Вы находитесь на странице: 1из 1

HumaLyzer PRIMUS

Settings for Multipurpose reagents


Software Version 3.2, Date 2015-06

Kit: CREATININE liquicolor & auto-CREATININE liquicolor


REF: 10051 & 10052

Standard included in kit; Urine sample require pre-dilution


Note: 1+49 with dist. water, multiply result with factor 50

Test Parameters CREA


Method: Two Point
Unit: mg/dl
Temperature: 37°C
First Wavelength: 500nm
Second Wavelength: None
Blank: Water
Factor: 1000.00
Norms High: 1.10
Norms Low: 0.50
Delay Time: 30s
Test Time: 120s
Decimal Digits: 2
Aspirate Volume: 500 µl

Standard Test Parameters


Method: Linear
Number of Standard: 1
Repeat Times: 1
Concentration1: 2.00

Control Test Parameters


No. AV SD Batch
1 0.00 0.0
2 0.00 0.0

Human Gesellschaft für Biochemica und Diagnostica mbH


Max-Planck-Ring 21 - D-65205 Wiesbaden – Germany
Tel.: +496122 99880 Fax: +496122 9988100 - e-mail: human@human.de