JG-INS-ANA-020
L0

Questions
The ABB distillation analysers correlate to which ASTM standard?
What is the length of the heater required to be adjusted in the boiler?

What is the distance between the tip of the thermocouple and the final coil of the heater in the distillation analyser.?
What is the flowrate required to be set for sample inlet pump in the distillation analyser?
What is the flowrate required to be set for sample outlet pump in the distillation analyser?
The Naphtha distillation analyser uses anticoking agent-
What is the PM frequency for diesel distillation analyser?
What is the wattage of the high temperature heater version in the distillation analyser?
What thermocouple is used in the distillation analyser?
Before doing maintenance on the distillation analyser the purge control mode should be switched to Maintenance mode
What is the ratio of anticoking agent to sample used in distillation analyser?
The _____________ heater version is used in Naphtha distillation analyser-
What is the limit of the sample temperature at the analyser input?
Depending on the level in the tranquiliser well, the pre-amp card generates an output voltage of ?
The LIC shows 0-100% indication for an input voltage of?
The air purge exit module lid opening is generally set at 20 mbars- True or False.
What is the maximum stroke frequency of the pneumatic anticoking pump ?
The hydraulic oil required to be filled in the sample pumps is Pulsa lube 8 G – True or False.
The pressure on the pressurization guage in the purge control unit is set at 2.5 bars -True or False.

The pressure on the purge guage in the purge control unit is set at
When will the RED indicator be ON in the purge control unit ?
When will the GREEN indicator be ON?
When will be the YELLOW indicator be ON ?
The sample pressure at the inlet of the analyser is normally set at ________ bars.
The diaphragm is located in the ______________ assembly of the sample pumps.
The ________ valve is to be pressed to remove gases from the hydraulic system of the pump.
The oil used in the gear box section of the pump is Pulsa lube 8 G.
The ABB distillation analyser can be used to measure distillation in the range of 0 to 100 %.
The heater can be switched ON even if the pumps are OFF
What is the PM frequency for Naphtha distillation analyser?
What is the PM frequency for LK distillation analyser?
What is the PM frequency for HK distillation analyser?

What is the PM frequency for HAGO distillation analyser?

The output parameter of a distillation analyser is


Which one of the following has the lowest 95% distillation point:- Diesel, HK, LK, Naphtha
The level sensor in the distillation analyser is a ______ type of a pressure transducer.
The pressure transducer should be flushed with high pressure inst. air during PM.

Boiler, tranquilliser well, level sensor, Pre amplifier PCB

The ABB pourpoint analyser correlates to which ASTM standard?

In the cooling cycle ,the sample in the sample cup is cooled with the help of _____.
The pendulum with mirror is located in the center of the cup
During the cooling cycle, the charge valve and drain valve are in _____ condition
The supply voltage to the sample cup heaters is______.
The operating sample pressure at the analyser inlet should be _____.
The operating sample flowrate to the analyser should be_____
The operating cooling water flowrate to the analyser should be____
The temperature of the sample in the sample cup is measured by ____
How many temperature probes are installed in the cell box assembly?
The pour point is detected by_____
The position of the pendulum in the sample cup is_____
The photo diode produces output voltage in the range of______
The light source in the pourpoint analyser is a______
The sample cup rotates at the speed of _____
The burette is heated.
The light source supply voltage is____ volts
In the "Flush" sequence, the product valve and the drain valve are____
The motor rotating the cup will be ON in the ______ cycle
In the ____ & ____ cycle, the heater will remain ON
The Unplug sequence starts when the instrument is powered on.
In the flushing sequence the sample is flushed through the sample cup.
The 4 LEDs located on the backplane board in the control box assembly indicate the status of_______.
When the 1st LED ( located in the control box ) from the top is glowing, it indicates that ______ power supply is healthy.
When the 2nd LED ( located in the control box ) from the top is glowing, it indicates that ______ power supply is healthy
What are the two operating modes of the analyser?
Which card is located adjacent to the CPU card?
If the difference between the ramp temperature and the effective cell temperature is higher than _____ deg C, a cooling
Which error is emitted when a pourpoint is detected with a value of more than 3 deg C above previous value.
When the cool-time out error is generated, cooling halts and the analyser switches to ______ state instead of Ready sta
The cooling rate may be set between _______ deg C / min in one minute increments.

The normal cooling rate set is _____ deg C / min


The pourpoint range minimum value should be set between -32 and +32 deg C by whole values of 1 deg C and be atlea

The flush temperature minimum should be set at ____ deg C

The pourpoint analyser can be calibrated by adjusting the ______


The relay board is located in the ____

If the difference between the ramp temperature and the effective cell temperature is higher than 10 deg C, a cooling err
Option 1 Option 2 Option 3 Option 4
ASTM D 86 ASTM D 56 ASTM D96 ASTM D 92
156 mm 158 mm 120 mm 160 mm

1-2 mm 6 mm 5 mm None
25 ml / min 25 lph 25 nm3 / hr 25 m3 / hr
1.25 lph 1.25 ml / min 1.8 lph None
True False.
3 weeks 8 weeks 15 days 12 weeks
300 watts 1000 watts 500 watts None
K- type R-type S-type J-type
True False.
0.048611111111111 0.090277777777778 0.042361111111111 0.084027777777778
High temperature Cartrige Standard temperature None
> FBP < IBP and also < 65 deg C > IBP None
0 – 2.5 VDC 0 – 5 VDC 0 – 5 VAC 0 – 2.5 VAC
0 – 10 mVDC 0 – 10 VDC 0 – 70 mVDC 0 – 100 mVDC
True False.
45 strokes / min 60 strokes / min 100 strokes / min None
True False.
True False.

5 bars 5 mbars 2.5 mbars 2.5 bars


Pressurzation mode Maintenance mode Purge mode None
Pressurzation mode Maintenance mode Purge mode None
Pressurzation mode Maintenance mode Purge mode None
1 2 3 4
Piston assy section Reagent head assy Gear box assy None
HPV PTP PTS None
True False.
True False.
True False
1 week 2 week 3 week 8 week
2 week 4 week 6 week 8 week
8 week 12 week 6 week 1 week

3 weeks 2 week 1 week 8 week

Temperature Level % distillation None


Diesel LK Naphtha HK
Capacitive Variable reluctance Load cell None
Atmosphere Boiler bottom Boiler top None
True False
Increasing sample lag t Reducing sample lag time Preheating the sample None
Boiler tranquilliser well, level sensor Pre amplifier PCB

ASTM D 86 ASTM D 56 ASTM D96 ASTM D 97


-32 to + 32 deg C -60 to + 60 deg C -10to + 10 deg C None
Freeze point Cloud point Distillation point Pourpoint
25 cst at 50 deg C 40 cst at 50 deg C 250 cst at 50 deg C None
Cooling water Peltier coolers Air None
1 0
Shut Open One is open and other isNone
110 VDC 110 VAC 230 VAC 24 VDC
0.5 to 2 bars 5 to 6 bars 8 to 10 bars 10 to 20 bars
1-5 lpm 1-5 lph 10-15 lph 50-60 lph
1-2 lph 5-8 lph 30-60 lph 1-5 lph
J-type t/c K-type t/c Pt-100 RTD PT-1000 RTD
1 2 3 4
Deflection of pendulum Temperature drop Temperature rise None
Off center In the center None Both
-10 VDC to +10VDC -5 VDC to +5 VDC -24 VDC to + 24 VDC None
LED Bulb Heated filament None
1/16 RPM 1 RPM 100 RPM 200 RPM
1 0
24 12 110 230
Closed Open One is open and other isNone
Flush Fill Settle Cool
Flush & fill Settle & Charge Ready & Stand by Heat & Unplug
1 0
1 0
Power supply CPU Temperature interface c Data accduisition card
-12 V +12 V +5V -5 V
-12 V +12 V +5V -5 V
Automatic & Manual Remote & Local Both None
Acquisition card Temperature sensor interface cSignal-A/O card Graphic display invert
20 10 30 25
Low difference error High difference error High sample error Low sample error
Flush Stand by Fill Unplug
0.1 and 9.9 2 and 4 3 and 6 10 and 15

2 to 3 4 5 6
10 8 6 1

60 -10 +57 None

Threshold voltage Lamp voltage Detector voltage Cooler voltage


Control box assy Power supply box assy Cell box assy SCS

1 0
Answers
Option 1
Option 2

Option 1
Option 1
Option 2
Option 2
Option 1
Option 3
Option 4
Option 1
Option 1
Option 3
Option 2
Option 1
Option 3
Option 1
Option 1
Option 2
Option 1

Option 3
Option 1
Option 3
Option 2
Option 2
Option 2
Option 2
Option 1
Option 2
Option 2
Option 4
Option 4
Option 1

Option 2

Option 1
Option 3
Option 2
Option 1
Option 2
Option 3
Option 4

Option 4
Option 1
Option 4
Option 3
Option 2
Option 2
Option 1
Option 2
Option 1
Option 3
Option 3
Option 3
Option 3
Option 1
Option 1
Option 2
Option 2
Option 1
Option 1
Option 2
Option 2
Option 4
Option 4
Option 1
Option 1
Option 1
Option 2
Option 3
Option 1
Option 3
Option 2
Option 2
Option 1
Option 1

Option 1
Option 2

Option 3

Option 1
Option 2

Option 1
Level L0
Sr no Questions Option 1
The courier 10 sxt sulphur in liquid analyser works on which
1 NDIR
principle
2 What is the full form of XRF? Xray fluorescence
3 What is the material used as reference std by the analyser Berylleum
Which element is a major inrterference in Sulphur
4 Nitrogen
measurement
Which purge gas is used to purge out argon from the sample
5 Nitrogen
cell?
When the sample cell is opened for maintenance, the Xray will
6 1
be automatically shut down
7 Which is the type of detector used in courier 10 sxt? Proportional counter
8 The sample cell window is made of which material? Titanium
9 The thickness of the Be window is______. 1 mm
The _______ is located between the protective window and
10 Proportional counter
the Be window.
What is the maximum sample pressure rating for the sample
11 1 bar
cell?
What command is used to open communication with the
12 A
analyser?
What is the command used to bring the analyser to active
13 1+
mode from passive mode?
The _____ paramenter defines the sample and reference
14 A
measurement sequence.
15 Command _____ puts the analyser in measurement mode. "-
16 To print the spectrum use the command _____. :
During a validation run, the std value was 200 ppm and the
analyser reading was 300 ppm.If your old intercept value was
17 "+300
-200, what will be your new intercept value to coorect for the
difference between the std value and the analyser reading?
During a validation run, the std value was 200 ppm and the
analyser reading was 150 ppm.If your old intercept value was
18 "+300
-150, what will be your new intercept value to coorect for the
difference between the std value and the analyser reading?
19 The sample measurement duration is set to _____. 1 min
The normal sample flowrate adjusted for the analyser is _____
20 10
lph
If the reference LED is glowing on the MHA box, then the
21 Reference
measuring head is in _______ position.
PPU is a pulse processing unit for input signals from ____ &
22 MMU
A/D converters.
For communicating with analyser from a PC, the analyser
23 Line
should be put in ______ mode.
24 Which parameter is a gain control parameter? G
25 Which command is used to output th normalised intensities. +
The local display software of the analyser operates in 5
26 different modes. These are _______, View mode,Result ASCII mode
mode,Gain mode,Alarm mode.
The analog current( 4-20 ma)signal is generated by _____
27 PPU
card.
What is the thickness of the yellow protection window in the
28 1 micron
SCA?
The sample bypass flowrate should be maintained between
29 10 to 100
_____ lph
While doing any maintenance on the SCA, the Maintenance
30
switch on the PES should be place in Off position. 1
Before starting validation of the analyser, the sample cell
31 should be flushed first with IPA and then with the standard
solution. 1
Nitrogen purge through the sample cell should be stopped
32
while carrying out validation . 1
Viscosity

1 The ABB viscosity analyser works on which principle? Faraday's law

As per Hagan Poisuille equation, dynamic viscosity is directly


2 proportional to ____ given the capillary bore,capillary Flashpoint
length,sample flow are constant.

3 Dynamic viscosity= Kinematic viscosity x______. Density


4 What is the unit of dynamic viscosity? Cst
5 Viscosity of a sample is extremely temperature dependent 1
6 The Hagan Poiseuille equation applies to ______ liquids Volatile
The oil bath of the viscosity analyser is maintained at___ deg
7 150
C
8 The mercury thermometer measures temperature at_______. Capillary inlet
9 The viscosity analyser requires variable sample flow 1
10 What is the capacity of the oil bath in the viscosity analyser? 25 ltrs
The MOC for capillary used in the viscosity analyser is
11 SS
______.
What is the sample flowrate delivered by the precision
12 25 lph
metering pump of the viscosity analyser ?
13 What is the application of viscosity analyser in our complex Diesel
The DP transmitter in the viscosity analyser measures
14 Pump
pressure across_________
What is the wattage capacity of the heater used in the viscosity
15 1500W
analyser?
Steam heating is provided in the viscosity analyser so that the
16 Vaporise
sample does not______ in the SCS
17 The oil level in the oil bath is monitored by a______. Level guage
What is the inlet sample flowrate required to be set in the SCS
18 20 lph
of the viscosity analyser ?
A thermal fuse is provided on the heater element to avoid
19 1
overheating of the heater.
20 How much oil is required to be filled in the oil bath? 20 ltrs
The viscometer pump motor should be started only after the
21 1
temp of the oil bath has reached its set point of 100 deg C
The "Alarm" LED will glow incase of low threshold of the
22 Output
_______ signal to SCAM unit
OCMA 25
Solvent extraction /
1 The OCMA-25 works on the principle of _______-.
NDIR absortiometry
2 Oil is soluble in water but insoluble in organic solvent. 1
Hydrocarbons ( CH bonds) absorb infrared radiations in the
3 1-2 microns
region of ______.
In OCMA 25 the measured value is unaffected by color of
4 1
sample.
The reference side of the IR cell in OCMA 25 is filled
5 Air
with______.
How much qty of B-Heavy oil is required to be diluted in the
6 1 ul
sample to make a standard solution of 24 ppm
A standard concentration of 24 ppm of B-heavy oil will
7 4
indicate_____ ppm of oil on the indicator.
" F.S Over" alarm lights when the indication exceeds the full
8 1
scale during the measurement / calibration.
9 What is the replacement frequency of adsorber.? Monthly
The PH adjuster pump stroke dial should be normally set to
10 10
______.
The PH adjuster pump stroke dial should be set to ______.to
11 10
purge out air bubbles from the line.
12 The solvent used in OCMA 25 is ______ S316
OIW Agar
The measuring principle of OIW 101 analyser is based on
1 _______. Resistance
2 The OIW 101 basically measures _____ Oil
3 The application of OIW 101 analyser in ETP is in____ Slop oil to RTF
4 How many OIW 101 analysers are installed in ETP? 1
5 Oil and water have same density. 1
6 The analyser should be installed in the horizontal position. 1
The presence of gas in the emulsion will be interpreted as
7 ____ by the instrument Water
What should be the velocity range of the sample through the
8 analyser 1-2 ft/sec

9 The OIW 101 analyser is suitable of pipe line sizes of 4" to 60". 1
The EPROMs also need to be replaced if the detector is
10 replaced. 1
To note the absolute maximum possible output that the
11 detector can generate, the antenna should be ____. Short cktd to gnd
To note the absolute minimum possible output that the
12 detector can generate, the antenna should be ____. Short cktd to gnd
13 The output of the detector is_____. Current
To check zero % oil indication, flow 100% water through the
14 analyser. 1
To check 100 % oil indication, flow 100% oil through the
15 analyser. 1
16 The detector's unit can be switched with that of a similar unit. 1
Which port should be adjusted during calibration as per
17 method 1 in the manual. R22
As per calibration method 1 in the manual, which terminals are
18 to be used to measure the output current? TP 6 & 11
What will be the instrument's % water reading on the display if
19 the 15 VDC to the detector has failed. 0%
What will be the instrument's % water reading on the display if
20 the antenna is shorted. 0%
What should be the resistance between the antenna jack and
21 the metal enclosure? 1 ohm
If the resistance between the antenna jack and the metal
enclosure is < 20 mega ohm, then it is possible that the
22 antenna is______. Open cktd
If there is inert material like sand , dirt etc in the sample ,then
23 the instrument will read it as _____. Water
The sample components with low dielectric constants will be
24 detected as______ OIl

Any conductive material like iron oxide, sulphides etc present


25 in the sample will be detected as _____ by the instrument. OIl
Water in oil
The measuring principle of Water in oil DC1000 analyser is
1 base on _______. Resistance
The temperature compensation is provided in the system
2 by______. RTD PT100

3 Change in the sample density can affect the analyser reading. 1


4 There are 2 nos of water in oil analysers installed in RTF 1
5 What is the power supply requirement for the analyser? 110 VDC
To select the calibration curve 0 , the terminals 11 and 12
6 should be connected to terminal no______. 2
For what range of water content, can the variation of dielectric
7 constant be considered linear? 0-10 %
To perform simple calibration on the instrument, use command
8 _________. CS
9 To display / modify calibration, use the command ____. CS
An emulsion of water in oil exhibits a temperature dependency
such that an increase in temperature causes a _______ in the
10 mixture's dielectric constant. Increase
TOC-Shimadzu
In TOC 4100, the carbon is converted to CO2 by method of
1 UV
_____.
Hydro carbon, on combustion in presence of air and catalyst
2 H2O
produces CO2 & _____.

3 What is the catalyst filled in the combustion tube? Ti


Total Organic
4 The fullform of TOC is _____
Content
5 The instrument measure TOC by the method_____. TC + IC
6 The detection principle used for CO2 in the analyser is _____. NDIR
The combustion tube in the instrument is maintained at _____
7 650
deg C
The halogen scrubber is provided in the system to
8 O2
remove_____.from the CO2 gas
9 What is the carrier gas flow rate set in analyser? 100 cc/min
10 What is the sparge gas flow rate set in analyser? 100 cc/min
11 What is the carrier gas pressure set in analyser? 100 kpa
What is the sample inlet flowrate set in the sample conditioning
12 1 lpm
system?
What is the sample inlet pressure set in the sample
13 1 barg
conditioning system
The Inorganic carbon in the sample is converted to CO2 by
14 UV
_____.
The catalyst in the combustion tube is to be replaced after
15 500
every _____ injections.
16 The CO2 absorbers remove CO2 from the carrier gas. 1
For the the range of 0-10 ppm of TOC, the baseline stability is
17 High sensitivity
to be checked in the _____ mode on the screen.
The specification of carrier gas with respect to CO2,HC,CO,
18 < 10 ppm
for better baseline stability is_____
Mol sieve outlet of N2
19 The carrier gas to TOC , in our complex, is fed from____.
plant
20 The rotary valve used in the analyser is a ___ port valve 5
The baseline position can be adjusted by rotating the optical
21 Pre amplifier card
adjustment screw on the _____.
What should be the dehumidifier temperaturedisplayed on the
22 < 10 deg C
screen?
The 5 parameters on which the availability of Ready state
23 depends are base line noise,baseline position,Furnace Carrier gas pressure
temperature, Dehumidifier temperature & _______.
If 2.125 gms of KHP is dissolve in 1 lit of distilled water, then
24 1000 ppm C
the TC content of the resulting solution will be _______.
25 The TC content of the sample is calculated by_____. Peak height
Potassium hydrogen pthalate is used to prepare TC standard
26 1
solution.
How much quantity of KHP should be dissolved in 1 lit of water
27 1.0625 gms
to produce a TC standard of 500 ppm C.
28 Zero water means water having ____ ppm of TC,IC or TN 1

29 What is the full form of NPOC. No Peak Of Carbon

Instrumentation &
30 What is the fullform of IC
Control
The water used by the analyser for rinsing the internal tubes is
31 1
tap water.
What is the height of the catalyst required to be filled in the
32 180 cm
combustion tube.
Conductivity detectors are also used in CO2 measurements in
33 1
TOC analysers.
Power supply to the analyser should be switched off
34 1
immediately after switching off the furnace.
The" Furnace Over Heat" error message appears when the
35 120
furnace temperature exceeds ____ deg C.
Option 2 Option 3 Option 4 Answers
Thermal
Chemiluminescence XRF Option 3
conductivity
Xenon ray
Both None Option 1
fluorescence
Yetrium Mollybdenum None Option 3
Oxygen Hydrogen Argon Option 4

Oxygen Hydrogen Helium Option 1

0 Option 1
TCD FID NDIR Option 1
Platinum Berylleum Carbon Option 3
2 mm 0.5 mm 25 um Option 4
X ray tube MHA Leak detector Option 4

0.3 bars 2 bars 5 bars Option 2

B # #A Option 4

#A 2+ None Option 3

B S Z Option 3
"+ ! ? Option 3
; " > Option 1

"-300 "-100 "+100 Option 2

"+100 "-100 None Option 3

2 min 10 min 5 min Option 4


20 30 60 Option 1

Sample Intermediate None Option 1

WIU Detector None Option 3

Local Both None Option 1


F A B Option 1
- > " Option 4
Hexadecimal
Binary mode Octal mode Option 1
mode
SPIU WIU CSU Option 4

75 micron 10 micron 1 mm Option 2

5 to 10 50 to 250 1 to 100 Option 3

Option 2
0

Option 1
0

Option 2
0

Hagan
Beer lambert's
Newton's laws Poiseuille Option 3
law
equation

Diff pressure across


Pourpoint None Option 2
capillary

Atomic
Atomic weight None Option 1
number
Cpo Both None Option 1
0 Option 1
Non volatile Newtonian None Option 3
200 250 100 Option 4
Capillary outlet Both None Option 2
0 Option 2
30 ltrs 18 ltrs 100 ltrs

Fused silica Glass Teflon Option 1

60.7 lph 65 lph 62 lph Option 2


LCO RFO KO Option 3
Filter Capillary None Option 3

2000W 1000W 500W Option 3

Congeal Liquify Solidify Option 2


Pressure
Level transmitter Level switch Option 4
transmitter
25 lph 5 lph 40 lph Option 4

0 Option 1
18 ltrs 15 ltrs 10 ltrs Option 2
0 Option 1
Input O/P & I/P None Option 2

Combustion / NDIR UV Condsuctivity Option 1


0 Option 2
3.4-3.5
2-3 microns 10-15 microns Option 3
microns
0 Option 1

Water Oil Pure solvent Option 4

2 ul 4.8 ul 4 ul Option 3

8 2 6 Option 2

0 Option 1
Twice in a No
Every 4 months Option 2
month replacement.
15 100 20 Option 4

15 120 20 Option 3
SS316 KHP IPA Option 1

Energy absorption Inductance Conductance Option 2


Water Sediments None Option 2
Crude Effluents from None Option1
2 3 4 Option 3
0 Option 2
0 Option 2

Sediments Oil All Option 3

3-4 ft/sec 0 ft/sec 5-7 ft/sec Option 4

0 Option 2

0 Option 1

Open to atmp Both None Option 1

Open to atmp Both None Option 2


Voltage Capacitance Inductance Option 1

0 Option 1

0 Option 1
0 Option 2

R21 R10 R19OC & R22 Option 4

TP 1 & 2 TP 3 & 4 None Option 1

50% 100% 25% Option 1

50% 100% 25% Option 3

2 ohm 3 ohm > 20 mega ohmOption 4

Short cktd Both None Option 2

Oil Both None Option 2

Water Both None Option 1

Water Both None Option 2

Capacitance Inductance Conductance Option 2

RTD PT1000 K type T/C J type T/C Option 1

0 Option 1
0 Option 1
110 VAC 24 VDC 5 VDC Option 3

3 4 1 Option 4

10-20 % 30-50 % None Option 1

MS TS None Option 1
MS BS DC Option 4

Decrease No effect None Option 2

UV/Persulphat
IR Combustion Option 3
e
Cl Br NaCl Option 1
Pt coated on
Ag Gold Option 4
alumina balls
Total Organic Total Oxygen Total Organic
Option 2
Carbon content Combustion.
IC-TC TC-IC TC/IC Option 3
UV IR Spectroscopy Option 1

680 850 950 Option 2

Halogens N2 H2 Option 2
150 cc/min 200 cc/min 50 cc/min Option 2
150 cc/min 200 cc/min 50 cc/min Option 1
150 kpa 200 kpa 250 kpa Option 3
3 lpm 5 lpm 2 lpm Option 4

4 barg 10 barg 20 barg Option 1


Acidification
IR Combustion Option 3
and sparging
1000 5000 10000 Option 4
0 Option 1
Medium sensitivity Low sensitivity None Option 1

> 10 ppm > 50 ppm <1 ppm Option 4

Inst air Plant air N2 header Option 1


2 3 8 Option 4
CPU NDIR detector None Option 3

<35 deg C <4 deg C < 100 deg C Option 3

Carrier gas Baseline


Sparge gas flow Option 4
flow fluctuations

10 ppm C 100 ppm C 200 ppm C Option 1


Area under the peak peak width Peak timing. Option 2
0 Option 1

2.125 gms 1 gm 5 gms Option 1


2 10 0 Option 4
Non
Non Purgeable Purgeable
None Option 3
Oxygen Content. OrganicCarbo
n.

Inorganic Carbon Integrated Ckt None Option 2

0 Option 2

150 mm 130 cm 130 mm Option 4


0 Option 1

0 Option 2

680 650 780 Option 4


Moisture

Sr no Questions Option 1

The type of sensor for moisture measurement used


1 by panametrics moisture analyzer is Aluminium oxide

2 The range of panametrics moisture moisture probe is -110 deg c to +20 deg C

3 The frequency of recalibration of moisture probe is Weekly

4 Moisture probe is used for measurement of Dewpoint

The type of temperature sensor used in M2LRT probe


5 is Thermistor

The Al2O3 sensor measures moisture due to change


6 in Capacitance
The dummy probe is used for testing the functionality
7 of Electronics
For calculating PPMv for gases , the electronics
8 requires the input of Capacitance

9 The moisture probe data sheet contains data of Dewpoint to MH


The M2LRT probe has temperature sensor inbuilt in
10 the probe 1
The decrease and increase in the moisture analyzer
reading with increase and decrease of temperature
11 over the day is called Diurnal swing
While checking the analyser with a dummy probe, the
_______ value is adjusted to match the analyser
12 display MH and the MH value on the dummy probe. MH High

Which is the front end process element for clauss


1 process? Reaction furnace
2 How much H2S is burnt in Clauss process ? 2/3
3 What is air demand? Oxygen demand
4 What is the ratio of H2S to SO2 ? 3:1
5 How many wavelengths are used in model 900? 5
6 What is the measurement wavelength for H2S? 228nm
7 What is the measurement wavelength for SO2? 228 nm
8 What is the measurement wavelength for Sv? 229 nm
9 What is the maximum limit for sulfur vapor ? 200 ppm
10 The Sv can be controlled by controlling the _____ Sample line temperature
11 How many UV lamps are installed in the analyser ? 1
Option 2 Option 3 Option 4 Answers

Zirconia Electrolyte Capacitance Option 1

-100 deg c to +100 deg C -200 deg c to +200 deg C -50 deg c to +50 deg C Option 1

Monthly Quarterly Six months Option 4

Oxygen Temperature Pressure Option 1

Thermocouple RTD Al2O3 Option 1

Pressure Current Voltage Option 1

Electronics & Interconnecting cables Interconnecting cables Analog o/p Option 2

Pressure Current Voltage Option 2

Dewpoint to AM Voltage to MH Pressure to MH Option 1

Peltier effect Thomson effect Newtons effect Option 1

MH low Zero Span Option 2

Demister Incinerator None Option 1


1/2 1/3 None Option 3
Concentration of H2S Concentration of SO2 None Option 1
4:1 2:1 None Option 1
3 6 None Option 3
214 nm 285 nm 383 nm Option 1
285 nm 214 nm 383 nm Option 2
286 nm 309 nm 384 nm Option 4
300 ppm 100 ppm 150 ppm Option 3
Vent line temperature SKO temperature None Option 3
2 3 4 Option 2