Академический Документы
Профессиональный Документы
Культура Документы
Introduction to
Pressure
Calibration
Three part pressure webinar
• 1st Feb 21: Introduction to pressure calibration.
1 Pa, 100 kPa, 350 kPa, 20, 30 MPa 100 MPa, 140 MPa, 350 MPa, 1000 MPa,
7.5 mTorr 14.5 psi 50 psi 3000, 4300 psi 15000 psi 20000 psi 50000 psi 150000 psi
©2018 Fluke Calibration
0 gauge, start of hydraulic pressure
Introduction to Fluid Pressure
• Types of pressure measuring devices
• Secondary – with approximate best specification
• Absolute, thermodynamic.
Gauge (+ or -)
Pressure Tube
Ref
• Because we are in a gravitational field, pressure is different at different heights even in the
same pressure vessel.
• If at all possible keep the reference and DUT at the same level. If this is not possible a
pressure head correction is made.
• Pressure gets lower further away from the source of gravity (higher).
• Quick estimate: Gas is about 0.00012% per cm, water (most oils) around 100 Pa/cm
(0.033 psi/inch).
Ref
©2018 Fluke Calibration
Pressure Units
• pascal (Pa) named after Blaise Pascal 1623-1662.
– SI unit of pressure
– newton per square meter (N/m2)
– N is kg m/s2 , Pa is kg/m s2
– Small force on large area equals very small pressure.
– Atmospheric pressure is around 100,000 Pa.
– Common are mPa, kPa and MPa.
• Unfortunately many confusing non-SI units are used worldwide. See NIST
Special Publication 811 for conversions.
– Pounds per square inch (psi), inches of mercury, inches of water and bar (ex)
– Sometimes you will see an ‘a’ or ‘g’ at the end to designate absolute or gauge
mode.
Filter
Reservoir
Supply
Valve Priming pump Variable Volume Vent Valve
(if necessary) Fine control Feed to priming pump – Probably should have
©2018 Fluke Calibration
a filter or come from a clean reservoir.
Equipment and Hardware
• Dos and Don'ts of Hardware and general safety
– Take your time, never be in a hurry.
– Always know the pressure rating.
– Replenish old fittings. Do not use fittings where the threads or seals have
worn. be wary of aluminum or brass threaded fittings for high pressure.
– 5S your hardware and label them for their type of media (gas or oil) and
pressure rating.
– Stay away from large volumes for high pressure.
– Try to stay away from using gas as a media if possible for higher pressures.
Ranges above 1000 psi (7MPa) I always try to use oil or water if compatible.
0.5
Pressure 0
0 20 40 60 80 100
-0.5
-1
-1.5
Time
2nd run for repeatability (if required)
Exercise, Preloading
not recorded
Zero adjust if available
Each step is a dwell (stabilization)
and a point taken. Many times averaged.
©2018 Fluke Calibration
Procedure and Points
Questions?