No part of this instruction manual may be reproduced, by any means, without the written consent of GEOKON. The
information contained herein is believed to be accurate and reliable. However, GEOKON assumes no responsibility
for errors, omissions or misinterpretation. The information herein is subject to change without notification.
The GEOKON® wordmark and logo are registered trademarks with the United States Patent and Trademark Office.
TABLE OF CONTENTS
1. INTRODUCTION............................................................................................................................................1
2. MODELS ................................................................................................................................................................2
I
6.4.2 CONNECTING SENSORS WITH BARE LEADS ................................................................... 17
7. DATA REDUCTION................................................................................................................................. 18
8. TROUBLESHOOTING ........................................................................................................................... 22
APPENDIX A. SPECIFICATIONS................................................................................................... 25
II
FIGURES
III
TABLES
IV
EQUATIONS
V
VI
1. INTRODUCTION
GEOKON vibrating wire embedment strain gauges are designed for direct
embedment in concrete. This can be accomplished by attaching the gauge to
rebar or tensioning cables and then casting the gauge into a concrete briquette,
which is subsequently cast into the structure, or grouting the gauge into
boreholes in the concrete.
Strains are measured using the vibrating wire principle. A length of steel wire is
tensioned between two end blocks that are firmly in contact with the mass
concrete. Deformations in the concrete will cause the two end blocks to move in
relation to each other, altering the tension in the steel wire. This change in
tension is measured as a change in the resonant frequency of vibration of the
wire.
Two coils, one with a magnet insert, the other with a pole piece insert, are
located close to the vibrating wire. In use, a pulse of varying frequency (swept
frequency) is applied to the coils causing the wire to vibrate primarily at its
resonant frequency.
Portable readouts and dataloggers are available from GEOKON. These models,
when used in conjunction with vibrating wire strain gauges, will provide the
necessary voltage pulses to pluck the wire. During vibration, a sinusoidal signal
is induced in the coils and transmitted to the readout box where it is conditioned
and displayed.
This manual contains installation instructions, readout and data reduction
procedures, and troubleshooting guidelines.
Note: Do not rotate or pull on the gauge end blocks, because this will alter the
readings and may cause permanent damage.
GEOKON vibrating wire strain gauges come in a variety of models, all easily
identifiable by the color of heat shrink which covers the protective tubes, as
shown in the table below.*
Model Heat Shrink Color
4200 (4,000 με Blue
4200L Black
4200 6 (5,000 με Red
4200 7 (10,000 με Green
4200X Yellow (dependent on the reason for the X designation)
The following sections describe in brief the various embedment strain gauges
available from GEOKON.
2 | MODELS | GEOKON
2.2 MODELS 4200-6 AND 4200-7
Model 4200-6 and 4200-7 strain gauges are supplied fully sealed and pre-
tensioned with the plucking coil mounted. Note the small collar under the shrink
tube at one end.
Collar
Collar
4 | MODELS | GEOKON
3. PRIOR TO INSTALLATION
Collar
Model 4200-6 / 4200-7 Note: Although the readings are taken on position D,
the digits shown must be converted to microstrain manually. To do this, multiply
the observed digit change by the gauge factor given on the calibration sheet
provided with the strain gauge.
Center Tensioning
Nut
4. The gauge end block will often turn with the tensioning nut. After making
the adjustment, grasp the end block and rotate it back to its original
position, so that the flats of the two end blocks are aligned. Remember to
hold the tube/coil assembly while doing this.
5. Check the reading. If okay, apply a spot of thread locking cement to
preserve the nut position and the tension.
Alternatively, the coil housing can be glued in place using cyanoacrylate glue. If
this method is chosen, it will no longer be possible to remove the gauge from
the coil housing.
1. Weld the two 9.5 x 51 mm (3/8 x 2") long hex bolts in place head down. The
bolts should be spaced at a nominal 530 mm (21") apart. A spacer jig is
available from GEOKON, or the cover plate can be flipped onto its back and
the holes in the cover plate can be used to mark the bolt locations. One hole
in the cover plate is slotted, so the spacing is not critical. Avoid welding
anywhere near the gauge as this will cause large local distortions in the
metal. Use either a special stud welder or an arc welder to weld the head of
the bolt to the surface.
2. Place the cover plate over the welded bolts.
3. Install washers, then nuts. Avoid excessive force while tightening the cover
retaining nuts, as this will distort the underlying steel surface and can give
rise to spurious strain readings.
■ Lighting arrestor boards and enclosures are available from GEOKON. These
units install where the instrument cable exits the structure being monitored.
The enclosure has a removable top to allow the customer to service the
components or replace the board in the event that the unit is damaged by a
lightning strike. A connection is made between the enclosure and earth
ground to facilitate the passing of transients away from the instruments.
See the figure below.
■ Plasma surge arrestors can be epoxied into the instrument cable, close to
the instrument. A ground strap then connects the surge arrestor to an earth
ground, such as a grounding stake or the steel structure.
Consult the factory for additional information on available lightning protection.
GEOKON strain gauges are typically set into concrete in one of two ways:
1. Cast the units directly into the concrete mix (see Section 5.1 through 5.3)
2. Cast them into briquettes that are subsequently cast into the concrete
structure (see Section 5.4).
When casting the gauge directly into the concrete mix, take care to avoid
applying any large forces to the end blocks during installation. This is most
imperative when installing Model 4202 gauges.
Model 4202 Note: Do not wrap an iron tie wire around the body of the gauge;
doing so could cause damage due to its delicate construction. Instead, use the
holes in the end blocks to affix the gauge to the rebar, being sure that the gauge
is not tensioned or compressed in the longitudinal direction.
Model 4200L Note: This gauge is specifically designed to allow strains to be
Nylon Cable Tie measured in curing concrete. However, do not bury the gauge more than one
Rebar or meter deep; doing so could damage the gauge.
Instrument Cable Tensioned Cable
Instrument Cable
Rebar or
4200 Tensioned Cable
Coil Assembly
Instrument Cable
Short Lengths
of Rebar
FIGURE 18: Alternative Method for Attaching Model 4200/4200HT Strain Gauges to
Rebar
MODEL 4200 SERIES VIBRATING WIRE STRAIN GAUGES | INSTALLING THE GAUGES IN CONCRETE | 13
5.4 USING PRE-CAST BRIQUETTES, SHOTCRETE, OR GROUTING
An alternative to casting the units into the concrete mix directly is to precast the
gauges into briquettes of the same mix as the mass concrete and then place the
briquettes in the structure prior to concrete placement. The briquettes should be
constructed no more than three days prior to installation, but no less than one
day prior, and should be continuously cured with water while awaiting
placement in the mass concrete.
GEOKON strain gauges may also be used in shotcrete, as well as in holes drilled
in rock or concrete that are subsequently grouted. When used in shotcrete,
special care should be taken to protect the lead wires, such as encasing them in
conduit or heavy tubing. The gauges may be placed by packing the immediate
area around the gauge by hand and then proceeding with the shotcrete
operation.
6.1 DATALOGGERS
Standard dataloggers, such as those from GEOKON or Campbell Scientific, can
read a vibrating wire instrument directly. GEOKON recommends using the
parameters in the table below when reading strain gauges with any datalogger.
It shows the recommended gauge type selection and gauge factor G entry to
convert to microstrain.
Alternately, if a calibration sheet is supplied with the strain gauge, the exact
values can be calculated from the start and end frequencies of the calibration.
Model 4200
Gauge Factor G Shown on calibration sheet
Start Frequency 4 (400 Hz)
End Frequency 12 (1200 Hz)
FIGURE 20: Lemo Connector to GK-404 ■ The current position: set by the Pos button, displayed as A through F.
■ The current reading: set by the Mode button, displayed as a numeric value
followed by the unit of measure.
Use the Pos and Mode buttons to select the correct position and display units
for the model of equipment purchased.
The GK-404 will continue to take measurements and display readings until the
unit is turned off, either manually or by the Auto-Off timer (if enabled).
For more information, consult the GK-404 manual.
■ Flying leads with alligator clips, if the sensor cable terminates in bare wires.
■ A 10 pin connector.
FIGURE 21: GK-405 Readout 6.4.1 CONNECTING SENSORS WITH 10-PIN BULKHEAD CONNECTORS
ATTACHED
Align the grooves on the sensor connector (male), with the appropriate
connector on the readout (female connector, labeled senor or load cell). Push
the connector into place, and then twist the outer ring of the male connector
until it locks into place.
1. Connect an ohmmeter to the green and white thermistor leads coming from
the instrument. Since the resistance changes with temperature are large,
the effect of cable resistance is usually insignificant. For long cables a
correction can be applied, equal to approximately 14.7Ω per 1000 feet
(48.5Ω per km) at 20 °C. Multiply these factors by two to account for both
directions.
2. Look up the temperature for the measured resistance in Appendix C. For the
4200HT use Appendix D.
The table below shows the readout position, theoretical gauge factors, and
experimental data derived from batch calibrations for each model of strain
gauge. (Individual calibrations are available at an additional cost; contact
GEOKON for more information.)
Model Readout Position Theoretical Gauge Factor Typical Batch Factor1 Typical Gauge Factor1 Experimental Data
4200 D 3.304 0.97 to 0.98 N/A 3.237
4200 6 D N/A N/A N/A None
4200 7 D N/A N/A N/A None
4202 E 0.391 0.91 N/A 0.356
4204 A N/A N/A 1.422 N/A
4210 B N/A N/A 0.3568 0.3423
4212 B N/A N/A 0.3624 N/A
4214 B N/A N/A 0.3665 N/A
1
The exact Gauge Factor is determined by actual calibration, there is no Batch Factor to apply.
TABLE 5: Embedment Strain Gauge Factors
■ Model 4204
■ GK-403 and GK-405 (convert the period to digits using this formula:)
2
Hz
Digits = -------
1000
-
2 –3
theory = G f x10
Where:
f is the frequency in digits.
G is the theoretical gauge factor, equal to 3.304 for the 4200 gauge and 0.3910
for the 4202 gauge.
In practice, the act of clamping shortens the vibrating wire slightly, causing it to
over-register the strain. You can compensate for this by applying the batch
gauge factor supplied with each gauge. With the batch gauge factor applied, the
apparent change in strain shown on the readout box is equal to:
apparent = R 1 – R 0 B
Note: For some gauges the reading may fluctuate by one digit, so this
resolution may not be useful.
T1 – T0 C1 –C2
EQUATION 4: Correction for Temperature Effects on the Gauge
Where:
T0 is the initial temperature.
T1 is the current temperature.
C1 is the coefficient expansion of steel: 12.2 microstrains/°C.
(C1 for Model 4200HT gauges is 17.3 microstrains/°C.)
C2 is the coefficient of expansion of concrete: ~10 microstrains/°C. (Users
should use their own values for C2 if known.)
Load-related strain in concrete (a composite of both external load and
temperature effects) corrected for temperature, is given by:
load = R 1 – R 0 B + T 1 – T 0 C 1 – C 2
Where:
R0 is the initial reading.
R1 is the current reading from the readout box, taken in position D or E.
If:
R0 = 3000 in position D
R1 = 2900 in position D
T0 = 20 °C
T1 = 30 °C
B = 0.975 (batch calibration factor)
Then:
The apparent strain = (2900 – 3000) 0.975 = –97.5 strain (compression).
The load-related strain, corrected for temperature effects on the gauge =
(2900 – 3000) 0.975 + (30 – 20) (12.2 – 10) = –75.5 mstrain (compression).
Note: The actual strain undergone by the concrete, (i.e., that which would be
measured by a measuring scale) is given by the formula:
actual = (R1 – R0) B + (T1 – T0) (C1)
Which in the current example = (2900 – 3000) 0.975 + (30 – 20) (12.2) = 24.5
strain (expansion).
See Appendix F for further information.
L = EA
Where:
L is the load.
E is the elastic modulus of the structural element in the appropriate units.
µ is the strain in microstrain.
A is the cross-sectional area in the appropriate units.
□ There may be an open circuit. Check all connections, terminals, and plugs. If
a cut is located in the cable, splice according to instructions in Section 4.1.
SYMPTOM: THERMISTOR RESISTANCE IS TOO LOW
□ There may be a short. Check all connections, terminals, and plugs. If a short
is located in the cable, splice according to instructions in Section 4.1.
□ Water may have penetrated the interior of the transducer. There is no
remedial action.
SYMPTOM: STRAIN GAUGE READINGS ARE UNSTABLE
□ Is the cable cut or crushed? Check the resistance of the cable by connecting
an ohmmeter to the sensor leads. Table 8 below shows the expected
resistance for the various wire combinations. Table 9 below is provided for
you to fill in the actual resistance found. Cable resistance is approximately
14.7Ω per 1000 ft (48.5Ω per km) of 22 AWG wire. Multiply this factor by
two to account for both directions.
□ If the resistance is very high or infinite, the cable is probably broken or cut. If
the resistance is very low, the gauge conductors may be shorted. If a cut or
a short is located in the cable, splice according to the instructions in Section
4.1.
□ Does the readout or datalogger work with another gauge? If not, it may
have a low battery or possibly be malfunctioning.
22 | TROUBLESHOOTING | GEOKON
Vibrating Wire Sensor Lead Grid - SAMPLE VALUES
Red
Black
White
Green
Shield
Notes:
1Possible resolution depends on the readout; the figures in the above table
pertain to the GK-404.
2
System Accuracy takes into account hysteresis, nonlinearity, misalignment,
batch factor variations, and other aspects of the actual measurement program.
System Accuracy to ±1.0% FS may be achieved through individual calibration
of each strain gauge.
3
Model 4200HT-T high-temperature strain gauge has a Thermal Coefficient of
17.3 / °C.
A.2 THERMISTOR
For more information, refer to Appendix C.
Range: -80 to +150 °C
Accuracy: ±0.5 °C
Where:
Lw is the length of the wire in inches.
F is the wire tension in pounds.
m is the mass of the wire per unit length (pounds, seconds2/inches2).
2. Note that:
Where:
W is the weight of Lw inches of wire in pounds.
g is the acceleration of gravity (386 inches/seconds2).
3. And:
Where:
ρ is the wire material density (0.283 pounds/inches3).
a is the cross-sectional area of the wire in inches2.
4. Combining the equations from steps one, two, and three gives:
5. Note that the tension (F) can be expressed in terms of strain, e.g.,
Where:
Ɛw is the wire strain (inches/inches).
E is the Young's Modulus of the wire (30 x 106 Psi).
10. The equation from step nine must now be expressed in terms of the strain in
the surface of the body to which the gauge is attached. Since the
deformation of the body must equal the deformation of the wire:
Where:
Ɛ is the strain in the body.
Lg is the gauge length in inches.
11. Combining the equations from steps nine and ten gives:
13. The display on position 'D' of the readout is based on the equation:
The squaring, inverting, and multiplication by the factor 3.304 x 109 is all done
internally by the microprocessor of the readout so that the displayed reading in
position D is given in microinches per inch (ε).
Note: In the previous steps, T is seconds x 106 and ε is microinches per inch.
An alternative is: ε = 3.304 x 10-3 f2 microstrain.
Where f is the frequency in Hz.
Ohms Temp Ohms Temp Ohms Temp Ohms Temp Ohms Temp
201.1K -50 15.72K -9 2221 32 474.7 73 137.2 114
187.3K -49 14.90K -8 2130 33 459.0 74 133.6 115
174.5K -48 14.12K -7 2042 34 444.0 75 130.0 116
162.7K -47 13.39K -6 1959 35 429.5 76 126.5 117
151.7K -46 12.70K -5 1880 36 415.6 77 123.2 118
141.6K -45 12.05K -4 1805 37 402.2 78 119.9 119
132.2K -44 11.44K -3 1733 38 389.3 79 116.8 120
123.5K -43 10.86K -2 1664 39 376.9 80 113.8 121
115.4K -42 10.31K -1 1598 40 364.9 81 110.8 122
107.9K -41 9796 0 1535 41 353.4 82 107.9 123
101.0K -40 9310 1 1475 42 342.2 83 105.2 124
94.48K -39 8851 2 1418 43 331.5 84 102.5 125
88.46K -38 8417 3 1363 44 321.2 85 99.9 126
82.87K -37 8006 4 1310 45 311.3 86 97.3 127
77.66K -36 7618 5 1260 46 301.7 87 94.9 128
72.81K -35 7252 6 1212 47 292.4 88 92.5 129
68.30K -34 6905 7 1167 48 283.5 89 90.2 130
64.09K -33 6576 8 1123 49 274.9 90 87.9 131
60.17K -32 6265 9 1081 50 266.6 91 85.7 132
56.51K -31 5971 10 1040 51 258.6 92 83.6 133
53.10K -30 5692 11 1002 52 250.9 93 81.6 134
49.91K -29 5427 12 965.0 53 243.4 94 79.6 135
46.94K -28 5177 13 929.6 54 236.2 95 77.6 136
44.16K -27 4939 14 895.8 55 229.3 96 75.8 137
41.56K -26 4714 15 863.3 56 222.6 97 73.9 138
39.13K -25 4500 16 832.2 57 216.1 98 72.2 139
36.86K -24 4297 17 802.3 58 209.8 99 70.4 140
34.73K -23 4105 18 773.7 59 203.8 100 68.8 141
32.74K -22 3922 19 746.3 60 197.9 101 67.1 142
30.87K -21 3748 20 719.9 61 192.2 102 65.5 143
29.13K -20 3583 21 694.7 62 186.8 103 64.0 144
27.49K -19 3426 22 670.4 63 181.5 104 62.5 145
25.95K -18 3277 23 647.1 64 176.4 105 61.1 146
24.51K -17 3135 24 624.7 65 171.4 106 59.6 147
23.16K -16 3000 25 603.3 66 166.7 107 58.3 148
21.89K -15 2872 26 582.6 67 162.0 108 56.8 149
20.70K -14 2750 27 562.8 68 157.6 109 55.6 150
19.58K -13 2633 28 543.7 69 153.2 110
18.52K -12 2523 29 525.4 70 149.0 111
17.53K -11 2417 30 507.8 71 145.0 112
16.60K -10 2317 31 490.9 72 141.1 113
Where:
T = Temperature in °C
LnR = Natural Log of Thermistor Resistance
A = 1.02569 x 10-3
B = 2.369 x 10-4
C = 1.019 x 10-7
Note: Coefficients calculated over the -50 to +150 C span.
Ohms Temp LnR LnR3 Calc. Temp Diff F.S. Error Ohms Temp LnR LnR3 Calc. Temp Diff F.S. Error
113898 -30 11.643 1578.342 -30.17 0.17 0.06 407.62 120 6.010 217.118 120.00 0.00 0.00
86182 -25 11.364 1467.637 -25.14 0.14 0.05 360.8 125 5.888 204.162 125.00 0.00 0.00
65805 -20 11.094 1365.581 -20.12 0.12 0.04 320.21 130 5.769 191.998 130.00 0.00 0.00
50684.2 -15 10.833 1271.425 -15.10 0.10 0.03 284.95 135 5.652 180.584 135.00 0.00 0.00
39360 -10 10.581 1184.457 -10.08 0.08 0.03 254.2 140 5.538 169.859 140.01 -0.01 0.00
30807.4 -5 10.336 1104.068 -5.07 0.07 0.02 227.3 145 5.426 159.773 145.02 -0.02 -0.01
24288.4 0 10.098 1029.614 -0.05 0.05 0.02 203.77 150 5.317 150.314 150.03 -0.03 -0.01
19294.6 5 9.868 960.798 4.96 0.04 0.01 183.11 155 5.210 141.428 155.04 -0.04 -0.01
15424.2 10 9.644 896.871 9.98 0.02 0.01 164.9 160 5.105 133.068 160.06 -0.06 -0.02
12423 15 9.427 837.843 14.98 0.02 0.01 148.83 165 5.003 125.210 165.08 -0.08 -0.03
10061.4 20 9.216 782.875 19.99 0.01 0.00 134.64 170 4.903 117.837 170.09 -0.09 -0.03
8200K 25 9.012 731.893 25.00 0.00 0.00 122.1 175 4.805 110.927 175.08 -0.08 -0.03
6721.54 30 8.813 684.514 30.01 -0.01 0.00 110.95 180 4.709 104.426 180.07 -0.07 -0.02
5540.74 35 8.620 640.478 35.01 -0.01 0.00 100.94 185 4.615 98.261 185.10 -0.10 -0.04
4592 40 8.432 599.519 40.02 -0.02 -0.01 92.086 190 4.523 92.512 190.09 -0.09 -0.03
3825.3 45 8.249 561.392 45.02 -0.02 -0.01 84.214 195 4.433 87.136 195.05 -0.05 -0.02
3202.92 50 8.072 525.913 50.01 -0.01 -0.01 77.088 200 4.345 82.026 200.05 -0.05 -0.02
2693.7 55 7.899 492.790 55.02 -0.02 -0.01 70.717 205 4.259 77.237 205.02 -0.02 -0.01
2276.32 60 7.730 461.946 60.02 -0.02 -0.01 64.985 210 4.174 72.729 210.00 0.00 0.00
1931.92 65 7.566 433.157 65.02 -0.02 -0.01 59.819 215 4.091 68.484 214.97 0.03 0.01
1646.56 70 7.406 406.283 70.02 -0.02 -0.01 55.161 220 4.010 64.494 219.93 0.07 0.02
1409.58 75 7.251 381.243 75.01 -0.01 0.00 50.955 225 3.931 60.742 224.88 0.12 0.04
1211.14 80 7.099 357.808 80.00 0.00 0.00 47.142 230 3.853 57.207 229.82 0.18 0.06
1044.68 85 6.951 335.915 85.00 0.00 0.00 43.673 235 3.777 53.870 234.77 0.23 0.08
903.64 90 6.806 315.325 90.02 -0.02 -0.01 40.533 240 3.702 50.740 239.69 0.31 0.11
785.15 95 6.666 296.191 95.01 -0.01 0.00 37.671 245 3.629 47.788 244.62 0.38 0.13
684.37 100 6.528 278.253 100.00 0.00 0.00 35.055 250 3.557 45.001 249.54 0.46 0.16
598.44 105 6.394 261.447 105.00 0.00 0.00 32.677 255 3.487 42.387 254.44 0.56 0.19
524.96 110 6.263 245.705 110.00 0.00 0.00 30.496 260 3.418 39.917 259.34 0.66 0.23
461.91 115 6.135 230.952 115.00 0.00 0.00
TABLE 13: Thermistor Resistance for 8.2kΩ
1
T= – 273.15
A+B(LnR)+C(LnR) 3 +D(LnR) 5
EQUATION 10: Thermistor Resistance for 10KΩ
Where:
T = Temperature in °C
LnR = Natural Log of Thermistor Resistance
A = 1.127670 x 10-3
B = 2.344442 x 10-4
C = 8.476921 x 10-8
D = 1.175122 x 10-11
Note: Coefficients optimized for a curve J Thermistor over the temperature
range of 0 °C to +250 °C.
Ohms Temp Ohms Temp Ohms Temp Ohms Temp Ohms Temp Ohms Temp Ohms Temp Ohms Temp
32,650 0 7,402 32 2,157 64 763.5 96 316.6 128 148.4 160 76.5 192 42.8 224
31,029 1 7,098 33 2,083 65 741.2 97 308.7 129 145.1 161 75.0 193 42.1 225
29,498 2 6,808 34 2,011 66 719.6 98 301.0 130 142.0 162 73.6 194 41.4 226
28,052 3 6,531 35 1,942 67 698.7 99 293.5 131 138.9 163 72.2 195 40.7 227
26,685 4 6,267 36 1,876 68 678.6 100 286.3 132 135.9 164 70.8 196 40.0 228
25,392 5 6,015 37 1,813 69 659.1 101 279.2 133 133.0 165 69.5 197 39.3 229
24,170 6 5,775 38 1,752 70 640.3 102 272.4 134 130.1 166 68.2 198 38.7 230
23,013 7 5,545 39 1,693 71 622.2 103 265.8 135 127.3 167 66.9 199 38.0 231
21,918 8 5,326 40 1,637 72 604.6 104 259.3 136 124.6 168 65.7 200 37.4 232
20,882 9 5,117 41 1,582 73 587.6 105 253.1 137 122.0 169 64.4 201 36.8 233
19,901 10 4,917 42 1,530 74 571.2 106 247.0 138 119.4 170 63.3 202 36.2 234
18,971 11 4,725 43 1,480 75 555.3 107 241.1 139 116.9 171 62.1 203 35.6 235
18,090 12 4,543 44 1,432 76 539.9 108 235.3 140 114.5 172 61.0 204 35.1 236
17,255 13 4,368 45 1,385 77 525.0 109 229.7 141 112.1 173 59.9 205 34.5 237
16,463 14 4,201 46 1,340 78 510.6 110 224.3 142 109.8 174 58.8 206 33.9 238
15,712 15 4,041 47 1,297 79 496.7 111 219.0 143 107.5 175 57.7 207 33.4 239
14,999 16 3,888 48 1,255 80 483.2 112 213.9 144 105.3 176 56.7 208 32.9 240
14,323 17 3,742 49 1,215 81 470.1 113 208.9 145 103.2 177 55.7 209 32.3 241
13,681 18 3,602 50 1,177 82 457.5 114 204.1 146 101.1 178 54.7 210 31.8 242
13,072 19 3,468 51 1,140 83 445.3 115 199.4 147 99.0 179 53.7 211 31.3 243
12,493 20 3,340 52 1,104 84 433.4 116 194.8 148 97.0 180 52.7 212 30.8 244
11,942 21 3,217 53 1,070 85 421.9 117 190.3 149 95.1 181 51.8 213 30.4 245
11,419 22 3,099 54 1,037 86 410.8 118 186.1 150 93.2 182 50.9 214 29.9 246
10,922 23 2,986 55 1,005 87 400.0 119 181.9 151 91.3 183 50.0 215 29.4 247
10,450 24 2,878 56 973.8 88 389.6 120 177.7 152 89.5 184 49.1 216 29.0 248
10,000 25 2,774 57 944.1 89 379.4 121 173.7 153 87.7 185 48.3 217 28.5 249
9,572 26 2,675 58 915.5 90 369.6 122 169.8 154 86.0 186 47.4 218 28.1 250
9,165 27 2,579 59 887.8 91 360.1 123 166.0 155 84.3 187 46.6 219
8,777 28 2,488 60 861.2 92 350.9 124 162.3 156 82.7 188 45.8 220
8,408 29 2,400 61 835.4 93 341.9 125 158.6 157 81.1 189 45.0 221
8,057 30 2,316 62 810.6 94 333.2 126 155.1 158 79.5 190 44.3 222
7,722 31 2,235 63 786.6 95 324.8 127 151.7 159 78.0 191 43.5 223
GEOKON's Model 4200-4 No Stress Strain Enclosure is made of double PVC walls
and is filled with polystyrene foam. The figure below shows a standard 4200-4
installation.
Instrument Cable
Concrete
Wire Tie
Coil Assembly
4200
Wire Tie
Enclosure Wall
Polystyrene Foam
1. Position the enclosure inside the concrete mass with the top open so that it
can be easily filled with concrete. (Often the 4200-4 will be placed next to a
strain gauge rosette.)
2. Mount the strain gauge by wrapping two turns of tie wire around the gauge
and then feeding the ends of the wire through the holes in the opposite
sides of the enclosure.
3. Wrap the ends of the tie wire around the outside of the enclosure and tie
together.
4. Use a tie wire tightening tool or similar device to tighten the wire.
MODEL 4200 SERIES VIBRATING WIRE STRAIN GAUGES | NO STRESS STRAIN ENCLOSURE | 31
5. Repeat this procedure, using another tie wire and the second set of holes in
the enclosure.
6. Arrange the instrument cable so that it exits through the top of the
enclosure.
7. When pouring concrete into the enclosure, remove aggregate that is too
large. Be careful not to disturb the gauges during the filling process.
Data interpretation is the same as that outlined in Section 7, with the exception
of the Temperature Coefficient, which for the 4200HT-T is 17.3 με/ °C. Coil
resistance for Model 4200HT-T is 50 ohms.
The thermistor included with the gauge is a high-temperature thermistor, which
uses the resistance to temperature conversion table shown in Table 14. When
using a readout box, it is not necessary to connect the green and white wires to
the readout because the temperature shown on the readout will be incorrect.
Instead, use a digital ohmmeter.
TO READ TEMPERATURES USING AN OHMMETER
1. Connect an ohmmeter to the green and white thermistor leads coming from
the strain gauge. (Since the resistance changes with temperature are large,
the effect of cable resistance is usually insignificant.)
2. Look up the temperature for the measured resistance in Table 14.
MODEL 4200 SERIES VIBRATING WIRE STRAIN GAUGES | MODEL 4200HT-T HIGH-TEMPERATURE STRAIN GAUGE | 33
APPENDIX F. MEASUREMENT AND CORRECTION OF
TEMPERATURE EFFECTS
If the ends of the structural member are free to expand or contract without
restraint, strain changes can take place without any change in stress. On the
other hand, if the ends of the structural member are restrained by some semi-
rigid medium, then any increase in temperature of the structural member will
result in a buildup of compressive load-related strain in the member, even
though the actual strain would be tensile.
The reason for this is because the member is restrained from expansion but the
vibrating wire is not. An increase in temperature will produce expansion in the
vibrating wire, which in turn will cause a reduction in wire tension. This results in
a decrease in the vibrational frequency. The magnitude of this temperature-
induced, compressive strain increase would be measured accurately by the
strain gauge, and can be determined using Equation 5.
These temperature-induced stresses can be separated from any external, load-
induced stresses by reading both the strain and temperature of the gauge at
frequent intervals. These readings should take place during a period when the
external loading from construction activity remains unchanging. When these
strain changes are plotted against the corresponding temperature changes, the
resulting graph will show a straight-line relationship, the slope of which yields a
factor KT microstrain/degree. This factor can be used to calculate the
temperature-induced stress, as shown by the following equation:
This can be subtracted, if desired, from the combined load related stress change
using the following equation:
To give that part of the stress change due to construction activity loads only, use
the following equation:
Note that the correction factor (KT) may change with time and with construction
activity as the rigidity of the restraint may change. In such a case, it would be
advisable to calculate a new temperature correction factor by repeating the
above procedure.
If, for whatever reason, the actual strain of the concrete member is required
(e.g., the change of unit length that would be measured by a dial gauge
attached to the surface), this is given by the equation:
MODEL 4200 SERIES VIBRATING WIRE STRAIN GAUGES | MEASUREMENT AND CORRECTION OF TEMPERATURE EFFECTS | 35
GEOKON Phone: +1 (603) 448-1562 GEOKON
48 Spencer Street Email: info@geokon.com is an ISO 9001:2015
Lebanon, New Hampshire Website: www.geokon.com registered company
03766, USA