Академический Документы
Профессиональный Документы
Культура Документы
SCM2500XRD
SCC3500XRD
Page 2
TABLE OF CONTENTS (Continued)
LIST OF FIGURES
When using this instrument, basic safety precautions shall always be followed to
reduce the risk of fire, electrical shocks and injury to persons, including the
following:
SAFETY PRECAUTIONS
In order to provide maximum user safety this instrument was designed with all electrical
circuitry enclosed within a protective non-conductive housing. The label below will be
visible at any location where high voltage is present.
The treated water sample flows into the sample cell where it is drawn into the bore
during the upstroke of the piston cycle and is expelled from the bore on the piston down
stroke. Particles contained in the water are temporarily immobilized on the piston and
cylinder surfaces. As the water is moved back and forth by the piston, charges
surrounding these particles (+ and -) are moved downstream to the electrodes. This
movement of like charges causes an alternating current to be generated, defined as
"streaming current." A signal gain selector, accessed through the menu screen, is used
to select the best signal amplification. The signal amplification should be set where a
normal change in dosage results in a desired deviation in streaming current (normally 30
units). The displayed streaming current value (SCV) should be considered as a relative
reading due to amplification of the primary signal.
The streaming current amplitude and polarity is a function of sampling location in the
treatment process. It is important to understand why the streaming current varies at
different points in the system to properly interpret the readings, therefore, Section 1.2
should be read very carefully.
The sample must be taken at a point where uniform distribution and mixing of coagulant
is obtained for all flow rate conditions, and at a point that allows for a quick response to
chemical feed changes as measured by the monitor (i.e. after the rapid mixer or static
mixer, and before the slow mixer or flocculation basin). The lag time, or the amount of
time it takes the water to travel from the point of chemical addition to the sensor, should
be no greater than 3 minutes (30 seconds to 1.5 minutes are typical lag times). If
uniform distribution and mixing is not being obtained at a selected sample point, the
streaming current reading will oscillate. If the chemical is not being properly mixed
(indicated by an oscillating monitor reading), the sample point needs to be moved
further downstream away from the mixing device to allow the chemical more time to mix
in with the water. Under rare circumstances, moving the sample point further
downstream in order to obtain a more stable reading may result in a lag time longer than
3 minutes. Steps should be taken to improve chemical mixing under these conditions.
When possible, avoid sampling from places where sludge, grit, etc., will be drawn into
the sensor sample cell. Sample lines must be sized to provide velocities that will
prevent floc/sludge accumulation. Keep sample lines as short as possible to minimize
delay in response time.
Figure 1, Streaming Current Profile, normally observed in a typical water plant. When
coagulant is added to raw water, the charge neutralization process begins immediately.
The time required for this neutralization process to go to completion is primarily a
function of mixing, time, raw water characteristics, type of coagulant, and to a lesser
Based upon the above discussion, the proper sample point for a specific plant depends
upon the following conditions:
1. Point or points of coagulant feed.
2. Mixing efficiency of raw water and coagulant.
3. Magnitude of raw water flow swings.
4. Type and quantity of coagulants used.
In order for the monitor to be used for automatically controlling the coagulant feed
(alum, polymer, ferric chloride), the following guidelines must be followed:
1. The coagulant must be thoroughly mixed with the raw water by the time the
sample is taken.
2. “Lag time” (i.e., the time it takes for the monitor to see a change in coagulant
feed) should be no greater than 5 minutes.
3. Variability in sample flow rates should be no greater than +/- 10 % of the initial
rate.
4. Coagulant pumps must be kept in good mechanical condition to respond quickly
and accurately to process changes.
5. Raw water quality should be in a stable condition (turbidity, pH, color, etc. should
not change rapidly or widely) when setting up the system.
The monitor should be operated a few days in manual control to observe how it
responds to normal process operation. Recording streaming current changes on a strip
chart recorder is very helpful in anticipating factors that may have to be dealt with when
automatic control is started. This period, if using the monitor with a recorder for
monitoring, is also a good time to determine the optimum coagulant dosage and
establish the operating “setpoint”. This enables the system to be put on automatic
control very simply.
SCC3500XRD
Power
Service
Diag.
Alarm
CHEMTRAC
SYSTEMS, INC.
CHEMTRAC CHEMTRAC
SYSTEMS, INC. SYSTEMS, INC.
DURA-TRAC TM
DURA-TRAC TM
DURA-TRAC II SENSOR
CHEMTRAC
SYSTEMS, INC.
OPERATOR
MANUAL
DURA-TRAC TM
II
PATENTED MADE IN U.S.A.
MONITOR/CONTROLLER
Power ...................................... 85-264 VAC, 47–63 Hz
Current Load ........................... 110 VAC @ 0.5A Max
220 VAC @ 0.25A Max
User Interface.......................... Backlit Liquid Crystal Display, Menu Driven Functions,
Keypad Interface
SCV Readout .......................... -1000 to +1000
Resolution ............................... Streaming Current Units
Signal Gain.............................. User Adjustable 1X to 20X
Zero Offset .............................. Automatic, Full Scale All Ranges
Accuracy ................................. 0.5% of Full Scale
Response Time ....................... 1 Second
SCV Output Signal .................. Standard 4-20 mA (300 ohm Load)
Optional 0-10 VDC
Alarm Output ........................... 24 VDC (Solid State Relay Optional) and Auditable
Alarm for Sensor Status, High/Low Streaming Current
Value and High/Low Pump Output - SC3500XRD Only
Digital Input ............................ Dry Contacts
Enclosure Type ....................... NEMA 250 Type 4X, Reinforced Fiberglass
Enclosure Size ........................ 9.2”W, 11.2”H, 6.3”D (234mm W, 285mm H, 161mm D)
Mounting Holes ....................... 11.25”W, 7.4”H (286mm W, 188mm H)
Weight..................................... 6 Pounds (2.72 kg)
Control Output Alarms............. Adjustable High/Low Alarms
Operating Temperature ........... 32° F to 120° F (0° C to 49° C)
The sensor can be located several hundred feet from the monitor. The sensor must be
mounted in a vertical position with the sample flowing into the inlet (3/4” barb fitting) and
exiting from the outlet (PVC elbow). Typically, the sensor is mounted as closely as
possible to the sampling point. Minimizing sample line lengths provides quicker
response to process changes. Sample may be obtained by using a sample pump,
tapping off a pressurized line, or using gravity feed system to get sample to the sensor.
The sample flow rate should not exceed 5.0 GPM for Dura-Trac sensor and 10.0 GPM
for Dura-Trac II sensor. Draining to atmosphere (unobstructed) is required; a
closed pipe (pressurized) drain is not recommended.
The Monitor/Controller module should be installed in a location, which will allow regular
viewing of the display as well as easy access to the front panel menu keys. See
Mounting Dimension Diagrams at end of Manual.
FUSE
FUSE: AGC-1.5A
WARNING TURN POWER OFF BEFORE MAKING ANY WIRE CONNECTIONS 17 18 19 20
OFF TO AVOID DAMAGING THE ELECTRONIC CIRCUIT.
SMM | Sensor
*
*
*
ON * *
L G N
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16
GND
N
L
POWER CORD
High
9 10 11 12
Cond.
Opto (W)
N
Opto (B)
Opto (O)
8
- Probe
L1
7
+ Probe
6
Shield
5
Opto
4
Signal
3
Vdc Gnd
2
+24 Vdc
1
High
9 10 11 12
Cond.
Opto (W)
Opto (B)
Opto (O)
8
- Probe
7
+ Probe
6
Shield
5
Opto
4
Signal
3
Vdc Gnd
2
+24 Vdc
CAUTIONS
R Make sure the monitor’s power switch is turn to off position before making any
signal wiring connections.
FUSE
FUSE: AGC-1.5A
WARNING TURN POWER OFF BEFORE MAKING ANY WIRE CONNECTIONS 17 18 19 20
OFF TO AVOID DAMAGING THE ELECTRONIC CIRCUIT.
1 9 17
SMM | Sensor
2 10 18
3 11
* 19
4
5
12
13
* 20
6 14
ON
7 15 *
8 16 * *
L G N 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16
GND
N
L
(WHITE)
(BLACK)
(BLUE)
(RED)
10 11 12
1 2 3 4 5 6 7 8 9 10
9
Opto (W)
Opto (O)
Opto (B)
+24 Vdc
Vdc Gnd
Signal +
+ Probe
Signal -
- Probe
Shield
8
7
6
(BLACK)
CHEMICAL PUMP
5
(BROWN)
FLUSH VALVE
4
(RED)
SAMPLE ISO VALVE
3
2
(WHITE)
24 VDC COMMON
1
Sample shall be connected to the sample cell inlet side. A 3/4" barbed fitting connection
is provided for hose connection. The PVC elbow opposite the inlet is the outlet. The
outlet sample must flow to an open drain (sample receiver). The cell cannot operate
properly if pressurized. Do not replace either fitting with any other type of fitting.
Start treated water sample through the cell at a rate of between 1 to 5 GPM. Make sure
the outlet is free of obstructions. Continuous reliability of the sample flow is essential.
Interruption of the sample flow will result in erroneous readings on the
Monitor/Controller. For Controller model SCC3500XRD operating in AUTO mode, loss
of flow to the Dura-Trac sensor will result in improper coagulant dosing. A suitable flow
alarm is recommended for maximum reliability.
IMPORTANT
On the SCC3500XRD, the pump control output of the controller must be set in the
MANUAL mode if the sensor is flushed or cleaned or if power to the sensor is
interrupted for any reason. The controller can be placed back in AUTO once power
and sample flow is restored. For system with Dura-Trac with flush with Sensor
Maintenance Module (SMM), during Automatic Flushing cycles, the pump control
output signal is maintained at a constant value (see Flush Menu).
DURA-TRAC TM
SAMPLE OUTLET
SAMPLE INLET
3/4" BARB FITTING
2" Line
DURA-TRAC TM
II
PATENTED MADE IN U.S.A.
SAMPLE OUTLET
45 DEGREE ELBOW
SAMPLE INLET
3/4" BARB FITTING
2" Line
The SCM2500XRD and SCC3500XRD are available with one of two options for
automatic sensor flushing/cleaning. The first option is the Dura-Trac with flush and is
composed of a valve mounted onto the Dura-Trac sensor, which is connected to flush
water. The second option is the SMM and is composed of a flush valve, sample
isolation valve, and chemical injection pump. The Dura-Trac with flush is recommended
for Drinking Water applications where high turbidity is often experienced. The SMM is
recommended for Waste Water applications. These options are only supplied if
specified at the time of purchase. The line of monitors offered by Chemtrac® can be
upgraded to include one of these two options. To upgrade to the Dura-Trac with either
options, the Dura-Trac sensor will need to be returned to the factory.
Potable water or the best quality plant water available shall be used for the flush water
supply. If potable water is used for flush water, a backflow prevention device should be
installed on the flush water line to prevent contamination of the potable water source.
The flush water needs to maintain a pressure of 35 PSI min to 80 PSI max.
Connect the male NPT end of the ¼’’ MNPT to ¼’’ Quick Connect Tube adapter fitting
(supplied with the unit) to the flush water supply. Connect a sufficient length of ¼’’
Teflon tubing between the flush water supply (¼’’ Quick Connect Tube adapter fitting)
and the valve’s ¼” quick connect fitting (see Figure 10 on the next page). The flush
valve’s outlet should already be connected to the probe flush assembly. The third valve
outlet (located on the bottom of the valve and pointed towards the ground) is the drain
outlet. Run this outlet to an appropriate drain using ¼’’ semi-rigid tubing. A manual
control valve should be installed on the flush water line in close proximity of the unit so
CHEMTRAC
SYSTEMS, INC.
DURA-TRAC TM
FLUSH WATER
CONNECTION
1/8" NPT TO 1/4"
COMPRESSION
FLUSH VALVE
25 PSI
CHECK
VALVE
Figure 11 show tubing connections between SMM and Dura-Trac II sensor. The semi-
rigid Teflon tubing is supplied with the unit.
During the automatic flush/cleaning cycle, the chemical pump injects a cleaning solution
into the sensor. This cleaning solution prevents the buildup of fouling agents, which can
overtime affect the sensitivity and response of the instrument. Periodic checks of the
chemical carboy should be made to ensure the chemical storage tank is at least 1/4 full.
CAUTIONS
R Acid concentration shall not exceed 15% concentration.
SENSOR
MAINTENANCE
MODULE
(SMM)
CHEMTRAC
DURA-TRAC II
SYSTEMS, INC.
SENSOR
DURA-TRAC TM
II
PATENTED MADE IN U.S.A.
PINCH VALVE
25 PSI
CHECK CARBOY
VALVE TANK
The Monitor/Controller is equipped with user interface keys, LED indicators, and audible
beeps. The interface keys are up (p), down (q), left (t), right (u), and function keys.
p = UP key
q = DOWN key
There are four LED lights to the left of the graphical display. These are:
q Power –Green LED that illuminates when there is power to the unit.
q Service - Yellow LED displays the status of the electronics.
q Diag. – Green LED that blinks to give a diagnostic status of the remote sensor.
When this LED stops blinking, the remote sensor diagnostic alarm is in alarm
condition.
q Alarm – Red LED that illuminates when the Monitor/Controller is in alarm condition.
The Monitor/Controller provides the user with an audible beep feedback every time a
button is pressed.
The menu navigation of the Monitor/Controller is by using buttons at the front panel.
The following steps explain general menu navigations.
1. From the MAIN screen, pressing any function key will gain access to other
menu screens.
2. At the menu screen, on the far left hand side of the text there is a text cursor (})
that indicates the active line on the menu. By pressing the p or q key, the text
cursor will move to a different menu text line.
3. To modify the parameter setting on the active menu text, first press the u key
to gain access to the parameter and then press the p or q key to scroll
through the parameter range. By holding down the p or q key, the number will
increased or decreased with larger unit changes.
4. When finished making the parameter change, press t key to store the value.
At this point, the new parameter has not been saved. After all parameters
modification are completed, press the SAVE function key to accept the change
or press the MAIN function key to discard the changes.
The menu system for the SCM2500XRD is simple to navigate, below you will find the
different screens accessible through the function keys (SETUP, ALARM, and FLUSH):
The MAIN screen on the SCM2500XRD is the screen used under normal operation of
the instrument. From this screen, all other features and menus can be accessed. The
menu screen shows with Flush option.
v u
w
x
y
Definitions
u Streaming Current Value (SCV) ranges from –1000 to +1000
v Gain, signal amplification factor, adjustable from 1X to 20X
w Next Flushing Cycle (optional): If the unit include sensor flush feature, the
menu display the time when the next flush will occur in days, hours, and minutes.
x Zero Offset Conditions
Offset: OUT Offset condition OUT, actual SCV is display
Offset: IN Offset condition IN, adjusted SCV is displayed and the actual
SCV value will be display to the right of OUT.
Offset Value: -1000 to +1000
The Offset function is used to automatically zero the current reading. Once an
optimum chemical dosage has been obtained, the user can null the reading so
that the setpoint becomes referenced to zero for easier interpretation (see Setup
Screen Menu).
y Function Keys – There are four main menu selections across the bottom for the
four buttons below. The four menus are:
SETUP – SCV Setup Menu
ALARM – Alarms Status
FLUSH – Sensor Flush Control Menu
Pressing one of these keys directly below these menu selections accesses the
functions within that menu.
From the MAIN screen, press SETUP function key, the SETUP screen will display.
Within the SETUP screen menu, the GAIN, OFFSET, and SCV ALARMS settings can
be modified.
u
v
w
x
y
z {
“
Definitions
u Sensor Gain: Signal amplification factor, adjustable from 1X to 20X. The gain
should be adjusted so that a regular change in dosage results in a noticeable
deviation (e.g. 20 to 30 units). To select the proper gain setting let the reading
stabilize for 10 minutes and record the reading. Make a normal dosage change
and then allow another 5 to 10 minutes for the reading to fully stabilize. If the
reading did not change noticeably, simply increase the gain setting and repeat
testing. If the reading changed, but did so by more than 50 units, it is suggested
to decrease the gain setting. Ideal change in Streaming Current for a normal
dosage change is anywhere between 20 to 50 units. If the reading seems
unstable when flow, dosage, NTU, and PH are stable, then check to ensure the
probe is clean and that the sensor is getting a well mixed sample.
v Offset Mode: Offset mode can be changed to IN or OUT. Changing the Offset
mode to from “OUT” to “IN” will automatically zero or null the current Streaming
Current reading. For example, if the current reading was –150 and the Zero
Offset was changed to “IN”, then the adjusted Streaming Current value is
displayed on the Main Screen would be zero. The reading will not stay exactly at
zero since the reading is always slightly changing as a result of slight changes in
charge. The Offset value will display at the line below. On the Main Display, the
raw Streaming Current value (with offset) will be displayed to the right of the
Zero Offset: IN. The different between the Streaming Current value and the raw
is the offset value. To disengage the Offset, simply change the Offset mode to
“OUT” and the true Streaming Current Reading will be displayed.
w Offset Value: The Offset value indicates the amount of offset that has been
applied to the reading in order to offset the reading to zero. So that, at any time
the operator can see what the offset value is and if anyone has changed it from a
previous offset. To disengage the offset, simply change the Offset Mode to
“OUT” and the true Streaming Current Reading will be displayed. This is also
the method used to re-zero the reading if the current offset no longer
corresponds to optimum coagulant feed. To re-zero the reading when Offset
mode is “IN”, increased or decreased the Offset Value then press Enter to
SAVE.
From the SETUP MENU (1), continue pressing the down arrow key until SETUP MENU
(2) menu screen display. SETUP MENU (2) menu screen is a continuation of the
SETUP MENU (1) menu screen.
u v
Definitions
u SCV Filter: The filter function activate rolling average of the Streaming Current
value over time. The SCV filter can be set to OFF, 30, 60, 90, or 120 seconds.
The default setting is OFF.
From the MAIN screen, press ALARM function key, the ALARM STATUS screen will
display. Within the ALARM STATUS menu you can view which Alarm has been
triggered, as well as, turn the Monitor/Controller’s audible alarm feature on and off.
z
u
v {
w
x
y
Definitions:
When any alarm is triggered, a +24 Vdc signal will be present on the Monitor’s digital
alarm output terminals. This signal can be connected to a SCADA/DCS system to alert
the operators to an Alarm condition. In lieu of a +24VDC signal, an optional solid state
relay output is available (must be specified with order).
After being properly adjusted by the operator to match up with the upper and lower SCV
readings that typically correlate to the acceptable window of process performance, the
Hi and Lo Alarms alert the operator as to whether the SCV has exceeded acceptable
limits. A Lo Alarm signify an underfeed of chemical (even possible loss of chemical
feed) and a Hi Alarm signify an overfeed.
This flush feature is an option that is available for the SCM2500XRD, see Section 3.3.7
for detail description.
The menu system for the SCC3500XRD is simple to navigate; below you will find the
different screens accessible through the function keys (SETUP, ALARM, FLUSH, and
CTRL). The menu screen shows with Flush option.
u
u
v
w {
{
x
y
z
Definitions
u Streaming Current Value (SCV) ranges from –1000 to +1000
v Gain, signal amplification factor, adjustable from 1X to 20X
w Next Flushing Cycle (optional): If the unit include sensor flush feature, the
menu display the time when the next flush will occur in days, hours, and minutes.
x Zero Offset Conditions
Offset: OUT Offset condition OUT, actual SCV is display
Offset: IN Offset condition IN, adjusted SCV is displayed and the actual
SCV value will be display to the right of OUT.
Offset Value: -1000 to +1000
From the MAIN screen, press SETUP function key, the SETUP screen will display.
Within the SETUP screen menu, the GAIN, OFFSET, SCV ALARMS, and SCV Filter
settings can be modified. The SETUP Menus are the same for SCM2500XRD and
SCC3500XRD see Sections 3.2.2 and 3.2.3 for detail descriptions.
From the MAIN screen, press ALARM function key, the ALARM STATUS screen will
display. Within the ALARM STATUS menu you can view which Alarm has been
triggered, as well as, turn the Monitor/Controller’s audible alarm feature on and off.
u |
v }
}
w
x
y
z
{
Definitions:
When any alarm is triggered, a +24 Vdc signal will be present on the Monitor’s
alarm terminals 13 (+) and 14 (-). This signal can be connected to a
SCADA/DCS system to alert the operators to an Alarm condition. In lieu of a
+24VDC signal, an optional solid-state relay output is available (must be
specified with order).
After being properly adjusted by the operator to match up with the upper and
lower Streaming Current readings that typically correlate to the acceptable
window of process performance, the Streaming Current Value (SCV) Hi and Lo
Alarms alert the operator as to whether the streaming current value has
exceeded acceptable limits. A Lo Alarm will signify an underfeed of chemical
(even possible loss of chemical feed) and a Hi Alarm will signify an overfeed.
From the MAIN screen, press CTRL function key, the PUMP CONTROL (1) Menu
screen will display. There are PUMP CONTROL (1), PUMP CONTROL (2), and FLOW
CONTROL (3) menus under CTRL function key. Use up or down arrow key to navigate
between menus. Within the PUMP CONTROL (1) Menu screen, the PID tuning
parameters can be adjusted.
u
v
w
x |
y
z
{
Definitions
u Pump Ctrl: The Controller’s pump output control has two mode of operation;
AUTO or Manual. In AUTO mode, the pump output percent will increased or
decreased automatically in response to deviations from the setpoint. In MAN
mode, the pump output percent can be increased or decreased manually by
setting the Man Output. The two mode can be toggle by pressing the p or q
key.
v Man Output: In MAN mode, the pump output percent % can be increased or
decreased by pressing the p or q key. The Man Output value can be changed
from 0 to 100% at 0.1% increment. This parameter is saved in non-volatile
memory. When running in Automatic mode, the Man Output setting is ignored.
However, when switching from AUTO to MAN, it is important to understand that
the Pump Output value will be changing from the value it was in MAN mode.
From the PUMP CONTROL (1) menu screen, continue pressing the down arrow key
until PUMP CONTROL (2) menu screen display. PUMP CONTROL (2) menu screen is
a continuation of the Pump Control (1) menu screen.
u
v
w z
x
y
Definitions
u Pump Out Min: The pump output limit define the minimum value of the pump
control output signal. The minimum percent (%) value can be increased or
decreased by pressing the p or q key. The minimum percent (%) value can be
changed from 0 to 100% at 0.1% increment.
v Pump Out Max: The pump output limit define the maximum value of the pump
control output signal. The maximum percent (%) value can be increased or
decreased by pressing the p or q key. The maximum percent (%) value can be
changed from 0 to 100% at 0.1% increment.
w Output Lo: The pump output limit define the low alarm value of the pump
control output signal. The low alarm percent (%) value can be increased or
decreased by pressing the p or q key. The low alarm percent (%) value can be
changed from 0 to 100% at 0.1% increment. When the pump control output
signal fall below the low alarm value, alarm condition will occur.
x Output Hi: The pump output limit define the high alarm value of the pump
control output signal. The low alarm percent (%) value can be increased or
decreased by pressing the p or q key. The high alarm percent (%) value can
be changed from 0 to 100% at 0.1% increment. When the pump control output
signal rise above the high alarm value, alarm condition will occur.
y MAIN Function Key: Pressing this key will ignore any setting modifications on
this screen and brings up the MAIN screen.
z SAVE Function Key: Pressing this key will save current setting modifications
and brings up the MAIN screen.
From the PUMP CONTROL (2) menu screen, continue pressing the down arrow key
until FLOW CONTROL (3) menu screen display. FLOW CONTROL (3) menu screen is
a continuation of the Pump Control (2) menu screen.
When ENABLED, the FLOW CONTROL function allows the user to control the chemical
feed based on the process FLOW rate and the Streaming Current Value PID control
u
v
w
x |
y
z
{
Definitions
u Flow Ctrl: The flow control modes are ENABLED or DISABLED. When the
flow control is ENABLED, pump output percent % will increased or decreased
automatically in response to process flow changes and SCV deviation from the
setpoint. When the flow control is DISABLED, the pump output percent will
increased or decreased automatically in response PID loop control only. By
pressing either p or q key to switch between ENABLED or DISABLED control
mode.
v SCV Control Band: When flow control is ENABLED, the proportional pump
output flow control will be adjusted (up and down) by the PID loop control. SCV
control band value (percent) set up how much the pump output control will be
effected by the PID loop control. For example, the SCV control band is set to
5% and the proportional pump output flow control is at 55%, the PID loop control
can adjust the pump output control up to 60% maximum and down to 50%
minimum. The SCV control band percent (%) value can be increased or
decreased by pressing the p or q key. The SCV control band percent (%)
value can be changed from 0 to 100% at 0.1% increment.
w Flow Min: Minimum process flow in percent set for the produce the minimum
pump control output for the parameter below (Pump Output Min). The percent
(%) value can be increased or decreased by pressing the p or q key. The
percent (%) value can be changed from 0 to 100% at 0.1% increment.
x Pump Output Min: Pump control output for the minimum process flow set
above (Flow Min). The percent (%) value can be increased or decreased by
pressing the p or q key. The percent (%) value can be changed from 0 to
100% at 0.1% increment.
y Flow Max: Maximum process flow in percent set to produce the maximum
pump control output for the parameter below (Pump Output Max). The percent
(%) value can be increased or decreased by pressing the p or q key. The
percent (%) value can be changed from 0 to 100% at 0.1% increment.
z Pump Output Max: Pump control output for the maximum process flow set
above (Flow Max). The percent (%) value can be increased or decreased by
pressing the p or q key. The percent (%) value can be changed from 0 to
100% at 0.1% increment.
{ MAIN Function Key: Pressing this key will ignore any setting modifications on
this screen and brings up the MAIN screen.
From the MAIN screen, press FLUSH function key, the FLUSH CTRL MENU screen will
display. The goal of the sensor flush is to minimize the need for operator involvement
for maintenance purposes. This feature greatly enhances the reliable operation of the
Streaming Current Monitor. Within the FLUSH CTRL MENU screen, the flush
parameter settings can be modified.
Sensor Flush is only available with the optional Sensor Maintenance Module or the
Dura-Trac Flush feature.
u
v
w
x }
y
z
{ |
Definitions
OFF
ON
SAMPLE ISOLATION
OFF
ON
SENSOR FLUSH FLUSH RINSE
AND RINSE
OFF
ON
CHEMICAL PUMP WASH PERIOD
OFF
TIME (SECONDS)
NOTES
R Sample Isolation and Chemical Wash is only available with the optional SMM.
The Dura-Trac with flush option does not include the sample isolation or chemical
pump feature; however, there will still be a delay between sensor flush and rinse.
After sample flow is established and power has been applied to the sensor and
streaming current monitor, allow 10-15 minutes or longer for the reading to stabilize.
Stable conditions differ from plant to plant. The reading may fluctuate 10 to 20 units,
even under stable conditions due to the chemical not being thoroughly mixed. If
readings fluctuate widely, refer to Troubleshooting Guide. The monitor’s reading may
be negative, zero, or positive. The monitor is simply indicating the streaming current
value of the sample. In most applications, the reading will be slightly negative (anionic).
After the reading has had ample time to stabilize, adjust the chemical dosage in normal
increments to test the response of the monitor. For model SCC3500XRD, enter the
Control menu to manually adjust chemical feed. For each dosage change, the reading
should respond by changing 20 to 50 SCV units. If the response is less than 20 units,
increase the gain. Likewise, if the response is greater than 50 units, it is recommended
to decrease the gain (especially if the monitor is to be used for automatic chemical feed
control).
Next, a test should be performed to determine the Raw SCV reading. This will help
determine the LO SCV Alarm setting and give a better understanding of instrument
response. To determine the Raw SCV reading, shut off sample flow to the sensor and
turn the outlet fitting up. Pour a sample of raw untreated water into the sensor and allow
the reading to stabilize and then record the reading. This reading should serve as an
absolute alarm value, which if measured by the monitor will indicate a loss of chemical
feed. It is recommended to set the Lo SCV Alarm value (see Alarm setup menu)
between the Raw SCV measurement and the setpoint value. Further testing will help to
After verifying proper response, the next step should be to optimize the process to
establish a SCV setpoint. The SCV setpoint can then be zero’d, or offset, so that the
setpoint will be adjusted to a reading of zero (0). This is done by entering the Setup
menu and changing the Offset mode to “IN”. Offsetting the SCV is not necessary for
normal monitor operation as it is only provided so that an easier reference of zero can
be used for monitoring changes in the process. For example, if the true SCV setpoint
reading was –150, and the user offset that reading to equal zero (0), if chemical dosage
was decreased the reading may go from 0 to –50. If the reading had not been zero’d,
then the true SCV would have gone from –150 to –200 (a –50 unit change). Once the
setpoint has been zero’d, it is easier to determine whether chemical is being under or
overdosed since going negative (anionic) from 0 would be an under-dose and going
positive (cationic) would be an overdose. A setpoint of zero (0) is also easier for
multiple operators (night and day shift) to remember. Since the setpoint will change
seasonally, adjusting the setpoint so that it equals zero (0) helps to avoid confusion.
A common mistake made with the monitor’s offset function is that it has been used to re-
zero the reading as it drifts from the previously offset reading of zero. Once the process
has been optimized and the Streaming Current Value has been zero’d at that point, it is
no longer necessary or correct to re-zero the reading. If the reading changes from the
setpoint of zero (or the true SCV setpoint if the offset is out), the chemical dosage
should be adjusted to bring the reading back to the setpoint and not the offset function.
Eventually, a new setpoint might need to be established due to seasonal changes or
large excursions in turbidity, but under normal conditions, the setpoint should not be
changed on a daily or even weekly basis. It is recommended to use the monitor for a
few weeks in the Offset Out mode to become familiar with the operation and response
of the unit.
The treatment optimization process should be done slowly and stepwise. Assuming that
the plant is producing acceptable water with present chemical dosages, trim
approximately 10%, wait for the reading to stabilize, and record the reading as the
setpoint. You may need to change GAIN setting to increase magnitude of response. If
the settled and finished water quality is still acceptable at the reduced dosages, trim
another increment, wait for the reading to stabilize, and record the reading as the new
setpoint. Continue this process, being sure to wait long enough each time to see the full
effect of the chemical feed changes on the process (i.e. settled water turbidity,
appearance of floc). In between these dosage changes it will be necessary to maintain
the setpoint. Failure to maintain the setpoint while trying to reach a lower dosage level
could result in under dosing the chemical if turbidity starts to increase. By maintaining
the setpoint, the chemical dosage is kept in check with process changes (i.e. turbidity,
flow rates).
If the Offset function is being used, re-zero the monitor reading as described in section
3.3 after each reduction in chemical feed. If a cationic coagulant is being used, the
streaming current value will become more negative with each reduction in dosage. The
"optimum setpoint" is obtained when a minimum dosage of coagulant is being fed and
produces desired results for the particular treatment process. This setpoint will remain
very close to the same reading even when raw water turbidities increase or decrease.
Simply adjust the coagulant dosage to maintain this setpoint reading on the monitor.
Make electrical connections between Streaming Current Controller and the chemical
feed pump. The AUTO or Manual control mode is indicated on the MAIN screen. To
switch between AUTO or Manual control mode can be done by pressing both p and q
keys from the MAIN screen. Once in Manual control mode, the pump control output
percentage can be increased or decreased by pressing the p or q key.
The AUTO or Manual control mode can also be change from the Pump Control (1)
menu. Once Pump Control (1) menu is display, move the text cursor on the far left hand
side of the screen (}) to “Pump Ctrl:” then press u the control text will blink. The
control mode can be set by pressing the pressing p or q key, this will toggle between
AUTO and MAN mode. Set the controller to MAN. Once completed, press t key then
press p or q key to get to next parameters on that screen. Next press q key to “Man
Output:” then press u the control number will blink. The percent (%) number will blink,
the manual chemical feed control value output can now be increased or decreased the
by pressing p or q key to the desire manual control value. When completed, press t
key to get to other parameters on that screen. The SAVE Function Key must be
pressed to save current setting modifications and brings up the MAIN screen.
Automatically controlling the treatment process with the Streaming Current Controller
(SCC3500XRD) is a very simple operation, once you know how the controller works.
Most controller manuals do not explain, in simple terms, the purpose of so-called "tuning
constants" termed Proportional Band, Integral Time, and Rate. Controllers with these
three functions are sometimes called "3-mode controllers," or "PID controllers". In
order to explain the purpose of these functions, a brief review of how a simple
proportional controller works may be helpful.
Many water treatment plants have "flow controlled" coagulant feed, i.e., when raw water
flow increases, a signal is sent to the coagulant pump that immediately increases the
feed rate. In other words, there is no "lag time" between water flow increasing and
coagulant increasing. This controller works extremely well in maintaining proper dosage
as long as the water flow is the only thing changing. However, if turbidity is changing,
the proper chemical dosage cannot be maintained with a simple proportional controller.
Since the SCV is taking its sample downstream of the coagulant feed point, there will be
some "lag time" before it can sense a change in turbidity, alum feed, raw water flow, etc.
If, for instance, the SCV reading should start going more negative due to an increase in
turbidity, it will cause the controller to increase coagulant feed. This will cause the
streaming current reading to go more positive, which will cause the pump to decrease
coagulant. The result is continuous cycling above and below the setpoint, thus, the
The following are description of parameters used in the PID control loop algorithm:
q Process Variable: The controlled process, meaning the value at the Controller
used to look at the process input (SCV).
q Setpoint: The target for the process variable, meaning the value at the setpoint
input.
q PID Output: The correction signal produced by the PID loop algorithm. The
output is used to control the process that effect the process variable.
q Proportional Band: The value of the property proportional constant (KP).
KP = Output range / effected process range.
q Integral Time: The value of the Integral constant (KI) expressed in repeats per
minute.
KI = Component of the loop output reacts to the amount setpoint error.
q PID Rate: The value is a function design to improve the PID loop response.
q Maximum Output: The maximum value produced by the PID loop output.
q Minimum Output: The minimum value produced by the PID loop output.
The Controller’s PID loop is designed for reverse acting; process variable goes down will
cause the output to go up. To set the loop to be direct acting, the minimum and
maximum values need to be swapped; set minimum at maximum value and set
maximum at minimum value.
Integral Time
A basic rule for setting the Integral Time constant setting is that the longer the lag time,
the higher integral time constant should be. Unlike Proportional Band which actually
responds to the changes in the SCV deviation, Integral Time only responds to the time
the deviation last and will continue to correct the pump output for as long as the SCV
stays away from the setpoint value (even if the deviation is not changing). A good
example of this would be to place the controller into manual while the SCV is at the
setpoint value and then decrease the Pump Output by 10%. Once the reading has
completely stabilized at a more negative value, place the controller into Automatic mode.
Since the reading will be stable at first, but away from the setpoint, the amount of Pump
Output change you note on the display is being controlled by the Integral Time constant
setting. Proportional Band will have no effect on the Pump Output until the SCV starts
to change as a result of the higher dosage of chemical finally reaching the sensor.
During the time when the SCV is stable (but not at the setpoint), you can note how much
the Pump Output is changing and if it appears to be making too small or too large of an
adjustment, simply change the Integral Time constant setting accordingly.
PID Rate
The PID Rate function serves to improve the controllers response when higher
Proportional Band and Integral Time settings appear to still be too responsive for the
application. The PID Rate setting is usually left at 1 for most applications since
proper setting of P Gain and Integral Time are capable of providing proper response.
Higher PID F settings will serve to buffer the responsiveness of the Controller. The PID
Rate setting should not exceed 10.
NOTE
R There will normally be a small amount of cycling for a few minutes after a process
change, even if Proportional Band, Integral Time, and PID Rate are set correctly.
The AUTO or Manual control mode is indicated on the MAIN screen. To switch between
AUTO or Manual control mode can be done by pressing both p and q keys from the
MAIN screen. When in AUTO control mode, the pump output percent will increase or
decrease automatically in response to deviations from the setpoint.
Flow control is a feature that allows the Controller to perform coagulant pump control in
the process that is varying their process flow. During large process flow variation, the
PID control alone may not be able to response to the flow changes. The Controller
monitors the process flow and uses it in the flow control calculation. The Controller is
equipped with auxiliary analog input (4-20mA). When the flow control is enabled, the
pump control output is proportionally changed by the amount of process flow changes in
conjunction with the PID loop based on the SCV setpoint. The following are parameters
used in the flow control:
q Flow Ctrl: The flow control modes are ENABLED or DISABLED. When the flow
control is ENABLED, pump output percent % will increase or decrease
automatically in response to process flow changes and SCV deviation from the
setpoint. When the flow control is DISABLED, the pump output percent will
increase or decrease automatically in response PID Loop control only.
q SCV Control Band: Percent band around the flow control proportional pump
output, based on the four parameters below, that the PID loop can effect.
q Flow Min: Minimum process flow in percent set for the produce the minimum
pump control output for the parameter below (Pump Output Min).
q Pump Output Min: Pump control output for the minimum process flow set above
(Flow Min).
q Flow Max: Maximum process flow in percent set for the produce the maximum
pump control output for the parameter below (Pump Output Max).
q Pump Output Max: Pump control output for the maximum process flow set
above (Flow Max).
Tuning constants are stored digitally in non-volatile memory and can be changed by
front panel keys. Table below shows default PID Parameters preset by Chemtrac®.
Experience has shown these values to be a good starting point for most applications.
However, significant changes may be necessary to tune the controller to your specific
operation.
Caused by:
R Decrease in pH
R Decrease in raw H20 flow
R Decrease in color
R Decrease in raw water turbidity
R Decrease in lime
R Decrease in caustic
R Decrease in anionic polymers Increase in aluminum sulfate, or PAC
R Increase in cationic polymers Increase in chlorine
Caused by:
R Increase in pH
R Increase in H20 flow
R Increase in color
R Increase in raw water turbidity
R Increase in lime
R Increase in caustic
R Increase in anionic polymers
R Decrease in aluminum sulfate, or PAC
R Decrease in cationic polymers
R Decrease in chlorine
NOTES
R Potassium Permanganate usually has no appreciable effect on the SCV (at
normal dosages of 1-2ppm).
R Increasing conductivity will drive the SCV toward zero.
R Change in lag time may affect the SCV.
Introduction
It will take several weeks or maybe even a few months for an operator who has
come to rely on Jar test, eye-balling, turbidity readings, or any other method for
determining coagulant dosage to become comfortable using the monitor as a
tool for controlling coagulant feed. Very few plants install an monitor and start
controlling in Automatic mode in the first few weeks. Monitoring the response
and repeatability of the monitor through several different turbidity excursions,
and comparing process results obtained by controlling dosage to maintain a
streaming current setpoint against the results of previous methods used to
determine dosage changes is the only way to prove the reliability of the
instrument and gain the confidence needed to realize the full potential of the
technology. The operator will soon realize that the time consuming jar testing
procedure is no longer where his comfort lies. The information provided by the
monitor is equivalent to a 5 second jar test once the operator understands how
to interpret the reading.
Maintenance Issues
To keep the monitor working properly, occasional maintenance will be
necessary. Cleaning only takes a few minutes to perform and is usually needed
once every 2 to 6 weeks (depending upon water quality conditions). An easy
test to see if the sensor needs to be cleaned is to first verify the reading has not
become “noisy” or “bouncy” and then make small changes in chemical feed to
make sure the sensor has not become unresponsive. As long as the monitor is
stable, responsive, and the response is no less than it has been noted in
previous response testing, the sensor does not need to be cleaned. If the
reading does appear to be less responsive or noisy, then record the amount of
response or noise (signal drift as measured in a one minute period) and then
remove the probe and piston for cleaning. First look to see if there is any
obvious buildup or solids in the probe or on the piston. Then, clean the probe
with a suitable cleaner (such as Comet/Ajax or Rover/CLR). Finally, rinse the
parts and install them back into the sensor. After allowing the sensor time to
stabilize, compare response and noise (if any) to previous results to verify the
cleaning did indeed improve response. If not, the piston and/or probe may need
to be replaced.
Eventually the moving parts of the monitor will wear. Most notably, the probe
and piston will wear the fastest and require replacement once every two years on
average. Once again, to determine if the probe and piston need to be replaced
simply make small chemical feed adjustments. If the probe is unresponsive and
cleaning does not improve the result, the probe and piston most likely need to be
replaced. Having a new set on hand to compare against is the best method of
determining if the parts need to be replaced. The motor in the monitor sensor
will also need replacing after 3 to 5 years of use. If the reading starts to become
slightly unstable, and cleaning or replacing the probe and piston does not
eliminate the unstable reading, check the shaft of the motor to see if it has
excessive movement. If so, replace the motor.
It may not be necessary to replace both the piston and the probe at the same
time as usually the piston wears faster than the probe. To check this, simply
install a new piston with the old probe and compare those readings to the ones
obtained with the new piston and new probe combination. If the old probe and
new piston combination produce a reading that is within 50% or more of the new
set, then the probe can still be used as long as the new piston is installed. The
same procedure should be followed with a new probe and old piston to
determine if the old piston is still suitable for use. Using this procedure, the
operator can determine which part, if not both, needs to be replaced. Be sure to
order a replacement part(s) to have on hand as a spare once you have installed
the new probe and/or piston to replace the worn part(s).
2. Display does not A. GAIN setting too low. A. Increase GAIN setting.
change with
B. Wrong sample point or B. Select correct sample point.
change in
insufficient mixing..
coagulant dosage. C. Clean cell (see Cleaning
C. Sample cell dirty. Procedures).
D. Excessive “lag time” D. Move sampling point closer to
between coagulant coagulant feed point and/or
injection point and sensor decrease sample TRANSPORT
sample cell. time (See Selecting Proper
Sample Point).
The Dura-Trac sensor is supplied with two (2) probes and (2) pistons so freshly cleaned parts
can always be on hand.
CAUTIONS
R If monitor is being used for automatic dosage control, put the controller into
MANUAL mode before starting replacement procedure.
Depending on the type of contamination, see table below for different cleaning
procedures recommended.
A “soak and brush” method is recommended for any contaminant. Soak only long
enough to dissolve/remove the material. Probe surfaces should be thoroughly washed
with clean water to remove any residual cleaning compound.
CAUTIONS
R Do not soak entire probe cartridge, only soak inside the bore. Acid concentration
shall not exceed 15% concentration.
Figure 12 AND 13 illustrates Dura-Trac and Dura-Trac II sensor parts identification with
part numbers. Table below lists recommended replacement parts.
WASHER
(1608)
VITON SEAL
110 CRANK (7701)
ASSEMBLY
(5302) MOUNTING NUT
(1607)
NEOPRENE
110 MOTOR
WASHER
(1301)
(5016)
OPTO
SWITCH 4-40 SET
(1308) SCREW
MOTOR BRACKET
ROD END (3311)
(3707)
DURA NUT
YOKE PIN (4200)
(5720)
EXTENSION
(5611)
YOKE
(5710)
GUIDE
(8711)
PISTON
(1701)
PROBE
WASHER
(7703)
PROBE
(8601)
DURA NUT
(4200)
O-RING
(1605)
DURA PLUG
(8606)
NEOPRENE WASHER
(5016) ROD END
110V MOTOR
OPTO (3707)
(5470)
SWITCH
4-40 SET (1308)
SCREWS
MOTOR YOKE PIN
BRACKET (5720)
(3311)
YOKE
(5710)
DURA NUT
(4200)
WW EXTENSION
(5611)
WW PROBE BLOCK
(5630)
GUIDE
(7720)
PISTON
(1710)
PROBE
WASHER
(7703)
PROBE
(8601)
DURA NUT
(4200)
O-RING
(1605)
DURA PLUG
(8606)
To place an order for spare parts you may either call, email, or fax Chemtrac® directly or
contact your local distributor or representative. The following information should be
included in your request: Model number of your monitor, part number(s), qty,
description of parts required, and purchase order number. Please contact Chemtrac®
for up to date pricing.
The following Figures 14 thru 19 illustrate enclosure and mounting dimensions for Streaming
Current Monitor/Controller, Dura-Trac Sensor, Dura-Trac Sensor with Flush, Dura-Trac II
Sensor, and Sensor Maintenance Module.
SCC3500XRD
13.6" [345mm]
11.2" [285mm]
12.5" [317mm]
Power
Service
Diag.
Alarm
CHEMTRAC
SYSTEMS, INC.
7.71" [196mm]
CHEMTRAC
SYSTEMS, INC.
8.5" [215mm]
7.2" [183mm]
DURA-TRAC TM
14.7" [373mm]
PATENTED MADE IN U.S.A.
11.2" [285mm]
9.7" [247mm] 6.3" [161mm]
CHEMTRAC
SYSTEMS, INC.
10.5" [266mm]
9.2" [234mm]
18.8" [477mm]
DURA-TRAC TM
II
PATENTED MADE IN U.S.A.
CHEMTRAC
SYSTEMS, INC.
8.5" [215mm]
7.2" [183mm]
DURA-TRAC TM
17.6" [447mm]
PATENTED MADE IN U.S.A.
CHEMTRAC
SYSTEMS, INC.
10.5" [266mm]
9.2" [234mm]
DURA-TRAC TM
II
PATENTED MADE IN U.S.A.
21.8" [554mm]
FIGURE 18. Dura-Trac II Sensor with SMM Dimensions
11.2" [285mm]
9.7" [247mm] 6.3" [161mm]
10.5" [266mm]
9.2" [234mm]
On specific model of the SCM2500XRD, the analog output (4-20mA) representing the Offset
value is provided at terminals 11 and 12; see figure below. Terminal 11 is labeled Offset Value
(AO+) and terminal 12 is labeled Offset Value (AO-). The scaling for the Offset value analog
output is from -1000 to +1000 streaming current value. The two analog outputs are provided to
allow the customer to monitor the SCV Signal and Offset value remotely. The explanation of
the Offset mode and Offset value are described in Section 3.2.2 SCM2500XRD Setup Menu
(1).
FUSE
FUSE: AGC-1.5A
WARNING TURN POWER OFF BEFORE MAKING ANY WIRE CONNECTIONS
17 18 19 20
OFF TO AVOID DAMAGING THE ELECTRONIC CIRCUIT.
1 9 17
SMM | Sensor
2 10 18
3 11
* 19
4
5
12
13
* 20
6 14
7 15 *
ON 8 16 * *
L G N 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16
GND
N
L