Вы находитесь на странице: 1из 12

Fluke 5790A/5790B

AC Measurement Standard

Mainframe Calibration Station Manual

Revision 3.0 – July, 2015


© Fluke Corporation.
Fluke 5790A/5790B Calibration Station Manual

Contents
Scope .............................................................................................................................................................. 3
Revision History .............................................................................................................................................. 3
System Overview ............................................................................................................................................ 3
Document Overview ....................................................................................................................................... 3
Referenced Documents .................................................................................................................................. 4
General System Requirements ....................................................................................................................... 4
System Components List ................................................................................................................................ 5
Station Accessories ......................................................................................................................................... 5
Software Environment.................................................................................................................................... 6
Environmental conditions .............................................................................................................................. 6
Personnel Requirements ................................................................................................................................ 6
Controlling Documents ................................................................................................................................... 6
Installation/Configuration .............................................................................................................................. 7
Operation ....................................................................................................................................................... 8
System Maintenance ...................................................................................................................................... 9
System Components Calibration .................................................................................................................. 10
Appendix A (Maintenance examles)............................................................................................................. 11
Appendix B (MET/CAL procedure map)........................................................................................................ 12

Danaher TM Company Confidential Page 2


Fluke 5790A/5790B Calibration Station Manual

Scope
To document the specific requirements for installation, operation and maintenance of the Fluke 5790A/5790B Input 1
and Input 2 MET/CAL Calibration Station at authorized service locations.

Revision History
1.0 Initial Release (November 3, 2013). Developed at Fluke Everett Service Center.
2.0 Additional clarifications on Input 2 and procedure map update in the appendix.
3.0 Update for Fluke 5790B release.

System Overview
The Fluke 5790X MET/CAL Calibration Station is designed to verify the performance of Fluke 5790A & Fluke 5790B
Mainframe (Device Under Test, abbreviated later in the text as “DUT”) to its published specifications with acceptable test
uncertainties for all verification parameters. The Calibration Station uses a Fluke 792A, Fluke 5730A/5720A/5700A (for
simplicity 57X0A is used in the document), Fluke 5725A, Agilent/Hewlett Packard 3458A or Fluke 8508A (referred to as
DMM in this document), Fluke PM6690 (optional. Can be replaced by 5730A with 25µHz/Hz oscillator), Fluke 752A, Fluke
732A/732B, Fluke 845A and MET/CAL procedures to perform an automated DUT verification. The automated software
monitors/logs/launches/sets tasks that are time sensitive including:

 Station maintenance reminder and instructions (once a week)


 DCV characterization of the source (Fluke 57X0A) – due every 30 days
 Time sensitive sections that are pre-requisite within the process.
o SL021 characterization within 96 hrs. of 2.2-22mV section.
o AC/DC for outputs ≥ 200mV within 96 hrs. of the Absolute run for outputs ≥ 60mV

Document Overview
This manual contains proprietary information. This manual and the software provided are property of Fluke Calibration and
should not be shared internally or externally without permission. Topics addressed in this manual are:
 Reference documentation list
 System requirements
 System component list
 Installation
 System operation
 Software overview
 Personnel requirements
 Accessories list
 Maintenance

Information contained in individual commercial instrument manuals, is not provided with the system. Please consult the
respective instrument manuals for detailed information.

Danaher TM Company Confidential Page 3


Fluke 5790A/5790B Calibration Station Manual
Referenced Documents
List of Referenced documents used in the station/uncertainty development.

Title Source

Fluke Metrology Software User’s Manual Fluke

MET/CAL Reference Manual Fluke

Fluke 8508A User Manual, PN 1673798, July 2002 Rev. 6, 3/13 Fluke
Agilent Technologies 3458A Multimeter User’s Guide Manual Part Number: 03458-90014, Rev Agilent
B.01 Issued Date: March 2001
Fluke 5790A Operators Manual, PN 893284 January 1992 Rev.4, 1/06 Fluke

Fluke 5790A Service Manual, PN 893292 January 1992 Rev.5, 1/07 Fluke

Fluke 5790A/AF Addendum, PN 857743 March 1992 Rev.4, 5/03 Fluke

Fluke 5790B Operators Manual (pending release at last update of this document) Fluke

Fluke 752A Instruction Manual, P/N 64509 May 1983, Rev. 4/84 Fluke

Fluke 792A Instruction Manual, P/N 871723 July 1990, Rev. 1 12/92 Fluke

Fluke 732A Instruction Manual, P/N 645051 May 1983 Fluke

Fluke 732B/734A Instruction Manual PN 869321 March 1992 Rev.3, 11/97 Fluke

Fluke 845A/845AB Instruction Manual P/N 294173, November 1967 Fluke

Fluke PM 6690 Fluke

Fluke 5790x Main Input Uncertainty Analysis.xlsx Fluke

General System Requirements


The Calibration Station consists of system components (list provided in table below). The station does not require a
custom built rack to hold the equipment provided that the following requirements are satisfied:

 Setup is station specific as long as it can accommodate DUT Mainframe inputs and Fluke 792A alignment
where the two devices are placed against each other to allow direct connection through type N Tee adapter. The
size of the cart and positioning of the DUT and 792A is such that their inputs are positioned at 90° angle
relative to the output of the source. Raisers (if needed) are used to elevate the 792A to the input under test
(Input 1, Input 2, Shunt input).
 The following equipment additional equipment is used and configuration and stacking is determined by the
carts being used:
Fluke 57X0A (bottom) Frequency counter
Fluke 5725A DMM(bottom)
 Main Power: Country specific (requires site specific hardware compatibility and set up). Minimum
requirements: 115VAC or 230VAC (max fluctuations of ±10%). Line frequency: 50 Hz to 60 Hz (with
maximum fluctuation of ± 3 Hz)
Danaher TM Company Confidential Page 4
Fluke 5790A/5790B Calibration Station Manual

System Components List


Manufacturer Model/Requirements Description
Fluke 8508A Reference Multimeter
Agilent 3458A Reference Multimeter (optional. Instead of Fluke 8508A)
Fluke PM6690 Timer/Counter (Can be substituted by 5730A with 25µHz/Hz osc.)
Tektronix FCA3000 series Alternatively Tektronix counter can be used instead of PM6690
Fluke 792A AC/DC Transfer Standard
Fluke 5720A/5700A Calibrator
Fluke 5725A Amplifier
Fluke 752A Reference Divider
Fluke 732A/732B DC Reference Cell
Fluke 845A/AB Null Detector
PC Specified below System Controller
Fluke MET/BASE FLUKE METROLOGY SOFTWARE (Version-site specific)
National Instruments GPIB PCI GPIB CONTROLLERS, QUANTITY. 2
Fluke N/A Custom knob turner for 792A. (optional)

Station Accessories
The station requires use of specified cables adapters to ensure measurements are carried within the specified uncertainty.
Included below is the list of the required accessories:

 Fluke SL021 (Fluke EPEL build reference divider with 12” shielded cable). Characterized before use with each
DUT. Quantity: 1
 36” two conductor shielded patch cable (Pomona 1167-36). Quantity: 1
 50 Ω Type N Tee (M-M-M). Quantity: 1
 Type N female to banana male adapter (Pomona 1740 or equivalent E-Z-HOOK). Quantity: 1
 Type N male to BNC female adapter. Quantity: 1
 Type N female to Type N female coupler. Quantity: 1
 BNC male to banana female adapter (Pomona 1469). Quantity: 1
 24” double banana leads (Pomona 2BA-24). Quantity: 2
 Heavy braid guard cable (in house made from braid spool). Quantity: 1
 ≈36” coaxial cable with choke (in house made using: 54” RG58 cable, Pomona 4898 double banana plug (Qty.
2), ferrite core, Fluke Part No. 474908). The cable is wound around the ferrite 5 times, thus resulting in ≈ 36”.
The cable is used to connect Fluke 792A DC output to 3458A Input for measurements ≤ 20mV.

Danaher TM Company Confidential Page 5


Fluke 5790A/5790B Calibration Station Manual

Software Environment
The automated procedure requires on-site installation of Fluke Calibration software MET/CAL (Revision 8.051 or later). The
system uses Fluke MET/CAL together with custom MET/CAL procedures saved as executables (MET/CAL pxe) - provided
to automate performance testing of Fluke 5790A and Fluke 5790B Mainframe to its published specifications. The automated
MET/CAL software uses csv (comma separated excel based file with a csv extension embedded in the pxe) file to propagate
DUT tolerance and measurement uncertainty. The csv file comes from a worksheet in the uncertainty analysis workbook. It is
the responsibility of each site to ensure that reported uncertainty is aligned with the site CMC claims and that non accredited
measurement are marked as such (driven through the uncertainty analysis file via the csv).

Environmental conditions
The calibration system and its uncertainty are designed to be used in a Calibration Laboratory with the following
environmental conditions:
 Temperature: 23°C ± 3°C
 Humidity: 15%RH to 70%RH
 Altitude: 0m to 3000m

Personnel Requirements
Personnel qualified to work on this station shall have sufficient Metrology Training (both theoretical and practical) in
ECAL products calibrations. The training compliance shall be documented under the site specific quality system. In
addition the operators of the Fluke 5790X Calibration Station shall be familiar with this document and the requirements
set within this document and the automated software instructions to operate the system.

Controlling Documents
The verification points included in the automated software are as specified by the commercially available Fluke 5790A
Service Manual listed in the References section of this document as well as the proposed verification points for Fluke
5790B (Service Manual not released at time of this revision). Devices sold under contract with the Air Force follow the
Fluke 5790A/AF Addendum listed in the Reference section of this document.

Fluke Calibration Company Confidential Page 6


Fluke 5790A/5790B Calibration Station Manual
Installation/Configuration

Fluke 5790X Main Input Uncertainty Analysis .xlsx file provided with this system drives the uncertainty reported by the
software. Review by Metrology is required for each site installation. Edit the file as follows:
Worksheet “Uncertainty”
 Edit column “Best Uncertainty” in the “Uncertainty” and “Unc 2mV – 20mV (100mV)” worksheets to reflect
the CMC of the site specific Scope of Accreditation.
 Edit any individual component of uncertainty if necessary to reflect site specific measurement uncertainty.
 Edit Colum “Accred. Step” in both worksheets to enable or disable the points accordingly to your Scope of
Accreditation. “1” stands for point that is accredited. “0” stands for non-accredited points.
Worksheet “Revision” – Document the changes to the document specific to your site. Save the xlsx file.
Open worksheet “5790_Main_Propagation” and save as “5790_Main_Propagation .csv” in the root directory of the
MET/CAL solution. Once the csv file is updated for lab capability each one of the project has to be rebuild into pxe file.
The csv file is embedded in each project and it is accessed during MET/CAL Run Time execution.

MET/CAL Station configuration (config.dat)


The following GPIB instruments have to be set on the station:

 DMM
Fluke 8508A – Choose from MET/CAL available list – “Fluke 8508A”. The alias will be set automatically
to 8508. Do add additional alias: “DMM” – required by this software. Set the MET/TRACK ID for this
device.
OR
Agilent/Hewlett-Packard 3458A – Choose from MET/CAL available list – “HP 3458A”. The alias will be
set automatically to 3458. Do add additional alias: “DMM” – required by this software. Set the
MET/TRACK ID for this device.

 Fluke PM 6690– Choose from MET/CAL available list for “Fluke PM 6690”! The alias will be set
automatically to 6690. Set the MET/TRACK ID for this device. If using Tektronix counter – still needs to be
configured as Fluke PM 6690.
 Fluke 5730A or 5720A or Fluke 5700A - choose from the list. Alias will automatically populate to 5720 or
5700. If using 5720A or 5730A add (Alias 2 = 5700). Select the checkmark for 5725A and add the
MET/TRACK ID for the Fluke 5725A. If using Fluke 5730A select emulation mode  5700A.

The following Manual standards have to be configured:


 Fluke 792A. Choose from list – Fluke 792A. Set the MET/TRACK ID for this device.
 Fluke 752A. Choose from list – Fluke 752A. Set the MET/TRACK ID for this device.
 Fluke 732A. Choose from list – Fluke 732A. Use the Fluke 732A configuration even if you have Fluke 732B.
Set the MET/TRACK ID for this device.

The following files (provided by the software developer) are required to be in place to allow the system to run:
 SL021_corrections_log.csv (to be placed in metcal.ini [MathFileDir] assigned location. This file logs
historical SL021 characterizations in a column based format for each DUT. This is used for troubleshooting.
 5790x_DCV_I2_log. csv (to be placed in metcal.ini [MathFileDir] assigned location. This file logs
historical Input 2 DCV measurements in a column based format for each DUT. For Eng. Stats. This file can be
bypassed by removing the data logging code in the Fluke 5790B I2 DCV ZCAL_ACAL VER 57X0A.mc
 SL021_Loading_divider.ini (to be placed in metcal.ini [RifWifDir] assigned location. This file contains
the loading corrections for Absolute Divider calibration. The corrections are derived during development and

Fluke Calibration Company Confidential Page 7


Fluke 5790A/5790B Calibration Station Manual
do not change when same design of the divider is used.
 5790x_Input2_Loading_Corrections.ini (to be placed in metcal.ini [RifWifDir] assigned location. This file
contains corrections whenever custom adapter for Input influences the results). The corrections are to be
experimentally obtained – corrections are the inverse of the errors of using custom adapter vs. short adapter at
Input 2.

Operation
To launch and operate the automated software, use the MET/CAL Run Time application to choose the MET/CAL
procedure. MET/CAL procedure files included with this system (hidden programming files are not included in this list).
Full procedure map is available as an appendix in this document.

(MET/CAL executable) Procedure Name Functionality


Fluke 57X0A DCV Characterization.pxe DCV characterization of the source. Required
5790A ACAL System 5700A Characterization
every 30 days. This procedure creates/updates
corrections in the 5700.cor file.
Characterization is saved in the database. The
pxe contains control chart tool tags for used by
sites with control chart tool.
Fluke 5790X Mainframe VER.pxe Calls other procedure from the sections below
Fluke 5790x: (90 Day) VER 792A/57X0A/SL021
to perform a complete Mainframe calibration
Fluke 5790x: (1 Year) VER 792A/57X0A/SL021 without need to combine the records. The
divider characterization is also executed within
Fluke 5790x/AF: (90 Day) VER 792A/57X0A/SL021 this combine procedure.
Fluke 5790x/AF: (1 Year) VER 792A/57X0A/SL021
Fluke SL021 Divider Characterization.pxe Characterization of the divider. Run for every
Fluke SL021 Characterization
DUT. This procedure creates
SNxxxxxxx_SL021_char.ini This pxe
does not collect data - do not save in
MET/TRACK!
Fluke 5790x Absolute with SL021 Divider.pxe ┌ Fluke 5790A 2mV - 20mV: 90 Day Divider (INPUT 1) 1 DUT Absolute calibration at amplitudes:
├ Fluke 5790A 2mV - 20mV: 1 Year Divider (INPUT 1) 1 600µV, 1mV, 2mV, 6mV, 20mV
├ Fluke 5790A/AF 2mV - 20mV: 1 Year Divider (INPUT 1) 1
Results are reported in deviation from nominal
├ Fluke 5790A 2mV - 20mV: 90 Day Divider (INPUT 2) 1
└ Fluke 5790A 2mV - 20mV: 1 Year Divider (INPUT 2) 1 µV/V (this is the error of the DUT). When used
by other processes this has to be used with “-
“sign.
Fluke 5790x AC-DC Calibration.pxe ┌ Fluke 5790A AC/DC: (2 year) ZCAL/ACAL VER 792A2 DUT AC-DC calibration. Results are reported
└ Fluke 5790A/AF AC/DC: (2 year) ZCAL/ACAL VER 792A2 in µV/V (this is a correction type, similar to the
Fluke 792A AC/DC). When used by other
processes this has to be used with “+“sign.
Fluke 5790x Absolute.pxe ┌ Fluke 5790A Absolute: (90 Day) ZCAL/ACAL VER 792A3 DUT Absolute calibration for ranges > 22mV.
├ Fluke 5790A Absolute: (1 year) ZCAL/ACAL VER 792A3 Results are reported in deviation from nominal
└ Fluke 5790A/AF Absolute: (1 year) ZCAL/ACAL VER3
µV/V (this is the error of the DUT). When used
by other processes this has to be used with “-
“sign.
Combine record procedure. This procedure
does not perform measurements. It saves a
record without results. Following saving this
record, combine as follows:
1 Mainframe record
2 Wideband record
3 Combine procedure record

Do not delete the original parts!

The automated procedure has built in logic to evaluate the validity of the Fluke 792A “sn.dat” calibration date against
the calibration date in the MET/TRACK database. It is critical that the specified Fluke 792A is used as configured and
called in the connection messages. Failure to follow these instructions will result in invalid calibration result and
traceability for the entire calibration. If new Fluke 792A calibration is available the calibration record in the database
and the corresponding “.dat” file obtained from Fluke EPEL has to be updated per the instructions in section “Software
maintenance”. If using Agilent 3458A reference accuracy/stability reported by the automated software and as
documented in the uncertainty analysis is valid upon satisfying the following conditions: ACAL performed within 24
Hours or anytime the temperature of the lab changes by > 1° C

Fluke Calibration Company Confidential Page 8


Fluke 5790A/5790B Calibration Station Manual

Note (only applies if calibration is run in sections (instead of running the Fluke 5790X Mainframe VER.pxe):
1. Procedures marked with ( 1 ) require “Fluke SL021 Characterization” procedure within 6 Hrs.
2. Procedures marked with ( 2 ) require “5790A ACAL System 5700A Characterization” characterization to be
completed within 30 days. This procedure also validates the 5700.cor/source match and monitors/enforces/logs
maintenance activities (weekly).
3. Procedures marked with ( 3 ) require “Fluke 5790A AC/DC: (2 year) ZCAL/ACAL VER 792A” procedure within
2 days.
The verification procedure follows the procedure listed in Fluke 5790A Service Manual (see references), where full
Input 1 verification is performed for all sections and abbreviated AC/DC and Absolute (> 220mV) is performed for
input 2. Exception to this rule is AF units – verified only for Input 1 and with special contracted AC/DC
specifications. This procedure allows selection of full calibration for Input 2 on all sections (including the divider
section). The full verification of Input 2 is optional and the added capability allows to accommodate special
customer requests. Lab owned devices are verified for full Input 1 and abbreviated Input 2, unless Input 2 full
verification is required by its use/Service Metrology. Additional option for downscale linearity on all ranges was
added for 5790B release. The full downscale linearity calibration is not performed for customer units. A short
downscale linearity set is built in the normal calibration run (Don’t have to select “Downscale Linearity”) in the
software. Additional option was added to support high frequency linearity in manufacturing – the 800kHz Check
Standard points are used whenever Check Standard is calibrated on the station.

System Maintenance
 Software maintenance:
This automated procedure uses corrections to derive traceable measurements for:
1. AC/DC Transfer Standard voltage measurements. The corrections reside in the “sn.dat” file located
in the “792_datadir” directory as defined in the metcal.ini. The station requires up to date file to
ensure traceability. The corrections are passed to/requested from Fluke EPEL after each calibration.
The file has to be relocated in the specified above directory each time the Fluke 792A has been
calibrated. No changes to the EPEL provide file are to be made! The automated software performs
time validation of the “.dat” file and the Fluke 792A status in the MET/TRACK database. Whenever
out of cycle changes are required – the file has to be manually updated. Calibration date in the file
when an extension is issued is an anticipated out of cycle update.
2. The automated procedure uses corrections to output corrected DCV signals from the Fluke 57X0A
calibrator. The corrections reside in the 5700.cor file located in the “correctdir” location as defined in
the metcal.ini. The correction file validity is verified with each DUT run for age (< 30 days) and for
file/source serial number match. The serial number match is performed to eliminate the opportunity of
using a wrong file/source combination. No user maintenance of this file. Operator will be prompted
when new characterization is required. The process uses a modified correction file that contains the
following section (inserted immediately below the # Generated line) to allow insertion of serial
number during characterization:

Mode: SERIAL

Range: 0.e-3 100.99999e+10 NA NA

3. Loading corrections for SL021 Divider (SL021_Loading_divider.ini). These corrections were derived
during the design of the system and are typical corrections for all SL021 dividers build under the
same BOM and dimension. The correction file is ini based and is to be placed in the metcal.ini
defined “RifWifDir” network location.
4. Loading corrections for Input 2 when a custom adapter with total length > 1.5” is used. The
corrections were derived through testing of typical device at multiple amplitudes and frequencies.
These corrections are lab specific and are a function of the adapter design. The corrections are saved
in the following file: 5790x_Input2_Loading_Corrections.ini
The correction file is ini based and is to be placed in the metcal.ini defined “RifWifDir” network

Fluke Calibration Company Confidential Page 9


Fluke 5790A/5790B Calibration Station Manual
location.
5. Station 792A run mode. When first launched the software will prompt for Fluke 792A run mode. The
software preserves the station run mode in the following ini file: Stat_5790x_Mode.ini
The file contains the following section/parameter:
[MODE]
PC-EVT-13B2=Auto (The parameter stands for the station name from metcal.ini and
its value defines whether the station is: “Manual” for station without knob turner, or “Auto” for
station with knob turner

 Accessories maintenance:
All cables and connections specified above in Station Accessories are cleaned weekly and evaluated for
wear by the operator. Maintenance tasks and log are executed within the automated software supplied with
the station. It is the responsibility of the operator to notify management (site specific organization
structure) when maintenance is required. It is the lab responsibility to maintain surplus of accessories to
accommodate replacement of worn out accessories. See Appendix A below shows examples of
acceptable/unacceptable states of cables and connectors.

System Components Calibration


The traceability of the system and its uncertainty is based on the following calibration conditions.

 Fluke 8508A Reference Multimeter. Calibrated by Fluke authorized service center.

 Agilent 3458A (optional. Instead of Fluke 8508A) – Used with its published specifications for 1 Year with
ACAL. This uncertainty assumes that the device is calibrated following the Agilent 3458A calibration
verification procedure and that calibration laboratory performing the calibration has low TUR in
relationship to the published specifications of the meter. It is recommended that the meter is calibrated at
following Fluke authorized service centers: Everett, WA (USA), Norwich, UK.

 Fluke 792A AC/DC Transfer Standard. Calibrated by Fluke EPEL in Everett, WA

 Fluke PM6690 – calibrated following the calibration procedure in its manual. ISO 17025 calibration is
required. The calibration laboratory performing the calibration must provide Time Base uncertainty at
10MHz sufficient for TUR of 4:1 or better (function of the installed oscillator option). It is recommended
that the Fluke PM6666 is calibrated at an authorized accredited Fluke service center.

 Fluke 57X0A/5725A calibration. Calibration should be performed at an authorized accredited Fluke


service center. In addition to calibration Fluke 57XA is used after it has been characterized using Fluke
752A, Fluke 732A/732B, Fluke 845A/AB (this is an automated MET/CAL procedure).

 Fluke 732A. Calibration should be performed at an authorized accredited Fluke service center. The current
state of the procedure at time of this release, the procedure requires that cc_ui.exe (control chart) is
installed in the lab and linear regression data through limit files is passing the reference cell value to the
procedure.

 Fluke 752A. Calibration should be performed at an authorized accredited Fluke service center. In addition
to calibration a self-calibration of the 752A should be performed within 8 hrs. of use or whenever the
752A has been subject to temperature change > 1°C since last self-calibration. Refer to Fluke 752A
Instruction Manual for self-calibration procedure.

 Fluke 845A/AB – Calibration should be performed at an authorized accredited Fluke service center.

Fluke Calibration Company Confidential Page 10


Fluke 5790A/5790B Calibration Station Manual
Appendix A (Maintenance examles)

Fluke Calibration Company Confidential Page 11


Fluke 5790A/5790B Calibration Station Manual
Appendix B (MET/CAL procedure map)
Procedure Name Read/Write Procedure Name Write Read Procedure Name Read

file to file from file from file


Fluke 5790B I2 DCV: (90 Day) ZCAL/ACAL VER/57X0A
Fluke 5790B I2 DCV: (1 Year) ZCAL/ACAL VER/57X0A
Call 1 15 SUB Get R Value cc_ui Sub_5790A_specs_unc_lookup 12
(5790B only). Fluke 5790B/AF I2 DCV: (90 Day) ZCAL/ACAL VER/57X0A
Optional for Fluke 5790B/AF I2 DCV: (1 Year) ZCAL/ACAL VER/57X0A
"A" model.

5790A ACAL System 5700A Characterization Sub DMM/Null Detector for 5700A Char Meas Loop 1 Sub Compute 792A & 5790A DC Average

Load Interpolated values for 5790A into 5700.cor 1 1 Sub Wait for 792A to Settle

Sub Fluke 5790A AC-DC Difference (2 year) VER 4, 5 3


Fluke 5790A AC/DC: (2 year) ZCAL/ACAL VER 792A Sub Compute 5790A 220mV-1000V AC-DC Difference
Call 2 13
Fluke 5790A/AF AC/DC: (2 year) ZCAL/ACAL VER 792A Sub Compute 5790A 70-700mV Absolute AC Error
Maintenance procedure 5790A Calibration Station 2 2

Fluke 5790A Absolute: (90 Day) ZCAL/ACAL VER 792A


Call 3 Fluke 5790A Absolute: (1 Year) ZCAL/ACAL VER 792A 13 Sub Fluke 5790A Absolute AC Error VER 6,7,8 3,4,5 Sub Average 10 5790 Readings
Fluke 5790A/AF Absolute: (1 Year) ZCAL/ACAL VER 792A
Fluke 5790x/AF: (90 Day) VER 792A/57X0A/SL021
Fluke 5790x/AF: (1 Year) VER 792A/57X0A/SL021

Place headers 5790A AC/DC Section


Sub Fluke 5790A SHUNT AC-DC Difference VER 8 Place headers 5790A Divider Section
Fluke 5790x: (90 Day) VER 792A/57X0A/SL021
Fluke 5790x: (1 Year) VER 792A/57X0A/SL021

Place headers 5790A Absolute Section


Place headers 5790B DCV Section

Sub 5790A 2mV Absolute AC 8 9,10 Unc Adder I2 Loading 14

Sub 5790A 2mV Absolute AC (600uV) 8 9,10 Fluke 792A/DMM Measure DC for Divider Section
Sub 5790A 2mV Absolute AC (1mV)
Fluke 5790A 2mV - 20mV DC 90 Day Divider (INPUT 1)
Fluke 5790A 2mV - 20mV: 1 Year Divider (INPUT 1)
Call 4
Fluke 5790A/AF 2mV - 20mV: 1 Year Divider (INPUT 1)
Fluke 5790A 2mV - 20mV DC 90 Day Divider (INPUT 2)
Fluke 5790A 2mV - 20mV: 1 Year Divider (INPUT 2)

SL021 Characterization called if needed. Sub 5790A 6mV Absolute AC 8 9,10 Sub 5790A 792A DC readings
Fluke SL021 Characterization

Sub 5790A 20mV Absolute AC 8 9,10


Sub 792A Output Low
Sub SL021 Divider Characterization 10,11

Files used File path/location Purpose

1 5700.cor [correctdir] metcal.ini Contains DCV corrections for 57X0A

2 5790A_Cal_Station_xxxxxx_log.txt [MathFiledir] metcal.ini Monitors/logs maintenance of station

3 792ASN.dat [792_datadir] metcal.ini Contains 792A calibration AC/DC values (READ ONLY)

4 uutSN_ACDC_I1.ini [workdir]\5790_UUT_Data metcal.ini Saves AC/DC Input 1 for use in Absolute Input 1
5 uutSN_ACDC_I2.ini [workdir]\5790_UUT_Data metcal.ini Saves AC/DC Input 2 for use in Absolute Input 2

6 uutSN_ABS_I1.ini [workdir]\5790_UUT_Data metcal.ini Saves stats for Absolute Input 1 for troubleshooting only.
7 uutSN_ABS_I2.ini [workdir]\5790_UUT_Data metcal.ini Saves stats for Absolute Input 2 for troubleshooting only.

8 5790A_uutAsset_Absolute.ini [RifWifDir] metcal.ini Absolute errors logging for User Code B units only. For use by other processes.
9 SL021_Loading_divider.ini [RifWifDir] metcal.ini Contains loading data for the divider (READ ONLY)

10 SN_xxxxxxxx_SL021_char.ini [workdir]\5790_UUT_Data metcal.ini SL021 divider corrections

11 SL021_corrections_log.csv [MathFiledir] metcal.ini Logging corrections for each characterization (for troubleshooting only).

12 5790A_Main_Propagation.csv Dynamic. Windows assigned temp location Pass uncertainty and spec to procedure.

during pxe launch/unpack. This csv is saved of a linked worksheet in the unc workbook.
13 Stat_5790x_Mode.ini [RifWifDir] metcal.ini 792A Knob Turning. Auto (knob turner), Manual (no knob turner)

14 5790x_Input2_Loading_Corrections.ini [RifWifDir] metcal.ini Loading correction for Input 2 whenever long adapter is used.
15 5790x_DCV_I2_log.csv [MathFiledir] metcal.ini Logging of positive and negative DCV for Input 2. Requested by Eng.

Block legend
Denotes visible procedure

Denotes hidden procedure

Fluke Calibration Company Confidential Page 12

Вам также может понравиться