9617723
Service instructions
10
DT-NDV
TopLine 20
30 DT-NDD
40
Sticker
Contents
Page
Please first read the Installation Instructions, the Operating Instructions and
these Service Instructions carefully.
These instructions will familiarise you with, the installation, the operating methods, its
use, the safety instructions and the maintenance of this cleaning and disinfection
appliance.
MEIKO is constantly working on the further development of all its models. We would
therefore ask you to understand that because of this, we must reserve the right to make
modifications at any time to any items covered by the contract in terms of their shape,
fittings and technical characteristics.
No claims may therefore be based on the details, the images or the descriptions
contained in these operating instructions.
You receive this Service Instruction free of charge. Further copies may be made
available against the payment of a fee.
This symbol warns that there is danger to human life and health.
IMPORTANT!
... … in order to preserve the operational and safety guarantees, whenever required, only
original parts supplied by the manufacturer are used.
…the user will lose the right to any possible claims if the appliance is modified using any
parts other than original parts.
...... a functional test on all safety systems of the cleaning and disinfection machine is
IMPORTANT
carried out during every regular maintenance.
Be careful when closing the wash-chamber door. Do not catch your hand!
We recommend that you always wear rubber gloves when cleaning or maintaining the
appliance (see Bio-materials Regulations - general activities)
You must not make any alterations after the assembly, commissioning and the hand-
over of the cleaning and disinfection machine to the customer/operator (e. g. electrical
connection, supply of water, steam, connection to waste water or position).
Modifications of the appliance, especially technical modifications without the written
IMPORTANT approval of the manufacturer and from non-authorized persons lead to the total loss of
the right to claim under the guarantee and invalidate our product liability.
IMPORTANT Parts carrying electric current as well as moving or rotating parts can cause
• Dangers to the user’s life and limb and
• Material damage
The cleaning and disinfection machine operates with hot water and the temperature in
the wash chamber exceeds 80°C during thermal disinfection (or at a higher temperature if
so determined by the customer). There exists therefore the danger of scalding.
The care utensils, parts of the wash-chamber and the wash-chamber door are still at a
high temperature. Entsprechende Vorsichtsmaßnahmen sind zu treffen.
WARNING!
When electrical appliances are operating, certain components in the appliance will
inevitably carry a dangerous electrical charge.
Before opening the covers of electrical or electronic components in the cleaning and
disinfection appliance or any electric motor, you MUST ensure that the appliance is
isolated from the power supply.
You can do this in any of the following ways:
1. Place the main switch to the appliance which is located in the building in the OFF
position and take adequate precautions to prevent the switch from being turned
on again.
2. Remove the contactor from the distribution box in the building and take adequate
precautions to prevent the power supply from being restored.
3. Open the inspection door on the cleaning and disinfection appliance and turn the
repair switch to "0".
Work on and rectification of malfunctions on the electrical portion of the cleaning and
disinfection appliance must only be carried out by specialists who have been trained and
authorised by MEIKO. The Health and Safety Regulations must be observed.
The machine, switch cabinets and other electrical components must NOT be sprayed
with a hose or a high pressure cleaner.
Repairs and maintenance on the cleaning and disinfection appliance must only be carried
out by specialists who have been trained and authorised by MEIKO.
ACHTUNG! Work and malfunction rectification on the steam installation must only be carried out by
specialists.
Work on the electrical part of the machine may only be undertaken by specialist
personnel.
The wiring diagram is located in the switch cabinet. This wiring diagram is part of the
machine and therefore must not be removed.
IMPORTANT
The manufacturer’s plate with the connected electrical loads is located inside the switch
cabinet. The manufacturer’s plate with the connected electrical loads (voltage, current
type, amperage, power etc.) is located inside the switch cabinet.
General Electrical Regulations must be observed when connecting the machine to the
power supply.
Topline 10/20/40
The machine's connection is located on the base frame as a screw and is designated
accordingly.
Topline 30
The machine's connection is located close to the connection box as a screw and is
designated accordingly.
When connecting the machine, make sure that the voltage and type of current are
correct. The mains supply must be correctly protected and equipped with a main switch.
If the neutral conductor (N) is not grounded, a 4 phase main switch must be used. A
repair switch is provided for connecting the appliance to the mains supply.
Tighten all electrical terminals in the switch cabinet; check that electrical plugs/jacks are
firmly in position.
After the cleaning and disinfection appliance has been installed in the
customer’s premises and its functions tested, the electrical
connections must be checked before commissioning and hand-over
to the customer by authorised specialists in accordance with the
appropriate specifications. You must comply with national and local
regulations.
1. Because of the voltage drop, the cable to the heating system contactor which is
to be used for the switching operation must be limited to a length of 100 metres
in the case of 1.5 mm2 cross-section.
3 Commissioning
Before delivery we have taken certain precautions in our factory to avoid damage to the
appliance due to improper commissioning.
After making all connections, you must send for a Medical Products Advisor from the
local MEIKO agency for the hand-over and commissioning of the appliance and to
IMPORTANT provide training in the use of the machine. Because of the guarantee, commissioning of
the appliance may only be carried out by the Medical Products Advisor based in the
MEIKO agency responsible for your area.
The advisor will check the whole appliance and will manage the commissioning in line
with the requirements of the Medical Devices Act.
The medical products advisor / customer service installation technician will introduce
you to the appliance and instruct the operator’s authorised representative and staff on
the machine. After these processes are complete, the installation will be handed over to
the authorized person who will sign for it. The appliance may only be used after the
instruction has been completed.
fully open and all condensate traps removed. All dirt collectors must be cleaned
afterwards.
Connection to the electricity supply
• Tighten all electrical terminals in the switch cabinet; check that electrical
plugs/jacks are firmly in position.
• Carry out a visual check on all electrical equipment (e.g. switches, cables,
housings, covers).
⇒ Insert the suction pipes for DEK (dosing unit for softening/rinse agent e.g. Doyen
SK22E) and DR (dosing unit for detergent, e.g. Doyen R 100) into the appropriate
containers.
⇒ Bleed the rinse agent and optional detergent pipes: see the chapter in the
Operating Instructions entitled “Bleed rinse agent” and “Bleed detergent”.
When the appliance is in “manual mode,” all safety functions are disabled.
Therefore, the service technician has a special responsibility of care to avoid any harm
to himself and to protect the machine from any damage.
The quantities of rinse agent and water softener are set using the configuration
parameters P201 and P202.
7 Operating information
7.1 Warnings
The cleaning and disinfection appliance displays warnings as well as errors.
In the case of warnings, the operator must immediately undertake the actions indicated
in the table.
Warning which are ignored lead to an unnecessary operational malfunction and
breakdown of the appliance. As soon as the indicated measure has been implemented,
the warning in the clear text display goes out.
The following warnings are displayed to avoid operating problems:
Works parameters active Works parameters were Switch the repair switch
reloaded off and on.
Emergency strategy is
EEPROM emergency strategy running. Only a limited
selection of programs is
possible without the
EEPROM.
Large amount of steam Fan not running (defective or wire Find cause and rectify
billows from wash- broken), non-return flap does not
chamber when wash- open.
chamber door opened
Wash-chamber door Switch S1 not recognised causing Important: the utensil was
closes almost overload on motor M2 when it reaches not cleaned and disinfected.
completely after the closed position. Result: the “PROGRAM END” does not
activation but controls open the wash-chamber door appear in the display.
immediately re-opens
again.S1 defective
S1 wrongly adjusted (switch actuator
bent)
Wire on S1 broken.
Mechanical obstruction just before
closed position reached (foreign body
in the door gap, door seal hard.
Wash-chamber door distorted.
Motor running on block (caused by
loose motor mounting)
8 User levels
The operation of the cleaning and disinfection appliance involves not only starting the
disinfection programs but other functions, such as accessing operational data, error lists
etc. Process parameters are also viewed / amended.
The user levels should be arranged as follows:
• Amend configuration
Several sub-menus:
- View / amend list of configuration parameters
- View inputs
- View / control outputs (see “Manual Operation“)
8.1.2 Display of inputs and outputs from the I/O circuit board
If the option “display input / output status“ has been activated, the LCD display appears
as follows (the process display, date, time etc is suppressed; only the information from
an input / output component can be displayed. The selection is open, see service
parameter liste P123):
I: 1234567890123456
Fill cold water
45°C 38°C 70mm
O: 123456789012
Line 1 shows the status of the following inputs and the status of the tank level detection.
Line 3 shows the temperatures of the sensors which are connected to the relevant input
/ output circuit board; it also shows the current tank level:
From the left: Temperature in the NDV
Temperature in the chamber
Tank level
The outputs are arranged as follows in line 4:
From the left: 1: M1 – Wash-pump
2: M3 – Rinse agent dosing pump
3: M4 – Detergent dosing pump
4: Door lock LOCKED, or, in the case of AT models: close door
5: Door lock OPEN, or, in the case of AT models: open door
6: M11 – Pumped rim flushing
7: K1 –NDV heating elements
8: Y1 – Cold water filling valve
9: Y2 – Hot water filling valve
0: Operating display / potential-free-connection contact
1: M6 – Chamber fan
2: Y13 – Release extending rotary jet
2: A0 value process
223 Activate demo mode Simulation on the display;
For exhibitions etc.
224 IR sensor reaction Reaction time of AT sensor
0: no reaction
1 .. 10: delay stages
225 Door close mode Different variants after activating foot pedal:
0 – only close (M2)
1 – close and start short program
2 – close and start normal program
3 – close and start intensive program
226 Door open mode Different variants when program end reached:
0 – do nothing (wait for foot pedal to be
activated)
Press the arrow key and keep it pressed until "Parameterize / please enter code"
appears on the display. The value flashes and can be changed within the
corresponding value range with or .
The next field is activated with the arrow key.
Enter the code and confirm with the key.
"Configuration level" appears on the display. Confirm the level by pressing the key.
"Change parameter list" appears on the display. Press the or arrow keys until
"Change outputs" appears on the display.
Confirm with the key. You can select the corresponding output according to the list
below with the and arrow keys.
Caution You are now activating the selected output. The safety functions are not
activated in this state. Confirm the selected output with the key. The value "off"
flashes and can be set to "on" with or if the changed value is accepted with .
Carry out the function test and then deactivate the changed outputs again by
reactivating the value "on" with the key. Change the value to "off" with the arrow
keys and confirm with the key.
To leave the menu press the arrow key until "Ready" or "Ready after error" appears
on the display.
All relevant A0 values and all A0 values required by Health Authorities can, of course, be
set.
PARAMETERISE
Configuration level
Confirm by pressing the key. Then select the ”Main program“ sub-menu with the
and keys, and confirm with the key.
PARAMETERISE
Main program no.: 1
Use the and keys to select the required main program number and confirm with
the key. Use the and keys to select sub-program 19 “Disinfection”.
PARAMETERISE
Main program no: 7
19 Disinfection
Required A0 value:
The A0 value is activated with the key. The A0 value can be changed by pressing the
and keys. Press the key to confirm and save the A0 value which you have just
set.
See the chapter entitled ”Amend service parameters“ to change the value in parameter
P109.
If you wish to save these combinations, enter the values (4+8+16 =) 28 as parameter set P109“.
11 Corrosion signs
The cleaning and disinfection machine is made from high-quality stainless steel.
Nevertheless, under special circumstances, corrosion may occur. The following points
must be observed:
• Ensure that no steel particles enter the appliance via the local fresh water supply
pipes. The same applies to the entry of other metal particles such as copper
filings or turnings.
• Corrosion may occur if unsuitable detergents are used.
12 Temporary decommissioning
Switch off the main power supply from the building.
Remove the drain cover on the ascending pipe and empty the wash-pump. Replace the
drain cover on the wash-pump.
Unscrew the drain plug on the low pressure steam generator and empty the steam
generator. Replace the drain plug in the low temperature steam generator.
Flush out the odour trap once per week with a bucket of clean water.
Bleed the dosing system: see the chapter ”Bleeding the rinse agent and detergent
hoses“.
Start a program and observe that the program runs correctly and that there are no
leaks, especially on the drain screw and the drain plug.
Use the remains of the dosing agents elsewhere or dispose of them properly.
The cleaning and disinfection appliance must be completely disconnected from the
power supply by placing the main switch supplying power from the building in the OFF
position or removing the main fuse in the building.
To avoid any accidents, shutdown cleaning and disinfection machine must be made
unusable immediately, e.g. by cutting the cable loom.
When you dispose of the installation (dismantling/scrapping), the parts and their
corresponding materials should preferably be re-used.
Here is a list of the materials that most frequently occur when dismantling:
• Stainless steel
• Brass
• Electrical and electronic parts
• PP and other plastics
14 Maintenance
“MAINTENANCE NECESSARY” appears in the display when the appliance has
completed a pre-set number of cleaning cycles. The standard setting is 5000 cycles:
Maintenance is necessary after 5000 cycles or at intervals of not longer than one year.
The operator has also the possibility of concluding a maintenance contract with a
MEIKO representative.
Thus, regular and duly maintenance is guaranteed at all times.
1 Faultless functioning
1 2 1 2
Controlling functions and facilities
Create image, readout history as necessary Air trapping: Check condition. ( TopLine )
Update parameter set. Image via digital service report
(DSB) to MEIKO OG.
Water tank: Check the overflow, level monitoring and built-in
Check program key (foil) to ensure proper functioning. dosing hoses and nipples. (TOPIC.2 every 24 months).
LED error display: The LED must light up for "errors". Check DEK pump
Tank level monitoring: Monitoring via M-Commander Replace the DEK dosing hose.
or visual inspection; the container must be empty
when the program ends.
After changing the dosing hose twice
DEK advance notice: Pull out the suction lance. replace the cassette.
Temperature probe: Check the position and fixing. Check the dosing screw fitting.
Door closing system: S1 comes and the door is locked. Chamber ……………… sec.
STB probe in the upper immersion tube. Hose lines / distribubutors / screw fiittings: Check condition
and fixing.
Replace perlators: (Depending on the water quality, Basin / urine bottles holders
the perlators must be replaced at brief intervals). Air
gap: Water outlet when filling. Remove foriegn matter
in the filling tubes. Put the basins in the holders and check every basin to
ensure proper
positioning, adjust the holder as necessary.
For NDD steam MV: Check during program
sequence.
Appearance
For NDD: Clean dirt catching filter.
Wash chamber / door: Removal of coatings,
Washing nozzles - completeness and function dirt and scaling residue.
Visual inspection whether all nozzles are there. The Electrical safety inspection
positioning and nozzle outlets need to be checked.
Check the spray pattern with the plexiglass panel;
manually unlock and remove lifting / rotating nozzle: carried out and consisting of:
Main rotating nozzle / lifting / rotating nozzle TopLine Protective conductor test
Options
1 Faultless functioning
2 Faulty : Arrange appointment for rectification
1 2
DR dosing unit
Check DR pump.
After changing the dosing hose for the second time, replace the cassette.
DR medium: Type:________________________________
Manufacturer:_____________________________________
AT
PRS
Miscellaneous
Note:
Date: Inspector:
16 Repair
Y2 Rinse agent
WW
Y1
KW
M3
Detergent
optional
M4
Heater
E2
optional
Heater
M1 E1
M6
Y13
Y2
WW Rinse agent
Y1
KW
M3
Detergent
optional
M4
Heater
E2
optional
M1
Y3
Steam max.
0,3 bar
M6
Y13