Вы находитесь на странице: 1из 404

KODAK 9000 And KODAK 9000 3D Extraoral Imaging System

Service Guide
Notice

The Service Guide for the KODAK 9000 and KODAK 9000 3D Extraoral Imaging System includes information both
on the Panoramic features as well as the 3D features. We recommend that you thoroughly familiarize yourself with this
Guide in order to carry out any necessary intervention on this radiological unit.

The information contained in this Guide may be subject to modification without notice, justification or notification to
the persons concerned.

No part of this Guide may be reproduced without the express permission of Carestream Health, Inc.

The US Federal law restricts this device to sale by or on the order of a physician.
This document is originally written in English.

Manual Name: KODAK 9000 and KODAK 9000 3D Extraoral Imaging System, Service Guide.
Document Code: SM712
Revision Number: Edition 02
Print Date: 05/2009

The brand names and logos reproduced in this Guide are copyright.
KODAK is a trademark of KODAK used under License.
KODAK 9000 and KODAK 9000 3D Extraoral Imaging System, complies with Directive 93/42/CEE relating to
medical equipment.

0086
Contents
1—About This Guide
Conventions in this Guide . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1–1
Note to the User. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1–2
Warning and Safety Instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1–2
Marking and Labeling Symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1–4
Responsibility and Warranty . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1–5

2—UNIT DIAGRAMS
Unit Functional Diagram. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2–1
Unit Wiring and Interconnections Diagram . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2–3
Blade Collimator Interconnections Diagram. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2–5
Cable Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2–6
Cable N° 1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2–6
Cable N° 2 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2–7
Cable N° 4 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2–8
Cable N° 5 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2–9
Cable N° 7 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2–10
Cable N° 8 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2–11
Cable N° 9 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2–12
Cable N° 10 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2–13
Cable N° 13 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2–14
Cable N° 14 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2–15
Cable N° 16 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2–16
Cable N° 19 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2–17
Cable N° 22 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2–18
Cable N° 24 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2–19
Cable N° 25 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2–20
Cable N° 31 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2–21
Cable N° 35 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2–22
Cable N° 37 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2–23
Cable N° 40 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2–24
Cable N°41 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2–25
Cable N° 42 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2–26
Cable N° 43 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2–27
Cable N° 60 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2–28
Cable N° 964 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2–29
Cable N° 965 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2–30
Cable N° 966 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2–31

KODAK 9000 and KODAK 9000 3D Extraoral Imaging System_Service Guide (SM712)_Ed 02 1
Contents

Cable N° 973 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2–32


Cable N° 974 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2–33
Cable N° 979 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2–34
Cable N°981 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2–35
Cable N° 982 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2–36
Cable N° LZ054. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2–37
Cable N° LZ055. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2–38

3—LIST OF SPARE PARTS


Column Spare Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3–1
Column and Head Cover Spare Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3–2
Column and Head “Fiber” Covers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3–2
Column and Head “Plastic” Covers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3–3
Head Assembly Spare Parts. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3–4
Rotative Arm Spare Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3–6
Rotative Arm Cover Spare Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3–8
Rotative Arm “Fiber” Covers for “Sector Collimator” Spare Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3–8
Rotative Arm “Plastic” Covers for “Sector Collimator” Spare Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3–9
Rotative Arm “Fiber” Covers for “Blade Collimator” Spare Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3–10
Rotative Arm “Plastic” Covers for “Blade Collimator” Spare Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3–11
Fixed Arm Spare Parts. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3–12
Miscellaneous Spare Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3–13

4—DRILLING HOLES FOR THE PLASTIC COVER MOUNTING SUPPORT (Ref. SPCGA75, CSPCGA76)
Description. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4–1

5—USING THE TECHNICAL MENU


Description. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5–1
KODAK 9000 3D Tool Kit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5–1
Panoramic Technical Menu Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5–3
? Menu Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5–3
“About” Submenu Description. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5–3
“Details” Submenu Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5–4
“Log” Submenu Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5–4
Maintenance Menu Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5–5
“General” Submenu Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5–5
“Communication” Submenu Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5–6
“Preferences” Submenu Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5–6
“Manufacturing Parameters” Submenu Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5–7
“Calibration” Submenu Description. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5–7
“Reset of Total Parameters” Submenu Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5–8
“Save/Restore” Submenu Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5–8
3D Technical Menu Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5–9
? Menu Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5–9
Maintenance Menu Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5–9
“Calibration” Submenu Description. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5–10

2
Contents

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5–10
Using the Technical Menu. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5–11
Accessing the Technical Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5–11
Calibration Procedures . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5–12
“mA Calibration” . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5–12
“Heating Voltage Calibration” . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5–14
“Rotation Axis Calibration” (Only for 3D Sector Collimator) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5–16
“Activate the 3D Collimator Calibration” (Only for 3D Sector Collimator). . . . . . . . . . . . . . . . . . . . . . . . . 5–17
“Margin Calibration” (Only for 3D Sector Collimator). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5–19
“Rotative Arm Calibration” (Unit Serial N°s: Up to WKAH046, WKAE030 & WKAF007) . . . . . . . . . . . 5–20
“Rotative Arm Calibration” (Unit Serial N°s: From WKAH047, WKAE031 & WKAF008) . . . . . . . . . . . 5–23
“Tube Head Calibration” (X-Ray Without Movements) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5–26
“Control of the X-Ray Beam” (Panoramic Only) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5–27
“Dose Measurement on Incisors” (Panoramic Only). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5–28
“Reset of the Calibration Parameters” . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5–29
Using the Calibration Tool Software. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5–30
Calibrating the Blade Collimator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5–30
Calibrating the Rotation Axis . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5–32
Controlling the Image Quality . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5–37

6—INFORMATION MESSAGES

7—ERROR MESSAGES
E 1: Handswitch was Released Before the End of Exposure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7–1
Definition of the Cause. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7–1
Solution of the Cause . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7–1
E2: Mains Power Failure Detected . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7–2
Definition of the Cause. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7–2
Solution of the Cause . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7–2
E3: Kv Measurement Problem (Kv High). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7–4
Definition of the Cause. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7–4
Solution of the Cause . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7–4
E4: Kv Measurement Problem (kV Low) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7–5
Definition of the Cause. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7–5
Solution of the Cause . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7–5
E6: Electric Arc Detection in the Generator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7–7
Definition of the Cause. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7–7
Solution of the Cause . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7–7
E7: mA Measurement Problem (mA High) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7–9
E8: mA Measurement Problem (mA Low) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7–9
Definition of the Cause. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7–9
Solution of the Cause . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7–9
E9: Filament Heating Problem (mA High) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7–11
E10: Filament Heating Problem (mA Low) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7–11
Definition of the Cause. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7–11
Solution of the Cause . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7–11

KODAK 9000 and KODAK 9000 3D Extraoral Imaging System_Service Guide (SM712)_Ed 02 3
Contents

E11: Safety Device Activated on High Voltage Capacitors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7–13


Definition of the Cause . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7–13
Solution of the Cause . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7–13
E12: Problem with Rotation Reference Detector . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7–14
Definition of the Cause . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7–14
Solution of the Cause . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7–14
E15: Movement Too Long. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7–15
Definition of the Cause . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7–15
Solution of the Cause . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7–15
E16: Movement Too Short. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7–17
Definition of the Cause . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7–17
Solution of the Cause . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7–17
E19: Distance Too Short . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7–19
Definition of the Cause . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7–19
Solution of the Cause . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7–19
E20: Distance Too Long . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7–20
Definition of the Cause . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7–20
Solution of the Cause . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7–20
E22: Acquisition is stopped the Interface is Masked . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7–21
Definition of the Cause . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7–21
Solution of the Cause . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7–21
E23: No movement is detected . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7–22
Definition of the Cause . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7–22
Solution of the Cause . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7–22
E24: Heating Voltage Self-calibration Time-out . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7–24
E25: Heating Voltage Lower Limit Reached. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7–24
E26: Heating Voltage Higher Limit Reached . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7–24
Definition of the Cause . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7–24
Solution of the Cause . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7–24
E27: No Detection of the Limit Switches X or Y . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7–26
Definition of the Cause . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7–26
Solution of the Cause . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7–26
E28: Sector Collimator Error Movement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7–27
Definition of the Cause . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7–27
Solution of the Cause . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7–27
E29: Digital Sensor Error Movement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7–28
Definition of the Cause . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7–28
Solution of the Cause . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7–28
E30: Limit Switch X . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7–29
Definition of the Cause . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7–29
Solution of the Cause . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7–29
E31: Distance Too Short X (Calibration) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7–30
Definition of the Cause . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7–30
Solution of the Cause . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7–30
E32: Distance Too Long X (Calibration). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7–31
Definition of the Cause . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7–31

4
Contents

Solution of the Cause . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7–31


E33: Limit Switch X . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7–32
Definition of the Cause. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7–32
Solution of the Cause . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7–32
E34: Distance Too Short X (Calibration) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7–33
E35: Distance Too Long X (Calibration) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7–33
Definition of the Cause. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7–33
Solution of the Cause . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7–33
E36: Limit Switch Y . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7–34
Definition of the Cause. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7–34
Solution of the Cause . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7–34
E37: Distance Too Short Y (Calibration) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7–35
Definition of the Cause. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7–35
Solution of the Cause . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7–35
E38: Distance Too Long Y (Calibration) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7–36
Definition of the Cause. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7–36
Solution of the Cause . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7–36
E39: Limit Switch Y . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7–37
Definition of the Cause. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7–37
Solution of the Cause . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7–37
E40: Distance Too Short Y (Calibration) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7–38
E41: Distance Too Long Y (Calibration) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7–38
Definition of the Cause. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7–38
Solution of the Cause . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7–38
E42: X or Y Position not Identified . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7–39
Definition of the Cause. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7–39
Solution of the Cause . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7–39
E47: Rotation Movement Too Short . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7–40
E48: Rotation Movement Too Long . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7–40
Definition of the Cause. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7–40
Solution of the Cause . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7–40
E49: Cephalostat Positioning Blocker Time-out. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7–42
Definition of the Cause. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7–42
Solution of the Cause . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7–42
E50: Watchdog Error. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7–43
Definition of the Cause. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7–43
Solution of the Cause . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7–43
E51: Communication Problem with serial EEPROM . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7–44
Definition of the Cause. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7–44
Solution of the Cause . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7–44
E52: Communication Problem with Console . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7–45
Definition of the Cause. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7–45
Solution of the Cause . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7–45
E53: Backup Battery out of Service . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7–46
Definition of the Cause. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7–46

KODAK 9000 and KODAK 9000 3D Extraoral Imaging System_Service Guide (SM712)_Ed 02 5
Contents

Solution of the Cause . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7–46


E54: Inconsistency in Calibration Parameters. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7–47
E55: Error in Stored Exposure Parameters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7–47
E56: Error in Manufacturing Parameters. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7–47
E57: Error in User Preferences . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7–47
Definition of the Cause . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7–47
Solution of the Cause . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7–47
E58: Exposure Context Error . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7–49
Definition of the Cause . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7–49
Solution of the Cause . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7–49
E59: Program Evolution Error. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7–50
Definition of the Cause . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7–50
Solution of the Cause . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7–50
E60: Bus I²C Error. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7–51
Definition of the Cause . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7–51
Solution of the Cause . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7–51
E62: Panoramic Trajectory Error. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7–52
Definition of the Cause . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7–52
Solution of the Cause . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7–52
E63: EPROM Detection Failure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7–53
Definition of the Cause . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7–53
Solution of the Cause . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7–53
E64: Transfer Failure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7–54
Definition of the Cause . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7–54
Solution of the Cause . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7–54
E65: RTC Failure. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7–55
Definition of the Cause . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7–55
Solution of the Cause . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7–55
E66: Bus I2C Error . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7–56
Definition of the Cause . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7–56
Solution of the Cause . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7–56
E70: Communication Loss with Collimator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7–57
Definition of the Cause . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7–57
Solution of the Cause . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7–57
E71: Data Error with the Blade Collimator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7–58
Definition of the Cause . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7–58
Solution of the Cause . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7–58
E72: Firmware Version not Compatible with Blade Collimator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7–59
Definition of the Cause . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7–59
Solution of the Cause . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7–59
E73: Time-Out for Collimator Positioning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7–60
Definition of the Cause . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7–60
Solution of the Cause . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7–60
E74: Blade Collimator Mechanical Failure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7–61
Definition of the Cause . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7–61
Solution of the Cause . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7–61

6
Contents

E75: Failure in Collimator Positioning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7–62


Definition of the Cause. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7–62
Solution of the Cause . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7–62
E76: Collimator Calibration Parameter not Valid . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7–63
Definition of the Cause. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7–63
Solution of the Cause . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7–63
No movement of the Column . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7–64
Definition of the Cause. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7–64
Solution of the Cause . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7–64
E101 to E116: JBUS Ethernet Communication Problem . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7–65
Definition of the Cause. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7–65
Solution of the Cause . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7–65
E201: Problem with Image Reception . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7–66
Definition of the Cause. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7–66
Solution of the Cause . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7–66
E301: Bad Driver Version . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7–67
Definition of the Cause. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7–67
Solution of the Cause . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7–67
E302: Driver Configuration Error. Please Reinstall the Driver . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7–68
Definition of the Cause. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7–68
Solution of the Cause . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7–68
E305: Mask File not Found. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7–69
Definition of the Cause. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7–69
Solution of the Cause . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7–69
E306: Unknown Radiological Unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7–70
Definition of the Cause. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7–70
Solution of the Cause . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7–70
E311: No Sensor Detected on the Radiological Unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7–71
Definition of the Cause. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7–71
Solution of the Cause . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7–71
E312: Demo Mode Error. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7–72
Definition of the Cause. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7–72
Solution of the Cause . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7–72
E315: Unable to Initialize Ethernet Communication . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7–73
Definition of the Cause. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7–73
Solution of the Cause . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7–73
E316: Unable to Connect to the Radiological Unit. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7–74
Definition of the Cause. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7–74
Solution of the Cause . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7–74
E321: Unable to Locate the Radiological Unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7–75
Definition of the Cause. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7–75
Solution of the Cause . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7–75
E400: Unable to Allocate Memory . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7–76
Definition of the Cause. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7–76
Solution of the Cause . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7–76

KODAK 9000 and KODAK 9000 3D Extraoral Imaging System_Service Guide (SM712)_Ed 02 7
Contents

E401: Unable to Connect to the Radiological Unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7–77


Definition of the Cause . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7–77
Solution of the Cause . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7–77
E402: Communication Interruption. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7–78
Definition of the Cause . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7–78
Solution of the Cause . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7–78
E405: Unknown Error Occurred During Acquisition . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7–79
Definition of the Cause . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7–79
Solution of the Cause . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7–79
E409: Cannot Open Serial Link. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7–80
Definition of the Cause . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7–80
Solution of the Cause . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7–80
E410: Acquisition Error Time-out. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7–81
Definition of the Cause . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7–81
Solution of the Cause . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7–81
E414: Mask Files not Found . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7–82
Definition of the Cause . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7–82
Solution of the Cause . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7–82
E416: Internal Fatal Error . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7–83
Definition of the Cause . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7–83
Solution of the Cause . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7–83
E417: The Radiological Unit Encountered a Fatal Error . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7–84
Definition of the Cause . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7–84
Solution of the Cause . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7–84
E419: Communication with the Radiological Unit Failed . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7–85
Definition of the Cause . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7–85
Solution of the Cause . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7–85
E420: Unable to Connect to the Radiological Unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7–86
Definition of the Cause . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7–86
Solution of the Cause . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7–86
E422: Error During Acquisition Process . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7–87
Definition of the Cause . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7–87
Solution of the Cause . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7–87

8—THE CJ653 BOARD TEST POINTS


Description. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8–1
CJ653 Board Test Points (TP) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8–1

9—THE CJ732 BOARD TEST POINTS AND LED


Description. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9–1
CJ732 Board Test Points (PT) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9–1
CJ732 Board LEDs (LD) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9–4

10—REPLACING THE COLUMN LINEAR ACTUATOR


Description. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10–1
Removing the Column Linear Actuator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10–2

8
Contents

Replacing the Column Linear Actuator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10–2


Testing the Column Linear Actuator. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10–3

11—REPLACING THE COLUMN CABLES


Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11–1
Removing the Column Cables. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11–2
Replacing the Column Cables . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11–2
Testing the Column Cables . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11–3

12—REPLACING THE COLUMN REINFORCEMENT JACKS


Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12–1
Removing the Column Reinforcement Jacks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12–2
Replacing the Column Reinforcement Jacks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12–3
Testing the Column Reinforcement Jacks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12–3

13—REPLACING THE COLUMN LIMIT SWITCHES


Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13–1
Removing the Column Limit Switches . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13–1
Replacing the Column Limit Switches . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13–2
Testing the Column Limit Switches . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13–2

14—REPLACING THE ON/OFF BUTTON


Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14–1
Removing the ON and OFF Button. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14–2
Replacing the ON and OFF Button . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14–2
Testing the ON and OFF Button . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14–2

15—REPLACING THE LF440 LINE FILTER


Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15–1
Removing the LF440 Line Filter. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15–2
Replacing the LF440 Line Filter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15–2
Testing the LF440 Line Filter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15–2

16—REPLACING THE CJ777 FILTER BOARD


Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16–1
Removing the CJ777 Filter Board. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16–2
Replacing the CJ777 Filter Board . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16–2
Testing the CJ777 Filter Board . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16–2

17—REPLACING THE LK218 HARMONIC FILTER


Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17–1
Removing the LK218 Harmonic Filter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17–2
Replacing the LK218 Harmonic Filter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17–2
Testing the LK218 Harmonic Filter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17–2

KODAK 9000 and KODAK 9000 3D Extraoral Imaging System_Service Guide (SM712)_Ed 02 9
Contents

18—REPLACING THE MICROPROCESSOR POWER SUPPLY JZ024 BOARD


Description. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .18–1
Removing the Microprocessor Power Supply JZ024 Board . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .18–2
Replacing the Microprocessor Power Supply JZ024 Board . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .18–3
Testing the Microprocessor Power Supply JZ024 Board . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .18–3

19—REPLACING THE DIGITAL SENSOR POWER SUPPLY JZ030 BOARD


Description. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .19–1
Removing the Digital Sensor Power Supply JZ030 Board . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .19–2
Replacing the Digital Sensor Power Supply JZ030 Board . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .19–2
Testing the Digital Sensor Power Supply JZ030 Board . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .19–2

20—REPLACING THE STEP MOTOR POWER SUPPLY JZ029 BOARD


Description. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .20–1
Removing the Step Motor Power Supply JZ029 Board . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .20–2
Replacing the Step Motor Power Supply JZ029 Board. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .20–2
Testing the Step Motor Power Supply JZ029 Board . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .20–2

21—REPLACING THE CJ699 MOTOR BOARD


Description. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .21–1
Removing the CJ699 Motor Board . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .21–1
Replacing the CJ699 Motor Board . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .21–2
Testing the CJ699 Motor Board. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .21–2

22—REPLACING THE CJ699 MOTOR BOARD FUSES


Description. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .22–1
Removing the CJ699 Motor Board Fuses . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .22–2
Replacing the CJ699 Motor Board Fuses . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .22–2
Testing the CJ699 Motor Board Fuses. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .22–2

23—REPLACING THE CJ653 INVERTER BOARD


Description. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .23–1
Removing the CJ653 Inverter Board . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .23–2
Replacing the CJ653 Inverter Board . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .23–2
Testing the CJ653 Inverter Board . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .23–2

24—REPLACING THE CJ653 INVERTER BOARD FUSES


Description. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .24–1
Removing the CJ653 Inverter board Fuses . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .24–2
Replacing the CJ653 Inverter board Fuses . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .24–3
Testing the CJ653 Inverter board Fuses. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .24–3

25—REPLACING THE CJ732 MICROPROCESSOR BOARD


Description. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .25–1
Removing the CJ732 Microprocessor Board. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .25–2

10
Contents

Replacing the CJ732 Microprocessor Board. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25–3


Testing the CJ732 Microprocessor Board . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25–4

26—REPLACING THE MICROPROCESSOR BOARD CJ732 BATTERY


Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26–1
Removing the Microprocessor Board CJ732 Battery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26–1
Replacing the Microprocessor Board CJ732 Battery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26–2
Testing the Microprocessor Board CJ732 Battery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26–2

27—REPLACING THE MICROPROCESSOR BOARD EEPROMS


Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27–1
Removing the Microprocessor Board CJ732 EEPROMS. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27–1
Replacing the Microprocessor Board CJ732 EEPROMS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27–3
Testing the Microprocessor Board CJ732 EEPROMS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27–4

28—REPLACING THE CJ708 BOARD for ROTATIVE MOVEMENT


Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28–1
Removing the CJ708 Board for Rotative Movement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28–1
Replacing the CJ708 Board for Rotative Movement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28–2
Testing the CJ708 Board for Rotative Movement. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28–2

29—REPLACING THE CJ708 BOARD for X and Y MOVEMENT


Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29–1
Removing the CJ708 Board for X and Y Movement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29–1
Replacing the CJ708 Board for X and Y Movement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29–2
Testing the CJ708 Board for X and Y Movement. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29–2

30—REMINDER: THE CJ804 BOARD for the CHIN REST


Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30–1

31—REPLACING THE CJ804 BOARD for the CHIN REST


Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31–1
Removing the CJ804 Board for the Chin Rest . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31–2
Replacing the CJ804 Board for the Chin Rest . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31–2
Testing the CJ804 Board for the Chin Rest. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31–2

32—REMINDER: THE CJ804 BOARD for The COLLIMATOR


Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32–1

33—REPLACING THE CJ804 BOARD for The COLLIMATOR


Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33–1
Removing the CJ804 Board for the Collimator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33–2
Replacing the CJ804 Board for the Collimator. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33–2
Testing the CJ804 Board for the Collimator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33–2

KODAK 9000 and KODAK 9000 3D Extraoral Imaging System_Service Guide (SM712)_Ed 02 11
Contents

34—REMINDER: THE CJ804 BOARD for the DIGITAL SENSOR


Description. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .34–1

35—REPLACING THE CJ804 BOARD for the DIGITAL SENSOR (WITHOUT CEPH)
Description. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .35–1
Removing the CJ804 Board for the Sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .35–2
Replacing the CJ804 Board for the Sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .35–2
Testing the CJ804 Board for the Sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .35–2

36—REPLACING THE CJ804 BOARD for the DIGITAL SENSOR (WITH CEPH)
Description. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .36–1
Removing the CJ804 Board for the Sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .36–2
Replacing the CJ804 Board for the Sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .36–2
Testing the CJ804 Board for the Sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .36–2

37—REPLACING THE ENCODER ASSEMBLY (Unit Serial N° All xxACxxx)


Description. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .37–1
Removing the Encoder Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .37–1
Replacing the Encoder Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .37–2
Testing the Encoder Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .37–3

38—REPLACING THE ENCODER ASSEMBLY (Up to Unit Serial N° WKAH046)


Description. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .38–1
Removing the Encoder Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .38–1
Replacing the Encoder Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .38–2
Testing the Encoder Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .38–3

39—REPLACING THE ENCODER ASSEMBLY (From Unit Serial N° WKAH047)


Description. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .39–1
Removing the Encoder Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .39–1
Replacing the Encoder Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .39–2
Testing the Encoder Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .39–2

40—REPLACING THE CJ691 LASER BOARD (HEAD)


Description. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .40–1
Removing the CJ691 Laser Board (Head). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .40–2
Replacing the CJ691 Laser Board (Head) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .40–2
Testing the CJ691 Laser Board (Head) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .40–2

41—REPLACING THE CJ691 LASER BOARD (ROTATIVE ARM)


Description. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .41–1
Removing the CJ691 Laser Board (Rotative Arm) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .41–2
Replacing the CJ691 Laser Board (Rotative Arm) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .41–2
Testing the CJ691 Laser Board (Rotative Arm) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .41–2

12
Contents

42—REPLACING THE HORIZONTAL POSITIONING LASER (Unit Serial N°s: xxACxxx, xxADxxx, xx-
AHxxx)
Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42–1
Removing the Horizontal Positioning Laser . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42–2
Replacing the Horizontal Positioning Laser . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42–2
Testing the Horizontal Positioning Laser . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42–2

43—REPLACING THE HORIZONTAL POSITIONING LASER (Unit Serial N°s: xxAExxx, xxAFxxx, xx-
AJxxx, xxAKxxx)
Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43–1
Removing the Horizontal Positioning Laser . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43–1
Replacing the Horizontal Positioning Laser . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43–2
Testing the Horizontal Positioning Laser . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43–3

44—REPLACING THE MID-SAGITTAL POSITIONING LASER


Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44–1
Removing the Mid-Sagittal Positioning Laser . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44–2
Replacing the Mid-Sagittal Positioning Laser . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44–2
Testing the Mid-Sagittal Positioning Laser. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44–2

45—REPLACING THE 3D CENTRAL POSITIONING LASER


Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45–1
Removing the 3D Central Positioning Laser. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45–2
Replacing the 3D Central Positioning Laser . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45–2
Testing the 3D Central Positioning Laser . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45–3

46—REPLACING THE 3D FOV POSITIONING LASER (Unit Serial N°s: xxACxxx, xxADxxx, xxAHxxx)
Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46–1
Removing the 3D FoV Positioning Laser . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46–2
Replacing the 3D FoV Positioning Laser . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46–2
Testing the 3D FoV Positioning Laser . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46–2

47—REPLACING THE CEPHALOMETRIC FRANKFURT POSITIONING LASER (Unit Serial N°s: xx-
AExxx, xxAFxxx, xxAJxxx, xxAKxxx)
Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47–1
Removing the Cephalometric Frankfurt Positioning Laser. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47–1
Replacing the Cephalometric Frankfurt Positioning Laser . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47–2
Testing the Cephalometric Frankfurt Positioning Laser . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47–3

48—REPLACING THE LF439 CAPACITORS


Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48–1
Removing the LF439 Capacitors. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48–2
Replacing the LF439 Capacitors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48–2
Testing the LF439 Capacitors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48–2

KODAK 9000 and KODAK 9000 3D Extraoral Imaging System_Service Guide (SM712)_Ed 02 13
Contents

49—REPLACING THE ROTATIVE ARM MOTOR ASSEMBLY (Unit Serial N° xxACxxx)


Description. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .49–1
Removing the Rotative Arm Motor Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .49–1
Replacing the Rotative Arm Motor Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .49–2
Testing the Rotative Arm Motor Assembly. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .49–2

50—REPLACING THE ROTATIVE ARM MOTOR ASSEMBLY (Unit Serial N°s: xxADxxx, xxAHxxx, xx-
AExxx, xxAFxxx, xxAJxxx, xxAKxxx)
Description. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .50–1
Removing the Rotative Arm Motor Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .50–2
Replacing the Rotative Arm Motor Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .50–2
Testing the Rotative Arm Motor Assembly. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .50–3

51—REPLACING THE ROTATIVE MOVEMENT LIMIT SWITCH (Unit Serial N° xxACxxx)


Description. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .51–1
Removing the Rotative Movement Limit Switch . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .51–2
Replacing the Rotative Movement Limit Switch . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .51–2
Testing the Rotative Movement Limit Switch. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .51–2

52—REPLACING THE ROTATIVE MOVEMENT CJ822 BOARD (Unit Serial N°s: xxADxxx, xxAHxxx, xx-
AExxx, xxAFxxx, xxAJxxx, xxAKxxx)
Description. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .52–1
Removing the Rotative Movement CJ822 Board . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .52–2
Replacing the Rotative Movement CJ822 Board . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .52–2
Testing the Rotative Movement CJ822 Board. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .52–2

53—REPLACING THE X OR Y MOTOR ASSEMBLY


Description. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .53–1
Removing the X or Y Motor Assembly. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .53–2
Replacing the X or Y Motor Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .53–2
Testing the X or Y Motor Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .53–2

54—REPLACING THE X OR Y FLEXIBLE MOTOR COUPLING


Description. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .54–1
Removing the X or Y Flexible Motor Coupling . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .54–1
Replacing the X or Y Flexible Motor Coupling . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .54–2
Testing the X or Y Flexible Motor Coupling. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .54–2

55—REPLACING THE X DOUBLE PITCH WORM


Description. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .55–1
Removing the X Double Pitch Worm . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .55–2
Replacing the X Double Pitch Worm . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .55–2
Testing the X Double Pitch Worm. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .55–3

56—REPLACING THE Y DOUBLE PITCH WORM

14
Contents

Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56–1
Removing the Y Double Pitch Worm . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56–1
Replacing the Y Double Pitch Worm . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56–2
Testing the Y Double Pitch Worm . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56–2

57—REPLACING THE X OR Y BEARINGS


Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57–1
Removing the X or Y Bearings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57–2
Replacing the X or Y Bearings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57–3
Testing the X or Y Bearings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57–3

58—REPLACING THE DIGITAL SENSOR


Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58–1
Removing the Digital Sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58–1
Replacing the Digital Sensor. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58–2
Testing the Digital Sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58–3

59—REPLACING THE DIGITAL SENSOR MOTOR ASSEMBLY


Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59–1
Removing the Digital Sensor Motor Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59–1
Replacing the Digital Sensor Motor Assembly. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59–2
Testing the Digital Sensor Motor Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59–2

60—REPLACING THE DIGITAL SENSOR LIMIT SWITCH


Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60–1
Removing the Digital Sensor Limit Switch . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60–1
Replacing the Digital Sensor Limit Switch. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60–2
Testing the Digital Sensor Limit Switch . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60–2

61—REPLACING THE SECTOR COLLIMATOR MOTOR ASSEMBLY (Unit Serial N°s: xxACxxx, xxA-
Dxxx, xxAHxxx)
Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61–1
Removing the Collimator Motor Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61–1
Replacing the Collimator Motor Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61–2
Testing the Collimator Motor Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61–2

62—REPLACING THE SECTOR COLLIMATOR PLATE ASSEMBLY (Unit Serial N°s: xxACxxx, xxADxxx,
xxAHxxx)
Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 62–1
Removing the 3D Collimator Plate Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 62–2
Replacing the 3D Collimator Plate Assembly. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 62–2
Testing the 3D Collimator Plate Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 62–3

63—REPLACING THE CJ666 SECTOR COLLIMATOR DETECTION BOARD (Unit Serial N°s: xxACxxx,

KODAK 9000 and KODAK 9000 3D Extraoral Imaging System_Service Guide (SM712)_Ed 02 15
Contents

xxADxxx, xxAHxxx)
Description. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .63–1
Removing the CJ666 Collimator Detection Board . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .63–2
Replacing the CJ666 Collimator Detection Board . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .63–2
Testing the CJ666 Collimator Detection Board. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .63–3

64—REPLACING THE BLADE COLLIMATOR (Unit Serial N°s: xxAExxx, xxAFxxx, xxAJxxx, xxAKxxx)
Description. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .64–1
Removing the Blade Collimator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .64–1
Replacing the Blade Collimator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .64–2
Testing the Blade Collimator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .64–3

65—REPLACING THE BLADE COLLIMATOR BOARD (Unit Serial N°s: xxAExxx, xxAFxxx, xxAJxxx,
xxAKxxx)
Description. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .65–1
Removing the Blade Collimator Board . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .65–2
Replacing the Blade Collimator Board . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .65–2
Testing the Blade Collimator Board . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .65–3

66—REPLACING THE BLADE COLLIMATOR FIXING RING (Unit Serial N°s: xxAExxx, xxAFxxx, xx-
AJxxx, xxAKxxx)
Description. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .66–1
Removing the Blade Collimator Fixing Ring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .66–2
Replacing the Blade Collimator Fixing Ring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .66–2
Testing the Blade Collimator Fixing Ring. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .66–4

67—REPLACING THE X-RAY GENERATOR (Unit Serial N°s: xxACxxx, xxADxxx, xxAHxxx)
Description. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .67–1
Removing the X-Ray Generator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .67–2
Replacing the X-Ray Generator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .67–2
Testing the X-Ray Generator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .67–3

68—REPLACING THE X-RAY GENERATOR (Unit Serial N°s: xxAExxx, xxAFxxx, xxAJxxx, xxAKxxx)
Description. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .68–1
Removing the X-Ray Generator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .68–2
Replacing the X-Ray Generator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .68–2
Testing the X-Ray Generator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .68–4

69—REPLACING THE PANORAMIC LASER ADJUSTMENT KNOB (Unit Serial N°s: xxACxxx, xxADxxx,
xxAHxxx)
Description. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .69–1
Removing the Panoramic Adjustment Laser Knob . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .69–2
Replacing the Panoramic Adjustment Laser Knob . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .69–2
Testing the Panoramic Adjustment Laser Knob . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .69–2

16
Contents

70—REPLACING THE FIXED ARM CABLE


Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 70–1
Removing the Fixed Arm Cable . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 70–2
Replacing the Fixed Arm Cable . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 70–3
Testing the Fixed Arm Cable . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 70–4

71—REPLACING THE CONTROL PANEL


Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 71–1
Removing the Control Panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 71–2
Replacing the Control Panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 71–2
Testing the Control Panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 71–2

72—REPLACING THE CHIN REST BASE


Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 72–1
Removing the Chin Rest Base. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 72–2
Replacing the Chin Rest Base . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 72–2
Testing the Chin Rest Base . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 72–2

73—REPLACING THE CHIN REST MOTOR


Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 73–1
Removing the Chin Rest Motor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 73–2
Replacing the Chin Rest Motor. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 73–2
Testing the Chin Rest Motor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 73–2

74—REPLACING THE CHIN REST MECHANIC


Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 74–1
Removing the Chin Rest Mechanic. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 74–2
Replacing the Chin Rest Mechanic . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 74–2
Testing the Chin Rest Mechanic . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 74–2

75—REPLACING THE CHIN REST LIMIT SWITCHES


Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 75–1
Removing the Chin Rest Limit Switches . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 75–1
Replacing the Chin Rest Limit Switches. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 75–2
Testing the Chin Rest Limit Switches . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 75–2

76—REPLACING THE HAND GRIPS


Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 76–1
Removing the Hand Grips. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 76–1
Replacing the Hand Grips . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 76–2
Testing the Hand Grips . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 76–2

77—REPLACING THE TEMPLE SUPPORT


Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 77–1

KODAK 9000 and KODAK 9000 3D Extraoral Imaging System_Service Guide (SM712)_Ed 02 17
Contents

Removing the Temple Support . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .77–1


Replacing the Temple Support. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .77–2
Testing the Temple Support . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .77–2

18
Chapter 1
About This Guide

Conventions in this Guide

The following special messages emphasize information or indicate potential risk to persons or
equipment:

WARNING
Warns you to avoid injury to yourself or others by following the
safety instructions precisely.

CAUTION
Alerts you to a condition that might cause serious damage.

IMPORTANT
Alerts you to a condition that might cause problems.

NOTE
Emphasizes important information.

TIP
Provides extra information and hints.

KODAK 9000 and KODAK 9000 3D Extraoral Imaging System_Service Guide (SM712)_Ed 02 1–1
Note to the User

Note to the User

WARNING
X-rays can be harmful and dangerous if not used properly. The
instructions and warnings contained in this guide must be therefore
carefully followed.

As a manufacturer of radiology units that conform to stringent radiological protection


standards in force throughout the world, we guarantee as low as reasonably achievable degree
of protection against radiation hazards. Nonetheless, you are handling a radiology unit
specially designed to emit x-ray doses in order to carry out a medical diagnosis.

The room in which your radiology unit is to be installed must comply with all official
regulations applicable to protection against radiation. You must install your radiology unit in a
room protected against x-ray emission. This room must reduce the frequency interferences of
the 30MHz to 1GHz band to at least 12db.

Your local representative will assist you in the initial use of your radiology unit and will
supply any relevant information you may require.

To use and operate your panoramic unit and your digital imaging software you must follow the
instructions contained in this guide.

Warning and Safety Instructions


When operating the radiology unit, observe the following warning and safety instructions:

DANGER OF ELECTRIC SHOCK

This is an electrical unit. DO NOT expose it to water spray. Such action


may cause an electric shock or a malfunction of the unit.

LASER WARNING

For maximum safety, advise the patient not to look at the beam.
Before turning on the beams, lower the Frankfurt plane beam to
the lowest level. While making adjustments, ensure that the beam
is not directed into the patient's eyes.

1–2 About This Guide


Warning and Safety Instructions

WARNINGS:

• You are responsible for the operation and maintenance of this unit. Only legally
qualified persons can operate this unit. They MUST have training to use the
radiological equipment. DO NOT open the cover of the unit.When necessary,
have a trained authorized service technician carry out inspection and
maintenance operations.
• Install this unit in an x-ray room that complies with current installation
standards. From this location, you must be able to maintain visual or audio
communication with the patient and be able to access the Acquisition interface
module during exposure. This unit must be permanently connected to the ground
with a fixed power supply cable.
• DO NOT place the PC and the peripheral equipment connected to it in the
immediate vicinity of the patient in the unit. Leave at least 1.5 m distance
between the patient and the unit. The PC and the peripheral equipment must
conform to the IEC60950 standard.
• See your computer installation guide for details of the data processing system,
PC and screen. Leave a sufficient amount of clear space around the CPU to
ensure that it is properly ventilated.
• To obtain maximum image quality and visual comfort, position the screen to
avoid direct light reflections from internal or external lighting.
• DO NOT operate the unit if there is the threat of an earthquake. Following an
earthquake, ensure that the unit is operating satisfactorily before using it again.
Failure to observe this precaution may expose patients to hazards.
• X-ray equipment is hazardous to patients and the operator if you do not observe
the exposure safety factors and operating instructions.
• DO NOT place objects within the field of operation of the unit.
• The patient should wear a protective lead-lined shoulder apron, unless other
Radiation Protection Protocols apply locally.
• Disinfect any parts of the unit that come into contact with the patient and the
operator after each patient has been exposed to x-rays.
• While adjusting the height of the unit, ensure that the patient is kept clear of the
mechanism.
• When the unit is not in use, ensure that the ON/OFF switch is set to OFF (O).
• If the unit develops a fault, switch it to off (O), display an “Unserviceable” notice
and contact a service technician.
• To dispose of the unit or its components, contact a service technician.
• Ask the patient to refrain from moving during the entire period of exposure.
• Ask the patient to remain still until the unit arm has stopped moving and the
RESET movement has completed.
• DO NOT use this unit in conjunction with oxygen-rich environments. This unit is
not intended for use with flammable anesthetics or flammable agents.

KODAK 9000 and KODAK 9000 3D Extraoral Imaging System_Service Guide (SM712)_Ed 02 1–3
Marking and Labeling Symbols

Marking and Labeling Symbols

Type B device symbol complying with the IEC 60601-1


standard

In the EEC, this symbol indicates: DO NOT discard this


product in a trash receptacle; use an appropriate recovery
and recycling facility.
Contact your local sales representative for additional
information on the collection and recovery programs
available for this product
WARNING and IONIZING RADIATION symbols warn
you about radiation dangers.

LASER WARNING
Laser radiation. DO NOT stare into the beam.
Class 2 laser product.
Maximum output power: 1mW, 650 nm (IEC 60825-1
standard)
This unit emits laser radiation.

N Connecting point to the neutral conductor


L Connecting point to the live conductor
Protection ground (G)

Functional ground

1–4 About This Guide


Responsibility and Warranty

Responsibility and Warranty

Only damage affecting products themselves will be considered. The client can in no case claim
damage compensation if it does not purely concern the product itself and provided that this
exclusion of responsibility does not infringe any legal measures.

This warranty does not cover damages and faults due to accident, incorrect usage, improper
use, negligence, or wear and tear due to normal use.

Carestream Health, Inc. can not be held liable for the consequences resulting from the
non-application of the instructions contained in the installation and user manuals, namely
bodily harm, profit loss, operational interruptions, data loss, loss of a financial nature, or any
direct or indirect damage.

Carestream Health, Inc. reserves the right to examine any presumed defect.

Clients are not entitled to delay the payment of their bills or any financial deductions by way
of products reputed to be faulty.

Due to the continuous development of its products, Carestream Health, Inc. reserves the right
to amend at any time the manual and products mentioned therein, without justification or
notification to the people involved.

Our products are guaranteed in their original packaging. Carestream Health, Inc. cannot be
held liable for damages resulting from transportation.

Consumable spares, software and accessories are excluded from the warranty.

KODAK 9000 and KODAK 9000 3D Extraoral Imaging System_Service Guide (SM712)_Ed 02 1–5
Responsibility and Warranty

1–6 About This Guide


Chapter 2
UNIT DIAGRAMS

WARNING
You MUST switch off the unit before removing the covers. ONLY an
approved Trophy technician is qualified to carry out any intervention on
the unit while it is switched on and while the covers are removed. NO
unqualified person must approach the unit in the above case.

Unit Functional Diagram

Figure 2–1 Unit Functional Diagram with Sector Collimator

INSIDE PATIENT ENVIRONMENT OUTSIDE PATIENT


AC FILTERS PANORAMIC IEC 601.2.7 ENVIRONMENT
Mains - CJ777 Mains
- LF440
- Harmonic filter Mains
INVERTER
HF Transformer Supply BOARD
230V
CJ653
Mains Mains
PSU
MOTOR BOARD Mains SENSORS
CJ699 +5V, +/-15V

BASIC INSULATION
PSU
AC/DC PSU
JZ024 AC/DC
+12V 15V Column
+/-15V for motor Actuator
board Motor

X-RAY PC IEC 950


TUBE Angle information
HEAD

Filament heating
STEP MOTOR X/Y and
DRIVER BOARD Rotation
kV, mA, Temp. Statif motor CJ708 Motors
feedback control

MICROPROCESSOR BOARD Image acquisition speed control


CJ732 Ethernet - RJ45
DIGITAL
RS232 SENSORS

Feedback control
MOTOR BOARD
CJ804
COLLIMATOR DC Motor
DC motor control
DIGITAL SENSORS
- SWITCHES CHIN REST
- SENSORS
- LASER
I2C BUS

CONTROL PANEL
CJ806

PSU : Power Supply Unit

KODAK 9000 and KODAK 9000 3D Extraoral Imaging System_Service Guide (SM712)_Ed 02 2–1
Unit Functional Diagram

Figure 2–2 Unit Functional Diagram with Blade Collimator

AC FILTERS INSIDE PATIENT ENVIRONMENT OUTSIDE PATIENT


PANORAMIC IEC 601.2.7 ENVIRONMENT
Mains - CJ777 Mains
- LF440
- Harmonic filter Mains

INVERTER Mains
HF Transformer Supply BOARD
230V
CJ653
Mains Mains
PSU PSU
MOTOR BOARD SENSORS SENSOR
Mains +6.5V,+/-15V
CJ699 +5V, +/-15V
JZ030 JZ031
BASIC INSULATION
PSU PSU
AC/DC AC/DC
JZ024 JZ029
15V Column
+12V for motor Actuator
+/-15V board Motor

X-RAY PC IEC 950


TUBE Angle information
HEAD

STEP MOTOR X/Y and


Filament heating DRIVER BOARD Rotation
Arm movement CJ708 Motors
kV, mA, Temp. motor control
feedback

MICROPROCESSOR
BOARD Image acquisition speed control
CJ732 3D/PANO DIGITAL
RS232 SENSORS
Feedback control
MOTOR BOARD
CJ804
CEPH DC Motor
DC motor POSITIONING
control LOCKER
- OPTICAL-SWITCH
- SWITCHES DIGITAL SENSORS
- SENSORS BLADE
- LASER I2C BUS CHIN REST
COLLIMATOR

CEPH DIGITAL Ethernet - RJ45


SENSOR
CONTROL PANEL
CJ806
PSU : Power Supply Unit

2–2 UNIT DIAGRAMS


Unit Wiring and Interconnections Diagram

Unit Wiring and Interconnections Diagram


Figure 2–3 Unit Wiring and Interconnections Diagram with Sector Collimator

KODAK 9000 and KODAK 9000 3D Extraoral Imaging System_Service Guide (SM712)_Ed 02 2–3
Unit Wiring and Interconnections Diagram

Figure 2–4 Unit Wiring and Interconnections Diagram with Blade Collimator

2–4 UNIT DIAGRAMS


Blade Collimator Interconnections Diagram

Blade Collimator Interconnections Diagram

3D FIELD OF VIEW (FoV)


LASER BOARD POSITIONING LASER BEAM
BLADE COLLIMATOR
CJ691 2

4 4 4 4 4 4 4 4

J4 J5 J6 J7 J8 J9 J10 J11
POWER SUPPLY
AC/DC
0, +15VRL
CN51 J1 BLADE COLLIMATOR BOARD
4
JZ024 CJ821
J17

J20
MICROPROCESSOR BOARD
CJ732 I2C BUS

KODAK 9000 and KODAK 9000 3D Extraoral Imaging System_Service Guide (SM712)_Ed 02 2–5
Cable Description

Cable Description

Cable N° 1

1 GND
2 IT-KBD +5V 5
3 LY LX 4
4 LX LY 3
5 +5V IT-KBD 2

GND 1

1 CJ732 J3 1 CJ806 J1

2–6 UNIT DIAGRAMS


Cable N° 2

Cable N° 2

1 0VRL +V LASER 2
2 +V LASER 0VRL 1

2 CJ732 J21 2 CJ691 J1

KODAK 9000 and KODAK 9000 3D Extraoral Imaging System_Service Guide (SM712)_Ed 02 2–7
Cable N° 4

Cable N° 4

1 0VRL +15VRL 4
2 ON/OFF CW 3
3 CW ON/OFF 2
4 +15VRL 0VRL 1

4 CJ804 J1 4 CJ732 J19

2–8 UNIT DIAGRAMS


Cable N° 5

Cable N° 5

1 SW-DOWN SW-UP 4
2 SW-DOWN SW-UP 3

3 SW-UP SW-DOWN 2
4 SW-UP SW-DOWN 1

5 CJ806 J6 5 CJ699 J6

5 CJ806 J5 5 CJ699 J7

KODAK 9000 and KODAK 9000 3D Extraoral Imaging System_Service Guide (SM712)_Ed 02 2–9
Cable N° 7

Cable N° 7

1 0VRL +12VRL 6
2 +12VRL NC 5
NC 4
NC 3
NC 2
0VRL 1

7 CJ732 J24 7 CJ699 J4

2–10 UNIT DIAGRAMS


Cable N° 8

Cable N° 8

1 - -
2 + +

8 CJ804 J2 8

KODAK 9000 and KODAK 9000 3D Extraoral Imaging System_Service Guide (SM712)_Ed 02 2–11
Cable N° 9

Cable N° 9

1 0VRL
2 NC +5VST 4
3 POS-STAT-A POS-STAT-B 3
4 +5VST POS-STAT-A 2
5 POS-STAT-B 0VRL 1

9 9 CJ732 J14

2–12 UNIT DIAGRAMS


Cable N° 10

Cable N° 10

1 0VRL 0VRL
2 +V LASER +V LASER

10 CJ691

LASER

KODAK 9000 and KODAK 9000 3D Extraoral Imaging System_Service Guide (SM712)_Ed 02 2–13
Cable N° 13

Cable N° 13

1 0VRL OUTB 14
2 12VSW OUTA 13
3 0VRL GND 12
4 Cde-Light +12VA 11
5 Cde-RE2 CS 10
6 Cde-RE1 0VRL 9
7 0VRL ERR-CAP 8
8 ERR-CAP 0VRL 7
9 0VRL Cde-RE1 6
10 CS Cde-RE2 5
11 +12VA Cde-Light 4
12 GND 0VRL 3
13 OUTA 12VSW 2
14 OUTB 0VRL 1

13 CJ732 J1 13 CJ653 J7

2–14 UNIT DIAGRAMS


Cable N° 14

Cable N° 14

-12VA 7

+12VA 6
1 +6V NC 5
2 GND GND 4
3 -12VA GND 3
4 +6V +6V 2
5 GND +6V 1
6 +12VA

14 CJ732 J4 14 JZ024 CN51

KODAK 9000 and KODAK 9000 3D Extraoral Imaging System_Service Guide (SM712)_Ed 02 2–15
Cable N° 16

Cable N° 16

1 0VRL 0VRL 1
2 RET-mA RET-kV 2
3 RET-mA ERR-TEMP 3
4 RET-kV RET-mA 4
5 ERR-TEMP V-HEAT 5
6 V-HEAT

16 16 CJ732 J12

2–16 UNIT DIAGRAMS


Cable N° 19

Cable N° 19

1 GND GND 8
2 +15VRL NC 7
3 NC GND 6

NC 5
1 GND NC 4
2 +15VRL NC 3
3 +15VSW +15VSW 2

NC 1

19 CJ653 J8

19 CJ732 J2 19 JZ029 CN51

KODAK 9000 and KODAK 9000 3D Extraoral Imaging System_Service Guide (SM712)_Ed 02 2–17
Cable N° 22

Cable N° 22

1 0VRL +15VRL 6
2 CLK EN 5
3 CCW RST 4
4 RST CCW 3
5 EN CLK 2
6 +15VRL 0VRL 1

22 CJ708 J1 22 CJ732 J11

2–18 UNIT DIAGRAMS


Cable N° 24

Cable N° 24

7 VTB VTB 1
8 VTA VTA 2

24 24 CJ653 J3

KODAK 9000 and KODAK 9000 3D Extraoral Imaging System_Service Guide (SM712)_Ed 02 2–19
Cable N° 25

Cable N° 25

0V-N-ISO
V-COM 0V-N-ISO 1
V-COM V-COM 2
+V-N-ISO +V-N-ISO 3

25 25 CJ653 J1

2–20 UNIT DIAGRAMS


Cable N° 31

Cable N° 31

1 0VRL +15VRL 6
2 CLK EN 5
3 CCW RST 4
4 RST CCW 3
5 EN CLK 2
6 +15VRL 0VRL 1

31 CJ708 J1 31 CJ732 J10

KODAK 9000 and KODAK 9000 3D Extraoral Imaging System_Service Guide (SM712)_Ed 02 2–21
Cable N° 35

Cable N° 35

1 - -
2 + +

35 CJ804 J2 35

2–22 UNIT DIAGRAMS


Cable N° 37

Cable N° 37

BK-DN
1 BK-DN COM-DN
2 COM-DN
3 BK-UP BK-UP
4 COM-UP COM-UP
5 NC

37

SW-DOWN

37 CJ699 J8

SW-UP

KODAK 9000 and KODAK 9000 3D Extraoral Imaging System_Service Guide (SM712)_Ed 02 2–23
Cable N° 40

Cable N° 40

FLAG PANO 2
1 GND GND 1
2 FLAG PANO
3 SW-END
4 SW-END SW-END
SW-END

40 CJ822 J1

40 CJ732 J13

(NC)

40 SW-END

2–24 UNIT DIAGRAMS


Cable N°41

Cable N°41

1 +15V +15V 1
2 +5V +5V 2
3 +5V +5V 3
4 GND GND 4
5 -15V -15V 5
6 GND GND 6

ALIM SENSOR 41 CJ785 J6

KODAK 9000 and KODAK 9000 3D Extraoral Imaging System_Service Guide (SM712)_Ed 02 2–25
Cable N° 42

Cable N° 42

1 GND GND 1

2 RXD/TXD RXD/TXD 2

3 TXD/RXD TXD/RXD 3

4 GND GND 4

5 PRE-ACQ PRE-ACQ 5

6 ACQ ACQ 6

7 CLK-K7 CLK-K7 7

8 GND GND 8

42 CJ732 J15 42 CJ785 J5

2–26 UNIT DIAGRAMS


Cable N° 43

Cable N° 43

1 0VRL +15VRL 4
2 ON/OFF CW 3
3 CW ON/OFF 2
4 +15VRL 0VRL 1

43 CJ804 J1 43 CJ732 J22

KODAK 9000 and KODAK 9000 3D Extraoral Imaging System_Service Guide (SM712)_Ed 02 2–27
Cable N° 60

Cable N° 60

1 0VRL +15VRL 6
2 CLK EN 5
3 CCW RST 4
4 RST CCW 3
5 EN CLK 2

6 +15VRL 0VRL 1

60 CJ708 J1 60 CJ732 J25

2–28 UNIT DIAGRAMS


Cable N° 964

Cable N° 964

GND 5
PRE-ACQ 4
ACQ 3
CLK-K7 2
1 GND GND 1
2 NC
3 NC
4 GND
5 PRE-ACQ
6 ACQ
7 CLK-K7
8 NC

964 CJ802 J5
964 CJ732 J29

KODAK 9000 and KODAK 9000 3D Extraoral Imaging System_Service Guide (SM712)_Ed 02 2–29
Cable N° 965

Cable N° 965

1 GND
2 HEAD-THICKNESS
3 3.3V-POT
4 NC
5 NC
6 SW-FRONT-LAT SW-FRONT-LAT 1
7 GND GND 4
8 NC
9 NC
10 FAN SENSE1
11 NC FAN SENSE1 (-)
12 NC +15V-FAN1 (+)

13 +15V-FAN1
14 NC
15 NC

965 CJ802 J10

SW-FRONT
LAT

2–30 UNIT DIAGRAMS


Cable N° 966

Cable N° 966

-15V 7
1 +15V +15V 6
2 +6.5V NC 5
3 +6.5V GND 4
4 GND GND 3
5 GND +6.5V 2
6 -15V +6.5V 1

966 ALIM LAMBDA CN2


966 CJ802 J6

KODAK 9000 and KODAK 9000 3D Extraoral Imaging System_Service Guide (SM712)_Ed 02 2–31
Cable N° 973

Cable N° 973

1 GND
2 NC NC 4
3 SDA SCL 3
4 SCL SDA 2
5 NC GND 1
6 NC

973 CJ732 J20 973 CJ821 J17

2–32 UNIT DIAGRAMS


Cable N° 974

Cable N° 974

- 5

7 - 4
NC 3
1 - 2
- 1

JZ029 CN51 974 974 CJ821 J1

KODAK 9000 and KODAK 9000 3D Extraoral Imaging System_Service Guide (SM712)_Ed 02 2–33
Cable N° 979

Cable N° 979

1 0VRL +12VRL 4
2 CEPH-1 CEPH-2 3
3 CEPH-2 CEPH-1 2
4 +12VRL 0VRL 1

979 CJ822 J2 979 CJ732 J16

2–34 UNIT DIAGRAMS


Cable N°981

Cable N°981

1 - -
2 + +

981 CJ804 J2 981

KODAK 9000 and KODAK 9000 3D Extraoral Imaging System_Service Guide (SM712)_Ed 02 2–35
Cable N° 982

Cable N° 982

1 0VRL 4 DOIGT CEPH


2 NC 3 0VRL
3 DOIGT CEPH 2 NC
4 +12VRL 1 +12VRL

982 CJ822 J3

2–36 UNIT DIAGRAMS


Cable N° LZ054

Cable N° LZ054

1 0VRL 0VRL
2 - -
3 - -
4 - -
NC
NC

- CJ708 J2

KODAK 9000 and KODAK 9000 3D Extraoral Imaging System_Service Guide (SM712)_Ed 02 2–37
Cable N° LZ055

Cable N° LZ055

1 0VRL 0VRL
2 - -
3 - -
4 - -
NC
NC

- CJ708 J2

2–38 UNIT DIAGRAMS


Chapter 3
LIST OF SPARE PARTS
Column Spare Parts
Figure 3–1 Column Spare Parts

SPCG796 SPCG463
SPHR332
SPDB209

SPLS133
SPHY300

SPJW153
SPHY301
SPCP581

SPCG492
SPCG865

Table 3–1 Column Spare Parts

Part No. Spare Part Name


SPCG463 Wall mounting bracket
SPDB209 End limit switch activator
SPHR332 ON/OFF power switch fixing cover
SPLS133 ON/OFF power switch
SPJW153 Column 2 reinforcement jack
SPCP581 Electric linear actuator
SPHY301 Column connecting terminals cover
SPCG865 Cable kit for motorized column
SPCG492 Wall mounting brackets cover
SPHY300 Column top cover
SPCG796 X-ray remote control and mounting hook

KODAK 9000 and KODAK 9000 3D Extraoral Imaging System_Service Guide (SM712)_Ed 02 3–1
Column and Head Cover Spare Parts

Column and Head Cover Spare Parts


Fiber and plastic covers are not compatible. You MUST identify the type of the cover before
placing an order. All the plastic covers have “ABS” logo inside the covers.

Column and Head “Fiber” Covers

Figure 3–2 Column and Head Fiber Cover Spare Parts

SPHY343

*SPHR611

SPHY339

SPHY338

SPHY340
SPHY341

IMPORTANT
Replace ANY of these “fiber” covers with ONLY “fiber” covers.

Or, if the unit serial N° is from WAxxxxx to WGxxxxx, you CAN upgrade the
COMPLETE set of the above fiber covers with a COMPLETE set of plastic covers
(except *SPHR611) ref. SPCGA18.

Table 3–2 Column and Head Fiber Cover Spare Parts

Part No. Spare Part Name


SPHY343 Unit head cover
SPHY339 Under the head back cover
SPHY340 Under the head front cover
SPHY338 Column inside cover
SPHY341 Column side covers
*SPHR611 Protection plate

3–2 LIST OF SPARE PARTS


Column and Head Cover Spare Parts

Column and Head “Plastic” Covers

Figure 3–3 Column and Head Plastic Cover Spare Parts

SPHY368

SPHY366

*SPHR611

SPHY367
SPHY371

SPHY370

SPHY369

IMPORTANT
Replace ANY of these “plastic” covers with ONLY “plastic” covers.

The reference number for the complete set of the above plastic covers (except
*SPHR611) is ref. SPCGA18.

Table 3–3 Column and Head Plastic Cover Spare Parts

Part No. Spare Part Name


SPHY366 Unit head cover
*SPHR611 Protection plate
SPHY370 Under the head back cover
SPHY369 Under the head front cover
SPHY368 Column left side cover
SPHY367 Column right side cover
SPHY371 Column inside cover

KODAK 9000 and KODAK 9000 3D Extraoral Imaging System_Service Guide (SM712)_Ed 02 3–3
Head Assembly Spare Parts

Head Assembly Spare Parts


Figure 3–4 Head Assembly Spare Parts (Unit Serial N° xxACxxx)

SPJ6991
SPJZ029 SPJ6992
SPJ6531 SPLF440
SPJ6532
SPLF439 SPCJ777

SPCG898 SPLK218

SPJ708X SPCG841

SPLZ052 SPJ708Z

SPDU513 SPJZ024

SPJ708X
SPJZ030
SPLZ052
SPCG880
SPCG898
SPJB181 SPJB181
SPDU514 SPCJ691
SPCI979 SPLP668

Figure 3–5 Head Assembly Spare Parts (Other than the Unit Serial N° xxACxxx)

SPJZ029

SPJ6531 SPJ6991
SPJ6532 SPJ6992

SPLF439 SPLF440

SPCG898 SPCJ777

SPJ708X SPLK218

SPLZ052 SPCG945

SPCJ822
SPJ708Z
SPDU513
SPJZ024
SPJZ030 SPJ708X
SPLZ052
SPCJ804
SPCG898
SPJB181 SPJB181
SPDU514 SPCJ691
SPCI979 SPLP668
*SPCI980 *SPLP669

3–4 LIST OF SPARE PARTS


Head Assembly Spare Parts

Table 3–4 Head Assembly Spare Parts

Part No. Spare Part Name Unit Serial No.


SPCI979 CJ732 microprocessor board for sector collimator • xxACxxx
SPLP668 EEPROM x2 (last version) for sector collimator • xxADxxx
• xxAHxxx

*SPCI980 CJ732 microprocessor board for blade collimator • xxAExxx


*SPLP669 EEPROM x2 (last version) for blade collimator • xxAFxxx
• xxAJxxx
• xxAKxxx

SPDU514 Double pitch worm Y


SPJB181 2 Bearings X or Y
SPCG880 Optical encoder assembly (with pulley belt) Up to WKAH046
SPCG985 Optical encoder assembly From WKAH047
SPJZ030 Digital sensor power supply JZ030 board
SPDU513 Double pitch worm X
SPLZ052 Flexible motor coupling X or Y
SPCG898 Motor assembly X or Y
SPLF439 2 Capacitors LF439
SPJ6531 CJ653 inverter board 110 V
SPJ6532 CJ653 inverter board 230 V
SPJZ029 Motor power supply JZ029
SPJ6991 CJ699 motor board 110 V
SPJ6992 CJ699 motor board 230 V
SPLF440 LF440 line filter
SPCJ777 CJ777 filter board
SPCG945 Rotative arm motor assembly Up to WKAH047
SPCG841 Rotative arm motor assembly All xxACxxx
SPLK218 LK218 Harmonic filter
SPJZ024 Microprocessor board power supply JZ024 board
SPJ708X Step motor driver 708 board for X and Y movement
SPJ708Z Step motor drive 708 board for rotative movement
SPCJ691 CJ691 laser board
SPCJ822 Optical switch board
SPCJ804 Ceph pin DC motor board

KODAK 9000 and KODAK 9000 3D Extraoral Imaging System_Service Guide (SM712)_Ed 02 3–5
Rotative Arm Spare Parts

Rotative Arm Spare Parts


Figure 3–6 Rotative Arm Spare Parts

SPCJ804
SPLP648

SPCG891

SPCJ691

SPCG879 SPCU227

SPCJ666
SPCJ804
SPCG876
SPLP648
SPHR622
SPCG872
SPCG878

SPCG990 (PANO)
SPCG873

Figure 3–7 Rotative Arm Spare Parts

SPCJ821

SPLP648

SPCG891
SPCU227

SPCJ691

SPCG956
SPCGA23

SPCG879 SPCJ804 SPHR669


SPLP648
SPCG957
SPCG990 (PANO)
SPCG873 (3D)

3–6 LIST OF SPARE PARTS


Rotative Arm Spare Parts

Table 3–5 Rotative Arm Spare Parts

Part No. Spare Part Name Unit Serial No.


SPCG891 Rotative arm cable kit
SPCG873 3D digital sensor
SPCG990 Panoramic digital sensor
SPCG879 Sensor select motor assembly
SPCJ691 Laser CJ691 board
SPCU227 X-ray generator
SPLP648 Laser
SPCJ804 DC motor CJ804 board
SPCJ666 Collimator detection board • xxACxxx
SPCG878 Horizontal positioning laser assembly • xxADxxx
• xxAHxxx
SPCG872 3D collimator plate assembly
SPCG876 Collimator select motor assembly
SPHR622 Panoramic laser knob
SPCG956 Horizontal positioning laser assembly • xxAExxx
SPCG957 Cephalostat Frankfort positioning laser • xxAFxxx
assembly • xxAJxxx
• xxAKxxx
SPCJ821 Blade collimator pilot board
SPHR669 Blade collimator
SPCGA23 Collimator fixing ring

KODAK 9000 and KODAK 9000 3D Extraoral Imaging System_Service Guide (SM712)_Ed 02 3–7
Rotative Arm Cover Spare Parts

Rotative Arm Cover Spare Parts


Fiber and plastic covers are not compatible. You MUST identify the type of the cover before
placing an order. All the plastic covers have “ABS” logo inside the covers.

Rotative Arm “Fiber” Covers for “Sector Collimator”


Spare Parts
Figure 3–8 Rotative Arm Fiber Covers for Sector Collimator Spare Parts
SPHY364

SPHY365

SPHY361

SPHY362 SPHY363

IMPORTANT
Replace ANY of these “fiber” covers with ONLY “fiber” covers.

Or, you CAN upgrade the COMPLETE set of fiber covers above with a COMPLETE set
of plastic covers ref. SPCGA76, if the unit has the following Serial N°s:

• >WAACxxx
• >WAADxxx
• >WAAHxxx

WARNING
If your unit Serial N° is <WJxxxxx, you must drill 2 holes in the cover
mounting support (see “Drilling 2 holes in the Cover Mounting Support ....”)

Table 3–6 Rotative Arm Fiber Covers for Sector Collimator Spare Parts

Part No. Spare Part Name


SPHY363 Generator front cover
SPHY364 Generator rear cover
SPHY365 Rotative arm top cover
SPHY362 Rotative arm bottom cover
SPHY361 Digital sensor cover

3–8 LIST OF SPARE PARTS


Rotative Arm Cover Spare Parts

Rotative Arm “Plastic” Covers for “Sector Collimator”


Spare Parts

IMPORTANT
This option is possible ONLY if you have already upgraded from
the “Fiber” covers.

Figure 3–9 Rotative Arm Plastic Covers for Sector Collimator Spare Parts

SPHY379

SPHY381

SPHY383

SPHY382 SPDU526

IMPORTANT
Replace ANY of these “plastic” covers with ONLY “plastic”
covers.

The reference number for the complete set of the above plastic covers for sector
collimator is ref. SPCGA76.

Table 3–7 Rotative Arm Plastic Cover for Sector Collimator Spare Parts

Part No. Spare Part Name


SPDU526 Generator front cover
SPHY379 Generator rear cover
SPHY381 Rotative arm top cover
SPHY382 Rotative arm bottom cover
SPHY383 Digital sensor cover

KODAK 9000 and KODAK 9000 3D Extraoral Imaging System_Service Guide (SM712)_Ed 02 3–9
Rotative Arm Cover Spare Parts

Rotative Arm “Fiber” Covers for “Blade Collimator”


Spare Parts
Figure 3–10 Rotative Arm Fiber Covers for Blade Collimator Spare Parts

SPHY364

SPHY365

SPHY361

SPHY387
SPHY362

IMPORTANT
Replace ANY of these “fiber” covers with ONLY “fiber” covers.

Or, you CAN upgrade the COMPLETE set of fiber covers above with a COMPLETE set
of plastic covers ref. SPCGA75, if the unit has the following Serial N°s:

• >WDAExxx
• >WDAFxxx
• >WDAJxxx
• >WDAKxxx

WARNING
ONLY if the unit Serial N° is <WJxxxxx, you must drill 2 holes in
the cover mounting support (see “Drilling 2 holes in the Cover
Mounting Support ....”).

Table 3–8 Rotative Arm Fiber Covers for Blade Collimator Spare Parts

Part No. Spare Part Name


SPHY387 Generator front cover
SPHY364 Generator rear cover
SPHY365 Rotative arm top cover
SPHY362 Rotative arm bottom cover
SPHY361 Digital sensor cover

3–10 LIST OF SPARE PARTS


Rotative Arm Cover Spare Parts

Rotative Arm “Plastic” Covers for “Blade Collimator”


Spare Parts
Figure 3–11 Rotative Arm Plastic Covers for Blade Collimator
SPHY379

SPHY381

SPHY383

SPHY382 SPHY380

IMPORTANT
Replace ANY of these “plastic” covers with ONLY “plastic”
covers.

The reference number for the complete set of the above plastic covers for blade
collimator is ref. SPCGA75.

Table 3–9 Rotative Arm Plastic Cover for Blade Collimator Spare Parts

Part No. Spare Part Name


SPHY380 Generator front cover
SPHY379 Generator rear cover
SPHY381 Rotative arm top cover
SPHY382 Rotative arm bottom cover
SPHY383 Digital sensor cover

KODAK 9000 and KODAK 9000 3D Extraoral Imaging System_Service Guide (SM712)_Ed 02 3–11
Fixed Arm Spare Parts

Fixed Arm Spare Parts


Figure 3–12 Fixed Arm Spare Parts

SPCG982

SPHY333
SPCG910
SPJP109

SPCG866
SPCG869

SPCJ804
SPCG889 SPHY357

SPHY360
SPHY335
SPCG854

SPHY334
SPHY332

SPHR596 SPCG863

Table 3–10 Fixed Arm Spare Parts

Part No. Spare Part Name


SPCG863 Cable kit - fixed arm
SPCG869 Control panel assembly
SPCJ804 DC motor CJ804 board
SPHR596 Hand grips
SPCG866 Chin rest base
SPCG889 Chin rest mechanical motor assembly
SPCG854 Chin rest mechanical assembly
SPHY332 Bottom chin rest mechanical assembly cover
SPHY333 Chin rest base cover
SPJP109 Chin rest label
SPHY334 Chin rest mechanical assembly front cover
SPHY335 Chin rest mechanical assembly back cover
SPCG982 Temple support
SPHY357 Chin rest bite block support (panoramic)
SPHY360 TMJ x4 nose rest
SPCG910 3D bite block and bite block support

3–12 LIST OF SPARE PARTS


Miscellaneous Spare Parts

Miscellaneous Spare Parts

Table 3–11 Miscellaneous Spare Parts

Part No. Spare Part Name


SPCG895 Panoramic accessories box
SPCG892 Low voltage cable kit
SPCG893 Power cable kit
SPCG894 Laser cable kit
SPCG858 Screw set for arm
SPCG888 Screw set for cover
SPLP621 Microprocessor board U15 battery (black)
SPIS206 Microprocessor board U14 battery (yellow)
SPLX018 2 Limit switchs for:
• Column
• XY rotation movement
• Sensor rotation movement
• Chin rest

KODAK 9000 and KODAK 9000 3D Extraoral Imaging System_Service Guide (SM712)_Ed 02 3–13
Miscellaneous Spare Parts

3–14 LIST OF SPARE PARTS


Chapter 4
DRILLING HOLES FOR THE PLASTIC COVER
MOUNTING SUPPORT (Ref. SPCGA75, CSPCGA76)

WARNING
You MUST switch off the unit before removing the covers. ONLY an
approved Trophy technician is qualified to inspect or maintain the unit
while it is switched on and while the covers are removed. In this case, NO
unqualified person must approach the unit.

Description
If your unit serial n° is <WJxxxxx, for the plastic head cover (Ref. SPCGA75 and SPCGA76),
you must drill 2 holes in the cover mounting support.

WARNING
ONLY if the unit Serial N° is <WJxxxxx, you must drill 2 holes in
the cover mounting support.

To drill 2 holes in the cover mounting support for the head plastic cover, follow these steps:

o 3 mm
o 6 mm

1. From the hole (A) measure 75mm and mark the hole (B) location.

2. Drill a 3mm hole at the indicated location for the hole (B).

3. Widen the hole (B) to 6mm.

4. Repeat the same procedure for the other side symmetrically.

KODAK 9000 and KODAK 9000 3D Extraoral Imaging System_Service Guide (SM712)_Ed 02 4–1
Description

4–2 DRILLING HOLES FOR THE PLASTIC COVER MOUNTING SUPPORT (Ref. SPCGA75, CSPCGA76)
Chapter 5
USING THE TECHNICAL MENU

Description
The technical menu enables the technician to:

• Get information on the unit and the software version.


• Get information on the system parameters.
• Set parameters.
• Calibrate certain parameters.

KODAK 9000 3D Tool Kit


The Kodak 9000 3D tool kit is composed of test tools that are used for post-installation and
calibration verifications.The following table lists the content of the Kodak 9000 3D tool kit.

Table 5–1 Kodak 9000 and Kodak 9000 3D Tool Kit Content

Part No. Description Quantity


SPCG903 3D test tool base 1

3D tubes 3

3D metal rod 1

SPCG901 Ball phantom tool 1

SPHR697 Leveling pad 1

KODAK 9000 and KODAK 9000 3D Extraoral Imaging System_Service Guide (SM712)_Ed 02 5–1
KODAK 9000 3D Tool Kit

Table 5–1 Kodak 9000 and Kodak 9000 3D Tool Kit Content (Continued)

Part No. Description Quantity


SPPE136 Rotative arm calibration tool 1

SPMN781 Ceph image quality control 1

SPHR763 K9500 metal rod 1

To use the test tools for post-installation control, see the Kodak 9000/9000 3D Installation
Guide (SM711)_Ed 02, Kodak 9000C/9000C 3D Installation Guide (SM714)_Ed 01 and
Kodak 9500 Installation Guide (SM718)_Ed 01. To use the test tools for calibration, see the
“Calibration Procedures” section in this chapter.

5–2 USING THE TECHNICAL MENU


Panoramic Technical Menu Overview

Panoramic Technical Menu Overview

? Menu Description
The ? menu is composed of the following submenus:

• About
• Details
• Log

“About” Submenu Description


The About submenu displays information on:

• The software version


• The firmware version

KODAK 9000 and KODAK 9000 3D Extraoral Imaging System_Service Guide (SM712)_Ed 02 5–3
Panoramic Technical Menu Overview

“Details” Submenu Description


The Details submenu displays information on the calibration parameters.

“Log” Submenu Description


The Log submenu displays information on the history of technical events.

5–4 USING THE TECHNICAL MENU


Panoramic Technical Menu Overview

Maintenance Menu Description


The Maintenance menu is composed of the following submenus:

• General
• Communication
• Preferences
• Manufacturing parameters
• Calibration
• Reset of total parameters
• Save/Restore

“General” Submenu Description


The General submenu displays information on:

• The number of images taken


• The energy usage of the x-ray tube

KODAK 9000 and KODAK 9000 3D Extraoral Imaging System_Service Guide (SM712)_Ed 02 5–5
Panoramic Technical Menu Overview

“Communication” Submenu Description


The Communication submenu displays information on the number of communications
exchanged between the PC and the unit.

“Preferences” Submenu Description


The Preferences menu displays information on:

• Radiation dose per image


• Initial position of the arm

The Preferences menu enables the:

• Reinitialization of the user preferences


• Resetting of the calibration values

5–6 USING THE TECHNICAL MENU


Panoramic Technical Menu Overview

“Manufacturing Parameters” Submenu Description


The Manufacturing Parameters menu displays information on:

WARNING
You MUST NOT modify the information displayed in the
Manufacturing Parameters submenu, except as follows:
In the “Type of Cephalometric image” field, you can select “with
Cephalostat” if a Cephalostat is added.

• Serial number: it corresponds to the serial number of the device.


• Voltage rating: it corresponds to the mains voltage of the unit. It should correspond to
the voltage of the x-ray room for the unit to operate correctly.
• Mains frequency: it corresponds to the mains frequency of the unit. It should
correspond to the frequency of the X-ray room for the unit to operate correctly.
• Tube reference: it corresponds to the type of x-ray tube label affixed to the unit. It
should correspond to the correct version of the x-ray tube for the unit to operate
correctly.
• Type of Cephalostat:

• “without Cephalostat” if no Ceph


• “with Cephalostat” if there is Ceph

“Calibration” Submenu Description


The Calibration menu is composed of the following submenus:

• kV calibration (manufacturer only)


• mA calibration
• Heating voltage calibration
• Arm calibration
• Tube head calibration
• Control of the x-ray beam
• Dose measurement on incisors
• Reset of the calibration parameters

KODAK 9000 and KODAK 9000 3D Extraoral Imaging System_Service Guide (SM712)_Ed 02 5–7
Panoramic Technical Menu Overview

“Reset of Total Parameters” Submenu Description


The Reset of Total Parameters enables you to reset the following saved parameters in the
CJ732 microprocessor board:

• Preferences
• Manufacturing parameters
• Calibration parameters
• Parameter (kV, mA) values

“Save/Restore” Submenu Description


The Save/Restore submenu displays information on:

• Save: saves the new calibration parameters in the CJ732 board EEPROMs.
• Restore: restores the saved calibration parameters in the CJ732 board EEPROMs.

5–8 USING THE TECHNICAL MENU


3D Technical Menu Overview

3D Technical Menu Overview

? Menu Description
The ? menu is composed of the following submenus:

• About
• Details
• Log

The ? menu functions are the same as the Panoramic Technical Menu. For more information
see Panoramic Technical Menu Description section.

Maintenance Menu Description


The Maintenance menu is composed of the following submenus:

• General
• Communication
• Preferences
• Manufacturing parameters
• Calibration
• Save/Restore

The Maintenance menu functions are the same as the Panoramic Technical Menu. For more
information see Panoramic Technical Menu Description section.

KODAK 9000 and KODAK 9000 3D Extraoral Imaging System_Service Guide (SM712)_Ed 02 5–9
“Calibration” Submenu Description
The Calibration menu is composed of the following submenus:

• kV calibration (manufacturer only)


• mA calibration
• Heating voltage calibration
• Arm calibration
• Rotation Axis Calibration
• Margin Calibration
• Tube head calibration
• Activate the 3D collimator calibration
• Reset of the calibration parameters

5–10 USING THE TECHNICAL MENU


Using the Technical Menu

Using the Technical Menu


You need to use the technical menu if you have the KODAK 9000/9000 3D and
KODAK9000C/9000C 3D units for different calibration procedures listed in the “Calibration
Procedures” section.

Accessing the Technical Menu


To access the technical menu, follow these steps:

1. Switch OFF the unit.

2. From the fixed arm, on the control panel, press and hold , at the same time
and turn ON the unit.

3. Access the Panoramic or 3D Acquisition Window. If an error message is displayed,


click Retry button until the screen is clear.

4. In the Panoramic or 3D Acquisition Window, click to access the technical menus


and submenus.

KODAK 9000 and KODAK 9000 3D Extraoral Imaging System_Service Guide (SM712)_Ed 02 5–11
Using the Technical Menu

Calibration Procedures

“mA Calibration”

You need to calibrate the mA if you replace the x-ray generator.

Before calibrating the mA, make sure that:

• You have a DC voltmeter.


• You have an ohmmeter.
• The unit is switched OFF.

To calibrate the mA follow these steps:

1. On the CJ732 board, disconnect cable 16 from connection J12 and use an ohmmeter to
measure the resistance between terminals 1 and 4. The resulting value must be between
247 Ω and 253 Ω . Re-connect the cable 16 to J12.

2. Switch ON the unit in technical mode.

3. Access the Panoramic/3D Acquisition Window. If an error message is displayed, click


Retry button until the screen is clear.

4. Check and/or adjust the +6.50VDC from the JZ024 power supply (see the “Replacing the
Microprocessor Power Supply JZ024 Board” chapter)

5. On the CJ732 board, place the DC voltmeter between the PT22 and the ground terminal.
The resulting value must be 4 mA per volt.

6. In the Panoramic/3D Acquisition Window, click and select Maintenance >


Calibration > mA Calibration.

7. The mA Calibration dialog box is displayed. It indicates the current mA parameter in the
Command field and the current mA measurement in the Measurement field. Click Start
to launch the calibration process.

5–12 USING THE TECHNICAL MENU


Using the Technical Menu

8. The mA Calibration - Step 1 dialog box is displayed to calibrate 4 mA. The set value in
the Value field indicates 4 mA. To calibrate the 4 mA, launch an x-ray using the x-ray
remote control. Check the result in the PanoMesurement field, the value indicated must
not be ± 1 mA of the set value. If this is not the case then check the DC voltmeter. Note the
value displayed on the voltmeter during the x-ray exposure and multiply this value by 4.
Enter the results in the miliAmpermeter field and click Next.

9. The mA Calibration - Step 2 dialog box is displayed to calibrate 12 mA. The set value in
the Value field indicates 12 mA. To calibrate the 12 mA, launch an x-ray using the x-ray
remote control. Check the result in the PanoMesurement field, the value indicated must
not be ± 2 mA of the set value. If this is not the case, then check the DC voltmeter. Note
the value displayed on the voltmeter during the x-ray exposure and multiply this value by
4. Enter the results in the miliAmpermeter field and click Next.

10. The mA calibration dialog box is displayed. It indicates the new mA parameter in the
Command field and the new mA measurement in the Measurement field. The mA
calibration process is complete. If you need to redo the calibration click Start, if the
calibration is correct, click Save.

KODAK 9000 and KODAK 9000 3D Extraoral Imaging System_Service Guide (SM712)_Ed 02 5–13
Using the Technical Menu

“Heating Voltage Calibration”

You need to calibrate the heating voltage if you replace the x-ray generator.

WARNING
You MUST calibrate the mA before calibrating the heating
voltage.

To calibrate the heating voltage, follow these steps:

1. Switch on the unit in technical mode.

2. Access the Panoramic/3D Acquisition Window. If any error message displays, click
Retry button of each error message dialog box until the screen is clear.

3. In the Panoramic/3D Acquisition Window, click and select Maintenance >


Calibration > Heating Voltage Calibration. The Heating Voltage Calibration dialog
box is displayed, indicating the kV set value in the kV field and the corresponding
correction to the mA set value in the Delta field.

5–14 USING THE TECHNICAL MENU


Using the Technical Menu

4. For each kV value press the x-ray remote control exposure button until the audible signal
ceases. The CJ732 microprocessor board, automatically measures the heating value for
each kV. A v is displayed next to each value to validate the calibration. Repeat the
operation for each kV value.

IMPORTANT
To avoid overheating of the generator:

• Wait 5 seconds between each x-rays


• Wait 3 minutes after every 3 x-rays

5. If the launching of the x-ray takes too long and you receive an E24, click Restart and
continue the operation.

6. When all the kV values have been calibrated, click Save. The calibration process is
complete.

KODAK 9000 and KODAK 9000 3D Extraoral Imaging System_Service Guide (SM712)_Ed 02 5–15
Using the Technical Menu

“Rotation Axis Calibration” (Only for 3D Sector Collimator)

You need to calibrate the rotation axis if you replace:

• The x-ray generator


• The sector collimator
• The digital sensor

To calibrate the rotation axis, follow these steps:

1. Switch on the unit in technical mode.

2. Remove the 3D collimator.

TIP
• Before removing the 3D collimator, mark the right and the left
side of the collimator with a pencil, so you can replace the
collimator correctly.
• After removal of the 3D collimator, mark the right side of the 3D
collimator with a pencil, so you can replace the collimator
correctly.

3. Access the 3D Acquisition Window. If an error message is displayed, click Retry button
until the screen is clear.

4. In the 3D Acquisition Window, click and select Maintenance > Calibration >
Rotation Axe Calibration.

5. The Calibration of the rotation axis box is displayed. Click Start.

6. Place the 3D tooling base on the chin rest base. Position the 3D metal rod on the 3D

tooling base. From the unit control panel, press . Position the central laser on the tip

of the 3D metal rod. Launch an x-ray using the remote control. Press to save the

5–16 USING THE TECHNICAL MENU


Using the Technical Menu

setting. The End of the rotation center calibration dialog box is displayed with the
parameter results. The calculation of the new parameter is also displayed in the System
Status Screen of the 3D Acquisition Window.

7. Click Save to save the new parameters in the unit system and the PC. The calibration of
the rotation axis is complete.

8. Close the technical menu. Position the 3D tubes on the 3D tooling base.

9. Access the 3D Acquisition Window. Launch an x-ray using the remote control. If the
rotation axis is calibrated correctly, the acquired image must display 3 circles.

10. Proceed with the 3D collimator calibration.

“Activate the 3D Collimator Calibration” (Only for 3D Sector Collimator)

You need to activate the 3D sector collimator calibration if you replace:

• The x-ray generator

KODAK 9000 and KODAK 9000 3D Extraoral Imaging System_Service Guide (SM712)_Ed 02 5–17
Using the Technical Menu

• The 3D sector collimator


• The digital sensor

To activate the 3D collimator calibration, follow these steps:

1. Switch on the unit in technical mode.

2. Access the 3D Acquisition Window. If an error message is displayed, click Retry button
until the screen is clear.

3. In the 3D Acquisition Window, click and select Maintenance > Calibration >
Activate the 3D collimator calibration.

4. Launch an x-ray using the remote control. Check that the image is centered correctly in the
margin.

5. If you are not satisfied with the result, adjust the collimator and retry again.

6. To end the process, in the 3D Acquisition Window, click and select Maintenance >
Calibration > Inactivate the 3D collimator calibration.

7. Proceed with the margin calibration.

5–18 USING THE TECHNICAL MENU


Using the Technical Menu

“Margin Calibration” (Only for 3D Sector Collimator)

You need to calibrate the margin if you replace:

• The sector collimator


• The x-ray generator
• The digital sensor

To calibrate the margin, follow these steps:

1. Switch on the unit in technical mode. Make sure no object is placed between the sensor
and the generator.

2. Access the 3D Acquisition Window. If an error message is displayed, click Retry button
until the screen is clear.

3. In the 3D Acquisition Window, click and select Maintenance > Calibration >
Margin Calibration.

4. The Calibration of the sensor margins box is displayed. Click Start.

5. Launch an x-ray using the remote control. Check the x-ray image to see the state of the
margin. The End of the margin calibration of the 3D sensor dialog box is displayed.

6. Click Save if the result is satisfactory. The result value indicated must be between 0 and
40 on every side.

KODAK 9000 and KODAK 9000 3D Extraoral Imaging System_Service Guide (SM712)_Ed 02 5–19
Using the Technical Menu

“Rotative Arm Calibration” (Unit Serial N°s: Up to WKAH046, WKAE030 &


WKAF007)

You need to calibrate the rotative arm if you replace:

• The X and Y limit switches


• The rotative arm limit switches
• The X and Y motor assembly
• The X and Y double pitch worm
• The encoder assembly

WARNING
You MUST correctly calibrate the rotative arm of KODAK
9000/9000 3D. If the arm calibration is not correctly executed the
cephalometric mode will not be operational.

To calibrate the rotative arm, follow these steps:

1. Remove the rotative arm covers.

2. Switch on the unit in technical mode.

3. Access the Panoramic Acquisition Window. If an error message is displayed, click


Retry button until the screen is clear.

4. In the Panoramic Acquisition Window, click and select Maintenance >


Calibration > Arm Calibration.

5–20 USING THE TECHNICAL MENU


Using the Technical Menu

5. The Calibration of the statif box is displayed. Click Start.

The rotative arm identifies the current parameters using the reference switches and the
limit switches for the X, Y and the rotative axes. The current parameters are displayed in
the Calibration of the statif dialog box.

6. Place a panoramic standard bite block on the chin rest. From the fixed arm, on the control

panel, press . Using position the central laser at the center of the bite block
incision.

7. Adjust manually the rotative arm to the mid-sagittal laser. Make sure that the mid-sagittal
laser is at the center of the digital sensor wheel (A) and cable access.

KODAK 9000 and KODAK 9000 3D Extraoral Imaging System_Service Guide (SM712)_Ed 02 5–21
Using the Technical Menu

8. Click Next. The rotative arm executes a full rotation and the new parameters are displayed
in the Calibration of the statif dialog box.

9. Click Save to save the new parameters.

10. Click Exit to end the calibration and to exit the menu.

5–22 USING THE TECHNICAL MENU


Using the Technical Menu

“Rotative Arm Calibration” (Unit Serial N°s: From WKAH047, WKAE031 &
WKAF008)

You need to calibrate the rotative arm if you replace:

• The X and Y limit switches


• The rotative arm limit switches
• The X and Y motor assembly
• The X and Y double pitch worm
• The encoder assembly

WARNING
You MUST correctly calibrate the rotative arm of KODAK
9000/9000 3D. If the arm calibration is not correctly executed the
cephalometric mode will not be operational.

To calibrate the rotative arm, follow these steps:

1. Remove the rotative arm covers.

2. Switch on the unit in technical mode.

3. Access the Panoramic Acquisition Window. If an error message is displayed, click


Retry button until the screen is clear.

4. In the Panoramic Acquisition Window, click and select Maintenance >


Calibration > Arm Calibration.

KODAK 9000 and KODAK 9000 3D Extraoral Imaging System_Service Guide (SM712)_Ed 02 5–23
Using the Technical Menu

5. The Calibration of the statif box is displayed. Click Start.

The rotative arm identifies the current parameters using the reference switches and the
limit switches for the X, Y and the rotative axes. The current parameters are displayed in
the Calibration of the statif dialog box.

6. Place a panoramic standard bite block on the chin rest. From the fixed arm, on the control

panel, press . Using position the central laser at the center of the bite block
incision.

7. Insert the arm calibration tool in the hole (A) Using , adjust the rotative arm to
enable the calibration tool to go through the holes until the arm is completely blocked.

5–24 USING THE TECHNICAL MENU


Using the Technical Menu

8. Remove the arm calibration tool. Click Next. The rotative arm executes a full rotation
and the new parameters are displayed in the Calibration of the statif dialog box. Check
that the calibration tool can go through the hole if not redo the procedures 6 through 8.

9. Click Save to save the new parameters.

10. Click Exit to end the calibration and to exit the menu.

KODAK 9000 and KODAK 9000 3D Extraoral Imaging System_Service Guide (SM712)_Ed 02 5–25
Using the Technical Menu

“Tube Head Calibration” (X-Ray Without Movements)

You use the tube head calibration menu, if you need to:

• Test the x-ray emission of the generator


• Execute an x-ray tube formation
• Measure the mA

Before starting, make sure that you have:

• An intensifier screen to test the x-ray emission of the generator


• A DC voltmeter to measure the mA

To start, follow these steps:

1. Switch ON the unit in technical mode.

2. Access the Panoramic/3D Acquisition Window. If an error message is displayed, click


Retry until the screen is clear.

3. In the Panoramic/3D Acquisition Window, click and select Maintenance >


Calibration > Tube Head Calibration.

• To test the x-ray emission of the generator: Attach an intensifier screen at the center of
the x-ray beam on the digital sensor and launch an x-ray.
• To execute an x-ray tube formation: See Error message E6 and follow the procedures.
• To measure the mA, connect the DC voltmeter as described in the “mA Calibration”
section, procedure 4.
• If needed, to calibrate the mA and the heating voltage: See the “mA Calibration” and
“Heating Voltage Calibration” section.

5–26 USING THE TECHNICAL MENU


Using the Technical Menu

“Control of the X-Ray Beam” (Panoramic Only)

You need to check that the x-ray beam is correctly centered on the CCD if you replace:

• The x-ray generator


• The sector collimator
• The digital sensor

To check the x-ray beam, follow these steps:

1. Switch on the unit in technical mode.

2. Access the Panoramic Acquisition Window. If an error message is displayed, click


Retry button until the screen is clear.

3. In the Panoramic Acquisition Window, click and select Maintenance >


Calibration > Control of the x-ray beam.

4. Launch an x-ray using the remote control. Check that the x-ray beam represented in black
is centered correctly in the white frame.

5. If the x-ray beam is not centered correctly, you must adjust the collimator and relaunch an
x-ray until the x-ray beam is centered correctly.

KODAK 9000 and KODAK 9000 3D Extraoral Imaging System_Service Guide (SM712)_Ed 02 5–27
Using the Technical Menu

“Dose Measurement on Incisors” (Panoramic Only)

You need to measure the dose of the incisors to check the radiation dose received by the soft
tissues. This is used for a consistency test.

To measure the dose of the incisors, follow these steps:

1. Switch on the unit in technical mode.

2. Access the Panoramic Acquisition Window. If an error message is displayed, click


Retry button until the screen is clear.

3. Attach the x-ray dosimeter at the center of the x-ray beam on the digital sensor and make
sure that the dosimeter cable does not block the rotation movement of the rotative arm.

4. In the Panoramic Acquisition Window, click and select Maintenance >


Calibration > Dose measurement on incisors.

5. Launch an x-ray using the remote control. Check the x-ray dosimeter to see if the dose is
compatible with the acceptable x-ray dose value (see your local regulations for the dose
value).

5–28 USING THE TECHNICAL MENU


Using the Technical Menu

“Reset of the Calibration Parameters”

You need to reset the calibration parameters when you change the microprocessor board
CJ732 EEPROMs.

To reset the calibration parameters, follow these steps:

1. Switch on the unit in technical mode.

2. Access the Panoramic/3D Acquisition Window. If an error message is displayed, click


Retry button until the screen is clear.

3. In the Panoramic/3D Acquisition Window, click and select Maintenance >


Calibration > Reset of the calibration parameters.

4. Click Yes to set all the parameters to 0.

5. You must now restore all the saved parameters. Select Maintenance > Calibration >
Save/Restore

6. Click Restore to restore all the saved calibration parameters in the new CJ732 board
EEPROMs.

KODAK 9000 and KODAK 9000 3D Extraoral Imaging System_Service Guide (SM712)_Ed 02 5–29
Using the Calibration Tool Software

Using the Calibration Tool Software


You need to use the calibration tool software to:

• Calibrate the blade collimator


• Calibrate the rotation axis
• Control the image quality

IMPORTANT
You must calibrate the blade collimator before calibrating the
rotation axis.

Calibrating the Blade Collimator


To calibrate the blade collimator, follow these steps:

1. On your desktop, double-click . The Calibrations and Quality Control Software


window is displayed.

5–30 USING THE TECHNICAL MENU


Using the Calibration Tool Software

2. Click . The Password dialog box is displayed. Enter the password


“Technician” and click OK. The Calibration Tools window is displayed.

3. In the central pane, click Collimator Calibration and follow the instructions.

KODAK 9000 and KODAK 9000 3D Extraoral Imaging System_Service Guide (SM712)_Ed 02 5–31
Using the Calibration Tool Software

Calibrating the Rotation Axis


To calibrate the rotation axis, follow these steps:

1. If the collimator is a sector collimator, you must remove the 3D collimator.

TIP
• Before removing the 3D collimator, mark the right and the left
side of the collimator with a pencil, so you can replace the
collimator correctly.
• After removal of the 3D collimator, mark the right side of the 3D
collimator with a pencil, so you can replace the collimator
correctly.

If the collimator is a blade collimator, you must open wide the blades.

2. Place the 3D tooling base on the chin rest base. Position the 3D metal rod on the 3D

tooling base. From the unit control panel, press . Position the central laser on the tip
of the 3D metal rod.

3. On your desktop, double-click . The Calibrations and Quality Control Software


window is displayed.

4. Click . The Calibration Tools window is displayed.

5–32 USING THE TECHNICAL MENU


Using the Calibration Tool Software

5. In the central pane, double-click Rotation Axis Calibration. The Rotation Center
Calibration dialog box is displayed.

6. In the menu bar, select Properties > Input from Acquisition. The 3D Acquisition
Window is displayed. The acquisition parameter settings are preselected by default.

7. Launch an x-ray using the remote control. The 3D Acquisition Window disappears and
the Rotation Center Calibration dialog box is displayed with the x-ray image of the
metal rod.

KODAK 9000 and KODAK 9000 3D Extraoral Imaging System_Service Guide (SM712)_Ed 02 5–33
Using the Calibration Tool Software

8. Click . The Computation In Progress appears. Wait until the


computation is finished.

9. The computed values are displayed. The Height Span and the Angle Span acceptable

values must be from 1 to 10. If the result is appropriate, click .

10. Close the Calibrations and Quality Control Software. Position the 3D tubes on the 3D
tooling base.

5–34 USING THE TECHNICAL MENU


Using the Calibration Tool Software

11. Access the 3D Acquisition Window. Launch an x-ray using the remote control. If the
rotation axis is calibrated correctly, the acquired image must display 3 circles.

12. If the acquired image is not satisfactory you can modify the parameters. In this case, open
the Rotation Center Calibration dialog box, in the menu bar, select Help > Tunes
Center Value. The Tunes Center V... dialog box is displayed.

13. Modify the value and click OK. The new value is displayed. Click .

14. Close the Calibrations and Quality Control Software. Access the 3D Acquisition
Window. Launch an x-ray using the remote control. Check that the acquired image is
satisfactory.

KODAK 9000 and KODAK 9000 3D Extraoral Imaging System_Service Guide (SM712)_Ed 02 5–35
Using the Calibration Tool Software

NOTE
If you need to compute the rotation axis values from a previous
acquisition, do the following:

In the Rotation Center Calibration dialog box, in the menu bar, select Properties >
Input from Directory.

The directory dialog box is displayed.

Find the path to your acquired 3D image with the 3D metal rode. Select the desired
image and follow the steps 7 through 10.

15. Replace correctly the 3D collimator if it is a sector collimator.

16. If the calibration is finished, proceed with the 3D blade collimator calibration.

5–36 USING THE TECHNICAL MENU


Using the Calibration Tool Software

Controlling the Image Quality


To control the image quality, follow these steps:

1. On your desktop, double-click . The Calibrations and Quality Control Software


window is displayed.

2. Click . The Quality Assurance Control Access window is displayed.

3. In the Tools pane, double-click Quality Assurance Control. The Image Quality Control
window is displayed.

KODAK 9000 and KODAK 9000 3D Extraoral Imaging System_Service Guide (SM712)_Ed 02 5–37
Using the Calibration Tool Software

4. In the Image Quality Control window, to perform the test, you can either:

• Select the desired device in the List of detected acquisition devices field.
Or

• Browse and import a file in the Import field.


Click Next and follow the instructions.

5–38 USING THE TECHNICAL MENU


Chapter 6
INFORMATION MESSAGES

In the event of an incorrect action, a message and the associated error code is displayed on the
Display Screen of the unit control panel and on System Status Screen of the Acquisition
Window. In some cases, an audible warning is also issued.

The following table lists the information error code, their message and the action to take.

Table 6–1 Information Messages

Information Information
Description Action
Error Code Message
I1 X-Ray tube cooling Cooling in progress. Wait until the Generator Cooling Indicator on the Acquisition
Window reaches zero.
I2 Thermal security Cooling in progress. Wait until the Generator Cooling Indicator on the Acquisition
Window reaches zero.
I3 Release handswitch The acquisition has Release the exposure button of the x-ray remote control.
ended.
I5 Head clamps position The head clamps Position manually the head clamps for the selected exam.
position does not
match the selected
exam program.
I6 Wrong rotative arm The exposure button of
position the x-ray remote
control is inactive Press to reset the rotative arm in start position.
because the rotative
arm is not in the start
position.
I9 Cooling Ceph sensor The Cephalometric Wait until the Ready Indicator LED is green.
sensor is in the cooling
process.
I 15 Interface inactive The Acquisition • Check that the unit is switched on.
Window cannot be • Wait for the connection between the unit and the PC.
accessed.
• Check that the Acquisition Window is not masked by another
application, in this case close the masking application.
I 16 Incorrect positioning of This error applies only Raise the chin rest to the maximum height.
the chin rest to panoramic mode.
I 17 Sensor in movement The sensor is in the Wait for the end of the positioning process.
positioning process for
either panoramic, 3D
or Ceph mode.
I 18 Unidentified Sensor The Sensor is Check that no object stops the correct positioning of the sensor, then,
positioning incorrectly positioned
for acquisition.
press to reset.

KODAK 9000 and KODAK 9000 3D Extraoral Imaging System_Service Guide (SM712)_Ed 02 6–1
6–2 INFORMATION MESSAGES
Chapter 7
ERROR MESSAGES

WARNING
You MUST switch off the unit before removing the covers. ONLY an
approved Trophy technician is qualified to carry out any intervention on
the unit while it is switched on and while the covers are removed. NO
unqualified person must approach the unit in the above case.

E 1: Handswitch was Released Before the End of


Exposure

Definition of the Cause


The microprocessor detects premature release of exposure button of the x-ray remote control
before the end of acquisition.

Solution of the Cause

Possible Cause: The user has released the exposure button prematurely.
Remarks: If the exposure button is released shortly before the end of the acquisition cycle, the
image can be saved.
Fault Identification:
Solution: • Save the image if possible.
• Relaunch the acquisition and hold the exposure button until the end of the
acquisition cycle.

Possible Cause: • The exposure button is defective.


• The door safety contact is defective.
• The wiring of the x-ray remote control is defective.

Remarks:
Fault Identification: • The fault occurs when the exposure button is activated.
• The problem occurs frequently and at random intervals.
• Check the wiring, the exposure button and the contacts.

Solution: Replace the defective component.

KODAK 9000 and KODAK 9000 3D Extraoral Imaging System_Service Guide (SM712)_Ed 02 7–1
E2: Mains Power Failure Detected

E2: Mains Power Failure Detected

Definition of the Cause


The microprocessor has detected a fault with the power supply voltage.

Solution of the Cause

Possible Cause: Spikes or interferences in the power supply line.


Remarks: In this case, there is a risk that the operation currently in progress may not be
performed correctly (for example, saving parameters). Therefore, it is recommended
to restart the operation.
Fault Identification: Check the following electrical installations:
• The wiring to and from the unit
• The grounding of the installation
• The section of the cables
• The routing of the unit cables close to components that emit significant
interference
• The presence of other devices connected to the unit power supply line

Solution: If fault is detected, the installation must be rendered compliant.

Possible Cause: The power supply to the unit is 20% higher or lower than the nominal value (see the
identification plate).
Remarks: The unit is not functioning correctly.
Fault Identification: Check the following electrical installations:
• The power supply socket voltage measured using a voltmeter
• The wiring to and from the unit
• The section of the cables

Solution: If fault is detected, the installation must be rendered compliant.

Possible Cause: One of the CJ653 board F3, F4 or F5 fuses are defective.
Remarks: • The F3 and F4 fuses protect the power supply of the Q5 transformer 220/12V.
• The F5 fuse protects the power supply.
Fault Identification: • Check the fuses.
• Measure the power supply voltage to see if it is excessively high or low.
Solution: Replace the fuses. If necessary, replace the CJ653 board.

7–2 ERROR MESSAGES


E2: Mains Power Failure Detected

Possible Cause: Cancellation of +12 V


Remarks: The U20 controller of CJ732 detects power supply voltage inferior to 12V.
Fault Identification: Measure the power supply voltage 12VA of the microprocessor power supply JZ024
(PT12 on J4 of the CJ732).
Solution: Replace the microprocessor power supply JZ024 if defective.

KODAK 9000 and KODAK 9000 3D Extraoral Imaging System_Service Guide (SM712)_Ed 02 7–3
E3: Kv Measurement Problem (Kv High)

E3: Kv Measurement Problem (Kv High)

Definition of the Cause


• The high voltage in the x-ray generator exceeds the specified value.
• The microprocessor has detected the fault and has interrupted the x-ray emission to
avoid damage to the x-ray generator.

Solution of the Cause

Possible Cause: A cable is defective.


Remarks:
Fault Identification: Check the connecting cable 16/CJ732 on J12 and cable 24/CJ653 on J3 for:
• Electric continuity
• Electric connections
• Cable condition

Solution: Replace the defective component.

Possible Cause: The following component have been replaced without being calibrated:
• The CJ732 board
• The x-ray generator

Remarks: The preheating value is incorrect.


Fault Identification:
Solution: Calibrate the heating voltage.

Possible Cause: • The CJ732 board is defective.


• The x-ray generator is defective.
Remarks:
Fault Identification: Measurements must be performed on the CJ732 board or/and x-ray generator.
Solution: • Replace the CJ732 board.
• Replace the x-ray generator.

7–4 ERROR MESSAGES


E4: Kv Measurement Problem (kV Low)

E4: Kv Measurement Problem (kV Low)

Definition of the Cause


• The high voltage in the x-ray generator is inferior to the specified value.
• The microprocessor has detected the fault and has interrupted the x-ray emission to
avoid damage to the x-ray generator.

Solution of the Cause

Possible Cause: A cable is defective.


Remarks:
Fault Identification: Check the connecting cable 13/CJ732 on J1, cable 16/CJ732 on J12 and cable
24/CJ653 on J3 for:
• Electric continuity
• Electric connections
• Cable condition

Solution: Replace the defective component.

Possible Cause: The following component have been replaced without being calibrated:
• The CJ732 board
• The x-ray generator

Remarks: The preheating value is incorrect.


Fault Identification:
Solution: Calibrate the heating voltage.

Possible Cause: The power supply to the unit is 20% lower than the nominal value (see the
identification plate).
Remarks: The unit is not functioning correctly.
Fault Identification: Check the following electrical installations:
• The power supply socket voltage measured using a voltmeter
• The wiring to and from the unit
• The section of the cables

Solution: If fault is detected, the installation must be rendered compliant.

Possible Cause: One of the CJ653 board F1 or F2 fuses are defective.

KODAK 9000 and KODAK 9000 3D Extraoral Imaging System_Service Guide (SM712)_Ed 02 7–5
E4: Kv Measurement Problem (kV Low)

Remarks: The F1 and F2 fuses protect the power supply of the CJ653 Inverter Board.
Fault Identification: • Check the fuses.
• If the fuses are burned, check the CJ653 board to see if it is defective.
Solution: • Replace the fuses.
• If necessary, replace the CJ653 board.

Possible Cause: • The CJ732 board is defective.


• The x-ray generator is defective.
Remarks: Measurements must be performed on the CJ732 board and the x-ray generator.
Fault Identification:
Solution: • Replace the CJ732 board.
• Replace the x-ray generator.

7–6 ERROR MESSAGES


E6: Electric Arc Detection in the Generator

E6: Electric Arc Detection in the Generator

Definition of the Cause


• An electric arc has been generated in the x-ray generator.
• The microprocessor has detected the fault and has interrupted the x-ray emission to
avoid damage to the x-ray generator.

Solution of the Cause

Possible Cause: The x-ray generator is defective


Remarks:
Fault Identification: Acquire 1 or more images at 90kV and 8mA. In many cases, when an electric arc is
generated a click can be heard in the x-ray generator.
Solution: • Replace the x-ray generator.

Possible Cause: The following component have been replaced without being calibrated:
• The CJ732 board
• The x-ray generator

Remarks: The preheating value is incorrect.


Fault Identification:
Solution: Calibrate the heating voltage.

Possible Cause: The power supply to the unit is 20% lower than the nominal value (see the
identification plate).
Remarks: The unit is not functioning correctly.
Fault Identification: Check the following electrical installations:
• The power supply socket voltage measured using a voltmeter
• The wiring to and from the unit
• The section of the cables

Solution: If fault is detected, the installation must be rendered compliant.

KODAK 9000 and KODAK 9000 3D Extraoral Imaging System_Service Guide (SM712)_Ed 02 7–7
E6: Electric Arc Detection in the Generator

Possible Cause: If the x-ray generator has not been used for 2 months or more, the x-ray tube may
degas, that means, the gas molecules escape from the x-ray tube.
Remarks:
Fault Identification:
Solution: You must recreate a vacuum inside the x-ray tube.
To recreate a vacuum inside the x-ray tube, follow these steps:
1. Switch off the unit.

2. From the fixed arm, on the control panel, press and hold at the same time

, and turn on the unit.

3. The display screen of the control panel runs a series of initialization messages,
then it indicates Mode: Technical.

4. Access the Panoramic Acquisition Window. Click select


Maintenance > Calibration > Tube head calibration.

5. Access the Panoramic Acquisition Window, in the Program pane, select


the following parameters:

• 60 kV, 2 mA and make 3 exposures. Wait 10 minutes.


• 60 kV, 6.3 mA and make 1 exposure.
• 60 kV, 15 mA and make 2 exposures. Wait 10 minutes.
• 60 kV, 15 mA and make 1 exposure.
• 70 kV, 6.3 mA and make 1 exposure.
• 70 kV, 15 mA and make 1 exposure. Wait 10 minutes.
• 70 kV, 10 mA and make 1 exposure.
• 80 kV, 10 mA and make 1 exposure.
• 90 kV, 10 mA and make 1 exposure.
• The end of procedures.

WARNING:
If the x-ray tube contains too much gas, there are risks of arcing, which can be
detected by the unit system. The following error messages could be displayed:
E6, E3, E4, E7, E8, E10.
In this case wait 15 minutes and restart again the procedures.
If after having carried out the procedure 3 times, the error message appears
again, then the x-ray generator is defective. Replace the x-ray generator.

7–8 ERROR MESSAGES


E7: mA Measurement Problem (mA High)

E7: mA Measurement Problem (mA High)


E8: mA Measurement Problem (mA Low)

Definition of the Cause


• The mA values in the x-ray generator is either superior or inferior to the specified
values.
• The microprocessor has detected the fault and has interrupted the x-ray emission to
avoid damage to the x-ray generator.

Solution of the Cause

Possible Cause: A cable is defective.


Remarks:
Fault Identification: Check the connecting cable 14/CJ732 on J4 and cable 16/CJ732 on J12 for:
• Electric continuity
• Electric connections
• Cable condition

Solution: Replace the defective component.

Possible Cause: The following component have been replaced without being calibrated:
• The CJ732 board
• The x-ray generator

Remarks: The preheating value is incorrect.


Fault Identification:
Solution: Calibrate the mA values.

Possible Cause: The value of the supply voltage to the heating circuit is incorrect.
Remarks:
Fault Identification: In this case, the mA values can only be too low (E 8 only).
Solution: • Measure with a voltmeter the supply voltage to the heating circuit PT14 on the
CJ732 board.

WARNING:
This value must be set to 6.5 VDC
• Adjust the voltage.

KODAK 9000 and KODAK 9000 3D Extraoral Imaging System_Service Guide (SM712)_Ed 02 7–9
E8: mA Measurement Problem (mA Low)

Possible Cause: The following is defective:


• The CJ732 board
• The x-ray generator

Remarks:
Fault Identification: Measurements must be performed on the defective components for fault
identification.
Solution: • Replace the CJ732 board.
• Replace the x-ray generator.

7–10 ERROR MESSAGES


E9: Filament Heating Problem (mA High)

E9: Filament Heating Problem (mA High)


E10: Filament Heating Problem (mA Low)

Definition of the Cause


• The current passing through the filament of the x-ray tube is either superior or inferior
to the specified values.
• The microprocessor has detected the fault and has interrupted the x-ray emission to
avoid damage to the x-ray generator.

Solution of the Cause

Possible Cause: A cable is defective.


Remarks:
Fault Identification: Check the connecting cable 14/CJ732 on J4 and cable 16/CJ732 on J12 for:
• Electric continuity
• Electric connections
• Cable condition

Solution: Replace the defective component.

Possible Cause: The value of the preheating voltage is incorrect


Remarks:
Fault Identification: The following component have been replaced without being calibrated:
• The CJ732 board
• The x-ray generator

Solution: Calibrate the heating voltage.

Possible Cause: The cause is one of the following:


• The value of the supply voltage to the heating circuit is incorrect.
• The power supply to the unit has failed.

Remarks:
Fault Identification: In this case, the mA can only be too low (E 10 only).

KODAK 9000 and KODAK 9000 3D Extraoral Imaging System_Service Guide (SM712)_Ed 02 7–11
E10: Filament Heating Problem (mA Low)

Solution: • Measure with a voltmeter the supply voltage to the heating circuit PT14 on the
CJ732 board.

WARNING:
This value must be set to 6.5 VDC
• Adjust the voltage

Possible Cause: The filament of the x-ray tube is open.


Remarks: In this case, the mA values can only be too low or absent.
Fault Identification: Measure the value of the filament resistor.
Solution: Replace the x-ray generator.

Possible Cause: The following is defective:


• The CJ732 board
• The x-ray generator

Remarks:
Fault Identification: Measurements must be performed on the CJ732 board and/or x-ray generator.
Solution: • Replace the CJ732 board.
• Replace the x-ray generator.

7–12 ERROR MESSAGES


E11: Safety Device Activated on High Voltage Capacitors

E11: Safety Device Activated on High Voltage


Capacitors

Definition of the Cause


The voltage at the terminals of the 2200 µF - 250V capacitors is incorrect.

Solution of the Cause

Possible Cause: The power supply to the unit is 20% higher or lower than the nominal value (see the
identification plate).
Remarks: The unit is not functioning correctly.
Fault Identification: Check the following electrical installations:
• The power supply socket voltage measured using a voltmeter
• The wiring to and from the unit
• The section of the cables

Solution: If a fault is detected, the installation must be rendered compliant.

Possible Cause: A cable is defective.


Remarks:
Fault Identification: Check the connecting cable 25/CJ653 on J1 for:
• Electric continuity
• Electric connections
• Cable condition

Solution: Replace the defective component.

Possible Cause: One of the 2200 µF 250 V capacities are unserviceable.


Remarks:
Fault Identification: Check the capacitors immediately after the x-ray emission, the voltage should be
between 127 V and 183 V on each capacitor.
Solution:
WARNING:
You MUST replace both capacitors.

KODAK 9000 and KODAK 9000 3D Extraoral Imaging System_Service Guide (SM712)_Ed 02 7–13
E12: Problem with Rotation Reference Detector

E12: Problem with Rotation Reference Detector

Definition of the Cause


The rotation reference was not detected during calibration or during reset movement. In this
case, the rotative arm and the driving wheel may skid (there is no risk of damage to the unit).

Solution of the Cause

Possible Cause: • A cable is defective.


• The optical switch CJ822 board is defective.
Remarks:
Fault Identification: Check the connecting cables 40/CJ732 on J13 and 979/CJ822 on J2 - 979/CJ732 on
J16, for:
• Electric continuity
• Electric connection
• Cable condition
Check the state of LD7 on the microprocessor CJ732 board.
Solution: Replace the defective component.

Possible Cause: • The rotative arm movement motor is defective.


• The driving wheel is skidding.
Remarks: Check to see if there is a mechanical locking.
Fault Identification: • The motor cannot be turned on.
• The motor is running but the driving wheel is skidding.
Solution: • Replace the motor.
• Clean the track with a soft damp cloth.
• Tighten the driving wheel on the motor shaft.

7–14 ERROR MESSAGES


E15: Movement Too Long

E15: Movement Too Long

Definition of the Cause


During calibration of the rotative arm movement, the microprocessor has detected a 10% value
superior to the theoretical value of 360° electric pulse waves of 6766.

Solution of the Cause

Possible Cause: Incorrect selection of the rotation wheel size.


Remarks:
Fault Identification:
Solution:

1. Access the technical menu. Click select Maintenance >


Manufacturing parameters.

2. In the Manufacturing parameters dialog box, check the “Size of the


transmission wheel” field to see if the size of the wheel indicated is correct.

3. If the size of the indicated wheel is incorrect, select the appropriate wheel
size.

4. Click Save.

Possible Cause: Incorrect tightening of the driving wheel.


Remarks:
Fault Identification:
Solution: • Tighten the driving wheel correctly.
• Calibrate the rotative arm movement.

Possible Cause: The driving wheel is skidding.


Remarks:
Fault Identification: The motor is turning and the driving wheel is skidding.
Solution: • Clean the track with a soft damp cloth.
• Tighten the driving wheel on the motor shaft.

KODAK 9000 and KODAK 9000 3D Extraoral Imaging System_Service Guide (SM712)_Ed 02 7–15
E15: Movement Too Long

Possible Cause: • The optical encoder is defective or incorrectly positioned.


• The code selector is defective or incorrectly positioned.
Remarks: Absence of the electric pulse wave detection during the rotative arm rotation.
Fault Identification: Using LD22 and LD26 on the CJ732, check the security and condition of:
• The optical encoder
• The optical encoder driving belt
• The encode wheel

Solution: Replace the defective component.

Possible Cause: The cable 40/CJ822 on J1 is defective.


Remarks:
Fault Identification: Check the cable 40/CJ822 on J1 for:
• Electric continuity
• Electric connection
• Cable condition

Solution: Replace the defective cable.

Possible Cause: The power supply cable is defective.


Remarks:
Fault Identification: Check the cable 979/CJ822/J2 - 979/CJ732/J16 for:
• Electric continuity
• Electric connection
• Cable condition

Solution: Replace the defective power supply cable.

Possible Cause: The flag detection problem.


Remarks:
Fault Identification: Check to see if the flag is:
• Installed
• Correctly positioned
• Correctly fixed with the fixing screws
• Correctly passing through the optical detector

Solution: • Install the flag.


• Position correctly the flag.
• Fix correctly the fixing screws of the flag.

7–16 ERROR MESSAGES


E16: Movement Too Short

E16: Movement Too Short

Definition of the Cause


During calibration of the rotative arm movement, the microprocessor has detected a 10% value
inferior to the theoretical value of 360° electric pulse waves.

Solution of the Cause

Possible Cause: Incorrect selection of the rotation wheel size.


Remarks:
Fault Identification:
Solution:

1. Access the technical menu. Click select Maintenance >


Manufacturing parameters.

2. In the Manufacturing parameters dialog box, check the “Size of the


transmission wheel” field to see if the size of the wheel indicated is correct.

3. If the size of the indicated wheel is incorrect, select the appropriate wheel
size.

4. Click Save.

Possible Cause: Mechanical locking due to incorrect tightening of the driving wheel.
Remarks:
Fault Identification: • Check the tightening of the screws of the rotative arm motor.
• Check the tightening of the screws of the drive roll.
• Check the rotation of the optical encoder.

Solution: If any mechanical locking change the optical encoder.

Possible Cause: The driving wheel is skidding.


Remarks:
Fault Identification: The motor is turning and the driving wheel is skidding.
Solution: • Clean the track with a soft damp cloth.
• Tighten the driving wheel on the motor shaft.

KODAK 9000 and KODAK 9000 3D Extraoral Imaging System_Service Guide (SM712)_Ed 02 7–17
E16: Movement Too Short

Possible Cause: • The optical encoder is defective or incorrectly positioned.


• The code selector is defective or incorrectly positioned.
Remarks: Absence of the electric pulse wave detection during the rotative arm rotation.
Fault Identification: Using LD22 and LD26 on the CJ732, check the security and condition of:
• The optical encoder
• The optical encoder driving belt
• The encode wheel

Solution: Replace the defective component.

Possible Cause: The cable 40/CJ822 on J1 is defective.


Remarks:
Fault Identification: Check the cable 40/CJ822 on J1 for:
• Electric continuity
• Electric connection
• Cable condition

Solution: Replace the defective cable.

Possible Cause: The power supply cable is defective.


Remarks:
Fault Identification: Check the cable 979/CJ822/J2 and 979/CJ732/J16 for:
• Electric continuity
• Electric connection
• Cable condition

Solution: Replace the defective power supply cable.

Possible Cause: The flag detection problem.


Remarks:
Fault Identification: Check to see if the flag is:
• Installed
• Correctly positioned
• Correctly fixed with the fixing screws
• Correctly passing through the optical detector

Solution: • Install the flag.


• Position correctly the flag.
• Fix correctly the fixing screws of the flag.

7–18 ERROR MESSAGES


E19: Distance Too Short

E19: Distance Too Short

Definition of the Cause


During calibration, detection of incorrect rotation movement distance of the rotative arm.

Solution of the Cause

Possible Cause: Damage to the unit head during transportation.


Remarks:
Fault Identification:
Solution: Check the rotative movement manually if damaged then change the unit head.

Possible Cause: The optical encode wheel is defective.


Remarks:
Fault Identification: Check the optical encoder wheel using LD22 and LD26 of the CJ732 board
Solution: Replace the encode wheel.

Possible Cause: The mechanical locking of the optical encode wheel.


Remarks:
Fault Identification:
Solution: Replace the optical encoder assembly.

Possible Cause: The optical encoder cable is defective.


Remarks:
Fault Identification:
Solution: Replace the optical encoder cable.

KODAK 9000 and KODAK 9000 3D Extraoral Imaging System_Service Guide (SM712)_Ed 02 7–19
E20: Distance Too Long

E20: Distance Too Long

Definition of the Cause


During calibration, detection of incorrect rotation movement distance of the rotative arm.

Solution of the Cause

Possible Cause: Damage to the unit head during transportation.


Remarks:
Fault Identification:
Solution: Check the rotative movement manually if damage then change the unit head.

7–20 ERROR MESSAGES


E22: Acquisition is stopped the Interface is Masked

E22: Acquisition is stopped the Interface is


Masked

Definition of the Cause


The Acquisition Window must be always visible and accessible during the x-ray emission. If
it is inactive or masked totally or partially by other applications, the x-ray emission will be
interrupted prematurely.

Solution of the Cause

Possible Cause: The Acquisition Window is not active or masked.


Remarks:
Fault Identification: • The user launches the acquisition using the exposure button of the x-ray remote
control but there is no response.
• The Control Panel of the fixed arm displays an Information Message.

Solution: You must open and keep visible the Acquisition Window.

KODAK 9000 and KODAK 9000 3D Extraoral Imaging System_Service Guide (SM712)_Ed 02 7–21
E23: No movement is detected

E23: No movement is detected

Definition of the Cause


During the x-ray emission, the microprocessor has not detected a rotative arm movement.

Solution of the Cause

Possible Cause: Rotation end limit switch defective.


Remarks: Switch located inside the rotative arm.
Fault Identification: • Check cable 40/CJ732/J13.
• Check the condition of the switch.
Solution: • Replace the switch.
• Replace cable 40/CJ732/J13.

Possible Cause: The rotative arm motor is defective.


Remarks:
Fault Identification: The rotative arm motor cannot be turned on.
Solution: Replace the rotative arm movement motor assembly.

Possible Cause: • The optical encoder is defective or incorrectly positioned.


• The code selector is defective or incorrectly positioned.
Remarks: Absence of the electric pulse wave detection during the rotative arm rotation.
Fault Identification: Using LD22 and LD26 on the CJ732, check the security and condition of:
• The optical encoder
• The optical encoder driving belt
• The encode wheel

Solution: Replace the defective component.

Possible Cause: Incorrect tightening of the driving wheel.


Remarks:
Fault Identification:
Solution: • Tighten the driving wheel correctly.
• Calibrate the rotative arm movement.

7–22 ERROR MESSAGES


E23: No movement is detected

Possible Cause: The driving wheel is skidding.


Remarks:
Fault Identification: The motor is turning and the driving wheel is skidding.
Solution: • Clean the track with a soft damp cloth.
• Tighten the driving wheel on the motor shaft.

Possible Cause: The cable 40/CJ822 on J1 is defective.


Remarks:
Fault Identification: Check the cable 40/CJ822 on J1 for:
• Electric continuity
• Electric connection
• Cable condition

Solution: Replace the defective cable.

Possible Cause: The power supply cable 979 is defective.


Remarks:
Fault Identification: Check the cable 979/CJ822/J2 - 979/CJ732/J16 for:
• Electric continuity
• Electric connection
• Cable condition

Solution: Replace the defective power supply cable.

Possible Cause: The flag detection problem.


Remarks:
Fault Identification: Check to see if the flag is:
• Installed
• Correctly positioned
• Correctly fixed with the fixing screws
• Correctly passing through the optical detector

Solution: • Install the flag.


• Position correctly the flag.
• Fix correctly the fixing screws of the flag.

KODAK 9000 and KODAK 9000 3D Extraoral Imaging System_Service Guide (SM712)_Ed 02 7–23
E24: Heating Voltage Self-calibration Time-out

E24: Heating Voltage Self-calibration Time-out


E25: Heating Voltage Lower Limit Reached
E26: Heating Voltage Higher Limit Reached

Definition of the Cause


During calibration, the microprocessor has failed to attain the specified mA values.

Solution of the Cause

Possible Cause: A bad heating voltage calibration.


Remarks:
Fault Identification:
Solution: Recalibrate the heating voltage correctly.

Possible Cause: A cable is defective.


Remarks:
Fault Identification: Check the connecting cable 14/CJ732 on J4 and cable 16/CJ732 on J12 for:
• Electric continuity
• Electric connections
• Cable condition

Solution: Replace the defective component.

Possible Cause: The cause is one of the following:


• The value of the supply voltage to the heating circuit is incorrect.
• The power supply to the unit has failed.

Remarks: In this case, the mA can only be too low.


Fault Identification: Measure the supply voltage at the output terminals of the power supply on the +5 V.

WARNING:
This value must be set to 6.5 V DC
Solution: • Adjust the voltage.
• Replace the power supply.

7–24 ERROR MESSAGES


E26: Heating Voltage Higher Limit Reached

Possible Cause: The following is defective:


• The CJ732 board
• The x-ray generator

Remarks:
Fault Identification: Measurements must be performed on the defective components.
Solution: • Replace the CJ732 board.
• Replace the x-ray generator.

Possible Cause: The filament of the x-ray tube is open.


Remarks: In this case, the mA values can only be too low or absent
Fault Identification: Measure the value of the filament resistor.
Solution: Replace the x-ray generator.

KODAK 9000 and KODAK 9000 3D Extraoral Imaging System_Service Guide (SM712)_Ed 02 7–25
E27: No Detection of the Limit Switches X or Y

E27: No Detection of the Limit Switches X or Y

Definition of the Cause


No detection of the limit switches X or Y after a 20 second time-out.

Solution of the Cause

Possible Cause: The X or Y limit switches are defective.


Remarks:
Fault Identification: The XY motor assembly runs but the limit switches are immobile.
Solution: Check the 4 bearings, if defective replace them.

Possible Cause: The following components can be defective:


• The motor assembly X or Y
• The motor assembly cables X or Y
• The motor power supply
• The motor board CJ708
• The microprocessor board CJ732

Remarks:
Fault Identification: • Check the X limit switches using LD4 and LD5 on the CJ732 board.
• Check the Y limit switches using LD10 and LD11 on the CJ732 board.
Solution: Replace the defective component.

7–26 ERROR MESSAGES


E28: Sector Collimator Error Movement

E28: Sector Collimator Error Movement

Definition of the Cause


Incorrect panoramic or 3D positioning of the sector collimator.

Solution of the Cause

Possible Cause: Mechanical locking of the sector collimator motor assembly.


Remarks:
Fault Identification:
Solution: Adjust the sector collimator motor assembly.

Possible Cause: The sector collimator select motor assembly is not active.
Remarks:
Fault Identification: • Check the motor activity on the J23 of the microprocessor CJ732 board.
• Check the cables and the connections.
• Check the DC collimator motor CJ804 board.
• Check the collimator motor assembly.

Solution: Replace the defective component.

KODAK 9000 and KODAK 9000 3D Extraoral Imaging System_Service Guide (SM712)_Ed 02 7–27
E29: Digital Sensor Error Movement

E29: Digital Sensor Error Movement

Definition of the Cause


Incorrect positioning of the panoramic or 3D digital sensor.

Solution of the Cause

Possible Cause: Mechanical locking of the sensor select motor assembly.


Remarks:
Fault Identification:
Solution: Adjust the sensor select motor assembly.

Possible Cause: The sensor select motor assembly is not active.


Remarks:
Fault Identification: • Check the motor activity on the J22 of the microprocessor CJ732 board.
• Check the cables and the connections.
• Check the DC sensor motor CJ804 board and the DIP-Switch configuration.
• Check the sensor select motor assembly.

Solution: Replace the defective components.

Possible Cause: • The digital sensor cover friction.


• Twisting of the digital sensor cables.
Remarks:
Fault Identification:
Solution: • Reposition correctly the digital sensor cover.
• Disconnect the digital sensor cables and reconnect them correctly.

Possible Cause: The digital sensor limit switch is defective.


Remarks:
Fault Identification:
Solution: Replace the defective digital sensor limit switch.

7–28 ERROR MESSAGES


E30: Limit Switch X

E30: Limit Switch X

Definition of the Cause


No detection of the limit switch X.

Solution of the Cause

Possible Cause: Check the limit switch X function using LD4 and LD5 on the CJ732 board.
Remarks:
Fault Identification:
Solution: • Replace the defective limit switch.
• Replace the defective cable.

KODAK 9000 and KODAK 9000 3D Extraoral Imaging System_Service Guide (SM712)_Ed 02 7–29
E31: Distance Too Short X (Calibration)

E31: Distance Too Short X (Calibration)

Definition of the Cause


The distance between the reference switch and the central positioning of the rotative arm ?????
is too short.

Solution of the Cause

Possible Cause: The central positioning of the rotative arm is not correctly adjusted.
Remarks:
Fault Identification:
Solution: Adjust central positioning of the rotative arm using the central positioning laser.

7–30 ERROR MESSAGES


E32: Distance Too Long X (Calibration)

E32: Distance Too Long X (Calibration)

Definition of the Cause


The distance between the reference switch and the central positioning of the rotative arm ?????
is too long.

Solution of the Cause

Possible Cause: The central positioning of the rotative arm is not correctly adjusted.
Remarks:
Fault Identification:
Solution: Adjust central positioning of the rotative arm using the central positioning laser.

Possible Cause: The following components can be defective:


• The motor assembly X
• The microprocessor board CJ732
There could be a mechanical locking of the rotative arm.
Remarks:
Fault Identification:
Solution: Replace the defective component.

KODAK 9000 and KODAK 9000 3D Extraoral Imaging System_Service Guide (SM712)_Ed 02 7–31
E33: Limit Switch X

E33: Limit Switch X

Definition of the Cause


No detection of the limit switch X.

Solution of the Cause

Possible Cause: Check the limit switch X function using LD4 and LD5 on the CJ732 board.
Remarks:
Fault Identification:
Solution: • Replace the defective limit switch.
• Replace the defective cable.

7–32 ERROR MESSAGES


E34: Distance Too Short X (Calibration)

E34: Distance Too Short X (Calibration)


E35: Distance Too Long X (Calibration)

Definition of the Cause


The rotative arm is incorrectly positioned at the end of the full travel movement.

Solution of the Cause

Possible Cause: • The Mechanical locking of the double pitch worm X.


• An incorrect motor cabling.
Remarks:
Fault Identification: Check the motor cabling.
Solution: Adjust the double pitch worm X.

Possible Cause: The following components can be defective:


• The motor assembly X
• The microprocessor board CJ732
There could be a mechanical locking of the rotative arm.
Remarks:
Fault Identification:
Solution: Replace the defective component.

KODAK 9000 and KODAK 9000 3D Extraoral Imaging System_Service Guide (SM712)_Ed 02 7–33
E36: Limit Switch Y

E36: Limit Switch Y

Definition of the Cause


No detection of the limit switch Y.

Solution of the Cause

Possible Cause: Check the limit switch Y function using LD10 and LD11 on the microprocessor
CJ732 board.
Remarks:
Fault Identification:
Solution: • Replace the defective limit switch.
• Replace the defective cable.

7–34 ERROR MESSAGES


E37: Distance Too Short Y (Calibration)

E37: Distance Too Short Y (Calibration)

Definition of the Cause


The distance between the reference switch and the central positioning of the rotative arm ?????
is too short.

Solution of the Cause

Possible Cause: The central positioning of the rotative arm is not correctly adjusted.
Remarks:
Fault Identification:
Solution: Adjust central positioning of the rotative arm using the central positioning laser.

KODAK 9000 and KODAK 9000 3D Extraoral Imaging System_Service Guide (SM712)_Ed 02 7–35
E38: Distance Too Long Y (Calibration)

E38: Distance Too Long Y (Calibration)

Definition of the Cause


The distance between the reference switch and the central positioning of the rotative arm ?????
is too long.

Solution of the Cause

Possible Cause: The central positioning of the rotative arm is not correctly adjusted.
Remarks:
Fault Identification:
Solution: Adjust central positioning of the rotative arm using the central positioning laser.

Possible Cause: The following components can be defective:


• The motor assembly Y
• The microprocessor board CJ732
There could be a mechanical locking of the rotative arm.
Remarks:
Fault Identification:
Solution: Replace the defective component.

7–36 ERROR MESSAGES


E39: Limit Switch Y

E39: Limit Switch Y

Definition of the Cause


No detection of the limit switch Y.

Solution of the Cause

Possible Cause: Check the limit switch Y function using LD10 and LD11 on the microprocessor
CJ732 board.
Remarks:
Fault Identification:
Solution: • Replace the defective limit switch.
• Replace the defective cable.

KODAK 9000 and KODAK 9000 3D Extraoral Imaging System_Service Guide (SM712)_Ed 02 7–37
E40: Distance Too Short Y (Calibration)

E40: Distance Too Short Y (Calibration)


E41: Distance Too Long Y (Calibration)

Definition of the Cause


The rotative arm is incorrectly positioned at the end of the full travel movement.

Solution of the Cause

Possible Cause: • The Mechanical locking of the double pitch worm Y.


• An incorrect motor cabling.
Remarks:
Fault Identification: Check the motor cabling.
Solution: Adjust the double pitch worm Y.

Possible Cause: The following components can be defective:


• The motor assembly Y
• The microprocessor board CJ732
There could be a mechanical locking of the rotative arm.
Remarks:
Fault Identification:
Solution: Replace the defective component.

7–38 ERROR MESSAGES


E42: X or Y Position not Identified

E42: X or Y Position not Identified

Definition of the Cause


The X or Y position not memorized or position unknown.

Solution of the Cause

Possible Cause: The microprocessor board CJ732 was replaced but not calibrated.
Remarks:
Fault Identification:
Solution: Calibrate the microprocessor board CJ732.
To calibrate the microprocessor board CJ732, follow these steps:

1. Switch on the unit in technical mode.

2. On the PC, click Ok button of the error message dialog box to clear the screen.

3. On the unit control panel, press and hold the X button to launch the X axe full
movement until the end.

4. On the unit control panel, press and hold the Y button to launch the Y axe full
movement until the end.

If needed replace the microprocessor board CJ732.

KODAK 9000 and KODAK 9000 3D Extraoral Imaging System_Service Guide (SM712)_Ed 02 7–39
E47: Rotation Movement Too Short

E47: Rotation Movement Too Short


E48: Rotation Movement Too Long

Definition of the Cause


During rotation movement, the microprocessor has detected during the 360° rotation, a 10%
superior or inferior value.

Solution of the Cause

Possible Cause: Incorrect selection of the rotation wheel size.


Remarks:
Fault Identification:
Solution:

1. Access the technical menu. Click select Maintenance >


Manufacturing parameters.

2. In the Manufacturing parameters dialog box, check the “Size of the


transmission wheel” field to see if the size of the wheel indicated is correct.

3. If the size of the indicated wheel is incorrect, select the appropriate wheel
size.

4. Click Save.

Possible Cause: The driving wheel is skidding.


Remarks:
Fault Identification: The motor is turning and the driving wheel is skidding.
Solution: • Clean the track with a soft damp cloth.
• Tighten the driving wheel on the motor shaft.

Possible Cause: A cable is defective


Remarks:
Fault Identification: Check the cable 22/CJ732/J11 and cable 22/CJ708/J1 for:
• Electric continuity
• Electric connection
• Cable condition

7–40 ERROR MESSAGES


E48: Rotation Movement Too Long

Solution: Replace the defective component.

Possible Cause: Mechanical locking due to bad tightening of the driving wheel.
Remarks:
Fault Identification: • Check the tightening of the screws of the rotative arm motor.
• Check the tightening of the screws of the drive roll.
• Check the rotation of the optical encoder.

Solution: If any mechanical locking change the optical encoder.

Possible Cause: Bad tightening of the drive roll.


Remarks:
Fault Identification:
Solution: • Tighten the drive roll.
• Calibrate the rotative arm movement.

KODAK 9000 and KODAK 9000 3D Extraoral Imaging System_Service Guide (SM712)_Ed 02 7–41
E49: Cephalostat Positioning Blocker Time-out

E49: Cephalostat Positioning Blocker Time-out

Definition of the Cause


There is a time-out detection by the system because the cephalostat positioning blocker is too
slow or does not respond to positioning request.

Solution of the Cause

Possible Cause: Incorrect or bad cabling of the positioning blocker.


Remarks:
Fault Identification: • Check to see if the LED on the induction cable of the positioning blocker is lit.
• Check cable 982/CJ822 on J3 for electric connection and cable condition.
• Check the board CJ822.

Solution: • Insert the induction cable further to see if the LED will lit.
• Replace the cable 982/CJ822/J3 if defective.
• Replace the board CJ822 if defective.

Possible Cause: Power supply cable is defective.


Remarks:
Fault Identification: • Check that the cable 981/CJ804/J2 is correctly soldered to the positioning
blocker.
• Check the cable 981/CJ804 on J2 for electric connection and cable condition.
• Check the board CJ804.

Solution: • Resolder the cable 981/CJ804/J2 to the positioning blocker.


• Replace the cable 981/CJ804/J2 if defective.
• Replace the board CJ804 if defective.

Possible Cause: A connection problem between the board CJ804 and CJ732.
Remarks:
Fault Identification: • Check cable 980/CJ804/J1 and 980/CJ732/J23 for electric connection and cable
condition.
• Check the board CJ804.
• Check the board CJ732.

Solution: • Replace the cable 980/CJ804/J1 and 980/CJ732/J23 if defective.


• Replace the board CJ804 if defective.
• Replace the board CJ732 if defective.

7–42 ERROR MESSAGES


E50: Watchdog Error

E50: Watchdog Error

Definition of the Cause


The microprocessor board CJ732 has detected a malfunctioning of the sequence of the
program.

Solution of the Cause

Possible Cause: Spikes or interferences in the power supply line.


Remarks:
Fault Identification: Check the electrical installation of:
• The wiring to and from the unit
• The grounding of the installation
• The section of the cables
• The routing of the unit cables close to components which emit significant
interference
• The presence of other devices connected to the unit power supply line

Solution: If a fault is detected, the installation must be rendered compliant.

Possible Cause: The microprocessor board CJ732 is defective.


Remarks:
Fault Identification:
Solution: Replace the CJ732 board.

KODAK 9000 and KODAK 9000 3D Extraoral Imaging System_Service Guide (SM712)_Ed 02 7–43
E51: Communication Problem with serial EEPROM

E51: Communication Problem with serial


EEPROM

Definition of the Cause


The microprocessor board CJ732 has detected a communication problem with the EEPROM.

Solution of the Cause

Possible Cause: The microprocessor board CJ732 is defective.


Remarks:
Fault Identification:
Solution: Replace the microprocessor board CJ732.

7–44 ERROR MESSAGES


E52: Communication Problem with Console

E52: Communication Problem with Console

Definition of the Cause


The microprocessor board CJ732 has detected a communication problem with the unit control
panel.

Solution of the Cause

Possible Cause: A cable is defective.


Remarks:
Fault Identification: Check connecting cable 1/C732 on J3 for:
• Electric continuity
• Electric connections
• Cable condition

Solution: Replace the defective cable.

Possible Cause: The Control Panel is defective.


Remarks:
Fault Identification: The display screen of the unit control panel is turned off.
Solution: Replace the Control Panel.

Possible Cause: The microprocessor board CJ732 is defective.


Remarks:
Fault Identification:
Solution: Replace the microprocessor board CJ732.

KODAK 9000 and KODAK 9000 3D Extraoral Imaging System_Service Guide (SM712)_Ed 02 7–45
E53: Backup Battery out of Service

E53: Backup Battery out of Service

Definition of the Cause


The microprocessor board CJ732 battery is exhausted. The operative life of the clock battery
in the unit is approximately 10 years.

WARNING
You MUST not incinerate or discard the batteries into the
environment. You MUST take them to the collection point for
used batteries.

Solution of the Cause

Possible Cause: The microprocessor board CJ732 battery quartz module is exhausted.
Remarks:
WARNING:
You can replace the exhausted battery quartz module with either U14 or U15
battery.
Fault Identification:
Solution: Replace the exhausted battery quartz module.

Possible Cause: The microprocessor board CJ732 is defective.


Remarks:
Fault Identification:
Solution: Replace the microprocessor board CJ732.

7–46 ERROR MESSAGES


E54: Inconsistency in Calibration Parameters

E54: Inconsistency in Calibration Parameters


E55: Error in Stored Exposure Parameters
E56: Error in Manufacturing Parameters
E57: Error in User Preferences

Definition of the Cause


There is an inconsistency in memorized parameter data.

Solution of the Cause

Possible Cause: Spikes or interferences in the power supply line.


Remarks:
Fault Identification: Check the following electrical installation:
• The wiring to and from the unit
• The grounding of the installation
• The section of the cables
• The routing of the unit cables close to components which emit significant
interferences
• The presence of other devices connected to the unit power supply line

Solution: If a fault is detected, the installation must be rendered compliant.

Possible Cause: • The microprocessor board CJ732 is defective.


• The microprocessor board CJ732 has been replaced but not initialized.
Remarks:
Fault Identification:

KODAK 9000 and KODAK 9000 3D Extraoral Imaging System_Service Guide (SM712)_Ed 02 7–47
E57: Error in User Preferences

Solution: To replace the microprocessor board CJ732, follow these steps:

1. Access the tech menu. click select Maintenance > Save / Restore.

2. Click Save to save the current calibration and system parameters.

3. Click select Maintenance > Manufacturing Parameters.

4. Write down all the displayed parameters. Click Cancel.

5. Replace the microprocessor board CJ732.

1. Access the tech menu. click select Maintenance > Save / Restore.

2. Click Restore. Check that the displayed parameters are similar to the initial
parameters that you had written down when saving the system parameters.

3. If the parameters are not similar to the initial parameters, you must redo the
following calibrations:

• mA calibration
• Heating voltage calibration
IS THIS THE SYSTEMATIC PROCEDURE EVERY TIME THE CJ732 IS
REPLACED????????????

7–48 ERROR MESSAGES


E58: Exposure Context Error

E58: Exposure Context Error

Definition of the Cause


There is an inconsistency in memorized parameter data.

WARNING
You MUST not incinerate or discard the batteries into the
environment. You MUST take them to the collection point for
used batteries.

Solution of the Cause

Possible Cause: The microprocessor board CJ732 battery is exhausted.


Remarks:
WARNING:
You can replace the exhausted battery with either U14 or U15 battery.
Fault Identification:
Solution: Replace the exhausted battery.

Possible Cause: The microprocessor board CJ732 is defective.


Remarks:
Fault Identification:
Solution: Replace the microprocessor board CJ732.

KODAK 9000 and KODAK 9000 3D Extraoral Imaging System_Service Guide (SM712)_Ed 02 7–49
E59: Program Evolution Error

E59: Program Evolution Error

Definition of the Cause


The microprocessor board CJ732 has detected a malfunctioning of the sequence of the
program.

Solution of the Cause

Possible Cause: Spikes or interferences in the power supply line.


Remarks:
Fault Identification: Check the electrical installation of:
• The wiring to and from the unit
• The grounding of the installation
• The section of the cables
• The routing of the unit cables close to components which emit significant
interference
• The presence of other devices connected to the unit power supply line

Solution: If a fault is detected, the installation must be rendered compliant.

Possible Cause: The microprocessor board CJ732 is defective.


Remarks:
Fault Identification:
Solution: Replace the CJ732 board.

7–50 ERROR MESSAGES


E60: Bus I²C Error

E60: Bus I²C Error

Definition of the Cause


The microprocessor board CJ732 detects a malfunction in the circuit which provides
communication with the unit control panel and/or the blade collimator.

Solution of the Cause

Possible Cause: This maybe a transient problem.


Remarks:
Fault Identification:
Solution: Turn off the unit and then turn it on again.

Possible Cause: A cable is defective.


Remarks:
Fault Identification: Check connecting cables 1/CJ732 on J3, 973/CJ732 on J20 - 973/CJ821 on J17 and
974/CJ821 on J1 - 974/JZ029 on CN51, for:
• Electric continuity
• Electric connections
• Cable condition

Solution: Replace the defective cable.

Possible Cause: • The microprocessor board CJ732 is defective.


• The Control Panel is defective.
• The blade collimator pilot board.

Remarks:
Fault Identification: Bus I²C is defective.
Solution: • Replace the CJ732 board.
• Replace the Control Panel.
• Replace the blade collimator pilot board.

KODAK 9000 and KODAK 9000 3D Extraoral Imaging System_Service Guide (SM712)_Ed 02 7–51
E62: Panoramic Trajectory Error

E62: Panoramic Trajectory Error

Definition of the Cause


The PC transmits incoherent panoramic trajectory data to the unit.

Solution of the Cause

Possible Cause: The PC transmits incoherent panoramic trajectory data to the unit.
Remarks:
Fault Identification:
Solution: • Close and reopen the Acquisition Window.
• Reinstall the Kodak dental imaging software.

7–52 ERROR MESSAGES


E63: EPROM Detection Failure

E63: EPROM Detection Failure

Definition of the Cause


The checksum of the PROGRAM EPROM data is incorrect.

Solution of the Cause

Possible Cause: • The microprocessor board CJ732 is defective.


• The EPROM is defective
Remarks:
Fault Identification:
Solution: • Replace the microprocessor board CJ732.
• Replace the 2 EEPROMs on the U27 and U28 of the CJ732.

KODAK 9000 and KODAK 9000 3D Extraoral Imaging System_Service Guide (SM712)_Ed 02 7–53
E64: Transfer Failure

E64: Transfer Failure

Definition of the Cause


The trajectory transfer time is too long, therefore there is transfer failure.

Solution of the Cause

Possible Cause: • The PC does not respond.


• The PC has crashed.
• The PC is processing other operations or applications.

Remarks:
Fault Identification:
Solution: • Relaunch an image acquisition.
• Reboot the PC.
• Close all the active applications.

Possible Cause:
Remarks:
Fault Identification:
Solution:

7–54 ERROR MESSAGES


E65: RTC Failure

E65: RTC Failure

Definition of the Cause


There is a Real Time Clock (RTC) failure.

The operative life of the clock battery in the unit is approximately 10 years.

Solution of the Cause

Possible Cause: The RTC is not correctly pugged in.


Remarks:
WARNING:
You can replace the battery quartz module with either U14 or U15 battery.
Fault Identification:
Solution: • Check that the RTC is correctly plugged in.
• Replace the battery quartz module.

Possible Cause: The microprocessor board CJ732 is defective.


Remarks:
Fault Identification:
Solution: Replace the microprocessor board CJ732.

KODAK 9000 and KODAK 9000 3D Extraoral Imaging System_Service Guide (SM712)_Ed 02 7–55
E66: Bus I2C Error

E66: Bus I2C Error

Definition of the Cause


The I2C bus is blocked.

Solution of the Cause

Possible Cause: One of the following Firmware version is incorrect:


• CJ732
• CJ806
• Blade collimator pilot board

Remarks:
Fault Identification:
Solution: Contact your representative.

7–56 ERROR MESSAGES


E70: Communication Loss with Collimator

E70: Communication Loss with Collimator

Definition of the Cause


The microprocessor board CJ732 has detected a communication problem with the blade
collimator.

Solution of the Cause

Possible Cause: A cable is defective.


Remarks:
Fault Identification: Check connecting cable 973/C732 on J20 - 973/CJ821 on J17, for:
• Electric continuity
• Electric connections
• Cable condition

Solution: Replace the defective cable.

Possible Cause: The blade collimator pilot board is defective.


Remarks:
Fault Identification:
Solution: Replace the blade collimator pilot board.

Possible Cause: The microprocessor board CJ732 is defective.


Remarks:
Fault Identification:
Solution: Replace the microprocessor board CJ732.

KODAK 9000 and KODAK 9000 3D Extraoral Imaging System_Service Guide (SM712)_Ed 02 7–57
E71: Data Error with the Blade Collimator

E71: Data Error with the Blade Collimator

Definition of the Cause


There is a communication problem between the microprocessor board CJ732 and the blade
collimator pilot board.

Solution of the Cause

Possible Cause: A cable is defective.


Remarks:
Fault Identification: Check connecting cable 973/C732 on J20 - 973/CJ821 on J17, for:
• Electric continuity
• Electric connections
• Cable condition

Solution: Replace the defective cable.

Possible Cause: The blade collimator pilot board is defective.


Remarks:
Fault Identification:
Solution: Replace the blade collimator pilot board.

Possible Cause: The microprocessor board CJ732 is defective.


Remarks:
Fault Identification:
Solution: Replace the microprocessor board CJ732.

7–58 ERROR MESSAGES


E72: Firmware Version not Compatible with Blade Collimator

E72: Firmware Version not Compatible with


Blade Collimator

Definition of the Cause


The blade collimator Firmware version is not supported by the Microprocessor CJ732 board.

Solution of the Cause

Possible Cause: Incompatible Firmware version.


Remarks:
Fault Identification: The blade collimator pilot board has been replaced and is no more compatible with
the microprocessor board CJ732.
Solution: Replace the microprocessor board CJ732 EEPROMS with the appropriate Firmware
version.

KODAK 9000 and KODAK 9000 3D Extraoral Imaging System_Service Guide (SM712)_Ed 02 7–59
E73: Time-Out for Collimator Positioning

E73: Time-Out for Collimator Positioning

Definition of the Cause


There is a long delay in the blade collimator positioning for the requested acquisition program.

Solution of the Cause

Possible Cause: Mechanical blockage of the blade collimator.


Remarks:
Fault Identification: During the power on initialisation of the unit, observe the blade movement for any
mechanical blockage.
Solution: Replace the blade collimator.

Possible Cause: Electrical problem with the blade collimator pilot board.
Remarks:
Fault Identification: Check cables 4 to 11, for:
• Electric continuity
• Electric connections
• Cable condition

Solution: Replace the defective blade collimator and/or blade collimator pilot board.

7–60 ERROR MESSAGES


E74: Blade Collimator Mechanical Failure

E74: Blade Collimator Mechanical Failure

Definition of the Cause


There is a problem with the blade collimator movement during the power on initialization of
the unit.

Solution of the Cause

Possible Cause: Mechanical blockage of the blade collimator.


Remarks:
Fault Identification: During the power on initialisation of the unit, observe the blade stroke of the
collimator for any mechanical blockage.
Solution: Replace the blade collimator.

Possible Cause: Electrical problem with the blade collimator pilot board.
Remarks:
Fault Identification: Check cables 4 to 11, for:
• Electric continuity
• Electric connections
• Cable condition

Solution: Replace the defective blade collimator and/or blade collimator pilot board.

KODAK 9000 and KODAK 9000 3D Extraoral Imaging System_Service Guide (SM712)_Ed 02 7–61
E75: Failure in Collimator Positioning

E75: Failure in Collimator Positioning

Definition of the Cause


There is a problem with the blade collimator positioning for the requested acquisition
program.

Solution of the Cause

Possible Cause: A cable is defective.


Remarks:
Fault Identification: Check connecting cable 973/C732 on J20 - 973/CJ821 on J17, for:
• Electric continuity
• Electric connections
• Cable condition

Solution: Replace the defective cable.

Possible Cause: The blade collimator is defective.


Remarks:
Fault Identification:
Solution: Replace the blade collimator.

Possible Cause: The microprocessor board CJ732 is defective.


Remarks:
Fault Identification:
Solution: Replace the microprocessor board CJ732.

7–62 ERROR MESSAGES


E76: Collimator Calibration Parameter not Valid

E76: Collimator Calibration Parameter not Valid

Definition of the Cause


Invalid calibration parameters.

Solution of the Cause

Possible Cause: Wrong calibration parameter of one of the following modes:


• Panoramic
• 3D
• Cephalometric

Remarks:
Fault Identification:
Solution: Recalibrate the wrong parameter for one of the following modes:
• Panoramic
• 3D
• Cephalometric

KODAK 9000 and KODAK 9000 3D Extraoral Imaging System_Service Guide (SM712)_Ed 02 7–63
No movement of the Column

No movement of the Column

Definition of the Cause


The column is immobile and no movement is detected.

Solution of the Cause

Possible Cause: One of the F1 or F2 fuses of the CJ699 board is burned.


Remarks:
Fault Identification: Check the fuses.
Solution: Replace the defective fuse.

Possible Cause: The electric linear actuator is defective.


Remarks:
Fault Identification: • Check the electric continuity of the electric linear actuator.
• Check the connecting cable 5/CJ699 on J6.
• Check the connecting cable 36/CJ699 on J5.
• Check connecting cable 37/CJ699 on J8.

Solution: Replace the electric linear actuator.

Possible Cause: The CJ699 board is defective.


Remarks:
Fault Identification:
Solution: Replace the CJ699 board.

7–64 ERROR MESSAGES


E101 to E116: JBUS Ethernet Communication Problem

E101 to E116: JBUS Ethernet Communication


Problem

Definition of the Cause


The Acquisition interface cannot communicate with the microprocessor CJ732 board.

Solution of the Cause

Possible Cause: There is no communication between the digital sensor RS232 cable with x-ray
synchro and the microprocessor CJ732 board.
Remarks:
Fault Identification: Check the digital sensor RS232 cable with x-ray synchro 42/CJ732 on J15, for:
• Electric continuity
• Electric connections
• Cable condition

Solution: Replace the defective cable.

Possible Cause: There is no communication between the digital sensor RS232 cable with x-ray
synchro and the microprocessor CJ732 board.
Remarks:
Fault Identification: Check the microprocessor CJ732 board.
Solution: Replace the defective CJ732 board.

KODAK 9000 and KODAK 9000 3D Extraoral Imaging System_Service Guide (SM712)_Ed 02 7–65
E201: Problem with Image Reception

E201: Problem with Image Reception

Definition of the Cause


The x-ray is launched for image acquisition but no image is acquired.

Solution of the Cause

Possible Cause: The possible cause is unknown and must be researched by Trophy.
Remarks:
Fault Identification:
Solution: Contact Trophy.

7–66 ERROR MESSAGES


E301: Bad Driver Version

E301: Bad Driver Version

Definition of the Cause


The K9000 firmware version and the driver version are not compatible.

Solution of the Cause

Possible Cause: The updated firmware was not correctly executed.


Remarks:
Fault Identification:
Solution: You must update the firmware. Follow these step to update the firmware:
1. Check to see if the DHCP/TFTP server is running. If not, restart the PC.

2. Restart the unit to start the firmware upload.

KODAK 9000 and KODAK 9000 3D Extraoral Imaging System_Service Guide (SM712)_Ed 02 7–67
E302: Driver Configuration Error. Please Reinstall the Driver

E302: Driver Configuration Error. Please


Reinstall the Driver

Definition of the Cause


There is an incoherent driver configuration.

Solution of the Cause

Possible Cause: • Some files are missing.


• There is an incorrect configuration in the registry.
Remarks:
Fault Identification:
Solution: You must reinstall the unit driver package.

7–68 ERROR MESSAGES


E305: Mask File not Found

E305: Mask File not Found

Definition of the Cause


When the Kodak dental imaging software is launched, the PC cannot find the mask files.

Solution of the Cause

Possible Cause: The mask files are missing.


Remarks:
Fault Identification:
Solution: Install the mask files using the mask file CD-ROM.

KODAK 9000 and KODAK 9000 3D Extraoral Imaging System_Service Guide (SM712)_Ed 02 7–69
E306: Unknown Radiological Unit

E306: Unknown Radiological Unit

Definition of the Cause


The PC failed to load the ODT2000.dll files.

Solution of the Cause

Possible Cause: The ODT2000.dll files are either missing or incompatible.


Remarks:
Fault Identification:
Solution: Reinstall the unit driver package.

7–70 ERROR MESSAGES


E311: No Sensor Detected on the Radiological Unit

E311: No Sensor Detected on the Radiological


Unit

Definition of the Cause


The PC cannot detect the digital sensor of the unit.

Solution of the Cause

Possible Cause: There is a hardware component problem.


Remarks:
Fault Identification:
Solution: Restart the unit. If the problem persists check for hardware defective components.

KODAK 9000 and KODAK 9000 3D Extraoral Imaging System_Service Guide (SM712)_Ed 02 7–71
E312: Demo Mode Error

E312: Demo Mode Error

Definition of the Cause


The demo files have not been found (pano, 3D, Ceph).

Solution of the Cause

Possible Cause: The demo files are missing.


Remarks:
Fault Identification:
Solution: To find the solution, follow these steps:
1. Check the demo files in Driver TrophyPan directory.

2. Re-install demo mode.

7–72 ERROR MESSAGES


E315: Unable to Initialize Ethernet Communication

E315: Unable to Initialize Ethernet


Communication

Definition of the Cause


The Ethernet initialization failed.

Solution of the Cause

Possible Cause: • The Ethernet interface configuration is not correctly executed in the PC.
• The firewall and/or anti-virus is blocking the transmission.
Remarks:
Fault Identification:
Solution: • Check the configuration of the Ethernet interface in the PC.
• Check the Firewall and/or anti-virus. Authorize TCP communication on port
502 and 33001 and UDP communication on ports 67, 69 and 1001.

KODAK 9000 and KODAK 9000 3D Extraoral Imaging System_Service Guide (SM712)_Ed 02 7–73
E316: Unable to Connect to the Radiological Unit

E316: Unable to Connect to the Radiological


Unit

Definition of the Cause


There is no connection between the PC and the unit.

Solution of the Cause

Possible Cause: The physical Ethernet connection failed.


Remarks:
Fault Identification:
Solution: • Make sure that the unit is ON.
• Check the Ethernet physical link between the PC and the unit.

7–74 ERROR MESSAGES


E321: Unable to Locate the Radiological Unit

E321: Unable to Locate the Radiological Unit

Definition of the Cause


The PC cannot locate the unit.

Solution of the Cause

Possible Cause: • There is no Ethernet connection between the PC and the unit.
• The unit IP address is either missing or not allocated.
Remarks:
Fault Identification:
Solution: 1. Check the Ethernet physical link between the PC and the unit.

2. Check to see if the DHCP/TFTP server is running. If not, restart the PC.

3. Restart the unit.

KODAK 9000 and KODAK 9000 3D Extraoral Imaging System_Service Guide (SM712)_Ed 02 7–75
E400: Unable to Allocate Memory

E400: Unable to Allocate Memory

Definition of the Cause


There is a memory allocation problem.

Solution of the Cause

Possible Cause: The PC driver is unable to allocate enough memory to acquire an image.
Remarks:
Fault Identification:
Solution: • Close unused running applications on the PC. If the problem is not resolved,
reboot the PC.
• Check to see if the PC is compatible with the PC system requirements.

7–76 ERROR MESSAGES


E401: Unable to Connect to the Radiological Unit

E401: Unable to Connect to the Radiological


Unit

Definition of the Cause


The PC is unable to connect to the unit.

Solution of the Cause

Possible Cause: • The PC is unable to connect to Modbus TCP.


• The firewall/antivirus is blocking the transmission.
Remarks:
Fault Identification:
Solution: • Make sure that the unit is ON.
• Check the Ethernet physical link between the PC and the unit.
• Check the Firewall and/or anti-virus. Authorize TCP communication on port
502 and 33001 and UDP communication on ports 67, 69 and 1001.

KODAK 9000 and KODAK 9000 3D Extraoral Imaging System_Service Guide (SM712)_Ed 02 7–77
E402: Communication Interruption

E402: Communication Interruption

Definition of the Cause


There is a communication interruption between the PC and the unit.

Solution of the Cause

Possible Cause: The Modbus TCP connection with the PC has been interrupted.
Remarks:
Fault Identification:
Solution: Check the Ethernet physical link between the PC and the unit.

7–78 ERROR MESSAGES


E405: Unknown Error Occurred During Acquisition

E405: Unknown Error Occurred During


Acquisition

Definition of the Cause


An unknown error occurred during the acquisition.

Solution of the Cause

Possible Cause: The possible cause is unknown and must be researched by Trophy.
Remarks:
Fault Identification:
Solution: Contact and send the log files to Trophy.

KODAK 9000 and KODAK 9000 3D Extraoral Imaging System_Service Guide (SM712)_Ed 02 7–79
E409: Cannot Open Serial Link

E409: Cannot Open Serial Link

Definition of the Cause


Unable to establish the communication between the PC and the rotative arm of the unit.

Solution of the Cause

Possible Cause: • The Ethernet cable between the PC and the unit is disconnected.
• The Ethernet communication is blocked by a firewall and/or anti-virus.
• the serial link between the CJ732 board and the digital sensor board is
disconnected.
Remarks:
Fault Identification:
Solution: 1. Check the Ethernet link between the PC and the unit.

2. Check the following configuration of the Ethernet interface in the PC:

• The authorized TCP communication on ports 502 and 33001.


• The authorized UDP communication on ports 67, 69 and 1001.

3. Check the serial link between the CJ732 board and the digital sensor board.

7–80 ERROR MESSAGES


E410: Acquisition Error Time-out

E410: Acquisition Error Time-out

Definition of the Cause


There is no data received from the unit.

Solution of the Cause

Possible Cause: There is an acquisition time-out.


Remarks:
Fault Identification: To identify the fault, follow these steps:
1. Restart the unit.

2. Reboot the PC.

3. Check the connections between the microprocessor board CJ732 and the digital
sensor.

4. If needed replace the digital sensor.


Solution: Replace the defective digital sensor.

KODAK 9000 and KODAK 9000 3D Extraoral Imaging System_Service Guide (SM712)_Ed 02 7–81
E414: Mask Files not Found

E414: Mask Files not Found

Definition of the Cause


The mask files cannot be found when accessing the Acquisition Windows for panoramic or
3D.

Solution of the Cause

Possible Cause: The missing mask files are invalid.


Remarks:
Fault Identification:
Solution: Install or reinstall the mask files.

7–82 ERROR MESSAGES


E416: Internal Fatal Error

E416: Internal Fatal Error

Definition of the Cause


An abnormal termination of a program caused by an internal fatal error.

Solution of the Cause

Possible Cause: The possible cause is unknown and must be researched by Trophy.
Remarks:
Fault Identification:
Solution: 1. Reboot the PC.

2. Restart the unit.

3. If problem persists, contact Trophy.

KODAK 9000 and KODAK 9000 3D Extraoral Imaging System_Service Guide (SM712)_Ed 02 7–83
E417: The Radiological Unit Encountered a Fatal Error

E417: The Radiological Unit Encountered a Fatal


Error

Definition of the Cause


The unit is unable to acquire an x-ray image.

Solution of the Cause

Possible Cause: The sensor firmware of the radiological unit has met a problem.
Remarks:
Fault Identification:
Solution: • Reboot the PC.
• Restart the unit.

7–84 ERROR MESSAGES


E419: Communication with the Radiological Unit Failed

E419: Communication with the Radiological Unit


Failed

Definition of the Cause


There is no ethernet communication between the PC and the sensor of the unit.

Solution of the Cause

Possible Cause: The Ethernet cable is disconnected.


Remarks:
Fault Identification:
Solution: Check the Ethernet physical link between the PC and the sensor of the unit.

KODAK 9000 and KODAK 9000 3D Extraoral Imaging System_Service Guide (SM712)_Ed 02 7–85
E420: Unable to Connect to the Radiological Unit

E420: Unable to Connect to the Radiological


Unit

Definition of the Cause


Unable to establish an ethernet communication between the PC and the sensor of the unit.

Solution of the Cause

Possible Cause: The Ethernet connection failed.


Remarks:
Fault Identification:
Solution: Check the Ethernet physical link between the PC and the sensor of the unit.

7–86 ERROR MESSAGES


E422: Error During Acquisition Process

E422: Error During Acquisition Process

Definition of the Cause


An error occurred during the acquisition, therefore, no image data has been received.

Solution of the Cause

Possible Cause: The firewall or anti-virus blocks the ethernet communication (UDP channel).
Remarks:
Fault Identification: Acquisition has started but no data is transmitted.
Solution: 1. Check the ethernet interface configuration in the PC.

2. Check the following Firewall and/or anti-virus configuration in the PC:

• The authorized TCP communication on ports 502 and 33001.


• The authorized UDP communication on ports 67, 69 and 1001.

KODAK 9000 and KODAK 9000 3D Extraoral Imaging System_Service Guide (SM712)_Ed 02 7–87
E422: Error During Acquisition Process

7–88 ERROR MESSAGES


Chapter 8
THE CJ653 BOARD TEST POINTS

WARNING
You MUST switch off the unit before removing the covers. ONLY an
approved Trophy technician is qualified to inspect or maintain the unit
while it is switched on and while the covers are removed. In this case, NO
unqualified person must approach the unit.

Description
The CJ653 inverter board has the following functions:

• Supplies power to the HT transformer of the x-ray tube.


• Supplies power to the relay frequency control and the input and output to and from the
CJ732 board.

CJ653 Board Test Points (TP)

The CJ653 Test Points (TP) enables you to safely test the relevant electric circuits. Figure 6-1
illustrates the location of the PTs.

Table 6-1 lists the PTs and their associated electric circuits.

KODAK 9000 and KODAK 9000 3D Extraoral Imaging System_Service Guide (SM712)_Ed 02 8–1
CJ653 Board Test Points (TP)

Figure 8–1 CJ653 Inverter Board Test Points (Unit Serial N°)

TP1

TP4

TP5

TP2
TP8
TP3

TP6

TP9
TP7

TP10

TP12 TP11

8–2 THE CJ653 BOARD TEST POINTS


CJ653 Board Test Points (TP)

Figure 8–2 CJ653 Inverter Board Test Points (Unit Serial N°)

TP1

TP4

TP5

TP2
TP8
TP3

TP6

TP9
TP7

TP10

TP12 TP11

KODAK 9000 and KODAK 9000 3D Extraoral Imaging System_Service Guide (SM712)_Ed 02 8–3
CJ653 Board Test Points (TP)

Table 8–1 CJ653 Inverter Board Test Points

TP n° Ref. Value Description


TP2 0VRL 0V from the: It is connected to the ground.
• 12VRL
• 12VSW
TP3 CD RE2 0V or 15VDC • Active during x-ray emission.
• Controls K2 (x-ray second step).
TP4 CD RE1 0V or 15VDC • Active during x-ray emission.
• Controls K2 (x-ray first step).
TP5 E CAPA 0V or 15VDC 0V in the event of E11
TP6 +12VRL 15VDC ±10% • Supplies power to the relays and the
input/output to/from the CJ732 board.
• Inactive in the event of an error.

TP7 CD LIGHT 0V or 15VDC Active when the x-ray warning lamp is in ready
state for x-ray emission.
TP8 12VSW 9 to 15VDC
TP9 OUT B • Square-wave signal • MOSFET command signal.
• Complementary to OUTA • The frequency increases when the power
• 0V or 12V, from 30 kHZ to demand decreases.
70 kHZ • Active during x-ray emission.

TP10 OUT A • Square-wave signal • MOSFET command signal.


• Complementary to OUT B • The frequency increases when the power
• 0V or 12V, from 30 kHZ to demand decreases.
70 kHZ • Active during x-ray emission.

TP11 R LIGHT Contact used to connect an x-ray 250 V max. / 1A max.


safety lamp
TP12 E LIGHT Contact used to connect an x-ray 250 V max. / 1A max.
safety lamp

8–4 THE CJ653 BOARD TEST POINTS


Chapter 9
THE CJ732 BOARD TEST POINTS AND LED

WARNING
You MUST switch off the unit before removing the covers. ONLY an
approved Trophy technician is qualified to inspect or maintain the unit
while it is switched on and while the covers are removed. In this case, NO
unqualified person must approach the unit.

Description
The CJ732 microprocessor board has the following functions:

• The master controller of the unit.


• Regulates the kV.
• Regulates the mA.
• Controls the 3 step motors.
• Controls the 4 laser beams.
• Manages the x-ray sequence process.
• Manages the image acquisition process.

CJ732 Board Test Points (PT)


The CJ732 board Test Points (PT) enables you to safely test the relevant circuits. Figure 7-1
illustrates the location of the PTs.

Table 7-1 lists the PTs and their associated electric circuit.

KODAK 9000 and KODAK 9000 3D Extraoral Imaging System_Service Guide (SM712)_Ed 02 9–1
CJ732 Board Test Points (PT)

Figure 9–1 Microprocessor Board CJ732 Test Points (PT) and LEDs (LD)

PT12 PT24 PT22 PT13 PT16

PT9

PT14 PT10

PT11

PT19 PT20

PT46 PT21
PT15
LD23
PT18
PT17 LD15

LD16

LD1
1

LD14

LD20
1

LD17
PT28
LD18 LD26

LD22

LD30
LD29
PT42

LD7

LD31
LD6

LD19

LD34

LD3 LD13 LD5 LD4 LD11 LD10

9–2 THE CJ732 BOARD TEST POINTS AND LED


CJ732 Board Test Points (PT)

Table 9–1 CJ732 Board Test Points (PT)

PT n° Ref. Value Description


PT9 +12 VSW 15 VDC
PT10 +12 VRL 15 VDC • Supplies power to the relays and the
inputs/outputs on the CJ732 board.
• Inactive in the event of an error.

PT11 0VRL • 0 V from the 12 VRL Connected to the ground.


• 0 V from the 12 VSW

PT12 +12 VA 11.5 to 12.5 VDC • Supplies power to the analog section of the
CJ732 board.
• Supplies power to the control of the
MOSFET’s of the CJ653 board.
• Generates the +5V DC

PT13 -12 VA -11.5 to 12.5 VDC Supplies power to the analog section of the CJ732
board
PT14 +6 V 6.45 to 6.55VDC Supplies power to the filament heating circuit on
the CJ732 board.
PT15 GND • 0 V from the +12 V Connected to the ground.
• 0 V from the -12 V
• 0 V from the +6 V

PT16 +5 V From 4.7 V to 5.3 VDC Supplies power to the digital section of the CJ732
board.
PT17 MA-REF 4 mA / V • Set value mA.
• Active during x-ray emission.
PT18 KV_REF 20 kV / V • Set value kV.
• Active during x-ray emission.
PT19 kV 20 kV / V • kV feedback voltage.
• Active during x-ray emission.
PT20 OUTA • A square-wave signal. • MOSFET command signal.
• Complementary to • The frequency increases when the power
OUTB. demand decreases.
• 0 or 12 V from 30kHz to • Active during x-ray emission.
70kHz
PT21 OUTB • A square-wave signal. • MOSFET command signal.
• Complementary to • The frequency increases when the power
OUTA. demand decreases.
• 0 or 12 V from 30kHz to • Active during x-ray emission.
70kHz
PT22 RET_MA 4 mA / V • mA feedback voltage.
• Active during x-ray emission.
PT24 V_HEAT 4 to 6 VDC Supplies power to the filament.
PT28 CLK-TDI • A square-wave signal. • The digital sensor acquisition clock signal.
• 0 or 12 V from 100kHz to • Active during x-ray emission.
300kHz.

KODAK 9000 and KODAK 9000 3D Extraoral Imaging System_Service Guide (SM712)_Ed 02 9–3
CJ732 Board LEDs (LD)

Table 9–1 CJ732 Board Test Points (PT)

PT n° Ref. Value Description


PT42 MES-mAs • A square-wave signal. • The mAs measure.
• 0 to 5 V from 0kHz to • Active during x-ray emission.
500kHz.

PT46 VREF1 4.6 V

CJ732 Board LEDs (LD)


The CJ732 board LEDs (LD) enables you to check the correct function of the indicated
components. Figure 7-1 illustrates the location of the LDs.

Table 7-2 lists the LDs and their associated component.

Table 9–2 CJ732 Board LED (LD)

LD n° Ref. Value Description


LD1 ERR_CAPA Lit when error message E 11 appears. This LED is red.
LD3 X_SW Lit when the exposure button of remote control is This LED is green.
active.
LD4 FDC_X_D Lit when the X right limit switch is active. This LED is green.
LD5 FDC_X_G Lit when the X left limit switch is active. This LED is green.
LD6 FDC_END Lit when the limit switch indicates the end of the This LED is green.
rotation.
LD7 FDC_R Lit when the complete rotation reference limit switch This LED is green.
is active.
LD10 FDC_Y_AV Lit when the front Y limit switch is active. This LED is green.
LD11 FDC_Y_AR Lit when the back Y limit switch is active. This LED is green.
LD13 ERR_TEMP • Lit during normal operation with thermal switch This LED is green.
closed.
• Not lit when the temperature of the generator is
greater than 65° C.
LD14 LIGHT • Lit when external safety lamp relay is activated. This LED is green.
• Lit during the x-ray emission.
LD15 RE2 • Lit when the relays K1 and K3 on the CJ653 This LED is green.
board are energized.
• Lit during the x-ray emission.

LD16 RE1 • Lit when the relay K2 on the CJ653 is energized. This LED is green.
• Lit during preheating.
• Lit during the x-ray emission.

9–4 THE CJ732 BOARD TEST POINTS AND LED


CJ732 Board LEDs (LD)

Table 9–2 CJ732 Board LED (LD)

LD n° Ref. Value Description


LD17 LASER_3D Lit when the 3D lasers are switched ON. This LED is green.
LD18 EN_R Lit when the rotative arm motor is active. This LED is orange.
LD19 SW_L Not lit when chin rest is at its minimum limit switch. This LED is green.
LD20 LASER_PANO Lit when the panoramic lasers are switched ON. This LED is green.
LD22 POS_STAT_A Blinking during arm rotation. This LED is green.
LD23 CDE_12V Lit when the 12 VRL relay is active This LED is green.
LD26 POS_STAT_B Blinking during arm rotation. This LED is green.
LD29 EN_Y Lit when the Y motor is active. This LED is orange.
LD30 LASER_CEPH Lit when the lasers are switched on. This LED is green.
LD31 EN_X Lit when the X motor is active. This LED is orange.
LD34 SW_H Not lit when chin rest is at its maximum limit switch. This LED is green.

KODAK 9000 and KODAK 9000 3D Extraoral Imaging System_Service Guide (SM712)_Ed 02 9–5
CJ732 Board LEDs (LD)

9–6 THE CJ732 BOARD TEST POINTS AND LED


Chapter 10
REPLACING THE COLUMN LINEAR ACTUATOR

WARNING
You MUST switch off the unit before removing the covers. ONLY an
approved Trophy technician is qualified to inspect or maintain the unit
while it is switched on and while the covers are removed. In this case, NO
unqualified person must approach the unit.

Description
The column linear actuator enables the up and down movement of the column mobile
component and the unit head.

x4

x4

KODAK 9000 and KODAK 9000 3D Extraoral Imaging System_Service Guide (SM712)_Ed 02 10–1
Removing the Column Linear Actuator

Removing the Column Linear Actuator


Before removing the column linear actuator, make sure that:

• There is enough space between the wall and the column. If there is not enough space,
then you must uninstall the unit head and the fixed arm (see the Kodak 9000 3D
Installation Guide (SM711)).
• The unit is OFF.
• You have the required tool.

To remove the column linear actuator, follow these steps:

1. Unscrew the fixing screws at the base of the column.

2. Loosen the 2 screws of the mobile section of the wall mounting bracket to be able to turn
the column and access the limit switches.

WARNING
DO NOT REMOVE the 2 screws of the mobile section of the wall
mounting brackets, because the unit will fall.

3. Measure the height of the (A) to (B). Prepare a wood block with the following dimensions:
5cm (L) x 5cm (D) x (H (A & B) + 2cm).

4. Remove the clips using a screw driver, then remove the pin (C).

5. Lift the mobile component with the help of 3 persons and place the wood stick under it.
Attach securely the wood stick and the column together.

6. Unscrew the fixing screws of the linear actuator retaining bracket.

7. Remove the column connecting terminals cover. Remove the clips using a screw driver,
then remove the pin (D). Disconnect the connectors.

8. Remove the faulty column linear actuator.

Replacing the Column Linear Actuator


Before replacing the column linear actuator, make sure that you have the appropriate
component.

To replace the column linear actuator, follow these steps:

1. Connect the connectors. Insert the pin (D) and fix with the clips.

2. Position correctly the linear actuator and screw the fixing screws of the retaining bracket.

3. Manually turn the actuator piston to match the height of the unit.

10–2 REPLACING THE COLUMN LINEAR ACTUATOR


Testing the Column Linear Actuator

4. Detach the wood stick from the column. Lift the unit with the help of 3 persons, remove
the wood stick and gently lower the unit in the fork of the linear actuator.

5. Insert the pin (C) and fix with the clips.

6. Position correctly the column using a spirit level. Screw the fixing screws at the base of
the column.

7. Tighten the screw of the mobile section of the wall mounting bracket.

Testing the Column Linear Actuator

From the fixed arm, on the control panel, press and hold to move up and down to the
maximum position to check that the column linear actuator is replaced correctly.

KODAK 9000 and KODAK 9000 3D Extraoral Imaging System_Service Guide (SM712)_Ed 02 10–3
Testing the Column Linear Actuator

10–4 REPLACING THE COLUMN LINEAR ACTUATOR


Chapter 11
REPLACING THE COLUMN CABLES

WARNING
You MUST switch off the unit before removing the covers. ONLY an
approved Trophy technician is qualified to inspect or maintain the unit
while it is switched on and while the covers are removed. In this case, NO
unqualified person must approach the unit.

Description
The column cables connect the column to the unit head and the x-ray remote control.

2.5 m m
3 mm

6 mm

x4

KODAK 9000 and KODAK 9000 3D Extraoral Imaging System_Service Guide (SM712)_Ed 02 11–1
Removing the Column Cables

Removing the Column Cables


Before removing the column cables, make sure that:

• The unit is OFF.


• You have uninstalled the unit head and the fixed arm (see the Kodak 9000 3D
Installation Guide (SM711)).
• You have the required tool.

To remove the column cables, follow these steps:

1. Remove the column reinforcement jack (see the “Replacing the Column Reinforcement
Jack” chapter).

2. Remove the column linear actuator (see the “Replacing the Column Linear Acutator”
chapter).

3. Remove the ON/OFF switch (see the “Replacing the column Limit Switch” chapter).

4. Facing the column, pull the black cables on the left side and the cables on the right side
through the hole.

5. Unscrew the fixing screws of the cable retaining bracket (A).

6. Disconnect all the cables at the base of the column.

7. Pull all the cables with the cable retaining bracket (A) from the base of the column to
remove the faulty cables.

Replacing the Column Cables


Before replacing the column cables, make sure that you have the appropriate component.

To replace the column cables, follow these steps:

1. Measure the length of the old cables from the cable bracket (A) and position the new
cables in the same length in the cable bracket. Attach all the cables in the cable bracket
(A).

2. Insert the cables from the hole on top of the column and let the cables fall inside the
column.

3. Facing the column, pull and position correctly the black cable on the left side and the
cables on the right side.

4. Screw the fixing screws of the cable bracket (A).

5. Connect all the cables at the base of the column.

6. Replace the ON/OFF switch (see the “Replacing the column Limit Switch” chapter).

11–2 REPLACING THE COLUMN CABLES


Testing the Column Cables

7. Replace the column linear actuator (see the “Replacing the Column Linear Acutator”
chapter).

8. Replace the column reinforcement jack (see the “Replacing the Column Reinforcement
Jack” chapter).

9. Reinstall the column, the unit head and the fixed arm (see the Kodak 9000 3D
Installation Guide (SM711)).

Testing the Column Cables


To check that the column cables are correctly replaced, do the following:

• Check the ON/OFF button: Turn on the unit.


• Check the column linear actuator: From the fixed arm, on the control panel, press and

hold to move the unit head, up and down to the maximum position.
• Check the security green lamp: See if the unit is in the ready state to launch an
acquisition from the PC.
• Check the Ethernet cable and the remote control cable: Launch an acquisition from the
PC.

KODAK 9000 and KODAK 9000 3D Extraoral Imaging System_Service Guide (SM712)_Ed 02 11–3
Testing the Column Cables

11–4 REPLACING THE COLUMN CABLES


Chapter 12
REPLACING THE COLUMN REINFORCEMENT
JACKS

WARNING
You MUST switch off the unit before removing the covers. ONLY an
approved Trophy technician is qualified to inspect or maintain the unit
while it is switched on and while the covers are removed. In this case, NO
unqualified person must approach the unit.

Description
The column reinforcement jacks support the extra weight of the Cephalometric component of
the Kodak 9000C and Kodak 9000C 3D units.

D
A

x2

B
2 mm
2.5 mm

x2

x1

x4

KODAK 9000 and KODAK 9000 3D Extraoral Imaging System_Service Guide (SM712)_Ed 02 12–1
Removing the Column Reinforcement Jacks

Removing the Column Reinforcement Jacks


Before removing the column reinforcement jack, make sure that:

• There is enough space between the wall and the column. If there is not enough space,
then you must uninstall the unit head and the fixed arm (see the Kodak 9000 3D
Installation Guide (SM711)).
• The unit is OFF.
• You have the required tool.

To remove the column reinforcement jack, follow these steps:

1. Unscrew the fixing screws at the base of the column.

2. Loosen the 2 screws of the mobile section of the wall mounting bracket to be able to turn
the column and access the limit switches.

WARNING
DO NOT REMOVE the 2 screws of the mobile section of the wall
mounting brackets, because the unit will fall.

3. Unscrew and remove the column connecting terminals cover.

WARNING
You MUST raise the unit to the maximum height before starting
the procedures.

4. Unscrew the fixing screws of the reinforcement jack fitting (A).

5. Remove the clips at the base of the reinforcement jack using a screw driver. Remove the
pin.

6. Unscrew the fixing screws of the reinforcement jack assembly protection pipe (B) and
remove.

7. Unscrew the fixing screw of the attaching pipe (C). Remove the attaching pipe (C), the
clips and the pin of the reinforcement jack fitting (A).

8. Keep the reinforcement jack fitting (A), the protection pipe (B) and the attaching pipe (C)
for the new reinforcement jack.

9. Remove the faulty reinforcement jack.

12–2 REPLACING THE COLUMN REINFORCEMENT JACKS


Replacing the Column Reinforcement Jacks

Replacing the Column Reinforcement Jacks


Before replacing the column reinforcement jack, make sure that you have the appropriate
component.

To replace the column reinforcement jack, follow these steps:

1. Insert the fitting (A) in the reinforcement jack. Replace the pin and fixing clips.

2. Insert the attaching pipe (C) to attach the reinforcement jack assembly and screw the
fixing screw.

3. Holding the protection pipe (B), insert the fitting (A) at the top behind the sliding frame
(D) of the column.

4. Position the protection pipe (B) correctly against the column and screw the fixing screws.

5. Insert the reinforcement jack assembly at the base. Insert the pin and the fixing clips.

6. Screw the fixing screws of the reinforcement jack fitting (A).

Testing the Column Reinforcement Jacks


Test only with the Cephalometric component, the up and down movement of the unit.

KODAK 9000 and KODAK 9000 3D Extraoral Imaging System_Service Guide (SM712)_Ed 02 12–3
Testing the Column Reinforcement Jacks

12–4 REPLACING THE COLUMN REINFORCEMENT JACKS


Chapter 13
REPLACING THE COLUMN LIMIT SWITCHES

WARNING
You MUST switch off the unit before removing the covers. ONLY an
approved Trophy technician is qualified to inspect or maintain the unit
while it is switched on and while the covers are removed. In this case, NO
unqualified person must approach the unit.

Description
The column limit switches limit the up and down movement of the column mobile component
and the unit head.

2.5 m m

x2

x2

Removing the Column Limit Switches


Before removing the column limit switches, make sure that:

• There is enough space between the wall and the column. If there is not enough space,
then you must uninstall the unit head and the fixed arm (see the Kodak 9000 3D
Installation Guide (SM711)).
• The unit is OFF.
• You have the required tool.

KODAK 9000 and KODAK 9000 3D Extraoral Imaging System_Service Guide (SM712)_Ed 02 13–1
Replacing the Column Limit Switches

To remove the column limit switches, follow these steps:

1. Unscrew the fixing screws at the base of the column.

2. Loosen the 2 screws of the mobile section of the wall mounting brackets to be able to turn
the column and access the limit switches.

WARNING
DO NOT REMOVE the 2 screws of the mobile section of the wall
mounting brackets, because the column will fall.

3. Unscrew the fixing screws of the limit switch and remove the metal plate.

4. Unsolder and disconnect the cables and remove the faulty limit switches.

CAUTION
Identify the orientation of the switch and the position of the cables
before removal.

Replacing the Column Limit Switches


Before replacing the column limit switches, make sure that you have the appropriate
component.

To replace the column limit switches, follow these steps:

1. Solder the cables to the limit switches. Cover the connectors with insulation sleevings.

CAUTION
DO NOT overheat the lugs.

2. Position the limit switches and the metal plate and screw the fixing screws.

3. Position correctly the column, using a spirit level. Screw the fixing screws at the base of
the column.

4. Tighten the 2 screws of the mobile section of the wall mounting brackets.

Testing the Column Limit Switches

From the fixed arm, on the control panel, press and hold to move up and down to the
maximum position to check that the limit switches are correctly replaced.

13–2 REPLACING THE COLUMN LIMIT SWITCHES


Chapter 14
REPLACING THE ON/OFF BUTTON

WARNING
You MUST switch off the unit before removing the covers. ONLY an
approved Trophy technician is qualified to inspect or maintain the unit
while it is switched on and while the covers are removed. In this case, NO
unqualified person must approach the unit.

Description
The ON/OFF button enables the unit to be powered.

x2

6 mm

2.5 mm

KODAK 9000 and KODAK 9000 3D Extraoral Imaging System_Service Guide (SM712)_Ed 02 14–1
Removing the ON and OFF Button

Removing the ON and OFF Button


Before removing the ON/OFF button, make sure that:

• The general mains to the unit is OFF.


• The column inside cover is removed.
• You have the required tool.

To remove the ON/OFF button, follow these steps:

1. Identify the ON/OFF position of the button in the plate (A).

2. Unclips the ON/OFF button (B) from the plate (A).

3. Disconnect the cables.

4. Remove the faulty ON/OFF button.

Replacing the ON and OFF Button


Before replacing the ON/OFF button, make sure that you have the appropriate component.

To replace the ON/OFF button, follow these steps:

1. Position the ON/OFF button (B) in the correct orientation and clips in the plate (A).

2. Replace the cable 29 on the lugs 2 and 5 in this order:

• N = blue on lug 2
• P = brown on lug 5
• Ground on ground lug

3. Replace the cable 30 on lugs 1 and 4 in this order:

• N = blue on lug 1
• P = brown on lug 4
• Ground on ground lug

4. Insert with correct orientation the ON/OFF button in the column inside cover and screw
the fixing screws.

Testing the ON and OFF Button


Turn ON the general mains. Turn ON the unit. Make sure that the ON/OFF button is lit only
when you push the ON button. If the ON/OFF button is lit when the unit is OFF you must
check the cable connections and redo the cabling correctly.

14–2 REPLACING THE ON/OFF BUTTON


Chapter 15
REPLACING THE LF440 LINE FILTER

WARNING
You MUST switch off the unit before removing the covers. ONLY an
approved Trophy technician is qualified to inspect or maintain the unit
while it is switched on and while the covers are removed. In this case, NO
unqualified person must approach the unit.

Description
The LF440 line filter enables line voltage noise filtering.

3 mm

x4

KODAK 9000 and KODAK 9000 3D Extraoral Imaging System_Service Guide (SM712)_Ed 02 15–1
Removing the LF440 Line Filter

Removing the LF440 Line Filter


Before removing the LF440 line filter, make sure that:

• The unit is powered OFF.


• The head cover is removed.
• You have the required tool.

To remove the LF440 line filter, follow these steps:

1. Disconnect the the brown (P) and the blue (N) wires to the mains input/output terminals.

2. Unscrew the fixing screws.

3. Remove the faulty Line Filter.

Replacing the LF440 Line Filter


Before replacing the LF440 line filter, make sure that you have the appropriate Line Filter.

To replace the LF440 line filter, follow these steps:

1. Position the LF440 line filter with “LINE” facing the Column.

2. Screw the 4 fixing screws

3. Connect the brown (P) and the blue (N) wires to the mains input/output terminals.

CAUTION
DO NOT connect any of the line voltage wires to the ground lug.

Testing the LF440 Line Filter


Turn ON the unit to check that the LF440 line filter is replaced correctly.

15–2 REPLACING THE LF440 LINE FILTER


Chapter 16
REPLACING THE CJ777 FILTER BOARD

WARNING
You MUST switch off the unit before removing the covers. ONLY an
approved Trophy technician is qualified to inspect or maintain the unit
while it is switched on and while the covers are removed. In this case, NO
unqualified person must approach the unit.

Description
The CJ777 filter board is the second line voltage noise filtering component of the unit.

2.5 mm

x4

KODAK 9000 and KODAK 9000 3D Extraoral Imaging System_Service Guide (SM712)_Ed 02 16–1
Removing the CJ777 Filter Board

Removing the CJ777 Filter Board


Before removing the CJ777 filter board, make sure that:

• The unit is powered OFF.


• The head cover is removed.
• You have the required tool.

To remove the CJ777 filter board, follow these steps:

1. Disconnect all the connectors.

2. Remove the fixing screws.

3. Remove the faulty board.

Replacing the CJ777 Filter Board


Before replacing the CJ777 filter board, make sure that you have the appropriate board.

To replace the CJ777 filter board, follow these steps:

1. Position the CJ777 filter board, with “J1” facing towards the column and screw the fixing
screws.

2. Connect all the connectors in their respective location.

Testing the CJ777 Filter Board


Turn ON the unit to check if the CJ777 filter board, is replaced correctly.

16–2 REPLACING THE CJ777 FILTER BOARD


Chapter 17
REPLACING THE LK218 HARMONIC FILTER

WARNING
You MUST switch off the unit before removing the covers. ONLY an
approved Trophy technician is qualified to inspect or maintain the unit
while it is switched on and while the covers are removed. In this case, NO
unqualified person must approach the unit.

Description
The LK218 harmonic filter enables smooth line voltage peaks.

IMPORTANT
The LK218 Harmonic Filter is not used for units operating at
110V.

7 mm

x2

KODAK 9000 and KODAK 9000 3D Extraoral Imaging System_Service Guide (SM712)_Ed 02 17–1
Removing the LK218 Harmonic Filter

Removing the LK218 Harmonic Filter


Before removing the LK218 harmonic filter, make sure that:

• The unit is powered OFF.


• The head cover is removed.
• You have the required tool.

To remove the LK218 harmonic filter, follow these steps:

1. Disconnect the J3 connector from the CJ777 board.

2. Remove the fixing nuts.

Replacing the LK218 Harmonic Filter


Before replacing the LK218 harmonic filter, make sure that you have the appropriate
Harmonic filter.

To replace the LK218 harmonic filter, follow these steps:

1. Position the J3 cable facing the CJ777 board and screw the 2 fixing nuts.

2. Connect the J3 connector to the CJ777 board.

Testing the LK218 Harmonic Filter


Turn ON the unit to check that the LK218 harmonic filter is replaced correctly.

17–2 REPLACING THE LK218 HARMONIC FILTER


Chapter 18
REPLACING THE MICROPROCESSOR POWER
SUPPLY JZ024 BOARD

WARNING
You MUST comply with the standards currently in force while installing
and using the radiology unit. Whenever necessary, you must wear a
lead-lined apron.

You MUST switch off the unit before removing the covers. ONLY an
approved Trophy technician is qualified to inspect or maintain the unit
while it is switched on and while the covers are removed. In this case, NO
unqualified person must approach the unit.

Description
The microprocessor power supply JZ024 board has the following functions:

• Supplies ± 12 VDC to the CJ732 board.


• Supplies + 6.5 VDC to the filament heating circuit.

KODAK 9000 and KODAK 9000 3D Extraoral Imaging System_Service Guide (SM712)_Ed 02 18–1
Removing the Microprocessor Power Supply JZ024 Board

2.5 mm

x4

Removing the Microprocessor Power Supply


JZ024 Board

Before removing the microprocessor power supply JZ024 board, make sure that:

• The unit is powered OFF.


• The head cover is removed.
• You have the required tool.

To remove the microprocessor power supply JZ024 board, follow these steps:

1. Disconnect the connectors.

2. Unscrew the fixing screws.

3. Remove the faulty JZ024 board.

18–2 REPLACING THE MICROPROCESSOR POWER SUPPLY JZ024 BOARD


Replacing the Microprocessor Power Supply JZ024 Board

Replacing the Microprocessor Power Supply


JZ024 Board

Before replacing the microprocessor power supply JZ024 board, make sure that:

• You have the appropriate JZ024 board.


• You have a DC voltmeter.

To replace the microprocessor power supply JZ024 board, follow these steps:

1. Position the JZ024 board respecting the position of the connectors and screw the fixing
screws.

2. Connect all the connectors to the respective locations.

DANGER OF ELECTRIC SHOCK


The power supply adjustment must be made while the unit is
switched ON. Hazardous voltage is present in the power supply
JZ024 board. Unqualified persons MUST NOT approach the unit.

3. If needed, check or adjust the power supply, connect a DC voltmeter between the PT15
(GND) and PT14 (6.5VDC) on the CJ732 board.

4. Using an insulated screwdriver, rotate the potentiometer (A) until you see 6.50 V.

Testing the Microprocessor Power Supply JZ024


Board
To test the power supply JZ024, follow these steps:

1. Access the Panoramic Acquisition Window. In the Program pane, select 75kV and
12mA.

2. Leave the x-ray room and use the remote control to launch an acquisition.

KODAK 9000 and KODAK 9000 3D Extraoral Imaging System_Service Guide (SM712)_Ed 02 18–3
Testing the Microprocessor Power Supply JZ024 Board

18–4 REPLACING THE MICROPROCESSOR POWER SUPPLY JZ024 BOARD


Chapter 19
REPLACING THE DIGITAL SENSOR POWER
SUPPLY JZ030 BOARD

WARNING
You MUST switch off the unit before removing the covers. ONLY an
approved Trophy technician is qualified to inspect or maintain the unit
while it is switched on and while the covers are removed. In this case, NO
unqualified person must approach the unit.

DANGER OF ELECTRIC SHOCK


The power supply adjustment must be made while the unit is
switched ON. Hazardous voltage is present in the power supply
JZ030 board. Unqualified persons MUST NOT approach the unit.

Description
The digital sensor power supply JZ030 board supplies power to the digital sensor.

1 : +15V
2 : +5.2V
3 : +5.2V
4 : GND
5 : GND
6 : -15V

2.5 mm

KODAK 9000 and KODAK 9000 3D Extraoral Imaging System_Service Guide (SM712)_Ed 02 19–1
Removing the Digital Sensor Power Supply JZ030 Board

Removing the Digital Sensor Power Supply


JZ030 Board
Before removing the digital sensor power supply JZ030 board, make sure that:

• The unit is powered OFF.


• The head cover is removed.
• You have the required tool.
• You have a voltmeter.

To remove the digital sensor power supply JZ030 board, follow these steps:

1. Disconnect the connectors.

2. Unscrew the fixing screws.

3. Remove the faulty power supply board.

Replacing the Digital Sensor Power Supply


JZ030 Board
Before replacing the digital sensor power supply JZ030 board, make sure that you have the
appropriate power supply board.

To replace the digital sensor power supply JZ030, follow these steps:

1. Position the power supply board with the cable 21 on the input CN1 facing the column and
screw the fixing screws.

2. Connect all the connectors

Testing the Digital Sensor Power Supply JZ030


Board
Use a voltmeter to check the following voltage:

• +5.2VDC check and adjust, if neccessary


• +15VDC
• -15VDC

19–2 REPLACING THE DIGITAL SENSOR POWER SUPPLY JZ030 BOARD


Chapter 20
REPLACING THE STEP MOTOR POWER SUPPLY
JZ029 BOARD

WARNING
You MUST switch off the unit before removing the covers. ONLY an
approved Trophy technician is qualified to inspect or maintain the unit
while it is switched on and while the covers are removed. In this case, NO
unqualified person must approach the unit.

Description
The step motor power supply JZ029 board, has the following functions:

• Supplies power to the step motor.


• Supplies power to the chin rest DC motor.
• Supplies power to the digital sensor DC motor.
• Supplies power to the collimator DC motor.
• Supplies power to the XY motor assembly.
• Supplies power to the rotative arm motor assembly.
• Supplies power to all the 12 VRL.

1-2-3-4 : +15V
5-6-7-8 : 0V

x4

2.5 mm

KODAK 9000 and KODAK 9000 3D Extraoral Imaging System_Service Guide (SM712)_Ed 02 20–1
Removing the Step Motor Power Supply JZ029 Board

Removing the Step Motor Power Supply JZ029


Board
Before removing the step motor power supply JZ029 board, make sure that:

• The unit is powered OFF.


• The head cover is removed.
• You have the required tools.

To remove the step motor power supply JZ029 board, follow these steps:

1. Disconnect all the connectors.

2. Unscrew the fixing screws.

3. Remove the faulty power supply.

Replacing the Step Motor Power Supply JZ029


Board
Before replacing the step motor power supply JZ029 board, make sure that you have the
appropriate power supply board.

To replace the step motor power supply JZ029 board, follow these steps:

1. Position the cable 20 on the CN1 facing the column.

2. Screw the fixing screws.

Testing the Step Motor Power Supply JZ029


Board
From the fixed arm, on the control panel, press and hold the X, Y, and rotation movement
buttons to test that the the new power supply JZ029 board is replaced correctly. Check the
value of the output voltage (see the above figure).

20–2 REPLACING THE STEP MOTOR POWER SUPPLY JZ029 BOARD


Chapter 21
REPLACING THE CJ699 MOTOR BOARD

WARNING
You MUST switch off the unit before removing the covers. ONLY an
approved Trophy technician is qualified to inspect or maintain the unit
while it is switched on and while the covers are removed. In this case, NO
unqualified person must approach the unit.

Description
The CJ699 motor board enables the up and down movement of the unit head.

2.5 mm

x3

Removing the CJ699 Motor Board


Before removing the CJ699 motor board, make sure that:

• The unit is powered OFF.


• The head cover is removed.
• You have the required tool.

KODAK 9000 and KODAK 9000 3D Extraoral Imaging System_Service Guide (SM712)_Ed 02 21–1
Replacing the CJ699 Motor Board

To remove the CJ699 motor board, follow these steps:

1. Disconnect all the connectors.

2. Unscrew the fixing screws of the board.

3. Remove the faulty board.

Replacing the CJ699 Motor Board


Before replacing the CJ699 motor board, make sure that:

• You have the appropriate board.


• The board operating voltage configuration matches your line voltage.

To replace the CJ699 motor board, follow these steps:

1. Position the board with “J5” facing the line filter and screw the fixing screws.

2. Connect all the connectors.

Testing the CJ699 Motor Board

From the control panel, press and hold to move up and down the unit head to check
that the CJ699 motor board is replaced correctly.

21–2 REPLACING THE CJ699 MOTOR BOARD


Chapter 22
REPLACING THE CJ699 MOTOR BOARD FUSES

WARNING
You MUST switch off the unit before removing the covers. ONLY an
approved Trophy technician is qualified to inspect or maintain the unit
while it is switched on and while the covers are removed. In this case, NO
unqualified person must approach the unit.

Description
• The F1 and F2 fuses protect the power supply of the column electric linear actuator.
• The F3 and F4 are not used.

CJ699
R8
R5
R7 R3 R1
R6
R4 R2

F1 F2

F5

FA CE E LE ME NT X 8

699 V3

If the unit is connected to a single-phase alternating current power supply line with a neutral
point, F2 on the CJ699 board fuse is the neutral link.

KODAK 9000 and KODAK 9000 3D Extraoral Imaging System_Service Guide (SM712)_Ed 02 22–1
Removing the CJ699 Motor Board Fuses

If the unit is connected to a double-phase alternating current power supply line, see the table
below:

Ref. Code Characteristics Remarks


F1 at 110V LW177 630mA - 5 x 20 - T- UL and IEC127 If the device is connected at a 2 phase
power supply, F1 must be replaced by
the same fuse as F2 (see the
accessories bag).
F1 at 230V LW178 315mA - 5 x 20 - T - UL and IEC127 If the device is connected at a 2 phase
power supply, F1 must be replaced by
the same fuse as F2 (see the
accessories bag).

Removing the CJ699 Motor Board Fuses


Before removing the fuse, make sure that:

• The unit is powered OFF.


• The head cover is removed.
• You have a flat screw driver.

To remove the fuse, follow these steps:

1. Unscrew the fuse counterclockwise.

2. Remove the faulty fuse.

Replacing the CJ699 Motor Board Fuses


Before replacing the fuse, make sure that you have the appropriate fuse.

To replace the fuse, follow these steps:

1. Replace the appropriate fuse.

2. Screw the fuse clockwise.

Testing the CJ699 Motor Board Fuses

From the control panel, press and hold to move the unit head up and down to check
that the CJ699 motor board fuse is replaced correctly.

22–2 REPLACING THE CJ699 MOTOR BOARD FUSES


Chapter 23
REPLACING THE CJ653 INVERTER BOARD

WARNING
You MUST switch off the unit before removing the covers. ONLY an
approved Trophy technician is qualified to inspect or maintain the unit
while it is switched on and while the covers are removed. In this case, NO
unqualified person must approach the unit.

Description
The CJ653 inverter board has the following functions:

• Supplies power to the HT transformer of the x-ray tube.


• Supplies the VDC to the MOS-FET power bridge.
• Supplies the mains measured voltage to the microprocessor.
• Drives the MOS-FET through the impulse transformers and respective amplifiers.

2.5 mm

x6

KODAK 9000 and KODAK 9000 3D Extraoral Imaging System_Service Guide (SM712)_Ed 02 23–1
Removing the CJ653 Inverter Board

Removing the CJ653 Inverter Board


Before removing the CJ653 inverter board, make sure that:

• The unit is powered OFF.


• The head cover is removed.
• You have the required tool.

To remove the CJ653 inverter board, follow these steps:

1. Disconnect all the connectors.

2. Unscrew the fixing screws.

3. Remove the faulty Board.

Replacing the CJ653 Inverter Board


Before replacing the CJ653 inverter board, make sure that:

• You have the appropriate board.


• The board operating voltage configuration matches your line voltage.

To replace the CJ653 inverter board, follow these steps:

1. Position the CJ653 board with the Q5 transformer facing the column and screw the fixing
screws.

2. Connect all the connectors.

Testing the CJ653 Inverter Board


Acquire an x-ray image to check that the inverter board CJ653 is replaced correctly and
functions properly.

23–2 REPLACING THE CJ653 INVERTER BOARD


Chapter 24
REPLACING THE CJ653 INVERTER BOARD FUSES

WARNING
You MUST switch off the unit before removing the covers. ONLY an
approved Trophy technician is qualified to inspect or maintain the unit
while it is switched on and while the covers are removed. In this case, NO
unqualified person must approach the unit.

Description
• The F1 and F2 fuses protect the CJ653 Inverter Board power supply.
• The F3 and F4 fuses protect the Q5 transformer 220/12V power supply.
• The F5 fuse protects the +12VSW.
• The F6 fuse protects the external warning lamp indicating the x-ray emission.

F1

F2

F4

F6

F3 F5

KODAK 9000 and KODAK 9000 3D Extraoral Imaging System_Service Guide (SM712)_Ed 02 24–1
Removing the CJ653 Inverter board Fuses

If the unit is connected to a single-phase alternating current power supply line with a neutral
point, F1 and F3 on CJ653 board are the neutral links.

If the unit is connected to a double-phase alternating current power supply lines, see the table
below:

CAUTION*
On the CJ653 board, the marking corresponding to fuses F1 and
F2 apply only to devices connected to a 230V power supply. The
20A* fuses must be inserted if the device is connected to a 110V
power supply.

Ref. Code Characteristics Remarks


F2 at 110V LW109 20A - 6.3 x 32 - F- UL* If the unit is connected at a 2 phase
see CAUTION power supply, F1 must be replaced by
the same fuse as F2 (see the
accessories bag).
F2 at 230V LW161 10A - 6.3 x 32 - F- UL If the unit is connected at a 2 phase
power supply, F1 must be replaced by
the same fuse as F2 (see the
accessories bag).
F4 at 110V LW179 125mA - 5 x 20 - T - UL and IEC127 If the unit is connected at a 2 phase
power supply, F3 must be replaced by
the same fuse as F4 (see the
accessories bag).
F4 at 230V LW180 63mA - 5 x 20 - T - UL and IEC127 If the unit is connected at a 2 phase
power supply, F3 must be replaced by
the same fuse as F4 (see the
accessories bag).
F5 LW156 1A - 5 x 20 - T - UL and IEC127
F6 LW156 1A - 5 x 20 - T - UL and IEC127

Removing the CJ653 Inverter board Fuses


Before removing the fuse, make sure that:

• The unit is powered OFF.


• The head cover is removed.
• You have a flat screw driver.

To remove the fuse, follow these steps:

1. Unscrew the fuse counterclockwise.

2. Remove the faulty fuse.

24–2 REPLACING THE CJ653 INVERTER BOARD FUSES


Replacing the CJ653 Inverter board Fuses

Replacing the CJ653 Inverter board Fuses


Before replacing the fuse, make sure that you have the appropriate fuse.

To replace the fuse, follow these steps:

1. Position the appropriate fuse.

2. Screw the fuse clockwise.

Testing the CJ653 Inverter board Fuses


Acquire an x-ray image to check that the CJ653 inverter board fuse is replaced correctly and
functions properly.

KODAK 9000 and KODAK 9000 3D Extraoral Imaging System_Service Guide (SM712)_Ed 02 24–3
Testing the CJ653 Inverter board Fuses

24–4 REPLACING THE CJ653 INVERTER BOARD FUSES


Chapter 25
REPLACING THE CJ732 MICROPROCESSOR
BOARD

WARNING
You MUST switch off the unit before removing the covers. ONLY an
approved Trophy technician is qualified to inspect or maintain the unit
while it is switched on and while the covers are removed. In this case, NO
unqualified person must approach the unit.

Description
The CJ732 microprocessor board has the following functions:

• The master controller of the unit


• Regulates the kV
• Regulates the mA
• Controls the 3 step motors
• Controls the 4 laser beams
• Manages the x-ray sequence process
• Manages the image acquisition process

KODAK 9000 and KODAK 9000 3D Extraoral Imaging System_Service Guide (SM712)_Ed 02 25–1
Removing the CJ732 Microprocessor Board

SW2
ON
2.5 mm 8
7
6

U30 5
4
3

C57 2
1
R120
P3
x4 PT46

CJ732
U36

J25 J10 J11 J4 J12


B1 P T37 P T1 S W2
C54 U25 M1
J 25 J 10 J 11
1 1 1
X 25 732 V3
U23 U30
U14
U50 U51 U21 U17 U26 U22

X1 U15 J4
C57
LD18
LD31 LD29 P T47
P T17 R120
U10 P3
ST 1 7 C23
P T39 P T46
C50 C79
D64 P T33

J23
LD25
J 23 E2
U36 P T14 P T2
D63 C24 C30
C95 X3
P T12 P T24
C66 P T32 Y1 U38 U28 Q4
U2
D62

J22
LD24 X U28 P T18
R81 P T31
J 22 R69 C43 DZ3
R74 P T22
D61
P2 P T13
X17 X15 X5
D74

J19
X16
LD33
J 19 X14 S T2 C40 PT 1 9 C65
R48 X13
J 12
D60 X18 P T15
U11 X12
P T41 D7 U35
R143 P T26
LD8 U27

J16
D47 C6 U19
J 16 LD9 X U27
P T27 B UZ1
D48
R141
LD3 Q3
C93
R151 LD21

J8 J8 D10

D52
LD13

LD5 P T45
X19

X20

U48 U46
LD30
J7

U37
P T23
P T43

P T40

U7
P T20
L3 L2 C9
P T5 1
C31
D15
U40 P4 P T30 P T25 E1

J9 C33 J3
S W1 C92
J9 LD4 J 15 C35 R52 P1
Q5
C32 D5 C8 U3
D13 P T34
Q1
C22 Q9
L1
R144 J 29 R19

J15
LD11 DZ1 C10
LD17 LD20 LD14 LD1 LD16 LD15
D50 J 28
J 17 LD10 E 10 E3
U52 U29 U8 U31 U44 U5 U4 K1 J 20
D49 U1
R142 Q6 C2 C80
Q8 Q2 DZ2 C82
M2 1

J17 Q7 P T38
J2

J 24 J 26 J 27 J 13 J6 J 18 J 14 J 21 J5 J1 X 26 X6
P T8 1
X4 PT1 1 PT1 0 PT9
P T7

J24
J26 J13 J18 J14 J29 J1 J2 J3 J20
J27 J6 J28 J21 J5

Removing the CJ732 Microprocessor Board


Before removing the CJ732 microprocessor board, make sure that:

• The unit is powered OFF.


• The head cover is removed
• You have the required tool.

To remove the CJ732 microprocessor board, follow these steps:

1. Disconnect all the connectors.

2. Unscrew the fixing screws of the board.

3. Remove the faulty board.

25–2 REPLACING THE CJ732 MICROPROCESSOR BOARD


Replacing the CJ732 Microprocessor Board

Replacing the CJ732 Microprocessor Board


Before replacing the CJ732 microprocessor board, make sure that:

• You have the appropriate board.


• The 2 EPROMs are mounted on the board.

To replace the CJ732 microprocessor board, follow these steps:

1. Position the DIP switch (SW2) in the same position as the replaced board.

Table 25–1 DIP Switch (SW2) Positions and Descriptions

DIP Switch DIP Switch


Function
Position State
S1 Emission of x-ray during the exposure OFF
No emission of x-ray during the exposure (Demo mode) ON
S2 Communication speed between the CJ732 and the fixed arm control
panel
9600 baud OFF
38400 baud ON
19200 baud OFF
57600 baud ON
S3 Communication speed between the CJ732 and the fixed arm control
panel
9600 baud OFF
38400 baud OFF
19200 baud ON
57600 baud ON
S4 NA
S5 NA
S6 NA
S7 NA
S8 NA

2. Position the CJ732 board with the connections J25, J10 and J11 facing the column and
screw the 6 fixing screws.

3. Connect all the connectors in their respective locations.

KODAK 9000 and KODAK 9000 3D Extraoral Imaging System_Service Guide (SM712)_Ed 02 25–3
Testing the CJ732 Microprocessor Board

To restore the memorized system data parameters, follow these steps:

4. To switch on the unit in Technical Mode, press and hold at the same time and

and turn ON the unit.

5. Access the Panoramic Acquisition Window. If an error message displays, click Retry
for each error message until the screen is clear.

6. In the Panoramic Acquisition Window, click and select Maintenance >


Save/Restore. Click Restore to transmit the technical data configuration to the new
CJ732 board.

Testing the CJ732 Microprocessor Board


Launch an x-ray acquisition using the test tools to see if the acquisition process is executed
correctly.

25–4 REPLACING THE CJ732 MICROPROCESSOR BOARD


Chapter 26
REPLACING THE MICROPROCESSOR BOARD
CJ732 BATTERY

WARNING
You MUST switch off the unit before removing the covers. ONLY an
approved Trophy technician is qualified to inspect or maintain the unit
while it is switched on and while the covers are removed. In this case, NO
unqualified person must approach the unit.

Description
The CJ732 board battery saves the system parameter configuration settings.

x2 U14
(B1)

U15 CJ732

Removing the Microprocessor Board CJ732


Battery
Before removing the microprocessor board CJ732 battery, make sure that:

• The unit is OFF.


• The unit head cover is removed.

KODAK 9000 and KODAK 9000 3D Extraoral Imaging System_Service Guide (SM712)_Ed 02 26–1
Replacing the Microprocessor Board CJ732 Battery

• You have the required tool.

To remove the microprocessor board CJ732 battery, follow these steps:

1. Insert the tip of a screw driver gently under each side of the battery to raise it.

WARNING
You MUST remove the battery from both sides without breaking
the integrated circuits.

2. Remove the faulty battery.

Replacing the Microprocessor Board CJ732


Battery
Before replacing the microprocessor board CJ732 battery, make sure that you have the
appropriate battery.

To replace the microprocessor board CJ732 battery, follow these steps:

1. Plug in:

• The U14 battery above the integrated circuit build-in battery socket.
• The U15 battery in the integrated circuit socket.

2. Restore the parameter settings (see the “Save/Restore” submenu in the Service Guide).

Testing the Microprocessor Board CJ732


Battery
To test the microprocessor board CJ732 battery, follow these steps:

1. Turn ON the unit. Position the 3D tooling base on the chin rest base. Position the phantom
tooling for the panoramic.

2. From the fixed arm, on the control panel, click to turn ON the mid-sagittal
positioning laser beam. Make sure that the mid-sagittal positioning laser is correctly
centered on the phantom tooling.

3. Access the Panoramic Acquisition Window. Acquire an image to check that the CJ732
battery is correctly replaced.

26–2 REPLACING THE MICROPROCESSOR BOARD CJ732 BATTERY


Chapter 27
REPLACING THE MICROPROCESSOR BOARD
EEPROMS

WARNING
You MUST switch off the unit before removing the covers. ONLY an
approved Trophy technician is qualified to inspect or maintain the unit
while it is switched on and while the covers are removed. In this case, NO
unqualified person must approach the unit.

Description
The CJ732 board EEPROMS saves the system parameter configuration settings.

CJ732
x2

LOW HIGH

U28 U27
1

Removing the Microprocessor Board CJ732


EEPROMS
Before removing the microprocessor board CJ732 EEPROMs, make sure that:

• The unit is OFF.


• The unit head cover is removed.
• You have the required tool.

KODAK 9000 and KODAK 9000 3D Extraoral Imaging System_Service Guide (SM712)_Ed 02 27–1
Removing the Microprocessor Board CJ732 EEPROMS

To remove the microprocessor board CJ732 EEPROMs, follow these steps:

1. You must save the system and calibration parameters before removing the EEPROMs.

From the fixed arm, on the control panel, press and hold , at the same time
and turn ON the unit.

2. Access the Panoramic Window. If an error message is displayed, click Retry button until
the screen is clear.

3. In the Panoramic Window, click and select Maintenance > Save / Restore. The
Save / Restore dialog box appears.

4. Click Save to save the current calibration and system parameters.

5. In the Panoramic Window, click and select Maintenance > Manufacturing


Parameters. The Manufacturing Parameters dialog box appears.

6. Write down all the displayed parameters to check that all the parameters are correct when
after replacing the EEPROMS you restore the system parameters. Click Cancel.

27–2 REPLACING THE MICROPROCESSOR BOARD EEPROMS


Replacing the Microprocessor Board CJ732 EEPROMS

7. Insert the tip of a screw driver gently under each side of the EEPROMs to raise it.

WARNING
You MUST remove the EEPROMs from both sides without
breaking the integrated circuits.

8. Remove the faulty EEPROMs.

Replacing the Microprocessor Board CJ732


EEPROMS
Before replacing the microprocessor board CJ732 EEPROMs, make sure that you have the
appropriate EEPROMs.

To replace the microprocessor board CJ732 EEPROMs, follow these steps:

1. Position correctly and plug in:

• EEPROM U27 with the indentation facing “HIGH”.


• EEPROM U28 with the indentation facing “LOW”.

1. From the fixed arm, on the control panel, press and hold , at the same time
and turn ON the unit.

2. Access the Panoramic Window. If an error message is displayed, click Retry button until
the screen is clear.

3. In the Panoramic Window, click and select Maintenance > Save / Restore. The
Save / Restore dialog box appears.

KODAK 9000 and KODAK 9000 3D Extraoral Imaging System_Service Guide (SM712)_Ed 02 27–3
Testing the Microprocessor Board CJ732 EEPROMS

4. If any error message is displayed, click Restore 2 times to clear the screen. The saved
system and calibration parameters are restored in the new EEPROMS.

5. Access the Panoramic Window. In the Panoramic Window, click and select
Maintenance > Manufacturing Parameters. The Manufacturing Parameters dialog
box appears.

6. Check that the displayed parameters are similar to the initial parameters that you had
written down when saving the system parameters.

7. If installing a cephalostat, select Cephalostat on the right-hand side in the Type of


Cephalometric image field. Click Save.

Testing the Microprocessor Board CJ732


EEPROMS
To test the microprocessor board CJ732 battery, follow these steps:

1. Turn ON the unit. Position the 3D tooling base on the chin rest base. Position the phantom
tooling for the panoramic.

2. From the fixed arm, on the control panel, click to turn ON the mid-sagittal
positioning laser beam. Make sure that the mid-sagittal positioning laser is correctly
centered on the phantom tooling.

3. Access the Panoramic Acquisition Window. Acquire an image to check that the CJ732
EEPROMs is correctly replaced.

27–4 REPLACING THE MICROPROCESSOR BOARD EEPROMS


Chapter 28
REPLACING THE CJ708 BOARD for ROTATIVE
MOVEMENT

WARNING
You MUST switch off the unit before removing the covers. ONLY an
approved Trophy technician is qualified to inspect or maintain the unit
while it is switched on and while the covers are removed. In this case, NO
unqualified person must approach the unit.

Description
The CJ708 step motor board drives the step motor for the rotation movement.

2.5 mm

x2

Removing the CJ708 Board for Rotative


Movement
Before removing the CJ708 step motor board for rotative movement, make sure that:

• The unit is powered OFF.


• The head cover is removed.
• You have the required tool.

KODAK 9000 and KODAK 9000 3D Extraoral Imaging System_Service Guide (SM712)_Ed 02 28–1
Replacing the CJ708 Board for Rotative Movement

To remove the CJ708 step motor board for rotative movement, follow these steps:

1. Disconnect all the connectors.

2. Unscrew the fixing screws.

3. Remove the faulty step motor board.

Replacing the CJ708 Board for Rotative


Movement
Before replacing the CJ708 step motor board for rotative movement, make sure that you have
the appropriate step motor board.

To replace the CJ708 step motor board for rotative movement, follow these steps:

1. Position the CJ708 step motor board making sure that the J2 faces the column.

2. Screw the fixing screws.

3. Connect all the connectors in their respective locations.

Testing the CJ708 Board for Rotative Movement

From the fixed arm, on the control panel press and hold the rotation movement
buttons, to test the correct function of the new CJ708 step motor board for rotative movement.

28–2 REPLACING THE CJ708 BOARD for ROTATIVE MOVEMENT


Chapter 29
REPLACING THE CJ708 BOARD for X and Y
MOVEMENT

WARNING
You MUST switch off the unit before removing the covers. ONLY an
approved Trophy technician is qualified to inspect or maintain the unit
while it is switched on and while the covers are removed. In this case, NO
unqualified person must approach the unit.

Description
The CJ708 step motor board drives the step motor for the X and Y movement.

2.5 mm

x4

Removing the CJ708 Board for X and Y


Movement
Before removing the CJ708 step motor board for X and Y movement, make sure that:

• The unit is powered OFF.


• The head cover is removed.
• You have the required tool.

KODAK 9000 and KODAK 9000 3D Extraoral Imaging System_Service Guide (SM712)_Ed 02 29–1
Replacing the CJ708 Board for X and Y Movement

To remove the CJ708 step motor board for X and Y movement, follow these steps:

1. Disconnect all the connectors.

2. Unscrew the fixing screws.

3. Remove the faulty step motor board.

Replacing the CJ708 Board for X and Y


Movement
Before replacing the CJ708 step motor board for X and Y movement, make sure that you have
the appropriate step motor board.

To replace the CJ708 step motor board for X and Y movement, follow these steps:

1. Position the CJ708 step motor board:

• For the X movement, make sure that the J2 faces the column and screw the fixing
screws.
• For the Y movement, make sure that the J2 faces the front of the unit head and screw
the fixing screws.

2. Connect all the connectors in their respective locations.

Testing the CJ708 Board for X and Y Movement

From the fixed arm, on the control panel press and hold the X and Y movement
buttons, to test the correct function of the new CJ708 step motor board for X and Y movement.

29–2 REPLACING THE CJ708 BOARD for X and Y MOVEMENT


Chapter 30
REMINDER: THE CJ804 BOARD for the CHIN REST

Description
The CJ804 board for the chin rest supplies power to the chin rest motor assembly.

IMPORTANT
This chapter is just a reminder of the configuration of the CJ804
board of the chin rest before the new CJ804 v3 board.

For replacing with the new CJ804 v3 board, see chapter,


“Replacing the CJ804 board for the chin rest”.

x2

2.5 m m

KODAK 9000 and KODAK 9000 3D Extraoral Imaging System_Service Guide (SM712)_Ed 02 30–1
Description

30–2 REMINDER: THE CJ804 BOARD for the CHIN REST


Chapter 31
REPLACING THE CJ804 BOARD for the CHIN REST

WARNING
You MUST switch off the unit before removing the covers. ONLY an
approved Trophy technician is qualified to inspect or maintain the unit
while it is switched on and while the covers are removed. In this case, NO
unqualified person must approach the unit.

Description
The CJ804 board for the chin rest supplies power to the chin rest motor assembly.

x2

S1

2.5 mm

KODAK 9000 and KODAK 9000 3D Extraoral Imaging System_Service Guide (SM712)_Ed 02 31–1
Removing the CJ804 Board for the Chin Rest

Removing the CJ804 Board for the Chin Rest


Before removing the CJ804 board for the chin rest, make sure that:

• The unit is powered OFF.


• The chin rest base cover is removed.
• You have the required tool.

To remove the CJ804 board for the chin rest, follow these steps:

1. Disconnect all the connectors.

2. Unscrew the fixing screws of the board.

3. Remove the faulty board.

Replacing the CJ804 Board for the Chin Rest


Before replacing the CJ804 board for the chin rest, make sure that you have the appropriate
board.

To replace the CJ804 board for the chin rest, follow these steps:

1. Check that the CJ804 board for the chin rest has the correct DIP-switch S1 configuration
as shown in the figure.

IMPORTANT
Make sure that on the welding side, the R1 is soldered as shown in
the figure.

2. Position the CJ804 board for the chin rest and screw the fixing screws.

3. Connect all the connectors in their respective locations.

Testing the CJ804 Board for the Chin Rest

From the fixed arm, on the control panel, press and hold to raise and lower the chin
rest several times to confirm the correct replacement of the CJ804 board.

31–2 REPLACING THE CJ804 BOARD for the CHIN REST


Chapter 32
REMINDER: THE CJ804 BOARD for The
COLLIMATOR

Description
The CJ804 board for the collimator supplies power to the collimator motor assembly.

IMPORTANT
This chapter is just a reminder of the configuration of the CJ804
board of the digital sensor before the new CJ804 v3 board.

For replacing with the new CJ804 v3 board, see chapter,


“Replacing the CJ804 board for the collimator”.

x2

2.5 m m

KODAK 9000 and KODAK 9000 3D Extraoral Imaging System_Service Guide (SM712)_Ed 02 32–1
Description

32–2 REMINDER: THE CJ804 BOARD for The COLLIMATOR


Chapter 33
REPLACING THE CJ804 BOARD for The
COLLIMATOR

WARNING
You MUST switch off the unit before removing the covers. ONLY an
approved Trophy technician is qualified to carry out any intervention on
the unit while it is switched on and while the covers are removed. NO
unqualified person must approach the unit in the above case.

Description
The CJ804 board for the collimator supplies power to the collimator motor assembly.

x2

S1

2.5 mm

KODAK 9000 and KODAK 9000 3D Extraoral Imaging System_Service Guide (SM712)_Ed 02 33–1
Removing the CJ804 Board for the Collimator

Removing the CJ804 Board for the Collimator


Before removing the CJ804 board for the collimator, make sure that:

• The unit is powered OFF.


• The rotative arm cover is removed.
• You have the required tool.

To remove the CJ804 board for the collimator, follow these steps:

1. Disconnect all the connectors.

2. Unscrew the fixing screws of the board.

3. Remove the faulty board.

Replacing the CJ804 Board for the Collimator


Before replacing the CJ804 board for the collimator, make sure that you have the appropriate
board.

To replace the CJ804 board for the collimator, follow these steps:

1. Check that the CJ804 board for the collimator has the correct DIP-switch S1 configuration
as shown in the figure.

2. On the welding side, unsolder the R1 as shown in the figure.

3. Position the CJ804 board for the collimator and screw the fixing screws.

4. Connect all the connectors in their respective locations.

Testing the CJ804 Board for the Collimator


From the PC, access the Panoramic Acquisition Window to check the automatic positioning
of the panoramic collimator. Access the 3D Acquisition Window to check the automatic
positioning of the 3D collimator.

33–2 REPLACING THE CJ804 BOARD for The COLLIMATOR


Chapter 34
REMINDER: THE CJ804 BOARD for the DIGITAL
SENSOR

Description
The CJ804 board for the sensor board supplies power to the digital sensor motor assembly.

IMPORTANT
This chapter is just a reminder of the configuration of the CJ804
board of the digital sensor before the new CJ804 v3 board.

For replacing with the new CJ804 v3 board, see chapters,


“Replacing the CJ804 board for the digital sensor (without ceph)”
and “Replacing the CJ804 board for the digital sensor (with
ceph)”.

x2

2.5 m m

KODAK 9000 and KODAK 9000 3D Extraoral Imaging System_Service Guide (SM712)_Ed 02 34–1
Description

34–2 REMINDER: THE CJ804 BOARD for the DIGITAL SENSOR


Chapter 35
REPLACING THE CJ804 BOARD for the DIGITAL
SENSOR (WITHOUT CEPH)

WARNING
You MUST switch off the unit before removing the covers. ONLY an
approved Trophy technician is qualified to inspect or maintain the unit
while it is switched on and while the covers are removed. In this case, NO
unqualified person must approach the unit.

Description
The CJ804 board for the digital sensor supplies power to the digital sensor motor assembly.

x2

S1

2.5 mm

KODAK 9000 and KODAK 9000 3D Extraoral Imaging System_Service Guide (SM712)_Ed 02 35–1
Removing the CJ804 Board for the Sensor

Removing the CJ804 Board for the Sensor


Before removing the CJ804 board for the digital sensor, make sure that:

• The unit is powered OFF.


• The rotative arm cover is removed.
• You have the required tool.

To remove the CJ804 board for the digital sensor, follow these steps:

1. Disconnect all the connectors.

2. Unscrew the fixing screws of the board.

3. Remove the faulty board.

Replacing the CJ804 Board for the Sensor


Before replacing the CJ804 board for the digital sensor, make sure that you have the
appropriate board.

To replace the CJ804 board for the digital sensor, follow these steps:

1. Check that the CJ804 board for the sensor has the correct DIP-switch S1 configuration as
shown in the figure.

2. On the welding side, unsolder the R1 as shown in the figure.

3. Connect all the connectors in their respective locations.

Testing the CJ804 Board for the Sensor


From the PC, access the Panoramic Acquisition Window to check the 180° automatic
rotation of the digital sensor. Access the 3D Acquisition Window to check the 180° automatic
rotation of the digital sensor.

35–2 REPLACING THE CJ804 BOARD for the DIGITAL SENSOR (WITHOUT CEPH)
Chapter 36
REPLACING THE CJ804 BOARD for the DIGITAL
SENSOR (WITH CEPH)

WARNING
You MUST switch off the unit before removing the covers. ONLY an
approved Trophy technician is qualified to inspect or maintain the unit
while it is switched on and while the covers are removed. In this case, NO
unqualified person must approach the unit.

Description
The CJ804 board for the digital sensor supplies power to the digital sensor motor assembly.

x2

S1

2.5 mm

KODAK 9000 and KODAK 9000 3D Extraoral Imaging System_Service Guide (SM712)_Ed 02 36–1
Removing the CJ804 Board for the Sensor

Removing the CJ804 Board for the Sensor


Before removing the CJ804 board for the digital sensor, make sure that:

• The unit is powered OFF.


• The rotative arm cover is removed.
• You have the required tool.

To remove the CJ804 board for the digital sensor, follow these steps:

1. Disconnect all the connectors.

2. Unscrew the fixing screws of the board.

3. Remove the faulty board.

Replacing the CJ804 Board for the Sensor


Before replacing the CJ804 board for the digital sensor, make sure that you have the
appropriate board.

To replace the CJ804 board for the digital sensor, follow these steps:

1. Check that the CJ804 board for the digital sensor has the correct DIP-switch S1
configuration as shown in the figure.

IMPORTANT
Make sure that on the welding side, the R1 is soldered as shown in
the figure.

2. Position the CJ804 board for the digital sensor and screw the fixing screws.

3. Connect all the connectors in their respective locations.

Testing the CJ804 Board for the Sensor


From the PC, access the Panoramic Acquisition Window to check the 180° automatic
rotation of the digital sensor. Access the 3D Acquisition Window to check the 180° automatic
rotation of the digital sensor.

36–2 REPLACING THE CJ804 BOARD for the DIGITAL SENSOR (WITH CEPH)
Chapter 37
REPLACING THE ENCODER ASSEMBLY (Unit Serial
N° All xxACxxx)

WARNING
You MUST switch off the unit before removing the covers. ONLY an
approved Trophy technician is qualified to inspect or maintain the unit
while it is switched on and while the covers are removed. In this case, NO
unqualified person must approach the unit.

Description
The encoder assembly has the following functions:

• Transmits information of the rotative arm position to the CJ732 microprocessor board.
• Detects the rotative arm movement during acquisition.

3 mm

Removing the Encoder Assembly


Before removing the encoder assembly, make sure that:.

• The unit is powered OFF.


• The unit head cover is removed.

KODAK 9000 and KODAK 9000 3D Extraoral Imaging System_Service Guide (SM712)_Ed 02 37–1
Replacing the Encoder Assembly

• You have the required tool.

To remove the encoder assembly, follow these steps:

1. Remove the pulley belt.

2. Disconnect the connector.

3. Unscrew the fixing screws.

TIP
Place a screw driver under the encoder to keep the brake pad (A)
under pressure when unscrewing the fixing screws. This avoids the
ejection of the brake pad (A) when the pressure is lifted.

4. Remove the faulty encoder assembly.

Replacing the Encoder Assembly


Before replacing the encoder assembly, make sure that you have the appropriate encoder
assembly.

To replace the encoder assembly, follow these steps:

WARNING
DO NOT touch the encoder wheel at any time in the replacement
procedure. The encoder is very fragile.

1. Position but do not tighten the right fixing screw.

2. Put the brake pad (A) in place and with your finger keep it under pressure.

3. Slide the brake pad (A) still under pressure to screw the left fixing screw.

4. Place a screw driver under the encoder and tighten the fixing screws. Remove the screw
driver.

5. Connect the connector.

6. Turn the encoder manually to check that the encoder wheel is free in movement.

7. Replace the pulley belt around the big pulley and then around the encoder.

37–2 REPLACING THE ENCODER ASSEMBLY (Unit Serial N° All xxACxxx)


Testing the Encoder Assembly

Testing the Encoder Assembly

From the fixed arm, on the control panel, click to check that the encoder is replaced
correctly.

KODAK 9000 and KODAK 9000 3D Extraoral Imaging System_Service Guide (SM712)_Ed 02 37–3
Testing the Encoder Assembly

37–4 REPLACING THE ENCODER ASSEMBLY (Unit Serial N° All xxACxxx)


Chapter 38
REPLACING THE ENCODER ASSEMBLY (Up to Unit
Serial N° WKAH046)

WARNING
You MUST switch off the unit before removing the covers. ONLY an
approved Trophy technician is qualified to inspect or maintain the unit
while it is switched on and while the covers are removed. In this case, NO
unqualified person must approach the unit.

Description
The encoder assembly has the following functions:

• Transmits information of the rotative arm position to the CJ732 microprocessor board.
• Detects the rotative arm movement during acquisition.

3 mm

Removing the Encoder Assembly


Before removing the encoder assembly, make sure that:.

• The unit is powered OFF.


• The unit head cover is removed.

KODAK 9000 and KODAK 9000 3D Extraoral Imaging System_Service Guide (SM712)_Ed 02 38–1
Replacing the Encoder Assembly

• You have the required tool.

To remove the encoder assembly, follow these steps:

1. Remove the pulley belt.

2. Disconnect the connector.

3. Unscrew the fixing screws.

TIP
Place a screw driver under the encoder to keep the brake pad (A)
under pressure when unscrewing the fixing screws. This avoids the
ejection of the brake pad (A) when the pressure is lifted.

4. Remove the faulty encoder assembly.

Replacing the Encoder Assembly


Before replacing the encoder assembly, make sure that you have the appropriate encoder
assembly.

To replace the encoder assembly, follow these steps:

WARNING
DO NOT touch the encoder wheel at any time in the replacement
procedure. The encoder is very fragile.

1. Position but do not tighten the right fixing screw.

2. Put the brake pad (A) in place and with your finger keep it under pressure.

3. Slide the brake pad (A) still under pressure to screw the left fixing screw.

4. Place a screw driver under the encoder and tighten the fixing screws. Remove the screw
driver.

5. Connect the connector.

6. Turn the encoder manually to check that the encoder wheel is free in movement.

7. Replace the pulley belt around the big pulley and then around the encoder.

38–2 REPLACING THE ENCODER ASSEMBLY (Up to Unit Serial N° WKAH046)


Testing the Encoder Assembly

Testing the Encoder Assembly

From the fixed arm, on the control panel, click to check that the encoder is replaced
correctly.

KODAK 9000 and KODAK 9000 3D Extraoral Imaging System_Service Guide (SM712)_Ed 02 38–3
Testing the Encoder Assembly

38–4 REPLACING THE ENCODER ASSEMBLY (Up to Unit Serial N° WKAH046)


Chapter 39
REPLACING THE ENCODER ASSEMBLY (From Unit
Serial N° WKAH047)

WARNING
You MUST switch off the unit before removing the covers. ONLY an
approved Trophy technician is qualified to inspect or maintain the unit
while it is switched on and while the covers are removed. In this case, NO
unqualified person must approach the unit.

Description
The encoder assembly has the following functions:

• Transmits information of the rotative arm position to the CJ732 microprocessor board.
• Detects the rotative arm movement during acquisition.

A 3 mm

Removing the Encoder Assembly


Before removing the encoder assembly, make sure that:.

• The unit is powered OFF.


• The unit head cover is removed.
• You have the required tool.

KODAK 9000 and KODAK 9000 3D Extraoral Imaging System_Service Guide (SM712)_Ed 02 39–1
Replacing the Encoder Assembly

To remove the encoder assembly, follow these steps:

1. Disconnect the connector.

2. Unscrew the fixing screws.

3. Remove the faulty encoder assembly.

Replacing the Encoder Assembly


Before replacing the encoder assembly, make sure that you have the appropriate encoder
assembly.

To replace the encoder assembly, follow these steps:

WARNING
DO NOT touch the encoder wheel at any time in the replacement
procedure. The encoder is very fragile.

1. Position and turn the encoder manually to check that the encoder wheel is free in
movement.

2. Tighten the fixing screws.

3. Connect the connector.

Testing the Encoder Assembly

From the fixed arm, on the control panel, click to check that the encoder is replaced
correctly.

39–2 REPLACING THE ENCODER ASSEMBLY (From Unit Serial N° WKAH047)


Chapter 40
REPLACING THE CJ691 LASER BOARD (HEAD)

WARNING
You MUST switch off the unit before removing the covers. ONLY an
approved Trophy technician is qualified to inspect or maintain the unit
while it is switched on and while the covers are removed. In this case, NO
unqualified person must approach the unit.

Description
The CJ691 laser board creates a circuit connection with the laser beams.

x4

2.5 mm

KODAK 9000 and KODAK 9000 3D Extraoral Imaging System_Service Guide (SM712)_Ed 02 40–1
Removing the CJ691 Laser Board (Head)

Removing the CJ691 Laser Board (Head)


Before removing the CJ691 laser board, make sure that:

• The unit is powered OFF.


• The unit head cover is removed.
• You have the required tool.

To remove the laser CJ691 laser board, follow these steps:

1. Disconnect all the connectors.

2. Unscrew the 2 fixing screws of the board.

3. Remove the faulty board.

Replacing the CJ691 Laser Board (Head)


Before replacing the CJ691 laser board, make sure that you have the appropriate board.

To replace the CJ691 laser board, follow these steps:

1. Position the laser board and screw the 2 fixing screws.

2. Connect all the connectors.

Testing the CJ691 Laser Board (Head)

From the fixed arm, on the control panel, click to turn ON the positioning laser beams
to check that the CJ691 laser board is replaced correctly.

40–2 REPLACING THE CJ691 LASER BOARD (HEAD)


Chapter 41
REPLACING THE CJ691 LASER BOARD (ROTATIVE
ARM)

WARNING
You MUST switch off the unit before removing the covers. ONLY an
approved Trophy technician is qualified to inspect or maintain the unit
while it is switched on and while the covers are removed. In this case, NO
unqualified person must approach the unit.

Description
The CJ691 laser board creates a circuit connection with the laser beams.

2.5 mm

x4

KODAK 9000 and KODAK 9000 3D Extraoral Imaging System_Service Guide (SM712)_Ed 02 41–1
Removing the CJ691 Laser Board (Rotative Arm)

Removing the CJ691 Laser Board (Rotative Arm)


Before removing the CJ691 laser board, make sure that:

• The unit is powered OFF.


• The generator front cover is removed.
• You have the required tool.

To remove the CJ691 laser board, follow these steps:

1. Disconnect all the connectors.

2. Unscrew the 2 fixing screws of the Board.

3. Remove the faulty Board.

Replacing the CJ691 Laser Board (Rotative Arm)


Before replacing the CJ691 laser board, make sure that you have the appropriate board.

To replace the CJ691 laser board, follow these steps:

1. Position the laser board and screw the 2 fixing screws.

2. Connect all the connectors.

3. Position the generator front cover and screw the fixing screws.

Testing the CJ691 Laser Board (Rotative Arm)

From the fixed arm, on the control panel, click to turn ON the positioning laser beams
to check that the laser board CJ691 is replaced correctly.

41–2 REPLACING THE CJ691 LASER BOARD (ROTATIVE ARM)


Chapter 42
REPLACING THE HORIZONTAL POSITIONING
LASER (Unit Serial N°s: xxACxxx, xxADxxx, xxAHxxx)

WARNING
You MUST switch off the unit before removing the covers. ONLY an
approved Trophy technician is qualified to inspect or maintain the unit
while it is switched on and while the covers are removed. In this case, NO
unqualified person must approach the unit.

Description
The horizontal positioning laser enables the correct position of the patient according to the
Frankfurt positioning plane.

x4

x1
1.5 mm
2 mm
2.5 mm

KODAK 9000 and KODAK 9000 3D Extraoral Imaging System_Service Guide (SM712)_Ed 02 42–1
Removing the Horizontal Positioning Laser

Removing the Horizontal Positioning Laser


Before remove the horizontal positioning laser, make sure that:

• The unit is powered OFF.


• You have the required tool.

To remove the horizontal positioning laser, follow these steps:

1. Unscrew and remove the laser knob.

2. Unscrew and remove the generator front cover.

3. Disconnect the laser connector.

4. Unscrew the fixing screws of the laser slide (A).

5. Unscrew the fixing screw of the laser.

6. Remove the faulty laser.

Replacing the Horizontal Positioning Laser


Before replacing the horizontal positioning laser, make sure that you have the appropriate
laser.

To replace the horizontal positioning laser, follow these steps:

1. Position the horizontal positioning laser, making sure that the laser does not touch the
metal frame. Screw the fixing screw.

2. Screw the fixing screws of the laser slide (A).

3. Connect the laser connector.

4. Turn ON the unit. On the control panel, click to turn ON the positioning laser
beams. Make sure that the horizontal laser beam is positioned correctly. If needed, adjust
the laser using the fixing screw of the laser slide.

5. Position and screw the fixing screws of the generator front cover.

6. Position and screw the fixing screw of the laser knob.

Testing the Horizontal Positioning Laser


See procedure 4 of the horizontal positioning laser replacement.

42–2 REPLACING THE HORIZONTAL POSITIONING LASER (Unit Serial N°s: xxACxxx, xxADxxx, xxAHxxx)
Chapter 43
REPLACING THE HORIZONTAL POSITIONING
LASER (Unit Serial N°s: xxAExxx, xxAFxxx, xxAJxxx,
xxAKxxx)

WARNING
You MUST switch off the unit before removing the covers. ONLY an
approved Trophy technician is qualified to inspect or maintain the unit
while it is switched on and while the covers are removed. In this case, NO
unqualified person must approach the unit.

Description
The horizontal positioning laser enables the correct position of the patient according to the
Frankfurt positioning plane.

3 mm

x1

Removing the Horizontal Positioning Laser


Before remove the horizontal positioning laser, make sure that:

• The unit is powered OFF.


• The generator front cover is removed.

KODAK 9000 and KODAK 9000 3D Extraoral Imaging System_Service Guide (SM712)_Ed 02 43–1
Replacing the Horizontal Positioning Laser

• You have the required tool.

To remove the horizontal positioning laser, follow these steps:

1. Disconnect the laser connector.

2. Unscrew the fixing screw of the laser.

3. Remove the faulty laser.

Replacing the Horizontal Positioning Laser


Before replacing the horizontal positioning laser, make sure that you have the appropriate
laser.

To replace the horizontal positioning laser, follow these steps:

1. Position the horizontal positioning laser and screw the fixing screw.

2. Connect the laser cable connector to CJ691 board on the unit head to the “LASER FACE”
connector.

3. Attach the cable with a cable tie.

43–2 REPLACING THE HORIZONTAL POSITIONING LASER (Unit Serial N°s: xxAExxx, xxAFxxx, xxAJxxx, xxAKxxx)
Testing the Horizontal Positioning Laser

4. Turn on the unit. From the fixed arm, on the control panel, press to turn on the
positioning laser beams.

5. Make sure that the horizontal positioning laser beam is projected above and parallel to the
3D digital sensor label.

6. If needed, adjust the laser using the fixing screw of the laser socket.

Testing the Horizontal Positioning Laser


See procedure 4 through 6 of the horizontal positioning laser replacement.

DRAFTKODAK 9000 and KODAK 9000 3D Extraoral Imaging System_Service Guide (SM712)_Ed 0243–3
Testing the Horizontal Positioning Laser

43–4 REPLACING THE HORIZONTAL POSITIONING LASER (Unit Serial N°s: xxAExxx, xxAFxxx, xxAJxxx, xxAKxxx)
Chapter 44
REPLACING THE MID-SAGITTAL POSITIONING
LASER

WARNING
You MUST switch off the unit before removing the covers. ONLY an
approved Trophy technician is qualified to inspect or maintain the unit
while it is switched on and while the covers are removed. In this case, NO
unqualified person must approach the unit.

Description
The mid-sagittal positioning laser beam enables the correct central positioning of the patient.

1.5 mm

x1

KODAK 9000 and KODAK 9000 3D Extraoral Imaging System_Service Guide (SM712)_Ed 02 44–1
Removing the Mid-Sagittal Positioning Laser

Removing the Mid-Sagittal Positioning Laser


Before removing the mid-sagittal positioning laser, make sure that:

• The unit is powered OFF.


• The head cover is removed.
• You have the required tool.

To remove the mid-sagittal positioning laser, follow these steps:

1. Disconnect the cable.

2. Unscrew the fixing screw.

3. Remove the faulty laser.

Replacing the Mid-Sagittal Positioning Laser


Before replacing the mid-sagittal positioning laser, make sure that you have the appropriate
laser.

To replace the mid-sagittal positioning laser, follow these steps:

1. Position the mid-sagittal positioning laser in the laser socket and screw the fixing screw.

2. Connect the connector.

3. Turn ON the unit. On the control panel, click to turn ON the positioning laser
beams. Make sure that the mid-sagittal positioning laser is correctly centered on the bite
block. If needed, adjust the laser using the fixing screw of the mid-sagittal laser.

4. Replace the head cover.

Testing the Mid-Sagittal Positioning Laser


See procedure 3 of the mid-sagittal positioning laser replacement.

44–2 REPLACING THE MID-SAGITTAL POSITIONING LASER


Chapter 45
REPLACING THE 3D CENTRAL POSITIONING
LASER

WARNING
You MUST switch off the unit before removing the covers. ONLY an
approved Trophy technician is qualified to inspect or maintain the unit
while it is switched on and while the covers are removed. In this case, NO
unqualified person must approach the unit.

Description
The 3D central positioning laser enables the horizontal positioning laser beam to correctly
target the region of interest for 3D acquisition using a dental plaster imprint.

1.5 mm
2 mm
3 mm

x1

x3

x8
A

KODAK 9000 and KODAK 9000 3D Extraoral Imaging System_Service Guide (SM712)_Ed 02 45–1
Removing the 3D Central Positioning Laser

Removing the 3D Central Positioning Laser


Before removing the 3D central positioning laser, make sure that:

• Unit is powered OFF.


• You have the required tool.

To remove the 3D central positioning laser, follow these steps:

1. Unscrew the fixing screw of the laser knob. Pull to remove the knob.

2. Unscrew the fixing screws of the generator front cover and remove the cover.

3. Unscrew the fixing screws of the rotative arm bottom cover and remove the cover.

4. Unscrew the fixing screws of the cable protection component (A) and remove the
protection component.

5. Disconnect the laser connector from the CJ691 board.

6. Unscrew and remove the laser socket.

7. Unscrew the fixing screw of the laser. Remove the faulty laser.

Replacing the 3D Central Positioning Laser


Before replacing the 3D central positioning laser, make sure that you have the appropriate
laser.

To replace the 3D central positioning laser, follow these steps:

1. Position the 3D central positioning laser in the socket and screw the fixing screw.

2. Position the laser socket and screw the fixing screws.

3. Connect the FoV laser cable to the CJ691 on the rotative arm on the “LASER LEFT”.

4. Place a white paper on the floor and attach with an adhesive tape. This is to correctly
adjust the 3D central positioning laser.

45–2 REPLACING THE 3D CENTRAL POSITIONING LASER


Testing the 3D Central Positioning Laser

5. Turn ON the unit. On the control panel, click to reset the unit rotative arm.

6. On the control panel, click to turn ON the positioning laser beams.

7. Mark the position of the laser with an X.

8. From the control panel, press and hold to turn the rotative arm 90 degrees at a
time. Mark at each 90 degrees, the position of the laser with an X. At the end of a full
rotation of the rotative arm, you will have 4 marked Xs. Draw a line between each 2 Xs.
The center of the crossed lines is the correct position of the laser.
90°

9. Using the fixing screws of the laser socket, adjust the 3D central positioning laser beam
until it is at the center of the crossed lines.

10. Position the cable protection component (A) and screw the fixing screws.

11. Position the rotative arm bottom cover and screw the fixing screws.

12. Position the generator front cover and screw the fixing screws.

13. Position the laser knob and screw the fixing screw.

Testing the 3D Central Positioning Laser


See procedures 4 to 9 of the 3D central positioning laser replacement.

KODAK 9000 and KODAK 9000 3D Extraoral Imaging System_Service Guide (SM712)_Ed 02 45–3
Testing the 3D Central Positioning Laser

45–4 REPLACING THE 3D CENTRAL POSITIONING LASER


Chapter 46
REPLACING THE 3D FOV POSITIONING LASER
(Unit Serial N°s: xxACxxx, xxADxxx, xxAHxxx)

WARNING
You MUST switch off the unit before removing the covers. ONLY an
approved Trophy technician is qualified to inspect or maintain the unit
while it is switched on and while the covers are removed. In this case, NO
unqualified person must approach the unit.

Description
The 3D FoV positioning laser enables the correct position of the patient for the 3D positioning
plane.

x1

1.5 mm
2 mm
2.5 mm

KODAK 9000 and KODAK 9000 3D Extraoral Imaging System_Service Guide (SM712)_Ed 02 46–1
Removing the 3D FoV Positioning Laser

Removing the 3D FoV Positioning Laser


Before removing the 3D FoV positioning laser, make sure that:

• The unit is powered OFF.


• You have the required tool

To remove the 3D FoV positioning laser, follow these steps:

1. Unscrew the fixing screw of the laser knob. Pull to remove the faulty knob.

2. Unscrew and remove the generator front cover.

3. Disconnect the laser connector.

4. Unscrew the fixing screw of the 3D FoV positioning laser.

5. Remove the faulty 3D FoV positioning laser.

Replacing the 3D FoV Positioning Laser


Before replacing the 3D FoV positioning laser, make sure that you have the appropriate laser.

To replace the 3D FoV positioning laser, follow these steps:

1. Position the 3D FoV positioning laser and screw the fixing screw.

2. Connect the laser connector.

3. Turn ON the unit. On the control panel, click to turn ON the positioning laser
beams. Make sure that the 3D FoV positioning laser is correctly projected on the extreme
right of the sensor Lexan. If needed, adjust the laser using the fixing screw of the laser
socket.

4. Position and screw the fixing screws of the generator front cover.

5. Position and screw the fixing screw of the laser knob.

Testing the 3D FoV Positioning Laser


See procedure 3 of the 3D FoV positioning laser replacement.

46–2 REPLACING THE 3D FOV POSITIONING LASER (Unit Serial N°s: xxACxxx, xxADxxx, xxAHxxx)
Chapter 47
REPLACING THE CEPHALOMETRIC FRANKFURT
POSITIONING LASER (Unit Serial N°s: xxAExxx,
xxAFxxx, xxAJxxx, xxAKxxx)

WARNING
You MUST switch off the unit before removing the covers. ONLY an
approved Trophy technician is qualified to inspect or maintain the unit
while it is switched on and while the covers are removed. In this case, NO
unqualified person must approach the unit.

Description
The horizontal positioning laser enables the correct position of the patient according to the
Frankfurt positioning plane.

x2
2.5 mm

Removing the Cephalometric Frankfurt


Positioning Laser
Before remove the cephalometric Frankfurt positioning laser, make sure that:

• The unit is powered OFF.

KODAK 9000 and KODAK 9000 3D Extraoral Imaging System_Service Guide (SM712)_Ed 02 47–1
Replacing the Cephalometric Frankfurt Positioning Laser

• The generator front cover is removed.


• You have the required tool.

To remove the cephalometric Frankfurt positioning laser, follow these steps:

1. Disconnect the laser connector.

2. Unscrew the fixing screws of the cephalometric Frankfurt positioning laser.

3. Remove the faulty laser.

Replacing the Cephalometric Frankfurt


Positioning Laser
Before replacing the cephalometric Frankfurt positioning laser, make sure that you have the
appropriate laser.

To replace the cephalometric Frankfurt positioning laser, follow these steps:

1. Position correctly the cephalometric Frankfurt positioning laser and screw the fixing
screws.

2. Connect the laser cable 976/laser ceph to cable 972/CJ732/J28. Attach the cable with a
cable tie.

3. Turn on the unit. From the fixed arm, on the control panel, press to turn on the
positioning laser beams.

4. Make sure that the cephalometric Frankfurt positioning laser beam is projected in the
middle of the ear cone.

5. If needed, adjust the laser using the fixing screw of the laser socket.

47–2 REPLACING THE CEPHALOMETRIC FRANKFURT POSITIONING LASER (Unit Serial N°s: xxAExxx, xxAFxxx, xxAJxxx, xxAKxxx)
Testing the Cephalometric Frankfurt Positioning Laser

Testing the Cephalometric Frankfurt Positioning


Laser
See procedure 3 through 5 of the cephalometric Frankfurt positioning laser replacement.

DRAFTKODAK 9000 and KODAK 9000 3D Extraoral Imaging System_Service Guide (SM712)_Ed 0247–3
Testing the Cephalometric Frankfurt Positioning Laser

47–4 REPLACING THE CEPHALOMETRIC FRANKFURT POSITIONING LASER (Unit Serial N°s: xxAExxx, xxAFxxx, xxAJxxx, xxAKxxx)
Chapter 48
REPLACING THE LF439 CAPACITORS

WARNING
You MUST switch off the unit before removing the covers. ONLY an
approved Trophy technician is qualified to inspect or maintain the unit
while it is switched on and while the covers are removed. In this case, NO
unqualified person must approach the unit.

Description
The LF439 Capacitors has the following functions:

• Filters the DC circuit.


• Stocks energy for the CJ653 inverter board.

WARNING
You MUST replace both capacitors.
The color of the connections are relevant to the polarity (positive
or negative) indicated on the capacitors.

D
-

B C
- +
A
+

KODAK 9000 and KODAK 9000 3D Extraoral Imaging System_Service Guide (SM712)_Ed 02 48–1
Removing the LF439 Capacitors

Removing the LF439 Capacitors


Before removing the LF439 capacitors, make sure that:

• The unit is powered OFF.


• The head cover is removed.
• You have the required tools.

To remove the LF439 capacitors, follow these steps:

1. Disconnect the connectors.

2. Unscrew the holding rods screws.

3. Remove BOTH capacitors.

Replacing the LF439 Capacitors


Before replacing the LF439 capacitors, make sure that you have the appropriate capacitors.

To replace BOTH LF439 capacitors, follow these steps:

1. Unscrew the holding rods screws and position the LF439 capacitors with the positive (+)
signs facing the column.

2. Tighten the holding rods screws.

3. Connect the following connectors to the appropriate polarity as shown in the figure:

• Red (A) to positive


• Orange (B) to negative
• Orange (C) to positive
• Black (D) to negative

Testing the LF439 Capacitors


Launch an x-ray with 70kV and 12mA parameters to check that the capacitors are replaced
correctly and functions properly.

48–2 REPLACING THE LF439 CAPACITORS


Chapter 49
REPLACING THE ROTATIVE ARM MOTOR
ASSEMBLY (Unit Serial N° xxACxxx)

WARNING
You MUST switch off the unit before removing the covers. ONLY an
approved Trophy technician is qualified to inspect or maintain the unit
while it is switched on and while the covers are removed. In this case, NO
unqualified person must approach the unit.

Description
The rotative arm motor assembly enables the rotative arm movement.

B
C

B D

2.5 mm
3 mm

Removing the Rotative Arm Motor Assembly


Before removing the rotative arm motor assembly, make sure that:

KODAK 9000 and KODAK 9000 3D Extraoral Imaging System_Service Guide (SM712)_Ed 02 49–1
Replacing the Rotative Arm Motor Assembly

• The unit is powered OFF.


• The unit head cover is removed.
• You have the required tool.

To remove the rotative arm motor assembly, follow these steps:

1. Disconnect J2 cable from the CJ708 board (A). Unscrew and remove the CJ708 board.

2. Unscrew the fixing screws of the rotative arm motor assembly cable clamps (B). Remove
the cable clamps.

3. Unscrew and remove the fixing screw under pressure (C).

4. Unscrew and remove the black shoulder fixing screw (D).

5. Remove the faulty rotative arm motor assembly.

Replacing the Rotative Arm Motor Assembly


Before replacing the rotative arm motor assembly, make sure that you have the appropriate
component.

To replace the rotative arm motor assembly, follow these steps:

1. Position the rotative arm motor assembly and screw the black shoulder (D) fixing screw
but do not over tighten.

2. Position and screw the fixing screw under pressure (C).

WARNING
DO NOT exceed the mechanical frame (see the figure).

3. Position the rotative arm motor assembly cable clamps (B) and screw the fixing screws.

4. Position and screw the fixing screw of the CJ708 board (A). Connect J2 cable to the CJ708
board.

5. Attach the cables with a cable tie.

Testing the Rotative Arm Motor Assembly

From the fixed arm, on the control panel, click to check that the rotative arm motor
assembly is replaced correctly

49–2 REPLACING THE ROTATIVE ARM MOTOR ASSEMBLY (Unit Serial N° xxACxxx)
Chapter 50
REPLACING THE ROTATIVE ARM MOTOR
ASSEMBLY (Unit Serial N°s: xxADxxx, xxAHxxx,
xxAExxx, xxAFxxx, xxAJxxx, xxAKxxx)

WARNING
You MUST switch off the unit before removing the covers. ONLY an
approved Trophy technician is qualified to inspect or maintain the unit
while it is switched on and while the covers are removed. In this case, NO
unqualified person must approach the unit.

Description
The rotative arm motor assembly enables the rotative arm movement.

C
E

B
A D

2.5 mm
3 mm

KODAK 9000 and KODAK 9000 3D Extraoral Imaging System_Service Guide (SM712)_Ed 02 50–1
Removing the Rotative Arm Motor Assembly

Removing the Rotative Arm Motor Assembly


Before removing the rotative arm motor assembly, make sure that:

• The unit is powered OFF.


• The unit head cover is removed.
• You have the required tool.

To remove the rotative arm motor assembly, follow these steps:

1. Disconnect the connectors from the CJ822 board (A). Unscrew and remove the CJ822
board

2. Disconnect J2 cable from the CJ708 board (B). Unscrew and remove the CJ708 board.

3. Unscrew and remove the fixing screws of the rotative arm motor assembly cable clamps
(C).

4. Unscrew and remove the fixing screw under pressure (D).

5. Unscrew and remove the black shoulder fixing screw (E).

6. Remove the faulty rotative arm motor assembly.

Replacing the Rotative Arm Motor Assembly


Before replacing the rotative arm motor assembly, make sure that you have the appropriate
component.

To replace the rotative arm motor assembly, follow these steps:

1. Position the rotative arm motor assembly and screw the black shoulder fixing screw (E)
but do not over tighten.

2. Position and screw the fixing screw under pressure (D).

WARNING
DO NOT exceed the mechanical frame (see the figure).

3. Position the rotative arm motor assembly cable clamps (C) and screw the fixing screws.

4. Position and screw the fixing screw of the CJ708 board (B). Connect J2 cable to the CJ708
board.

5. Position and screw the fixing screw of the CJ822 board (B). Connect the connectors.

6. Manually turn the rotative arm to check that the metal flags pass through the optical limit
switch appropriately.

7. Attach the cables with a cable tie.

50–2 REPLACING THE ROTATIVE ARM MOTOR ASSEMBLY (Unit Serial N°s: xxADxxx, xxAHxxx, xxAExxx, xxAFxxx, xxAJxxx, xx-
Testing the Rotative Arm Motor Assembly

8. Calibrate the rotative arm (see Chapter Using the Technical Menu, section Rotative Arm
Calibration).

Testing the Rotative Arm Motor Assembly

From the fixed arm, on the control panel, click to check that the rotative arm motor
assembly is replaced correctly.

KODAK 9000 and KODAK 9000 3D Extraoral Imaging System_Service Guide (SM712)_Ed 02 50–3
Testing the Rotative Arm Motor Assembly

50–4 REPLACING THE ROTATIVE ARM MOTOR ASSEMBLY (Unit Serial N°s: xxADxxx, xxAHxxx, xxAExxx, xxAFxxx, xxAJxxx, xx-
Chapter 51
REPLACING THE ROTATIVE MOVEMENT LIMIT
SWITCH (Unit Serial N° xxACxxx)

WARNING
You MUST switch off the unit before removing the covers. ONLY an
approved Trophy technician is qualified to inspect or maintain the unit
while it is switched on and while the covers are removed. In this case, NO
unqualified person must approach the unit.

Description
The rotative movement limit switch limits the movement of the rotative arm.

2.5 mm

x2

KODAK 9000 and KODAK 9000 3D Extraoral Imaging System_Service Guide (SM712)_Ed 02 51–1
Removing the Rotative Movement Limit Switch

Removing the Rotative Movement Limit Switch


Before removing the rotative movement limit switch, make sure that:

• The unit is powered OFF.


• The unit head cover is removed.
• You have the required tool.

To remove the rotative movement limit switch, follow these steps:

1. Unsolder and unscrew the fixing screws.

CAUTION
Identify the orientation of the switch and the position of the cables
before removal.

2. Disconnect the limit switch (A).

3. Remove the faulty limit switch (A).

Replacing the Rotative Movement Limit Switch


Before replacing the rotative movement limit switch, make sure that you have the appropriate
limit switch.

To replace the rotative movement limit switch, follow these steps:

1. Position and screw the fixing screws.

2. Solder the cables to the lugs.

CAUTION
DO NOT overheat the lugs.

Testing the Rotative Movement Limit Switch

From the fixed arm, on the control panel, click to check that the rotative movement
limit switch is replaced correctly.

51–2 REPLACING THE ROTATIVE MOVEMENT LIMIT SWITCH (Unit Serial N° xxACxxx)
Chapter 52
REPLACING THE ROTATIVE MOVEMENT CJ822
BOARD (Unit Serial N°s: xxADxxx, xxAHxxx, xxAExxx,
xxAFxxx, xxAJxxx, xxAKxxx)

WARNING
You MUST switch off the unit before removing the covers. ONLY an
approved Trophy technician is qualified to inspect or maintain the unit
while it is switched on and while the covers are removed. In this case, NO
unqualified person must approach the unit.

Description
The rotative movement limit optical switch board CJ822 limits the movement of the rotative
arm.

2.5 mm

x2

A
A

KODAK 9000 and KODAK 9000 3D Extraoral Imaging System_Service Guide (SM712)_Ed 02 52–1
Removing the Rotative Movement CJ822 Board

Removing the Rotative Movement CJ822 Board


Before removing the rotative movement optical limit switch board, make sure that:

• The unit is powered OFF.


• The unit head cover is removed.
• You have the required tool.

To remove the rotative movement optical limit switch board, follow these steps:

1. Disconnect the connectors.

2. Unscrew the fixing screws of the optical limit switch board.

3. Remove the faulty optical limit switch board.

Replacing the Rotative Movement CJ822 Board


Before replacing the rotative movement optical limit switch board, make sure that you have
the appropriate limit switch.

To replace the rotative movement optical limit switch board, follow these steps:

1. Position correctly and screw the fixing screws.

2. Connect the connectors.

3. Manually turn the rotative arm to check that the metal flags (A) pass appropriately through
the optical sensor.

4. Calibrate the rotative arm (see Chapter Using the Technical Menu, section Rotative Arm
Calibration).

Testing the Rotative Movement CJ822 Board

From the fixed arm, on the control panel, click to check that the rotative movement
optical limit switch board is replaced correctly.

52–2 REPLACING THE ROTATIVE MOVEMENT CJ822 BOARD (Unit Serial N°s: xxADxxx, xxAHxxx, xxAExxx, xxAFxxx, xxAJxxx, xx-
Chapter 53
REPLACING THE X OR Y MOTOR ASSEMBLY

WARNING
You MUST switch off the unit before removing the covers. ONLY an
approved Trophy technician is qualified to inspect or maintain the unit
while it is switched on and while the covers are removed. In this case, NO
unqualified person must approach the unit.

Description
The X and Y motor assembly enables the rotative arm movement.

2 mm
2.5 mm

KODAK 9000 and KODAK 9000 3D Extraoral Imaging System_Service Guide (SM712)_Ed 02 53–1
Removing the X or Y Motor Assembly

Removing the X or Y Motor Assembly


Before removing the X or Y motor assembly, make sure that:

• The unit is powered OFF.


• The unit head cover is removed.
• You have the required tool.

To remove the X or Y motor assembly, follow these steps:

1. loosen but do not remove, the fixing screw of the flexible X or Y motor coupling closest to
the X or Y motor assembly.

2. Disconnect the X or Y motor assembly cable.

3. Unscrew the fixing screws of the X or Y motor assembly.

4. Remove the faulty X or Y motor assembly.

Replacing the X or Y Motor Assembly


Before replacing the X or Y motor assembly, make sure that you have the appropriate
component.

To replace the X or Y motor assembly, follow these steps:

1. Insert the X or Y motor assembly shaft inside the flexible motor coupling. Screw the
fixing screws of the X or Y motor assembly.

2. Connect the X or Y motor assembly cable.

3. Tighten the fixing screw of the flexible X or Y motor coupling.

Testing the X or Y Motor Assembly

From the fixed arm, on the control panel, press and hold to test:

• The X movement of the rotative arm


• The Y movement of the rotative arm

Make sure that you retry several times to check that the X or Y motor assembly is replaced
correctly.

53–2 REPLACING THE X OR Y MOTOR ASSEMBLY


Chapter 54
REPLACING THE X OR Y FLEXIBLE MOTOR
COUPLING

WARNING
You MUST switch off the unit before removing the covers. ONLY an
approved Trophy technician is qualified to inspect or maintain the unit
while it is switched on and while the covers are removed. In this case, NO
unqualified person must approach the unit.

Description
The X and Y flexible motor coupling connect the X and Y motor assembly and the X and Y
double pitch worm allowing some degree of misalignment.

1.5 mm
2.5 mm

Removing the X or Y Flexible Motor Coupling


Before removing the X or Y flexible motor coupling, make sure that:

• The unit is powered OFF.


• The unit head cover is removed.
• You have the required tool.

KODAK 9000 and KODAK 9000 3D Extraoral Imaging System_Service Guide (SM712)_Ed 02 54–1
Replacing the X or Y Flexible Motor Coupling

To remove the X or Y flexible motor coupling, follow these steps:

1. loosen but do not remove the fixing screw of the flexible motor coupling closest to the X
or Y motor assembly.

2. Disconnect the X or Y motor assembly cable.

3. Unscrew the fixing screws of the X or Y motor assembly.

4. Remove the X or Y motor assembly.

5. Unscrew the fixing screws of the X or Y flexible motor coupling.

6. Remove the faulty X or Y flexible motor coupling.

Replacing the X or Y Flexible Motor Coupling


Before replacing the X or Y flexible motor coupling, make sure that you have the appropriate
component.

To replace the X or Y flexible motor coupling, follow these steps:

1. Position, but do not tighten the fixing screws of the X or Y flexible motor coupling.

2. Position and screw the fixing screws of the X or Y motor assembly.

3. Tighten the fixing screws of the X or Y flexible motor coupling.

4. Connect the X or Y motor assembly cable.

Testing the X or Y Flexible Motor Coupling

From the fixed arm, on the control panel, press and hold to test:

• The X movement of the rotative arm


• The Y movement of the rotative arm

Test the X or Y flexible motor coupling several times to check that it is replaced correctly.

54–2 REPLACING THE X OR Y FLEXIBLE MOTOR COUPLING


Chapter 55
REPLACING THE X DOUBLE PITCH WORM

WARNING
You MUST switch off the unit before removing the covers. ONLY an
approved Trophy technician is qualified to inspect or maintain the unit
while it is switched on and while the covers are removed. In this case, NO
unqualified person must approach the unit.

Description
The X double pitch worm enables a high precision rotation movement of the X axis of the
rotative arm.

A
1.5 mm
2.5 mm

KODAK 9000 and KODAK 9000 3D Extraoral Imaging System_Service Guide (SM712)_Ed 02 55–1
Removing the X Double Pitch Worm

Removing the X Double Pitch Worm


Before removing the X double pitch worm, make sure that:

• The unit is powered OFF.


• The unit head cover is removed.
• You have the required tool.

To remove the X double pitch worm, follow these steps:

1. Loosen, but do not remove, the fixing screw of the flexible motor coupling closest to the X
motor assembly. Push the motor coupling towards the X motor assembly to free the X
double pitch worm.

2. Unscrew and remove the front support (A).

3. Unscrew the central support (B) and remove it from the X double pitch worm.

4. Remove the faulty X double pitch worm.

5. Keep the central support (B) and the fixing screws for the new X double pitch worm.

Replacing the X Double Pitch Worm


Before replacing the X double pitch worm, make sure that you have the appropriate
component.

To replace the X double pitch worm, follow these steps:

1. Loosely mount the central support (B) on the new X double pitch worm.

2. Insert the back support (A).

3. Insert the front support (C).

4. Position the X double pitch worm with the longer tip towards the motor.

5. Position the front support (C) and loosely screw the fixing screws.

6. Position the central support (B) and loosely screw the fixing screws.

7. Align the back support (A) with the main frame and tighten the fixing screws.

8. Align the front support (C) with the main frame and tighten the fixing screws.

9. Put the flexible motor coupling in place and screw the fixing screws.

10. Manually turn the X double pitch worm towards the motor. Untighten the support (B)
fixing screws, align correctly and tighten them again. Tighten the black nut of the worm
(D) fixing screws.

11. Check the end-to-end movement and realign if necessary.

55–2 REPLACING THE X DOUBLE PITCH WORM


Testing the X Double Pitch Worm

Testing the X Double Pitch Worm


If you receive an E35 message on the PC, click Restart. This is because you have manually
moved the X double pitch worm.

From the fixed arm, on the control panel, press and hold several times for a full X
movement to check that the X double pitch worm is replaced correctly. Try also several times
by pressing and releasing to make sure that there is no mechanical blockage.

KODAK 9000 and KODAK 9000 3D Extraoral Imaging System_Service Guide (SM712)_Ed 02 55–3
Testing the X Double Pitch Worm

55–4 REPLACING THE X DOUBLE PITCH WORM


Chapter 56
REPLACING THE Y DOUBLE PITCH WORM

WARNING
You MUST switch off the unit before removing the covers. ONLY an
approved Trophy technician is qualified to inspect or maintain the unit
while it is switched on and while the covers are removed. In this case, NO
unqualified person must approach the unit.

Description
The Y double pitch worm enables a high precision rotation movement of the Y axis of the
rotative arm.

C
B

1.5 mm A
2.5 mm

Removing the Y Double Pitch Worm


Before removing the Y double pitch worm, make sure that:

• The unit is powered OFF.


• The unit head cover is removed.
• You have the required tool.

To remove the Y double pitch worm, follow these steps:

KODAK 9000 and KODAK 9000 3D Extraoral Imaging System_Service Guide (SM712)_Ed 02 56–1
Replacing the Y Double Pitch Worm

1. Loosen, but do not remove, the fixing screw of the flexible motor coupling closest to the Y
motor assembly. Push the motor coupling towards the Y motor assembly to free the Y
double pitch worm.

2. Unscrew and remove the front support (A).

3. Unscrew the central support (B) and remove it from the Y double pitch worm.

4. Remove the faulty Y double pitch worm.

5. Remove the central support (B) and keep with the fixing screws for the new Y double
pitch worm.

Replacing the Y Double Pitch Worm


Before replacing the Y double pitch worm, make sure that you have the appropriate
component.

To replace the Y double pitch worm, follow these steps:

1. Loosely mount the central support (B) on the new Y double pitch worm.

2. Insert the back support (A).

3. Insert the front support (C).

4. Position the Y double pitch worm with the longer tip towards the motor.

5. Position the front support (C) and loosely screw the fixing screws.

6. Position the central support (B) and loosely screw the fixing screws.

7. Align the back support (A) with the main frame and tighten the fixing screws.

8. Align the front support (C) with the main frame and tighten the fixing screws.

9. Put the flexible motor coupling in place and screw the fixing screws.

10. Manually turn the Y double pitch worm towards the motor. Untighten the support (B)
fixing screws, align correctly and tighten them again. Tighten the black nut of the worm
(D) fixing screws.

Testing the Y Double Pitch Worm


If you are prompted with an E35 message on the PC, click Restart. This is because you have
manually moved the Y double pitch worm.

From the fixed arm, on the control panel, press and hold several times for a full Y
movement to check that the Y double pitch worm is replaced correctly. Try also several times
by pressing and releasing to make sure that there is no mechanical blockage.

56–2 REPLACING THE Y DOUBLE PITCH WORM


Chapter 57
REPLACING THE X OR Y BEARINGS

WARNING
You MUST switch off the unit before removing the covers. ONLY an
approved Trophy technician is qualified to inspect or maintain the unit
while it is switched on and while the covers are removed. In this case, NO
unqualified person must approach the unit.

Description
The X or Y bearings enable the rotation movement of the X or Y double pitch worm.

Figure 57–1 X Bearings

1.5 mm
2.5 mm

A
B

KODAK 9000 and KODAK 9000 3D Extraoral Imaging System_Service Guide (SM712)_Ed 02 57–1
Removing the X or Y Bearings

Figure 57–2 Y Bearings

B
1.5 mm
2.5 mm

Removing the X or Y Bearings


Before removing the X or Y bearings, make sure that:

• The unit is powered OFF.


• The unit head cover is removed.
• You have the required tool.

To remove the X or Y bearings, follow these steps:

1. Loosen but do not remove, the fixing screw of the flexible motor coupling closest to the X
or Y motor assembly. Push the flexible motor coupling towards the X or Y motor
assembly to free the X or Y double pitch worm.

2. Unscrew and remove the back support (A). Push out the faulty bearing.

3. Loosen the fixing screws of the central support (B) but do not remove it from the X or Y
double pitch worm.

4. Unscrew the fixing screws of the front support (C).

5. Remove the X or Y double pitch worm with the central support (B).

6. Push out and remove the faulty bearing of the front support (C).

57–2 REPLACING THE X OR Y BEARINGS


Replacing the X or Y Bearings

Replacing the X or Y Bearings


Before replacing the X or Y bearings, make sure that you have the appropriate component.

To replace the X or Y bearings, follow these steps:

1. Insert the bearing of the back support (A).

2. Insert the bearing of the front support (C).

3. Place the front support (C) on the X or Y double pitch worm.

4. Insert the X or Y double pitch worm with the longer tip towards the motor.

5. Position the front support (C) and loosely screw the fixing screws.

6. Position the central support (B) and loosely screw the fixing screws.

7. Align the front support (C) with the main frame and tighten the fixing screws.

8. Align the central support (B) with the main frame and tighten the fixing screws.

9. Put the flexible motor coupling in place and screw the fixing screws.

10. Manually turn the X or Y double pitch worm towards the motor. Untighten the fixing
screws of the support (B) and tighten again to align correctly.

Testing the X or Y Bearings


If you receive an E35 message on the PC, click Restart. This is because you have manually
moved the double pitch worm X or Y.

From the fixed arm, on the control panel, press and hold several times for a full X or Y
movement to check that the X or Y bearings are replaced correctly. Try also several times by
pressing and releasing to make sure that there is no mechanical blockage.

KODAK 9000 and KODAK 9000 3D Extraoral Imaging System_Service Guide (SM712)_Ed 02 57–3
Testing the X or Y Bearings

57–4 REPLACING THE X OR Y BEARINGS


Chapter 58
REPLACING THE DIGITAL SENSOR

WARNING
You MUST switch off the unit before removing the covers. ONLY an
approved Trophy technician is qualified to inspect or maintain the unit
while it is switched on and while the covers are removed. In this case, NO
unqualified person must approach the unit.

Description
The digital sensor enables an x-ray to be converted into electrical signals which in turn are
converted to digital information and transmitted to the PC for x-ray imaging.

1.5 mm
2 mm
2.5 mm
x2

x3

x3
x2

Removing the Digital Sensor


Before removing the digital sensor, make sure that:

• The unit is powered OFF.


• You have the required tool.

KODAK 9000 and KODAK 9000 3D Extraoral Imaging System_Service Guide (SM712)_Ed 02 58–1
Replacing the Digital Sensor

To remove the digital sensor, follow these steps:

1. Unscrew the fixing screw of the laser knob. Pull to remove the knob.

2. Unscrew the fixing screws of the generator front cover and remove the cover.

3. Unscrew the fixing screws of the rotative arm bottom cover and remove the cover.

4. Unscrew the fixing screws of the rotative arm top cover and remove the cover.

5. Unscrew the fixing screws of the cable attachment (A).

6. Disconnect all the connectors.

IMPORTANT
Identify the position of the cables before disconnecting.

7. Unscrew the fixing screws of the sensor attachment (B).

8. Remove the faulty digital sensor.

Replacing the Digital Sensor


Before replacing the digital sensor, make sure that you have the appropriate digital sensor.

To replace the digital sensor, follow these steps:

1. Position the sensor attachment (B) and screw the fixing screws.

IMPORTANT
Make sure that the digital sensor is not tilted sideways.

2. Connect all the connectors.

3. Position the cable attach (A) and screw the fixing screws.

4. Calibrate the panoramic and 3D digital sensor in this order (see the “Calibration
Procedures” section):

• “Control of the X-Ray Beam”


• “Rotation Axis Calibration”
• “Activate the 3D Collimator”
• “Margin Calibration”

58–2 REPLACING THE DIGITAL SENSOR


Testing the Digital Sensor

5. Turn ON the unit. In the Imaging Window, click to check the correct positioning
of the panoramic digital sensor. Then, click to check the correct positioning of
the 3D digital sensor. If any error message is displayed see the “Error Message”
chapter in the Service Guide.

6. Position the rotative arm top cover and screw the fixing screws.

7. Position the rotative arm bottom cover and screw the fixing screws.

8. Position the generator front cover and screw the fixing screws.

9. Position the laser knob and screw the fixing screw.

Testing the Digital Sensor


See procedure 5 of the digital sensor replacement.

KODAK 9000 and KODAK 9000 3D Extraoral Imaging System_Service Guide (SM712)_Ed 02 58–3
Testing the Digital Sensor

58–4 REPLACING THE DIGITAL SENSOR


Chapter 59
REPLACING THE DIGITAL SENSOR MOTOR
ASSEMBLY

WARNING
You MUST switch off the unit before removing the covers. ONLY an
approved Trophy technician is qualified to inspect or maintain the unit
while it is switched on and while the covers are removed. In this case, NO
unqualified person must approach the unit.

Description
The digital sensor motor enables the automatic positioning of the digital sensor in the
panoramic or 3D position.

2 mm
2.5 mm

x3

x2

Removing the Digital Sensor Motor Assembly


Before removing the digital sensor motor assembly, make sure that:

• The unit is powered OFF.


• You have the required tool.

KODAK 9000 and KODAK 9000 3D Extraoral Imaging System_Service Guide (SM712)_Ed 02 59–1
Replacing the Digital Sensor Motor Assembly

To remove the digital sensor motor assembly, follow these steps:

1. Unscrew the fixing screw of the laser knob. Pull to remove the knob.

2. Unscrew and remove the generator front cover.

3. Unscrew and remove the rotative arm bottom cover.

4. Unscrew and remove the rotative arm top cover.

5. Disconnect the digital sensor motor assembly connector.

6. Unscrew the fixing screws of the digital sensor motor assembly.

7. Remove the faulty digital sensor motor assembly.

Replacing the Digital Sensor Motor Assembly


Before replacing the digital sensor motor assembly, make sure that you have the appropriate
component.

To replace the digital sensor motor assembly, follow these steps:

1. Position and screw the fixing screws of the digital sensor motor assembly.

2. Position and connect the digital sensor motor assembly connector.

3. Gently and manually position the digital sensor in the Panoramic and 3D position to check
that the digital sensor motor assembly is correctly positioned. Make sure there is no play
between the gears.

4. Turn ON the unit and wait for the connection. In the Imaging Window, click to
check the correct positioning of the panoramic digital sensor. Then, click to
check the correct positioning of the 3D digital sensor. If an error message is
displayed, see the “Error Message” chapter in the Service Guide.

5. Position and screw the fixing screws of the rotative arm top cover.

6. Position and screw the fixing screws of the rotative arm bottom cover.

7. Position and screw the fixing screws of the generator front cover.

8. Position and screw the fixing screw of the laser knob.

Testing the Digital Sensor Motor Assembly


See procedures 3 to 4 of the digital sensor motor assembly replacement.

59–2 REPLACING THE DIGITAL SENSOR MOTOR ASSEMBLY


Chapter 60
REPLACING THE DIGITAL SENSOR LIMIT SWITCH

WARNING
You MUST switch off the unit before removing the covers. ONLY an
approved Trophy technician is qualified to inspect or maintain the unit
while it is switched on and while the covers are removed. In this case, NO
unqualified person must approach the unit.

Description
The digital sensor limit switch restricts the digital sensor maximum rotation movement for
panoramic and 3D positioning.

2 mm
3 mm

x2

Removing the Digital Sensor Limit Switch


Before removing the digital sensor limit switch, make sure that:

• The unit is powered OFF.


• You have the required tool.

To remove the digital sensor limit switch, follow these steps:

KODAK 9000 and KODAK 9000 3D Extraoral Imaging System_Service Guide (SM712)_Ed 02 60–1
Replacing the Digital Sensor Limit Switch

1. Unscrew the fixing screw of the laser knob. Pull to remove the knob.

2. Unscrew and remove the generator front cover .

3. Unscrew and remove the rotative arm bottom cover.

4. Unscrew and remove the rotative arm top cover.

1. Unsolder and disconnect the limit switch (A). Unscrew the fixing screws.

CAUTION
Identify the orientation of the switch and the position of the cables
before removal.

2. Remove the faulty limit switch (A).

Replacing the Digital Sensor Limit Switch


Before replacing the digital sensor limit switch, make sure that you have the appropriate limit
switch.

To replace the digital sensor limit switch, follow these steps:

1. Position and screw the fixing screws of the limit switch (A).

2. Solder the limit switch (A).

CAUTION
DO NOT overheat the lugs.

3. Turn ON the unit. In the Imaging Window, click to check the correct positioning
of the panoramic digital sensor. Then, click to check the correct positioning of
the 3D digital sensor. If any error message is displayed see the “Error Message”
chapter in the Service Guide.

4. Position and screw the fixing screws of the rotative arm top cover.

5. Position and screw the fixing screws of the rotative arm bottom cover.

6. Position and screw the fixing screws of the generator front cover.

7. Position and screw the fixing screw of the laser knob.

Testing the Digital Sensor Limit Switch


See procedure 3 of the digital sensor limit switch replacement.

60–2 REPLACING THE DIGITAL SENSOR LIMIT SWITCH


Chapter 61
REPLACING THE SECTOR COLLIMATOR MOTOR
ASSEMBLY (Unit Serial N°s: xxACxxx, xxADxxx,
xxAHxxx)

WARNING
You MUST switch off the unit before removing the covers. ONLY an
approved Trophy technician is qualified to inspect or maintain the unit
while it is switched on and while the covers are removed. In this case, NO
unqualified person must approach the unit.

Description
The collimator motor enables the automatic positioning of the collimator in the panoramic or
3D position.

1.5 mm
2 mm
2.5 mm

x4

x1

Removing the Collimator Motor Assembly


Before removing the collimator motor assembly, make sure that:

• The unit is powered OFF.


• You have the required tool.

KODAK 9000 and KODAK 9000 3D Extraoral Imaging System_Service Guide (SM712)_Ed 02 61–1
Replacing the Collimator Motor Assembly

To remove the collimator motor assembly, follow these steps:

1. Unscrew the fixing screw of the laser knob. Pull to remove the knob.

2. Unscrew and remove the generator front cover.

3. Unscrew and remove the generator rear cover.

4. Disconnect the collimator motor connector from the CJ804 board.

5. Unscrew the fixing screws of the collimator motor assembly.

6. Remove the faulty collimator motor assembly.

Replacing the Collimator Motor Assembly


Before replacing the collimator motor assembly, make sure that you have the appropriate
component.

To replace the collimator motor assembly, follow these steps:

1. Position and screw, but do not tighten, the fixing screws of the collimator motor assembly.

2. Tighten the fixing screws of the collimator motor to fit the 3D collimator plate assembly
allowing a normal mechanical movement. To check that there is no play between the gears
of the collimator motor and the 3D collimator plate, gently and manually bring the
collimator to the maximum mechanical position and then release it. The gears should have
a normal mechanical movement.

3. Connect the collimator motor connector to the CJ804 board.

4. Position and screw the fixing screws of the generator rear cover.

5. Position and screw the fixing screws of the generator front cover.

6. Position and screw the fixing screw of the laser knob.

Testing the Collimator Motor Assembly


To test that the collimator motor assembly is replaced correctly, check the correct positioning
of the panoramic and 3D collimator.

To check the correct positioning of the collimators, follow these steps:

1. In the Imaging Window, click to position the panoramic collimator. In the


Panoramic Acquisition Window, check that is green. If E28 is displayed,
see the “Error Message” chapter in the Service Guide.

61–2 REPLACING THE SECTOR COLLIMATOR MOTOR ASSEMBLY (Unit Serial N°s: xxACxxx, xxADxxx, xxAHxxx)
Testing the Collimator Motor Assembly

2. Click to check the positioning of the 3D collimator. In the 3D Acquisition


Window, check that is green. If E28 is displayed, see the “Error Message”
chapter in the Service Guide.

KODAK 9000 and KODAK 9000 3D Extraoral Imaging System_Service Guide (SM712)_Ed 02 61–3
Testing the Collimator Motor Assembly

61–4 REPLACING THE SECTOR COLLIMATOR MOTOR ASSEMBLY (Unit Serial N°s: xxACxxx, xxADxxx, xxAHxxx)
Chapter 62
REPLACING THE SECTOR COLLIMATOR PLATE
ASSEMBLY (Unit Serial N°s: xxACxxx, xxADxxx,
xxAHxxx)

WARNING
You MUST switch off the unit before removing the covers. ONLY an
approved Trophy technician is qualified to inspect or maintain the unit
while it is switched on and while the covers are removed. In this case, NO
unqualified person must approach the unit.

Description
The collimator plate assembly directs the stream of x-ray beams for either the panoramic or the
3D image acquisition.

22

1.5 mm
3 mm
5.5 mm

B x2

x2

x4

x1
A

x1

KODAK 9000 and KODAK 9000 3D Extraoral Imaging System_Service Guide (SM712)_Ed 02 62–1
Removing the 3D Collimator Plate Assembly

Removing the 3D Collimator Plate Assembly


Before removing the 3D collimator plate assembly, make sure that:

• The unit is powered OFF.


• You have the required tool.

To remove the 3D collimator plate assembly, follow these steps:

1. Unscrew the fixing screw of the laser knob. Pull to remove the knob.

2. Unscrew and remove the generator front cover.

3. Disconnect the horizontal laser connector. Unscrew the fixing screws of the laser slide
(A).

4. Disconnect the 3D FOV laser connector. Unscrew the fixing screws of the laser socket and
remove.

5. Loosen the fixing screws of the collimator motor (B) and pull it slightly down to free the
collimator plate assembly.

6. Unscrew the fixing screw of the 3D collimator plate assembly.

7. Remove the faulty 3D collimator plate assembly.

Replacing the 3D Collimator Plate Assembly


Before replacing the 3D collimator plate assembly, make sure that you have the appropriate
component.

To replace the 3D collimator plate assembly, follow these steps:

1. Position the 3D collimator plate assembly and screw the fixing screw.

2. Tighten the fixing screws of the collimator motor (B) to fit the 3D collimator plate
assembly allowing a normal mechanical movement. To check that there is no play
between the gears of the collimator motor (B) and the 3D collimator plate, gently and
manually bring the collimator to the maximum mechanical position and then release it.
The gears should have a normal mechanical movement.

3. Calibrate the panoramic and 3D collimator in this order (see the “Calibration
Procedures” section):

• “Control of the X-Ray Beam”


• “Rotation Axis Calibration”
• “Activate the 3D Collimator”
• “Margin Calibration”

4. Position the 3D FOV laser socket and screw the fixing screws. Connect the connector.

62–2 REPLACING THE SECTOR COLLIMATOR PLATE ASSEMBLY (Unit Serial N°s: xxACxxx, xxADxxx, xxAHxxx)
Testing the 3D Collimator Plate Assembly

5. Position the horizontal laser slide (A) and screw the fixing screws.

6. Position the generator front cover and screw the fixing screws.

7. Position the knob and screw the fixing screw.

Testing the 3D Collimator Plate Assembly


To test that the 3D collimator plate assembly is correctly replaced, check the correct
positioning of the panoramic and 3D collimator.

To check the correct positioning of the collimators, follow these steps:

1. In the Imaging Window, click to position the panoramic collimator. In the


Panoramic Acquisition Window, check that is green. If E28 is displayed,
see the “Error Message” chapter in the Service Guide.

2. Click to check the positioning of the 3D collimator. In the 3D Acquisition


Window, check that is green. If E28 is displayed, see the “Error Message”
chapter in the Service Guide.

KODAK 9000 and KODAK 9000 3D Extraoral Imaging System_Service Guide (SM712)_Ed 02 62–3
Testing the 3D Collimator Plate Assembly

62–4 REPLACING THE SECTOR COLLIMATOR PLATE ASSEMBLY (Unit Serial N°s: xxACxxx, xxADxxx, xxAHxxx)
Chapter 63
REPLACING THE CJ666 SECTOR COLLIMATOR
DETECTION BOARD (Unit Serial N°s: xxACxxx,
xxADxxx, xxAHxxx)

WARNING
You MUST switch off the unit before removing the covers. ONLY an
approved Trophy technician is qualified to inspect or maintain the unit
while it is switched on and while the covers are removed. In this case, NO
unqualified person must approach the unit.

Description
The CJ666 collimator detection board detects the correct positioning of the panoramic and 3D
collimator.

22
1.5 mm
2.5 mm
3 mm

x2
x2
B

x1

A
x4

x2

KODAK 9000 and KODAK 9000 3D Extraoral Imaging System_Service Guide (SM712)_Ed 02 63–1
Removing the CJ666 Collimator Detection Board

Removing the CJ666 Collimator Detection Board


Before removing the CJ666 collimator detection board, make sure that:

• The unit is powered OFF.


• You have the required tool.

To remove the CJ666 collimator detection board, follow these steps:

1. Unscrew the fixing screw of the laser knob. Pull to remove the knob.

2. Unscrew and remove the generator front cover.

3. Disconnect the horizontal laser connector. Unscrew the fixing screws of the laser slide
(A).

4. Disconnect the 3D FOV laser connector. Unscrew the fixing screws of the laser socket.

5. Loosen the fixing screws of the collimator motor (B) and pull it slightly down to free the
collimator plate assembly.

6. Unscrew the fixing screw of the 3D collimator plate assembly.

7. Disconnect the connector. Unscrew the fixing screws and remove the faulty collimator
CJ666 board.

Replacing the CJ666 Collimator Detection Board


Before replacing the CJ666 collimator detection board, make sure that you have the
appropriate board.

To replace the CJ666 collimator detection board, follow these steps:

1. Position and screw the fixing screws of the collimator CJ666 board. Connect the
connector.

2. Position and screw the fixing screw of the 3D collimator plate assembly.

3. Tighten the fixing screws of the collimator motor (B) to fit the 3D collimator plate
assembly allowing a normal mechanical movement. To check that there is no play
between the gears of the collimator motor (B) and the 3D collimator plate, gently and
manually bring the collimator to the maximum mechanical position and then release it.
The gears should have a normal mechanical movement.

4. Screw the fixing screws of the laser socket. Connect the connector of the 3D FOV laser.

5. Position and screw the fixing screws of the horizontal laser slide (A). Connect the laser
connector.

6. Position and screw the fixing screws of the generator front cover.

7. Position and screw the fixing screw of the knob.

63–2 REPLACING THE CJ666 SECTOR COLLIMATOR DETECTION BOARD (Unit Serial N°s: xxACxxx, xxADxxx, xxAHxxx)
Testing the CJ666 Collimator Detection Board

Testing the CJ666 Collimator Detection Board


To test that the CJ666 collimator detection board is replaced correctly, check the correct
positioning of the panoramic and 3D collimator.

To check the correct positioning of the collimators, follow these steps:

1. In the Imaging Window, click to position the panoramic collimator. In the


Panoramic Acquisition Window, check that is green. If E28 is displayed,
see the “Error Message” chapter in the Service Guide.

2. Click to check the positioning of the 3D collimator. In the 3D Acquisition


Window, check that is green. If E28 is displayed, see the “Error Message”
chapter in the Service Guide.

KODAK 9000 and KODAK 9000 3D Extraoral Imaging System_Service Guide (SM712)_Ed 02 63–3
Testing the CJ666 Collimator Detection Board

63–4 REPLACING THE CJ666 SECTOR COLLIMATOR DETECTION BOARD (Unit Serial N°s: xxACxxx, xxADxxx, xxAHxxx)
Chapter 64
REPLACING THE BLADE COLLIMATOR (Unit Serial
N°s: xxAExxx, xxAFxxx, xxAJxxx, xxAKxxx)
WARNING
You MUST switch off the unit before removing the covers. ONLY an
approved Trophy technician is qualified to inspect or maintain the unit
while it is switched on and while the covers are removed. In this case, NO
unqualified person must approach the unit.

Description
The blade collimator directs the x-ray beam for either the panoramic, the 3D or the
cephalometric image acquisition.

2 mm

x3

Removing the Blade Collimator


Before removing the blade collimator, make sure that:

• The unit is powered OFF.


• The generator front cover is removed.
• You have the required tool.

To remove the blade collimator, follow these steps:

KODAK 9000 and KODAK 9000 3D Extraoral Imaging System_Service Guide (SM712)_Ed 02 64–1
Replacing the Blade Collimator

1. Unscrew and remove the blade collimator cover.

2. Disconnect cables 4 to 11 from the blade collimator pilot board.

3. Disconnect the FoV laser cable from the CJ691 on the “LASER FACE”.

4. Unscrew the screws.

5. Remove the faulty blade collimator.

Replacing the Blade Collimator


Before replacing the blade collimator, make sure that you have the appropriate blade
collimator.

To replace the blade collimator, follow these steps:

1. Remove the blade collimator cover.

2. Position the blade collimator with the cables on the left side of the generator. Screw the
fixing screws but do not tighten.

3. Place a spirit level on top of the blade collimator. The bottom of the collimator assembly
(A) should be parallel to the bottom of the generator (B). Tighten the fixing screws making
sure that the blade collimator is correctly positioned.

A
B

x4

64–2 REPLACING THE BLADE COLLIMATOR (Unit Serial N°s: xxAExxx, xxAFxxx, xxAJxxx, xxAKxxx)
Testing the Blade Collimator

4. Pass cables 4 to 11 through the hole above the generator. Connect all the connectors 4 to
11 with their associated J numbers on the board. For example, cable number 5 on J5 of the
board. Attach all the cables with a cable tie.

5. Connect the FoV laser cable to the CJ691 on the rotative arm on the “LASER TUBE
HEAD”.

6. Position the blade collimator cover and screw the fixing screws.

7. Calibrate the panoramic, 3D and cephalometric collimator (see Chapter Using the
Technical Menu, section Using the Calibration Tool Software):

8. Check the panoramic blade collimator calibration:

• Check the x-ray beam, see “Control of the x-ray beam”.


• Check the panoramic image using the test tools.

9. Check the 3D blade collimator calibration:

• Calibrate the margin, see “Calibrating the margin”.


• Check the 3D image using the test tools.

10. Check the Cephalometric blade collimator calibration by launching an image.

Testing the Blade Collimator


See procedures 8 through 10 of the blade collimator replacement.

KODAK 9000 and KODAK 9000 3D Extraoral Imaging System_Service Guide (SM712)_Ed 02 64–3
Testing the Blade Collimator

64–4 REPLACING THE BLADE COLLIMATOR (Unit Serial N°s: xxAExxx, xxAFxxx, xxAJxxx, xxAKxxx)
Chapter 65
REPLACING THE BLADE COLLIMATOR BOARD
(Unit Serial N°s: xxAExxx, xxAFxxx, xxAJxxx,
xxAKxxx)

WARNING
You MUST switch off the unit before removing the covers. ONLY an
approved Trophy technician is qualified to inspect or maintain the unit
while it is switched on and while the covers are removed. In this case, NO
unqualified person must approach the unit.

Description
The blade collimator pilot board enables the blades of the collimator to be positioned for the
panoramic, 3D and cephalometric mode.

x8

x4

x4

2.5 mm

KODAK 9000 and KODAK 9000 3D Extraoral Imaging System_Service Guide (SM712)_Ed 02 65–1
Removing the Blade Collimator Board

Removing the Blade Collimator Board


Before removing the blade collimator board, make sure that:

• The unit is powered OFF.


• The generator rear cover is removed.
• The generator front cover is removed.
• You have the required tool.

To remove the blade collimator board, follow these steps:

1. Disconnect all the connectors.

2. Unscrew and remove the fixing screws.

3. Remove the faulty blade collimator board and the spacers.

Replacing the Blade Collimator Board


Before replacing the blade collimator board, make sure that you have the appropriate limit
switch.

To replace the blade collimator board, follow these steps:

1. Position the spacers. Position the blade collimator board with J11 towards you on top of
the cables (A). Screw the fixing screws.

2. Connect all the connectors 4 to 11 with their associated J numbers on the board. For
example, cable number 5 on J5 of the board.

3. Make sure that the jumpers are only on J12 and J13.

4. Connect cable CJ821/J17 on J17 of the blade collimator board.

5. Connect Cable CJ821/J1 on J1 of the blade collimator board.

6. Attach all the cables with a cable tie.

65–2 REPLACING THE BLADE COLLIMATOR BOARD (Unit Serial N°s: xxAExxx, xxAFxxx, xxAJxxx, xxAKxxx)
Testing the Blade Collimator Board

Testing the Blade Collimator Board


To check that the blade collimator board is replaced correctly, you must:

• Launch a panoramic x-ray image.


• Launch a 3D x-ray image.
• Launch a cephalometric x-ray image.

DRAFTKODAK 9000 and KODAK 9000 3D Extraoral Imaging System_Service Guide (SM712)_Ed 0265–3
Testing the Blade Collimator Board

65–4 REPLACING THE BLADE COLLIMATOR BOARD (Unit Serial N°s: xxAExxx, xxAFxxx, xxAJxxx, xxAKxxx)
Chapter 66
REPLACING THE BLADE COLLIMATOR FIXING
RING (Unit Serial N°s: xxAExxx, xxAFxxx, xxAJxxx,
xxAKxxx)

WARNING
You MUST switch off the unit before removing the covers. ONLY an
approved Trophy technician is qualified to inspect or maintain the unit
while it is switched on and while the covers are removed. In this case, NO
unqualified person must approach the unit.

Description
The blade collimator fixing ring enables the blade collimator to be correctly positioned and
fixed on the x-ray generator.

2 mm

x1

x2

x3

x4

KODAK 9000 and KODAK 9000 3D Extraoral Imaging System_Service Guide (SM712)_Ed 02 66–1
Removing the Blade Collimator Fixing Ring

Removing the Blade Collimator Fixing Ring


Before removing the blade collimator fixing ring, make sure that:

• The unit is powered OFF.


• The generator front cover is removed.
• You have the required tool.

To remove the blade collimator fixing ring, follow these steps:

1. Disconnect the FoV laser cable from the CJ691 on the “LASER FACE”.

2. Unscrew and remove the blade collimator cover.

3. Unscrew the screws of the blade collimator. Remove the blade collimator with the cables
still connected to the pilot board.

4. Unscrew the screws of the fixing ring.

5. Remove the faulty fixing ring.

Replacing the Blade Collimator Fixing Ring


Before replacing the blade collimator fixing ring, make sure that you have the appropriate
blade collimator fixing ring.

To replace the blade collimator, follow these steps:

1. Position and screw the fixing screws of the blade collimator fixing ring.

2. Position the blade collimator with the cables on the left side of the generator. Screw the
fixing screws but do not tighten.

66–2 REPLACING THE BLADE COLLIMATOR FIXING RING (Unit Serial N°s: xxAExxx, xxAFxxx, xxAJxxx, xxAKxxx)
Replacing the Blade Collimator Fixing Ring

3. Place a spirit level on top of the blade collimator. The bottom of the collimator assembly
(A) should be parallel to the bottom of the generator (B). Tighten the fixing screws making
sure that the blade collimator is correctly positioned.

A
B

x4

4. Connect the FoV laser cable to the CJ691 on the rotative arm on the “LASER TUBE
HEAD”.

5. Calibrate the panoramic, 3D and cephalometric collimator (see Chapter Using the
Technical Menu, section Using the Calibration Tool Software):

6. Check the panoramic blade collimator calibration:

• Check the x-ray beam, see “Control of the x-ray beam”.


• Check the panoramic image using the test tools.

KODAK 9000 and KODAK 9000 3D Extraoral Imaging System_Service Guide (SM712)_Ed 02 66–3
Testing the Blade Collimator Fixing Ring

7. Check the 3D blade collimator calibration:

• Calibrate the margin, see “Calibrating the margin”.


• Check the 3D image using the test tools.

8. Check the Cephalometric blade collimator calibration by launching an image.

9. Position the blade collimator cover and screw the fixing screws.

Testing the Blade Collimator Fixing Ring


NA

66–4 REPLACING THE BLADE COLLIMATOR FIXING RING (Unit Serial N°s: xxAExxx, xxAFxxx, xxAJxxx, xxAKxxx)
Chapter 67
REPLACING THE X-RAY GENERATOR (Unit Serial
N°s: xxACxxx, xxADxxx, xxAHxxx)
WARNING
You MUST comply with the standards currently in force while installing
and using the radiology unit.

You MUST switch off the unit before removing the covers. ONLY an
approved Trophy technician is qualified to inspect or maintain the unit
while it is switched on and while the covers are removed. In this case, NO
unqualified person must approach the unit.

Description
The x-ray generator has the following functions:

• Supplies high-voltage circuit power to the x-ray tube.


• Enables selection of kVp, mA and exposure time.

22

1.5 mm
2.5 mm
3 mm
5 mm

x4

x1

x3
x2
A B
x4 x2

x1 x2

x4

KODAK 9000 and KODAK 9000 3D Extraoral Imaging System_Service Guide (SM712)_Ed 02 67–1
Removing the X-Ray Generator

Removing the X-Ray Generator


Before removing the x-ray generator, make sure that:

• The unit is powered OFF.


• You have the required tool.

To remove the x-ray generator, follow these steps:

1. Unscrew the fixing screw of the laser knob. Pull to remove the knob.

2. Unscrew and remove the generator front cover.

3. Disconnect the horizontal laser connector. Unscrew and remove the laser slide (A).

4. Disconnect the 3D FoV laser connector. Unscrew the fixing screws of the laser socket.

5. Loosen the fixing screws of the collimator motor (B) and pull it slightly down to free the
collimator plate assembly.

6. Unscrew and remove the 3D collimator plate assembly.

7. Disconnect the connector of the CJ666 board. Unscrew and remove the CJ666 board.

8. Unscrew and remove the generator rear cover.

9. Disconnect the collimator motor connector from the CJ804 board. Unscrew and remove
the collimator motor assembly.

10. Unscrew the fixing screws of the fixed primary collimator (B) and remove.

IMPORTANT
Make sure you identify the position and the orientation of the
primary collimator before removal.

11. Disconnect and remove the ground and power supply cables (C).

12. Unscrew and remove the faulty generator.

Replacing the X-Ray Generator


Before replacing the x-ray generator, make sure that:

• You have the appropriate x-ray generator.


• You have an ohmmeter.
• You have a DC voltmeter.

To replace the x-ray generator, follow these steps:

67–2 REPLACING THE X-RAY GENERATOR (Unit Serial N°s: xxACxxx, xxADxxx, xxAHxxx)
Testing the X-Ray Generator

1. Position the generator and screw the fixing screws.

2. Position the power supply block (C) and connect the ground cable.

3. Position the fixed primary collimator (B) in the correct orientation and screw the fixing
screws.

4. Connect the collimator motor connector to the CJ804 board.

5. Connect the CJ666 board connector and screw the fixing screws.

6. Position the collimator motor assembly and screw but do not tighten the fixing screws.

7. Position the laser socket and screw the fixing screws. Connect the connector of the 3D
FoV laser.

8. Position the 3D collimator plate assembly and screw the fixing screw.

9. Tighten the fixing screws of the collimator motor (B) to fit the 3D collimator plate
assembly allowing a normal mechanical movement. To check that there is no play
between the gears of the collimator motor (B) and the 3D collimator plate, gently and
manually bring the collimator to the maximum mechanical position and then release it.
The gears should have a normal mechanical movement.

10. Calibrate the panoramic and 3D collimator in this order (see the “Calibration
Procedures” section):

• “mA Calibration”
• “Heating Voltage Calibration”
• “Control the X-Ray Beam”
• “Rotation Axis Calibration”
• “Activatate the 3D collimator Calibration”
• “Margin Calibration”

11. Position the generator rear cover and screw the fixing screws.

12. Position the horizontal laser slide (A) and screw the fixing screws. Connect the laser
connector.

13. Position the generator front cover and screw the fixing screws.

14. Position the knob and screw the fixing screw.

Testing the X-Ray Generator


Launch an x-ray with different kV and mA combinations to make sure that the x-ray generator
replacement and calibration is correctly executed.

KODAK 9000 and KODAK 9000 3D Extraoral Imaging System_Service Guide (SM712)_Ed 02 67–3
Testing the X-Ray Generator

67–4 REPLACING THE X-RAY GENERATOR (Unit Serial N°s: xxACxxx, xxADxxx, xxAHxxx)
Chapter 68
REPLACING THE X-RAY GENERATOR (Unit Serial
N°s: xxAExxx, xxAFxxx, xxAJxxx, xxAKxxx)
WARNING
You MUST comply with the standards currently in force while installing
and using the radiology unit.

You MUST switch off the unit before removing the covers. ONLY an
approved Trophy technician is qualified to inspect or maintain the unit
while it is switched on and while the covers are removed. In this case, NO
unqualified person must approach the unit.

Description
The x-ray generator has the following functions:

• Supplies high-voltage circuit power to the x-ray tube.


• Enables selection of kVp, mA and exposure time.

x4

D x1

x1
x3

x3 x2
x2
C
B

A
2 mm
3 mm
5 mm
6 mm

KODAK 9000 and KODAK 9000 3D Extraoral Imaging System_Service Guide (SM712)_Ed 02 68–1
Removing the X-Ray Generator

Removing the X-Ray Generator


Before removing the x-ray generator, make sure that:

• The unit is powered OFF.


• The generator front cover is removed.
• The generator rear cover is removed.
• You have the required tool.

To remove the x-ray generator, follow these steps:

1. Disconnect the FoV laser cable from the CJ691 on the “LASER TUBE HEAD”.

2. Unscrew and remove the blade collimator cover (A). Unscrew the screws of the blade
collimator. Remove the blade collimator with the cables still connected to the pilot board.

3. Unscrew the screws of the fixing ring (B). Remove the fixing ring.

4. Disconnect the laser connector. Unscrew the fixing screws of the cephalometric Frankfurt
positioning laser (C). Remove the laser assembly.

5. Disconnect the horizontal positioning laser (D) connector. Unscrew the fixing screw of the
laser. Remove the laser assembly.

6. Disconnect and remove the power supply block (E) and the ground cables.

7. Unscrew and remove the faulty generator.

Replacing the X-Ray Generator


Before replacing the x-ray generator, make sure that:

• You have the appropriate x-ray generator.


• You have an ohmmeter.
• You have a DC voltmeter.

To replace the x-ray generator, follow these steps:

1. Position and screw the fixing screws of the blade collimator fixing ring.

2. Position the blade collimator with the cables on the left side of the generator. Screw the
fixing screws but do not tighten.

68–2 REPLACING THE X-RAY GENERATOR (Unit Serial N°s: xxAExxx, xxAFxxx, xxAJxxx, xxAKxxx)
Replacing the X-Ray Generator

3. Place a spirit level on top of the blade collimator. The bottom of the collimator assembly
(A) should be parallel to the bottom of the generator (B). Tighten the fixing screws making
sure that the blade collimator is correctly positioned.

A
B

x4

4. Connect the FoV laser cable to the CJ691 on the “LASER TUBE HEAD”.

5. Position the blade collimator cover and screw the fixing screws.

6. Position correctly the cephalometric Frankfurt positioning laser (C) and screw the fixing
screws.

7. Connect the laser cable 976/laser ceph to cable 972/CJ732/J28. Attach the cable with a
cable tie.

8. Position the power supply block (E) and connect the ground cable.

9. If Ceph option, adjust the cephalometric Frankfurt positioning laser beam (see Chapter
“Replacing the Cephalometric Frankfurt Positioning Laser”).

10. Adjust the horizontal positioning laser beam (see Chapter “Replacing the Horizontal
Positioning Laser”).

11. Calibrate the x-ray generator in this order (see the “Calibration Procedures” section):

• “mA Calibration”
• “Heating Voltage Calibration”

12. Calibrate the panoramic, 3D and cephalometric collimator (see Chapter Using the
Technical Menu, section Using the Calibration Tool Software):

KODAK 9000 and KODAK 9000 3D Extraoral Imaging System_Service Guide (SM712)_Ed 02 68–3
Testing the X-Ray Generator

13. Check the panoramic blade collimator calibration:

• Check the x-ray beam, see “Control of the x-ray beam”.


• Check the panoramic image using the test tools.

14. Check the 3D blade collimator calibration:

• Calibrate the margin, see “Calibrating the margin”.


• Check the 3D image using the test tools.

15. If Ceph option, check the Cephalometric blade collimator calibration by launching an
image.

16. Position the generator rear cover and screw the fixing screws.

17. Position the generator front cover and screw the fixing screws.

Testing the X-Ray Generator


Launch an x-ray with different kV and mA combinations to make sure that the x-ray generator
replacement and calibration is correctly executed.

68–4 REPLACING THE X-RAY GENERATOR (Unit Serial N°s: xxAExxx, xxAFxxx, xxAJxxx, xxAKxxx)
Chapter 69
REPLACING THE PANORAMIC LASER
ADJUSTMENT KNOB (Unit Serial N°s: xxACxxx,
xxADxxx, xxAHxxx)

WARNING
You MUST switch off the unit before removing the covers. ONLY an
approved Trophy technician is qualified to inspect or maintain the unit
while it is switched on and while the covers are removed. In this case, NO
unqualified person must approach the unit.

Description
The panoramic laser adjustment knob enables you to adjust the patient position to the
Frankfurt plane.

x1

1.5 mm

KODAK 9000 and KODAK 9000 3D Extraoral Imaging System_Service Guide (SM712)_Ed 02 69–1
Removing the Panoramic Adjustment Laser Knob

Removing the Panoramic Adjustment Laser


Knob
Before removing the panoramic adjustment laser knob, make sure that:

• The unit is powered OFF.


• You have the required tool.

To remove the panoramic adjustment laser knob, follow these steps:

1. Unscrew the fixing screw of the laser knob.

2. Pull to remove the faulty knob.

Replacing the Panoramic Adjustment Laser


Knob
Before replacing the panoramic adjustment laser knob, make sure that you have the
appropriate component.

To replace the panoramic adjustment laser knob, follow these steps:

1. Push the knob in position.

2. Screw the fixing screw.

Testing the Panoramic Adjustment Laser Knob


Move the panoramic adjustment laser knob up and down to make sure that it is correctly
replaced.

69–2 REPLACING THE PANORAMIC LASER ADJUSTMENT KNOB (Unit Serial N°s: xxACxxx, xxADxxx, xxAHxxx)
Chapter 70
REPLACING THE FIXED ARM CABLE

WARNING
You MUST switch off the unit before removing the covers. ONLY an
approved Trophy technician is qualified to inspect or maintain the unit
while it is switched on and while the covers are removed. In this case, NO
unqualified person must approach the unit.

Description
The fixed arm cable enables the connection between the CJ732 board, the CJ699 board, the
control panel and the chin rest mechanics.

KODAK 9000 and KODAK 9000 3D Extraoral Imaging System_Service Guide (SM712)_Ed 02 70–1
Removing the Fixed Arm Cable

X4 X3 X2 X1

2.5 mm CJ699 C2 C1
115V 230V 115V

X5

3 mm F4 F2
D3
C4 C3

5
A
R9
J3
K1 F3 F1 R11
R10
K2 J5
T3
125 m A / 2 50m A 3 15m A / 6 3 0mA
1
K4 K3 1
(2 30 V) / (1 15 V) (23 0 V) / (11 5V)

D1
IN VE R TER B OA R D A LIM D IGIPA N
S TATIF M OTO R

D2
C OL U M N M OTO R
A LIM C N T J12 4 J1 4 J2 4 1 5
D5 D4
J8

Q2 Q1
X7
S W TR AV EL

J10 J11
J7
J9 S W_D OWN S W_U P
U P B OA R D 1 1 1
J7 J6
1 J4 6 SW
U P/D O WN
S W_D OWN S W_U P
J6
B1 PT37 PT 1 SW2
C5 4 U2 5 M1

X25
1
J25
1
J10
1
J11
CJ732
732 V3
U2 3 U3 0
U1 4
U5 0 U5 1 U2 1 U1 7 U2 6 U2 2
X1 U1 5 J4
LD 18 C5 7
LD 31 LD 29 PT47
PT 17 R1 20
U1 0 P3
ST1 7
C2 3
PT 39 PT46
C5 0 C7 9
D6 4 PT33
LD 25
J23 E2
U3 6 PT 14 PT2
D6 3 C9 5 C2 4 C3 0 X3
PT 12 PT24
C6 6 PT 32 Y1 U3 8 U2 8 Q4
U2

B
D6 2 XU2 8
LD 24 R8 1 PT31 PT18
J22 R6 9 C4 3 DZ3
R7 4 PT22
D6 1
P2 PT 13
X 17 X 15
X5
D7 4
X 16
LD 33
C4 0

J19
J19 X 14 ST2 PT19 C6 5
R4 8 J12
X 13
D6 0 PT15
X 18 U1 1
X 12 D7
PT41 U3 5
R1 43 PT26
LD 8 U2 7
D4 7 C6 U1 9
J16 LD 9 XU2 7
PT27 BUZ1
D4 8
R1 41
LD 3 Q3 C9 3
R1 51 LD 21
LD 30
J8 D1 0 X 19 J7 PT43
LD 13 X 20
U3 7 PT 40 PT51
D5 2 PT23 PT20
L3 L2 C9 C3 1
LD 5 PT45 U4 8 U4 6
U7
D1 5
U4 0 P4 PT30 PT25 E1

J3
C3 3 J3
SW1 C9 2
J9 J15 C3 5 R5 2 P1
LD 4
Q5 C8
D1 3 C3 2 D5 U3
PT34
Q1
C2 2 Q9
L1
R1 44 J29 R1 9
LD 11 DZ1 C1 0

J27 J17
D5 0

D4 9
LD 10 U5 2 U2 9 U8
J28
U3 1
LD 17
E10
U4 4
LD 20 LD 14
E3
U5
LD 1 LD 16 LD 15

U4 K1
U1 J20

R1 42 Q6 C2 C8 0
Q8 Q2 DZ2 C8 2
M2 1
Q7 PT38
J2
J24 J26 J27 J13 J6 J18 J14 J21 J5 J1 X26 X6
1

J26
PT8
X4 P T11 P T10 P T9
PT7

x2

F D

x2
H
x4

G
H
E x2

x2 x1

Removing the Fixed Arm Cable


Before removing the fixed arm cable, make sure that:

• The unit is powered OFF.


• The unit head cover is removed.
• The column inside cover is removed.
• You have the required tool.

To remove the fixed arm cable, follow these steps:

70–2 REPLACING THE FIXED ARM CABLE


Replacing the Fixed Arm Cable

1. Disconnect the following connectors:

• J6 and J7 from the CJ699 board (A)


• J3, J19, J26 and J27 from the CJ732 board (B)

2. With an adhesive tape, attach new cable at the switch end to the old cable connectors in the
unit head.

3. Use the screwdriver handle to unclip the control panel from the hole below (C).
Disconnect all the connectors.

4. Unscrew and remove the chin rest back cover (D).

5. Unscrew and remove the chin rest bottom cover (E).

6. Unscrew and remove the chin rest base along with the chin rest base cover (F) and the chin
rest front cover (G).

7. Unsolder and disconnect the limit switches (H).

8. Pull the old cord from the switch end until the new cord with the switches appear. Remove
the adhesive tape and the old cable.

Replacing the Fixed Arm Cable


Before replacing the fixed arm cable, make sure that you have the appropriate cable kit.

To replace the fixed arm cable, follow these steps:

1. Solder and connect the limit switches (H).

CAUTION
DO NOT overheat the lugs.

2. Connect all the connectors and clip the control panel (C) back into the mount.

3. Connect the following connectors:

• J6 and J7 on the CJ699 board (A)


• J3, J19, J26 and J27 on the CJ732 board (B)

4. Turn ON the unit and test the functions of the control panel to make sure that the cable is
correctly replaced.

5. Position and screw the fixing screws of the chin rest base with the chin rest base cover (F)
and the chin rest front cover (G).

6. Position and screw the fixing screws of the chin rest bottom cover (E).

7. Position and screw the fixing screws of the chin rest back cover (D).

KODAK 9000 and KODAK 9000 3D Extraoral Imaging System_Service Guide (SM712)_Ed 02 70–3
Testing the Fixed Arm Cable

Testing the Fixed Arm Cable


Power ON the unit. Make sure that:

• All the LEDs are lit.


• The display screen displays all the parameters. If the display screen is inactive, unclip
the control panel and using the potentiometer (B), activate the display screen.
• Test the buttons to check that they function correctly.

70–4 REPLACING THE FIXED ARM CABLE


Chapter 71
REPLACING THE CONTROL PANEL

WARNING
You MUST switch off the unit before removing the covers. ONLY an
approved Trophy technician is qualified to inspect or maintain the unit
while it is switched on and while the covers are removed. In this case, NO
unqualified person must approach the unit.

Description
The control panel enables you to operate patient positioning and rotative arm adjustment
functions. It also displays error messages in case of unit malfunctioning.

KODAK 9000 and KODAK 9000 3D Extraoral Imaging System_Service Guide (SM712)_Ed 02 71–1
Removing the Control Panel

Removing the Control Panel


To remove the control panel, make sure that:

• The unit is powered OFF.


• You have the required tool.

To remove the control panel, follow these steps:

1. Use the screwdriver handle (A) to unclip the control panel from the hole below.

2. Disconnect all the connectors and remove the faulty control panel.

Replacing the Control Panel


To replace the control panel, make sure that you have appropriate component.

To replace the control panel, follow these steps:

1. Connect all the connectors.

2. Clip in the control panel.

Testing the Control Panel


Power ON the unit. Make sure that:

• All the LEDs are lit.


• The display screen displays all the parameters. If the display screen is inactive, unclip
the control panel and using the potentiometer (B), activate the display screen.
• Test all the buttons to check that they function correctly.

71–2 REPLACING THE CONTROL PANEL


Chapter 72
REPLACING THE CHIN REST BASE

WARNING
You MUST switch off the unit before removing the covers. ONLY an
approved Trophy technician is qualified to inspect or maintain the unit
while it is switched on and while the covers are removed. In this case, NO
unqualified person must approach the unit.

Description
The chin rest base holds the chin rest supports for patient positioning.

2 mm C
2.5 m m
3 mm

x2

x2
D

x4

x1 x2

KODAK 9000 and KODAK 9000 3D Extraoral Imaging System_Service Guide (SM712)_Ed 02 72–1
Removing the Chin Rest Base

Removing the Chin Rest Base


Before removing the chin rest base, make sure that:

• The unit is powered OFF.


• You have the required tool.

To remove the chin rest base, follow these steps:

1. Unscrew and pull down the chin rest bottom cover (A).

2. Push and hold the temple support adjustor (B) to unscrew the temple supports.

3. Unscrew and remove the tip of the temple support adjustor (B).

4. Remove the chin rest base cover (C).

5. Unscrew and remove the chin rest base and the chin rest front cover (D).

Replacing the Chin Rest Base


Before replacing the chin rest base, make sure that you have the appropriate component.

To replace the chin rest base, follow these steps:

1. Position and screw the fixing screws of the chin rest base and the chin rest front cover (D).

2. Replace the chin rest base cover (C).

3. Screw in the tip of the temple support adjustor (B).

4. Push and hold the temple support adjustor (B) and position the temple supports. Screw the
fixing screws of the temple supports.

5. Position and screw the fixing screws of the chin rest bottom cover (A).

Testing the Chin Rest Base


Push the temple support adjustor to see if the temple supports open easily. Release the temple
support adjustor to check if the temple supports close automatically.

72–2 REPLACING THE CHIN REST BASE


Chapter 73
REPLACING THE CHIN REST MOTOR

WARNING
You MUST switch off the unit before removing the covers. ONLY an
approved Trophy technician is qualified to inspect or maintain the unit
while it is switched on and while the covers are removed. In this case, NO
unqualified person must approach the unit.

Description
The chin rest motor enables the up and down movement of the chin rest.

E
x4

x1
A
C x3

x2
6

2.5 m m
3 mm
B

KODAK 9000 and KODAK 9000 3D Extraoral Imaging System_Service Guide (SM712)_Ed 02 73–1
Removing the Chin Rest Motor

Removing the Chin Rest Motor


Before removing the chin rest motor, make sure that:

• The unit is powered OFF.


• You have the required tool.

To remove the chin rest motor, follow these steps:

1. Unscrew and remove the chin rest bottom cover (A).

2. Remove the pulley belt (B).

3. Unscrew and remove the pulley (C).

4. Disconnect the connectors and remove the motor (D).

5. Unscrew and remove the motor fixing component (E).

Replacing the Chin Rest Motor


Before replacing the chin rest motor, make sure that you have the appropriate component.

To replace the chin rest motor, follow these steps:

1. Position and screw the fixing screws of the motor fixing component (E).

2. Position the motor and connect the connectors (D).

3. Position and screw the fixing screw of the pulley (C).

4. Position the pulley belt (B).

5. Position and screw the fixing screws of the chin rest bottom cover (A).

Testing the Chin Rest Motor

From the fixed arm, on the control Panel, press and hold to raise and lower the chin
rest to the maximum to check that the chin rest motor is replaced correctly and functions
correctly.

73–2 REPLACING THE CHIN REST MOTOR


Chapter 74
REPLACING THE CHIN REST MECHANIC

WARNING
You MUST switch off the unit before removing the covers. ONLY an
approved Trophy technician is qualified to inspect or maintain the unit
while it is switched on and while the covers are removed. In this case, NO
unqualified person must approach the unit.

Description
The chin rest mechanic enables the up and down movement of the chin rest.

2.5 m m
3 mm
A

x2

D
E
C

G
F
x2
x4

F
x4
B
x2

x2 x1

KODAK 9000 and KODAK 9000 3D Extraoral Imaging System_Service Guide (SM712)_Ed 02 74–1
Removing the Chin Rest Mechanic

Removing the Chin Rest Mechanic


Before removing the chin rest mechanic, make sure that:

• The unit is powered OFF.


• You have the required tool.

To remove the mechanic, follow these steps:

1. Unscrew and remove the chin rest back cover (A).

2. Unscrew and remove the chin rest bottom cover (B).

3. Unscrew and remove the chin rest base with the chin rest base cover (C) and the chin rest
front cover (D).

4. Disconnect the J1 cable (E) from the CJ804 board.

5. Unsolder and disconnect the limit switches (F).

6. Unscrew and remove the complete chin rest mechanic (G).

Replacing the Chin Rest Mechanic


Before replacing the chin rest mechanic, make sure that you have the appropriate component.

To replace the chin rest mechanic, follow these steps:

1. Position and screw the fixing screws of the chin rest mechanic (G).

2. Connect and solder the limit switches (F).

3. Connect the J1 cable (E) on the CJ804 board.

4. Turn the unit ON. On the control Panel, press and hold to raise and lower the chin
rest to the maximum to check that the limit switches and the connectors are correctly
replaced.

5. Position and screw the fixing screws of the chin rest base with the chin rest base cover (C)
and the chin rest front cover (D).

6. Position and screw the fixing screws of the chin rest bottom cover (B).

7. Position and screw the fixing screws of the chin rest back cover (A).

Testing the Chin Rest Mechanic


See procedure 4 of the chin rest limit mechanic replacement

74–2 REPLACING THE CHIN REST MECHANIC


Chapter 75
REPLACING THE CHIN REST LIMIT SWITCHES

WARNING
You MUST switch off the unit before removing the covers. ONLY an
approved Trophy technician is qualified to inspect or maintain the unit
while it is switched on and while the covers are removed. In this case, NO
unqualified person must approach the unit.

Description
The chin rest limit switches limit the chin rest maximum up and down movement.

2.5 m m
3 mm
A

x2

E
x4
x2

B
x2
x2 x1 E

Removing the Chin Rest Limit Switches


Before removing the chin rest limit switches, make sure that:

• The unit is powered OFF.


• You have the required tool.

To remove the chin rest limit switches, follow these steps:

KODAK 9000 and KODAK 9000 3D Extraoral Imaging System_Service Guide (SM712)_Ed 02 75–1
Replacing the Chin Rest Limit Switches

1. Unscrew and remove the chin rest back cover (A).

2. Unscrew and remove the chin rest bottom cover (B).

3. Unscrew and remove the chin rest base with the chin rest base cover (C) and the chin rest
front cover (D).

1. Unsolder and disconnect the limit switches (E). Unscrew the fixing screws.

CAUTION
Identify the orientation of the switch and the position of the cables
before removal.

2. Remove the faulty limit switches (E).

Replacing the Chin Rest Limit Switches


Before replacing the chin rest limit switches, make sure that you have the appropriate limit
switches.

To replace the chin rest limit switches, follow these steps:

1. Position and screw the fixing screws of the limit switches (E).

2. Solder the limit switches (E).

CAUTION
DO NOT overheat the lugs.

3. Turn ON the unit. From the fixed arm, on the control panel, press and hold to raise
and lower the chin rest to the maximum position to check that the limit switches are
correctly replaced.

4. Turn OFF the unit.

5. Position and screw the fixing screws of the chin rest base with the chin rest base cover (C)
and the chin rest front cover (D).

6. Position and screw the fixing screws of the chin rest bottom cover (B).

7. Position and screw the fixing screws of the chin rest back cover (A).

Testing the Chin Rest Limit Switches


See procedure 3 of the chin rest limit switches replacement.

75–2 REPLACING THE CHIN REST LIMIT SWITCHES


Chapter 76
REPLACING THE HAND GRIPS

WARNING
You MUST switch off the unit before removing the covers. ONLY an
approved Trophy technician is qualified to inspect or maintain the unit
while it is switched on and while the covers are removed. In this case, NO
unqualified person must approach the unit.

Description
The hand grips enables the patient to stand in the correct position.

12

x2

Removing the Hand Grips


Before removing the hand grips, make sure you have the right tools.

• The unit is powered OFF.


• You have the required tool.

KODAK 9000 and KODAK 9000 3D Extraoral Imaging System_Service Guide (SM712)_Ed 02 76–1
Replacing the Hand Grips

To remove the hand grips, follow these steps:

1. Use the back of a screwdriver (A) to unclip the control panel from the hole below.

2. Unscrew the screws and remove the faulty hand grips

Replacing the Hand Grips


Before replacing the hand grips, make sure you have the appropriate component.

To replace the hand grips, follow these steps:

1. Screw in the hand grips.

2. Clip in the control panel

Testing the Hand Grips


NA

76–2 REPLACING THE HAND GRIPS


Chapter 77
REPLACING THE TEMPLE SUPPORT

WARNING
You MUST switch off the unit before removing the covers. ONLY an
approved Trophy technician is qualified to inspect or maintain the unit
while it is switched on and while the covers are removed. In this case, NO
unqualified person must approach the unit.

Description
The temple support retains the head in the correct position for acquisition.

2 mm

x2

x2
A

Removing the Temple Support


Before removing the temple support, make sure that:

• The unit is powered OFF.


• You have the required tool.

To remove the temple support, follow these steps:

1. Push and hold the temple support adjustor (A) to unscrew the temple supports.

2. Remove the faulty temple support.

KODAK 9000 and KODAK 9000 3D Extraoral Imaging System_Service Guide (SM712)_Ed 02 77–1
Replacing the Temple Support

Replacing the Temple Support


Before replacing the temple support, make sure that you have the appropriate component.

To replace the temple support, follow these steps:

1. Push and hold the temple support adjustor (A).

2. Screw in the temple supports one at a time.

Testing the Temple Support


NA

77–2 REPLACING THE TEMPLE SUPPORT

Вам также может понравиться