Вы находитесь на странице: 1из 14

COMPATIBLE AEC CHAMBERS

FOR AEC BOARD 734614xx: 4 CHANNEL ION


CHAMBER COMPATIBILITY NOTES
COMPATIBILITY: AEC Boards, 734614-xx, are compatible with Ion Chambers requiring / providing
the following connections.
CAUTION: Ensure board is correctly wired before applying power. Field Select and Reset outputs are
open collector and are easily damaged if wired to a low impedance supply.

Select= 0v low 734614-09


Output = + or -Ramp
734614-01 734614-05 734614-13
(Notes 3, 4)
Plug style 12 pin Function 12 pin Function 9 pin -D Function 12 pin Function

Field select - (C) 10 0V low 10 0v low 3 0v low 10 0v low


Field select - (L) 11 0V low 11 0v low 2 0v low 11 0v low
Field select - (R) 9 0V low 9 0v low 6 0v low 9 0v low
Reset / Start 8 0V low 8 0v low 4 0v low 8 0v low
Output signal 12 + Ramp 12 + Ramp 5 + Ramp 12 - Ramp
Bias Voltage 1/2 ------- 1/2 - 300v 1 ------- 1/2 +300v
+ 12 vdc supply 4 + 12 vdc 4 + 12 vdc 8 + 12 vdc 4 + 12 vdc
- 12 vdc supply 5 - 12 vdc 5 - 12 vdc 7 - 12 vdc 5 - 12 vdc
+ 45 vdc supply
3 ------- 3 +45 vdc ------- ------- 3 +45 vdc
(Note 1)
0 vdc supply 7 0 vdc 7 0 vdc 9 0 vdc 7 0 vdc
- 24 vdc supply
6 -24 vdc 6 -24 vdc ------- ------- 6 -24 vdc
(Note 2)

Select= 12v high 734614-04 734614-07 734614-11


734614-08 734614-12
Output = +Ramp (Note 5) Philips - old (Note 4)
Plug style 15 pin-D Function 12 Pin Function 12 Pin Function 9 pin -D Function 12 Pin Function

Field select - (C) 12 12v high 10 12v high 10 12v high 3 12v high 10 12v high
Field select - (L) 3 12v high 11 12v high 11 12v high 2 12v high 11 12v high
Field select - (R) 11 12v high 9 12v high 9 12v high 6 12v high 9 12v high
Reset / Start 4 12v high 8 12v high 8 12v high 4 12v high 8 12v high
Output signal 7 +Ramp 12 + Ramp 12 + Ramp 5 + Ramp 12 + Ramp
Bias Voltage 1 + 500v 1 + 500v 1 ------- 1 ------- 1 +300v
Bias Voltage 2 ------- 2 (+ 300v) 2 ------- ------- ------- 2 +300v
+ 12 vdc supply 14 +12 vdc 4 + 12 vdc 4 + 12 vdc 8 + 12 vdc 4 + 12 vdc
- 12 vdc supply 15 -12 vdc 5 - 12 vdc 5 - 12 vdc 7 - 12 vdc 5 - 12 vdc
+ 45 vdc supply
5 +45 vdc 3 + 45 vdc 3 ------- ------- ------- 3 + 45 vdc
(Note 1)
0 vdc supply 8 / 13 0 vdc 7 0 vdc 7 0 vdc 9 0 vdc 7 0 vdc
- 24 vdc supply
6 -24 vdc 6 -24 vdc 6 -24 vdc ------- ------- 6 -24 vdc
(Note 2)

Notes: (1) +45vdc supply is only present on boards having a 12 pin WECO or 15 pin D-Sub and a Bias Voltage.
(2) -24vdc is not available on boards having 9 pin D-Sub connectors.
(3) 734614-09, AEC board has been replaced by, 737998-00 AEC board for AID Chambers only.
(4) AEC Board with 9-pin D-Sub connectors: Jumpers, JW1 to JW8, reverse the L and R field selections.
In Normal position (jumper 1-2) - Pin 2 of J11 to J14 selects the left field.
In Reverse position (jumper 2-3) - Pin 6 of J11 to J14 selects the left field.
(5) 734614-04 uses four WECO to 15 pin D-Sub adapters, that are included with the board assembly

DWG NO: SUP73461400 REV C PAGE 1 of 7


COMPATIBLE AEC CHAMBERS
FOR AEC BOARD 734614xx: 4 CHANNEL ION

Select= 12v high 734614-10


734614-03 734614-06
Output = + or - DC (Note 4)
Plug style 12 pin Function 12 pin Function 9 pin -D Function

Field select - (C) 10 12v high 10 12v high 3 12v high


Field select - (L) 11 12v high 11 12v high 2 12v high
Field select - (R) 9 12v high 9 12v high 6 12v high
Reset / Start 8 ------- 8 ------- 4 0v low
Output signal 12 + DC 12 - DC 5 +DC
Bias Voltage 1/2 + 300v 1/2 + 300v 1 +300v
+ 12 vdc supply 4 + 12 vdc 4 + 12 vdc 8 +12 vdc
- 12 vdc supply 5 - 12 vdc 5 - 12 vdc 7 - 12 vdc
+ 45 vdc supply
3 + 45 vdc 3 + 45 vdc ------- -------
(Note 1)
0 vdc supply 7 0 vdc 7 0 vdc 9 0 vdc
- 24 vdc supply
6 -24 vdc 6 -24 vdc ------- -------
(Note 2)

Select= 24v high


734614-02
Output = +Ramp
Plug style 12 pin Function

Field select - (C) 10 24v high


Field select - (L) 11 24v high
Field select - (R) 9 24v high
Reset / Start 8 24v high
Output signal 12 + Ramp
Bias Voltage 1/2 + 300v
+ 12 vdc supply 4 + 12 vdc
- 12 vdc supply 5 - 12 vdc
+ 45 vdc supply
3 + 45 vdc
(Note 1)
0 vdc supply 7 0 vdc
- 24 vdc supply
6 - 24 vdc
(Note 2)

Notes: (1) +45vdc supply is only present on boards having a 12 pin WECO or 15 pin D-Sub and a Bias Voltage.
(2) -24vdc is not available on boards having 9 pin D-Sub connectors.
(3) 734614-09, AEC board has been replaced by, 737998-00 AEC board for AID Chambers only.
(4) AEC Board with 9-pin D-Sub connectors: Jumpers, JW1 to JW8, reverse the L and R field selections.
In Normal position (jumper 1-2) - Pin 2 of J11 to J14 selects the left field.
In Reverse position (jumper 2-3) - Pin 6 of J11 to J14 selects the left field.
(5) Connector Types on AEC board (734614-04 - uses four WECO to 15 pin D-Sub adapters, that are included with the board)

(for Philips only)

9 PIN D-SUB 12 PIN WECO 15 PIN D-SUB

DWG NO: SUP73461400 REV C PAGE 2 of 7


COMPATIBLE AEC CHAMBERS
FOR AEC BOARD 734614xx: 4 CHANNEL ION

Chamber 9 Pin 12 Pin


Manufacturer Preamplifier Model Compatibility ‘D-Sub’ ‘WECO’ Comments
Notes (Pg. 6)
Board # Board #
Preampli Cellule (i) N/A 734614-13 Requires +300vdc;
• negative ramp output
Apelem
A-B-C= 1-2-3

60667I (ii) 737998-00 734614-01 Check Preamp jumpers.


60917F, H, P (iii), (iv)

61116A, C, D Preamplifiers must be


61135A, B, C, D, G, J, N (iii), (iv) configured for:
61137A (iv) • active low field select
and reset
61150A (iv)

61154A, C, J, N (iii), (iv) • a positive ramp output


705147
705172
705185
705207

61135A, B, C, D, G, J, L, N For all 734614-11 734614-08 Check Preamp jumpers.


61137A, 61170A, C, J preamps
61140A, 61116A, C listed Preamplifiers must be
Advanced See note
61154A, C, J, L, N configured for:
Instrument (vi)
Development • active high field select
and reset
• a positive ramp output

61136A, C, E 734614-02 Used in ICX200 series


to verify Ion Chambers, GEMS
Compatible

Preamplifier must be
configured for:
• active +24V high field
select and reset
• a positive Ramp
output

A-B-C= 1-2-3

DWG NO: SUP73461400 REV C PAGE 3 of 7


COMPATIBLE AEC CHAMBERS
FOR AEC BOARD 734614xx: 4 CHANNEL ION

Chamber 9 Pin 12 Pin


Manufacturer Preamplifier Model Compatibility ‘D-Sub’ ‘WECO’ Comments
Notes (Pg. 6)
Board # Board #
61176C, 61170C, S, T with (v) 734614-10 N/A Preamplifier is
Advanced ICX3122 Ion Chamber configured for:
Instrument
• active high field select
Development
• active low reset select
(Continued)
• a positive DC output
SSMC 1001 Pre-Amplifier (vii) 734614-09 734614-01 Preamplifier must be
Claymount (Telamat PA 101 Pre-Amp) configured for:
(Comet AG) • active low field and
reset select
Solid State Chambers • positive Ramp output

XP3-VSC N/A 734614-12 Preamplifier must be


(Chamber and Preamp) configured for:
Comet Ion Chamber • positive Ramp output

A-B-C= 3-1-2

CPI CPI Vision System P.D. Tab No’s yet to be


model # xxxxxxx determined
Vision System
Platinum One P.D. model
InfiMed-Digital System #3801.401.023.5.0
Quantamat II (3 Field) N/A 734614-02 24 volt.
General Electric
(GEMS)
Quantamatic (3 Field) N/A 734614-12 12 volt

109xx & 119xx Chambers N/A 734614-13 Requires +300vdc;


Keithley Instruments
with PC-435 Integrator • negative ramp output

3 field Ion Chambers N/A 734614-02 24 Volt


Model #2252611-11 GEMS Compatible
Medys (GEMS) (1v/mR)
Model #2252611-20 A-B-C= 1-2-3
(4v/mR)

DWG NO: SUP73461400 REV C PAGE 4 of 7


COMPATIBLE AEC CHAMBERS
FOR AEC BOARD 734614xx: 4 CHANNEL ION

Chamber 9 Pin 12 Pin


Manufacturer Preamplifier Model Compatibility ‘D-Sub’ ‘WECO’ Comments
Notes (Pg. 6)
Board # Board #
9803 509 30202 Amplimat N/A 734614-04 Scopomat.
(Chamber and Preamp) (15 pin –D)
9803 509 10002 Amplimat N/A 734614-04 Bucky.
Philips Medical (Chamber and Preamp) (15 pin –D)
Systems
9803 509 10102 Amplimat N/A 734614-04 Child’s Bucky.
15 pin D-Sub Connector
(Chamber and Preamp.) (15 pin –D)
9803 509 50002 Amplimat N/A 734614-04 Chest Stand.
(Chamber and Preamp) (15 pin –D)
9803 509 30202 Amplimat N/A 734614-07 Scopomat.
(Chamber and Preamp)
Philips Medical 9803 509 10002 Amplimat N/A 734614-07 Bucky.
Systems (Chamber and Preamp)
12 Pin WECO Connector 9803 509 10102 Amplimat N/A 734614-07 Child’s Bucky.
(Chamber and Preamp.)
9803 509 50002 Amplimat N/A 734614-07 Chest Stand.
(Chamber and Preamp)
168-03-111-1 Preamp (viii) N/A 734614-05 Requires − 300V
See note.
with 3 field Ion Chamber (or +300V. See note.)
RMS Fischer Medical
41376G Preamp N/A 734614-12 Requires + 300V
Systems
with 3 field Ion Chamber.
Note: The 168-03-111-1 preamplifier may have later versions that mimic the function of the 41376G preamplifier.
Verify the bias voltage requirement with the customer to ensure the correct version of AEC board is chosen.
81 67 538 X1651 Preamp N/A 734614-03 3 field;
with 3 field Ion Chamber Positive DC output
Siemens AG
Adaptor III with X1560 N/A 734614-06 3 field;
Preamp. Negative DC output.
70900 preamplifier 734614-03
(DC output).
70901 preamplifier 734614-09 734614-01 Select + Ramp on
VacuTec
(Ramp output). chamber
Meßtechnik Gmbh.
70902 preamplifier (ix) 734614-09 734614-01 Must use Ramp Module,
(Digital output) VacuTec Part# 9020011
Note: The above Preamplifiers will use one of the following chambers: 70140, 70141, 70145, 70151

L R
A C
C
2 B 3

Field Selections
Normal (N) position
A, B, C = 2, 1, 3 = Fields L, C, R

DWG NO: SUP73461400 REV C PAGE 5 of 7


COMPATIBLE AEC CHAMBERS
FOR AEC BOARD 734614xx: 4 CHANNEL ION

ION CHAMBER COMPATIBILITY NOTES

The Apelem “Preampli Cellule” Preamp is compatible with the 734614-13 AEC board provided that
the following requirement is met:

(i) Apelem “Preampli Cellule”


• The AEC type setting (in the x-ray generator) must be set to Apelem using the console’s
AEC setup menu or CPI Genware software.
• Apelem’s J1 pin 7 must be connected to J1 pin 7 of the 734614-13 AEC board. The
negative ramp output is at Apelem’s J1 pin 8 connector pin.

AID compatible Chambers with Preamps come with a standard 9 pin D-Sub connector. If a D-Sub
connector is not available, chamber cable has flying leads or easy access to the preamp’s jumpers is
not available, then AEC boards with WECO connectors are used. All AID Preamps, listed in Table 1,
are compatible with ICX Series 100 and 900 Chambers with the following exceptions or inclusions:

(ii) AID Preamp 60667I


• Chambers ICX150 to ICX152 and ICX176 have chamber field configuration set as 1-2-3
(field II= pin 2 and field I= pin 3).

(iii) AID Preamps 60917F, H, 61135A and 61154C


• Chamber ICX176 has chamber field configuration set as 1-2-3
(field II= pin 2 and field I= pin 3).

(iv) AID Preamps 60917F, H, 61135C, D, J, 61137A, 61150A and 61154C, J


• Chambers ICX187, ICX192R, ICX904, ICX935R, ICX963 has chamber field configuration set as
3-1-2 (field II= pin 6 and field III= pin 2).

(v) AID Preamp 61170S, T


• Chambers ICX2122, ICX3122, ICX3954 are compatible.
• The older versions of the 61170S and 61170T Preamps had the OUTPUT (Positive or
Negative) which is not selectable and was specified at time of purchase.

(vi) For all the AID Preamps listed in Table 1


• Chamber D-Sub cables are provided by AID
• A 737998-00 AEC board can be used provided that the jumper settings to the Preamp
can be set to active low field select and reset, with a positive ramp.
• Preamp 61135L does not have a 9 pin D-Sub connector and therefore must use the
734614-08 AEC board having the WECO connector.

DWG NO: SUP73461400 REV C PAGE 6 of 7


COMPATIBLE AEC CHAMBERS
FOR AEC BOARD 734614xx: 4 CHANNEL ION

(vii) The following statements apply to the compatibility of the chambers to the preamps:
• The Claymount Preamp SSMC 1001 is a direct replacement/equivalent for the Comet AG
Preamp PA 101.
• Both the above Preamps allow connection to either the Claymount SSMC 500 series Solid
State Chambers or the Comet AG “Telamat” series Solid State Chambers.

RMS Fischer Medical Systems 168-03-111-1 Preamp and Chamber is compatible provided that the
jumpers are correctly set as follows:

• Jumpers should be set to: 1-2, 5-6, 11-12 and 15-16.

VacuTec Preamp 70901, listed in Table 1, is compatible with 70140, 70141, 70145, 70151
Chambers. Chamber D-Sub cables are provided by VacuTec.

The 70140, 70141, 70145, 70151 Chambers may be used with Preamp 70902 if the following
requirement is met:

• A VacuTec Ramp Module, 9020011, is provided.

CONSULT FACTORY FOR COMPATIBILITY WITH CHAMBERS / PREAMPLIFIERS


NOT LISTED IN THIS DOCUMENT

DWG NO: SUP73461400 REV C PAGE 7 of 7


COMPATIBLE AEC CHAMBERS
FOR AEC BOARD 737992xx: 4 CHANNEL SOLID STATE

CHAMBER COMPATIBILITY NOTES


AEC Board, 737992-xx, is compatible with Solid State Chambers requiring / providing the following
connections.

Standard/Regular Dose
737992-00 737992-01 737992-02 737992-04
Plug & Function Weco Function Lemo (1) Function Lemo (2) Function Lemo (4) Function

3 + (Anode) 1 + 1 + 1 +
Field - Center
4 - (Cathode) 6 - 6 - 6 -
5 + (Anode) 2 + 2 + 2 +
* Field - Left 6 - (Cathode) 5 - 5 - 5 -
1 + (Anode) 3 + 3 + 3 +
* Field - Right 2 - (Cathode) 4 - 4 - 4 -
Shield 7 Ground Housing Ground Housing Ground Housing Ground

High Dose Range


737992-05 737992-06
Plug & Function Weco Function Lemo (1) Function

3 + (Anode) 1 +
Field - Center
4 - (Cathode) 6 -
5 + (Anode) 2 +
* Field - Left 6 - (Cathode) 5 -
1 + (Anode) 3 +
* Field - Right 2 - (Cathode) 4 -
Shield 7 Ground Housing Ground

* AEC Board Jumpers, JW1 to JW8, reverse the L and R field selections.
L R
In Normal (N) position;
Pins 5/6 of J11-J14, Weco (or Pins 2/5 of J1-J4, Lemo) connects to left field. A C
C
Pins 1/2 of J11-J14, Weco (or Pins 3/4 of J1-J4, Lemo) connects to right field.
In Reverse (R) position; 2 B 3
Pins 1/2 of J11-J14, Weco (or Pins 3/4 of J1-J4, Lemo) connects to left field.
Pins 5/6 of J11-J14, Weco (or Pins 2/5 of J1-J4, Lemo) connects to right field. 1

Field Selections
Normal (N) position
A, B, C = 2, 1, 3 = Fields L, C, R
Reverse (R) position
A, B, C = 3, 1, 2 = Fields R, C, L

DWG NO: SUP73799200 REV A PAGE 1 OF 2


COMPATIBLE AEC CHAMBERS
FOR AEC BOARD 737992xx: 4 CHANNEL SOLID STATE

CHAMBER COMPATIBILITY NOTES

Assembly Quantity &


Manufacturer Chamber Model Dose Range
No. Connector Type

1 – Lemo
737992-01 Standard/Regular
4 – Weco

SSMC 501 (Telamat B3-C)


2 – Lemo
SSMC 502 (Telamat B3-AC) 737992-02 Standard/Regular
4 – Weco
Claymount SSMC 503 (Telamat B3)
(Comet AG) SSMC 504 (Telamat B3A.12)
4 – Lemo
SSMC 505 (Telamat XB1) 737992-04 Standard/Regular
4 – Weco
SSMC 506 (Telamat XB3)

1 – Lemo
737992-06 High Dose
4 – Weco

All AEC Dosemat Solid-State 737992-00 4 – Weco Standard/Regular


Gilardoni
Chambers
737992-05 4 – Weco High Dose

Notes: (1) All Channels must be configured for NO preamplifier.


(2) When the 737992-xx AEC board is installed, the Chamber type must be set to ION in
the generator AEC setup menu.
(3) Connector Types on AEC board

LEMO WECO

DWG NO: SUP73799200 REV A PAGE 2 OF 2


COMPATIBLE ION AEC PREAMPLIFIERS / CHAMBERS
FOR AEC BOARD 737998-00: 4 CHANNEL ION

COMPATIBILITY: AEC Board, 737998-00, is compatible with Preamp / Chamber combinations


requiring / providing the following signals and levels.
CAUTION: Ensure board is correctly wired before applying power. Field Select and Reset outputs are
open collector and are easily damaged if wired to a low impedance supply.

737998-00
Description L R
J1 to J4
Function
D-Sub pin-out A C
Not Connected 1 None C
Field select - (C) 3 0v – Active Low 2 B 3
Field select - (L) 2* 0v – Active Low
Field select - (R) 6* 0v – Active Low 1
Reset / Start 4 0v – Active Low
Chamber Output signal 5 + Ramp Field Selections
- 12 vdc supply 7 - 12 vdc Normal (N) position
+ 12 vdc supply 8 + 12 vdc A, B, C = 2, 1, 3 = Fields L, C, R
0 vdc supply 9 0 vdc (common) Reverse (R) position
A, B, C = 3, 1, 2 = Fields R, C, L
∗ AEC Board Jumpers, JW1 to JW8, reverse the L and R field selections.
In Normal (N) position, pin 2 of J1 to J4 selects the left field.
In Reverse (R) position, pin 6 of J1 to J4 selects the left field.

PREAMPLIFIER COMPATIBILITY NOTES


Chamber
Manufacturer Preamplifier Model Compatibility Comments
Notes (Pg. 2)
60667I (i)
60917F, H, P (ii), (iii)
61135A, B, C, D, G, J, N (ii), (iii)
61137A (iii) Interfaces directly with 9-Pin
61150A (iii) Male D-Sub connectors
Advanced Instrument
61154A, C, J, N (ii), (iii)
Development
61170C, S, T (iv)
705147
705172
705185
705207 D-Sub socket on AEC Board
Claymount SSMC 1001 Pre-Amplifier (v) 737998-00
(Comet AG) (Telamat PA 101 Pre-Amp)
70901
VacuTec
70902 (vi)

Notes: (1) All preamplifiers must be configured for:


• active low field select and reset/start signals
• positive ramp output signal
(2) Interface cables (AID# 75035-xx, Ion Chamber Cable, 9 pin D-Sub, M-F, Straight-thru,
1 to 1 pin-out, or similar) are required to interface directly to the above preamplifiers.
(3) Due to mechanical interference, do not use right angle connectors to connect the
preamplifiers to the 737998-00 AEC board.

DWG NO: SUP73799800 REV A PAGE 1 of 2


COMPATIBLE ION AEC PREAMPLIFIERS / CHAMBERS
FOR AEC BOARD 737998-00: 4 CHANNEL ION

ION CHAMBER COMPATIBILITY NOTES


All AID Preamps, listed in Table 1, are compatible with ICX Series 100 and 900 Chambers with the
following exceptions or inclusions:

(i) AID Preamp 60667I


• Chamber ICX110 can be used if an AID# 75075-xx ICX110 Adapter Cable (14-pin AMP M
Series connector to 9-pin Male D-Sub connector) is used.
• Chambers ICX150 to ICX152 and ICX176 can be used provided that an AID# 75020-xx
“Continental 2-1-3 / 1-2-3 Conversion Ion Chamber Cable” is used (cable swaps pins 2 & 3).

(ii) AID Preamps 60917F, H, 61135A and 61154C


• Chamber ICX176 can be used provided that an AID# 75020-xx Conversion Cable is used.

(iii) AID Preamps 60917F, H, 61135C, D, J, 61137A, 61150A and 61154C, J


• Chambers ICX187, ICX192R, ICX904, ICX935R, ICX963 can be used if Jumpers JW1 to JW8
on the 737998-00 AEC board are set to Reverse (R). This will provide Normal (N), Left/Right
field orientation as described on Page 1.

(iv) AID Preamp 61170S


• Chambers ICX2122, ICX3122, ICX3954 are also compatible with Preamp 61170S.

(v) The following statements apply to the compatibility of the chambers to the preamps:
• The Claymount Preamp SSMC 1001 is a direct replacement/equivalent for the Comet AG
Preamp PA 101.
• Both the above Preamps allow connection to either the Claymount SSMC 500 series Solid
State Chambers or the Comet AG “Telamat” series Solid State Chambers.

VacuTec Preamp 70901, listed in Table 1, is compatible with 70140, 70141, 70145, 70151 Chambers.

The 70140, 70141, 70145, 70151 Chambers may be used with Preamp 70902 if the following
requirement is met:

• A VacuTec Ramp Module, 9020011, is provided.

CONSULT FACTORY FOR COMPATIBILITY WITH CHAMBERS / PREAMPLIFIERS


NOT LISTED IN THIS DOCUMENT

DWG NO: SUP73799800 REV A PAGE 2 OF 2


COMPATIBLE ION AEC PREAMPLIFIERS / CHAMBERS
FOR AEC BOARD 739389-00: 3 OF 5 FIELD ION

COMPATIBILITY: AEC Board, 739389-00, is compatible with Preamp / Chamber combinations


requiring / providing the following signals and levels.
CAUTION: Ensure board is correctly wired before applying power. Field Select and Reset outputs are
open collector and are easily damaged if wired to a low impedance supply.

ASSY 739389-00
Description J1 to J2
Function
D-Sub pin-out
Not Connected 1 None
Field select - (C) 3 0v – Active Low
Field select - (L) 2 (Note 1) 0v – Active Low
Field select - (R) 6 (Note 1) 0v – Active Low
Reset / Start 4 0v – Active Low
Chamber Output signal 5 + Ramp
- 12 vdc supply 7 - 12 vdc
+ 12 vdc supply 8 + 12 vdc
0 vdc supply 9 0 vdc (common)
No connection 10 No connection
Portrait 11 (Note 2) Portrait / Landscape output to chamber.
No connection 12 No connection
Inverted 13 (Note 2) Inverted / Not inverted output to chamber.
No connection 14 No connection
No connection 15 No connection

Notes: (1) AEC Board Jumpers, JW5 to JW8, reverse the L and R field selections.
In Normal (N) position, pin 2 of J1 to J4 selects the left field.
In Reverse (R) position, pin 6 of J1 to J4 selects the left field.
(2) FIELD SELECTION: The Portrait / Landscape and Inverted / Not inverted
output lines (active low) select the AEC fields per the following table:

PORTRAIT INVERTED LEFT FIELD CENTER FIELD RIGHT FIELD


Low High 1 2 3
High High 3 2 5
Low Low 5 2 4
High Low 4 2 1

Field Selections
Normal (N) Position

DWG NO: SUP73938900 REV B PAGE 1 of 2


COMPATIBLE ION AEC PREAMPLIFIERS / CHAMBERS
FOR AEC BOARD 739389-00: 3 OF 5 FIELD ION

PREAMPLIFIER COMPATIBILITY NOTES


Assembly Chamber
Manufacturer Preamplifier Model Compatibility Comments
Number Notes (Pg. 2)

Interfaces directly
Pre-Amp board 61135E with with 15-Pin
61141A Input Isolation
Male D-Sub
Board
Advanced connectors
Instrument 739389-00
Pre-Amp board 61135E with
Development 61165A Input Isolation
Board

61174A, D, S, T

VacuTec 139 00 03 739389-00 (i)


D-Sub socket
on AEC Board
Claymount SSMC 504 (D3/5-CN) 739389-01 739389-00

Notes: (1) All preamplifiers must be configured for:


• active low field select and reset/start signals
positive ramp output signal
(2) The 739389-00 AEC board may be connected to (2 only) active low, positive output, ramp
type, 3-field AEC chambers by use of a custom (15-Pin Male to 9 Pin Female) cable adaptor.
Refer to connector pin out for details (portrait and inverted signals are not connected).
Reference SUP73799800 for 3-field AEC preamp/chamber compatibility.

ION CHAMBER COMPATIBILITY NOTES


(i) Chamber must have a ramp module (Vacutec 903 00 11) or equivalent to convert digital AEC
pulses to an analog positive ramp signal.

CONSULT FACTORY FOR COMPATIBILITY WITH CHAMBERS / PREAMPLIFIERS


NOT LISTED IN THIS DOCUMENT

DWG NO: SUP73938900 REV B PAGE 2 OF 2