TC 220
Automatic Chemistry Analyzer
Technician Manual
Ver. 1.15
Tecom Science Corporation
Tel.: 0086-791-88116898
Fax: 0086-791-88111989
Web: www.tecom-cn.com
Email:service@tecom-cn.com
1
TC220 Chemistry Analyzer Technician Manual(Ver1.15) YeMengRui
CONTENT
Chapter 1 Analyzer Introduction ···········································································································4
1.1 System Structure······························································································································4
1.1.1Appearance pic of Analyzer ···································································································4
1.1.2 Composition of Analyzer ·······································································································6
1.1.2.1 Sample/Reagent Tray Assembly ·················································································6
1.1.2.2 Reation Tray Assembly and Temperature Control System·····································10
1.1.2.3 Photometric System ··································································································12
Chapter 2 Installation ························································································································12
2.1 Unpacking ·······································································································································12
2.2 Installation Environment Requirements ·······················································································13
2.2.1 Installation Environment Requirements ·············································································13
2.2.2 Power Requirements············································································································14
2.2.3 Temperature and Humidity Requirements ·········································································15
2.2.4 Space Requirements ············································································································15
2.3 Installation Process························································································································15
Chapter 3 Fluidic System·····················································································································28
3.1 TC220 Analyzer Fluidic System·····································································································28
3.1.1 Fluidic Diagram ····················································································································28
3.1.2 View Pictures of Analyzer Parts ··························································································29
3.1.3 Fluidic system steps instruction·························································································31
Chapter 4 Hardware Structure·············································································································33
4.1 Sample /reagent tray ······················································································································33
4.2 Reaction tray and drive assembly ································································································33
4.3 Optical Sensor ································································································································35
4.4 Sample/Reagent probe and drive assembly ················································································36
4.5 Mixer probe and drive assembly ···································································································37
4.6 Sample/reagent syringe and drive assembly···············································································38
4.7 7-steps washing unit and drive assembly ····················································································38
4.8 Photometric System·······················································································································39
4.9 Temperature control process of Reaction Tray ···········································································41
4.10 Water Temperature Control process of water tank····································································43
4.11 System Power Supply ··················································································································44
4.12 Circuit board function description······························································································46
4.12.1 Main controller board function··························································································46
4.12.2 Bus board····························································································································48
4.12.3 Signal Process Board ········································································································49
4.12.4 Liquid Level Detection Board····························································································50
4.12.5 Power Board ·······················································································································52
4.12.6 Water Controller Board ······································································································53
4.12.7 Reagent Refrigeration Pinboard ·······················································································54
4.12.8 M7 Motor Subdivision Board·····························································································55
4.12.9 Stepping Motor Drive Board······························································································56
4.12.10 Mixer Motor Power Board ································································································58
Chapter 5 Component Adjustment and Replacement·······································································59
2
TC220 Chemistry Analyzer Technician Manual(Ver1.15) YeMengRui
5.1 Signal process board voltage adjustment····················································································59
5.2 Reaction tray position adjustment································································································60
5.3 Tungsten-Halogen Lamp replacement··························································································61
5.4 Sample/reagent probe position adjustment·················································································63
5.4.1 Wash well position for sample/reagent probe ···································································64
5.4.2 Cuvette position for sample/reagent probe ·······································································65
5.4.3 Sample position for sample/reagent probe········································································66
5.4.4 Reagent position for sample/reagent probe ······································································67
5.5 Sample/reagent probe replacement······························································································69
5.6 Mixer probe position adjustment ··································································································70
5.6.1 Wash well position for mixer probe ····················································································71
5.6.2 Cuvette position for mixer probe ························································································72
5.7 Mixer probe replacement ···············································································································73
5.8 Synchronous belt replacement ·····································································································75
5.9 Light filter replacement··················································································································75
Chapter 6 Maintenance ························································································································80
6.1 Daily Maintenance ··························································································································80
6.1.1 Check the distilled water volume of water bucket·····························································80
6.1.2 Check waste solution···········································································································80
6.1.3 Check the signal value of all cuvettes················································································81
6.1.4 Check sample/reagent probe clogging ··············································································81
6.2 Weekly Maintenance ······················································································································82
6.2.1 Clean sample/reagent probe and mixer probe···································································83
6.2.2 Clean two wash well·············································································································83
6.2.3 Clean sample and reagent tray ···························································································84
6.2.4 Clean the reaction tray·········································································································84
6.2.5 Clean 7-steps washing needles ··························································································84
6.2.6 Clean cuvettes using wash solution···················································································84
6.2.7 Check the signal value of all cuvettes.···············································································85
6.3 Monthly Maintenance ·····················································································································86
6.3.1 Clean incubation groove of reaction tray···········································································86
6.3.2 Clean the panel of machine ·································································································87
6.3.3 Clean distilled water bucket ································································································87
6.3.4 Clean waste solution bucket ·······························································································88
6.3.5 Clean water filter ··················································································································89
6.3.6 Clean dust screen ················································································································89
Appendix A Spare Parts Price List ······································································································91
Appendix B Preventive Maintenance Kit ····························································································97
Appendix C Error Code Message········································································································98
Appendix D Diagram ····························································································································99
3
TC220 Chemistry Analyzer Technician Manual(Ver1.15) YeMengRui
Upper cover
(Front View)
(Left View)
4
TC220 Chemistry Analyzer Technician Manual(Ver1.15) YeMengRui
(Right View)
Air outlet
Air inlet
(Rear View)
Sample/reagent
probe assembly Mixer probe assembly
Reaction tray
Reagent tray
TC220 main unit is mainly consists of sample/reagent tray assembly 、 reaction tray
Reagent
position
Sample
position
6
TC220 Chemistry Analyzer Technician Manual(Ver1.15) YeMengRui
dispenser system and related mechanical parts ,its function is made sample and reagent to
Sample/Reagent Tray
● Specification
Sample tray
Reagent tray has 25 reagent positions,The reagent bottle volume is 20ml.NO.1 reagent
position is defined by software for distilled water bottle.it is used for predilution or
post-dilution.
Reagent tray
When you load reagent bottle,it must be attention to make bottom of bottle to be on a level
with bottom of sample/reagent tray ,don’t appear the different height.The reagent bottle
volume is 20ml.
Attention
● Don’t load/unload sample cup or reagent bottle when the sample/reagent is running.
7
TC220 Chemistry Analyzer Technician Manual(Ver1.15) YeMengRui
Sample/reagent
syringe assembly
● Specification
8
TC220 Chemistry Analyzer Technician Manual(Ver1.15) YeMengRui
● Motion Specification
Sample/reagent dispenser mechanical system:it moves again and again with the sequence
Liquid level detection:sample/reagent probe have auto liquid level detection funtion,it is
Crash protection (vertical & horizontal):When the probe crashs in horizontal or vertical
Reagent refrigeration system:reagent refrigeration system starts to work when the cooling
switch is on.
Attention
When the analyzing unit is in operation, do not place any part of your body or any obstacle in the route the
Mixer System
Mixer arm
Mixer probe
9
TC220 Chemistry Analyzer Technician Manual(Ver1.15) YeMengRui
● Function
It is used to stir the reaction liquid (reagents and sample) in reaction cuvette to make them
react completely.
● Motion
When stirring is finished, the mixer moves automatically to its wash well for cleaning.
For single-reagent test, the sample mixer starts to work once sample is dispensed.
For double-reagent test, the reagent mixer starts to work once R2 is dispensed.
Reaction cuvette
holder lock knob
Reaction cuvette
holder lock knob
Reaction tray assembly is consist of reaction tray and temperature control parts.
Reaction tray is used to load reaction cuvettes. Reaction cuvette is as reaction liquid
● Specification
Temperature control parts:via heating loop and temperature sensor to make the reaction
temperature is in the 37±0.1℃.
Reaction cup size:5mm×6mm×25mm (L×D×H),,optical path 6.1mm.
● Motion
Reaction tray only make counterclockwise rotation。Reaction tray make appointed reaction
cuvette stop in adding sample position、adding reagent position and measure optical axis
position.
11
TC220 Chemistry Analyzer Technician Manual(Ver1.15) YeMengRui
● Composition
● Specification
Chapter 2 Installation
Attention
The system should be installed by our authorized personnel only.
The system should be installed by our authorized personnel only, and you should prepare a
If you need to move the system to another site, please contact our Customer Service
Department or your local distributor, who are the appropriate people for the moving job.
2.1 Unpacking
When you receive the system, carefully inspect the package. If you see any signs of
mishandling or damage, file a claim immediately with our Customer Service Department or
12
TC220 Chemistry Analyzer Technician Manual(Ver1.15) YeMengRui
After opening the package, check the delivered goods against the packing list as well as
the appearance of the system. If you find anything missing or damaged, alert our Customer
Attention
Make sure the system is installed in a place meeting the following requirements. Otherwise,
⊙ The bearing platform (or ground) should be level (gradient less than 1/200).
⊙ The bearing platform (or ground) should be able to bear 30Kg weight.
Attention
The system radiates heat while operating. A well-ventilated environment helps keep the
room temperature stable. Use ventilation equipment if necessary. But if so, be sure not to
expose the system to the direct draft that may lead to unreliable results.
⊙ The installation site should be free of corrosive gas and flammable gas.
13
TC220 Chemistry Analyzer Technician Manual(Ver1.15) YeMengRui
⊙ The site should not be disturbed by large noise or power supply.
⊙ The system should not be placed near brush-type motors and electrical contacts that are
⊙ Do not use such devices as mobile phones or radio transmitters near the system.
Electromagnetic waves generated by those devices may interfere with operation of the
system.
⊙ The distance between the power socket and the system should be less than 2.5 meters.
⊙ The equipment’s maximum output is 350 W, customizable with 1000W on-line UPS.
Warning
Make sure the power socket is grounded correctly. Improper grounding may lead to electric
Be sure to connect the system to a power socket that meets the above-mentioned
14
TC220 Chemistry Analyzer Technician Manual(Ver1.15) YeMengRui
Attention
Operating the system in an environment other than the specified may lead to unreliable test
results.
If the temperature or relative humidity does not meet the above-mentioned requirements,
The system should be installed and used meeting the space and accessibility requirements
as shown below.
wall
Minimum500
Maximum2000
Analyzer PC
383
548 Unit:mm
Front
1.Make sure that in the hospital, there is a proper installation area with qualified
environments.
15
TC220 Chemistry Analyzer Technician Manual(Ver1.15) YeMengRui
3. In the installation place, unpack the package, check the goods according the packing-list
5. Install computer, printer and then test the printer whether it can work normally.
Attention
The TC220 software is better used in the Windows XP and WIN7 operating system.
6. Inspect the equipment appearance to see whether it is OK. Open the top cover, check
7. Holding the sample/reagent probe assembly arm, pull several times up and down & left
and right respectively at wash well position, reaction cup position, reagent position, sample
position and other positions around, to see whether the movement resistance is uniform
8. Connect serial port communication cable, equipment power line, ground wire, distilled
9. Fill up the distilled water bucket with distilled water,then connect the distilled water
Water tube
Water sensor
16
TC220 Chemistry Analyzer Technician Manual(Ver1.15) YeMengRui
Waste sensor
Waste tube
When install the machine,please put water and waste bucket on the lower level place than
machine.otherwise water will flow out from machine.the connection pictures are shown as
followings.
17
TC220 Chemistry Analyzer Technician Manual(Ver1.15) YeMengRui
Tube connection
Water bucket
Waste bucket
10. Check whether the 60pc reaction cuvettes are damaged in the reaction tray. Please
observe whether each reaction cups height are at the same level on the reaction tray.
Remarks:
The reaction tray is consisted by 6 connected cup holders and each holder is loaded by
11. Please open the power switch as follow: analyzer power switch→computer power
12. Copy the Chemistry analyzer installation software from installation CD into the
Attention
18
TC220 Chemistry Analyzer Technician Manual(Ver1.15) YeMengRui
there is not 9-pins serial port in the computer,the convertor should be used,otherwise it is
not needed.
Step1: there are two ports in the USB convertor.one is USB port.the other is male 9-pins
serial port.
Step 2: connect the serial cable with male 9-pins port of convertor,and connect the USB
Step 4: Insert driver CD of USB convertor into CD drive.and then the automatic installation
19
TC220 Chemistry Analyzer Technician Manual(Ver1.15) YeMengRui
Step 5: The screen will indicate “Found new hardware wizard” see the following picture.
20
TC220 Chemistry Analyzer Technician Manual(Ver1.15) YeMengRui
21
TC220 Chemistry Analyzer Technician Manual(Ver1.15) YeMengRui
Step 7: after install the driver of convertor,enter the hardware information menu which is in
the control panel to check whether the added com number is 1 to 3 or not.if not,it
should be done to modify it through advanced properties menu of the added com
port.
13. There are two folders in the installation CD , one folder is “TC200_V43_EN” which is
operating software folder, the other folder is “PIC”.there are three factory setting menu
Attention
① There is some important factory setting information in the CD,if you change the
computer or reinstall the windows operating system,you should set some menu again.so
22
TC220 Chemistry Analyzer Technician Manual(Ver1.15) YeMengRui
② Each machine have own factory setting pictures,so it is necessory to keep each
14. Open the “TC200_V43_EN” folder and double click the icon of TC200 application
program which is .then select login ID “Admin” and input “Admin” password to
enter the software operation interface.
Note:in the original software ,the default login ID have two ID.one is Admin,the password is
Admin.the other is 1,the password neednot be inputted.
15. Enter “Maintenance”→ Parameters setting →input password“777”,set up related
parameters.
23
TC220 Chemistry Analyzer Technician Manual(Ver1.15) YeMengRui
Parameter Meaning
Com Serial Port It is the serial port number connected between machine and computer.
Select Generally it is automatically detected by software.
Language Normally choose English.
Barcode Length Generally, the item should be set up by engineers according the Barcode requirement.
Separate Mixing TC220 haveseparate mixing system. select “Mixing system”
System
If select “power on run”,when customer enters the operating software,the machine automatically
Power On Wash
washes all reaction cuvettes.
If select “power off run”,when customer exits the operating software,the machine automatically
Power Off Wash
washes all reaction cuvettes.
Screen Color Click “Back color” button to choose background color for software operating interface according
Setup customer’s favor.
Syringe Drive Only select “Drive Board”,otherwise sample/reagent syringe movement is abnormal.
Mode The drive board is M7 Motor Subdivision Board and refer chapter “Circuit board function description”.
Result too small If the result is more less than lower limit of reference range,the
parameter will be retested.
Balance has not been If the blank absorbance is not stable,the parameter will be retested.
Auto-check tested
setup Beyond the scope of If the result is beyond upper limit of linearity range,the parameter will be
linear retested with auto-dilution.
Substrate depleted If substrate is out, the parameter will be retested with auto-dilution.it is
valid for kinetic and two-points method.
Server IP According the LIS system setting.
If the sample/reagent probe cannot detect the corresponding reagent, the machine donot test the
corresponding parameter in the next measurement cycle.
Rgt.Sample
If the sample/reagent probe cannot detect the sample some times,the machine donot test the sample
Alarm
in the next measurement cycle.this time can be set in the “instrument settings”menu.its name is
serum times.normally we set four times.
16. Click “Initialize” to observe the Initialization status of each parts and the lamp should
facus the light point into the NO.11 cuvette.otherwise you should adjust the reaction tray
24
TC220 Chemistry Analyzer Technician Manual(Ver1.15) YeMengRui
Attention
① If the water tank can’t be injected water, please unscrew the air plug which is on the left
side of analyzer about 20 seconds.Screw the air plug tightly when the water arrives at
② Donot test CRU wash valve in the Motion detection menu,if you click it,the water will be
dispensed into cuvettes in a very short time,and then flow out cuvettes and flow into the
incubation groove.
18. Input password “999” to enter “Instrument settings” menu. Then confirm the setting
25
TC220 Chemistry Analyzer Technician Manual(Ver1.15) YeMengRui
Attention
There are one sample/reagent probe and one mixture probe in the TC220,arm setup
colume and stiring setup colume are used to adjust probe’s positions.if positions are not
19. Enter “Temperature pressure” menu. Then confirm the setting values is the same with
factory picture .
20. Enter “Cuvette cleaning” then perform washing all cuvettes three times.
21. Enter cuvettes checking menu, click “Add water” and then “check cuvettes” to get the
signal value , observe each wavelength signal value ,it should be between 30000-56000,
If the signal value is lower than 30000 or more than 56000, please check related
wavelength cuvettes.
26
TC220 Chemistry Analyzer Technician Manual(Ver1.15) YeMengRui
27
TC220 Chemistry Analyzer Technician Manual(Ver1.15) YeMengRui
28
TC220 Chemistry Analyzer Technician Manual(Ver1.15) YeMengRui
Drain pump
Sample/reagent syringe
Attention
Before you detect CRU wash valve(water dispensing valve),you should put one empty cup
under the washing needles,otherwise some water will be dispensed into cuvettes in the
short time,and then overflow into the incubation groove.the overflow water will damage the
rotating stepping motor of reaction tray.
29
TC220 Chemistry Analyzer Technician Manual(Ver1.15) YeMengRui
Power board
(Right View)
Water pump
Water tank
(Left View)
30
TC220 Chemistry Analyzer Technician Manual(Ver1.15) YeMengRui
Backwater valve
Drain pump
Backwater pump
(Rear View)
1.Water pump1(P3) and water pump2(P4)to take distilled water into water tank, and when
When the liquid level of water tank reaches certain height, air valve(EV4) will open to take
in air. Then when the pressure and liquid level reach certain level and keep constant, water
The liquid level of water tank is controlled by its inner float switch whereas the pressure is
2.1 When the seven washing needles goes down to the bottom of reaction cuvettes, the
drain pump(P0) will be on ,then aspirate and pump water from cuvettes.
2.2 Water dispensing valve(EV1) is opened, pressure in water tank makes the distilled
water of tank dispense into reaction cuvettes from five short washing needles through water
filter. at the meantime, backwater pump(P1) and backwater valve(EV5) are working ,then
31
TC220 Chemistry Analyzer Technician Manual(Ver1.15) YeMengRui
pumpback completely the water of front end five short washing needles and prevent water
Sample/ reagent probe rotates above the reagent bottle or sample serum cup, and
sample/ reagent probe rotates above the reaction cuvettes, and then goes down into it, and
Then sample/reagent probe goes up and rotates to washing well, EV2 and EV3 will be
opened, the water which is from water tank wash the inner and outer side of
Note: there is different sample/reagent probe movement route according single or double
reagent.
Mixer probe rotates above the reaction cuvettes, and then goes down into it for mixing,and
32
TC220 Chemistry Analyzer Technician Manual(Ver1.15) YeMengRui
cooling fans.one side with words of semiconductor is cold side.the hot side of
with the bottom of reagent tray.two cooling fans release the heat which come from the
refrigeration semiconductor
(hot side)
Size:30×30×3.5
Refrigeration semiconductor
(cold side)
Reaction tray have 60 cuvettes which supported by six cuvette holders with each holder
Temperature control of reaction tray:the heating loop which is on the bottom of incubation
33
TC220 Chemistry Analyzer Technician Manual(Ver1.15) YeMengRui
groove heats the incubation groove. Temperature sensor and temperature controller board
The type of stepping motor which control the reaction tray rotation is 23HY109-20B2.
Initialization slot
Code disk
Initialization slot
When you click initialization button which is in the operation software, the reaction tray
rotates until the initialization slot of code disk is detected by optical sensor.the light of
34
TC220 Chemistry Analyzer Technician Manual(Ver1.15) YeMengRui
Optical sensor is used to make main controller board know the position of mechanical part
movement.if the optical sensor is not working well,the corresponding mechanical part
35
TC220 Chemistry Analyzer Technician Manual(Ver1.15) YeMengRui
Specification:
The type of stepping motor which control the probe to be up and down or left and right is
17HD432Y-22B.
The type of Optical Sensor is TP808,it is used to detect the position of the probe
36
TC220 Chemistry Analyzer Technician Manual(Ver1.15) YeMengRui
movement.
Sample/reagent probe assembly drive sample/reagent probe to move up and down, left
and right, and make sample/reagent probe to the corresponding reagent position, sample
Specification:
The type of stepping motor which control the mixer probe to be vertical movement is
17HD432Y-22B.
The type of stepping motor which control the mixer probe to be horizontal movement is
14HY115Y-22B.
The type of Detection Sensor is ST135,it is optical switch,it is used to detect vertical
The type of Detection Sensor is ST135,it is optical switch,it is used to detect horizontal
It is used to stir the reaction liquid (reagents and sample) in reaction cuvette to make them
react completely.
37
TC220 Chemistry Analyzer Technician Manual(Ver1.15) YeMengRui
Specification:
The type of stepping motor which control the syringe to be up and down is 17HA403Y-18B.
The type of Optical Sensor is ST157,it is used to detect the up or down position of the
syringe.
Sample/reagent syringe is used to aspirate the proper amount of reagent and sample, the
stepping motor .
Specification:
The type of stepping motor which control the washing arm to be up and down is
17HD432Y-22B.
The type of Optical Sensor is TP808,it is used to detect the up or down position of the
washing arm.
Washing unit is used to wash cuvettes and keep cuvettes clean, make the reaction result
more accurate.
There are 7 wash needles on the washing unit. five short needles are used to dispense
distilled water into cuvettes,seven long needles are used to aspirate water from cuvettes.
38
TC220 Chemistry Analyzer Technician Manual(Ver1.15) YeMengRui
The lamp gives out the lights and the lights go into the cuvettes of the reaction tray, transmit
through the cuvettes and separated by light filters to be specific wavelengths, and then go
to the signal process board. the light signal are converted to electric signal through
photodiod .the electric signal is amplified and converted to the A/D conversion circuit, and
then sent to the main controller board,which will submit the absorbance to the PC for
calculation.
The absorbance is taken when the reaction cuvette rotates to the corresponding optical
path.
39
TC220 Chemistry Analyzer Technician Manual(Ver1.15) YeMengRui
40
TC220 Chemistry Analyzer Technician Manual(Ver1.15) YeMengRui
Code disk
Optical Fiber
Lamp
Incubation groove
Temperature switch will cut off ,when temperature of reaction tray reaches to be more than
Temperature sensor (LM95071) is surface mounted IC,when install it,it should be done to
daub some heat transfer silicone on the induction side and donot fix so tight that damage
the chip.
main controller board, The main controller board makes temperature signal compare with
temperature setted up by software, then main controller board transmits signal to power
board,making LR3410(V1) breakover,DC 24V power will load to the heating loop, the
heating loop will heat the reaction groove,the D8 indicator of power board always lights. If
temperature of reaction tray reaches to the temperature setted up by software, then main
controller board transmits signal to power board,making LR3410(V1) cut off, the heating
+24v power
output interface
for heating loop
V1(LR3410)
D8 indicator
(Power Board)
42
TC220 Chemistry Analyzer Technician Manual(Ver1.15) YeMengRui
(Temperature sensor.LM95071)
When the water temperature of tank is higher than 60℃,the temperature switch will be
main controller board, The main controller board makes temperature signal compare with
temperature setted up by software, then main controller board transmits signal to power
stick.it will heat water,the D10 of power board always lights. If water temperature reaches to
the temperature setted up by software, then main controller board transmits signal to power
board,making LR3410(V2) cut off, the calefaction stick stops heating, the D10 indicator of
43
TC220 Chemistry Analyzer Technician Manual(Ver1.15) YeMengRui
V2(LR3410)
D10 indicator
(Power Board)
TC220 system power supply consists of one switch power supply (SP-320-24).it provides
1.Power board
3.Bus board
1. Bus board
1. Bus board
45
TC220 Chemistry Analyzer Technician Manual(Ver1.15) YeMengRui
⑷ Detect the signal of distilled water level sensor and waste level sensor;
46
TC220 Chemistry Analyzer Technician Manual(Ver1.15) YeMengRui
D10 indicator
D9 indicator
D7 indicator
D12 indicator
D6 indicator
D13 indicator
D18 indicator
Communication
D3 indicator
Port with Water
Connected with
controller board
temperature sensor
of reaction tray
Communication
Port with Power
board
Connected with
temperature sensor
Connected with level sensor
of water tank
of water and waste
D2 indicator
D1 indicator
Power inlet D4 indicator
Communication Port
with bus board RS323 Port
D5 indicator
47
TC220 Chemistry Analyzer Technician Manual(Ver1.15) YeMengRui
D6 indicator When the reaction tray stop on the initialization position,it is bright
Four motor drive boards insert into slots which are on the bus board.and the bus board
collects and send mechanical motor position signal to the main controller board.and the
main controller board send back command to bus board according software operation.
48
TC220 Chemistry Analyzer Technician Manual(Ver1.15) YeMengRui
Signal Process Board Function: the light signal are converted to electric signal through
photodiod .the electric signal is amplified and converted to the A/D conversion circuit, and
then sent to the main controller board,which will submit the absorbance to the PC for
calculation.When the board is added with power,the D9 and D10 indicator always light.
Communication Port
D10 D9
with Main controller
Board
Signal Gain Signal Offset
49
TC220 Chemistry Analyzer Technician Manual(Ver1.15) YeMengRui
When aspirating samples/reagents and dispensing into reaction tray, the chemistry
analyzer requires the probe to dip into the liquid for a specific depth(2mm below the liquid
level), so as to avoid carryover that has impacts on test results, and to avoid air aspiration
when the reagent/sample is insufficient. Also when washing the probe in wash well, probe
also need to stay at a special depth to wash the inner side and outer side.
When the sample/reagent probe touches the sample/reagent level, there will be a change
discriminator then outputs a voltage variation,which will be filtered by a low-pass filter. After
that, a band-pass filter takes the variation and reshapes it to pulse signals, which are finally
There are four groups of jumper on the liquid level detection board,sensitivity of the liquid
level detection can be adjusted through add or remove jumper cap.more jumper caps
Four groups of
jumper.normally we
install two jumper caps.
50
TC220 Chemistry Analyzer Technician Manual(Ver1.15) YeMengRui
D1
D2
Anti-Crash subassembly
When the probe crashs in horizontal or vertical direction,the anti-crash subassembly will be
active,and it sends signal to motor drive board and make sample/reagent probe arm stop
51
TC220 Chemistry Analyzer Technician Manual(Ver1.15) YeMengRui
main controller board, The main controller board makes temperature signal compare with
temperature setted up by software, then main controller board transmits signal to power
board,making LR3410(V1) breakover,DC 24V power will load to the heating loop, the
heating loop will heat the reaction groove,the D8 indicator of power board always lights. If
temperature of reaction tray reaches to the temperature setted up by software, then main
controller board transmits signal to power board,making LR3410(V1) cut off, the heating
loop stops heating, the D8 indicator of temperature control board always blink.
main controller board, The main controller board makes temperature signal compare with
52
TC220 Chemistry Analyzer Technician Manual(Ver1.15) YeMengRui
temperature setted up by software, then main controller board transmits signal to power
stick.it will heat water,the D10 of power board always lights. If water temperature reaches to
the temperature setted up by software, then main controller board transmits signal to power
board,making LR3410(V2) cut off, the calefaction stick stops heating, the D10 indicator of
This board is connected with valve,pump and control their working condition according
Communication port
Inner probe washing valve
with main controller
board
Backwater peristaltic pump
Drain pump
Air valve
53
TC220 Chemistry Analyzer Technician Manual(Ver1.15) YeMengRui
Cooling fan of
refrigeration
semiconductor
Communication port
with power board
Connected with
temperature sensor
of reagent tray
Connected with
Connected with temperature switch
temperature switch
main controller board, The main controller board makes temperature signal compare with
temperature setted up by software, then main controller board transmits signal to power
temperature of reagent tray is lower than the temperature setted up by software, then main
controller board transmits signal to refrigeration pinboard ,making LR3410(V1) cut off, the
accurate.
Program switch
1 2 Number
ON OFF 16 8
55
TC220 Chemistry Analyzer Technician Manual(Ver1.15) YeMengRui
MotorStyle is 8
This board is used to drive the corresponding stepping motor according command from
main controller board.there are four pieces motor drive boards,see the following
pictures.
56
TC220 Chemistry Analyzer Technician Manual(Ver1.15) YeMengRui
Four stepping motor drive boards have own program switch position except the board
When you replace motor drive board,you should notice the program switch
K2-5:off.K2-4:on.K2-3:on.K2-2:off.K2-1:off K1-5:off.K1-4:on.K1-3:off.K1-2:on.K1-1:on
K2-5:off.K2-4:off.K2-3:on.K2-2:on.K2-1:on K1-5:off.K1-4:off.K1-3:off.K1-2:on.K1-1:off
K2-5:off.K2-4:off.K2-3:on.K2-2:offK2-1:on K1-5:off.K1-4:off.K1-3:on.K1-2:on.K1-1:off
57
TC220 Chemistry Analyzer Technician Manual(Ver1.15) YeMengRui
When the stepping motor movement problem is occurred,the error message will
appear and give “Motor Number is error”.in the following table,it is shown that the
DC+2.4~2.5V output
DC+12V input
58
TC220 Chemistry Analyzer Technician Manual(Ver1.15) YeMengRui
Steps:
Remove the power connector of Tungsten-Halogen Lamp and switch on the machine,then
set the multimeter to DC 200mV grade. The black pen of the multimeter is connected to
ground point, and the red pen measure the corresponding test points of wavelength, adjust
5~10mv.
according this adjustment procedure,to adjust the offset voltage of all the wavelengths.
Firstly initialize the machine,and add 400ul distilled water into the NO.11 reaction
cuvette,waiting thirty minutes, set the multimeter to DC 20V grade. The black pen of the
multimeter is connected to ground point, and the red pen measure the corresponding test
according this adjustment procedure,to adjust the gain voltage of all the wavelengths.
59
TC220 Chemistry Analyzer Technician Manual(Ver1.15) YeMengRui
Not used
After initialize the machine,the light of lamp should focus the middle of the NO.11 cuvette.if
the reaction tray position is not suitable,it is necessary to adjust it as following steps.
60
TC220 Chemistry Analyzer Technician Manual(Ver1.15) YeMengRui
Three retaining
screws
3.Rotate the reaction tray and make light of lamp focus the middle of the NO.11 cuvette.
Steps
61
TC220 Chemistry Analyzer Technician Manual(Ver1.15) YeMengRui
The connector of
lamp
4.Loosen the retaining screw of lamp and then take out the lamp.
5.Put the new lamp into lamp hole and connect the power connector of lamp.
8.Adjust the lamp position and observe whether the light focus the middle of incubation
groove hole or not.if it is,tighten the retaining screw of lamp and reinstall the cover of lamp.
9.Finish replacement.
62
TC220 Chemistry Analyzer Technician Manual(Ver1.15) YeMengRui
Note:after you replace the new lamp,you should adjust signal process board according 5.1
settings”.then input password”999” then enter this menu which is shown in the following
picture.
CRU setup colume and time setup colume are only filled in with factory value.
The arm setup colume is used to adjust sample/reagent probe position which is related with
63
TC220 Chemistry Analyzer Technician Manual(Ver1.15) YeMengRui
1.Click “Wash Pos.”button,and then click “Start”button,the probe tip should be upon in the
2.Click “Deep” button ,then the probe will insert into wash well.
3.Increase or decrease the depth value to make the probe tip just touch the mouth of water
64
TC220 Chemistry Analyzer Technician Manual(Ver1.15) YeMengRui
4.Use the depth value which is got from step 3 and then fill in final depth value.
tray.and the probe tip should be upon in the middle of NO.1 cuvette.if not,increase or
2. Click “Deep” button ,then the probe will insert into the NO.1 cuvette.
3. Increase or decrease the depth value to make the probe tip just touch the bottom of the
NO.1 cuvette.
Depth value of cuvette
4. Use the depth value which is got from step 3 to subtract 2,and then fill in final depth
value.
65
TC220 Chemistry Analyzer Technician Manual(Ver1.15) YeMengRui
1. Put one empty serum cup on the NO.1 sample position,and the other empty serum cup is
tray.and the probe tip should be upon in the middle mouth of NO.1 serum cup.if
tray.and the probe tip should be upon in the middle mouth of NO.18 serum cup.if
5. Click “Deep” button ,then the probe will insert into the NO.18 serum cup.
6. Increase or decrease the depth value to make the probe tip just touch the bottom of the
7. Use the depth value which is got from step 6 to subtract 2,and then fill in final depth
value.
Note:our software allow customer to add different tube in the “data dictionary menu”.when
customer add new tube,service engineer should adjust depth value for the new tube
according step 6.and notice when they load new sample in the “edit” menu,then should
select correct “cup type” for the sample.otherwise some wrong result will be occurred.
1. Put one empty reagent bottle on the NO.1 reagent position,and the other empty reagent
tray.and the probe tip should be upon in the middle mouth of NO.1 reagent bottle.if
67
TC220 Chemistry Analyzer Technician Manual(Ver1.15) YeMengRui
tray.and the probe tip should be upon in the middle mouth of NO.26 reagent bottle.if
5. Click “Deep” button ,then the probe will insert into the NO.26 reagent bottle.
6. Increase or decrease the depth value to make the probe tip just touch the bottom of the
7. Use the depth value which is got from step 6 to subtract 2,and then fill in final depth
68
TC220 Chemistry Analyzer Technician Manual(Ver1.15) YeMengRui
value.
If the probe is bent or damaged, it must be replaced immediately. follow the procedure
2. Take out the sample tray,and then pull the probe arm to its highest point. rotate the probe
arm to move the probe to a position above the sample tray and convenient to operate.
3. Use a small screwdriver to loosen the two retaining screws on the probe.
One screw
69
TC220 Chemistry Analyzer Technician Manual(Ver1.15) YeMengRui
WARNING:
NOTE:
Exercise caution when pulling the probe away from the arm.
BIOHAZARD:
settings”.then input password”999” then enter this menu which is shown in the following
picture.
The stiring setup colume is used to adjust mixer probe position which is related with wash
well,cuvette position.
70
TC220 Chemistry Analyzer Technician Manual(Ver1.15) YeMengRui
1.Click “Wash Pos.”button,and then click “Start”button,the probe tip should be upon in the
Mixer motor
Retaining screw
2.Click “Deep” button ,then the probe will insert into wash well.
3.Increase or decrease the depth value to make the probe tip just touch the mouth of water
Wash well
Mixer probe
4.Use the depth value which is got from step 3 and then fill in final depth value.
5. Click save and exit the menu then initialize the machine.
tray.and the probe tip should be upon in the middle of NO.1 cuvette.if not,increase or
72
TC220 Chemistry Analyzer Technician Manual(Ver1.15) YeMengRui
2. Click “Stiring deep” button ,then the probe will insert into the NO.1 cuvette.
3. Increase or decrease the depth value to make the probe tip just touch the bottom of the
NO.1 cuvette.
4. Use the depth value which is got from step 3 to subtract 4,and then fill in final depth
value.
5. Click save and exit the menu then initialize the machine.
If the probe is bent or damaged, it must be replaced immediately. follow the procedure
73
TC220 Chemistry Analyzer Technician Manual(Ver1.15) YeMengRui
3. Remove the retaining screw of mixer probe arm and then take out the mixer probe arm.
74
TC220 Chemistry Analyzer Technician Manual(Ver1.15) YeMengRui
There are synchronous belts for the left/right and up/down of the sample/reagent probe arm
and 7-steps washing unit. It also exists for the reaction tray.
Steps:
3. Unscrew the nine screws which fix the cover of the machine ,and then carefully lift and
4. Unscrew the three retaining screws which fix the shield cover of signal process board.
5. Unplug the signal connector of signal process board from main controller board.
7. Carefully remove the signal process board and put it on the dry and clean place.
76
TC220 Chemistry Analyzer Technician Manual(Ver1.15) YeMengRui
9. Loosen the retaining screw of 690nm wavelength locating sleeve and carefully take out
Fiber of 690nm
10. Push out the locating sleeve and take out the whole locating sleeve.
77
TC220 Chemistry Analyzer Technician Manual(Ver1.15) YeMengRui
11. Open the locating sleeve and push out the filter using tool.
Light filter
12. Install the new light filter according the following direction.
78
TC220 Chemistry Analyzer Technician Manual(Ver1.15) YeMengRui
Filter
13. Install the whole locating sleeve on the filter bracket.and tighten the retaining screw.
Note:during tighten the retaining screw,donot screw too tight,otherwise filter will be
damaged by fracturing.
14. Adjust the signal process board according 5.1 signal process board voltage adjustment.
79
TC220 Chemistry Analyzer Technician Manual(Ver1.15) YeMengRui
Chapter 6 Maintenance
performance.
2. Check the level of distilled water. If fluid level is low, replace it as follows.
3. Check the tubing of distilled water whether there is some leakage or oppressed tubing or
not.
2. Check the level of waste solution. If the level is low, this procedure needn’t to be
NOTE: It is recommended that you should execute this procedure every day.
80
TC220 Chemistry Analyzer Technician Manual(Ver1.15) YeMengRui
Ready to test
3. In this menu,we set the minimum signal value is 25000 and maximum signal value is
65535.if the signal value of cuvette is less than 25000 or more than 65535.the software will
mark red color on the corresponding cuvette number.the best signal value range of cuvette
When you find the cuvette which is bad signal value.you should take out the cuvette,and
observe what’s wrong with it.and clean it using wash solution or replace it.
When the probe is clogged, the fluid flow will become abnormal and the sample/reagent
81
TC220 Chemistry Analyzer Technician Manual(Ver1.15) YeMengRui
3. Click “internal probe wash” button,and then click “Start” button.immediately observe the
probe tip.
4. Normally very thin liquid flows out from probe.and the liquid column is straight.otherwise
the probe is clogged.you can use a needle to unclog the probe from tip.
WARNING:
The probe tip is sharp and can cause puncture wounds. To prevent injury, exercise caution when working around the
probe.
82
TC220 Chemistry Analyzer Technician Manual(Ver1.15) YeMengRui
BIOHAZARD:
Wear gloves and lab coat and, if necessary, goggles.Dispose of the used gauze in accordance with your local or national
guidelines for biohazard waste disposal.Dispose of the waste in accordance with your local or national guidelines for
biohazard waste disposal and consult the manufacturer or distributor of the reagents for details.
3. Pull the sample/reagent and mixer probe arm to the highest point by hand. Rotate the
4. Pinch acid or alkaline detergent-soaked gauze with tweezers and gently clean the
exterior of probes.
6. After cleaning, gently pull the probe arms to its highest point and rotate the probe arms to
2. Pull the sample/reagent probe and mixer arm to its highest point. Rotate the arms to
3. Clean the inside of and the place around wash well with cotton swabs.
4. Pull the sample/reagent probe and mixer arm to its highest point and rotate them to be
83
TC220 Chemistry Analyzer Technician Manual(Ver1.15) YeMengRui
2. Take out reagent bottle and sample cup from sample or reagent tray.
3. Wash sample and reagent tray with clean water and wipe it dry with clean gauze.
4. Wash sample and reagent rack with clean water and wipe it dry with clean gauze.
Using gauze moistened with distilled water, wipe the surface and around of the reaction
tray.
NOTE: DO NOT clean the reaction tray using gauze moistened with ethanol or detergent
Clean the individual nozzles with gauze moistened with distilled water from up to down.
NOTE: DO NOT touch the surface with tweezers directly preventing from scratch of
nozzles.
Normally the machine clean cuvettes using distilled water during testing.so the customer
Note:wash solution should be bought from TECOM corporation with machine’s order.
3. Put the wash solution bottle on the one empty reagent position.
6. Click the washing button and then the machine automatically aspirate and dispense
Washing button
7.This procedure makes the cuvettes soak with wash solution soak in and clean it.
85
TC220 Chemistry Analyzer Technician Manual(Ver1.15) YeMengRui
3.Click “Add water”button,and the machine automatically add some water into every
cuvette.
4. after the process of add water,click”Check cuvette” then the machine automatically
5. if the signal value of cuvette is less than 25000 or more than 65535.the software will
mark red color on the corresponding cuvette number.the best signal value range of cuvette
When you find the cuvette which is bad signal value.you should take out the cuvette,and
observe what’s wrong with it.and clean it using wash solution or replace it.
6. Click “remove water” and the machine empty water from all cuvettes.
2. Remove all cuvettes holder from reaction tray and put it on a clear table.
NOTE: DO NOT scratch the light path when wipe the incubator.
86
TC220 Chemistry Analyzer Technician Manual(Ver1.15) YeMengRui
Incubation groove
2. Wipe the panel of the machine with clean gauze (water or disinfector-dipped gauze if
necessary).
2. Unscrew the cap of distilled water bucket(together with the distilled water tube and the
sensor).
CAUTION:
After removing the cap (together with the tube and sensor), place it on a clean table.
3. Wash the bucket interior with deionized water. Use a clean brush to clean the interior if
necessary.
4. Wash the water tube and the sensor with deionized water. Use clean gauze to wash
them if necessary.
87
TC220 Chemistry Analyzer Technician Manual(Ver1.15) YeMengRui
5. Wipe water off the bucket exterior, water tube and sensor cable with clean gauze.
7. Screw the cap (together with the tube and sensor) back onto the bucket until secure.
BIOHAZARD:
Exercise caution and do not spill the waste onto other people or things.
2. Unscrew the cap (together with the waste tube and the sensor).
BIOHAZARD:
After removing the cap of the waste tank (together with the waste tube and sensor), place it on an appropriate place to
4. Wash the bucket interior with clean water. Soak the bucket with disinfector if necessary.
5. Wash the waste tube and the sensor with clean water.
6. Wipe water off the bucket exterior, waste tube and sensor cable with clean gauze.
7. Screw the cap (together with the waste tube and sensor) back onto the bucket until
secure.
CAUTION:
Ensure the waste tube is over the tank and not blocked, bent, or twisted. A blocked, bent or twisted waste tube may lead
88
TC220 Chemistry Analyzer Technician Manual(Ver1.15) YeMengRui
The water filter is connected with outlet of water tank.it is used to filtrate some dirty material
from water tank.it is necessary to clean it monthly.otherwise it will be clogged and affect
water flow.
Water filter
Water filter
Filter screen
Dust screen
90
TC220 Chemistry Analyzer Technician Manual(Ver1.15) YeMengRui
91
TC220 Chemistry Analyzer Technician Manual(Ver1.15) YeMengRui
92
TC220 Chemistry Analyzer Technician Manual(Ver1.15) YeMengRui
93
TC220 Chemistry Analyzer Technician Manual(Ver1.15) YeMengRui
94
TC220 Chemistry Analyzer Technician Manual(Ver1.15) YeMengRui
95
TC220 Chemistry Analyzer Technician Manual(Ver1.15) YeMengRui
96
TC220 Chemistry Analyzer Technician Manual(Ver1.15) YeMengRui
97
TC220 Chemistry Analyzer Technician Manual(Ver1.15) YeMengRui
98
TC220 Chemistry Analyzer Technician Manual(Ver1.15) YeMengRui
Appendix D Diagram
99
TC220 Chemistry Analyzer Technician Manual(Ver1.15) YeMengRui
100
TC220 Chemistry Analyzer Technician Manual(Ver1.15) YeMengRui
101
TC220 Chemistry Analyzer Technician Manual(Ver1.15) YeMengRui
102
TC220 Chemistry Analyzer Technician Manual(Ver1.15) YeMengRui
103
TC220 Chemistry Analyzer Technician Manual(Ver1.15) YeMengRui
104
TC220 Chemistry Analyzer Technician Manual(Ver1.15) YeMengRui
105
TC220 Chemistry Analyzer Technician Manual(Ver1.15) YeMengRui
106
TC220 Chemistry Analyzer Technician Manual(Ver1.15) YeMengRui
107
TC220 Chemistry Analyzer Technician Manual(Ver1.15) YeMengRui
108
TC220 Chemistry Analyzer Technician Manual(Ver1.15) YeMengRui
109
TC220 Chemistry Analyzer Technician Manual(Ver1.15) YeMengRui
110
TC220 Chemistry Analyzer Technician Manual(Ver1.15) YeMengRui
111
TC220 Chemistry Analyzer Technician Manual(Ver1.15) YeMengRui
112
TC220 Chemistry Analyzer Technician Manual(Ver1.15) YeMengRui
113
TC220 Chemistry Analyzer Technician Manual(Ver1.15) YeMengRui
114
TC220 Chemistry Analyzer Technician Manual(Ver1.15) YeMengRui
115
TC220 Chemistry Analyzer Technician Manual(Ver1.15) YeMengRui
116
TC220 Chemistry Analyzer Technician Manual(Ver1.15) YeMengRui
117
TC220 Chemistry Analyzer Technician Manual(Ver1.15) YeMengRui
118
TC220 Chemistry Analyzer Technician Manual(Ver1.15) YeMengRui
119
TC220 Chemistry Analyzer Technician Manual(Ver1.15) YeMengRui
120
TC220 Chemistry Analyzer Technician Manual(Ver1.15) YeMengRui
121
TC220 Chemistry Analyzer Technician Manual(Ver1.15) YeMengRui
122
TC220 Chemistry Analyzer Technician Manual(Ver1.15) YeMengRui
123
Figure D-26 Syringe Subassembly