Вы находитесь на странице: 1из 14

Standard DS-1® Fourth Edition, Volume 3 - ADDENDUM 4

Applies to First, Second and Third Printing

In this Volume 3 - Addendum 4, the following paragraphs have been modified:


Paragraph Page Paragraph Page
3.24.6.e ........................................................... 1 Appendix A.6 ................................................... 5
3.25.6.i ............................................................. 2 3.14.4.e .......................................................... 6
3.25.9 .............................................................. 2 3.14.4.f ........................................................... 6
Figure 3.25.2 ................................................... 3 3.15.4.d .......................................................... 6
Figure 3.25.3 ................................................... 3 Figure 3.14.3 ................................................... 7
Table 3.10 ....................................................... 4 Figure 3.14.4 ................................................... 7
Table 3.10.1 .................................................... 5 Figure 3.14.5 ................................................... 7

*See page 8 for additional changes.

Page 85
¶ 3.24.6.e - Add new paragraph
ADD:
e. Float bore dimensions: On stabilizers equipped with
float bores, the ID shall be free of flaws or pitting that
will interfere with the valve’s ability to seal. For box up
float bores without stress relief features and for box
down float bores, the float bore dimensions shall meet
the dimensions given in figure 3.25.2 and table 3.10. If
a baffle plate is to be used with the float valve the bore
shall be visually inspected for the presence of a baffle
recess as seen in figure 3.25.3. If a baffle plate is to
be used with a baffle plate recess (figure 3.25.3), the
baffle recess should be of sufficient depth to include
the baffle plate and the float bore dimensions shall
meet the dimensions given in figure 3.25.3 and table
3.10. If a baffle plate is to be used without a baffle
plate recess (figure 3.25.2) the float bore dimensions
shall meet the dimensions given in figure 3.25.2 and
the float bore depth shall not exceed the bore depth
listed in table 3.10 plus the height of the baffle plate.
Note: If the Float Bore Diameter (R) and Connection
Size do not match table 3.10 values, calculate the
Float Bore Depth (A) using table 3.10.1.

Issue Date: November 3, 2015 Page 1 of 14


Standard DS-1® Fourth Edition, Volume 3 - ADDENDUM 4
Applies to First, Second and Third Printing

Page 86
¶ 3.25.6.i
CURRENT: (Text to be removed is underlined) CHANGE TO: (Amended text is underlined)
i. Float bore dimensions: On bit subs equipped with i. Float bore dimensions: On subs equipped with float
float bores, the ID shall be free of flaws or pitting that bores, the ID shall be free of flaws or pitting that will
will interfere with the valves ability to seal. Float bore interfere with the valve’s ability to seal. For box up float
dimensions shall meet the float bore dimensions in bores without stress relief features and for box down
figure 3.25.2 and table 3.10. float bores, the float bore dimensions shall meet the
Note: If the Float Bore Diameter (R) and Connection dimensions given in figure 3.25.2 and table 3.10. If a
Size do not match table 3.10 values, calculate the baffle plate is to be used with the float valve the bore
Float Bore Depth (A) using table 3.10.1. shall be visually inspected for the presence of a baffle
recess as seen in figure 3.25.3. If a baffle plate is to
be used with a baffle plate recess (figure 3.25.3), the
baffle recess should be of sufficient depth to include
the baffle plate and the float bore dimensions shall
meet the dimensions given in figure 3.25.3 and table
3.10. If a baffle plate is to be used without a baffle
plate recess (figure 3.25.2) the float bore dimensions
shall meet the dimensions given in figure 3.25.2 and
the float bore depth shall not exceed the bore depth
listed in table 3.10 plus the height of the baffle plate.
Note: If the Float Bore Diameter (R) and Connection
Size do not match table 3.10 values, calculate the
Float Bore Depth (A) using table 3.10.1.

Page 87
¶ 3.25.9
CURRENT: (Text to be removed is underlined) CHANGE TO: (Amended text is underlined)
3.25.9 Magnetic Particle Body Inspection: Inspect the 3.25.9 Magnetic Particle Body Inspection: Inspect the
outside diameter from shoulder to shoulder in accordance outside diameter from shoulder to shoulder in accordance
with procedure 3.9, MPI Slip/Upset Inspection. Any crack, with procedure 3.9, MPI Slip/Upset Inspection. Any crack
regardless of orientation, shall be rejected. If the sub is shall be rejected. If the sub is manufactured from non-
manufactured from non-magnetic materials, procedure magnetic materials, procedure 3.17, Liquid Penetrant
3.17, Liquid Penetrant Inspection, shall be substituted for Inspection, shall be substituted for Magnetic Particle
Magnetic Particle Inspection. Inspection.

Issue Date: November 3, 2015 Page 2 of 14


Standard DS-1® Fourth Edition, Volume 3 - ADDENDUM 4
Applies to First, Second and Third Printing

Page 87 Page 87
Figure 3.25.2 Figure 3.25.3 - Add new figure
CHANGE TO: (Amended text is underlined) ADD:

INTERNAL SHOULDER DIMENSION:


IF R ≤ BORE, DISREGARD
IF BORE < R ≤ (BORE+0.5 in.), MINIMUM SHOULDER = 1/8 in.
OTHERWISE, MINIMUM SHOULDER = 1/4 in.

(NOTE: ID'S OF THE BORE AND BIT PIN MUST BE SMALL


ENOUGH TO HOLD THE VALVE ASSEMBLY)

INTERNAL SHOULDER

BORE

Corner
broken
45° max
1/8 in.
max
rad.
r
R 1/8 in.
max
30 rad.
A deg.
R
A

Figure 3.25.2 Float bore profile. Figure 3.25.3 Float bore profile with baffle plate recess.

Issue Date: November 3, 2015 Page 3 of 14


Page 212
Table 3.10
CHANGE TO: (Amended text is underlined)

Table 3.10 Float Bore and Float Valve Data

Bore Data Valve Data


Valve Assembly Connection Bore Diameter Bore Depth Valve Diameter Valve Length

Issue Date: November 3, 2015


Size Size (R, +1/64, -0) (A, ±1/16) (D) (LV)
(in) (in) (in) (in)
Box Down Box Up
Configuration Configuration
1R 2-3/8 Reg 1-11/16 9-1/8 9-13/32 1-21/32 5-7/8
1F-2R 2-7/8 Reg 1-15/16 10 10-1/4 1-29/32 6-1/4
2F-3R 3-1/2 Reg 2-7/16 10-1/2 10-1/2 2-13/32 6-1/2
4R 4-1/2 Reg 3-1/2 12-13/16 12-13/16 3-15/32 8-5/16
5R 5-1/2 Reg 3-29/32 14-3/4 15-1/16 3-7/8 9-3/4
4R 6-5/8 Reg 3-1/2 13-9/16 15-13/32 3-15/32 8-5/16
5F-6R 6-5/8 Reg 4-13/16 17 17-13/32 4-25/32 11-3/4
5F-6R 7-5/8 Reg 4-13/16 17-1/4 18-1/8 4-25/32 11-3/4
5F-6R 8-5/8 Reg 4-13/16 17-3/8 19-1/4 4-25/32 11-3/4
6F 8-5/8 Reg 5-23/32 20-1/4 20-9/16 5-11/16 14-5/8
1R NC 23 1-11/16 9-1/8 9-13/16 1-21/32 5-7/8
1F-2R NC 26 1-15/16 9-1/2 10-7/32 1-29/32 6-1/4
2F-3R NC 31 2-7/16 10-1/4 10-29/32 2-13/32 6-1/2
3 ½ IF NC 38 3-5/32 14-1/4 14-1/4 3-1/8 10
Applies to First, Second and Third Printing

4R NC 44 3-1/2 13-1/16 13-17/32 3-15/32 8-5/16


4R NC 46 3-1/2 13-1/16 13-3/4 3-15/32 8-5/16
4F NC 46 3-11/16 16-3/4 16-3/4 3-21/32 12
Standard DS-1® Fourth Edition, Volume 3 - ADDENDUM 4

4R NC 50 3-1/2 13-1/16 14-3/16 3-15/32 8-5/16


5R NC 50 3-29/32 14-1/2 15-3/32 3-7/8 9-3/4
5F-6R NC61 4-13/16 17-1/2 17-23/32 4-25/32 11-3/4
3F 3-1/2 FH 2-27/32 14 14 2-13/16 10
5F-6R 5-1/2 FH 4-13/16 17 17-7/32 4-25/32 11-3/4
5F-6R 5-1/2 IF 4-13/16 17 17-5/8 4-25/32 11-3/4

Page 4 of 14
6F 6-5/8 IF 5-23/32 19-7/8 20-1/16 5-11/16 14-5/8
Standard DS-1® Fourth Edition, Volume 3 - ADDENDUM 4
Applies to First, Second and Third Printing

Page 213
Table 3.10.1
CHANGE TO: (Amended text is underlined)

Table 3.10.1 Float Bore Calculation


(for cases not covered in Table 3.10)

To determine Float Bore Depth (A): Where:

1) Measure Float Bore Diameter (R) R = Bore Diameter

2) Obtain Valve Size and Valve Length (LV) LV = Valve Length


corresponding to (R) from table 3.10 and the Max Pin LP = Max Pin Length (table 3.8)
Length (LP) for the Sub’s connection from table 3.8.
3/16" = Margin for proper Pin and Valve Separation
3) Float Bore Depth
A = Bore Depth
Box Down: (A) ± 1/16" = LV + LP + 3/16"

Box Up: (A) ± 1/16" = LV + LP + 3/16" + minimum length needed to have all seals seated within the float bore1

1 See Section A.6 in the Appendix for more information on calculating float bore depth for box up configurations.

Page 328
Appendix A.6 - Add new paragraph
ADD:
A.6 Float Bore Depth Calculations for Box
Up Configurations

The float bore depth required to ensure all float valve Take Length of Box dimension calculated from either
seals are located within the float bore for box up con- equation A.22 or A.24:
figurations (Abu) can be calculated using the following
equations: Valve Stick Out = LV - (Abd - 0.0625 - Length of Box).............(A.25)

Additional Length Required =


Valve Stick Out - Top of Valve to 1st Seal...........(A.26)
.......(A.21)
Abu = Abd + Additional Length Required..............(A.27)
If X < R:
Length of Box =
Round Abu up to the nearest 1/32’’.

..........(A.22) Where:
If X ≥ R:
C = Gage point pitch diameter (in) (API Spec. 7-2)
tpr = Thread taper (in/ft) (API Spec. 7-2)
............................................(A.23) H = Non-truncated thread height (in) (API Spec. 7-2)
Srs = Root truncation (in) (API Spec. 7-2)
Length of Box = LBC + L..................................(A.24) LBC = Depth of box (in) (API Spec. 7-2 evaluated at
max tolerance)
R = Bore diameter (in) (DS-1 Table 3.10)
LV = Valve length (in) (DS-1 Table 3.10)
Abd = Box down configuration float bore depth (in)
(DS-1 Table 3.10)
Top of Valve to 1st Seal = Distance measured from the
top of the float valve to the top of the first seal (in)

Issue Date: November 3, 2015 Page 5 of 14


Standard DS-1® Fourth Edition, Volume 3 - ADDENDUM 4
Applies to First, Second and Third Printing

Page 62
¶ 3.14.4.e
CURRENT: (Text to be removed is underlined) CHANGE TO: (Amended text is underlined)
e. Pin Stress Relief Groove: The diameter and width of e. Pin Stress Relief Groove: Unless waived by the
the API pin stress relief groove (if present) shall be customer, all API connections NC38 and larger that
measured and shall meet the requirements of table make up to BHA components shall be equipped with
3.8 for drill collars and table 3.9.1 for HWDP. pin stress relief grooves or they shall be rejected. The
diameter and width of the API pin stress relief groove
shall be measured and shall meet the requirements
of table 3.8 for drill collars and table 3.9.1 for HWDP.
The pin stress relief groove length shall be measured
from the connection shoulder to the first full thread by
placing a metal rule on the thread taper and squared
against the connection shoulder, as depicted in figures
3.14.3, 3.14.4 and 3.14.5. “First full thread” is defined
as the thread that is closest to the pin shoulder and
reaches the same height and thread profile as the
second thread. The location of the first full thread
can be identified by rotating the profile gage until the
absolute minimum amount of light is visible between
the thread form and the profile gage.

Page 62
¶ 3.14.4.f
CURRENT: (Text to be removed is underlined) CHANGE TO: (Amended text is underlined)
f. Boreback: The diameter and length of the boreback f. Box Boreback: Unless waived by the customer, all API
cylinder (if present) shall be measured and shall meet connections NC38 and larger that make up to BHA
the requirements of table 3.8 for drill collars and table components shall be equipped with boreback boxes
3.9.1 for HWDP. or they shall be rejected. The diameter and length of
the boreback cylinder shall be measured and shall
meet the requirements of table 3.8 for drill collars and
table 3.9.1 for HWDP.

Page 67
¶ 3.15.4.d
CURRENT: CHANGE TO: (Amended text is underlined)
d. Determine the polarity of the existing magnetic field d. Determine the polarity of the existing magnetic field
(if any) in each end if the test piece using the pocket (if any) in each end if the test piece using the pocket
magnetometer. Mark each end “+” (positive) or “-” magnetometer. Mark each end “+” (positive) or “-”
(negative), whichever applies. The magnetizing coil (negative), whichever applies. The magnetizing coil
shall be placed on the connection so as to reinforce (not shall be placed on the connection so as to reinforce (not
oppose) any field already present. Magnetizing current oppose) any field already present. Magnetizing current
activation and magnetic particle solution application activation and magnetic particle solution application
shall be performed simultaneously. The solution shall shall be performed simultaneously. The solution shall
be distributed over the area described in paragraph be distributed over the area described in paragraph
3.15.3. The magnetizing current shall remain on for at 3.15.3. The magnetizing current shall remain on for at
least 2 seconds after the solution has been distributed. least 2 seconds after the solution has been distributed.
The solution shall be agitated before each application. The solution shall be agitated before each application.
Note: For tools racked in close proximity, the solution
shall be agitated within 3 minute intervals during
application.

Issue Date: November 3, 2015 Page 6 of 14


Standard DS-1® Fourth Edition, Volume 3 - ADDENDUM 4
Applies to First, Second and Third Printing

Page 61
ADD: Figures 3.14.3, 3.14.4, and 3.14.5

Figure 3.14.3 Thread not fully formed as seen with light Figure 3.14.4 Lay thread profile gage along thread taper and
showing between profile gage and thread. rotate around the thread form until absolute minimum light
is visible between the profile gage and first thread. Thread is
fully formed (First Full Thread).

Figure 3.14.5 Square scale at the point of the “First Full


Thread” and take the measurement from the shoulder side of
thread profile to pin shoulder.

Issue Date: November 3, 2015 Page 7 of 14


Standard DS-1® Fourth Edition, Volume 3 - ADDENDUM 4
Applies to First, Second and Third Printing

The current requirements regarding calibration in the Standard DS-1® Fourth Edition, Volume 3 stipulate that a tag
or sticker be attached to most calibrated devices/instruments. There is often oil, grease, dirt, etc. at rig sites and in
machine shops that can smudge, cover, or remove evidence of calibration on a tag or sticker attached to a calibrated
device. Particularly for devices that only require calibration every year or longer, it can be unrealistic to assume that
the tag or sticker will be legible with the required information for such a period of time. Because of this problem,
many inspection companies carry proof of calibration in the form of a certificate or other verifiable documentation
that shows the necessary information about the calibration of the device. If the device is traceable to verifiable proof
of calibration, this evidence should be satisfactory to ensure a properly calibrated device, whether or not a tag or
sticker is present or legible.

In this Volume 3 - Addendum 4, the following paragraphs have been modified with requirements regarding proof of
calibration.

Paragraph Page Paragraph Page


3.4.2 ............................................................... 9 3.17.2.d ........................................................... 9
3.5.2.b ............................................................ 9 3.17.2.e ......................................................... 12
3.6.2.c ............................................................. 9 3.19.2.e ......................................................... 10
3.6.2.g ............................................................ 9 3.19.2.h ........................................................... 9
3.7.6.1 ............................................................. 9 3.19.2.k .......................................................... 12
3.8.2.1.b ........................................................ 10 3.21.2.2 ......................................................... 12
3.8.2.2.b ........................................................ 10 3.22.2.2 ......................................................... 13
3.8.2.2.d .......................................................... 9 3.23.2.2 ......................................................... 13
3.9.2.b ........................................................... 10 3.23.2.3 ......................................................... 13
3.9.2.c ............................................................ 10 3.24.2 ............................................................ 13
3.9.2.d ............................................................. 9 3.26.2.c ............................................................ 9
3.10.2.b ......................................................... 10 3.27.2.c ............................................................ 9
3.11.2 ........................................................ 9, 11 3.29.3 ............................................................ 13
3.13.2.a ......................................................... 11 3.30.2.1.d ........................................................ 9
3.13.2.b ......................................................... 11 3.30.2.2.c ....................................................... 10
3.13.2.c .......................................................... 11 3.31.2.1.f ....................................................... 10
3.13.2.d ......................................................... 11 3.32.2.1.f ....................................................... 10
3.13.2.e ......................................................... 11 3.33.2.2.a ...................................................... 13
3.14.2.a ......................................................... 11 3.33.2.2.b ...................................................... 14
3.14.2.b ......................................................... 11 3.33.2.2.c ....................................................... 14
3.14.2.c .......................................................... 11 3.33.2.2.d ...................................................... 14
3.15.2.b ......................................................... 10 3.33.7.2 ......................................................... 14
3.15.2.f ............................................................ 9 3.36.2.2.a ...................................................... 10
3.16.2.b ......................................................... 12

Issue Date: November 3, 2015 Page 8 of 14


Standard DS-1® Fourth Edition, Volume 3 - ADDENDUM 4
Applies to First, Second and Third Printing

Paragraph Page # CHANGE TO: (Amended text is underlined)


3.4.2 32 ... The light intensity meter shall have a tag or sticker
affixed showing calibration in the past six months. If a tag
3.8.2.2.d 37
or sticker is not present, the meter’s unique identification
3.9.2.d 39 number shall be traceable to a proof of calibration docu-
3.11.2 41 ment. The tag, sticker, or proof of calibration document
3.15.2.f 67 shall show the date of calibration, the due date of the
3.17.2.d 69 next calibration, and the company and individual that
performed the most recent calibration.
3.19.2.h 73
3.26.2.c 87
3.27.2.c 88
3.30.2.1.d 100
CURRENT:
... Light intensity meter shall have a tag or sticker
attached showing calibration in the past six months. The
tag or sticker shall show the date of calibration, the due
date of the next calibration, as well as the company and
individual performing the calibration.

Paragraph Page # ADD TO END OF PARAGRAPH:


(Amended text is underlined)
3.5.2.b 33 / 33(d)
... If a tag or sticker is not present, the device’s unique
3.6.2.g 34
identification number shall be traceable to a proof of cali-
bration document. The tag, sticker, or proof of calibration
document shall show the date of calibration, the due date
of the next calibration, and the company and individual
that performed the most recent calibration.

Paragraph Page # CHANGE TO: (Amended text is underlined)


3.6.2.c 34 ... The instrument shall have a calibration tag or sticker
affixed. If a tag or sticker is not present, the instrument’s
CURRENT: unique identification number shall be traceable to a proof
... The instrument shall have a tag or sticker attached of calibration document. The tag, sticker, or proof of
showing the date of calibration, the due date of the next calibration document shall show the date of calibration,
calibration, as well as the company and individual per- the due date of the next calibration, and the company
forming the calibration. and individual that performed the most recent calibration.

Paragraph Page # ADD TO END OF PARAGRAPH:


(Amended text is underlined)
3.7.6.1 35
... A calibration tag or sticker shall be affixed to the gage
or micrometer. If a tag or sticker is not present, the gage’s
or micrometer’s unique identification number shall be
traceable to a proof of calibration document. The tag,
sticker, or proof of calibration document shall show the
date of calibration, the due date of the next calibration,
and the company and individual that performed the most
recent calibration.

Issue Date: November 3, 2015 Page 9 of 14


Standard DS-1® Fourth Edition, Volume 3 - ADDENDUM 4
Applies to First, Second and Third Printing

Paragraph Page # CHANGE TO: (Amended text is underlined)


3.8.2.1.b 37 ... The ten pound weight used for the lift test shall have
been calibrated in the past three years to standards trace-
3.9.2.c 39 able to the National Institute of Standards and Technology
3.19.2.e 73 (NIST), or an equivalent body. A calibration tag or sticker
3.36.2.2.a 118 shall be affixed to the weight. If a tag or sticker is not
present, the weight’s unique identification number shall
CURRENT:
be traceable to a proof of calibration document. The tag,
... The ten pound weight used for the lift test shall show sticker, or proof of calibration document shall show the
evidence of certification in the past three years and the date of calibration, the due date of the next calibration,
certification must be traceable to the National Institute of and the company and individual that performed the most
Standards and Technology (NIST) or equivalent body. ... recent calibration. ...
<last paragraph> A tag or sticker shall be attached to the <last paragraph> A calibration tag or sticker shall be
yoke [showing/ verifying] the date of calibration, the due affixed to the yoke. If a tag or sticker is not present, the
date of the next [calibration/test], as well as the company yoke’s unique identification number shall be traceable to
and individual performing the [calibration/test]. a proof of calibration document. The tag, sticker, or proof
of calibration document shall show the date of calibration,
the due date of the next calibration, and the company
and individual that performed the most recent calibration.

Paragraph Page # CHANGE TO: (Amended text is underlined)


3.8.2.2.b 37 ... The blacklight intensity meter shall have a tag or sticker
affixed showing calibration in the past six months. If a tag
3.9.2.b (4th bullet) 38 or sticker is not present, the meter’s unique identification
3.15.2.b 67 number shall be traceable to a proof of calibration docu-
3.30.2.2.c 100 ment. The tag, sticker, or proof of calibration document
shall show the date of calibration, the due date of the
CURRENT: next calibration, and the company and individual that
... The blacklight intensity meter shall have a tag or sticker performed the most recent calibration.
attached showing calibration in the past six months. The
tag or sticker shall show the date of calibration, the due
date of the next calibration, as well as the company and
individual performing the calibration.

Paragraph Page # CHANGE TO: (Amended text is underlined)


3.10.2.b 40 ... Linearity calibration shall be indicated by a tag or
sticker affixed to the unit. If a tag or sticker is not present,
3.31.2.1.f 103 the unit’s unique identification number shall be traceable
3.32.2.1.f 107 to a proof of calibration document. The tag, sticker, or
proof of calibration document shall show the date of
CURRENT:
calibration, the due date of the next calibration, and the
... Linearity calibration shall be indicated by a sticker or company and individual that performed the most recent
tag attached to the unit, showing the date of calibration, calibration.
due date of next calibration and signature and company
of the individual performing the calibration.

Issue Date: November 3, 2015 Page 10 of 14


Standard DS-1® Fourth Edition, Volume 3 - ADDENDUM 4
Applies to First, Second and Third Printing

Paragraph Page # CHANGE TO: (Amended text is underlined)


3.11.2 41 ... The lead gage shall have a tag or sticker showing
calibration in the past six months, and the standard lead
3.13.2.a 50 template shall have a tag or sticker showing calibration in
3.14.2.a 60 the past year. The profile gage [and angle gages] shall
have a tag or sticker affixed showing calibration in the
CURRENT:
past three years. If a tag or sticker is not present, the
... The lead gage shall show evidence of calibration in gages’ and template’s unique identification numbers shall
the past six months and the standard lead template shall be traceable to proof of calibration documents. The tag,
show evidence of calibration in the past year. The profile sticker, or proof of calibration documents shall show the
gage [and angle gages] shall show evidence of calibration date of calibration, the due date of the next calibration,
in the past three years. Calibration must be traceable and the company and individual that performed the most
to the National Institute of Standards and Technology recent calibration. Calibration shall also be traceable
(NIST), or equivalent body. ... to the National Institute of Standards and Technology
(NIST), or an equivalent body.

Paragraph Page # CHANGE TO: (Amended text is underlined)


3.13.2.b 50 ... The measuring devices shall have been calibrated
in the past six months to standards traceable to the
3.13.2.c 51 National Institute of Standards and Technology (NIST),
3.13.2.e 51 or an equivalent body. A calibration tag or sticker shall
3.14.2.b 60 be affixed to the devices. If a tag or sticker is not pres-
ent, the devices’ unique identification numbers shall be
3.14.2.c 61
traceable to proof of calibration documents. The tag,
CURRENT: sticker, or proof of calibration documents shall show the
... The measuring devices shall show evidence of cali- date of calibration, the due date of the next calibration,
bration in the last six months, traceable to the National and the company and individual that performed the most
Institute of Standards and Technology (NIST), or equiva- recent calibration. ...
lent body. ...

Paragraph Page # CHANGE TO: (Amended text is underlined)


3.13.2.d 51 ... An ID micrometer and setting standard for the ID mi-
crometer are also required. The measuring devices shall
CURRENT: have been calibrated in the past six months to standards
... An ID micrometer and setting standard for the ID traceable to the National Institute of Standards and Tech-
micrometer are also required and shall show evidence nology (NIST), or an equivalent body. A calibration tag or
of calibration in the last six months, traceable to the Na- sticker shall be affixed to the devices. If a tag or sticker
tional Institute of Standards and Technology (NIST), or is not present, the devices’ unique identification numbers
equivalent body. ... shall be traceable to proof of calibration documents. The
tag, sticker, or proof of calibration documents shall show
the date of calibration, the due date of the next calibration,
and the company and individual that performed the most
recent calibration. ...

Issue Date: November 3, 2015 Page 11 of 14


Standard DS-1® Fourth Edition, Volume 3 - ADDENDUM 4
Applies to First, Second and Third Printing

Paragraph Page # CHANGE TO: (Amended text is underlined)


3.16.2.b 68 ... Calibration shall be indicated by a tag or sticker affixed
to the unit. If a tag or sticker is not present, the unit’s
CURRENT: unique identification number shall be traceable to a proof
... Calibration shall be indicated by a sticker or tag, at- of calibration document. The tag, sticker, or proof of
tached to the unit, showing the date of calibration, due calibration document shall show the date of calibration,
date, and signature and company name of the individual the due date of the next calibration, and the company
performing calibration. and individual that performed the most recent calibration.

Paragraph Page # CHANGE TO: (Amended text is underlined)


3.17.2.e 69 A blacklight source and a blacklight intensity meter are
required. The blacklight intensity meter shall have a tag or
CURRENT: sticker affixed showing calibration in the past six months.
A blacklight source and an ultraviolet light meter are If a tag or sticker is not present, the meter’s unique iden-
required. The ultraviolet light intensity meter shall have tification number shall be traceable to a proof of calibra-
a tag or sticker attached showing calibration in the past tion document. The tag, sticker, or proof of calibration
six months. The tag or sticker shall show the date of document shall show the date of calibration, the due date
calibration, the due date of the next calibration, as well of the next calibration, and the company and individual
as the company and individual performing the calibration. that performed the most recent calibration.

Paragraph Page # CHANGE TO: (Amended text is underlined)


3.19.2.k 73 ... The thickness gage shall have been calibrated in the
past year to standards traceable to the National Institute
CURRENT: of Standards and Technology (NIST), or an equivalent
... The thickness gage shall show evidence of calibration body. A calibration tag or sticker shall be affixed to the
in the past year and the calibration must be traceable gage. If a tag or sticker is not present, the gage’s unique
to the National Institute of Standards and Technology identification number shall be traceable to a proof of cali-
(NIST), or equivalent body. ... bration document. The tag, sticker, or proof of calibration
document shall show the date of calibration, the due date
of the next calibration, and the company and individual
that performed the most recent calibration. ...

Paragraph Page # CHANGE TO: (Amended text is underlined)


3.21.2.2 76 A thermometer or temperature measuring device for
measuring the temperature of the phosphating solution
CURRENT: shall have been calibrated in the past year to standards
A thermometer or temperature measuring device for traceable to the National Institute of Standards and Tech-
measuring the temperature of the phosphating solution nology (NIST), or an equivalent body. A calibration tag or
that has been calibrated in the past year to standards sticker shall be affixed to the thermometer or temperature
traceable to the National Institute of Standards and measuring device. If a tag or sticker is not present, the
Technology (NIST), or equivalent body. thermometer’s or device’s unique identification number
shall be traceable to a proof of calibration document. The
tag, sticker, or proof of calibration document shall show
the date of calibration, the due date of the next calibra-
tion, and the company and individual that performed the
most recent calibration.

Issue Date: November 3, 2015 Page 12 of 14


Standard DS-1® Fourth Edition, Volume 3 - ADDENDUM 4
Applies to First, Second and Third Printing

Paragraph Page # CHANGE TO: (Amended text is underlined)


3.22.2.2 78 ... calibrated within the past year to standards traceable
to the National Institute of Standards and Technology
3.23.2.2 82 (NIST) or an equivalent body. A calibration tag or sticker
CURRENT: shall be affixed to the temperature sensor. If a tag or
... calibrated within the past year to standards traceable to sticker is not present, the temperature sensor’s unique
the National Institute of Standards and Technology (NIST) identification number shall be traceable to a proof of cali-
or equivalent body. A sticker or tag shall be affixed to bration document. The tag, sticker, or proof of calibration
the temperature sensor to display evidence of calibration. document shall show the date of calibration, the due date
of the next calibration, and the company and individual
that performed the most recent calibration.

Paragraph Page # CHANGE TO: (Amended text is underlined)


3.23.2.3 82 ... calibrated within the past year to standards traceable
to the National Institute of Standards and Technology
CURRENT: (NIST) or an equivalent body. A calibration tag or sticker
... calibrated within the past year to standards traceable shall be affixed to the hardness tester. If a tag or sticker
to the National Institute of Standards and Technology is not present, the hardness tester’s unique identification
(NIST) or equivalent body. A sticker or tag shall be affixed number shall be traceable to a proof of calibration docu-
to the hardness tester to display evidence of calibration. ment. The tag, sticker, or proof of calibration document
shall show the date of calibration, the due date of the
next calibration, and the company and individual that
performed the most recent calibration.

Paragraph Page # CHANGE TO: (Amended text is underlined)


3.24.2 84 ... The micrometer shall have been calibrated in the past
six months to standards traceable to the National Institute
3.29.3 92 of Standards and Technology (NIST), or an equivalent
CURRENT: body. A calibration tag or sticker shall be affixed to the
micrometer. If a tag or sticker is not present, the microm-
... The micrometer shall show evidence of calibration in
eter’s unique identification number shall be traceable to a
the past six months. Calibration must be traceable to the
proof of calibration document. The tag, sticker, or proof
National Institute of Standards and Technology (NIST),
of calibration document shall show the date of calibration,
or equivalent body. ...
the due date of the next calibration, and the company and
individual that performed the most recent calibration. ...

Paragraph Page # CHANGE TO: (Amended text is underlined)


3.33.2.2.a 111 Calibrated ring and plug gages meeting the requirements
of API 7-2 (latest edition). A calibration tag or sticker shall
CURRENT: be affixed to the gages. If a tag or sticker is not present,
Calibrated ring and plug gages meeting the requirements the gages’ unique identification numbers shall be traceable
of API 7-2 (latest edition). to proof of calibration documents. The tag, sticker, or proof
of calibration documents shall show the date of calibration,
the due date of the next calibration, and the company and
individual that performed the most recent calibration.

Issue Date: November 3, 2015 Page 13 of 14


Standard DS-1® Fourth Edition, Volume 3 - ADDENDUM 4
Applies to First, Second and Third Printing

Paragraph Page # CHANGE TO: (Amended text is underlined)


3.33.2.2.b 111 Dial caliper that has been calibrated within the past six
months to standards traceable to the National Institute of
CURRENT: Standards Technology (NIST), or an equivalent body. A
Dial caliper that has been calibrated within the past six calibration tag or sticker shall be affixed to the device. If
months to standards traceable to the National Institute a tag or sticker is not present, the device’s unique identi-
of Standards Technology (NIST). A tag or sticker shall fication number shall be traceable to a proof of calibration
be affixed to the device indicating the date of calibration, document. The tag, sticker, or proof of calibration docu-
the due date of next calibration, and the company and ment shall show the date of calibration, the due date of
individual performing the calibration. the next calibration, and the company and individual that
performed the most recent calibration.

Paragraph Page # CHANGE TO: (Amended text is underlined)


3.33.2.2.c 111 ... Each device shall have been calibrated to standards
traceable to the National Institute of Standards and Tech-
CURRENT: nology (NIST), or an equivalent body. A calibration tag or
... Calibrations must be traceable to the NIST or equiva- sticker shall be affixed to each device. If a tag or sticker
lent body, and a tag or sticker shall be affixed to each is not present, each device’s unique identification number
device indicating the date of calibration, the due date of shall be traceable to a proof of calibration document. The
next calibration, and the company and individual perform- tag, sticker, or proof of calibration document shall show
ing the calibration. the date of calibration, the due date of the next calibra-
tion, and the company and individual that performed the
most recent calibration.

Paragraph Page # CHANGE TO: (Amended text is underlined)


3.33.2.2.d 111 Internal and external taper gages that have been cali-
brated within the past six months to standards traceable
CURRENT: to the NIST, or an equivalent body. A calibration tag or
Internal and external taper gages that have been cali- sticker shall be affixed to each device. If a tag or sticker
brated within the past six months to standards traceable is not present, each device’s unique identification number
to the NIST. A tag or sticker shall be affixed to each device shall be traceable to a proof of calibration document. The
indicating the date of calibration, the due date of next tag, sticker, or proof of calibration document shall show
calibration, and the company and individual performing the date of calibration, the due date of the next calibra-
the calibration. tion, and the company and individual that performed the
most recent calibration.

Paragraph Page # CHANGE TO: (Amended text is underlined)


3.33.7.2 113 ... A calibration tag or sticker shall be affixed to the gage.
If a tag or sticker is not present, the gage’s unique identi-
CURRENT: fication number shall be traceable to a proof of calibration
... A tag or sticker shall be attached to the gage indicating document. The tag, sticker, or proof of calibration docu-
the test date, due date, company name and signature of ment shall show the date of calibration, the due date of
the individual performing the calibration. ... the next calibration, and the company and individual that
performed the most recent calibration. ...

<< END >>

Issue Date: November 3, 2015 Page 14 of 14

Вам также может понравиться