Академический Документы
Профессиональный Документы
Культура Документы
INTRODUCTION
The deleterious effects of scattered radiation on image quality in x-ray imaging
have long been realized and led to the invention of the grid by Bucky in 1 9 1 3 (1)
and the development of a practical grid assembly by Potter a few years later (2).
However, although the scatter problem is well known, the magnitude of its effect
is not generally appreciated. Figure 1 shows lumbosacral spine radiographs made
for the physics section of the American College of Radiology Teaching File to
demonstrate this effect. One radiograph was obtained without a grid (Fig la) and
the second with a highly efficient prototype grid (3) (Fig ib) The same x-ray .
tube potential and incident beam quality were employed for both; thus, the only
difference between the two images is due to scatter. The nongrid image provides
virtually no information and is useless for diagnostic purposes. The grid image is
markedly improved, and spondylolisthesis is evident.
Figure 1 demonstrates that when scatter is controlled, image contrast is im-
proved. The degree of improvement depends on the amount of scatter present and
the efficiency with which it is controlled. However, controlling scatter also me-
sults in increases in x-ray tube load and patient dose. The reasons for this are two-
fold. First, the scatter reaching the imaging receptor is significantly less when a
Abbreviations: CIF contrast improvement factor, SDF scatter degradation factor, S/P ratio of scattered radia-
tion to primary radiation
I From the Department of Radiology, University of Alabama Hospitals and Clinics, University of Alabama at Birming.
ham, 619 S 19th St, Birmingham, AL 35233. Received December 6, 1990; revision requested December 18; revision
receivedjanuary 8, 1991; acceptedJanuary 10. Address reprint requests to the author.
RSNA,
#{163} 1991
307
grid is employed. Second, a grid also absorbs approximately 40% of the primary ma-
diation emerging from the patient. The absorption of scatter radiation, as well as
the absorption of primary radiation, makes it necessary to increase the radiation
incident on the patient to achieve the desired image receptor exposure level and
film density. For example, in Figure 1 the patient’s, entrance skin exposure was
increased by a factor of seven when the grid was used compared with when it was
not. Thus, scatter control improves image quality at the penalty of increased pa-
tient dose. The objective of x-ray equipment manufacturers and the medical phys-
ics community is to control scatter efficiently, that is, to maximize improvement
in image contrast and minimize the associated dose penalty. This article reviews
the degree of success that has been achieved in this regard and the potential for fu-
tune improvements.
a. b.
Figure 1. Comparison of lumbosacra! spine radiographs obtained at 1 00 kVp (34
[three phase]) with a 23 X 43-cm radiation field. (a) Nongrid radiograph acquired with
use of 100 mA at 0. 1 2 second, 102-cm source-to-image receptor distance, 0.33-R (0.09.
mC/kg) entrance skin exposure, and 1 .2-mm effective focal size. (b) Scanning grid ra-
diograph obtained with use of 200 mA at 0.8 second, 1 22-cm source-to-image receptor
distance, 2.2-R (0.57-mC/kg) entrance skin exposure, and 1 .2-mm effective focal spot
size. Both exposures were made in the same cassette with use of DuPont (Wilmington,
Del) HiP!us screens combined with Eastman-Kodak (Rochester, NY) X-L film. The radio-
graphs were developed within minutes of each other in the same processor, and the half-
value layers were matched.
where ‘2 the transmitted x-ray intensity associated with the object of interest
and I is the transmitted intensity of its adjacent surroundings. Display contrast is a
function of the display and, for digital images, postprocessing. In conventional
screen-film systems, the display is film, and display contrast is synonymous with
::iiiIIIIIIIIII:IIIIIIIIiii:: Patient
the radiographic
graphic images,
film density,
tion.
I
image (x-ray
=
assuming
x-ray intensity,
and radio-
no scatter)
x
D
posi-
.
X-Ray Image
x -
Image Receptor
t
D
Radiographic Image
x -#{248}
where .P is the difference in primary intensity between the object and it surround
and P is the intensity of the surround. That is, I P, ‘2 P + iSP, and subject
-
contrast is the fractional variation in primary x-ray intensity emerging from the pa-
tient.
Figure 3. Dia-
grams illustrate
x-ray to radio-
graphic image
conversion and
screen-film expo- t
sure response.
D film density, x -0.
I = x-ray intensi- 3
ty,x=position.
where S = scatter, S/P = ratio of scattered radiation to primary radiation and SDF =
(1 + S/P) 1 The latter is known as the scatter degradation factor and is the frac-
tion of primary beam (possible) contrast that is imaged due to the presence of
ii:ii Patient
Figure
image).
ty, x
4.
the radiographic
D =
position.
Diagrams
film
image
illustrate
density,
(primary
I
formation
and scatter
x-ray intensi-
of
t
X-Ray Image
Image Receptor
t
D
Radiographic Image
creases as the tube potential is increased over the diagnostic range. The data pre-
sented in Figure 6 indicate that the scatter emerging from the patient is less when
the radiation field size is reduced (ie, proper collimation is employed) and when
patient thickness is reduced (ie, compression is employed).
Because of the differences in patient thickness, field size, and anatomic attenua-
tion, as well as in x-ray tube potential, the relative intensity of scatter emerging
from patients varies greatly in clinical practice. Typical ranges for common exami-
nations are listed in Table 1 . Also listed are the corresponding SDFs, which mdi-
cate that for many routine examinations, the fraction of possible contrast imaged
is quite small if scatter is not controlled. Radiography of the chest, abdomen, and
spine is associated with extremely high S/Ps. As illustrated in Figure 7, these high
values result from the greater attenuation of the spine and mediastinum, which me-
sults in reduced primary radiation levels compared with the levels in adjacent me-
LI.
0
Cl)
Scatter/Primary nw si#{149}
(cm)
5- 6.
Figures 5, 6. (5) Plot of the fraction of possible or primary-beam (SDF) contrast that is
imaged versus the S/P. (6) As seen in the plot of S/P and field size, the relative intensity
of scatter (ie, S/P) increases with patient thickness and field size (based on 1 05-kVp sin-
gle-phase data from reference 4).
Table 1
Ranges of Patient S/P for Selected
focal spot Examinations and Associated SDFs
primary
radiation
scattered
radiation
U)
C
0
C
)1
0
position
Figure 7. Diagram shows the effect of
variations of primary beam intensity on im-
aged S/P. High S/Ps occur under the spine
because of its greater attenuation and be-
cause of the presence of x rays scattered
from the more intense primary fields in the
adjacent soft-tissue regions.
RADIOPAQUE INTERSPACE
STRIPS MATERIAL
focal spot
collimator
h
primary radiation
D
Figure 8. Conventional grid. b height of
grid, d = thickness of radlopaque strips, D
= thickness of radiolucent interspace mate-
mial.
scatt#{149}rsd
radiation
- grid
imagi
rceptor
Table 2
Typical Grid Design Parameters
Thickness of
Thickness of Radiolucent
Grid Lead Interspace
Interspace Height Strips Material Strip Density
Ratio Material (mm) (zm) (sm) (lines/cm)
If one substitutes Equation (7) in Equation (8) (and notes that the primary beam
contrast P/P is not greatly affected by the grid) , one obtains the equation
The CIF of a grid depends on its scatter and primary transmission (T5 and T) and
also on the amount of scatter that is present. As illustrated in Figure 5 the greater ,
the scatter level, the smaller the fraction of possible contrast that is imaged and
the greater the room for improvement. In addition, the CIF of a grid will also de-
pend on the penetrating quality of the scatter and primary radiation emerging
from the patient.
The dose penalty or Bucky factor of a grid is the factor by which the milliampeme
second product has to be increased (at a given tube potential) when a grid is em-
ployed to achieve the same film density that was achieved without the grid. That
is,
Grid The performance results for three common grids are given in Table 3 In compar- .
Performance ing the 8: 1 and 1 2: 1 aluminum interspace grids, as one would expect, the 12:1
in General grid achieves greater contrast improvement and has a higher dose penalty. The
Radiography 1 2 : 1 fiber interspace grid achieves better contrast improvement with a lower
Table 3
Typical Grid Performance at 80 and 100 kVp (34)
Tube S/P
Gnid/ Potential T T, Without With Bucky
Interspace (kVp) (%) (%) Grid Grid CIF Factor
Table 4
CIF and BuCky Factor of Aluminum-Interspace 8:1 and 12:1 Grids as a
Function of S/P
Table 5
Mammography Grid Performance
Phantom S/P Bucky
Thickness (cm) Without Grid With Grid C!F Factor
Source -Reference 8.
Note-Measured for a 1 6 X 8-cm BR- 1 2 breast phantom at 30 kVp on a Mam-CP
mammography unit (Transwon!d X-ray, Charlotte, NC).
a. b.
Figure 11. Images of Memorial breast phantom (6 cm) obtained without (a) and with
(b) a grid on the same mammography unit at 30 kVp (30) with the same cassette and
with Eastman-Kodak Min-R screen combined with DuPont Microvision film. The films
were developed within minutes of each other in the same processor. The techniques and
associated average glandular doses were 49 mAs and 73 mrad (730 Gy) for a and 117
mAs and 1 74 mrad (1 ,740 tGy) for b.
Figure 13 Lateral sacral spine radiographs obtained at 1 16 kVp (3q5) with a 33-cm
subject thickness and 20 X 30-cm radiation field. (a) Radiograph obtained with a 12:1
grid with 40.6 lines per centimeter at 300 mA and 0.07 second. The source-to-image ne-
ceptor distance was 1 02 cm and the entrance exposure, skin
5 20 mR (0 1 34 mC/kg). .
(b) Radiograph obtained on a scanning multiple slit device at 600 mA and ‘/ second.
The source-to-image receptor distance was 1 22 cm and the entrance skin exposure, 760
mR (0. 1 96 mC/kg). Both exposures employed a nominal 1 .2-mm focal spot with the
same cassette, Eastman-Kodak Lanex screens, and Eastman-Kodak Ortho-G film. The
films were developed within minutes of each other in the same processor. The same gen-
erator was employed for both films, and the half-value layers of the two source assem-
blies used were matched.
Scanning Slit The use of a single pair of moving slits to suppress scatter was suggested at the
Radiography turn of the century (1 1) . Such a device is illustrated in Figure 1 4 and consists of a
beam-defining fore slit between the x-ray tube and the patient and a similar scat-
ten-limiting aft slit between the patient and image receptor. The two slits move
synchronously and scan the area of interest during an exposure. The principle of
scanning slit radiography is that the small radiation field defined by the fore slit
produces little scatter in the primary field and the x rays scattered out of the irma-
Table 6
Comparison of Performance of Conventional and Ideal Mammography Grids
Focal Spot
Coll,mator/ Shutters
Primary
Scattered Radiation
Patient
Table . Eliminating
Cassette I Aft Slit
An alternative way of Improving scatter control is to employ more efficient grids. Novel Grid
Two techniques that have clinical potential have been described in the scientific Techniques
literature. One is a scanning grid that consists of radlopaque slats situated between
the patient and film (Fig 1 5) (3,20). However, unlike a conventional grid, there is
no interspace material, and, during an exposure, the assembly scans across the ra-
diation field and the grid slats articulate so that they remain focused at the x-ray
source. The scanning motion is synchronized with the exposure time so
pivot point
grid slats
receptor
CONCLUSIONS Scatter control plays an important role in diagnostic x-ray imaging. Without scatter
control, image quality is severely compromised. In many instances, conventional
techniques yield acceptable results, especially if attention is paid to detail-colli-
mating to the area of interest and, when feasible, compression. However, in imag-
ing thick body parts, the capability of conventional grids is limited. Novel scan-
ning grids and low-strip-density air-interspace tantalum grids offer considerable
promise, and it is highly likely that these techniques could virtually eliminate
Table 7
Comparison of Conventiona 1 and Novel Scatt em Co ntrol
TeChniques at 100 kVp
Tp Ts Relative
Technique (%) (%) S/P CIF Dose
1. Bucky G. Method and apparatus for pro- 1 3. Barnes GT, Brezovich IA. The design REFERENCES
jecting roentgen images. U.S. patent and performance of a scanning multiple
1,164,987, 1915. slit assembly. Med Phys 1979; 6:197-
2. Potter HE. The Bucky diaphragm princi- 204.
plc applied to roentgenognaphy. AJR 1 4 . Sorenson JA, Nelson JA, Niklason LT, et
1920; 7:292-295. al. Rotating disk device for slit radiogra-
3. Barnes GT, Moreland RF, Yester MV, et al. phy ofthe chest. Radiology 1980;
The scanning grid: a novel and effective 134:227-231.
Bucky movement. Radiology 1980; 15 . Korbuly D, Moore R, Formanek A, et al.
135:765-767. Scanography with rotation of the radio-
4 . Stargardt A, Angerstein W. Uber die graphic tube: a new method. Radiology
streustrahlenverminderung bei den ront- 1980; 135:495-499.
genvergrosserungstecknik. ROFO 1974; 16. Rudin 5, Bednarek DR. Improved con-
123:364-369. trast in special procedures using a rout-
5. Friednich M. Neure entwicklungsten- ing aperture wheel (RAW) device. Radi-
denzen den mammographie technik: dir ology 1980; 137:505-5 10.
raster-mammographie. ROFO 1978; 17. King MA, Barnes GT, Yester MV. A mam-
128:207-222. mographic scanning slit assembly: design
6. Barnes GT, Brezovich IA. The intensity considerations and preliminary results.
of scattered radiation in mammography. In: Logan \W, Muntz EP, eds. Reduced
Presented at the Fourth International dose mammography. New York: Masson,
Conference on Medical Physics, Ottawa, 1979; 243-252.
July25-30, 1976. 1 8. Yester MV, Barnes GT, King MA. Experi-
7, Barnes GT, Brezovich IA. The intensity mental measurements of the scatter re-
of scattered radiation in mammography. duction obtained in mammography with
Radiology 1978; 126:243-247. a scanning multiple slit assembly. Med
8. Wagner AJ. Contrast and grid perfor- Phys 1981; 8:155-162.
mance in mammography. In: Barnes GT, 1 9. Barnes GT, Wu X, Wagner AJ, Rubin E.
Frey GD, eds. Screen-film mammogra- Scatter control in mammography: past,
phy: imaging considerations and medical present, and future. Presented at the
physics responsibilities. Madison, Wis: 75th Scientific Assembly and Annual
Medical Physics (in press). Meeting of the Radiological Society of
9. Synder RE, Masterson ME, Miller DW, et North America, Chicago, November 26-
al. Image quality and radiation expo- December 1, 1989.
sure in mammography. In: Logan WW, 20. Barnes GT. Scanning grid apparatus for
ed. Breast carcinoma: the radiologist’s suppressing scatter in radiographic imag-
expanded role. New York: Wiley, 1977; ing. U.S. patent 4,340,818, 1982.
167- 184. 2 1 . Sorenson JA, Niklason LT. Performance
10. Barnes GT, Brezovich IA, Witten DM. characteristics of improved antiscatter
Scanning multiple slit assembly: a practi- grids. Med Phys 1980; 7:525-528.
cal and efficient device to reduce scatter. 22. SorensonJA, Niklason LT, Jacobsen SC, et
AJR 1977; 129:497-501. al . Tantalum-air interspace crossed
1 1. Pasche 0. Uber eine neue blendenvor- grid: design performance characteristics.
nichtung den rontgentechnik. Dtsch Med Radiology 1982; 145:485-492.
Wochenschr 1903; 29:266-267. 23. Barnes GT, Sones RA, Tesic MM. Digital
12. Tesic MM, Mattson BA, Barnes GT, et al. chest radiography: performance evalua-
Digital radiography of the chest: design tion ofa prototype unit. Radiology 1985;
features and considerations for a prom- 154:801-806.
type unit. Radiology 1983; 148:259-
264.