Control Unit
controls the analyzer box
Analyzer box and displays the measure-
ment values
robust, stable housing with
integrated rubber edges,
protects the sensors,
pumps, analysis and
memory electronics
¸
Colour graphic display
innovative operating convenience
thanks to an easy, clear menu
structure
Engine l ≤ 1
User-defined
Options OK
2
testo 350
Li-ion battery,
mains-independent
˝Service opening
power supply for seve-
ral hours
Output of
external
cooling loop
The measurement
˛ gas pump
Sensor chamber transports the gas
to the gas sensors
Input of
external
cooling loop
—
ˇGas sensors Condensate container
The fresh air and dilution The condensate pump
pump provides the gas transports occurring
sensors with fresh air as condensate into the
required condensate container
Plug & play: Easy gas sensor replacement Condensate container: Condensate container quickly and easily
emptied
3
testo 350 – Flue gas measurement at the highest level, thanks to:
Filters can be easily exchanged Service opening with easy access Continuous condensate drain-off Control unit connections
without tools to wearing parts for long -term measurements
4
We measure it.
Flue gas analysis – brilliantly easy: testo 350, the only one that thinks ahead!
The portable flue gas analyzer testo 350 is the ideal tool for professional Typical fuels, a practicable order of the exhaust gas parameters in the dis-
flue gas analysis. Helpful instrument settings guide the user safely through play, the corresponding calculations as well as useful instrument
typical measurement tasks such as: pre-settings, are stored under each of these measurement objects. Exam-
ples of these are the activation of the dilution in measurements on l ≤ 1
3 Flue gas analysis in commissioning, setting, optimization or operational
regulated industrial engines and gas turbines, or the testing of the relevant
measurements on industrial burners, stationary industrial engines, gas
gas sensor in the dilution slot.
turbines and flue gas purification systems.
3 Control and monitoring of officially prescribed emission limits in exhaust The advantages of the application-specific menu
gas.
3 Information in the display guides the user through the menu.
3 Function testing of stationary emission measuring instruments.
3 Easy operation without previous knowledge of the instrument
3 Control and monitoring of defined gas atmospheres in furnace rooms or
3 Reduction of the work steps before the start of the measurement.
kilns in different processes.
Control unit is simply clicked in Control unit turned over: safe trans- Robust flue gas analyzer for industrial use, even under the toughest condi-
port to the measurement site tions
5
testo 350
testo 350
Part no.
testo 350 Control Unit 0632 3511
Option BLUETOOTH® wireless transmission
testo 350 Analyzer box 0632 3510
Option CO (H2-compensated) sensor, 0 to 10000 ppm
Option NO sensor, 0 to 4000 ppm, resolution 1 ppm
Option NO2 sensor, 0 to 500 ppm, resolution 0.1 ppm
Option CxHy sensor (Pellistor)
Option Peltier gas preparation incl. hose pump
Option BLUETOOTH® wireless transmission
Option fresh air valve for long-term measurement, incl.
measuring range extension with dilution factor 5 for all
sensors
Option measuring range extension for individual slot
with the following selectable dilution factors:
0, 2, 5, 10, 20, 40
Flue gas probe for industrial engines, probe shaft 0600 7550
length 335 mm, incl. cone, heat protection shield, spe-
cial hose for NO2-/SO2 measurement, Tmax probe
shaft 1000 °C, hose length 2.2 m
Set BLUETOOTH® printer with wireless Bluetooth inter- 0554 0553
face, incl. 1 roll of thermal paper, rech. battery, mains unit
easyEmission software 0554 3334
100-240 V AC / 6.3 V DC international mains unit 0554 1096
Transport case for safe and tidy storage for flue gas 0516 3510
analyzer testo 350, gas sampling probe and accesso-
ries, dimensions 570 x 470 x 210 mm (LxWxH)
Part no.
testo 350 Control Unit 0632 3511
Option BLUETOOTH® wireless transmission
testo 350 Analyzer box 0632 3510
Option CO (H2-compensated) sensor, 0 to 10000 ppm
resolution 1 ppm)
Option NO sensor, 0 to 4000 ppm, resolution 1 ppm
Option NO2 sensor, 0 to 500 ppm, resolution 0.1 ppm
Option SO2 sensor, 0 to 5000 ppm, resolution 1 ppm
Option Peltier gas preparation incl. hose pump
Option BLUETOOTH® wireless transmission
Option measuring range extension for individual slot
with the following selectable dilution factors:
0, 2, 5, 10, 20, 40
Gas sampling probe, modular, incl. special hose for 0600 8764
NO2-/SO2 measurement, cone, thermocouple NiCr-Ni
(TI), probe shaft length 335 mm, Tmax probe shaft
1000 °C, hose length 2.2 m
100-240 V AC / 6.3 V DC international mains unit 0554 1096
Transport case for safe and tidy storage for flue gas 0516 3510
analyzer testo 350, gas sampling probe and accesso-
ries, dimensions 570 x 470 x 210 mm (LxWxH)
Advantages in emission measurement on industrial engines
3 Highly accurate NOx measurement 3Special instrument menu for the testing of exhaust gas pre-
The testo 350 allows the separate measurement of NO and NO2. The high paration systems
and fluctuating NO2 components in engine exhaust gas make this measu- This exhaust gas menu allows the simultaneous measurement of exhaust
rement necessary, in order to display the real NOx value of the engine. In gas concentrations before and after the catalytic converter. For this purpo-
addition to this, the integrated gas preparation and the exhaust gas probe se, two analyzer boxes are connected to each other with the Testo
with a special hose provide protection from NO2 and SO2 absorption. databus cable. The measurement values of the two analyzer boxes are
shown parallel to each other in the control unit's display, enabling a fast
3Automatic measurement range extension for unexpectedly overview of the status of the converter.
high CO concentrations
In measurements on unfamiliar systems, or under less than ideal operating 3 Spatial distances
conditions, unexpectedly high emission values (e.g. CO concentrations up For greater distances between the gas sampling site and the adjustment
to 50,000 ppm) can occur. In these cases, the measurement range exten- site, the control unit can be connected with the analyzer box either via the
sion is automatically activated. This means maximum sensor life. These Testo databus cable or by Bluetooth®
useful pre-settings are already stored application-specifically in the instru-
ment – the testo 350 thinks ahead! testo 350: Flue gas analysis
Options OK
5. Documentation 4. Start measurement 3. Exhaust gas selection 2. Fuel selection 1. Application selection
05.05.10 09:06 05.05.10 09:06 05.05.10 09:06 05.05.10 09:06
Testo AG/Burner Testo AG/Burner Testo AG/Burner Testo AG/Burner
Light fuel oil Flue gas Light fuel oil Flue gas Light fuel oil Flue gas Light fuel oil 05.05.10 09:06
Measurement Measurement Measurement type Ready Testo AG/Burner
Light fuel oil
Light fuel oil
Flue gas Burner
Flue gas Heavy fuel oil
(before + after cat.) Turbine
Natural gas
Flue gas + m/s Engine λ > 1
LPG
Flue gas + ∆P Enginer λ ≤ 1
Wood 15%w
User-defined
Draught measurement
Briquette
Options Options OK Coeff. OK Change list
Options OK
6
Advantages in emission measurement on gas turbines
3 Easy, precise test gas adjustment by the user 3 Highly accurate NOx measurements at low concentrations
In order to fulfil the highest accuracy and comparability requirements, the Emission measurement during monitoring and adjustment work on Low-
testo 350 can, if needed, be adjusted using test gas on site. NOx gas turbines requires a very high level of measurement accuracy
because of the low NO concentrations. Thanks to the combination of the
3 Use under tough conditions
NO2 sensor and the special NOlow sensor with a resolution of 0.1 ppm,
Special chambers and closed cooling loops isolate the instrument electro-
exactly these requirements are fulfilled. In addition to this, the integrated
nics and sensors from the ambient air. This means the sensor chamber is
gas preparation and the special exhaust gas probe for industrial engines
thermally separated from the other instrument components, and possible
with a special hose provides protection from NO2 absorption.
drift of the sensors due to thermal influences are reduced.
3 Combination of measuring range extension and COlow
sensor
Thanks to the freely selectable dilution levels, concentrations of up to a
maximum of 20,000 ppm can also be measured without any problems
with the COlow sensor (measuring range 500 ppm).
in only 5 steps
Options OK
1. Application selection 2. Fuel selection 3. Exhaust gas selection 4. Start measurement 5. Documentation
Options OK
7
We measure it.
Part no.
testo 350 Control Unit 0632 3511
Option BLUETOOTH® wireless transmission
testo 350 Analyzer box 0632 3510
Option COlow (H2-compensated) sensor, 0 to 500 ppm,
resolution 0.1 ppm
Option NOlow-Sensor, 0 to 300 ppm, resolution
0.1 ppm
Option NO2 sensor, 0 to 500 ppm, resolution 0.1 ppm
Option Peltier gas preparation incl. hose pump
Option BLUETOOTH® wireless transmission
Option fresh air valve for long-term measurement, incl.
measuring range extension with dilution factor 5 for all
sensors
Option measuring range extension for individual slot
with the following selectable dilution factors:
0, 2, 5, 10, 20, 40
Flue gas probe for industrial engines, probe shaft 0600 7550
length 335 mm, incl. cone, heat protection shield, spe-
cial hose for NO2-/SO2 measurement, Tmax probe
shaft 1000 °C, hose length 2.2 m
100-240 V AC / 6.3 V DC international mains unit 0554 1096
Transport case for safe and tidy storage for flue gas 0516 3510
analyzer testo 350, gas sampling probe and accesso-
ries, dimensions 570 x 470 x 210 mm (LxWxH)
Part no.
testo 350 Control Unit 0632 3511
Option BLUETOOTH® wireless transmission
testo 350 Analyzer box 0632 3510
Option CO (H2-compensated) sensor, 0 to 10000
ppm, resolution 1 ppm
Option CO2 (NDIR) sensor, 0 to 50 Vol %, resolution
0.01 Vol %, infrared measurement principle, incl. abso-
lute pressure measurement, condensate container
filling level monitoring and CO2 absorbtion filter with fil-
ler pack
Option NO sensor, 0 to 4000 ppm, resolution 1 ppm
Option NO2 sensor, 0 to 500 ppm, resolution 0.1 ppm
Option Peltier gas preparation incl. hose pump
Option BLUETOOTH® wireless transmission
Gas sampling probe, modular, incl. special hose for 0600 8764
NO2-/SO2 measurement, cone, thermocouple NiCr-Ni
(TI), probe shaft length 335 mm, Tmax probe shaft
1000 °C, hose length 2.2 m
easyEmission software 0554 3334
100-240 V AC / 6.3 V DC international mains unit 0554 1096
Transport case for safe and tidy storage for flue gas 0516 3510
analyzer testo 350, gas sampling probe and accesso-
ries, dimensions 570 x 470 x 210 mm (LxWxH)
We measure it.
testo 350
05.05.10 09:06
ment is necessary
3 Specific fuels are pre-set for the application
3 Application-specific flue gas parameters are stored in the menu
Box 3
3 Instrument settings such as the dilution factor of gas sensors are
activated by application
In this measurement, very high CO con-
centrations can occur at short notice. In 3 Automatic testing of whether the relevant gas sensors are con-
order to protect the sensors, and to achie- nected to the intended dilution slot
ve a longer sensor life, please plug the
CO sensor into the dilution slot. The sen-
3 Special measurement mode for the testing of catalytic converters
sor may only be exchanged when the
with two flue gas analyzers
instrument is switched off! 3 Faster and easier analysis of the system status thanks to graphic
and coloured presentation of the measurement values – a glance
is enough!
Replacing filter
8
We measure it.
Using the "easyEmission" software, data can be read out, conveniently processed, archived
and managed.
Advantages of easyEmission
Wireless transfer*
• Bluetooth®
2.0 (up to 100 m without obstruction)
• USB
• Infrared interface
(Communication with the Testo printer)
• Testo databus
(up to 800m cable length) for the simultane-
Testo databus
ous operation of up to 16 analyzer boxes.
Control optionally via PC, Testo databus con-
troller or control unit.
Testo databus
controller
Databus
connection cable
9
testo 350
Spare thermal paper for printer (6 rolls), permanent, measure- 0554 0568
Option CxHy sensor, methane 100 to 40000 ppm, propane 100 to 21000
ppm, butane 100 to 18000 ppm, resolution 10 ppm. Pellistor is adjusted to ment data documentation can be read for up to 10 years
methane ex-works 0554 0569
Spare thermal paper for printer (6 rolls)
Option H2S sensor, 0 to 300 ppm, resolution 0.1 ppm
Calibration certificates
Option BLUETOOTH® wireless transmission ISO calibration certificate flue gas, calibration points 2.5 Vol% O2 0554 0003
100 and 1000 ppm CO 800 ppm NO 80 ppm NO2 1000 ppm
Option Peltier gas preparation incl. peristaltic pump for automatic condensa- SO2
te evacuation
ISO calibration certificate flow velocity, hot wire/vane anemo- 0554 0004
Option fresh air valve for long-term measurement, incl. measuring range meter, Pitot tube; calibration points 1, 2, 5, 10 m/s
extension with dilution factor 5 for all sensors. For measurements >2 hours,
the option Peltier gas preparation is additionally recommended. ISO calibration certificate flow velocity, hot wire/vane anemo- 0554 0034
meter, Pitot tube; calibration points 5, 10, 15, 20 m/s
Option measuring range extension for individual slot with the following selec-
table dilution factors: 0, 2, 5, 10, 20, 40
Option DC voltage input 11V to 40V
Option special gas pump for long-term measurements with extended gua-
rantee. For measurements >2 hours, the option Peltier gas preparation is
additionally recommended.
Transport case for safe and tidy storage for flue gas analyzer 0516 3510
testo 350, gas sampling probe and accessories, dimensions
570 x 470 210 mm (LxWxH)
Carrying strap set for analyzer box testo 350
Carrying strap set for analyzer box testo 350 0554 0434
Spare dirt filter for analyzer box testo 350 (20 off) 0554 3381
Hose set to convey flue gas from analyzer box testo 350, 0554 0451
length 5m
Wall holder for flue gas analyzer testo 350, lockable 0554 0203
10
We measure it.
Gas sampling probe, modular, incl. special hose for NO2-/SO2 0600 8764 Flue gas probe with pre-filter for industrial engines, probe shaft length 0600 7551
measurement, cone, thermocouple NiCr-Ni (TI), probe shaft length 335 mm, incl. cone, heat protection shield, special hose for NO2-/SO2
335 mm, Tmax probe shaft 1000°C, hose length 2.2 m measurement, Tmax probe shaft 1000 °C, hose length 2.2 m
Gas sampling probe with pre-filter, modular, incl. special hose for 0600 8766 Flue gas probe accessories for industrial engines Part no.
NO2-/SO2 measurement, cone, thermocouple NiCr-Ni (TI), probe
Thermocouple for flue gas temperature measurement, NiCr-Ni, 0600 8894
shaft length 335 mm, Tmax probe shaft 1000°C, hose length 2.2
length 400 mm, Tmax. +1000 °C with 2.4 m connection cable
m, Ø pre-filter 14 mm
and additional temperature protection
Gas sampling probe, modular, incl. special hose for NO2-/SO2 0600 9767 0600 8895
Thermocouple for flue gas temperature measurement, NiCr-Ni,
measurement, cone, thermocouple NiCr-Ni (TI), probe shaft length length 400 mm, Tmax. +1000 °C with 5.2 m connection cable
700 mm, Tmax probe shaft 500°C, hose length 2.2 m and additional temperature protection
Gas sampling probe, modular, incl. special hose for NO2-/SO2 0600 8765 0554 7455
Spare probe shaft with pre-filter for flue gas probe for industrial
measurement, cone, thermocouple NiCr-Ni (TI), probe shaft length engines, probe shaft length 335 mm, Tmax probe shaft 1000°C
700 mm, Tmax probe shaft 1000°C, hose length 2.2 m
Temperature sensors Part no.
Gas sampling probe with pre-filter, modular, incl. special hose for 0600 8767
NO2-/SO2 measurement, cone, thermocouple NiCr-Ni (TI), probe Combustion air temperature probe, immersion depth 60 mm 0600 9797
shaft length 700 mm, Tmax probe shaft 1000°C, hose length 2.2 m,
Ø pre-filter 14 mm Pitot tubes Part no.
Accessories standard gas sampling probes Part no. 350 mm / 1000 mm Ø 7 mm
Hose extension, length 2.8 m
0554 1202
Pitot tube stainless steel, 350 mm long, for measurement of
Probe shaft with pre-filter, probe shaft length 335 mm, Tmax 0554 2145
0554 8766 flow velocity
probe shaft 1000°C, Ø pre-filter 14 mm
Pitot tube, 1000 mm long, stainless steel, measures flow velocity 0554 2345
Probe shaft with pre-filter, probe shaft length 700 mm, Tmax 0554 8767
probe shaft 1000°C, Ø pre-filter 14 mm
Connection hose, silicon, length 5 m, load to maximum
0554 0440
0554 3372 700 hPa (mbar)
Spare sintered filters (2)
Probe shaft, length 700 mm, Tmax probe shaft 500°C 0554 9767 Straight Pitot tube, stainless steel, length 350 mm for measuring
0653 2041
flow velocity, incl. temperature measurement, 3-fold hose (5 m
Probe shaft, length 335 mm, Tmax probe shaft 1000°C 0554 8764 length) and heat protection shield
Probe shaft, length 700 mm, Tmax probe shaft 1000°C 0554 8765 Straight Pitot tube, stainless steel, length 750 mm for measuring
0653 2042
flow velocity, incl. temperature measurement, 3-fold hose (5 m
0554 3385 length) and heat protection shield
Spare dirt filter for probe handle
Unheated sampling probes +600 °C, stainless steel 1.4571, length 1 m Connection: G1/4“ 0600 7801
1000 mm
Weight: 400 g
Ø 12 mm 0600 7803
Unheated sampling probe up to +1200 °C, Inconel 625, length 1 m Ø 20 mm
Connection: G1/4“
Unheated sampling probe up to +1800 °C, Al oxide, length 1 m 1000 mm
Weight: 400 g 0600 7805
Ø 12 mm
Ø 20 mm
Heating: > +180 °C; power consumption: 650 watts; Connec-
Heated sampling probe, voltage supply 230 V / 50 Hz, stainless steel tion: electr. connection to heated handle, connection adapter 0600 7820
1000 mm
1.4571, heating > 180°C, exhaust gas temperature max. +600 °C with thread connection/screw socket G1/4“; Max. flue gas
Ø 25 mm temp.: +600 °C
1000 mm 0600 7802
Extension shaft up to +600 °C, stainless steel 1.4571, length 1 m Connection: Thread screw/screw socket G1/4“; Weight:
Extension shaft up to +1200 °C, Inconel 625, length 1 m 0.45 kg
Ø 12 mm 0600 7804
Ø 20 mm
Pre-filter for dusty flue gases, ceramic, dust load max. 20g/m³, filter 50 mm Dust load: max. 20 g / m3; filter fineness: 20 µm; Temperature: 0554 0710
pore size 20µm, temperature max. 1000 °C max. 1000 °C; Material: ceramic; Connection: G1/4“ thread
nipple; Weight: 0.2 kg
Ø 23 mm
Thermocouple, NiCr-Ni, -200 to +1200 °C, Inconel 625, 1.2 m long 0430 0065
Connection: To analyser via 4 m connection cable with 8 pin
plug; Weight: 0.15 kg.
Thermocouple, NiCr-Ni, -200 to +1200 °C, Inconel 625, 2.2 m long The length depends on the number of sampling and extension 0430 0066
Ø 4 mm pipes used.
0554 0760
Mounting flange, stainless steel 1.4571, adjustable quick-action fit-
ting, suitable for all sampling7extension pipes 130 Ø 160 mm
mm
Transport case for industrial probes, aluminium profile, space for: handle, probes, flange and accessories, dimensions 1270 x 320 x 140 mm (LxWxH) 0516 7900
11
testo 350
Protection class IP 40 –
Individual dilution with selectable dilution factor (x2, x5, x10, x20, x40)
CO (H2 compen- COlow meas. (H2 NO measure- NOlow measu- SO2 measure- CxHy measurement
sated) compensated) ment rement ment
Meas. range dilution factor- dilution factor- dilution factor- dilution factor- dilution factor- Methane: 100 to 40,000 ppm
dependent dependent dependent dependent dependent Propane: 100 to 21,000 ppm
Butane: 100 to 18,000 ppm
Accuracy ±2 % of m.v. ±2 % of m.v. ±2 % of m.v. ±2 % of m.v. ±2 % of m.v. ±2 % of m.v.
(additional error) (additional error) (additional error) (additional error) (additional error) (additional error)
CO (H2 compen- COlow meas. (H2 NO measure- NOlow measure- SO2 measure- NO2 measure- H2S measure-
sated) compensated) ment ment ment ment ment
Meas. range 2500 to 50000 500 to 2500 ppm 1500 to 20000 300 to 1500 ppm 500 to 25000 500 to 2500 ppm 200 to 1500 ppm
ppm ppm ppm
Accuracy ±5 % of m.v. ±5 % of m.v. ±5 % of m.v. ±5 % of m.v. ±5 % of m.v. ±5 % of m.v. ±5 % of m.v.
(additional error) (additional error) (additional error) (additional error) (additional error) (additional error) (additional error)
Pressure range Pressure range Pressure range Pressure range Pressure range Pressure range Pressure range
-100 to 0 mbar -100 to 0 mbar -100 to 0 mbar -100 to 0 mbar -100 to 0 mbar -100 to 0 mbar -100 to 0 mbar
at probe tip at probe tip at probe tip at probe tip at probe tip at probe tip at probe tip
Resolution 1 ppm 0.1 ppm 1 ppm 0.1 ppm 1 ppm 0.1 ppm 0.1 ppm
12
We measure it.
Accuracy ±0.4 °C (-100 to +200 °C) ±1 °C (0 to +1760 °C) ±0.2 °C (-10 to +50 °C)
± 1 Digit ±1 °C (-200 to +100.1 °C)
±1 °C (+200.1 to +1370 °C)
Resolution 0.1 °C (-200 to +1370 °C) 0.1 °C (0 to +1760 °C) 0.1 °C (-20 to +50 °C)
13
testo 350
We measure it.