Академический Документы
Профессиональный Документы
Культура Документы
Biorex Diagnostics Limited, Unit 2C Antrim Technology Park, Antrim, BT41 1QS (United Kingdom)
Tel: +44 (0) 2894 468786 | Fax: +44 (0) 2894 469933 | Website: www.biorexdiagnostics.com BXC0431 – LDL CHOLESTEROL| Revision No.10 AUG/11 | Page 1 of 2
Calculation: Limitations - Interference: Determination of LDL-Cholesterol in Serum Based on Precipitation
A sample x Calibrator conc. = LDL conc. Criterion: Recovery within 10% of initial value. of LDL with Dextran Sulfate. Ärztl. Lab. 1985;31 :325-330.
A Calibrator Icterus: No significant interference up to an index I of 50 (approximate 2. Bablok W. et al. A General Regression Procedure for Method
50 mg/dl bilirubin) Transformation. J Clin Chem Clin Biochem 1988;26:783-790.
Manual Procedure: End Point: Haemolysis: No significant interference up to an index H of 500 3. Bachorik P.S., Ross J.W. National cholesterol education program
Wavelength Temperature Cuvette Measurement (approximate haemoglobin concentration: 500 mg/dl). recommendations for measurement of low-density lipoprotein
Hg 600 nm Reagent Blank Lipemia (Intralipid): No significant interference up to an index L of 750. cholesterol: executive summary. Clin Chem 1995;41 :1414-1420.
(Side (one reagent No significant interference from native triglycerides up to 1500 mg/dl. 4. Cohn J.S., Mc Namara J.R., Schaefer E.J.. Lipoprotein Cholesterol
37oC 1 cm light path No significant interference from HDL, VLD, or chylomicrons. Concentrations in the Plasma of Human Subjects as Measured in
wavelength blank per
700 nm) series only) In rare cases, elevated immunoglobulin concentrations can lead to the Fed and Fasted States. Clin Chem 1988;34:2456-2459.
falsely elevated LDL-cholesterol results. 5. Cremer F., Nagel D., Mann H. et al. Ten-year follow-up results from
Pipette into test tubes as follows: Abnormal liver function does affect lipid metabolism; consequently the Goettingen Risk, Incidence and Prevalence Study (GR(PS). 1.
Reagent Blank Sample / Calibrator HDL and LDL results are of limited diagnostic value Risk factors for myocardial infarction in a cohort of 5790 men.
Sample / Calibrator --- 5 µl Atherosclerosis 1997;129:221-230.
R1 450 µl 450 µl Normal Values:
Mix well and incubate at: 37°C for 5 minutes.
Levels in terms of risk for coronary heart disease: BIOSYSTEMS A25 / A15 APPLICATION:
R2 150 µl 150 µl Adult Levels
Incubate at 37°C for 5 minutes. Measure the absorbance of the Recommended < 130 mg/dl
Calibrator and the sample against the reagent blank. (desirable) (<3.37 mmol/l) A 25 A 15
130-159 mg/dl TEST NAME LDL LDL
Moderate Risk
LDL Calculation: (3.37-4.12 mmol/l) ANALYSIS MODE BIR DIFF BIR DIFF
Absorbance sample x Calibrator conc. = LDL conc. ≥ 160 mg/dl SAMPLE TYPE SERUM SERUM
Absorbance Calibrator Risk
(≥ 4.14 mmol/l) UNITS mg/dl mg/dl
Recommended values according to the GRIPS study: GENERAL
REACTION TYPE INCREASING INCREASING
Measuring/reportable range: mg/dl mmol/l
1 - 400 mg/dl (0.02 – 10.4 mmol/l) DECIMALS 1 1
For patients with
Determine samples with LDL-cholesterol concentration> 400 mg/dl via 145 3,8 manifested coronary
NO.OF REPLICATES 1 1
the rerun function. On instruments without rerun function, manually heart disease TEST NAME IN PT REP
dilute the samples with 0.9% NaCI or distilled/deionized water (e.g. 1 + For patients having PROCEDURE READING MONOCHR MONOCHR
9). Multiply the result by the appropriate dilution factor (e.g. factor 10). 170 4,4 one or more risk VOLUMES SAMPLE 3 3
factors REAGENT 1 300 300
Automated Analysis:
For persons exhibiting REAGENT 2 100 100
no risk factors and WASHING 1.2 1.2
This reagent is suitable for all Automated systems. Applications for
200 5,2 without manifested
automated systems can be requested from our technical department. PREDIL FACTOR
coronary heart
POST DIL FACTOR 2 2
disease
Imprecision: FILTERS MAIN 600 600
Each laboratory should investigate the transferability of the expected
Reproducibility was determined using controls. The following results REFERENCE
values to its own patient population and if necessary determine its
were obtained: TIMES READING 1 285s 168s
own reference range.
Intra Assay – Within run
For diagnostic purposes the LDL-cholesterol results should always be READING 2 600s 600s
MW SD CV assayed in conjunction with the patient's medical history, clinical
Sample REAGENT 2 300s 192s
mg/dl mg/dl % examinations and other findings. CALIBRATION TYPE MULTIPLE MULTIPLE
Control serum 1 89.6 0.39 0.43
Control serum 2 120.2 0.63 0.54 CALIBRATOR REPLICATES 3 3
Quality Control: CALIBRATION
Control serum 3 152.9 0.81 0.53 BLANK REPLICATES 3 3
Lipid Controls BXC0330A / BXC0316 A
The control intervals and limits must be adapted to the individual CALIBRATION CURVE
Inter Assay – Between Run laboratory and country-specific requirements. Values obtained should BLANK ABS LIMIT 0.15 0.15
MW SD CV fall within established limits. Each laboratory should establish corrective OPTIONS KINETIC BLANK LIMIT
Sample
mg/dl mg/dl % measures to be taken if values fall outside the limits. LINEARITY LIMIT 400 400
Control serum 1 89.6 0.45 0.51
Control serum 2 120.2 0.65 0.54 Health & Safety:
Control serum 3 152.9 0.51 0.33 This kit is designed for use by suitably qualified laboratory personnel
only. Exercise the normal precautions required for the handling of
Method comparison: laboratory reagents. Do not ingest the material. Dispose of material
A comparison of the Biorex LDL-D (y) with a commercial obtainable according to local guidelines.
assay (x) gave following result (mg/dl):
y = 0.97 x + 5.12; r = 0.983
References:
1. Armstrong V., Seidel D. Evaluation of a Commercial Kit for the
Biorex Diagnostics Limited, Unit 2C Antrim Technology Park, Antrim, BT41 1QS (United Kingdom)
Tel: +44 (0) 2894 468786 | Fax: +44 (0) 2894 469933 | Website: www.biorexdiagnostics.com BXC0431-LDL CHOLESTEROL | Revision No.10 AUG/11 | Page 2 of 2