the spill
is contained and a mercury vapor monitor has
demonstrated that vapor levels are acceptable.

Page 20
IPM Safety

References

American National Standards Institute. ANSI Z136.1 (2007) Safe Use of Lasers standard.

Canadian Standards Organization (CSA). Laser safety in health care facilities [standard]. CAN/CSA-Z386-
08. 2008.

Centers for Disease Control and Prevention (CDC). Universal precautions for prevention of transmission of
HIV and other bloodborne infections are discussed in the following documents:

CDC. Guidelines for prevention of transmission of human immunodeficiency virus and hepatitis B virus to
health-care and public-safety workers. MMWR 1989;38(S-6):1-36.

CDC. Recommendations for prevention of HIV transmission in health-care settings. MMWR 1987;36(suppl
no. 2S).

CDC. Update: Universal precautions for prevention of transmission of human immunodeficiency virus,
hepatitis B virus, and other bloodborne pathogens in health-care settings. MMWR 1988;37:377-88.

Clinical and Laboratory Standards Institute (CLSI) Wayne (PA): Clinical laboratory safety. Approved
guideline (2nd ed) GP17-A2; 2004 and Protection of laboratory workers from occupationally acquired
infections. Approved guideline M29-A3; 2005.

ECRI Institute. Laser safety eyewear. Health Devices 1993;22(4).

International Commission on Radiological Protection (ICRP). 1990 recommendations of the ICRP.


Publication 60, 1991.

Laser Institute of America. Laser safety Guide, Orlando (FL). 2007

National Academy of Sciences (NAS). Health effects of exposure to low levels of ionizing radiation (BEIR
VII). Washington (DC): National Academy of Sciences Press; 2006.

National Council on Radiation Protection and Measurements (NCRP). Ionizing radiation exposure of the
population of the United States. Bethesda (MD): NCRP; 1987. NCRP Report 93.

Rutala WA. (ed) Disinfection, sterilization and antisepsis: principles, practices, current issues, and new
research. Association for Professionals in Infection Control and Epidemiology, Washington (DC). 2007.

9 CFR 1910.120. 40 CFR 261.24.

29 CFR 1910.10002. 40 CFR 261.33.

29 CFR 1910.1200. 42 CFR 72.3.

29 CFR 1910.147. 49 CFR 173.2.

Page 21

Вам также может понравиться