Академический Документы
Профессиональный Документы
Культура Документы
上传
下载 打印 收藏 加入文辑 1 /7 在当前文档中搜索...
AATCC takes no positwn respecting the validity of any patent rights asserted in connection
with any item mentioned in this test method. Users of this test method are expressly advised
that determination of the validity of any such patent rights, and the risk of infringement of
such rights, are entirely their own responsibility.
This test method is subject to revision at any time by the responsible technical committee
and must be reviewed every five years and if not revised, either reaffirmed or withdrawn.
Your comments are invited either for revision of this test method or for additional
procedures and should be addressed to the AATCC Technical Center. Your comments will
receive careful consideration at a meeting of the responsible technical committee, which
you may attend. If you feel that your comments have not received a fair hearing, you
should make your views known to Diana A. Wyman, Technical Director, at the address
shown below.
This test method is copyrighted by AATCC, PO Box 12215, Research Triangle Park, NC
27709 USA. Individual reprints (single or multiple copies) of this method may be obtained by
contacting AATCC at the above address or tel: 919-549-3526; fax: 919-549-8933, or e-mail:
orders@aatcc.org.
AATCC License Agreement: This test method is copyrighted by the American Association
of Textile Chemists and Colorists (AATCC), PO Box 12215, 1 Davis Drive, Research
Triangle Park, NC USA. All rights reserved.
AATCC grants you a license as follows: The right to download one electronic file of this
AA TCC test method for temporary storage on one computer for purposes of viewing, and/or
printing one copy of the AATCC test method for individual use. Neither the electronic file nor
the hard copy print may be reproduced in any way. In addition, the electronic file may not be
distributed elsewhere over computer networks or otherwise. The hard copy print may only be
distributed to other employees for their internal use within your organization.
The American Association of Textie Chemists and Colorists Licensee-ZHEJIANG INST OF STANOARDl2ATION 5950617
Prov!OOCI by IH$ uncle< llcense with AATCC Noc 10< Resale. 2011n114 oo:46:t7
No reproduclion or nei'llOfttjng permitted without license from IHS
AATCC Evaluation Procedure 6-2016
Developed in 1995 by AATCC Commit ing] at a 45° angle to/from the specimen. All OSHA standards are rules must also
tee RA36; reaffirmed 1996, 1997, 2003, NOTE: Instruments of differe nt geo be consulted and followed.
2008; reaffirmed and editorially re metrics may produce different colorimet 1.3.2 Good laboratory practices should
vised 1998; revised 2016. Technically ric results on most textile materials. be followed. Wear eye protection in all
equivalent to ISO 105-J01. 1.2.6 reflectance, n.-the ratio of the laboratory areas.
reflected radiant or luminous energy 1.3.3 The manufacturer's safety rec
(light) to the i ncident energy in the given ommendations should be followed when
1. General Description ol Instrumental
conditions. operating laboratory testing equipment.
Color Measurement
1.2.7 reflectance factor, n.-the ratio 1.4 General Guidelines
1.1 Purpose of the light reflected from the specimen 1.4.1 In general, instrumental color
I.I .I This evaluation procedure is a to the light reflected from the perfect measurement procedures are dictated by
reference document to support the proper reflecting diffuser under the same the type of specimen to be measured and
measurement of the color stimulus of geometric and spectral conditions of the instrument with which it will be mea
specimens by instrumental means as re measurement. sured. Many types of color measuring in
quired in many of the current AATCC 1.2.8 specular reflection, n.-the re struments are available, differing in such
test methods. The document contains flection without diffusion, in accordance features as area-of-view, illumination
three major subsections covering reflec with the laws of optical reflection, as in a method and geometry. The user is cau
tance measurement, transmittance mea mirror. tioned that conflicting results may be
surement and related calculations. In ad 1.2.9 standardization, n.-of color obtained by comparing data acquired on
dition, an appendix is supplied which measuring instrumellt, the act of measur
instruments of different designs.
details specific techniques and specimen i ng one or more standard materials with 1.5 Procedure
handli ng procedures. a color measuring instrument for the 1.5.1 Operators who utilize Evaluation
1.2 Terminology purpose of calculating a set of correction Procedure 6 as a reference for color mea
surement should generally conduct such
J.2.1 area-of-view, n.-o/ color mea factors to be applied to all subsequent
suring instrument, the dimensions of the measurements. measurements as follows:
surface area that a color measuring in 1.5.2 Refer to the a ppropriate section
1.2.10 transmittance, n.-the ratio
(Reflectance or Transmittance) for speci
strument is capable of covering in a sin of transmitted light to incident light under
men to be measured.
gle color measurement. specified geometric and spectral conditions.
1.5.3 Standardize instrument (see 2.4).
1.2.2 col or m easu ring instrument, NOTE: Regular transmittance (of trans
Maintain a record of the procedure and of
n.- any device, such as a colorimeter or parent materials) is the ratio of undifuf sed
the results for any verification standards
spectrophotometer, used to meas ure the transmit.led light to incident light.
measured.
relative amounts of energy reflected from 1.2.11 transmittance factor, n.-the
1.5.4 Obtain and prepare specimen,
(or transmitted through) a specimen in ratio of the light measured after bei ng
noting any special sampli ng and/or con
the visible region of the energy spectrum transmitted through a transparent or trans
ditioni ng procedures that may be required
(comprising the wavelengths from 360- lucent specimen to that measured in the
780 nm, and including as a minimum the
(see Appendix A.1.2).
absence of the specimen. In case of liq
1.5.5 Present specimen to color measur
region from 400-700 nm). uid samples, the test measurement is usu
ing insnument, again following any spe
1.2.3 color measurement, n.-a nu ally that of the colored solution in a cu
cial techniques required for the type of ma
merical representation of the color stimu vette and the reference measurement is
terial being measured (see A ppendix A).
lus of an object obtai ned by use of a color that of the identical cuvette filled with
1.5.6 Measure the color of the speci
measuring instrument. A single measure the solvent used for dissolution (usually
men, obtaining the appropriate spectral
ment may represent an average of multi di stilled water).
reflectance or transmittance factors or
ple readings of a specimen . 1.2.12 verification standard, n.- in
colorimetric values.
1.2.4 fluorescence, n.-a phenomenon color measurement, any stable material
1.5.7 Perform any appropriate numeri
in which radiant flux of certain wave which is used for the purpose of con
cal calculations as specified in Section 4 or
lengths is absorbed and re-emitted at other firming (or verifying) the validity of an
as is required by the specific test method.
visible, usually longer, wavelengths. instrument standardization. Color mea
1.2.5 geometry, n.-of a color mea surements, which are made immediately
2. Measurement ol Color by Reflectance
suring instrumenl, one of the following following a standardization, are compared
Methods
terms (diffuse/O, O/diffuse, 0/45 or 45/0) to original measurements of the standard
which describe the angle or manner in to detect improper standardization. 2.1 Principle
which a color measuring instrument: 1.3 Safety Precautions 2.1.l Materials of an opaque or nearly
(1) iHumfoates the specimen (diffuse, 1.3.1 T hese safet y precautions are for opaque nature (but not translucent) are
oo, 450). information purposes only. The precau measured by reflectance methods to ob
(2) views the resulting reflected light tions are ancillary to the testing tain a numerical representation of the
(0 [0°-10°], diffuse, 45°, 0°). prOcedures and are not intended to be all color of the specimen. Proper equipment
Diffuse/O and O/diffuse geometry instru inclusive. It is the user's responsibility to setup, standardization of the color mea
ments contain a sphere used to diffuse the use safe and proper techniques in handling suring instrument, and proper presenta
light illuminating [or reflected from] the materials in this test procedure. Manufac tion of the test specimens to the instru
specimen, while 0/45 and 45/0 geometry turers MUST be consulted for specific ment are required to achieve consistent,
instrumenis generally use mirrors or fiber deiails suc h as safe ty data sheets and reliable, and meaningful reflectance mea
optics to direct the illumination [or view- other manufacturer's recommendations . surement results. In addition, calculation
of the colorimetric values required for lows for the inclusion or exclusion of the tion from their original values, the stan
evaluating the results must be performed specular reflecrance. The majority of dardization is not considered valid. The
in the prescribed manner. spectrophotometers in use in the textile in number of verification standards and
2.2 Uses and Limitations dustry use diffuse/0, where the actual the acceptability limits depend on user re
2.2. l This section is restricted to the viewing angle is 8°. Measurements made quirements, but are typically 1-3 stan
measurement of the color of opaque and on otber geometries will likely produce dards and an acceptance lin:tit of 0.20
nearly opaque specimens by reflectance different spectral or colorimetric results. t1Ec,11c (2: I) (DM/ I 0°) unil5.
spectrophotometers or reflectance color 2.3.6 O/diffuse (sphere) instruments are 2.5 Sa mpling
imeters. The use of special procedures similar, but die path of illumination and 2.5.1 All measurements taken on
and/or aids to the presentation of speci viewing are reversed. This method illumi color measuring instrumentation involve
mens is covered in the appendix. nates the sample at an angle between 0° "sampling. The area-ot�view of the in
"
2.3 Appara tus and Materials and 10°, and measures the an1ount of light strument, the number of presentations
2.3.1 Reflectance color measuring in reflected from the surface into the sphere. averaged to produce a single measure
strumentation illuminates a specimen imd 2.3.7 Instruments with 45/0 or (0/45) ment, the difficulty of presenting the
measures the amount of light which is re geometry illuminate the specimen at the specimen to the instrument, and the accu
flected from the surface of the specimen. first angle and view the specimen at the racy with which the sample represents the
Illumination is normally done in a poly second. These two geometries can be ei object of concern (garment, roll, dye lot,
chromatic manner (white light), however ther circumferential (viewing or illumi etc.) all play important parts in achieving
monochromatic mode is acceptable for nating at 45° to the specimen in a com meaningful and reproducible results. Re
nonfluorescent specimens. Reflectance plete circle) or directional. For most fer to ASTM E l 345 (see 7. I) and SAE
color measuring instruments may be textile samples, either 45/0 or 0/45 yield Jl 545 (see 7.3) for techniques in estab
broadly divided between two groups: equivalent results. lisbing sampling procedures.
Spectrophotometers and Colorimeters. 2.3.8 All color measuring instruments 2.6 Specimen Preparation
2.3.2 Specu·opbotometers (typically require a white calibrated standard witb 2.6.1 The ideal specimen to measure is
diffuse/0°, using polychromatic illumina which to standardize the instrument. The a rigid, non-textured, inert, opaque speci
tion) separate and measure the spectrum colorimen·ic values for this calibration men of uniform color. Such ideal speci
of light reflected from the specimen, rela standard are stored in the instrument or mens do not exist in textiles, so it be
tive to a reference, white at regular inter the software and require only that a spe comes necessary to employ techniques
vals (wavelength intervals of 5 nm, l 0 cific standard be used to standardize the and practices when measuring most tex
nm, and 20 nm are most common). This instrument. The co1Tect white standard is tiJe materials which eliminate or reduce
data may be used to calculate the needed usually identified with a serial number. the effect any objectionable characteris
tristimulus values (X,Y,Z) for the selected 2.4 Standardization tics have on the insm1mental color mea
.ill un:tinant/observer combination. Some 2.4.1 Proper standardization of any surement. Specific procedures and tech
spectrophotometers (typically 0/diffuse) color measuriog instrument is absolutely niques for handling specimens which
illuminate the sample with monochro necessary to achieve unifonn, accurate meet the folJowing characteristics are
matic light, and measure the an1ount of results. While different types of instru presented in Appendix A.
1ight reflected from the surface as the ments require varying methods of stan 2.6.2 Fluorescence of the specimen
sample is illuminated ar regular wave dardization, tbere are common principles (from dyes or fluorescent whitening
length intervals. which must be observed. agents [FWAs]) will influence the results
2.3.3 Colorimeters measure the tristim 2.4.2 ln general, an instrument stan depending on the amount of fluorescing
ulus values (X, Y,Z) directly through dardization involves measuring a clean material present and the amount and
broadband filters which are designed to white surface of known reflectance fac quality of ultraviolet and visib.le energy
produce colorimetric values for one illu tors (referenced to a perfect reflecting in the instrument light source. Results
minant and observer (typically C/2°). diffuser) and calculating (through soft may be pa1ticularly hard to duplicate be
Measurement of reflectance factors at ware built into the instrument or com tween instruments. Example materials
specific wavelengths is not possible with puter program) a series of correction fac are white or lightly colored materials
a colorimeter. tors which will be applied to all treated with FWAs (see Al .1).
2.3.4 Witbio these two categories, the subsequent measurements. Some instru 2.6.3 Moisture Content of textile mate
instrument5 are further defined by their ments also require a black tile (or light rials can affect their color and appearance
geometry. Two types of geometties are trap), and possibly a gray tile. Each of characteristics. Tbe amount of condition
prevalent: Sphere (also referred to as dif these materials must be maintained in it� ing time necessary to achieve a stable
fuse/O [d/0] or 0/diffuse [0/d]) aod 45/0 original clean, unscratched condition. Re moisture state varies with fiber, fabric
or 0/45. The first term in each geometry fer ro manufacturer's recommendations construction, dyes and surrounding con
type refers to the method (or angle) by for cleaning instructions. ditions. Examples of materials which are
which the sample is illuminated (ex: 45° 2.4.3 The frequency with which this typically affected by moisture content are
in a 45/0 instrument). The second term standardization must be performed depends cotton and rayon fabrics (see A1.2).
refers to the angle at wbich the instru on many factors including the type of in 2.6.4 Non-Rigid specimens tend to pro
ment views the illuminated specimen (ex: strument, the environmental conditions in trude (or "pillow") into the viewing port
0° in a 45/0 instrument). which the instrument operates, and the re of instruments which contact the speci
2.3.5 Diffuse/O (sphere) instruments il quired accuracy of the results. For most ap men at the measurement port. The
luminate the specimen indirectly when tl1e plications, an interval of 2-4 h is typical. amount of intrusion may vary depending
specimen is placed against a port opening 2.4.4 Once the standardization step has on number of layers, softness of material
into a diffusely illuminated sphere, and been performed, it is important to verify and the backing pressure applied to
view the specimen at an angle between 0° the success of the procedure by measur mount the specimen. Variations in the
and 10° from the perpendicular. This ar ing some colored materials (verification amount of intrusion will result in signifi
rangement is designed to capture all light standards) and comparing the resulting cant deviations in the resulting color
reflected from the specimen. Some sphere colorimetric values to tbe original values measurement which are both unpredict
instruments with a viewing angle greater for these materials. lf the measured val able and non-reproducible. Examples of
than 0° include a specular port wh.ich al- ues do not fall within an acceptable varia- these materials are fiber, yam, knits and
3. Measurement of Color by if
The use of one or more colored fllters as
Transmittance Methods Y!Y,, > 0.008856
a verification standard to check both
3.1 Principle photometric and wavelength accuracy is but
, 3.1.1 Materials of a transparent nature recommended. L* = 903.3 (YIY,,)
are measured by transmittance methods 3.5 Sampling if
to obtain a numerical representation of 3.5.1 Sampling procedures for trans YIY,, $ 0.008856
their color and/or their relative strength mission methods depend on the nature of a* =5001/(X/X,,) -f(YIY,,)]
(concentration) compared to that of a the material and the type of specimen col b* = 200[/(Y/Y,,) - f('ZJZ,,)]
standard material. The most common ap lected. Powder and paste specimens
plication is the measurement of solutions should be taken from enough locations to where :
of dyes in clear solvents contained in a ensure a reproducible sample within the f(X!X,,) = (X/X,,)'13
glass cuvette or flow cell for the pur population being sampled and thoroughly if
poses of determining colorant identity, mixed prior to proceeding with weighing XIX,,> 0.008856
relative strength or color difference. and dissolving. or
3.2 Uses and Limitations 3.6 Specime n Preparation f(XIX,,) = 7.787(X/X,,) + 16/ l l 6
3.2. l This method is generally suitable 3.6. l Specimens must be prepared by if
for the measurement of true solutions prescribed laboratory procedures using XIX,,$ 0.008856
which are transparent and void of particu analytical balances and volumetric glass f(YIY,,) = (YIY,,)'13
late matter. Although translucent or tur ware. Solutions should be prepared at a if
bid liquids are sometimes measured by concentration at which the percent trans YIY,,> 0.008856
transmittance methods, such methods are mittance of the solution in the transmit or
beyond the scope of this evaluation tance cell is between I 0% and 50% at the j(YIY,,) = 7.787(Y/Y,,) + 16/116
procedure . wavelength of maximum absorbance (see if
3.3 Apparatus, Reagents, and Materials 7.4). Each specimen must be in true solu YIY,, � 0.008856
3 .3 .1 Transmittance instruments may tion, not in suspension, di spersion or as- f(ZIZ,,) = (Z/Z,,)113
if wavelength O.) is: where:
ZIZ,, > 0.008856 KIS = ((1.0 - R1)2)/(2.0R,) K/S = KJS as calculated in equation
or 4.3.2 above
where:
f(ZIZ.) =
7.787(Z/Z,,) + 16/116 E = Energy of selected illuminant
Ri. is the reflectance factor of the spec
if (normally 065)
imen at wavelength A (the %R value is
ZIZ,,::; 0.008856 x,y,z = tristimulus weighting values
normally measured by a specu·ophotometer
C*,w (a*2 + b*2)1n
= for selected observer (normally
and normalized to LO [i.e., 100% I .OJ ) .
hob= arctan( b*/a*) expressed on a 0° to
=
JOO)
* 4.3.3 For applications where the Pineo
n number of wavelength inter-
360° scale with the a positive axis be correction form of the equation is re
=
*
ing 0° and the b positive axis at 90°. vals used
quired (generally for deeply dyed tex
For these equations, X,,, Y. and Z,, are tiles), the fomrnla used is: (TSVSTR) tristimulus color strength
the tristimulus values of the illuminant. value (X,Y,Z function). Normally the Y
KIS = (1.0 - (R1• - s)2)/2.0(R,_ - s)
For daylight the preferred illuminant/ value is used as a gross color strength
where: value related to the visual lightness
observer combination is 065/10°. Table I
s is the minimum reflectauce achiev function, although either the X or Z
gives the values for all combinations in
able in the deepest dyed shade and is used may be chosen for measurement of
ASTM E308.
for all wavelengths. components where the absorption
4.3 Color Strength Value using Re
4.3.4 The four types of Color Strength characteristics are known and fall into
flectance Measurement
Value calculations that are most com widely separated regions of the visible
4.3.1 Color Strength Value is a single
monly used are: spectrum. In most applications, the
numerical value related to the amount of
color absorbing material (colorant) con smallest of the 3 tristimulus values (X,
(SWL) KJS at single wavelength (usu
tained in a specimen. It is most often used Y or Z) is chosen and is used in place of
ally wavelength of maximum absor
to calculate the difference in strength (% the %R value in equation in 4.3.2
bance). Use equation 4.3.2 and calcu
strength) between two colored speci above. While the TSVSTR equation is
late K/S for a single wavelength.
mens. Color Strength Value may be cal used in the textile industry, no specific
culated by any one of four acceptable (SUM) K/S summed over wavelength scientific suppon of it can be found in
methods. The value which results from interval within the visible spectrum. normally accepted reference books.
Use equation 4.3.2, calculate K/S at
one method may not agree with any other NOTE: See 4.5 below to calculate the
method. The choice of method is usually each desired wavelength interval and
relative difference in strength between
dependent on the nature of the specimen sum. The resulting value may be nor
two specimens using one of these calcu
and the need for obtaining a color malized by dividing it by the number lation methods.
of intervals summed.
strength value. Colorimeters must calcu 4.4 Color Strength Value using
late this value using a tristimulus function (WSUM) K/S weighted by visual Transmittance Measurement
method. The methods used are labeled as function (such as the x,y,z functions 4.4.1 Color Strength Value is a single
SWL, SUM, WSUM and TSVSTR, and and the 065 illuminant energy function) numerical value related to the amount of
are calculated as follows: and summed over a wavelength inter color absorbing material (colorant) con
4.3.2 Color Strength Value of speci val within the visible spectrum, then tained in a solution. It is most often used
mens measured on a spectrophotometer divided by the number of wavelength to calculate the difference in strength (%
most often involves calculation of a K/S intervals summed. strength) between two colored solutions.
value at one or more wavelength inter Color Strength Value may be calculated
vals. The commonly used equation for WSUM = Li. [(K/S,_ * xi. * E1) by any one of four acceptable methods.
*
(K/S1, + Y1-. E1-.)
calculation of the K/S value for opaque +
The value which results from one method
+ (K/S1. * z,
* E,_)]ln
specimens (i.e., textiles) at a specified may not agree with any other method.
The choice of method is usually depen
dent on the nature of the specimen and
Table 1-Tristimulus Values for Illuminant/Observer Combinations the need for obtaining a color strength
value. Colorimeters must calculate this
Trislimulus Values
Illuminant/Observer value using a tristimulus function
Combinations x. Y, z. method. The methods used are labelled as
110° observer) SWL, SUM, WSUM and TSVSTR, and
are calculated as follows:
N10° 111.146 100.000 35.203
4.4.2 Color Strength Value of specimens
C/10° 97.285 100.000 116.145
o:.o110• measured on a spectrophotometer most ot�
96.720 100.000 81.427
Dssf10° 95.799 100.000 90.926 ten involves calculating an absorbance
Dss/10° 94.811 100.000 107.304 value at one or more wavelengths. The
D7J10' 94. 41 6 100.000 120.641 eq ua tion for calculating the absorbance
F2/10° (Cool White Fluorescent) 103.279 100.000 69.027 value at a specified wavelength (1.) is:
F7/10° (D ayl ight Fluorescent) 95.792 100.000 107.686 A1-. = log,0 (l.O/t1)
F11/10° (Ultralume 4000, TL84) 103863 100 000 65.607 where:
12° observer) ,;,_ is the internal transmittance of the
A/2' 109.850 100.000 35.585 specimen (the %T value normally mea
C/2° 98.074 100.000 118.232 sured by a spectrophotometer and nor
D:J2° 96.422 100.000 82.521 malized to 1.0 (i.e., 100% = 1.0)).
Dssf2' 95 682 100000 92.149 4.4.3 The four types of Color Strength
DssJ2° 95.047 100.000 108.883
Value calculations that are most com
07,)2° 94.97 2 100.000 122.638
monly used are:
F2/2'(Cool White Fluorescent) 99.186 100.000 67.393
F7/2°(0aylight Fluorescent) 95.041 100.000 108.747 (SWL) absorbance at single wavelength
Fl 1/2° (Ultralume 4000, TL84) 100.962 100.000 64.350 (usually wavelength of maximum ab-
sorbancc). Use equation in 4.4.2 and calculate the colorimetric values. induced fluorescence. Note that this tech
calculate A for a single wavelength. 5.4 Identification of specimen. nique may result in disagreement with vi
(SUM) absorbance- summed over 5.5 The color strength value calcula sual results. Note also that this technique
wavelength inte� within the visible tion (if any) used (SWL, SUM, WSUM, is only applicable in those cases where
spectrum. U e equation in4.4.2. calcu TSVSTR). the fluorescence is caused by absorption
late A at each desired wavelength in 5.6 The specimen presentation and aver of ultraviolet radiation. Instruments
terval and sum. The resulting value aging methods employed, such as area-of which can control the UV energy will
may be normalized by dividing it by view, number of thick'llesses. temperature produce results more consistent with vi
the number of intervals summed. and % relative humidiry if not at standard sual observation. but the.�e result� may be
conditions and number of readings in a harder to reproduce on other non-similar
(WSUM) absorbance weighted by
measurement. instruments. in either case when speci
visual function (such as the x.y,z func
mens are fluorescent, all specimens to be
tiornc and the 06.1 illuminant energy 6. Precision and Bias compared must be measured on the same
function) and summed over a wave
instrument at nearly the same time. Previ
length interval within the visible spec 6.1 Evaluation Procedure 6 is meant to
ously measured data (standards, controls.
trum, then divided by the number of be used as a supportive reference method.
etc.) should not be used for direct com
wavelength intervals summed. Specific precision and bias statements
parison. Fluorescent specimens which
contained within each referring AATCC
WSUM = I:,_ [(A, * x, * E,) absorb visible energy cannot be measured
test method must be used ro evaluate the
+(Ai, * Yi. * E,) consistently except on an instrument
significance of any colorimetric results
+(Ai.* q* E,)J/ n where the quantity and quality of incident
obtained as pan of that test method pro
where: energy can be controlled. While such in
cedure. The user is cautioned that preci
A Absorb:mce calculated in equa- struments do not currently exist commer
= sion and bias are affected in part by sam
tion in 4.4.2 above cially, the best alternative is to use a 04
/ 5,
ple presentation, averaging methods, and
E Energy of selected illuminant 4510 (circumferential or bidirectional)
= the repeatability and accuracy of the indi
(normally 06�) type measuring instrument.
vidual color measuring instrument used.
x.y.z = tristimulus weighting values A 1.2 Moisture Cont ent-If the exist
for selected observer (normally 7. References ence of moisture affects the measurement
)Qo) of color it is necessary 10 "condition" the
7.1 ASTM standards arc available from
n number of wavelength intervals specimens to allow the moisture content
ASTM. W. Conshohocken. PA; web �ite:
=