The document provides specifications for a mobile LED OT light. It describes the light's intended function of illuminating surgical areas. Key requirements include operating on 220VAC with battery backup, producing over 120,000 lux intensity, having adjustable illumination and compatible lighting dome, and including a 2-year warranty. The bidder must meet all technical specifications and provide documentation and installation.
Исходное описание:
specification for OT light, single dome bacis,mobile
The document provides specifications for a mobile LED OT light. It describes the light's intended function of illuminating surgical areas. Key requirements include operating on 220VAC with battery backup, producing over 120,000 lux intensity, having adjustable illumination and compatible lighting dome, and including a 2-year warranty. The bidder must meet all technical specifications and provide documentation and installation.
The document provides specifications for a mobile LED OT light. It describes the light's intended function of illuminating surgical areas. Key requirements include operating on 220VAC with battery backup, producing over 120,000 lux intensity, having adjustable illumination and compatible lighting dome, and including a 2-year warranty. The bidder must meet all technical specifications and provide documentation and installation.
LED OT light (mobile) Manufacturer Brand Type/Model Country of origin 1 Description of function 1.1 OT lights also referred to as an surgical light or surgical lighthead is a medical device intended to assist medical personnel during a surgical procedure by illuminating a local area or cavity of the patient. A combination of several surgical lights is often referred to as a “surgical light system”. 2 Operation requirements 2.1 Should operate on 220VAC and shall have battery backup system for emergency during mains failure. 2.2 Movement should be easier of all the arms and wheels. 2.3 Temperature rise at the surgeon head level should be less than 2 o C. 3 Technical Specification 3.1 Lux intensity should not be less than 1,20,000 Lux 3.2 Light field diameter shall be above 24cm or better 3.3 Colour temperature should be between 4000 to 4500 K 3.4 Colour rendering index should not be less than 90. 3.5 Depth of illumination should not be less than 100 cm. 3.6 Should have illumination adjustment. 3.7 The light dome shall be compatible for laminar air flow. 3.8 Should have heavy base with atleast 5 swivel Casters. Casters should be of good quality. 4 System Configuration 4.1 LED OT light (mobile) with complete accessories 5 Accessories, spares and consumables 5.1 All standard accessories and parts required to operate the equipment, including all standard tools and lubrication materials, to be included in the offer. 6 Standards and safety requirements 6.1 Should have safety certificate from a compenent authority CE/FDA/STQC CB certificate/ STQC S certificate or valid detailed electrical and functional safety test report from ERTL. Copy of the certificate/ test report shall be produced along with the technical bid. 7 Warranty 7.1 Comprehesive warranty for 2 years after the date of acceptance. 8 Installation and commissioning 8.1 The bidder must arrange for the equipment to be installed and commissioned by certified or qualified personnel. 9 Documentation 9.1 Operating manual in English 9.2 Service manual in English 9.3 List of important spare parts and accessories with their part number and costing dder's Remarks