Вы находитесь на странице: 1из 11

HEALTHCARE Enclosure

Imaging Services

Document No: DD+DIS019.10E


This number is needed for production line only:
F7.5170.1088.0 DX-M
Type 5170 / 200

DX-G
Type 5170 / 100
as of Software Version NIM_2000

Installation Instructions for


DICOM Digitizer Software NIM_2XXX*
*XXX = Version number of the software
Always use the installation instructions delivered with the Software that will be installed.
It is recommended to download the latest published version of the software from
MEDNET GSO Library Path: Computed Radiography => CR Digitizer => DX-M => Software

► Purpose of this Document

This enclosure is part of the DX-M DICOM digitizer software and provides the
installation instructions.

► Referenced Documents

Document Title
DD+DIS001.10E DX-G / DX-M Service Manual,
Chapter 3.7 - Software Menus, Settings
DD+DIS001.10E DX-G / DX-M Service Manual,
Chapter 3.8 - Software & Releases.

NOTE:
Dispose the enclosure document after finishing your task
(observe the confidentiality note below).

CONFIDENTIALITY NOTE:
Use, dissemination, distribution or reproduction of this document by unauthorized personnel is not permitted and may be unlawful.
DOCUMENT CONTROL NOTE:
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.

Revision 0
04-2010 printed in Germany Document Node ID: 29717238
Enclosure_e_template_v07
Agfa Company Confidential Copyright © 2010 Agfa HealthCare N.V.
DD+DIS019.10E Enclosure

► Manufacturer
Agfa HealthCare N.V.

Publisher
Agfa-Gevaert HealthCare GmbH
Tegernseer Landstraße 161
D - 81539 München
Germany
Copyright © 2010 Agfa HealthCare N.V.
All rights reserved.
Technical modifications reserved.
AGFA and the Agfa-Rhombus are trademarks of Agfa HealthCare N.V.

WARNING:
Improper operation or service activities may cause damage or injuries.

INSTRUCTION:
(1) Read the "Generic Safety Directions" document
(see MEDNET GSO => General Info => Agfa HealthCare => Publications =>
Service Manual) prior to attempting any operation, repair or maintenance task on
the equipment.
(2) Strictly observe all safety directions within the "Generic Safety Directions" and on
the product.

DOCUMENT CONTROL NOTE:


The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.

Revision 0 DX-G/DX-M Page 2 of 11


04-2010 Type 5170/100/200 as of SW NIM_2000 Agfa Company Confidential
DD+DIS019.10E Enclosure

LIST OF CONTENTS

1 APPLICATION RANGE OF DICOM DIGITIZER SOFTWARE .................................................4


2 PREREQUISITES FOR INSTALLATION .................................................................................4
3 PREPARING THE INSTALLATION..........................................................................................5
4 INSTALLATION ........................................................................................................................5
4.1 Installing Digitizer Software on the Digitizer via USB Memory Stick ........................................6

4.2 Creating a Backup ....................................................................................................................7

5 VERIFICATION OF SUCCESSFUL INSTALLATION...............................................................7


5.1 Checking Software Version via Information Button ..................................................................8

5.2 Checking Software version via Service Mode Button ...............................................................9

5.2.1 Device Info (Main Device Info)................................................................................................10

5.2.2 Modification (Software Installation).........................................................................................11

DOCUMENT CONTROL NOTE:


The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.

Revision 0 DX-G/DX-M Page 3 of 11


04-2010 Type 5170/100/200 as of SW NIM_2000 Agfa Company Confidential
DD+DIS019.10E Enclosure

1 Application Range of DICOM Digitizer Software

The DICOM Digitizer software can be used for the following CR applications:
• General Radiography (GenRad),
• Mammography (Mammo) and
• Full Leg – Full Spine (FLFS)

NOTE:
For more information regarding the features of the latest software refer to
DX-G / DX-M Service Manual, Chapter 3.8 - Software & Releases.

NOTE:
Software is upgradeable from every previous version of NIM_2xxx* software.
*xxx = Version number of the software

2 Prerequisites for Installation

IMPORTANT:
It is recommended to download the latest published version of the software from
MEDNET GSO Library: Computed Radiography => CR Digitizer => DX-M => Software
Always use the installation instructions delivered with the Software that will be installed.

REQUIRED TOOLS:
• Service PC with USB port
• USB Memory Stick with at least 20 MB free disk space (The USB Memory Stick
delivered with the digitizer is recommended. The Stick can also be ordered via
order number*: CM+9 5170 0760 0)
* The last digit in the spare part number indicates the spare part revision at release of this
document. When ordering, the actual revision of the spare part is delivered.

• Software to extract the zip-files (e.g. Winzip)

REQUIRED TIME:
Approximately 15 minutes for the installation via USB Memory Stick

NOTE:
Register this software installation in the local service management system as
modification, e.g. In SMS as “MOD”.

DOCUMENT CONTROL NOTE:


The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.

Revision 0 DX-G/DX-M Page 4 of 11


04-2010 Type 5170/100/200 as of SW NIM_2000 Agfa Company Confidential
DD+DIS019.10E Enclosure

3 Preparing the Installation

NOTE:
The digitizer software can be downloaded from the MEDNET GSO Library and is also
available as part of USB Memory Stick with order number*: CM+9 5170 0760 0
* The last digit in the spare part number indicates the spare part revision at release of
this document. When ordering, the actual revision of the spare part is delivered.

It is recommended to download the latest version of the software from


MEDNET GSO Library.

(1) Download the latest Digitizer Software (File size approximately 12 MB) from
MEDNET GSO Library, Path: Computed Radiography => CR Digitizer =>
DX-M => Software

(2) Extract all files from the downloaded zip file into a folder on the Service PC.
The downloaded zip file contains the software package as zip file named:
<NIM_2XXX.zip>* and this document.
* XXXX = Version number of the software

(3) Extract all files from <NIM_2XXX.zip>* with extraction mode <Extract to
here>. Folder NIM_2XXX* is created automatically.
* XXX = Version number of the software

4 Installation

Following steps are necessary to install the software via USB Memory stick:

# Type of software installation Section


1 Installing Digitizer Software on the Digitizer via USB Memory Stick 4.1
2 Creating a Backup 4.2
3 Verification of installed Software Versions 5

DOCUMENT CONTROL NOTE:


The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.

Revision 0 DX-G/DX-M Page 5 of 11


04-2010 Type 5170/100/200 as of SW NIM_2000 Agfa Company Confidential
DD+DIS019.10E Enclosure

4.1 Installing Digitizer Software on the Digitizer via USB Memory Stick

(1) Copy the folder “NIM_2XXX“ and its contents to the USB Memory Stick root
directory.
(2) Navigate to “\NIM_2XXX” on the USB Memory Stick.
* XXX = Version number of the software

(3) Double click on <Prepare_Installation.cmd>.


If the solid state disk of digitizer should be formatted additionally, double click on
<Prepare_Formatting.cmd> instead.
(4) Release the USB Memory Stick from Service PC.
(5) Switch off digitizer.
(6) Open front cover.
(7) Plug the USB Memory Stick into the USB slot of the digitizer.
NOTE:
The USB port of the digitizer does not accept all types of USB sticks. The
USB Memory Stick delivered with the digitizer is recommended. The Stick
can also be ordered via order number*: CM+9 5170 0760 0
* The last digit in the spare part number indicates the spare part revision at release of this
document. When ordering, the actual revision of the spare part is delivered.

USB Port

Figure 1

(8) Close digitizer.


(9) Switch on digitizer.
(10) Wait until digitizer has booted with the new software.
(11) Remove USB Memory Stick from digitizer.
IMPORTANT:
Do not remove the USB stick as long as it booted and the status
indicator do not continuously lighting up in green.

DOCUMENT CONTROL NOTE:


The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.

Revision 0 DX-G/DX-M Page 6 of 11


04-2010 Type 5170/100/200 as of SW NIM_2000 Agfa Company Confidential
DD+DIS019.10E Enclosure

4.2 Creating a Backup

NOTE
The backup procedure will be executed automatically by the digitizer.
To create a backup manually refer to chapter 3.7 Software Menus, Settings.

5 Verification of successful Installation

If the self-test is completed successfully, the digitizer enters the operator mode and the
status indicator is continuously lighting up in green.

ACTION:
Check the installed software version using the touch panel of the digitizer.

The Standby screen with following buttons


is shown in the local display of the digitizer:

c Erasure Button

d Configuration Button 1 2 3 4
e Information Button

f Service Mode Button

Figure 2: Menu in the local display of the digitizer

To check the installed Software Version of the digitizer use the following buttons:

• e Information Button, see section 5.1.


• f Service Mode Button, see section 5.2.

For additional information about the Touch Panel refer to


DX-G / DX-M Service Manual chapter 3.7 Software Menus, Settings.

DOCUMENT CONTROL NOTE:


The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.

Revision 0 DX-G/DX-M Page 7 of 11


04-2010 Type 5170/100/200 as of SW NIM_2000 Agfa Company Confidential
DD+DIS019.10E Enclosure

5.1 Checking Software Version via Information Button

The information button displays following information of the digitizer:


• IP Address
• Digitizer Name
• Software version

(1) Click the information button to display


the information screen.
The software version of the digitizer
is displayed.

Figure 3

(2) Click the button to exit the digitizer


information screen.

Figure 4

DOCUMENT CONTROL NOTE:


The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.

Revision 0 DX-G/DX-M Page 8 of 11


04-2010 Type 5170/100/200 as of SW NIM_2000 Agfa Company Confidential
DD+DIS019.10E Enclosure

5.2 Checking Software version via Service Mode Button

The service mode button allows the field service engineer to access the service
software of the digitizer.

(1) Click on the service mode button in


the local display of the digitizer.

Figure 5

(2) Enter password if applicable. Service Menu

Service menu will be displayed in Device Info


the local display.
Reporting

NOTE: Modification

Navigate through the service Analysis & Repair


menu by touching on the
several menu buttons or Preventive Maintenance

options with one finger. Configuration

Backup / Restore

Figure 6: Service Menu Main Screen

To check the installed Software Version of the digitizer use the following submenus:
• Device Info, see section 5.2.1.
• Modification, see section 5.2.2.

For additional information about the Touch Panel refer to


DX-G / DX-M Service Manual chapter 3.7 Software Menus, Settings.

DOCUMENT CONTROL NOTE:


The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.

Revision 0 DX-G/DX-M Page 9 of 11


04-2010 Type 5170/100/200 as of SW NIM_2000 Agfa Company Confidential
DD+DIS019.10E Enclosure

5.2.1 Device Info (Main Device Info)

The Menu Main Device Info shows the device specific data configured in
production.

(1) Select <Device Info – Main Service Menu > Device Info > Main Device Info
Device Info>.
Device specific data will be displayed.
Type Number: 5170
Subtype Number: 200
Serial Number: 12345
Manufacturer Date: 28 Jan 2010
The Device specific Data Software Version: NIM 2007
are “read-only” and just for Number of IP Cycles: 23456

identification of the device.

Figure 7

(2) Check installed Software Version.

(3) Select <Cancel> to go back to the menu.

DOCUMENT CONTROL NOTE:


The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.

Revision 0 DX-G/DX-M Page 10 of 11


04-2010 Type 5170/100/200 as of SW NIM_2000 Agfa Company Confidential
DD+DIS019.10E Enclosure

5.2.2 Modification (Software Installation)

The Menu Software Installation show the installed software of the following
boards on the digitizer:
• PMI Board,
• IP Handling Board and
• Cassette Handling Board

(1) Select <Modification – Service Menu > Modification > Software Installation
Software Installation>. Installed Software: NIM_2007
Installed software version is shown Board Software Version

in the local display. PMI Board


IP Handling Board
NIM_2007
NIM_2007
Cassette Handling Board NIM_2007

Instructions for Software Upload:


1. Switch off the Digitizer.
2. Insert USB Stick with prepared Software.
3. Switch Digitizer on to install Software.
4. Wait until the Digitizer shows the READY screen.
5. Switch Digitizer off and remove USB stick.

Figure 8

(2) Check installed software versions.


For upload Software refer to chapter 3.7 Software Menus, Setting.

(3) Select <Cancel> to go back to the menu.

Result Software Installation of DICOM digitizer software was successful.


NOTE:
If the solid state disk was formatted during installation, the digitizer needs to be
adapted to the hospital network again. Refer to DX-G / DX-M Service Manual,
Chapter 1 - "Controls, Connections and Setup" for more information.

DOCUMENT CONTROL NOTE:


The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.

Revision 0 DX-G/DX-M Page 11 of 11


04-2010 Type 5170/100/200 as of SW NIM_2000 Agfa Company Confidential