Mac-Lab/CardioLab UPS

Installation Instructions

Introduction
Document Use
Use the following instructions to install the IEB UPS. These instructions only apply to Mac-Lab®/
CardioLab® systems that require an IEB UPS installation.

Revision History

Revision Date Comments

A 18 September 2006 Initial release of document.

B 20 May 2009 Updated with current photos and reformatting.

C 22 August 2012 Updated with new part numbers and reformatting.

D 25 June 2014 Updated to correct a label name on the UPS. Updated part numbers for RoHS
compliance.

Required Parts for 100V Systems

Qty. Part Number Description

1 2034079-002 UPS POWERCOM, 100V, 50/60HZ, 1500VA, BNT

1 2033852-001 IEB, W/VIDEO, ANALOG I/O, NO MDM 100V

1 2079326-009 IEB 100V MAC-LAB

1 2079326-010 IEB 100V CARDIOLAB

1 2079326-011 IEB 100V SPECIALSLAB

1 2079326-012 IEB 100V COMBOLAB

1 2003927-001 IEC to IEC Power Cable

1 405535-013 NEMA to IEC Power Cable

2033819-011D EN Mac-Lab/CardioLab UPS Installation Instructions 1


25 June 2014
Required Parts for 120V Systems

Qty. Part Number Description

1 2034077-002 UPS POWERCOM, 120V, 50/60HZ, 1500VA, BNT

1 2033852-002 IEB, W/VIDEO, ANALOG I/O, NO MDM 120V

1 2079326-001 IEB 110V MAC-LAB

1 2079326-002 IEB 110V CARDIOLAB

1 2079326-003 IEB 110V SPECIALSLAB

1 2079326-004 IEB 110V COMBOLAB

1 2003927-001 IEC to IEC Power Cable

1 80274-006 NEMA to IEC Power Cable

Required Parts for 240V Systems

Qty. Part Number Description

1 2034078-002 UPS POWERCOM, 230V, 50/60HZ, 1500VA, BNT

1 2033852-003 IEB, W/VIDEO, ANALOG I/O, NO MDM 240V

1 2059901-004 IEB, MAC-LAB, C14 CONN, 240V CHINESE

1 2059901-005 IEB, CARDIOLAB, C14 CONN, 240V CHINESE

1 2059901-006 IEB, COMBOLAB, C14 CONN, 240V CHINESE

1 2059901-007 IEB, CARDIOLAB, C14 CONN, 240V KOREAN

1 2059901-008 IEB, MAC-LAB, C14 CONN, 240V KOREAN

1 2059901-009 IEB, COMBOLAB, C14 CONN, 240V KOREAN

1 2079326-005 IEB 240V MAC-LAB

1 2079326-006 IEB 240V CARDIOLAB

1 2079326-007 IEB 240V SPECIALSLAB

1 2079326-008 IEB 240V COMBOLAB

2 2003927-001 IEC to IEC POWER CABLE

2 Mac-Lab/CardioLab UPS Installation Instructions 2033819-011D EN


UPS Set Up Instructions
Install the IEB UPS on 100V Systems and on 120V Systems
Use the following procedure to install the current IEB UPS with a Powercom 1500VA UPS.

WARNING: SHOCK HAZARD

Do not remove equipment covers or otherwise work on equipment unless it is


de-energized. Ensure all equipment is de-energized and unplugged before
proceeding.

1. Shut down the Mac-Lab/CardioLab computer connected to the IEB.

2. Unplug the main power cord and turn off both power switches on the IEB and the UPS.

Refer to the Mac-Lab/CardioLab/Centricity Cardiology INW Server System Service Manual


for switch locations and detailed disassembly instructions.

3. Place the Powercom UPS from the kit on a level surface near the IEB.

2033819-011D EN Mac-Lab/CardioLab UPS Installation Instructions 3


Qty Part Number Description

1 2003927-001 IEC to IEC Cable

1 405535-013 NEMA to IEC Power Cable (100V)

1 80274-006 NEMA to IEC Power Cable (120V)

4. Connect the PN 2003927-001 IEC to IEC cable to the UPS.

5. Connect the other end of the PN 2003927-001 IEC to IEC cable to the IEB power receptacle
labeled 'UPS Power to Comp'.

6. Connect the appropriate end of PN 80274-006 for 120V or PN 405535-013 for 100V power
cord to a UPS output power receptacle (refer to the previous figure).

7. Connect the other end of the power cord PN 80274-006 for 120V or PN 405535-013 for 100V
from the UPS into the HP Computer.

8. Turn on both IEB power switches and turn on the UPS.

9. Start up the Mac-Lab/CardioLab computer.

10. Log on to the Mac-Lab/CardioLab Custom Shell as local mlcluser.

11. Test the unit by turning off the main power switch on the rear of the IEB for 10 seconds.

12. Verify power to the computer is not interrupted and the computer power LED is still green.

13. Verify the UPS is emitting audible beeps.

14. Return power to the IEB by turning on the main power switch on the rear of the IEB.

15. Verify the UPS is no longer emitting audible beeps.

16. Verify power to the computer was not interrupted by verifying the computer is still logged on
to the Mac-Lab/CardioLab Custom Shell.

17. If the verification steps above do not execute successfully, contact Tech Support.

4 Mac-Lab/CardioLab UPS Installation Instructions 2033819-011D EN


Install the IEB UPS on 240V Systems
Use the following procedure to install the current IEB UPS with a Powercom 1500VA UPS.

WARNING: SHOCK HAZARD

Do not remove equipment covers or otherwise work on equipment unless it is


de-energized. Ensure all equipment is de-energized and unplugged before
proceeding.

1. Shut down the Mac-Lab/CardioLab computer connected to the IEB.

2. Unplug the main power cord and turn off both power switches on the IEB and the UPS.

Refer to the Mac-Lab/CardioLab/Centricity Cardiology INW Server System Service Manual


for switch locations and detailed disassembly instructions.

3. Place the Powercom UPS from the kit on a level surface near the IEB.

4. Connect the IEC to IEC Cable PN 2003927-001 to the UPS.

5. Connect the other end of the PN 2003927-001 IEC to IEC cable to the IEB power receptacle
labeled 'UPS Power to Comp'.

6. Connect the appropriate end of the PN 2003927-001 power cord to a UPS output power
receptacle (refer to the previous figure).

2033819-011D EN Mac-Lab/CardioLab UPS Installation Instructions 5


GE Medical Systems Information Technologies. • 8200 W. Tower Ave.
Milwaukee, WI 53223 U.S.A.
© 2012 - 2014 GE Medical Systems Information Technologies. All rights reserved.
GE Medical Systems Information Technologies, a General Electric Company.

CAUTION: POWER OUTAGE

The white UPS output power receptacle labeled “BYPASS OUTLET” does not
supply battery backup power in the event of a power outage. Do not use the
BYPASS OUTLET power receptacle. The HP computer must be plugged into
one of the UPS output receptacles to ensure that power is maintained to the
computer during a power outage.

7. Connect the other end of the PN 2003927-001 power cord to the HP Computer.

8. Turn on both IEB power switches and turn on the UPS.

9. Start up the Mac-Lab/CardioLab computer.

10. Log on to the Mac-Lab/CardioLab Custom Shell as local mlcluser.

11. Test the unit by turning off the main power switch on the rear of the IEB for 10 seconds.

12. Verify power to the computer is not interrupted and the computer power LED is still green.

13. Verify the UPS is emitting audible beeps.

14. Return power to the IEB by turning on the main power switch on the rear of the IEB.

15. Verify the UPS is no longer emitting audible beeps.

16. Verify power to the computer was not interrupted by verifying the computer is still logged on
to the Mac-Lab/CardioLab Custom Shell.

17. If the verification steps above do not execute successfully, contact Tech Support.