Oscillometry (Luisa)
Impulse oscillometry (IOS) is a respiratory function test that allows you to evaluate
the impedance of the respiratory system at different oscillation frequencies.
Furthermore, this test evaluates respiratory mechanics through the application of
small pressure pulses generated by a horn, which produces fluctuations in flow at a
certain frequency that overlap with the natural pattern of respiratory flow. (Luisa)
1.Calibration:
The oscillometer must be calibrated or verified in daily calibration for volume,
before
to perform the test.
2.Syringe
Use a 3.00 L syringe with accuracy ± 0.5% of absolute volume (15 mL).
3.Connection
Ensure a firm and tight connection between the syringe and oscillometer.especially
when do not use the original syringe of the equipment. (cami)
4.Special Programs
Most current oscillometers have special menus in the program to perform and store
the calibration data.
5.Verification Of Calibration
For verification of pressure calibration the rear grid (resistor) must be closed, place
the resistance administered by the manufacturer and perform a conventional
oscillometry (Feli)
1.Presentation
The technician who performs the test receives and introduces himself with the
patient.
2.Order verification
The medical request should be reviewed and the full name and registration
number.
3. Explain
The technician explains to the patient the purpose of the test.
Consider (Luisa)
In patients who cannot keep from foot or suffer from rib cage deformity, it
you can use the arm extension as a height estimation.
The test should be performed with the subject sitting. Chairs without wheels
and with support should be used for arms.
Consider (Luisa)