System
System
Start time: 4/22/2020 1:26:09 PM
Instrument: iCAP Q
Operator: ICP-MS-PC\ICP-MS
Template: STD BHIOS 4
Instrument Serial Number: 03310R
Last Autotune: Autotune-KED Autotune Line 1-20200422-132020411.imatdat
Solution: No solution specified
Sensitivity & Stability Test
Sensitivity
Sensitivity
Analyte Result Value Condition Limit
Bkg4.5 Passed 0.13 CPS Less than 1.0 CPS
Bkg220.7 Passed 0.0 CPS Less than 2.0 CPS
7Li Passed 83,867.0 CPS Greater than 50,000.0 CPS
59Co Passed 203,306.0 CPS Greater than 100,000.0 CPS
238U Passed 495,699.0 CPS Greater than 300,000.0 CPS
140Ce.16O/140Ce Passed 0.0193 Less than 0.025
137Ba++/137Ba Passed 0.026 Less than 0.03
115In Passed 376,975.0 CPS Greater than 220,000.0 CPS
Stability
Stability
Analyte Value Limit
7Li 0.8 % 2
59Co 0.5 % 2
238U 0.4 % 2
115In 0.4 % 2
Analyte Result Centroid Mass [u] Offset Peak width [u] Peak width min [u] Peak width max [u]
7Li Passed 6.9946 0.0214 0.726 0.650 0.850
59Co Passed 58.9177 0.0155 0.711 0.650 0.850
115In Passed 114.8442 0.0597 0.768 0.650 0.850
238U Passed 237.9889 0.0619 0.777 0.650 0.850
Tunesettings
Tune Settings
Parameter Value
Additional Gas Flow 1 0.00
Additional Gas Flow 2 0.00
Additional Gas Flow 3 0.00
Angular Deflection -382.40
Auxilliary Flow 0.80
CCT Bias -2.01
CCT Entry Lens -94.00
CCT Exit Lens -160.01
CCT Focus Lens -1.92
CCT1 Flow 0.00
CCT1 Shut-Off Valve 0.00
CCT2 Flow 0.00
CCT2 Shut-Off Valve 0.00
Cool Flow 14.00
D1 Lens -198.39
D2 Lens -80.00
Deflection Entry Lens -35.00
Extraction Lens 1 Negative 0.00
Extraction Lens 1 Polarity 0.00
Extraction Lens 1 Positive 0.00
Extraction Lens 2 -144.33
Focus Lens 21.70
Nebulizer Flow 0.81
Peristaltic Pump Speed 40.00
Plasma Power 1350.00
Pole Bias -1.00
Quad Entry Lens -28.01
Sampling Depth 5.00
Spray Chamber Temperature 2.50
Torch Horizontal Position -0.50
Torch Vertical Position -0.77
Virtual CCT Mass Maximum Dac Limit Set 4095.00
Virtual CCT Mass parameter b 0.65
Virtual CCT Mass to Dac Factor 130.00
Virtual CCT Mass to Dac Offset -190.00
Vacuum Check
Vacuum Check
Parameter Result Value
Analyzer Pressure Vacuum ok 4.032e-7
Interface Pressure 1.581e+0
Detector Voltages
Detector Voltages
Analog Counting
-2125.00 1500.00
Analyte: 7Li
59Co
Analyte: 59Co
238U
Analyte: 238U
115In
Analyte: 115In