Thank you very much for using SUNX sensors.

Please read 3 SETTING OF PRESSURE LEAD


this Instruction Manual carefully and thoroughly for the DIRECTION AND PIPING
correct and optimum use of this sensor. Kindly keep this
manual in a convenient place for quick reference. Standard type
INSTRUCTION MANUAL 䢇 This product is not a safety sensor. Its use is not 䢇 Setting of pressure lead direction Main body
• The pressure lead direc-
Hexagon-
socket-
intended or designed to protect life and prevent body head bolts
LED Display • Digital Pressure Sensor injury or property damage from dangerous parts of
tion can be changed by
dismantling the pressure
DP2 Series For use outside Japan machinery. It is a normal pressure detection sensor.
䢇 In case this sensor is used within Japan, SI unit most port attachment and
changing the mounting
MJE-DP2 No.0580-05 be used since use of pressure units in Japan is
direction. The tightening
restricted to SI units.
torque of the hexagon-
socket-head bolt (length
1 SPECIFICATIONS 9mm or less) should be Pressure port
0.29N䡠m or less. Pressure
Vacuum pressure Positive pressure attachment ports
Type ⳮ101kPa type 100kPa type 1MPa type Hexagon-socket-
head plug
Standard Flat IP67 Light weight Standard Flat IP67 Standard Flat IP67
Note: Please make sure to close any unused pressure port with
Model No.

Asian DP2-20 DP2-60 DP2-80 DP2-21 DP2-41 DP2-61 DP2-22 DP2-42 DP2-62
the hexagon-socket-head plug supplied as accessory.
North American (Note) DP2-20F(-P) DP2-40N DP2-60N DP2-21F(-P) DP2-41N DP2-61N DP2-22F(-P) DP2-42N DP2-62N
Item European DP2-40E DP2-60E DP2-41E DP2-61E DP2-42E DP2-62E 䢇 Piping
• When connecting a 12mm spanner
Type of pressure Gauge pressure
hexagon-socket-head plug
Rated pressure range 0 toⳮ101.3kPa 0 to 100.0kPa 0 to 1.000MPa or coupling to the pressure
Set pressure range 5.1 toⳮ101.3kPa ⳮ5.0 to 100.0kPa ⳮ0.050 to 1.000MPa port, hold the hexagonal
part of the pressure port
Pressure withstandability 490kPa 1.47MPa
with a 12mm spanner and
Applicable fluid Non-corrosive gas make sure that the
Selectable units kgf/cm2, bar, psi, mmHg, inHg kgf/cm2, bar, psi tightening torque is 9.8N䡠m
or less. Also, in order to prevent any leakage, wind a sealing
Supply voltage 12 to 24V DC Ⳮ 10
ⳮ15 % Ripple P-P 10% or less
tape on the coupling when connecting.
Current consumption 50mA or less However, sealing tape is not required for North American type
<Asian, North American (Standard NPN output • Flat • IP67 types)> <North American (Standard PNP output type), European> using NPTF 1/8 coupling. (Sealing tape is required if NPT1/8
NPN open-collector transistor PNP open-collector transistor coupling is used.)
Comparative outputs
Comparative Output 1 • Maximum sink current: 100mA • Maximum source current: 100mA
(Comparative Output 2 ) • Applied voltage: 30V DC or less (between comparative output and 0V) • Applied voltage: 30V DC or less (between comparative output and ⳭV) Flat type, Light weight type
• Residual voltage: 1V or less (at 100mA sink current) • Residual voltage: 2V or less (at 100mA source current)
0.4V or less (at 16mA sink current) 䢇 Setting of pressure lead direction
• The pressure lead direction can be changed by dismantling
Equipped with 4 types of modes: hysteresis mode, window comparator mode, dual output mode, automatic sensitivity the pressure port attachment and changing the mounting
Output modes
setting mode (selectable by key operation) direction. The tightening torque of the hexagon-socket-head
Hysteresis 1 digit (However, variable in hysteresis mode and 2 digits when using psi unit) bolt (length 9mm or less) should be 0.29N䡠m or less.
Repeatability WithinⳲ0.2% F.S.Ⳳ1 digit Pressure port attachment
Response time 2.5ms or less Hexagon-socket-
head bolts
Short-circuit protection Incorporated
Main body
Output voltage (V)

Output voltage: 1 to 5V (over rated pressure range) 5


Zero-point: within 1VⳲ5% F.S.
High pressure
Analog voltage output Span: within 4VⳲ5% F.S.
(Positive pressure type)
Linearity: withinⳲ1% F.S. 1 High vacuum
Output impedance: 1kΩ approx.
Pressure (Vacuum pressure type) 2-"4.3mm mounting holes
Display 31/2 digit red LED display (Sampling rate: 4 times/sec. approx.)
Pressure port
Displayable pressure range 5.1 toⳮ101.3kPa ⳮ5.0 to 100.0kPa ⳮ0.050 to 1.000MPa
Analog bar display LED bar display in steps of 10% F.S. approx. 䢇 Piping
Operation Comparative Output 1 Orange LED (lights up when Comparative Output 1 is ON) • When connecting a cou-
indicators Comparative Output 2 Green LED (lights up when Comparative Output 2 is ON) pling to the pressure port,
hold the pressure port
Protection Standard • Flat • Light weight types: IP40 (IEC), IP67 type: IP67 (IEC) attachment with a 16mm
Ambient temperature ⳮ10 toⳭ50⬚C (No dew condensation or icing allowed), Storage: ⳮ10 toⳭ60⬚C (Light weight type: 10mm)
spanner and make sure that 16mm
Ambient humidity 35 to 85% RH, Storage: 35 to 85% RH
the tightening torque is (light weight type: 10mm)
Temperature characteristics Over ambient temperature range ⳮ10 toⳭ50⬚C: withinⳲ1% F.S. of detected pressure at 20⬚C 9.8N䡠m or less (Light weight spanner
Asian Standard • Flat • IP67 types: Rc (PT) 1/8 female thread, Light weight type: M5 female thread type: 1.47N䡠m or less). Also, in order to prevent any leakage,
Pressure wind a sealing tape on the coupling when connecting.
North American Standard type: NPTF 1/8 female thread, Flat • IP67 types: NPT 1/8 female thread
port
European Flat • IP67 types: G (PF) 1/8 female thread
Front case: ABS, Rear case: PPS (glass fiber reinforced), Display surface: Acrylic IP67 type
Material Pressure port attachment: Die-cast zinc alloy [Light weight type: POM (glass fiber reinforced), pressure port is brass (nickel plated)]
䢇 Piping for pressure measurement inlet port
Front cover (IP67 type only): Polycarbonate
• When connecting a coupling to the pressure measurement
Cable 0.15mm2 5-core oil resistant cabtyre cable, 2m long (IP67 type: 5m long) inlet port, hold the pressure port attachment with a spanner
Cable extension Extension up to total 100m is possible with 0.3mm2, or more, cable. and make sure that the tightening torque is 9.8N䡠m or less.
Also, in order to prevent any leakage, wind a sealing tape on
Weight Standard type: 95g approx., Flat type: 120g approx., IP67 type: 370g approx., Light weight type: 70g approx.
the coupling when connecting.
Accessories Hexagon-socket-head plug for pressure port: 1 No. (Standard type only), Pressure unit label: 1 No.
Note: Model Nos. of North American standard type having the suffix ‘-P’ are PNP output type. 䢇 Piping for atmospheric pressure inlet port
• If there is a possibility of
2 CAUTIONS water entering into the Safe place,
䢇 Do not use during the initial transient time (0.5 sec.) after the sensor enclosure through where there is no
DP2 series is designed for use with non-corrosive gas. the atmospheric pressure exposure to water
power supply is switched on.
It cannot be used for liquid or corrosive gas. inlet port, connect a tube to Atmospheric
䢇 Do not run the wires together with high-voltage lines or power pressure inlet port
the atmospheric pressure
lines or put them in the same raceway. This can cause mal-
䢇 Use within the rated pressure range. inlet port through a M5 M5 coupling
function due to induction.
䢇 Do not apply pressure exceeding the pressure coupling and extend the
䢇 Cable extension, using 0.3mm2, or more, cable, should be other end of the tube to a
withstandability value. The diaphragm will get damaged and
100m or less overall. safe place. In this case,
correct operation shall not be maintained.
䢇 Avoid use of standard type, flat type and light weight type of ensure that this end of the Tube
䢇 Make sure to carry out the wiring in the power supply off tube does not get clogged.
sensor in places where steam and dust is excessive.
condition. Pressure measurement
䢇 Take care that the sensor does not come in direct contact inlet port
䢇 Take care that wrong wiring will damage the sensor. 䢇 Fitting of front cover
with water, oil, grease, or organic solvents, such as, thinner,
䢇 Verify that the supply voltage variation is within the rating. etc. • Insert the bosses on the front cover into the guide holes at
䢇 If power is supplied from a commercial switching regulator, 䢇 Do not insert wires, etc, into the pressure port. The the bottom of the pressure port attachment, and push in the
ensure that the frame ground (F.G.) terminal of the power diaphragm will get damaged and correct operation shall not direction of the arrow to fit the hook.
supply is connected to an actual ground. be maintained. When removing the front cover, release the hook first.
䢇 In case noise generating equipment (switching regulator, 䢇 Do not operate the keys with pointed or sharp objects. Hook
inverter motor, etc.) is used in the vicinity of this sensor,
connect the frame ground (F.G.) terminal of the equipment to
an actual ground.

Guide holes

Downloaded from www.Manualslib.com manuals search engine


4 MOUNTING OF STANDARD TYPE SENSOR 6 FUNCTIONAL DESCRIPTION OF 9 OUTPUT MODES & THEIR CHARACTERISTICS
䢇 A sensor mounting bracket MS-DPX and MS-DPX-4 OPERATION PANEL Hysteresis mode ( )
(optional) may be used.
1 DP2-20 䢇 The common hysteresis of the comparative outputs can be
When mounting the sensor with the sensor mounting bracket, –101.3kPa
etc., the tightening torque should be 1.2N·m or less. 3 set, as desired, with the set values.
2
<MS-DPX> Hysteresis
M4 (length 6mm) pan head screw ON
(attached with MS-DPX) Comparative
OUT1 OUT2 Output 1 OFF Hysteresis: 1 digit or more
Spring washer 2 digits or more when
(attached with MS-DPX) MODE Comparative
ON (
using psi unit )
0-ADJ Output 2 OFF
Sensor mounting bracket 6
MS-DPX (optional)
4 0 Set Value 1 (P1) Set Value 2 (P2)
High pressure
5 (Positive pressure type)
06H0129 High vacuum
(Vacuum pressure type)
<MS-DPX-4>
Description Function
31/2 digit LED Displays measured pressure, settings, error mes- Window comparator mode ( )
Sensor mounting 1 display (Red) sages and key-protect status.
bracket 䢇 The comparative outputs can be turned ON or OFF by a pressure
Comparative Output 1
MS-DPX-4 (optional) 2 operation indicator Lights up when Comparative Output 1 is ON. which is within the pressure range set by Set Value 1 and Set Value 2.
(Orange)
Hysteresis Hysteresis
Comparative Output 2 Hysteresis: 1 digit
Spring washer M4 (length 6mm) pan head screw 3 operation indicator Lights up when Comparative Output 2 is ON. ON
2 digits when
(attached with MS-DPX-4) (attached with MS-DPX-4)
(Green) Comparative
Output 1
( )
using psi unit
• In the initial setting mode, pressing the OFF

In the sensing mode, if both the keys are pressed


key changes the settable digit. ON
Note: In case mounting brackets or screws other than the sensor
• In the Set Value 1, 2 modes, pressing Comparative
mounting bracket shown in the fiqure above are used, the length the key changes the set value to the Output 2 OFF
of the screws inserted into the pressure port attachment should high pressure side in case of positive
Increment key

simultaneously, zero-point adjustment is done.


be 5mm or less. If the length of the screws is longer than 5mm, 4 ( )
pressure type sensor and to the high 0 Set Value 1 (P1) Set Value 2 (P2)
the sensor may be damaged. vacuum side in case of vacuum pressure High pressure
type sensor. (Positive pressure type)
• In the sensing mode, if the key is pressed High vacuum
continuously for 4 sec. or more, the display (Vacuum pressure type)
䢇 Panel mounting bracket MS-DPX-2 (optional) and a front shows peak hold value.
protection cover DPX-04 (optional) are also available.
• In the initial setting mode, pressing the
key changes the set conditions. • When operating in window comparator mode ( ) Set Value
• In the Set Value 1, 2 modes, pressing 1 (P1) and Set Value 2 (P2) should be set with a difference
the key changes the set value to the of 3 digits or more. However, when the pressure unit is set
Front protection cover low pressure side in case of positive
Decrement key pressure type sensor and to the low to ‘psi’, the difference should be 6 digits or more.
DPX-04 (optional) DP
2-2
0

–10
1.3k
Pa
5 ( ) vacuum side in case of vacuum pressure
OU

type sensor.
T1

OU
T2

• In the sensing mode, if the key is pressed


0-A
DJ
MO
DE

continuously for 4 sec. or more, the display Dual output mode ( )


shows bottom hold value.
• Pressing the key changes the selected mode to 䢇 The outputs can be put to different use such as detection of
sensing mode, Set Value 1 (P1) set mode and Set different kinds of objects, control function, alarm function etc.
Panel mounting bracket Value 2 (P2) set mode. Hysteresis
MS-DPX-2 (optional) • In the sensing mode, if the key is pressed continu-
6 ously for about 3 sec., key-protect can be ON
(Suitable for 1 to 3.2mm thick panel) Mode selection key
Comparative Hysteresis: 1 digit
set/released.
( MODE ) 2 digits when
• In the sensing mode, if the mode selection key is
pressed while pressing the increment key ( ) , the
Output 1 OFF
(
using psi unit )
ON
5 I/O CIRCUIT DIAGRAM initial setting mode is obtained. Comparative
Output 2 OFF
NPN output type 7 ERROR MESSAGES 0 Set Value 1 (P1) Set Value 2 (P2)
Color code 䢇 When an error occurs, take the following corrective action. High pressure
(Positive pressure type)
D (Brown)ⳭV High vacuum
Error message Cause Corrective action
(Gray) Analog (Vacuum pressure type)
Sensor circuit

1kΩ voltage output


(Note)
Load Overcurrent due to short- Switch off the power supply
Z D1 (Black) Comparative Output 1 Load Ⳮ 12 to 24V DC circuit. and check the load.
Tr1 Automatic sensitivity setting mode ( )
ZD2 100mA max. ⳮ Ⳮ 10
ⳮ15 %
0
Applied pressure at the
(White) Comparative Output 2
pressure port should be 䢇 Using actual objects, if the pressure values for OK objects
100mA max. Pressure is being applied
Tr2 ZD3 (Blue) 0V during zero-point adjust-
brought to atmospheric and NG objects are input, then the sensor is automatically set
pressure and zero-point to the optimum pressure value (mid-value).
0 ment.
0 adjustment should be done
Internal circuit Users’ circuit again. Hysteresis
pressure

Applied pressure exceeds ON


Positive

Note: The analog voltage output is not incorporated with a short-circuit the upper limit of dis- Comparative
type

protection circuit. Do not directly connect a power supply or a playable pressure range. Output 1 OFF Hysteresis: 1 digit
2 digits when
capacitive load. When using the analog voltage output, take ON (
using psi unit )
pressure

0
Vacuum

care to connect external equipment of proper input impedance. Applied pressure exceeds Comparative
0
Also, when a cable extension is used, voltage drop due to cable the lower limit of display- Output 2 OFF
type

able pressure range. Applied pressure should be


resistance should be taken into account. 0 Set Value 1 (P1)
brought within the rated Set Value 2 (P2)
Set Value 3 (P3) High pressure
pressure

Applied pressure exceeds pressure range.


Positive

Symbols …D: Reverse supply polarity protection diode the lower limit of display- (Positive pressure type)
type

ZD1, ZD2, ZD3: Surge absorption zener diode able pressure range. High vacuum
Tr1, Tr2: NPN output transistor (Vacuum pressure type)
pressure

0
Vacuum

0 Applied pressure exceeds


the upper limit of dis- 10 PRESSURE UNITS
type

playable pressure range.


PNP output type 䢇 The pressure unit can be selected as per customer’s
requirement.
Color code 8 ANALOG BAR DISPLAY 䢇 In case of positive pressure type, the pressure unit can be
D (Brown)ⳭV
䢇 Pressure changes are displayed in an analog fashion by using LED bars. changed from International System of Units (SI) ‘kPa’ or
Hence, any sudden changes in pressure can be detected at a glance. ‘MPa’ to ‘kgf/cm2’, ‘bar’ or ‘psi’. In case of vacuum pressure
Tr1
䢇 The analog bar display shows the measured pressure, irrespective type, the pressure unit can be changed from International
Sensor circuit

ZD1 (Black) Comparative Output 1


Tr2 ZD2 100mA max.
Ⳮ 12 to 24V DC of the pressure unit, in steps of 10% F.S. approx. System of Units (SI) ‘kPa’ to ‘kgf/cm2’, ‘bar’, ‘psi’, ‘mmHg’ or
(White) Comparative Output 2
Ⳮ10 • Please refer to 11 SETTING 2 Initial setting for the ‘inHg’.
1kΩ
100mA max.
Load ⳮ ⳮ15 % procedure to change to analog bar display. • When the pressure unit is changed, the set values and the
(Gray) Analog
voltage output (Note) Load measured value are automatically converted.
ZD3 (Blue) 0V Analog bar display for positive pressure type sensor • Please refer to 11 SETTING 2 Initial setting for the
procedure to change the pressure unit.
Internal circuit Users’ circuit Atm. pressure condition High pressure condition
International System
DP2-21 100kPa DP2-21 100kPa DP2-21 100kPa of Units (SI)
Note: The analog voltage output is not incorporated with a short-circuit
protection circuit. Do not directly connect a power supply or a OUT1 OUT2 OUT1 OUT2 OUT1 OUT2
inHg kPa (Note) kgf/cm2
capacitive load. When using the analog voltage output, take 0-ADJ
MODE
0-ADJ
MODE
0-ADJ
MODE

care to connect external equipment of proper input impedance.


Also, when a cable extension is used, voltage drop due to cable
resistance should be taken into account. 06H0132

Analog bar display for vacuum pressure type sensor


Symbols … D: Reverse supply polarity protection diode
ZD1, ZD2, ZD3: Surge absorption zener diode Atm. pressure condition High vacuum condition mmHg psi bar
Tr1, Tr2: PNP output transistor DP2-20 DP2-20 DP2-20
–101.3kPa –101.3kPa –101.3kPa
: Vacuum pressure type : Positive pressure type

OUT1 OUT2 OUT1 OUT2 OUT1 OUT2


Note: MPa in case of DP2-22䡺, DP2-42䡺 and DP2-62䡺.
MODE MODE MODE
0-ADJ 0-ADJ 0-ADJ

06H0133
Downloaded from www.Manualslib.com manuals search engine
11 SETTING 3 Setting of pressure values

䢇 If key-protect has been set, make sure to release key- For the case when output mode is set to either the hysteresis mode For the case when the output mode is set to automatic
protect before operating the keys. (Please refer to 15 KEY- ( ), window comparator mode ( ) or dual output mode ( ). sensitivity setting mode ( )
PROTECT FUNCTION for the procedure.)
䢇 The conditions which are set are stored in an EEPROM. 䢇 [Set Value 1 (P1)] and [Set Value 2 (P2)] of the comparative 䢇 Comparative outputs’ [Set Value 1 (P1)], [Set Value 2 (P2)]
Kindly note that the EEPROM has a life span and its outputs are set. and [Set Value 3 (P3)] are set.
guaranteed life is 100,000 write operation cycles.
• The setting of Set Value 2 (P2) with respect to Set Value 1 • The setting of Set Value 2 (P2) with respect to Set Value 1
Setting procedure (P1) can only be towards the high pressure side in case of (P1) can only be towards the high pressure side in case of
the positive pressure type sensor and only towards the high the positive pressure type sensor and only towards the high
1 2 3
Zero-point Initial Pressure Measurement vacuum side in case of the vacuum pressure type sensor. vacuum side in case of the vacuum pressure type sensor.
adjustment setting value setting • Set Value 1 (P1) and Set Value 2 (P2) can be made common • Set Value 3 (P3) is automatically set to the mid-value of Set
for all the output modes. However, when a changeover is Value 1 (P1) and Set Value 2 (P2). However, if Set Value 1
Adjust Set [Display], Enter Commence
zero-point [Output mode], Set Value 1 (P1) measurement made to the automatic sensitivity setting mode, since Set (P1) is set to a value on the vacuum pressure side for a
and [Unit] Set Value 2 (P2) on completion Value 3 (P3) has not been set, make sure to carry out the positive pressure type sensor or to the positive pressure side
Set Value 3 (P3) of setting pressure value settings for the automatic sensitivity mode. for a vacuum pressure type sensor, Set Value 3 (P3) is
automatically set to the mid-value of ‘zero’(atmospheric
1 Zero-point adjustment pressure) and Set Value 2 (P2). Further, if both, Set Value 1
Set to Set Value 1 (P1) set mode
(P1) and Set Value 2 (P2) are set to a value on the vacuum
䢇 The displayed pressure when the pressure port is left open is DP2-20 –101.3kPa
• In the sensing mode, press MODE key. pressure side for a positive pressure type sensor or to the
adjusted to zero. • and Set Value 1 (P1) which is being positive pressure side for a vacuum pressure type sensor, Set
Set to sensing mode set are displayed alternately. Value 3 (P3) is automatically set to ‘zero’(atmospheric
• The figure on the left shows the display of
OUT1 OUT2

pressure).
• The sensor will automatically enter the
MODE

a vacuum pressure type sensor when the • The automatically set Set Value 3 (P3) can be changed manually.
0-ADJ
DP2-20 –101.3kPa

sensing mode when power is supplied. pressure unit has been set to ‘kPa’. • Since display of error messages is not possible during
• The figure on the left shows the display pressure value setting in the automatic sensitivity setting
OUT1 OUT2
Displayed
MODE when the pressure unit and display are set mode, make sure that the sensor is used within the rated
0-ADJ
alternately
to ‘kPa’ and ‘digital display’, respectively. pressure range.
DP2-20 –101.3kPa

Perform zero-point adjustment OUT1 OUT2 Set to Set Value 1 (P1) set mode
MODE

DP2-20 –101.3kPa • Let the pressure port be at atmospheric


0-ADJ

DP2-20 –101.3kPa
• In the sensing mode, press MODE key.
pressure (i.e., no applied pressure condi- • and Set Value 1 (P1) which is being
tion), and press, simultaneously, the incre- set are displayed alternately.
OUT1 OUT2
Enter Set Value 1 (P1) OUT1 OUT2

• The figure on the left shows the display of


MODE ment and decrement keys continuously. MODE

• Enter using a vacuum pressure type sensor when the


0-ADJ

• key and key.


0-ADJ

is displayed and, when the finger is DP2-20 –101.3kPa

released, zero-point adjustment is • In case of the positive pressure type pressure unit has been set to ‘kPa’.
completed and the sensor returns to the sensor, if key is pressed once the set Displayed
sensing mode.
OUT1 OUT2

MODE
value changes towards the high pressure alternately
DP2-20 –101.3kPa • If pressure has been applied during zero-
0-ADJ
side by 1 digit and if key is pressed DP2-20 –101.3kPa

point adjustment, is displayed when once the set value changes towards the
the keys are pressed. Bring the applied low pressure side by 1 digit.
OUT1 OUT2 Displayed OUT1 OUT2

pressure to atmospheric pressure (i.e., no In case of the vacuum pressure type MODE
MODE
alternately
0-ADJ

applied pressure condition) and carry out sensor, if key is pressed once the set 0-ADJ

DP2-20 –101.3kPa

the zero-point adjustment once again. value changes towards the high vacuum
side by 1 digit and if key is pressed
2 Initial setting
OUT1 OUT2
once the set value changes towards the Enter Set Value 1 (P1)
MODE
0-ADJ low vacuum side by 1 digit.
䢇 Pressure [Unit], [Display] and [Output mode] of the • If key or key is pressed con- DP2-20 –101.3kPa
• Within the required permissible pressure
comparative outputs are set. tinuously, the set value changes quickly. range, having created a pressure state
which is nearest to the atmospheric
Set to initial setting mode • If the set pressure range is exceeded, OUT1 OUT2
pressure, press key.
either (upper limit exceeded) or MODE

DP2-20 –101.3kPa • In the sensing mode, press MODE key while (lower limit exceeded) is displayed.
0-ADJ

• The pressure value at the time of pressing


pressing key. key is entered as Set Value 1 (P1). Set
• Initial setting is displayed. Value 1 (P1) and are displayed
OUT1 OUT2 Displayed
• If sensor is being used for the first time, alternately.
MODE
Set to Set Value 2 (P2) set mode alternately
0-ADJ

is displayed. • If the set pressure range is exceeded,


• In the Set Value 1 (P1) set mode, press
DP2-20 –101.3kPa
either (upper limit exceeded) or
DP2-20 –101.3kPa
MODE
key. (lower limit exceeded) are displayed
and Set Value 1 (P1) is set automatically

OUT1 OUT2

and Set Value 2 (P2) which is MODE


to the upper or lower limit of the set
Set initial conditions OUT1 OUT2

being set are displayed alternately. 0-ADJ


MODE
pressure range.
• The settable digit blinks.
0-ADJ

DP2-20 –101.3kPa
• The setting of Set Value 1 (P1) can be
• The settable digit changes when key repeated several times in the Set Value 1
OUT1 OUT2
is pressed. Displayed (P1) set mode.
0-ADJ
MODE
alternately
DP2-20 –101.3kPa

06H0143 Set to Set Value 2 (P2) set mode


DP2-20 –101.3kPa
• Change the setting of each digit as desired. OUT1 OUT2

• The setting is changed when key is 0-ADJ


MODE
DP2-20 –101.3kPa
• In the Set Value 1 (P1) set mode, press
OUT1 OUT2 pressed. key.
MODE

0-ADJ
MODE
OUT1 OUT2
• and Set Value 2 (P2) which is being
Enter Set Value 2 (P2) MODE set are displayed alternately.
0-ADJ

DP2-20 –101.3kPa
• Using key and key, enter in a
Change with key. manner similar to that for entering Set
Value 1 (P1). Displayed
3rd digit 2nd digit 1st digit OUT1 OUT2

MODE • If the set pressure range is exceeded, alternately


0-ADJ

either (upper limit exceeded) or DP2-20 –101.3kPa

(lower limit exceeded) is displayed.


Unit Output mode Display Displayed OUT1 OUT2

MODE
: kPa or : Hysteresis : Digital alternately 0-ADJ

MPa mode display DP2-20 –101.3kPa

: kgf/cm 2 : Window : Analog


pressure type pressure type

comparator bar OUT1 OUT2


Enter Set Value 2 (P2)
mode display MODE

• Within the required permissible pressure


0-ADJ
Positive

: bar : Dual output DP2-20 –101.3kPa

mode range, having created a pressure state


which is nearest to the high pressure end
: psi : Automatic OUT1 OUT2

(for a positive pressure type sensor) or


• If the output mode has been set to the window comparator
MODE
sensitivity
setting mode
0-ADJ
the high vacuum end (for a vacuum
mode ( ) in the initial setting mode, Set Value 1 (P1) and pressure type sensor), press key.
Vacuum

: mmHg Set Value 2 (P2) should be set with a difference of 3 digits


Displayed • The pressure value at the time of pressing
or more. However, when unit is set to ‘psi’, the difference
alternately key is entered as Set Value 2 (P2). Set
: inHg should be 6 digits or more. Value 2 (P2) and are displayed
DP2-20 –101.3kPa

alternately.
06H0145
• If the set pressure range is exceeded,
Set to sensing mode OUT1 OUT2
either (upper limit exceeded) or
MODE
Set to sensing mode (lower limit exceeded) are displayed
DP2-20 –101.3kPa • Press MODE key. 0-ADJ

• Press key. and Set Value 2 (P2) is set automatically


• The sensor returns to sensing mode after DP2-20 –101.3kPa MODE

to the upper or lower limit of the set


OUT1 OUT2
the initial conditions have been set. • The sensor returns to sensing mode after pressure range.
0-ADJ
MODE • Since the initial conditions which have OUT1 OUT2
Set Value 1 (P1) and Set Value 2 (P2) • The setting of Set Value 2 (P2) can be
been set are stored in an EEPROM, they MODE have been set. repeated several times in the Set Value 2
• Since the values which have been set are
0-ADJ

are not erased even if the power supply is (P2) set mode.
switched off. stored in an EEPROM, they are not
• The figure on the left shows the display erased even if the power supply is
when the unit and display are set to ‘kPa’ switched off.
and ‘digital display’, respectively.
Downloaded from www.Manualslib.com manuals search engine
13 CONVERSION OF PRESSURE UNITS
䢇 In the DP2 series, the conversion to different units is automatically done on changing the setting of the pressure unit. However,
this conversion can also be obtained by multiplying the values by the coefficients given in the following table.
Set to Set Value 3 (P3) set mode
Conversion procedure
DP2-20 –101.3kPa • In the Set Value 2 (P2) set mode, press
MODEkey. • For example, if 2kPa is to be expressed in kgf/cm2, since 1kPa=1.01972⳯10ⳮ2kgf/cm2, 2kPa becomes 2⳯1.01972⳯10ⳮ2 c 0.020kgf/cm2
OUT1 OUT2 • and the automatically set Set Value Conversion table for pressure units
0-ADJ
MODE
3 (P3) are displayed alternately.
• In case mmHg
Set Value 1 (P1) = ⳮ50.0kPa kPa MPa kgf/cm2 bar psi inHg atm
(Torr)
Displayed Set Value 2 (P2) = ⳮ100.0kPa
alternately then
ⳮ50.0Ⳮ(ⳮ100.0) 1kPa 1 1⳯10ⳮ3 1.01972⳯10ⳮ2 1⳯10ⳮ2 1.45038⳯10ⳮ1 7.50062 0.2953 9.86923⳯10ⳮ3
DP2-20 –101.3kPa Set Value 3 (P3) = = ⳮ75.0kPa
2
Digits smaller than the displayed digits
OUT1 OUT2
(
are discarded. ) 1MPa 1⳯103 1 1.01972⳯10 1⳯10 1.45038⳯102 7.50062⳯103 0.2953⳯103 9.86923
MODE
0-ADJ

1kgf/cm2 9.80665⳯10 9.80665⳯10ⳮ2 1 9.80665⳯10ⳮ1 1.42234⳯10 7.35559⳯102 2.8959⳯10 9.67841⳯10ⳮ1


In case Set Value 3 (P3) is to be changed
1bar 1⳯102 1⳯10ⳮ1 1.01972 1 1.45038⳯10 7.50062⳯102 2.953⳯10 9.86923⳯10ⳮ1
• The automatically set Set Value 3 (P3) can be manually changed to
a value between Set Value 1 (P1) and Set Value 2 (P2). However
1psi 6.89473 6.89473⳯10ⳮ3 7.03065⳯10ⳮ2 6.89473⳯10ⳮ2 1 5.17147⳯10 2.036 6.80457⳯10ⳮ2
Set Value 3 (P3) cannot be set to a value on the vacuum pressure
side for a positive pressure type sensor or to the positive pressure
1mmHg
side for a vacuum pressure type sensor. 1.33322⳯10ⳮ1 1.33322⳯10ⳮ4 1.35951⳯10ⳮ3 1.33322⳯10ⳮ3 1.93368⳯10ⳮ2 1 3.9370⳯10ⳮ2 1.31579⳯10ⳮ3
(1Torr)
Hysteresis
ON 1inHg 3.3864 3.3864⳯10ⳮ3 3.4531⳯10ⳮ2 3.3864⳯10ⳮ2 0.4912 2.5400⳯10 1 3.342⳯10ⳮ2
Comparative
Output 1 Hysteresis: 1 digit
OFF Changeable
2 digits when 1atm 1.01325⳯102 1.01325⳯10ⳮ1 1.03323 1.01325 1.46960⳯10 7.60000⳯102 2.9921⳯10 1
Comparative
ON range (
psi unit is used )
Output 2 OFF 14 PEAK HOLD & BOTTOM HOLD FUNCTIONS 15 KEY-PROTECT FUNCTION
0 Set Value 1 (P1) Set Value 2 (P2) 䢇 Peak hold and bottom hold functions enable the display of the 䢇 Key-protect is a function which prevents any unintentional
Set Value 3 (P3) High pressure peak value (maximum pressure value in case of the positive change in the conditions which have been entered in each
(Positive pressure type) pressure type sensor and maximum vacuum pressure value setting mode by making the sensor not to respond to the key
High vacuum in case of the vacuum pressure type sensor) and the bottom operations.
(Vacuum pressure type) value (minimum pressure value in case of the positive
Setting of key-protect
• Enter using key and key. pressure type sensor and minimum vacuum pressure value in
DP2-20 –101.3kPa
case of the vacuum pressure type sensor) of the varying • In the sensing mode, press key
• In case of a positive pressure type
MODE
DP2-20 –101.3kPa

measured pressure. continuously for about 3 sec. and release


sensor, if key is pressed once, the set
OUT1 OUT2

value changes towards the high pressure 䢇 These functions are convenient for finding the pressure OUT1 OUT2
it immediately when is displayed.
MODE
0-ADJ

side by 1 digit and if key is pressed variation range or for determining the reference for pressure
0-ADJ
MODE • Key-protect is set and the sensor returns
once, the set value changes towards the settings. to the sensing mode.
Displayed low pressure side by 1 digit.
In case of a vacuum pressure type • Please note that the peak value and the bottom value data
alternately
is erased when it is no longer displayed.
sensor, if key is pressed once the set • Since the key-protect information is stored in an EEPROM,
DP2-20 –101.3kPa

value changes towards the high vacuum • The response time of the comparative outputs becomes
it is not erased even if the power supply is switched off.
slower during the peak hold and bottom hold display.
side by 1 digit and if key is pressed • Please take care to remember if the key-protect function
once the set value changes towards the
OUT1 OUT2

MODE has been set.


0-ADJ
low vacuum side by 1 digit. Peak hold display
• If key or key is pressed conti-
Release of key-protect
nuously, the set value changes quickly. Initiating peak hold display
• If the set pressure range is exceeded, • In the sensing mode, keep key pressed DP2-20 –101.3kPa • In the sensing mode, press key
MODE

either (upper limit exceeded) or


DP2-20 –101.3kPa

until is displayed. (4 sec. approx.) continuously for about 3 sec. and release
(lower limit exceeded) is displayed. When the finger is released after is OUT1 OUT2
it immediately when is displayed.
OUT1 OUT2

MODE
displayed, the peak value and are 0-ADJ
MODE
• Key-protect is released and the sensor
0-ADJ
displayed alternately. returns to the sensing mode.
Set to sensing mode
• If the applied pressure exceeds the
DP2-20 –101.3kPa • Press MODE key.
Displayed displayable pressure range, error
• The sensor returns to the sensing mode message ( or ) and are • When the keys are to be operated, make sure that key-
alternately
OUT1 OUT2 after Set Value 1 (P1), Set Value 2 (P2) displayed alternately. In this case, bring protect is released.
MODE and Set Value 3 (P3) have been set. DP2-20 –101.3kPa
back the applied pressure to within the
• Since the values which have been set are
0-ADJ

stored in an EEPROM, they are not erased


rated pressure range. 16 LABEL FOR CHANGE IN PRESSURE UNIT
OUT1 OUT2
• The figure on the left shows the display of
even if the power supply is switched off. 0-ADJ
MODE
a vacuum type sensor when the pressure 䢇 When a pressure unit other than ‘kPa’ or ‘MPa’ has been
unit has been set to ‘kPa’. selected in the initial setting mode, the label (supplied as
12 PROCEDURE FOR CHECKING SET VALUES accessory) which corresponds to the selected unit should be
䢇 The conditions which have been set in the initial setting and the stuck at the position shown in the figure below.
pressure settings can be checked by the following procedure.
Ending peak hold display
Procedure to check initial conditions Procedure to check set values Pressure unit label (accessory)
DP2-20 –101.3kPa • Press key.
[• Sensor returns to sensing mode.]

Set to sensing mode Set to sensing mode OUT1 OUT2

MODE
0-ADJ

Press MODE key


while pressing Press MODE key.
key.
Bottom hold display
Sensor enters initial setting mode Sensor enters Set Value 1 (P1) set mode
Initiating bottom hold display
• Initial conditions which have • Set Value 1 (P1) which has
• In the sensing mode, keep key pressed
DP2-20 –101.3kPa
been set are displayed and can been set is displayed and can DP2-20 –101.3kPa

be checked. be checked. until is displayed. (4 sec. approx.)


Press MODE key. Press MODE key. When the finger is released after is OUT1 OUT2
OUT1 OUT2

MODE displayed, the bottom value and are MODE


0-ADJ 0-ADJ
Sensor returns to sensing mode Sensor enters Set Value 2 (P2) set mode displayed alternately.
• Set Value 2 (P2) which has • If the applied pressure exceeds the
been set is displayed and can Displayed displayable pressure range, error
be checked. alternately message or and are
Press MODE key. DP2-20 –101.3kPa
displayed alternately. In this case, bring
back the applied pressure to within the
In case of auto- rated pressure range.
matic sensitivity OUT1 OUT2
• The figure on the left shows the display of
setting mode only 0-ADJ
MODE
a vacuum type sensor when the pressure
unit has been set to ‘kPa’.
SUNX Limited http://www.sunx.co.jp/
Sensor enters Set Value 3 (P3) set mode
• Set Value 3 (P3) which has Head Office
been set is displayed and can 2431-1 Ushiyama-cho, Kasugai-shi, Aichi,
be checked.
Press MODE key. Ending bottom hold display 486-0901, Japan
DP2-20 –101.3kPa • Press key. Phone: ⴐ81-(0)568-33-7211
Sensor returns to sensing mode [• Sensor returns to sensing mode.] FAX: ⴐ81-(0)568-33-2631
OUT1 OUT2

MODE
Overseas Sales Dept.
• Please note that if any key, except key, is pressed in any MODE 0-ADJ
Phone: ⴐ81-(0)568-33-7861
setting mode, the set conditions shall get changed. FAX: ⴐ81-(0)568-33-8591