AD-4404 ........................................... Page 70
AD-4405........................................... Page 71
AD-4406 ........................................... Page 72
AD-4407........................................... Page 73
AD-4408A/C..................................... Page 74
AD-4531/2….................................... Page 75
AD-4532B........................................ Page 76
Others
M Series Moisture Analyzers........... Page 77
AD-4712........................................... Page 78
AD-4713........................................... Page 79
AD-4714........................................... Page 80
AD-4715 ........................................... Page 81
SV/SV-A Series................................ Page 82
3
Simple Calibration Flow Chart
End
AD-4212A/B Series
4
Simple Calibration Flow Chart
AD-4212C Series
5
Simple Calibration Flow Chart
WARM UP BALANCE FOR
AT LEAST 30 MINUTES
END
EK Series
6
Simple Calibration Flow Chart
WARM UP BALANCE FOR
AT LEAST 10 MINUTES
END
EK-G/EW-G Series
7
Simple Calibration Flow Chart
REMOVE COVER AND DISPLAYS
PRESS CAL SWITCH
009.798 GRAVITY
OR SLIDE SWITCH
(IF FITTED) SETTING
ENSURE P AN DISPLAYS
EMPTY AND REQUIRED
STABLE PRESS CALIBRATION
PRINT MASS
PLACE MASS ON
PAN AND PRESS DISPLAYS END
PRINT
END
EK-H Series
8
Simple Calibration Flow Chart
WARM UP THE DISPLAYS
BALANCE FOR 009.798 GRAVITY
30 MINUTES SETTING
RELEASE THE
CAL SWITCH DISPLAYS CAL 0
PRESS PRINT TO
PERFORM ZERO DISPLAYS A MASS
CALIBRATION
END
EK-i/EW-i Series
9
Simple Calibration Flow Chart
WARM UP BALANCE FOR
AT LEAST 30 MINUTES
END
EP-A Series
10
Simple Calibration Flow Chart
WARM UP BALANCE FOR
AT LEAST 30 MINUTES
END
EP-B Series
11
Simple Calibration Flow Chart
WARM UP BALANCE FOR
AT LEAST 30 MINUTES
CHECK DISPLAY IS
DISPLAYS CAL in
0.0000 & PRESS CAL
LEVER ON SIDE OF
GENTLY LOWER CAL
BALANCE
MASS
DISPLAYS CAL. . .
RETURNS TO NORMAL
DISPLAYS 0.0000
WEIGHING
END
ER Series
12
Simple Calibration Flow Chart
WARM UP BALANCE FOR
AT LEAST 30 MINUTES
END
EW Series
13
Simple Calibration Flow Chart
WARM UP BALANCE FOR
AT LEAST 30 MINUTES
(AT REAR OF
CHECK PAN EMPTY &
INSTRUMENT)
SLIDE CAL SWITCH ON
DISPLAYS CAL 0
END
FA/FX/FY Series
14
Simple Calibration Flow Chart
ENSURE BALANCE IS
LEVEL & WARMED UP
FOR 1 HOUR
END
FP Series
15
Simple Calibration Flow Chart
ENSURE BALANCE IS
LEVEL & WARMED UP
FOR 1 HOUR
DISPLAYS CAL
FOLLOWED BY LAST SET
CAL MASS
RETURNS TO NORMAL
REMOVE CAL MASS
WEIGHING
END
FR Series
16
Simple Calibration Flow Chart
Warm up balance for
at least 30 minutes
FZ-CT/FX-CT Series
17
Simple Calibration Flow Chart
Warm up balance for
at least 30 minutes
FZ-i/FX-i Series
18
Simple Calibration Flow Chart
Warm up balance for
at least 30 minutes
FZ-iWP/FX-iWP Series
19
Simple Calibration Flow Chart
PRESS AND HOLD
CAL UNTIL CAL out
IS DISPLAYED
RELEASE THE
CAL KEY DISPLAYS CAL 0
ENSURE P AN DISPLAYS
EMPTY AND REQUIRED
STABLE PRESS CALIBRATION
PRINT MASS
PLACE MASS ON
PAN AND PRESS DISPLAYS End
PRINT
RETURNS TO
REMOVE MASS WEIGHING MODE
NOTE IF GLP MODE
IN USE THEN
RESULTS ARE
OUTPUT FIRST
END
GF Series
20
Simple Calibration Flow Chart
WARM UP BAL ANCE
ENSURE PAN IS FOR AT
EMPTY LEAST 30 MINUTES
UNTIL CALout
PRESS & HOLD
IS DISPL AYED
CAL
THEN RELEASE
ZERO ST ORED
DISPLAYS CAL PRESS PRINT
MASS
REMOVE CAL
MASS
END
GF-K Series
21
Simple Calibration Flow Chart
WARM UP
BAL ANCE FOR AT
LEAST 1 HOUR
END
GH Series
22
Simple Calibration Flow Chart
WARM UP THE DISPLAYS
BALANCE FOR 009.798 GRAVITY
30 MINUTES SETTING
PRESS PRINT TO
PERFORM ZERO DISPLAYS A MASS
CALIBRATION
END
GP Series
23
Simple Calibration Flow Chart
PRESS AND HOLD
CAL UNTIL CAL out
IS DISPLAYED
RELEASE THE
DISPLAYS CAL 0
CAL KEY
ENSURE P AN DISPLAYS
EMPTY AND REQUIRED
STABLE PRESS CALIBRATION
PRINT MASS
PLACE MASS ON
DISPLAYS End
PAN AND PRESS
PRINT
RETURNS TO
WEIGHING MODE
REMOVE MASS
NOTE IF GLP MODE
IN USE THEN
RESULTS ARE
OUTPUT FIRST
END
GR Series
24
Simple Calibration Flow Chart
PRESS AND HOLD
CAL UNTIL CAL out
IS DISPLAYED
RELEASE THE
DISPLAYS CAL 0
CAL KEY
ENSURE P AN DISPLAYS
EMPTY AND REQUIRED
STABLE PRESS CALIBRATION
PRINT MASS
PLACE MASS ON
PAN AND PRESS DISPLAYS End
PRINT
RETURNS TO
REMOVE MASS WEIGHING MODE
NOTE IF GLP MODE
IN USE THEN
RESULTS ARE
OUTPUT FIRST
END
GX Series
25
Simple Calibration Flow Chart
WARM UP BALANCE
ENSURE P AN IS FOR AT
EMPTY LEAST 30 MINUTES
UNTIL CALout
IS DISPLAYED PRESS & HOLD
THEN RELEASE CAL
ZERO STORED
DISPLAYS CAL PRESS PRINT
MASS
REMOVE CAL
MASS
END
GX-K Series
26
Simple Calibration Flow Chart
PRESS RE-ZERO
ZERO CALIBRATION
DISPLAYS CAL 0
DISPLAYS CAL F
CALIBRATION
DISPLAYS CAL End
COMPLE TE D
RETURNS TO NORMAL
REMOVE CAL MASS
WEIGHING
END
HA Series
27
Simple Calibration Flow Chart
WARM UP BALANCE FOR
AT LEAST ONE HOUR
NO
SELECT CAL
MASS?
YES
ENTERS SPAN
CALIBRATION THEN PRESS RE-ZERO
DISPLAYS CAL END
END
HF/HF-G/HP Series
28
Simple Calibration Flow Chart
WARM UP BALANCE FOR
AT LEAST ONE HOUR &
ENSURE P AN IS EMPTY
DISPLAYS CC in
ENTERING CAL MODE PRESS PRINT
IMMEDIATELY
CALIBRATION
DISPLAYS tSt End
COMPLE TE D
END
HM Series
29
Simple Calibration Flow Chart
WARM UP BALANCE FOR
AT LEAST ONE HOUR &
ENSURE P AN IS EMPTY
CALIBRATION
DISPLAYS CAL End
COMPLE TE D
RETURNS TO NORMAL
REMOVE CAL MASS
WEIGHING
END
HR Series
30
Simple Calibration Flow Chart
DISPLAYS CAL F
CALIBRATION
DISPLAYS CAL End
COMPLE TE D
RETURNS TO NORMAL
REMOVE CAL MASS
WEIGHING
END
HX Series
31
Simple Calibration Flow Chart
WARM UP SCALE FOR AT
LEAST 30 MINUTES
CALIBRATION
SLIDE CAL SWITCH OFF
COMPLE TE D
END
FC Series
32
Simple Calibration Flow Chart
WARM UP SCALE
FOR AT LEAST 30
MINUTES
ENSURE P AN
DISPL AYS CAL F EMPTY & PRESS
ENTER
PRESS CAL
END
FC-i Series
33
Simple Calibration Flow Chart
WARM UP SCALE
FOR AT LEAST 30
MINUTES
ENSURE EMPTY
PAN AND WAIT
FOR STABLE
CALIBRATES &
PRESS MODE
DISPLAYS End
END
NEW FG Series
34
Simple Calibration Flow Chart
DISABLE AUTO OFF WARM UP SCALE FOR AT
FUNCTION BY PLACING
MASS ON PAN LEAST 10 MINUTES
DISPLAY SHOWS
GRAVITY VALUE PRESS CAL SWITCH
( TYPICAL 09.978)
CALIBRATION
PRESS CAL SWITCH
COMPLE TE D
END
FG Series
35
Simple Calibration Flow Chart
WARM UP THE DISABLE THE
SCALE FOR AT AUTO POWER
LEAST 30 MINS. OFF
2. PRESS ZERO
HERE
TO SET POUNDS
CAL MASS -
PRESS THE DISPLAYS
SHOWS CAL 2
CAL SWITCH CAL 0
ENSURE P AN DISPLAYS
EMPTY & STABLE CAL 1
PRESS ZERO
DISPLAYS
PRESS ZERO
CAL MASS
RETURNS TO
PRESS CAL
WEIGH MODE
END
FG-D Series
36
Simple Calibration Flow Chart
DISPLAY SHOWS
GRAVITY VALUE PRESS CAL SWITCH
(TYPICAL 09.978)
PRESS AUXILIAR Y
DISPLAYS CAL MASS BUTTON TO SELECT CAL
MASS
RESUMES NORMAL
PRESS ON/OFF
WEIGHING
END
FJ Series
37
Simple Calibration Flow Chart
“CAL” displayed Press & hold
calibration switch.
To calibrate with a
different value,
change the displayed
value using the 10-
key pad.
FS-i Series
38
Simple Calibration Flow Chart
ENSURE SCALE IS
WARMED UP FOR 30
MINUTES
SCALE CAN BE
PRESS UNITS TO SELECT
CALIBRATED WITH KG
CAL UNIT
OR LB MASS
END
FS/FS-KL Series
39
Simple Calibration Flow Chart
ALLOW 10 MINUTE WARM
UP PERIOD
YES
DISPLAY SHOWS CAL 1 FULL CAPACITY
(FULL SCALE CAL) CAL?
NO
DISPLAY SHOWS CAL 2 PRESS MODE TO SELECT
(2/3 FULL SCALE CAL) CAL 2
E ND
FV/FW Series
40
Simple Calibration Flow Chart
“CAL” displayed Press calibration
(CAL) switch
Display e.g.
Count
Press the ENTER
. CAL . key. The calibration
Weight Unit Weight
o value stops flashing.
CAL 0 5.000kg
Press ON/OFF to
exit CAL mode.
HC-i Series
41
Simple Calibration Flow Chart
DISABLE AUTO POWER
WARM UP SCALE FOR AT
OFF BY PLACING MASS
LEAST 30 MINUTES
ON PAN
DISPLAYS CAL F
END
HC/HD Series
42
Simple Calibration Flow Chart
ENSURE PAN
EMPTY
CALIBRATES
ZERO & DISPLAYS RELEASE KEY
100.0 g
END
HJ-150
43
Simple Calibration Flow Chart
DISABLE THE WARM UP SCALE
AUTO POWER FOR AT LEAST 10
OFF FUNCTION MINUTES
END
HL Series
44
Simple Calibration Flow Chart
ENSURE THE
SCALE IS OFF
ENSURE P AN
CAL F
EMPTY & ST ABLE
IS DISPLAYED
PRESS RE-ZERO
END
HL-i Series
45
Simple Calibration Flow Chart
DISABLE THE WARM UP SCALE
AUTO POWER FOR AT LEAST 10
OFF FUNCTION MINUTES
END
HL-WP Series
46
Simple Calibration Flow Chart
Turn scale ON
End
HT Series
47
Simple Calibration Flow Chart
ALLOW 10 MINUTE WARM
UP PERIOD
YES
DISPLAY SHOWS CAL 1 FULL CAPACITY
(FULL SCALE CAL) CAL?
NO
HV/HW Series
48
Simple Calibration Flow Chart
HV-G Series
49
Simple Calibration Flow Chart
HVG-WP Series
50
Simple Calibration Flow Chart
WARM UP THE DISABLE THE
SCALE FOR AT AUTO POWER
LEAST 10 MINS. OFF FEATURE
DISPLAYS
PRESS ZERO
CAL 0
ENSURE SCALE
DISPLAYS
UNLOADED &
CAL 1
STABLE THEN
NOTE FOR USA PRESS ZERO
PRESS MODE
HERE
TO SELECT
CALIBRATION IN
POUNDS, CAL 3
RETURNS
PRESS CAL TO WEIGHING
END
HW-D
51
Simple Calibration Flow Chart
HW-G Series
52
Simple Calibration Flow Chart
REMOVE COVER
LOWER DISPLA Y
AND PRESS CAL
SHOWS CAL
SWITCH
LOWER DISPLA Y
PRESS * KEY SHOWS GRAVITY
SETTING G9.XXX
LOWER DISPLA Y
PRESS C KEY
SHOWS CAL 0
XX REPRESENTS
LOWER DISPLA Y
THE SCALE MAX.
SHOWS CAL XX
CAPACITY
PLACE MASS ON
PAN AND WAIT
FOR STABLE
LOWER DISPLA Y
PRESS THE SHOWS -------
* KEY FOR 2 SECS. AND
THEN CAL END
SF/SG Series
53
Simple Calibration Flow Chart
Display off.
Press the
“CAL 0” displayed RE-ZERO key
Scale automaticlly
returns to weighing
mode.
SJ-H Series
54
Simple Calibration Flow Chart
DISABLE THE WARM UP SCALE
AUTO POWER FOR AT LEAST 10
OFF FUNCTION MINUTES
END
SK Series
55
Simple Calibration Flow Chart
DISABLE THE WARM UP SCALE
AUTO POWER FOR AT LEAST 10
OFF FUNCTION MINUTES
END
SK-WP Series
56
Simple Calibration Flow Chart
WARM UP SCALE FOR
AT LEAST ONE HOUR
END
AD-4316
57
Simple Calibration Flow Chart
WARM UP SCALE FOR AT
LEAST ONE HOUR
PRESS STANDBY/
OPERATE TO TURN
DISPLAY OFF
PRESS STANDBY/
OPE RATE
YES
SPAN AT MAX?
NO
PRESS STANDBY/
7 DECIMALS DISPLA YED
OPE RATE
END
AD-4321
58
Simple Calibration Flow Chart
WARM UP SCALE FOR AT
LEAST ONE HOUR
SET MAXIMUM
DISPLAYS
CAPACITY (nnnnn) &
CAP nnnnn
PRESS ENTER
END
AD-4322
59
Simple Calibration Flow Chart
WARM UP SCALE FOR AT
LEAST 30 MINUTES
PRESS STANDBY/
DISPLAYS PCAL 0
OPE RATE
DISPLAYS . . . . FOR A
SET CAL MASS TO BE
FEW SECONDS
USED, IF NOT MAX.
FOLLOWED BY CAL SPn
PLACE MASS ON SCALE
FOLLOWED BY
PRESS STANDBY/
XXXX - THE MAXIMUM
OPE RATE
CAPACITY
END
AD-4323
60
Simple Calibration Flow Chart
WARM UP SCALE FOR AT
LEAST 30 MINUTES
DISPLAY SHOWS
SLIDE CAL SWITCH ON
CAL d 01
ENSURE SCALE
DISPLAYS PCAL 0 UNLOADED & PRESS
ENTER
END
AD-4324
61
Simple Calibration Flow Chart
WARM UP SCALE FOR AT
LEAST 30 MINUTES
DISPLAY SHOWS
CAL d 01 SLIDE CAL SWITCH ON
ENSURE SCALE
DISPLAYS PCAL 0 UNLOADED & PRESS
ENTER
END
AD-4325
62
Simple Calibration Flow Chart
WARM UP SCALE FOR AT
LEAST 10 MINUTES
ENSURE SCALE
DISPLAYS CAL 0 UNLOADED & PRESS
ENTER
END
AD-4326/7
63
Simple Calibration Flow Chart
WARM UP SCALE FOR AT
LEAST 10 MINUTES
YES
CHANGE
DIVISION?
SELECT DIVISION
NO WITH TARE BUTTON
DISPLAYS CAP
YES
CHANGE
CAPACITY?
DISPLAYS CAL 0
DISPLAYS SPAn
END
AD-4328
64
Simple Calibration Flow Chart
DISPLAYS CAL in
PRESS CAL KEY FOR 2 SECONDS
THEN d-01
SELECT REQUIRED
DIVISION WITH
TARE KEY
DISPLAYS MAX.
PRESS PRINT KEY CAPACITY AS
C 10000
SET REQUIRED
MAXIMUM WITH
TARE AND ZERO
KEYS
ENSURE SCALE
DISPLAYS FXXXX
EMPTY AND NO MD
XXXX=CAP ACITY
THEN PRESS PRINT
PLACE MASS ON
SCALE AND PRESS DISPLAYS CAL End
PRINT WHEN NO
MOTION
AD-4329
65
Simple Calibration Flow Chart
WARM UP SCALE FOR AT
LEAST 30 MINUTES
MOMENTARILY SHORT
CLEAR ANY STORED
B10 TO B12 ON THE A/D
TARE ON THE A/D CARD
CARD I/O CONNECTOR
END
AD-4335/35C
66
Simple Calibration Flow Chart
CAL SEt
PRESS CAL DISPLAYED
CAL 0
PRESS ENTER DISPLAYED
PLACE MASS
PRESS ENTER ----- DISPLAYED
FOR 2 SECS.
DISPLAYS
CAL ENd
PRESS ENTER
TWICE TO EXIT
AD-4401
67
Simple Calibration Flow Chart
HOLD ENTER DISPLAYS
& PRESS THE MENU
USE TO SHOWS
SELECT CAL. Calibration mode
PRESS ENTER CAL G
SHOWS
PRESS ENTER Set Unit
0:None 1:g 2:kg 3:t
CONTINUED ON
NEXT PAGE
AD-4402
68
Simple Calibration Flow Chart
FROM
PREVIOUS
PAGE
PRESS ESC
SEVERAL TIMES
TO RETURN
TO WEIGHING
END
OF
PROCEDURE
AD-4402
69
Simple Calibration Flow Chart
WARM UP SCALE
FOR AT LEAST 30
MINUTES
PRESS TO
SELECT CAL &
PRESS ENTER
ENSURE EMPTY
PAN & ST ABLE
PRESS ENTER
PLACE MASS
WAIT STABLE
PRESS ENTER
ESC SEVERAL
TIMES TO STOP PRESS ESC
MODE
END
AD-4404
70
Simple Calibration Flow Chart
WARM UP SCALE
FOR AT LEAST 30
MINUTES
PLACE MASS
DISPLAYS CALEnd WAIT STABLE
PRESS ENTER
CALIBRATION DATA
STORED PRESS CAL
DISPLAYS CALoFF
PRESS ON/OFF
END
AD-4405
71
Simple Calibration Flow Chart
WARM UP SCALE
FOR AT LEAST 30
MINUTES
ENSURE P AN
CALIBRATES ZERO, DISPLAYS CAL F
EMPTY & STABLE
THEN FULL CAPACITY MASS
PRESS ENTER
IF NECESSARY
SELECT DIGIT WITH KEY SET CAL MASS
SET VALUE WITH KEY TO BE USED
PLACE MASS
DISPLAYS CALEnd WAIT STABLE
PRESS ENTER
CALIBRATION
DATA STORED PRESS CAL
DISPLAYS End
PRESS ON/OFF
END
AD-4406
72
Simple Calibration Flow Chart
WARM UP SCALE
FOR AT LEAST 30
MINUTES
ENSURE PAN
CALIBRATES ZERO, DISPLAYS CAL F EMPTY & STABLE
THEN FULL CAPACITY MASS PRESS ENTER
PLACE MASS
DISPLAYS CALEnd WAIT STABLE
PRESS ENTER
CALIBRATION DATA
STORED PRESS CAL
DISPLAYS CALoFF
PRESS ON/OFF
END
AD-4407
73
Simple Calibration Flow Chart
Warm up scale for
at least 30 minutes
Press ENTER
If necessary, set cal
mass to be used
Press ON/OFF
End
AD-4408A/C
74
Simple Calibration Flow Chart
WARM UP SCALE FOR AT
LEAST 30 MINUTES
UNIT REVERTS TO
NORMAL WEIGHING
END
AD-4531/2
75
Simple Calibration Flow Chart
Warm up scale for
at least 30 minutes
Displays CAL
Press key
Press ESC
End
AD-4532B
76
Simple Calibration Flow Chart
PRESS & HOLD
PROGRAM UNTIL DISPLAYS CH
CH
RETURNS TO
REMOVE MASS
WEIGH MODE
END
END
AD-4712
78
Simple Calibration Flow Chart
WARM UP SCALE FOR AT
LEAST 30 MINUTES
END
AD-4713
79
Simple Calibration Flow Chart
WARM UP SCALE FOR AT
LEAST 30 MINUTES
DISPLAYS REMOVE
RELEASE THE DOWN
PLATE FOLLOWED BY
ARROW KEY
CAL ZERO PUSH KEY
END
AD-4714
80
Simple Calibration Flow Chart
ALLOW 1 HOUR
WARM UP TIME
DISPLAYS
>CAL W:200.000g PRESS SELECT TO
PERFORMS SPAN CALIBRATE SPAN
CALIBRATION
DISPLAYS
CALIBRATION END
REMOVE MASSES
E ND
AD-4715
81
Simple Calibration Flow Chart
CAL IS PRESS & HOLD
DISPLAYED THE HOLD KEY
PRESS PRINT
PLACE STANDARD
VISCOSITY FLUID
MEASUREMENT IN SAMPLE CUP
STARTS
PRESS START
USE MODE,
START, STOP TO
CORRECT VALUE
PRESS PRINT
END
SV/SV-A Series
82
Simple Calibration Flow Chart
Simplified Calibration
(For the SV-10/1A/10A Only) PLACE PURIFIED
WATER IN THE
SAMPLE CUP
ENSURE
VISCOSITY & TEMP
ARE STABLE
RETURNS TO
End DISPLAYED
MEASURING
ON COMPLETION
STATUS
END
SV-10/1A/10A
83