Академический Документы
Профессиональный Документы
Культура Документы
EUM-Vet-06 (Rev.0)
ULTRA 100HF Operational & Service Manual
Revision
Reason for
Chapter Page Document Rev Date
changes
1
ULTRA 100HF Operational & Service Manual
Table of Contents
Chapter i Safety Information
Introduction
Manual Conventions
Qualification of personnel
Contraindications
Side effect
Identification labels
Symbols Information
Procedural Safety
Disposal
User Maintenance/Servicing
Environmental Safety and Operating Environment
Warranty Scope
Manufacturer’s Responsibility
Chapter 1 Introduction
1.1 Introduction
1.2 Description
2
ULTRA 100HF Operational & Service Manual
1.4 Dimensions
1.5 Specification
1.6 Abbreviations
Chapter 2 Operation
2.1 Safety Precaution
2.2 Specifications
3.2 Calibration
3.2.2 Auto mA
3.2.3 Auto-Gain
3.3 DR Interface
Chapter 4 Troubleshooting
4.1 Introduction
Chapter 5 Maintenance
3
ULTRA 100HF Operational & Service Manual
4
ULTRA 100HF Operational & Service Manual
Safety Information
Introduction
i
The policy of EcoRay Co., Ltd is to provide X-ray equipments that meet high standards of
performance and reliability. We enforce strict quality control techniques to eliminate the
potential for defects and hazards in our products.
The safety guidelines provided in this section of the manual are intended to educate the
operator on all safety issues in order to operate and maintain in a safe manner.
5
ULTRA 100HF Operational & Service Manual
Manual Conventions
ULTRA 100HF is a diagnostic system to obtain the information necessary for the diagnosis
of the patient through the anatomical structure of human body obtained by a film or an
image processing (workstation) after process of examination using radiation exposure with
cassette IP, CR or DR (Portable flat panel).
This manual includes the contents regarding service and operation integrated together.
This manual provides the operational guidelines of the ULTRA 100HF to service
technicians, operators and other people related with ULTRA 100HF.
This manual cannot be used for training, repair or licensor of radiation equipment inside of
the hospital. This equipment must be handled by proper radiologists own radiation devices
license.
ULTRA 100HF is a portable X-ray generating device that is designed to be used with
supporting devices or shooting table. This equipment can receive maintenance service of
the equipment since the installation.
This manual uses three types of messages to emphasize information or potential risks to
personnel or equipment. To ensure safety, read all Safety Information provided herein and
observe all.
WARNING
CAUTION
NOTE
Warning
An instruction that draws attention to the risk of injury or death.
Caution
An instruction that draws attention to the risk of damage to the product or
proper use of the product.
Note
Provides additional information such as expanded explanations, hints or
reminders.
6
ULTRA 100HF Operational & Service Manual
* Please refer the technical data of each components for more detailed information.
Note
Only a qualified, authorized Service Provider should service/repair electrical or
mechanical components.
Qualifications of personnel
Operating personnel
Before using the ULTRA 100HF, it is required that the operating personnel is
thoroughly familiar with the operating instructions, in particular:
Safety
Function and Safety Checks
Acceptance tester
7
ULTRA 100HF Operational & Service Manual
The equipment should be tested by a qualified person that can perform the test, evaluate
the result for compliance and knowledge of radiation safety to national regulations.
Caution
Federal law restricts this device to sale by or on the order of a physician.
Service personnel
Note
It is the responsibility of the owner to ensure that the system is operated only by
trained radiologist, service technicians or product specialists.
Note
It is the responsibility of the supplier to ensure that the technicians have the right
training and knowledge to perform service and maintenance.
Note
When Service or Maintenance will be performed the technician shall lock the
equipment from all energy sources.
To ensure the safety of the patients, the operator and third parties, the equipment must be
checked according to the Function and Safety Checks instructions to maintain reliability
and serviceability.
If national rules or regulations specify more frequent checks and/or maintenance, such
regulations must be observed.
8
ULTRA 100HF Operational & Service Manual
Caution
No modification of this equipment is allowed.
Do not modify this equipment without authorization of the manufacturer.
If this equipment is modified, appropriate inspection and testing must be
conducted to ensure continued safe use of equipment.
9
ULTRA 100HF Operational & Service Manual
- Be sure that all personnel authorized to operate the equipment are familiar with the
established regulations of the authorities named above. All personnel should be
monitored to ensure compliance with the above.
- Although X-Ray radiation is hazardous, X-Ray equipment does not pose any danger
when properly used. Be certain all operating personnel are properly educated
concerning the hazards of X-Ray radiation.
- Persons responsible for the equipment must understand the safety requirements
and special warnings for X-Ray operation.
- All users must equip with the protective devices in prior to using the equipment.
10
ULTRA 100HF Operational & Service Manual
Table 1. The recommended radiation dose for patient in examination (IAEA BSS No. 115)
PA 5
Head
LAT 3
PA 0.4
Chest
LAT 1.5
PA 7
Thoracic vertebrae
LAT 20
PA 10
lumbar LAT 30
OBL 40
Abdomen AP 10
Pelvis AP 10
Femur AP 10
11
ULTRA 100HF Operational & Service Manual
The shielding can be classified as protective equipment for operator and patient:
12
ULTRA 100HF Operational & Service Manual
Treatment group
* Refer diagnostic codes in "ICD-10 Version 2014 site".
13
ULTRA 100HF Operational & Service Manual
Contraindications
Expected except group of treatment
For the purpose of diagnosis on diseases using radiation, it is not recommendable for the
listed patients as following on radiation diagnosis.
Side effect
1) Exposing high dose of radiation directly to patient's eyeballs might cause eye
diseases.
2) Exposing radiation to infant or pregnant without protection might cause genetic
disorder.
3) Exposing radiation to patient equipped with cardiac pacemaker due to cardiovascular
disease might cause stop of the cardiac pacemaker.
Caution
Operate the equipment outside of its operating environment limits may lead to
malfunction of or damage to the equipment.
14
ULTRA 100HF Operational & Service Manual
Identification labels
Identification label on each component is located in different location individually.
Through the labels, it is available to be informed regarding manufacturer and EU
representative and manufactured date with serial numbers and approvals belong to the
product.
Since damages of the identification labels might cause delays on getting service, avoid
from danger causing any damage of the labels.
15
ULTRA 100HF Operational & Service Manual
16
ULTRA 100HF Operational & Service Manual
System Label
1 Side of Equipment
3 Location of Collimator
17
ULTRA 100HF Operational & Service Manual
Symbols Information
This manual uses three types of messages to emphasize information or potential risks to
personnel or equipment. To ensure safety, read all Safety Information provided herein and
observe all.
symbols Description
Warning; Electricity
: To warn of electricity
No pushing
: To prohibit pushing against an object
No stepping on surface
: To prohibit pushing against an surface.
18
ULTRA 100HF Operational & Service Manual
19
ULTRA 100HF Operational & Service Manual
"OFF" (power)
: To indicate disconnection from the mains, at
least for mains switches or their positions, and 5008
all those cases where safety is involved..
Alternating current
: To indicate on the rating plate that the
5032
equipment is suitable for alternating current
only; to identify relevant terminals.
X-ray source assembly, emitting
: To indicate the emission or the imminent
emission of X-radiation. 5339
Patient, thin
: To indicate a reference to a thin patient.
5389
Patient, normal
: To indicate a reference to a normal patient.
5390
Patient, obese
: To indicate a reference to an obese patient.
5391
20
ULTRA 100HF Operational & Service Manual
Packing symbol
Symbol Description ( ISO 7000 ) Number
Fragile, handle
with care 0621
Keep dry
0626
This way up
0623
Temperature limitation
0632
Humidity
limitation 2620
3082
Authorized representative
in the European Community
Date of manufacture
2497
Serial number
2498
21
ULTRA 100HF Operational & Service Manual
Electrical Safety
Sources of potentially lethal voltages/currents are physically and electrically isolated from
both the operator and patient by insulated cables and mechanically secured covers and
access panels..
Mechanical Safety
22
ULTRA 100HF Operational & Service Manual
Procedural Safety
The procedures in this manual are designed for the safety of the patient, operator and the
equipment as well as ensuring the acquisition of diagnostic quality images. Failure to follow
the procedures in this manual (or cautions, warnings and notes) can result in potential
health and safety hazards and/or damage to the equipment.
I. X-RAY equipment must be handled by the proper radiologists own radiation devices
licenses. Users must equip with protective equipments protects from the exposure of
radiation and use the radiation exposure measuring device for safety.
II. ULTRA 100HF cannot be used for perspective examination or others does not belong
to the general uses.
III. ULTRA 100HF must be used for diagnosis and cannot be used for treatment.
IV. ULTRA 100HF is a equipment that is in accordance with IEC 60601-1-2 CLASS
TYPE-B.
V. Users cannot arbitrarily change this product and if necessary or unavoidable to
change this product, the product change must be requested to either the headquarter
or sales agencies.
VI. The necessary adjustment for operation is made at factory before shipping out.
VII. If any problem occurs to the product, users immediately stop using the product and
ask for service to either the headquarter or sales agencies.
VIII. In case of combining ULTRA 100HF with other products from other companies, it is
recommended to make an inquiry either to the headquarter or sales agencies.
IX. Ground wires must be used for AC power supply at all times and it is required to use
the approved ground wires and plugs. The plugs must not be dismantled and
disassembling cases of the product must be avoided due to any electric shock may
caused by the AC power supply.
X. Do not use this equipment in the presence of explosive liquid or gas, etc and if such
those are found, do not turn the power ON.
If any explosive liquid or gas is found when the power is ON, do not turn the power OFF
and eliminate the substance first.
Caution
This equipment is not suitable for use in the presence of flammable anesthetic
mixtures with air, oxygen or nitrous oxide.
23
ULTRA 100HF Operational & Service Manual
Disposal
This product may contain hazardous substances such as lead. Improper disposal can
cause environmental pollution. When disposing this equipment, please contact
EcoRay Co., Ltd.
Do not dispose some or all of the product without prior consultation should for any
damages caused thereby disposing the product without consulting EcoRay Co., Ltd shall
not be responsible in law.
Warning
Information of WEEE label
- The following information is for EU members :
The use of this symbol indicates that this product should not be treated as
household waste. Do not dispose electrical and electronic equipment into municipal
waste and must be collected separately. Handle this product correctly to prevent
environmental contamination caused by improper handling of waste.
Please contact EcoRay Co., Ltd or sales agency to get more details on disposing the
product.
24
ULTRA 100HF Operational & Service Manual
User Maintenance/Servicing
Unless specifically instructed herein, all maintenance and serving of the equipment should
be performed by qualified, authorized Service Providers. None of the electronic assemblies,
cables or mechanical assemblies is user serviceable.
Some cleaning solutions, disinfectants and other chemicals can be destructive to the
equipment or pose a risk of fire. The manufacturer is not responsible for damages or
injuries that may result from the use of non-authorized chemicals on or near the equipment.
Warning
Do not operate the equipment when cleaning the equipment.
Before cleaning the equipment, please turn the power off.
Caution
Do not spray cleaning solution directly onto the equipment. Moisten a cloth with
a 70% isopropyl alcohol solution for use on plastics and enameled metal. Apply to
patient contact areas after each contact.
Caution
Do not immerse the equipment in liquid, do not use autoclave equipment.
The following environmental conditions are appropriate for the equipment. Operation under
conditions outside of these ranges could damage the equipment.
Warning
Operate the equipment outside of its operating environment limits may lead to
malfunction of or damage to the equipment.
25
ULTRA 100HF Operational & Service Manual
Electromagnetic Compatibility(EMC)
The ULTRA 100HF system complies with the requirements of IEC 60601-1-2:2001
+A1:2004/EN 60601-1-2:2001 +A1:2006 regarding electromagnetic compatibility.
Surrounding equipment shall follow the standard IEC 60601-1-2:2001
+A1:2004/EN 60601-1-2:2001 +A1:2006.
Caution
Mobile telephones or other radiating equipment can interfere with the function of
the ULTRA 100HF and can therefore cause safety hazards.
ULTRA 100HF must be installed in a shield room with more than 20 db on
attenuation factor of electromagnetic waves.
26
ULTRA 100HF Operational & Service Manual
27
ULTRA 100HF Operational & Service Manual
NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected
by absorption and reflection from structures, objects and people.
a Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless)
telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast
cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to
fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field
strength in the location in which the ULTRA 100HF system is used exceeds the applicable RF compliance level
above, the ULTRA 100HF system should be observed to verify normal operation. If abnormal performance is
observed, additional measures may be necessary, such as reorienting or relocating the ULTRA 100HF system.
b Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 10 V/m.
28
ULTRA 100HF Operational & Service Manual
Warranty Scope
The warranty is limited to the components for any problems that may occur within a
specified period and it may be excluded from the warranty in the following cases :
If users use the product with other purposes besides diagnosis or excessively
used or user errors.
Damages caused by using wrong power.
Natural disasters such as lightning, earthquakes, hurricanes, flood, and etc.
If any problem occurs due to exterior renovation or changes particular part of the
product with unknown origin without permission.
Any damaged product during transportation.
Any damages due to the external factors such the conditions of use and storage
conditions described in the specification are not satisfied.
If any products purchased from abnormal route besides the manufacturer or
sales agency is broken or damaged.
Manufacturer’s Responsibility
Allow only authorized, properly trained personnel to operate the equipment. Ensure that
all individuals authorized to use the equipment are aware of the danger of excessive
exposure to X-Ray radiation.
This equipment is sold with the understanding that the manufacturer, its agents, and
representatives do not accept any responsibility for overexposure of patients or
personnel to X-Ray radiation. No responsibility is assumed for any unit that has not been
serviced by qualified authorized service personnel.
29
ULTRA 100HF Operational & Service Manual
ii
30
ULTRA 100HF Operational & Service Manual
Introduction
1.1 Introduction
1
ULTRA 100HF is a diagnostic system to obtain the information necessary for the
diagnosis of the patient through the anatomical structure of human body obtained by a film
or an image processing (workstation) after process of examination using radiation
exposure with cassette IP, CR or DR (Portable flat panel).
31
ULTRA 100HF Operational & Service Manual
1.2 Description
ULTRA 100HF is a medical diagnostic X-RAY generator equipped with ease-use in indoor
or outdoor and also a product that can be used in applications where requires x-ray
diagnosis.
Specifications subject to change without notice.
Key features :
32
ULTRA 100HF Operational & Service Manual
Note
The mount for ULTRA 100HF is an option that can be provided according to any
request from customers.
33
ULTRA 100HF Operational & Service Manual
1.4 Dimensions
1.4.1 Basic Dimensions
34
ULTRA 100HF Operational & Service Manual
Caution
It is recommended to fold the portable head when using ULTRA 100HF with the
mount during transportation for safety.
35
ULTRA 100HF Operational & Service Manual
1.5 Specification
Specification
Inverter 80KHz
Inverter Frequency
Filament 40KHz
Manufacturer TOSHIBA
X-ray tube
Model No. DF-182
Weight 18 kg
36
ULTRA 100HF Operational & Service Manual
1.6 Abbreviations
% Percent
Btu British Thermal Unit
°C Degree Celsius
CE Communautés Européenes
cm Centimeter
°F Degree Fahrenheit
ga Gauge
hPa Hecto Pascal
inHg Inches Mercury
Kg Kilogram
kJ Kilojoule
M Meter
max. Maximum
min. Minimum
mm Millimeter
kVp Kilo voltage Peak
mAs Milliampere per sec
37
ULTRA 100HF Operational & Service Manual
Operation
Warning
If anyone who might operate this product recognizes the unintended
condition of the unit such as mechanical failure, noise, unexpected vibration,
and so on must inform directly to the service engineer in charge.
38
ULTRA 100HF Operational & Service Manual
2.2 Specifications
Specifications
Input power 200~240VAC(50Hz/60Hz)
Inverter 80KHz
Inverter Frequency
Filament 40KHz
KV range 40KV~110KV(1KV STEP)
mAs range 0.32mAs~100mAs
Manufacture TOSHIBA
x-ray tube
Model DF-182
Focal spot 1.8mm
Anode heat storage capacity 28KJ
Maximum anode cooling rate 265W
Peak wavelength 650nm
Laser point
Optical power <1mV , Class I
Lamp(LED) Over 100 lux.
Filtration total Al. equiv. 2.5mm
Weight 18 kg
39
ULTRA 100HF Operational & Service Manual
- Lead Glass
- Lead gloves
- Lead Apron
- Lead Neck Apron
2m
Caution
If any problem occurs after turning the power ON, please reboot of the
equipment. If the problem keep occurring, stop using the equipment and
request for a service.
40
ULTRA 100HF Operational & Service Manual
Color of conductors
brown: L
blue : N
green and yellow : GND
41
ULTRA 100HF Operational & Service Manual
11
12 14 15
13
42
ULTRA 100HF Operational & Service Manual
1 X-RAY Ready X-RAY ready status displayed with the LED lamp
2 X-RAY Exposure X-RAY exposure status displayed with the LED lamp
X-RAY exposure availability displayed with the LED lamp after the
3 X-RAY Standby
previous exposure.
4 Error The Light's ON when an error occurs while using the product.
13 APR Saving To save the different positions in order to recall by pushing this button.
43
ULTRA 100HF Operational & Service Manual
Range 0.32mAs~100mAs
mAs
0.32, 0.4, 0.5, 0.63, 0.8, 1.0, 1.25, 1.6, 2.0, 2.5, 3.2, 4.0, 5.0, 6.3, 8.0, 10,
STEP 12, 16, 20, 25, 32, 40, 50, 63, 80, 100mAs
Defaults KV mA mAs
100mA 0.32~2.0mAs
80mA 2.5~6.3mAs
40KV~50KV
60mA 8.0~50mAs
50mA 63~100mAs
80mA 0.32~3.2mAs
70mA 3.2~8.0mAs
51KV~60KV
50mA 10~50mAs
40mA 63mAs~100mAs
70mA 0.32~2.5mAs
60mA 4.0~8.0mAs
61KV~70KV
50mA 10~32mAs
35mA 40~100mAs
Allowable 60mA 0.32~1.6mAs
value 40KV/2mAs 50mA 2.5~10mAs
71KV~80KV
40mA 12~32mAs
30mA 40~100mAs
50mA 0.32~1.6mAs
81KV~90KV 40mA 2.0~5.0mAs
30mA 6.3~40mAs
25mA 50~100mAs
50mA 0.32~0.8mAs
40mA 1.0~2.5mAs
91KV~100KV
30mA 3.2~25mAs
20mA 32~100mAs
40mA 0.32~0.8mAs
30mA 1.0~10mAs
101KV~110KV
25mA 12~25mAs
20mA 32~100mAs
44
ULTRA 100HF Operational & Service Manual
Caution
Input power of the portable X-RAY equipment must be used in the range of
the input power marked on the equipment within +/- 10%.
The power cable of the product must be separated not to be mixed with
other cables.
Before the exposure of patients using the ULTRA 100HF, it is required to Shoot
3-5 times with low doses (40kvp 5mAs) as warm-up operation to ensure the
safe use.
45
ULTRA 100HF Operational & Service Manual
"Ready" button
"Exposure" button
Ready Exposure
46
ULTRA 100HF Operational & Service Manual
1) Turn the power ON while pressing kVp DOWN button and kVp UP button
at the same time as following :
( mAs mode )
* There is no separate button for saving and it automatically saves 3 seconds after
entering the mode.
2. Auto Calibration
- Depending on the equipment input current, it corrects the values such as mA, mAs and kVp
automatically.
1) Turn the power ON while pressing both mAs DOWN button and mAs UP button at the same
time as following :
III. In order to change the previously saved values, select one of those (the body size, body part
and position) and press the "Memory" button for more than 1 second after changing the kVp
and mAs values.
47
ULTRA 100HF Operational & Service Manual
48
ULTRA 100HF Operational & Service Manual
* Test Distance 1m
Test value Dose Area Calculated value Ultra 100 display value DAP value
kV mAs uGy 가로 세로 uGy*m^2 uGy*m^2 Tolerance(%) uGy*m^2 Tolerance(%)
40 2.5 26.36 0.3 0.3 2.3724 2.5084955 -5.42538335 2.4 -1.163378857
40 5 53.87 0.3 0.3 4.8483 5.01699099 -3.3623938 4.9 -1.066353155
40 10 114.3 0.3 0.3 10.287 10.033982 2.52161118 11.1 -7.903178769
55 2.5 63.17 0.3 0.3 5.6853 5.92828037 -4.09866535 6.2 -9.053172216
55 5 128.3 0.3 0.3 11.547 11.8565607 -2.61088147 12.6 -9.119251754
55 10 261.8 0.3 0.3 23.562 23.7131215 -0.6372906 25.9 -9.922756982
65 2.5 91.72 0.3 0.3 8.2548 8.2799949 -0.30428645 9.1 -10.23889131
65 5 183.3 0.3 0.3 16.497 16.5599898 -0.38037345 18.2 -10.32308905
65 10 369.4 0.3 0.3 33.246 33.1199796 0.38049659 37 -11.29158395
75 2.5 125.8 0.3 0.3 11.322 11.0236619 2.70634343 12.5 -10.40452217
75 5 246.7 0.3 0.3 22.203 22.0473237 0.70610065 24.6 -10.7958384
75 10 489.6 0.3 0.3 44.064 44.0946474 -0.06950369 49.2 -11.65577342
85 2.5 158.2 0.3 0.3 14.238 14.1592812 0.55595179 15.9 -11.67298778
85 5 315.7 0.3 0.3 28.413 28.3185624 0.33348287 31.5 -10.86474501
85 10 646.1 0.3 0.3 58.149 56.6371249 2.66940652 64 -10.06208189
90 2.5 176.1 0.3 0.3 15.849 15.8740731 -0.1579498 17.2 -8.52419711
90 5 354 0.3 0.3 31.86 31.7481461 0.35231621 34.6 -8.600125549
90 10 719 0.3 0.3 64.71 63.4962923 1.91146237 70.5 -8.947612425
100 2.5 213.8 0.3 0.3 19.242 19.5976211 -1.81461346 20.7 -7.57717493
100 5 435.4 0.3 0.3 39.186 39.1952421 -0.02357975 42.4 -8.201908845
100 10 875.2 0.3 0.3 78.768 78.3904843 0.4815836 84.7 -7.530977047
49
ULTRA 100HF Operational & Service Manual
50
ULTRA 100HF Operational & Service Manual
51
ULTRA 100HF Operational & Service Manual
3.2 Calibration
3.2.1 Auto kVp
1) Control panel의 Turn the power ON while pressing the "Collimator LED" button.
2) It matches the input values of kVp SET with output values of kVp Feedback
automatically.
3.2.2 Auto mA
1) It matches the input values of mA SET with output values of mA Feedback
automatically.
2) When it is 100mA output. it automatically adjusts V-F Frequency values
(mAs measured values) to 10,000 Hz.
3) 1 or 2 minutes after the completion of the auto calibration, it automatically
reboots and turns out to be the operational mode.
(FND Display)
3.2.3 Auto-Gain
1) Turn the power ON while pressing the "Collimator Laser" button.
2) When the auto-gain calibration is finished, there is a message shows up, “PUSH” on
the FND. Then press the hand switch with 2 steps, "Ready" and "Exposure".
4) 10 seconds of countdown and the auto exposure starts with a message, “OUT” on
the FND of the operational panel. Leave the room within 10 seconds.
52
ULTRA 100HF Operational & Service Manual
5) Through 5 times for the each section of kVp and mA, total 35 times of X-ray exposure,
correct the values with the feedback values.
6) 5 or 6 minutes after the auto calibration, it automatically reboots and becomes the
operation mode.
(FND Display)
3.2.4 Manual kVp
1) Turn the power ON while pressing the "Reverse" button.
2) The Auto kVp calibration runs about 4-5 minutes. Check the progress with the
connected oscilloscope.
3) When the auto kVp calibration is over, FND on the operational panel displays a
message, “PUSH”. Then press the hand switch with 2 steps, "Ready" and "Exposure".
53
ULTRA 100HF Operational & Service Manual
6) After the X-ray exposure, check the kVp value measured by the connected instrument
(oscilloscope) and if the kVp value is higher or lower than the set value, adjust the "kVp
gain" value with carrying out manual calibration.
7) If the kVp value measured by the oscilloscope is higher than 40 kVp, press the "kVp
Down" button on the operational panel once. Then the FND displays “40” and “-1” with
exposing X-ray. If you could 40 kVp correctly from the oscilloscope, press the Memory"
button to save the setting.
8) If the kVp value measured by the oscilloscope is lower than 40 kVp, press the "kVp UP"
button on the operational panel once. Then the FND displays “40” and “1” with exposing X-
ray. If you could 40 kVp correctly from the oscilloscope, press the Memory" button to save
the setting.
9) By saving the setting like above with 40 kVp, then the 50 kVp automatically becomes
changed. Check the each value of 40, 50, 60, 70, 80, 90 and 100kvp and adjust with the
same method.
10) After the adjustments on 40, 50, 60, 70, 80, 90, 100 kVp, it automatically reboots and
ready to operate again as following :
(Manual cal)
* There is no separate button for saving and it automatically saves 3 seconds after
entering the mode.
54
ULTRA 100HF Operational & Service Manual
3.3 DR Interface
In case of using wired detector, connect to an external output terminal of the exposure
signal. In order to use the external output terminal, you must first activate the external
output in the ULTRA 100HF.
1. Activating DR interface
Turn the power ON while pressing the "Patient size" buttons. Then “dRP on “ , “dRP off”
messages will be displayed and the DR interface will be activated.
55
ULTRA 100HF Operational & Service Manual
56
ULTRA 100HF Operational & Service Manual
57
ULTRA 100HF Operational & Service Manual
58
ULTRA 100HF Operational & Service Manual
Troubleshooting
4
4.1 Introduction
This chapter is intended to provide the information about the errors that may occur while
using ULTRA 100HF by displaying the error codes on the FND display so that user can
cope on the errors, but it is recommended to contact the manufacturer or the service
engineer in the event of any error occurs.
59
ULTRA 100HF Operational & Service Manual
ERROR
Symptom Solution
CODE
kV_LIMIT_OVER_ERROR Carry out CALIBRATION & REBOOT,
4 : Occurs when using min. or max. kVp. If the problem still appears, replace the main
(INVERTER FEED BACK) board.
mA_LIMIT_OVER_ERROR Carry out CALIBRATION & REBOOT,
5 : Occurs when using min. or max. mA. If the problem still appears, replace the main
(INVERTER FEED BACK) board.
READY_TIMEOVER_ERROR Replace the hand switch.
: If there is no exposure signal during If the problem still appears, replace the main
7
certain standby time after the input of board.
"READY" signal.
XRAY_TIMEOVER_ERROR Carry out CALIBRATION & REBOOT,
: If there is no exposed X-ray during If the problem still appears, replace the main
8
certain standby time after the input of board.
"Exposure" signal.
NO_EXP_OK_NO SIGNAL_Error Turn the power ON while pressing the patient
: Occurs if DR Interface is not connected.. size (Small) button. Then “dRP off”
9
and "DR interface is not used" messages will
be displayed on the FND display.
OVER_CPU_ERROR Carry out CALIBRATION & REBOOT,
10 : Occurs when using max. mA If the problem still appears, replace the main
board.
ALRAM_CPU_ERROR Carry out CALIBRATION & REBOOT,
20 : Occurs when there is an error on TEMP If the problem still appears, replace the main
or KV. board.
TEMP_FB_CPU_ERROR - If the error occurs when it is "COLD START",
: It occurs when the measured the temperature sensor has a problem so, it is
temperature is higher than the set required to replace the board.
25
temperature or if there is no measure - If the error occurs during operation, just turn
value on the Anode temperature during the power OFF and turn the power ON after
certain standby time. certain time.
MON_kV_OVER_ERROR Carry out CALIBRATION & REBOOT,
: Occurs if there is higher kVp values than If the problem still appears, replace the main
30
the set kVp values. board.
(CONTROL BOARD FEED BACK)
MON_kV_UNDER_ERROR Carry out CALIBRATION & REBOOT,
31 : Occurs if there is lower kVp values than If the problem still appears, replace the main
the set kVp values. board.
MON_mA_OVER_ERROR Carry out CALIBRATION & REBOOT,
: Occurs if there is higher mA values than If the problem still appears, replace the main
40
the set mA values. board.
(CONTROL BOARD FEED BACK)
MON_mA_UNDER_ERROR Carry out CALIBRATION & REBOOT,
: Occurs if there is lower mA values than If the problem still appears, replace the main
41
the set mA values. board.
(CONTROL BOARD FEED BACK)
MON_PRE_mA_OVER_ERROR Carry out CALIBRATION & REBOOT,
: Occurs if there is higher PRE_mA values If the problem still appears, replace the main
50
than the set PRE_mA values. board.
(CONTROL BOARD FEED BACK)
51 MON_PRE_mA_UNDER_ERROR Carry out CALIBRATION & REBOOT,
60
ULTRA 100HF Operational & Service Manual
: Occurs if there is lower PRE_mA values If the problem still appears, replace the main
than the set PRE_mA values. board.
(CONTROL BOARD FEED BACK)
SET_mA_ERROR Carry out CALIBRATION & REBOOT,
: Occurs if the CONTROL BOARD cannot If the problem still appears, replace the main
55 output as the value set in mA to the board.
inverter.
(CONTROL BOARD FEED BACK)
SET_kV_ERROR Carry out CALIBRATION & REBOOT,
: Occurs if the CONTROL BOARD If the problem still appears, replace the main
56 cannot output as the value set in KV to board.
the inverter.
(CONTROL BOARD FEED BACK)
FB_START_ERR_ERROR Carry out CALIBRATION & REBOOT,
: Occurs if there is no CURRENT FEED If the problem still appears, replace the main
57 BACK to the inverter for certain time after board.
the input of exposure signal.
(CONTROL BOARD FEED BACK)
OP_COMM_ERROR 1) Reconnect the communication line after
: Occurs when there is a communication turning the power OFF and reboot.
error with external devices 2) If the problem still appears, check the
60
(CONTROL BOARD FEED BACK) cables first.
3) If the problem still appears, replace the
main board.
Rotor_Current_ERROR Replace the ROTOR board.
: Occurs if there is an error on the current
61
of the rotor.
(ROTOR BOARD FEED BACK)
POWER_LEVEL_ERROR Check on the input power if it's correct or
: Occurs when there is an input power move on to another power. If the problem
70 error such as lack of power or higher keeps occurring, replace the main board.
power.
(CONTROL BOARD FEED BACK)
TUBE_TEMP_ERROR - Replace the temperature sensor.
: If there is no measured temperature on - If the error occurs during operation due to
the tube for certain time. high temperature of the tube, turn the power
80 If the measured temperature is higher OFF and turn the power ON after certain time.
than the working temperature.
(CONTROL BOARD or TUBE FEED
BACK)
DC_OUT_ERROR 1) Replace the main board.
90
: Occurs if more than 24 VDC output. 2) Replace the hand switch.
EEPROM_TIMEOVER_ERROR 1) Carry out CALIBRATION & REBOOT,
: Occurs if it is unable to recall the saved 2) If the problem still appears, replace the
100 values or save values. EEPROM.
(CONTROL BOARD FEED BACK) 3) If the problem still appears, replace the
main board.
EEPROM_VALUE_ERROR 1) Carry out CALIBRATION & REBOOT,
: Occurs if the set value is out of the 2) If the problem still appears, replace the
110 range. EEPROM.
(CONTROL BOARD FEED BACK) 3) If the problem still appears, replace the
main board.
61
ULTRA 100HF Operational & Service Manual
62
ULTRA 100HF Operational & Service Manual
Maintenance
5
5.1 Periodic Maintenance Schedule
Refer to the schedule below for information on when to perform periodic maintenance on
the crane.
Note : Due to varying operating conditions, the procedures listed below may have to be
performed at greater or lesser intervals. You may have to adjust intervals according to your
crane’s performance.
63
ULTRA 100HF Operational & Service Manual
Warning
All peripheral equipment must be turned off. You could get seriously
injured if you do not.
Warning
This equipment is not classified as anesthetic-proof and may ignite
inflammable anesthetics. Flammable agents used for skin cleaning or
disinfecting may also produce an explosion hazard.
Ensure the power has been disconnected and that the emergency switches have
been activated, before starting any cleaning operation..
Ensure no liquid gets into the unit.
Do not immerse the equipment, including any components or accessories, in
liquid.
Do not autoclave the equipment, including any component or accessories.
Do not use water. Water can short-circuit the electrical installation and cause
corrosion to mechanical parts.
Do not use acid or abrasive products.
Use only a dry cloth to clean chrome-plated parts.
For safety reasons, do not spray disinfectants.
Do not spray cleaning or disinfection solution directly on the equipment.
When disinfecting the examination room, ensure the unit is covered with plastic
sheets.
64
ULTRA 100HF Operational & Service Manual
Warning
All service and maintenance including the procedures described within this
chapter are to be performed by qualified service personnel only.
Warning
This equipment is not classified as anesthetic-proof and may ignite
inflammable anesthetics. Flammable agents used for skin cleaning or
disinfecting may also produce an explosion hazard.
Inspection Comments
65
ULTRA 100HF Operational & Service Manual
Note : If the product fails any portion of this operational check, stop the operational check
and correct the problem. Once the problem has been corrected, restart the operational
check from the beginning.
Check if the laser pointer and collimator lights are working fine at the "READY" of
the hand switch.
Check the condition of the angle indicator and if the indicator is working fine
without a problem.
66
ULTRA 100HF Operational & Service Manual
ITEM
Electrical insulating oil in HT tank. This is a mineral oil with trace additives
(1.5 Litre )
Solder (lead/tin).
Epoxy fiberglass circuit board materials, tracks are solder on copper.
Wire, tinned copper. Insulated with PVC, Insulation Rubber, or silicone.
Steel and / or aluminum (Product cabinet and console chassis).
Plastic (console enclosure and console membrane).
Electrical and electronic components: IC’s, transistors, diodes, resistors,
capacitors, etc.
Caution
Do not disassemble, incinerate or short-circuit the battery(s) in this product.
Do not put it in trash that is disposed of in landfills: dispose of it as required
by local ordinances.
The fluorescent lamp in the LCD display contains mercury. Do not put it in
trash that is disposed of in landfills: dispose of it as required by local
ordinances.
The LCD is made of glass. If the LCD breaks due to rough handing or
dropping, and the internal fluid gets in your eyes or on tour hands,
immediately wash the affected areas with water for at least 15 minutes.
Seek medical attention if any symptoms are present after washing.
67
ULTRA 100HF Operational & Service Manual
Electrical Schematics
6
6.1 Introduction
This chapter provides the electrical schematics of ULTRA 100HF.
Drawing Current
Drawing Description #Sheets
Number Rev
68
ULTRA 100HF Operational & Service Manual
- Block Diagram
69
ULTRA 100HF Operational & Service Manual
- Inverter
70
ULTRA 100HF Operational & Service Manual
- Main Control
71
ULTRA 100HF Operational & Service Manual
72
ULTRA 100HF Operational & Service Manual
73
ULTRA 100HF Operational & Service Manual
74
ULTRA 100HF Operational & Service Manual
75
ULTRA 100HF Operational & Service Manual
76
ULTRA 100HF Operational & Service Manual
77
ULTRA 100HF Operational & Service Manual
78
ULTRA 100HF Operational & Service Manual
79
ULTRA 100HF Operational & Service Manual
80
ULTRA 100HF Operational & Service Manual
81
ULTRA 100HF Operational & Service Manual
82
ULTRA 100HF Operational & Service Manual
- Power
83
ULTRA 100HF Operational & Service Manual
- Tank
84
ULTRA 100HF Operational & Service Manual
7.1 INTRODUCTION
7
This chapter contains the list of spare parts for the ULTRA 100HF of Diagnostic
radiography system.
The matrix on the next page lists the recommended spare parts for the LaVid FMTS
System of Diagnostic radiography system.
85
ULTRA 100HF Operational & Service Manual
SPARE TABLE
SUGGESTED
DESCRIPTION Part Number NOTE
QTY
Power cable SY-U-001 1
Power connector SY-U-002 1
Power switch SY-U-003 1
DC fan SY-U-004 1
Hand switch SY-U-005 1
Collimator lamp SY-U-006 1
Laser module SY-U-007 1
Noise filter SY-U-008 1
Main board SY-U-009 1
Power board SY-U-010 1
Inverter board SY-U-011 1
Mono block SY-U-012 1
Membrane sheet SY-U -013 1
Interface cable SY-U-014 1
86
ULTRA 100HF Operational & Service Manual
87