Seminarscript
Version 3.0_GB
Overview Seminar workflow
1. „Preparation“
2
2. „Introduction
Introduction“
3. B. Braun Presentation
4. Quality assurance
5. General Info’s/risks of CI
6. System Overview
7. Service manual
8. Disassembly
9. Assembly
10
10. Service program and Calibration
11. Checking
12. Documentation und Certifying
Disassembly/Assembly
1
1.3 Checklist Special-/Calibration tools Preparation
Disposables
Calibration/Checking/Others
• PC
• Interface line (PC connection Interface line)
• Service software
• Service plug
• Allen key 3mm (mech. pressure cut off)
• Measuring cylinder 25ml
• Stop watch
• Patient leakage current test adapter
• Manometer (pressure gauge)
• etc.
2
6. System Overview
6.1 In general
6 2 The Infusomat Space
6.2
6.3 Handling overview
6.4 Powering on
6.5 Start of Infusion
6.6 Customized Options
6.7 Activating functions
6.8 Alarm
6.9 Pump principle
3
6.1 In general System Overview
68 mm
68 mm
m
Weight of:
Perfusor = approx. 1,4kg
Infusomat = 1,4 kg
Infusomat Space
p
Perfusor Space
Perfusor- and Infusomat Space are offering exactly the same handling structure.
4
6.1 In general System Overview
release button
5
6.1 In general System Overview
Pole clamp
Position 1:
Position 2:
Power Supply
Integrated
handle
12 V
power supply
6
6.1 In general System Overview
Power Supply
pp y
Power Supply
• Input:
Voltage 100….240V~
/50-60 Hz/ 550 mA
7
6.1 In general System Overview
Euro-Plug UK-Plug
AUS-Plug
US-Plug
8
6.1 In general System Overview
Mech. Concept Infusomat Space (continuing)
Space PCA-Kit:
1 1 On demand Button
1.
2. 1 PCA-syringe holder lock
3. 1 Fixation band
4. 1 Fixation clip
Order data:
• Art.-Nr. 8713554
9
6.2 The Infusomat® Space System Overview
6.2.1 Back
Cover of battery
Port for drop sensor compartment
P3
P3 P2
P2
P3 Socket for
P2 Socket for 12 V,
Guide rails connection with
Interface, PCA,
Space Control
nurse call
10
6.2.1 Back The Infusomat® Space
Th drip
The d i sensor
• Drip chamber adapter replaceable
depending on drip chamber size
• Infrared light pulse with about 1 A current
• Pulse frequency 1000 Hz
• Wide angle optic
• Shielding for a better EMC protection
(see production stamp)
Th crank
The k for
f emergency opening
i (only)
( l )
In every case turn the crank clockwise (only) like shown at the crank opening
(left side when looking at the front)
11
6.2.1 Back The Infusomat® Space
Battery
connector (P1)
Device feet
Battery pack
Battery pack
• Recharging time 6 h
• Infusomat and Perfusor are using the same battery, this minimizes
risk of confusion
12
6.2.1 Back The Infusomat® Space
After battery is fully charged it switches from charging current to “trickle” charging
t1 t2 t3 t
Type
plate
l t
Glued Seal
13
6.2.3 The Front The Infusomat® Space
Emergency door
opening slot
Pump-based anti-
free-flow
free flow clamp Upstream sensor
Line guide
cover
Release
mechanism of Anti-
Free-flow clamp
Downstream
pressure sensor
Air sensor incl.
Temp. sensor
14
Air Alarm
Infusomat SP S Infusomat SP P
15
6.2.3 The Front The Infusomat®Space
Infusomat SP S Infusomat SP P
Door Jam
Door Jam
in combination
with the older
version of
extension hook
made of plastic
16
6.2.3 The Front The Infusomat®Space
17
6.2.3 The Front The Infusomat®Space
Start-/Stop key
Graphic display
Communication key
Cursor for navigation and (PC ,Space Control or
value inputs OK key to confirm barcode data receipt
inputs
18
6.3 Handling Overview System Overview
Anti-Freeflow-
Anti Freeflow Fixation hook of the
clamp Set related line pump segment
19
6.3 Handling Overview System Overview
Pl
Please watch
t h th
the
respective Instruction
for use !
20
6.4 Powering on System Overview
- Battery status
- Last therapy
P
Press “O /Off” switch
“On/Off” it h on unit
it and
d note
t the
th automatic
t ti self
lf check:
h k
21
6.5 Start of Infusion System Overview
• Observable things
22
6.6 Customized Options System Overview
Customized
Options
23
6.6.2 Data Lock Customized Options
Customized
Options
Occlusion
pressure Data Lock
Data Lock Prevention against non-authorized
Date manipulations.
Clock
Alarm Volume
Patient Data
Bolus Rate
Macro Mode
Help texts on
Level 1:
No changes of values possible, stop, start, turning on/-off allowed
Level 2:
Same as Level 1, additionally no change of the disposable possible.
24
6.6.3 Date & Clock Customized Options
Customized
Options
Occlusion
pressure Date & Clock
Data Lock For documentation reasons and the
Date History Protocol.
Automatic adjustment if connected to
Clock
SpaceStation together with SpaceCom.
Alarm Volume
Patient Data
Bolus Rate
Macro Mode
Help texts on
Customized
Options
Occlusion
pressure Alarm Volume
The loudness of the audible alarm signal can be
Data Lock adjusted to the environmental conditions in
Date the hospital
Clock
Alarm Volume
Patient Data
Bolus Rate
Macro Mode
Help texts on
25
6.6.5 Patient Data Customized Options
Customized
Options
Occlusion
pressure Patient Data
Data Lock are a safety feature which makes it possible to treat
the right patient with the right drug
Date
Clock
Alarm Volume
Patient Data
Bolus Rate
Macro Mode
Help texts on
Customized
Options
Occlusion
pressure Bolus Rate
The max.bolus rate can be limited for safety reasons
Data Lock If the bolus rate is set to zero the bolus
Date function is disabled.
Clock
Alarm Volume
Patient Data
Bolus Rate
Macro Mode
Help texts on
26
6.6.8 Macro Mode Customized Options
Customized
Options
Occlusion
pressure Macro Mode
Individual adjustment of the display content.
Data Lock The user can decide if the delivery rate shall be
Date screened in big numbers or smaller.
Clock
Alarm Volume
Patient Data
Bolus Rate
Macro Mode
Help texts on
Customized
Options
27
6.7 Activating functions System Overview
Overview:
- Bolus mode („Others“)
- Prime („Rates“)
28
6.8 Alarm System Overview
1
1. Air alarm
2. Temperature alarm
•Disposables
•etc.
29
6.9 Pump principle (general) System overview
The Infusomat SP is developed out of the Perfusor SP
and gear Cµ
P
30
6.9 Pump principle (Infusomat SP) System overview
Acoustic User interface
Acoustic alarm via Alarm-/Signal-LEDs
Watchdog (Matrix)
Power supply alarm via
with monitoring
Loudspeaker
piezzo buzzer
(>65dBa) Graphic display
(<65dBa)
CµP
seriell
Functional- synchron Control- Real time clock
Processor processor FµP & CµP FµP CµP
(M32) (H8S)
32kHz
Stepper motor- Movement sensor
output stage (2 light barriers) FµP
12MHz 7,37MHz
Front cover
• Downstream sensor detection of an occlusion at the outlet side to avoid a tube rupture
(overpressure)
• Air and temperature sensor for detection of air bubbles inside the tubing (rupture or an
empty reservoir) and compensation of temperature depending tube parameters.
31
6.9 Pump principle System overview
In order to support the user and to achieve a reproducible closing of the
pump cover the front cover will
b closed
be l d motor
t ddriven.
i FµP & CµP FµP CµP
Front cover
Schließriegel
Cover motor
FµP DC-Motor- FµP
(DC-Motor)
Linear potentiometer Output
CµP and gear
The position of the looking bolt will be detected and controlled via a linear potentiometer.
The circuit and the control equals the function of the syringe detection at the syringe pump.
FµP
Pumpenkopf
Safety clamp Pumpenkopf
Sensor Peristaltic
Downstream (Stepper
e stamotor)
Peristaltic
tc
Sensor (Stepper motor
Tube
Front cover
Locking bolt
Cover motor
FµP DC-Motor- FµP
(DC-Motor)
Linear potentiometer Output
CµP and gear
32
6.9 Pump principle System overview
Occlusions test
FµP FµP
Pumpenkopf
Safety clamp
Pumpenkopf
Peristaltic
sensor
Down stream (Stepper motor) Upstream
Peristaltic
Sensor Sensor
(Stepper motor)
Tube
Front cover
Locking bolt
Cover motor
FµP DC-Motor- FµP
(DC-Motor)
Linear potentiometer Output
CµP and gear
33
7. Service manual
7 1 Overview
7.1
7.2 Specialties
7.3 Features
7.4 Updates
Chapter 0 to 10
Chapter 0 = Important preliminary remark
• Addresses
34
7.2 Specialties Service manual
• Amendment service
• Index/ directory
• Real pictures
• Exploded drawing
• etc.
35
7.4 Updates Service manual
686 C 030001
type of device = Infusomat SP
group
subgroup
PFA E xxxxxx
= Infusomat SP
next sw group
Device SW Code
Infusomat Space S 686………
Infusomat Space P 687………
Perfusor
f S
Space 688………
Space Control 689……….
Space Station 690……… No version display
36
8. Disassembly
8 1 Required tools
8.1
37
8.2.1 Demo/exercise phase A Disassembly
• Remove battery
38
User Interface Disassembly
39
User Interface (Downstream, Air detector) Disassembly
Possible failures:
40
8.2.2 Demo/ exercise phase B Disassembly
Front
41
8.2.2 Demo/ exercise phase B Disassembly
Frame:
•Locking bolt
•Pump Chassis
remove two fixing clips
press three springs (one unit)
remove spring holder wing
Springs now removable
9. Assembly
9 1 The procedure
9.1
9.2 Caution
42
9.1 The procedure Assembly
43
9.2 Caution Assembly
44
9.2.3 Cable routing Caution
45
9.2.5 Housing Caution
When putting upper housing and lower housing together be aware of the following:
46
9.2.5 Housing (Loudspeaker) Caution
47
10. Service Program and Calibration
10.1 In General
48
10.1 In general - CAN driver Service Prog./Calib.
On Service CD V3.5 located in the tool folder, you will find CAN driver 4.14.
This driver is compatible with WIN 7
The driver installation routine will detect an existing driver on your PC and will stop the
installation
You need to uninstall the old CAN driver first before you install the new one
The old (big) CAN Box is able to work with the latest driver
(see next slide with pictures of old/new CAN Box)
49
10.1 In general - CAN driver Service Prog./Calib.
50
10.3 The Calibration Service Prog./Calib.
The only thing which is going to be calibrated (state March 05) is the locking bolt.
51
10.4 Opp. inside Service Prog. Service Prog./Calib.
10.4.1.Disposables
10.4.3 Tests
Inside „Disposables“ you have the opportunity to create your own table of disposables,
adapted to your
you special
spec a needs/situation
eeds/s tuat o in you
your hospital!
osp ta
The procedure is similar to the one described in the Perfusor Space Script.
52
10.4.2 Modification Data Others in Service Prog
Documentation of customizations
With HiBaSeD 5.0 the Modification
Data can be printed out in a clear design
Within HiBaSeD 6.1.0 instruction manual a pump default setting list can be seen
and /or printed out for documentation of customized changes.
53
10.4.3 Tests Others in Service Prog..
A test will be started with “Initializing” the device (with device switched on!).
You can choose the History viewer in the star up window of the Service Program
The Password to get the complete access to the History viewer (incl
(incl. „Service
Service Editor“)
Editor )
is “sp3”
54
10.4.5 Clone Data - Upload Service Prog./Calib.
5. By pressing „Start“ button (full screen mode) the file will be uploaded
55
10.4.5 Clone Data - Drug file download Service Prog./Calib.
Caution:
For the SW Update please always read and carefully
follow the respective „Instruction for SW Update”
(see Software IFU J04)
56
10.5 Software Update – Before update Service Prog./Calib.
Hint:
During the update process the user specific settings (like “Option”,
Option , limitations of infusion
rates and so on) are going to be reset to the default setting.
Affected settings are Modification data and the disposable articles as well as the drug
library (the following procedure is not valid for the drug library).
How to reactivate the previous saved user specific settings (after the SW update):
2. Open the matching HiBaSeD service program (usually a new, higher version) and
chose your pump within HiBaSeD
3. Take the previous saved hard copy of the Modification data (and/or the disposable
article data) and modify all respective settings.
5. Now the file can be taken to upload this settings to all other Space pumps which should
get the same settings (see slide 10.5.5 Clone Data - Upload)
57
10.5 Software Update Service Prog./Calib.
4
4. “F”
F 4
4… RS 232 or CAN 30
3.0
5. “G” 5… RS 232 or CAN 3.0
6. “H” 6… RS 232 or CAN 5.0
7. “J” 7…. RS 232 or CAN SpOS*1
58
10.5 Software Update – The procedure Service Prog./Calib
1. Update itself:
- Several pumps in parallel via CAN cable with SpSt and SpaceUpdate tool
3. Adaptation of user specific data like “Modification Data”, “Disposable Articles” and
“Drug
Drug Library
Library” when required (use Clone function within HiBaSeD)
Hint: All the above mentioned sw tools respectively programs comes together as part of
the “HiBaSeD Service-CD” (Article number: 8713301).
IMPORTANT: After a successful update the following steps must be carried out !
1. After the pump has been used with the update program, the device must be switched
off and disconnected from mains power supply before it can be used.
2. After a successful update the User settings, Modification and Calibration data of the
device need to be re-adjusted with the HiBaSeD-Tool.
59
10.5 Software Update – After update Service Prog./Calib.
• Sometimes the Disposables and the Drug data has to be uploaded to the device,
even if they are not marketed in red (see complete “Instruction for SW Update”).
• After the updade of the user- and adjust data sometimes (see complete “Instruction
for SW Update”) a CalPatchAkku or a CalPatch.ini has to be uploaded to the pump.
Note: If you do not download the CalPatchAkku an operation of the pump in battery
mode might not be possible
• To have the same customized settings in the Space pumps like they have been
before the Software update it is recommended to take the setting print out from and
After all steps are carried out, finally the pump should run thru a functional test
to verify the correct operation.
operation
• Insert an infusion line into the pump and start the infusion.
• Check in the Menu “Status” of the pump if the software version of the pump is
like desired (e.g. 686 J 03 0006)
If no error messages or alarm are coming up, the software update has been successful!
HC-XX-DE08 | Service Department
Page CoE AIS | Infusomat SP | Page 132
60
10.5 Software Update Service Prog./Calib.
Attention:
For calibration after the update from SW G direct to SW J please use for:
For calibration after the update from SW E/F direct to SW J please use for:
Therapy Modes:
- Ramp and Taper
- Multi Does Mode
- Program Mode
- Dose Over Time
61
10.5 Software Update – News in SW “J” Service Prog./Calib.
W-LAN Battery
It is possible to load (download) an existing drug file from a single pump (new in
HiBaSeD)
Serial cable SP
PC (No.: 8713234)
Space pump
62
How to downgrade Device Software
5. After the pump software has been erased it is important to initiate the connection
between the SpaceUpdate program and the device by pressing the on/off button for
several seconds (the update program will show the identified pump when connected) .
11. Checking
63
11.1 Checking after repair Checking
1. Mains connection
2. Primary side
3
3. Secondary side
4. Test adapter (SP)
5. Measuring device
64
11.2 Performing the TSC Checking
In general:
Measuring- and cut off values are only valid when using a new (unused) tubing set!!
The electronic pressure cut off is realized via force sensor
Caution:
Don't stress the sensors with hard objects.
Forces over 15N leading to damage of the sensors
Attention:
When inserting the 3 mm allenkey be
aware not to damage the loudspeaker!
Test setup for mechanical pressure cut off adjustment
65
11.2 Performing the TSC Checking
Air sensor
The air sensor check will be done in the „Service Info“ - mode!
Connect service plug to device and press the „“ key inside the line changing
menu.
Hint:
The air alarm triggering depends on the adjusted sensor sensitivity for single
bubble and for cumulated air alarm.
66
11.2 Performing the TSC Checking
Delivery accuracy
For example:
25 ml of fluid has been delivered at a rate of
250 ml/h within 5 min. 42,0 sec.
the delivery speed is to high (+5%)!
For example:
25 ml of fluid has been delivered at a rate of
250 ml/h within 5 min. 42,0 sec.
the delivery speed is to high (+5%)!
67
Typical measuring mistakes
OK!!
68
12. Documentation und Certifying
• summary
• open questions
69