Вы находитесь на странице: 1из 35

CTA R1.

1 Page 1

RECOMMENDED CT CONNECTIVITY CONFIGURATION

Version(s):
All models of PC platform 4, 8, & 16 slice scanners – v1.6 or later
All models of AQ32 & AQ64 scanners – v2.04 or later

Background …
Based on many customers’ input and industry standard, this document is intended to “standardize”
Toshiba CT scanner (PC platform only) connectivity configuration. Although they are not required,
these steps are strongly recommended and will solve many commonly reported connectivity issues.

Note: If there is any discrepancy, always go by NASU published document.

Recommend to implement following configurations on all CT scanners …

DICOM Connect Tool Page 3


Frequently Used Devices (For AQ32/64 v2.04 or later only) Page 10
Send “^” As The Delimiter Between Name Components Page 12
Private Tag Configuration Page 15
Study Description Tag (0008,1030) Page 17
Series Description Tag (0008,103E) Page 23

If additional DICOM options are installed, below procedures are recommended …

Query / Retrieve SCU service Page 26


Query / Retrieve SCP service Page 29
Modality Worklist Mgmt & Modality Performed Procedure Step SCU services Page 30
• Study Number in Modality Worklist Mgmt Service Page 34

National Solution Support Toshiba America Medical Systems, Inc.


CTA R1.1 Page 2

Definitions

Below is the easy-to-understand explanation of commonly seen DICOM definitions. Please refer to
DICOM v3.0 technical documentation published by NEMA for the official definition of below key terms.

Terms Description
1 Service Class User (SCU) Refers to the device that initiates a DICOM connection.
2 Service Class Provider (SCP) Refers to the device that fulfills the DICOM request.
3 Store Service A DICOM service that can archive or receive images in DICOM
format.
Store SCU A DICOM store device that “sends” images in DICOM
Store SCP A DICOM store device that “receives” images in DICOM
4 Print SCU Service A DICOM service that sends images to DICOM imagers in DICOM
format.
5 Query/Retrieve Service A DICOM service that allows querying and retrieving images
between SCP and SCU devices
Query/Retrieve SCU Technologists sit on the modality to perform query and retrieve
images stored on the image archive device.
Query/Retrieve SCP Technologists sit on the image archive device to perform query
and retrieve images stored on the modality.
6 Modality Worklist SCU Service The modality performs query to RIS via RIS Broker and retrieves
all matching work orders.
7 Modality Performed Procedure The modality sends a message to RIS via RIS Broker about the
Step SCU Service status of the selected work order, i.e In Progress or Completion,
so that RIS can remove it upon completion and automatically
begin the billing and other processes.
8 Storage Commitment SCU Immediately after storing images, the modality sends a message
Service to the image archive requesting the confirmation of each image
just sent. Once the confirmation message is received from the
image archive, the modality can safely delete images from its local
storage folder.

National Solution Support Toshiba America Medical Systems, Inc.


CTA R1.1 Page 3

DICOM CONNECT TOOL

Complaint(s):
N/A

Version(s):
All models of PC platform 4, 8, & 16 slice scanners – v1.6 or later
All models of AQ32 & AQ64 scanners – v2.04 or later

Resolution(s):
Below is the procedure in configuring various DICOM options available on CT A-series scanners:
1) Store SCU
2) Print SCU
3) Storage SCP
4) Query/Retrieve SCU
5) Query/Retrieve SCP
6) Modality Worklist SCU (MWM) & Performed Procedure Step SCU (MPPS)

1) Install appropriate DICOM license(s) on the scanner.


Refer to the appropriate installation manual provided by Nasu for detail procedure.

Verify that the license is installed and working by clicking UTILITY => ENGINEERING =>
OPTION LICENSE => LIST and the License List window opens.

Note: Click UTILITY => OPTION => LICENSE LIST will also open the License List window.

Scan console

Description License Name Status


Store SCU STRG_SCU ON
Print SCU PRINT_SCU ON
Modality Worklist SCU MWM_SCU ON
Modality Performed Procedure Step SCU PPS_SCU ON

Display console

Description License Name Status


Store SCU STRG_SCU ON
Print SCU PRINT_SCU ON
Storage SCP STRG_SCP ON
Query/Retrieve SCU QR_SCU ON
Query/Retrieve SCP QR_SCP ON

National Solution Support Toshiba America Medical Systems, Inc.


CTA R1.1 Page 4

Perform following procedures on the Scan and/or Display consoles based on the installed
DICOM services as specified above.

2) Click UTILITY => ENGINEERING => CONSOLE to open a command prompt window.

3) At the prompt, execute: notepad <space> c:/winnt/system32/drivers/etc/hosts <enter>

4) At the end of this hosts file, enter IP address and hostname (separated at least by one space) of
the DICOM node.

Note: Make sure to press <ENTER> key after each entry, including the last entry on the list.

5) When finishing, click FILE => SAVE to save the change.

6) Click FILE => EXIT to exit and close this notepad program.

7) Execute: exit <enter> to close this command prompt window.

For 4, 8, and 16 slice scanners (PC platform), skip to Step #17.


For AQ32/64 scanners, continue steps below.

8) Click UTILITY => ENGINEERING => DICOM CONNECT TOOL to bring up “DICOM Nodes
Settings” window.

* This figure is taken from AQ64 v2.04

National Solution Support Toshiba America Medical Systems, Inc.


CTA R1.1 Page 5

Button Description
ADD Allow adding another DICOM node to this list
MODIFY Allow to edit the setting of the selected DICOM node on this list
DELETE Allow deleting one existing DICOM node from this list
SAVE Save the change(s)
EXIT Exit this DICOM Nodes Settings window
INITIALIZE Remove all entries from this list and reset back to the factory default setting.

Note: The order of entries entered in this DICOM Connect Tool window is the order that will
appear within CT application. The first 10 nodes in this window of the Display console are
reserved for auto-archiving. One can use UTILITY => ENGINEERING => FREQUENTLY USED
DEVICE on both Scan and Display consoles to rearrange the order later if necessary.

9) Click ADD or MODIFY button and Add/Edit Application Entities window opens.

* This figure is taken from AQ64 v2.04

Button Description
Mnemonic Easy to recognized name appeared on CT application
AE Title Configured device’s AE title
Hostname Hostname of this configured device as entered in
c:/winnt/system32/drivers/etc/hosts file.
Note: Do not enter actual IP address of configured device here.
Port Number Configured device’s port number.
Services Supported Display a list of supported DICOM services.
Syntaxes Supported Display a list of supported transfer syntaxes for the selected
service.
Edit Services Add/remove/modify the supported DICOM services for this node.
Edit Transfer Add/remove/modify the supported transfer syntax of selected
Syntaxes DICOM services in Services Supported window.

National Solution Support Toshiba America Medical Systems, Inc.


CTA R1.1 Page 6

OK Save the change made here and close this window


Cancel Discard the change made here and close this window.
Connect Check Perform C-ECHO check to this configured device.

10) Click EDIT SERVICES button to add DICOM services under Service Supported (right) column
of Edit Services window.

* This figure is taken from AQ64 v2.04

Description DICOM Services Supported Services


Storage SCU STRG_SCU Standard_Echo
Standard_CT
Standard_Sec_Capture
Print SCU PRINT_SCU Basic_Grayscale_Print_Management
Modality Worklist Mgmt MODALIT_SCU Modality_Worklist_Find
Modality Performed Procedure MPPS_SCU Performed_Procedure_Step
Step Performed_Procedure_Step_Retr
Storage SCP STRG_SCP Storage_SCP
Query/Retrieve SCU QR_SCU Study_Root_QR_Find
Study_Root_QR_Move
Query/Retrieve SCP QR_SCP <No need to add any service>

11) Click OK button to close this Edit Services window.

12) In Add/Edit Application Entities window, click each service in Services Supported (left) column
and make sure only IMPLICIT_LITTLE_ENDIAN appears under Syntaxes Supported (right)
column.

Note: For VA installation, add EXPLICIT_BIG_ENDIAN under EDIT TRANSFER SYNTAXES


button.

13) Click OK button to close Add/Edit Application Entities window.

National Solution Support Toshiba America Medical Systems, Inc.


CTA R1.1 Page 7

14) Click SAVE button to save the change.

15) Click EXIT button to close DICOM Nodes Settings window.

16) Reboot the entire system.

Below procedure is for 4, 8, and 16 slice scanners (PC platform) only …

17) Click UTILITY => ENGINEERING => DICOM CONNECT TOOL to bring up “DICOM Connect
Tool” window.

* This figure is taken from AQ16 PC v1.6

Note: The first 10 nodes entered in this DICOM Connect Tool window of the Display console are
reserved for auto-archiving. The order of entries entered in this window is the order that will
appear within CT application. To change this order, all entries in this window must be deleted
and re-created in the desirable order.

18) Click HOST => ADD or HOST => MODIFY button and Add/Edit Application Entities window
opens.

* This figure is taken from AQ16 PC v1.6

National Solution Support Toshiba America Medical Systems, Inc.


CTA R1.1 Page 8

Button Description
Edit Service … Allow to add and remove DICOM service(s)
Edit Transfer … Allow to add and remove transfer syntax of a selected DICOM
service
Host Name Hostname of this configured device as entered in
c:/winnt/system32/drivers/etc/hosts file.
Note: Do not enter actual IP address of configured device here.
Nemonic Easy to recognized name shown in CT application
Application Title Configured device’s AE title
Port No. Configured device’s port number
OK Click this button to save changes made in this screen
Cancel Click this button to exit this screen without saving any change

19) Click EDIT SERVICE … button to open Edit Service window and add DICOM services from
“Services List” to “Added Services” column.

* This figure is taken from AQ16 PC v1.6

Description DICOM Services Supported Services


Storage SCU STRG_SCU None Verification
STR_SCP CT Image Storage
STR_SCP MR Image Storage
Print SCU PRINT_SCU PRINT_SCP Basic Grayscale Print
Modality Worklist Mgmt MODALIT_SCU MWM_SCP Modality Worklist Management
Modality Performed MPPS_SCU PPS_SCU Modality Performed Procedure Step
Procedure Step PPS_SCU MPPS Retrieve
Storage SCP STRG_SCP STR_SCU
Query/Retrieve SCU QR_SCU QR_SCP Study Root Query/Retrieve – Find
QR_SCP Study Root Query/Retrieve – Move
Query/Retrieve SCP QR_SCP QR_SCU

20) Click OK button to save changes and close this Edit Services window.

21) In Hosts window, make sure each DICOM service has the transfer syntax name:
IMPLICIT_LITTLE_ENDIAN.

Note: For VA installation, add EXPLICIT_BIG_ENDIAN under EDIT TRANSFER … button.

National Solution Support Toshiba America Medical Systems, Inc.


CTA R1.1 Page 9

22) Click OK button to close this Hosts window.

23) Click FILE => SAVE button to save the change. When prompting to overwrite the file, click OK
button to proceed.

* This figure is taken from AQ16 PC v1.6

24) Click QUIT => QUIT to exit this DICOM Connect Tool window.

25) Reboot the entire system.

National Solution Support Toshiba America Medical Systems, Inc.


CTA R1.1 Page 10

FREQUENTLY USED DEVICE

Complaint(s):
Users request most frequently archive devices to display at the beginning of Destination Host window
under Image Directory => Transfer

* This figure is taken from AQ64 v2.04ER001

Or display in the particular order under auto-archive list window (Only the first 10 DICOM archive nodes
configured under DICOM Connect Tools in Display console will appear in this list).

* This figure is taken from AQ64 v2.04ER001

Version(s):
All models of AQ32 & AQ64 scanners – v2.04 or later

National Solution Support Toshiba America Medical Systems, Inc.


CTA R1.1 Page 11

Resolution(s):

This procedure applies to both Scan and Display consoles.

Auto archive list, although appears on the Scan console, is actually pulled from the Display
console.

Procedure …

1) Click UTILITY => ENGINEERING => FREQUENTLY USED DEVICE to open the File Media
Tool window.

2) Move each entry from the left to right column in the proper order by clicking each archived node.

* Above figure is taken from AQ64 v2.04ER001

3) When finish, click OK button to save the change and exit this window.

4) Click QUIT button to close Engineering Utility window.

National Solution Support Toshiba America Medical Systems, Inc.


CTA R1.1 Page 12

SEND “ ^ “ AS THE DELIMITER BETWEEN NAME COMPONENTS

Complaint(s):
A space is used as the delimiter between name components in DICOM tag (0010,0010). As the result,
the entire patient name appears in the last name field of a Storage SCP device. DICOM v3.0 specifies
to use “ ^ “ as the delimiter between name components.

Version(s):
All versions of CT A-series scanners

Resolution(s):
By enabling this function, only the patient name, DICOM tag (0010,0010), will be send with “ ^ “
between each of 5 name components as shown below …

[last name]^[first name]^[middle name]^[prefix]^[suffix]

Example:
SMITH^JOHN^J^MR^MD
BROWN^MARY^M^MS^MD

Procedure #1:

Do below steps on both scan and display consoles if it is a multislice scanner

Repeat this procedure when adding a new Storage SCP node in DICOM Connect Tool
or WinDcmSetup

1) Click UTILITY => ENGINEERING => CONSOLE to open a command prompt window.

If this is a Unix platform scanner, go to step 4).


If this is a PC platform scanner, go to next step.

2) For PC platform scanners, execute sh <space> -L <enter>

3) For PC platform scanners, execute x: <enter>

4) Dending on the scanner platform (Unix or PC), execute below command under the root
privilege:

For all Unix systems:


/toshiba/system_util/dcm_setup/DcmSetRemoteLang <enter>

National Solution Support Toshiba America Medical Systems, Inc.


CTA R1.1 Page 13

For all PC base systems:


/toshiba/system_util/dcm_setup/DcmSetRemoteLang.exe <enter>

Below text menu will appear on the command prompt window:

===== DICOM Language Table =====


1: Dump Lang. Table
2: Change Lang. Table
q: Quit
Input >>

5) At “input>>” prompt, enter 1 to dump the entire table.

This table contains a list of Storage SCP devices that this CT scanner console is configured to
archive images to.

Below is an example of the text output that will appear on the screen:
-------------+------------------+--------+-------------------------------------
Hostname | AE-Title | Send ^ | Supported Charactor Set(s)
-------------+------------------+--------+-------------------------------------
PACS | EFILM | OFF | [ISO_IR 6]
-------------+------------------+--------+-------------------------------------
EFILMR | EFILMR | ON | [ISO_IR 6]

Note: Depending on the software version, “Send ^” column might not appear on the screen.

6) For each node in this table, do …


a. At the prompt, enter 2 to change the language table.
b. Enter the hostname of a Storage SCP device as appear on the above table.
c. Enter 1 for ISO IR6
d. Enter 2 for ISO IR100
e. Enter q to quit this submenu
f. Enter y for yes when prompting to send “ ^ “ character.

Repeat above steps for the next node until all nodes have the value of ON in “Send ^”
column in the above table.

7) Enter q to quit this DICOM Language Table menu.

8) Execute: exit <exit> to exit from the shell.

9) Execute: exit <exit> to close this command prompt window and reboot the system.

National Solution Support Toshiba America Medical Systems, Inc.


CTA R1.1 Page 14

Procedure #2:

Do below steps on both scan and display consoles if it is a multislice scanner.

This procedure is taken from FSM-CT3204 released on Jan. 28, 2003.

1) Click UTILITY => ENGINEERING => SCHEDULER TOOL => TAG SETUP =>
MWMtoSCHEDULER1 to open the Tag Setup window.

2) Scroll down the Tag-code List to see if 00111080 entry exists.

If yes,
Click that entry and confirm the value of 00111080 appears in the Reference Code and
Output Code #1 fields below.

* This figure is taken from AQ64 v2.04

If no,
a) In the Reference Code and Output Code #1 fields, enter 00111080
b) Click ADD button to add this entry.

3) Click OK button to close this Tag-code List window

4) Click QUIT button to close Patient Reserve Preference window.

5) Click QUIT button to close Engineer Utility window.

6) Reboot the system.

National Solution Support Toshiba America Medical Systems, Inc.


CTA R1.1 Page 15

PRIVATE TAG CONFIGURATION

Complaint(s):
N/A

Version(s):
All models of PC platform CT scanners
All models of Unix platform CT scanners

Resolution(s):
The CT scanner, by default, is set to “ OFF “ in DICOM Private Table menu for each of DICOM network
nodes that the system is archiving images to. Unless there is a specific reason to set this value to “ ON
“, the recommended setting is “ OFF “.

Note: Please contact CT NSS before changing this setting to “ ON “.

Below is the procedure …

Do this procedure on both Scan and Display consoles if it is a multislice scanner.

1) Click UTILITIES => ENGINEERING => CONSOLE to open a command prompt window.

If this is a Unix platform scanner, go to step 4).


If this is a PC platform scanner, go to next step.

2) For PC platform scanners, execute sh <space> -L <enter>

3) For PC platform scanners, execute x: <enter>

4) Execute: /toshiba/system_util/dcm_setup/DcmSetPrivate.exe <enter>

A DICOM Private Table menu is open on the screen.

5) Enter 1 to dump the entire table with a list of devices this CT scanner is archiving images to.
-------------+------------------+-------------------
hostname | AE-Title | Set Private ?
-------------+------------------+-------------------
EFILMR | EFILMR | Not Set
-------------+------------------+-------------------
ADVT | ADVT | Not Set

* Above figure is taken from AQ64 v2.04

6) All values under “ Set Private ?” column should be “ Not Set “.

National Solution Support Toshiba America Medical Systems, Inc.


CTA R1.1 Page 16

If there is “ Set “ value under this column, do following steps. Otherwise, you are done.

7) Enter 2 <enter> to change private tag setting.

8) Enter hostname as appeared under the first column of Dump Private Table.

9) Enter n <enter> for no when prompting to send “private tag”.


===== DICOM Private Table =====
1: Dump Private Table
2: Change Private Table
q: Quit
Input >> 2
hostname >> EFILMR
Set Private ? <Y/N> >> n

* Above figure is taken from AQ64 v2.04

Note: Repeat Step #7 to #9 for next network node as necessary.

10) Enter q <enter> to quit this DICOM Private Table menu.

11) Execute: exit <exit> to exit from the shell.

12) Execute: exit <exit> to close this command prompt window.

National Solution Support Toshiba America Medical Systems, Inc.


CTA R1.1 Page 17

STUDY DESCRIPTION TAG (0008,1030)

Complaint(s):
Technologists first manually or use Modality Worklist function to register a patient. After the image
acquisition and / or post processing, images are archived to a Storage SCP device but the study
description tag (0008,1030) is blank.

Version(s):
All models of PC platform 4, 8, & 16 slice scanners – v1.6 or later
All models of AQ32 & AQ64 scanners – v2.04 or later

Resolution(s):
RIS Broker can send the description of a work order or study in any valid DICOM Modality Worklist tag
as defined in DICOM v3.0. However, in most scenarios, RIS Broker will send this information in one or
both of following tags:

Scheduled Procedure Step Description (0040,0007)


Requested Procedure Description (0032,1060)

If this data is not stored in either of these two DICOM tags, please contact CT NSS for further support.

Below is the procedure to solve this complaint:

Procedure #1

Do this on the scan console to display multiple tabs configuration after clicking DETAIL button in
Patient Registration screen.

1) Click UTILITY => ENGINEERING => SCHEDULER TOOL => INITIAL SETTING => NEXT

National Solution Support Toshiba America Medical Systems, Inc.


CTA R1.1 Page 18

2) In the Detail Display Level entry, check MAIN/CE/STUDY/PATIENT TAG

* Above figure is taken from AQ64 v2.04ER001.

3) Click OK button to save this change.

4) Click QUIT button to close Patient Reserve Preference window.

5) Click QUIT button to close Engineer Utility window.

6) Reboot the entire system.

7) After the reboot, click DETAIL button in the Patient Registration screen, and there should be 4
or 5 tabs appeared to the right of it.

* This figure is taken from AQ64 v2.04ER001.

National Solution Support Toshiba America Medical Systems, Inc.


CTA R1.1 Page 19

* This figure is taken from AQ16 v1.6ER002

Procedure #2

Do this on the Scan console to ensure the scanner pulls proper data from RIS Broker.

1) From Patient Registration screen, click APPOINTMENT => FROM SERVER => OPTION.

2) On the Get Item Select List (left) column in Option window, scroll down and ensure both
Scheduled Procedure Step Description and Requested Procedure Description tags are
highlight. If not, click the entry to highlight it.

* This figure is taken from AQ64 v2.04ER001.

National Solution Support Toshiba America Medical Systems, Inc.


CTA R1.1 Page 20

3) Click SAVE button to save this change.

4) Click OK button to close this Option window. Appointment window will refresh to display the
latest scheduled work orders pulled from RIS Broker.

Procedure #3

Do this on Scan console to map the data from either Scheduled Procedure Step Description or
Requested Procedure Description tag to Study Description tag (0008,1030).

After performing Procedure #4 below, if the Study Description field in the Patient Registration
screen is still blank, then re-do the entire Procedure #3 but replace 00321060 with 00400007.

1) Click UTILITY => ENGINEERING => SCHEDULER TOOL => TAG SETUP =>
MWMtoSCHEDULER1

2) In the Tag-Code List window, scroll down to see if there is an entry for 00321060.
If yes,
a) Click that entry to highlight it.
b) In the Reference Code and Output Code #1 fields, make sure they have a value of
00321060.
c) In the Output Code #2 field, make sure they have a value of 00081030.

If no,
a) In the Reference Code and Output Code #1 fields, enter 00321060.
b) In the Output Code #2 field, enter 00081030.

* Above figure is taken from AQ64 v2.04ER001.

National Solution Support Toshiba America Medical Systems, Inc.


CTA R1.1 Page 21

3) Click ADD button to add this entry or MODIFY button to save the change.

4) Click OK button to close this Tag Setup window.

5) Click QUIT button to close this Patient Reserve Preference window.

6) Click QUIT button to close this Engineer Utility window.

7) Reboot the entire system.

Procedure #4

Do this on Scan console to ensure changes done on Procedure #3 is working. If not working,
repeat Procedure #3 but replace 00321060 with 00400007.

1) From the Patient Registration screen, click APPOINTMENT => FROM SERVER => RELOAD to
pull the latest matching work orders from RIS Broker.

2) Click one of these work orders in the Appointment window and click OK button to import it into
the Patient Registration screen.

3) From the Patient Registration screen, click DETAIL => PATIENT tab. The description of the
selected work order should now appear in the dedicated Study Description field.

Procedure #5

Do this on Scan console to display the description of each work order in the Appointment
window.

1) Click UTILITY => ENGINEERING => CONSOLE

2) Execute: sh <space> -L <enter>

3) Execute: x: <enter>

4) Execute: DEFtool.ksh <enter>

National Solution Support Toshiba America Medical Systems, Inc.


CTA R1.1 Page 22

5) Enter 3 <enter> for Display Data

* This figure is taken from AQ64 v2.04.

6) Enter 1 <enter> for ADD

In the next step, use whichever DICOM tag number (00321060 or 00400007) that is working as
shown in Procedure #4 above.

7) Enter 00321060 <enter> or 00400007 <enter> when prompting to enter tag number.

8) Enter 30 <enter> when prompting to enter the field length.

9) Enter LO <enter> when prompting to enter VR.

10) Enter Study Description <enter> when prompting to enter the tag name.

11) Enter n <enter> for no when prompting to add another dispdata entry.

* This figure is taken from AQ64 v2.04..

Note: Use # 5 for MOVE under Dispdata Edit Menu to move this tag to a different location within
the Appointment window for ease of display purpose.

12) Enter 0 (zero) to quit Dispdata Edit Menu.

13) Enter 0 (zero) to quit DICOM MWM Define Set Tool.

14) Execute: exit <enter> to exit from the shell.

15) Execute: exit <enter> to close this command prompt window.

Note: To see this new change, users must close the Appointment window and re-open again.

National Solution Support Toshiba America Medical Systems, Inc.


CTA R1.1 Page 23

SERIES DESCRIPTION TAG (0008,103E)

Complaint(s):
The customer requests a customized proper description of each series archived to DICOM Storage
SCP devices. Some of advantages with this request are:
1) SCP devices can automatically do hanging protocol and other tasks.
2) It enables physicians to know what images are stored in each series without loading them on
the screen.

CT scanners’ specific requirements to meet this request are as follow …


1) Data must be sent in (0008,103E) or series description tag.
2) Users must be able to enter their own description of each series.
3) This customized description must be saved with the protocol setting under Exam Preset.
4) If the system generates its own series description, this value must be consistent.

Version(s):
All models of PC platform 4, 8, & 16 slice scanners – v1.6 or later
All models of PC platform AQ32 & AQ64 scanners – v2.04 or later

Resolution(s):
The scanner now provides an ability to place multiple fields stored with an image in the series
description tag (0008,103E) as long as the total length in this tag is less than 64 characters, including
spaces and all other none-alphanumeric characters.

The scanner, by default, places following tags in this series description tag:
0x00204000 Image Comment
0x00211016 Series Comment
0x00180010 Cntrast agent
0x00181210 Convolution Kernel or Filter Setting
0x7ff1110a Filter setting
0x00211003

Although this configuration works, it needs to fine tune slightly to meet the needs of most US
customers. Below is the recommended configuration …

IMPORTANT: Please make sure following DICOM dictionary files are from v2.04 or later of
AQ32/64 or v1.6 of all PC platform scanners. Do not restore these files from the site specific
backup. If there are other special mappings in these files, except series description mapping,
manually reapply those mappings again after completing this procedure.

1) /ct/KBProj/develop/toshiba/program_data/etc/dicom/NEMAtoDICOM/NEMAtoDICOM.CT
2) /ct/KBProj/develop/toshiba/program_data/etc/dicom/NEMAtoDICOM/NEMAtoDICOM.SC
3) /ct/KBProj/develop/toshiba/program_data/etc/dicom/NEMAtoDICOM/NEMAtoDICOM.W1.CT
4) /ct/KBProj/develop/toshiba/program_data/etc/dicom/NEMAtoDICOM/NEMAtoDICOM.W3.CT
5) /ct/KBProj/develop/toshiba/program_data/etc/dicom/NEMAtoDICOM/NEMAtoDICOM.TriggerTime.CT

National Solution Support Toshiba America Medical Systems, Inc.


CTA R1.1 Page 24

Do this procedure on both Scan and Display consoles if it is a multislice scanner.

1) Click UTILITIES => ENGINEERING => CONSOLE to open a command prompt window.

2) Execute sh <space> -L <enter>

3) Execute x: <enter>

4) Execute: notepad <space>


/ct/KBProj/develop/toshiba/program_data/etc/dicom/NEMAtoDICOM/0x0008103E <enter>

5) At the bottom of this file, make following changes …

Before

TAGS:0x00204000,0x00211016,0x00180010,0x00181210,0x7ff1110a,0x00211003
DELIMITER:/
REPLACE:,

After

TAGS:0x00204000,0x00211016
DELIMITER:/
REPLACE:SPACE

Note: Do not use “ \ “ (back slash) and “, “ (comma) in both Delimiter and Replace fields above.
They are invalid characters in Study Description (00081,103E) tag and CT applications.

Note: Character specified in the Delimiter field is used as the separator between fields specified
in the Tag field above.

Note: Character specified in the Replace field is used as the separator within a multi-value field
specified in the Tag field above.

6) Click FILE => SAVE to save this change.

7) Click FILE => EXIT to exit notepad application.

8) From the command prompt, execute exit <enter> to exit the shell.

9) From the command prompt, execute exit <enter> to close this command prompt window

10) Reboot the entire system.

National Solution Support Toshiba America Medical Systems, Inc.


CTA R1.1 Page 25

Optional method

This method only works on the Display console of AQ32/64 v2.04 or later scanner …

1) Click UTILITY => SETUP UTILITY => SERIES DESCRIPTIN TOOL

* Above figure is taken from AQ64 v2.04ER001

2) Click ADD and/or DEL buttons in the middle to rearrange fields so that only Image Comment and
Series Comment fields appeared on the SetTag (left) column as shown above.

Note: Fields displayed under SeeTag (right) column are preconfigured at the factory. If there are
fields that customers would like to place in the series description tag but not available to select from
here, please contact CT NSS to get the procedure.

3) Ensure Delimiter field at the bottom has a value of “ / “.

4) Ensure Replace field at the bottom has a value of “ ; “ (semi-colon).

5) Click OK button to save this change and close this window.

6) Reboot the entire system.

National Solution Support Toshiba America Medical Systems, Inc.


CTA R1.1 Page 26

QUERY / RETRIEVE SCU SERVICE

Complaint(s):
N/A

Version(s):
All models of PC platform 4, 8, & 16 slice scanners – v1.6 or later
All models of PC platform AQ32 & AQ64 scanners – v2.04 or later

Resolution(s):

Toshiba only guarantee that this DICOM function will work with Toshiba CT images.
Other vendors’ CT images might work but no guarantee.

Perform this configuration only on Display console.

1) Implement 6 recommended procedures as described on the first page of this document.

2) Provide Display console DICOM configuration information to Query / Retrieve SCP devices (ex:
PACS) where images are stored.

Display console default settings are:

AQ32/64 v2.04 …
Local AE Title: TM_CT_CMW_V3.00 (This is configurable)
Port Number: 2700
IP Address: <Vary depending on the system configuration>

4, 8, 16 Slice Scanners (PC platform) v1.6 …


Local AE Title: TM_CT_CMW_V2.00 (This is configurable)
Port Number: 2700
IP Address: <Vary depending on the system configuration>

National Solution Support Toshiba America Medical Systems, Inc.


CTA R1.1 Page 27

To test this function …

1) Click the appropriate icon (see below in red circle) to bring up Study Directory window.

* This figure is taken from AQ64 v2.04

* This figure is taken from AQ16 PC v1.6ER002

2) Click Source Host … icon to bring up Source Host window.

* This figure is taken from AQ64 v2.04

3) In Source Host window, click an appropriate DICOM node (ex: PACS) and click OK button

Note: The exact description of icons and their order in this window varies depending on the setup.

* This figure is taken from AQ16 PC v1.6ER002

National Solution Support Toshiba America Medical Systems, Inc.


CTA R1.1 Page 28

4) In this “Setting DICOM Query / Retrieve Find Key” window, enter the appropriate search key(s). “ * “
wildcard search is acceptable.

* This figure is taken from AQ16 PC v1.6ER002

5) Click OK button to begin the query process. A list of matching studies should appear in the Study
Directory window.

Note that not every field in this Study Directory window will populate with the data.

6) Select the appropriate study to retrieve at the study level

7) Click TRANSFER (in AQ32/64) or COPY (for all other PC platform CT scanners) button.

8) In the Destination Host window, click “Anet Images” icon (exact description is configurable) and
then OK button. Display console now begins the retrieval process.

9) When the process is complete, ensure that study is displayable on Display console.

National Solution Support Toshiba America Medical Systems, Inc.


CTA R1.1 Page 29

QUERY / RETRIEVE SCP SERVICE

Complaint(s):
N/A

Version(s):
All models of PC platform 4, 8, & 16 slice scanners – v1.6 or later
All models of PC platform AQ32 & AQ64 scanners – v2.04 or later

Resolution(s):

Perform this configuration only on Display console.

1) Implement 6 recommended procedures as described on the first page of this document.

2) Provide Display console DICOM configuration information to Query / Retrieve SCU devices (ex:
Vitrea2) that will query and retrieve images stored on Display console.

Display console default settings are:

AQ32/64 v2.04 …
Local AE Title: TM_CT_CMW_V3.00 (This is configurable)
Port Number: 2700
IP Address: <Vary depending on the system configuration>

4, 8, 16 Slice Scanners (PC platform) v1.6 …


Local AE Title: TM_CT_CMW_V2.00 (This is configurable)
Port Number: 2700
IP Address: <Vary depending on the system configuration>

To test this function …

On the Query/Retrieve SCU device, perform query and retrieve processes to Display console. (The
exact procedure varies depending on the type of DICOM workstations or PACS).

Ensure the entire study is displayable on that workstation or PACS.

National Solution Support Toshiba America Medical Systems, Inc.


CTA R1.1 Page 30

MODALITY WORKLIST MGMT / MODALITY PERFORMED PROCEDURE STEP SCU


SERVICES

Complaint(s):
N/A

Version(s):
All models of PC platform 4, 8, & 16 slice scanners – v1.6 or later
All models of PC platform AQ32 & AQ64 scanners – v2.04 or later

Resolution(s):

Modality Worklist Management (MWM) and Modality Performed Procedure Step (MPPS) services are
two separate but related DICOM services. For these two services to work on the scanner, RIS and RIS
Broker must support these functions as well.

The scanner can have only MWM function, which enables it to pull all scheduled work orders from RIS
Broker. If MPPS function is also supported, then the scanner can automatically inform RIS that a work
order is complete after finishing the image acquisition.

Procedure #1

Note: The default setting under Getdata list should be appropriate in most situations. Please consult
with CT NSS before making further changes.

1) Implement 6 recommended procedures as described on the first page of this document.

2) Click UTILITY => ENGINEERING => CONSOLE to open a command prompt window.

3) Execute: sh <space> -L <enter>

4) Execute: x: <enter>

5) Execute: DEFtool.ksh <enter>

6) In “DICOM MWM Define Tool Set” menu, enter 1 for Getdata.

7) In “GETDATA EDIT MENU” menu, enter 7 for Modify MWM Server Information.

8) Enter 1 when prompting to enter an entry to modify.

9) Enter the hostname of MWM server as configured in DICOM Connect Tool window.

10) Enter 2 when prompting to enter an entry to modify

National Solution Support Toshiba America Medical Systems, Inc.


CTA R1.1 Page 31

11) Enter COMPARE <enter>


Note:
COMPARE – The scanner compares the last and this retrieve data and only update the
difference in its local database.
STARTUP – The scanner will automatically perform MWM query to RIS after the system starts
up.
CYCLIC.nnn – The scanner will automatically perform MWM query to RIS after “nnn” time
interval in second.

12) Enter 0 (zero) to quit

13) Enter 0 (zero) to quit “GETDATA EDIT MENU” menu.

14) Enter 2 for Keydata.

Note: The default setting for key data is Modality = CT. Do not modify or delete this tag.

15) In “KEYDATA EDIT MENU” menu, enter 1 to add a tag in the matching key list.

16) Enter the item number (1st column in this list) of “>Scheduled Procedure Step Start Date(SPSS)”
entry.

17) Enter Today <enter> when prompting to enter the matching key data

18) Enter n <enter> for no when prompting to enter another keydata.

19) Enter 0 (zero) to quit

20) Enter 0 (zero) to quit the “DICOM MWM Define Tool Set” menu.

21) Execute: exit <enter> to exit shell.

22) Execute: exit <enter> to close this command prompt window.

Procedure #2
1) Click UTILITY => ENGINEERING => SCHEDULER TOOL => INITIAL SETTING to open
InitialSetting(1) window.

2) In the Accession Number field, enter “ – “ (dash). This signal the scanner to pull the data from RIS
Broker when using the worklist function. If this field is blank on the retrieved work order or
registering the patient via EMERGENCY button, then the scanner will leave this field blank.

3) In the Study ID field, enter the appropriate starting study number. The scanner should increment
this number by 1 when registering a patient for the scan.

National Solution Support Toshiba America Medical Systems, Inc.


CTA R1.1 Page 32

* This figure is taken from AQ64 v2.04

4) In RIS Connection field, check DICOM MWM option.

5) In MWM List Select field, check Multiple Selection option.

* This figure is taken from AQ64 v2.04

6) If MPPS is set to ON, then in DICOM MPPS Transfer field, check Completed When An
Examination Is Complete option. If MPPS is set to OFF, then check None.

7) If MPPS is set to ON, then in MPPS Server field, enter the hostname of RIS Broker as configured in
the DICOM Connect Tool window.

* This figure is taken from AQ64 v2.04

8) In MWM Reloading Key Data field, check Input Window option. This enables users to query RIS
Broker by a specific field in Appointment window.

* This figure is taken from AQ64 v2.04

9) Click NEXT button at the bottom of this window

10) In Sequence Create Data field, check Scheduled Protocol option.

* This figure is taken from AQ64 v2.04

11) Click OK button to close this window.

12) Click QUIT button to close Patient Reserve Preference window.

13) Click QUIT button to close Engineering Utility window.

14) Reboot the system.

National Solution Support Toshiba America Medical Systems, Inc.


CTA R1.1 Page 33

How Modality Performed Procedure Step Works …


After registering the patient in Patient Registration screen, when technologists click a protocol to scan
in Scan Plan window,
1) The scanner will send “MPPS In Progress” message to RIS.

2) When technologists click NEXT PATIENT button in the Scan Plan window to complete the scan,
the scanner will send “MPPS Complete” message to RIS.

With this recommended MPPS configuration, the completion of a select work order means the image
acquisition of that work order is complete. There could be other post-processing jobs that need to be
done on Display console, which is not part of this MPPS completion message.

National Solution Support Toshiba America Medical Systems, Inc.


CTA R1.1 Page 34

STUDY NUMBER IN MODALITY WORKLIST MGMT SERVICE

Complaint(s):
Customers request that the study number of each study should be assigned sequentially by the
scanner rather than a value assigned by RIS via Modality Worklist Management service

Version(s):
All models of PC platform 4, 8, & 16 slice scanners – v1.6 or later
All models of PC platform AQ32 & AQ64 scanners – v2.04 or later

Resolution(s):

Perform this procedure on Scan console only.

1) Click APPOINTMENT => FROM SERVER => OPTION to open Patient Information window.

3) Scroll down “Get Item Select List” (left column) window to locate Requested Procedure ID.

4) Click this entry to remove the highlight.

5) Click SAVE button to save this change.

6) Click OK button to exit this window. The scanner will now query RIS Broker again to update the list.

7) Click UTILITY => ENGINEERING => SCHEDULER TOOL => INITIAL SETTING to open
InitialSetting(1) window.

National Solution Support Toshiba America Medical Systems, Inc.


CTA R1.1 Page 35

8) In the Study ID field, enter the starting study ID number.

9) Click OK button to save the change and close InitialSetting(1) window.

10) Click QUIT button to close Patient Reserve Preference window.

11) Click QUIT button to close Engineering Utility window.

12) Reboot the system.

Note: When importing a work order into Patient Registration screen from Appointment window, this
Study Number field will remain blank. The scanner will automatically fill in the appropriate study number
after the image acquisition is complete.

National Solution Support Toshiba America Medical Systems, Inc.

Вам также может понравиться