Академический Документы
Профессиональный Документы
Культура Документы
Operator’s Manual
151 Graham Road · P.O. Box 9010 · College Station, Texas 77842-9010
Telephone (979) 690-1711 · FAX (979) 690-0440 · www.oico.com · oimail@oico.com
Notice The information contained in this document may be revised without notice.
OI Analytical shall not be liable for errors contained herein or for incidental or
consequential damages in connection with the furnishing, performance, or use of this
material.
Publication 21860309
Copyright 2005, 2008 OI Analytical
ii
Limited OI Analytical warrants each OI Analytical manufactured product against defects in
Warranty materials and workmanship under normal use and service for a period of one year.
Equipment installed by OI Analytical is warranted from the installation date; all other
equipment is warranted from the ship date. If purchaser schedules or delays installation
more than 90 days after delivery, then the warranty period starts on the 91st day from
date of shipment. This warranty extends only to the original purchaser. OI Analytical
will, at its option, repair or replace equipment that proves to be defective during the
warranty period, provided the equipment is returned to OI Analytical at the expense of
the purchaser.
Consumables, expendables, and parts are warranted for 30 days and are not covered
under extended warranties or service contracts.
OI Analytical warrants for a period of one year from the date of delivery: (i) the
Software, when installed and used with an OI Analytical recommended hardware
configuration, will perform in substantial conformance with the documentation
supplied with the Software; and (ii) the physical media on which the Software is
furnished will be free from defects in materials and workmanship under normal use.
This warranty shall not apply to defects originating from, but not limited to, the
following:
iii
iv
Table of Contents
Notice.......................................................................................................................... ii
Limited Warranty ..................................................................................................... iii
Chapter 1: Introduction.................................................................................. 1
System Design ............................................................................................................ 1
Operating Principles ................................................................................................... 1
Available Cyanide Method.......................................................................................1
Total Cyanide Method..............................................................................................2
Postdistillation Total Cyanide Method .....................................................................3
Features....................................................................................................................... 3
Specifications.............................................................................................................. 3
General Specifications..............................................................................................3
Dimensions .......................................................................................................3
Weight ...............................................................................................................3
Analytical Configurations .................................................................................4
Options ..............................................................................................................4
Instrument Specifications .........................................................................................4
3090 Autosampler .............................................................................................4
Mini 8 Pump .....................................................................................................4
Valve .................................................................................................................5
Amperometric Detector ....................................................................................5
Photometric Detector Option ............................................................................5
UV Digestor Module Option ............................................................................5
Requirements............................................................................................................5
Space Requirements ..........................................................................................5
Power Requirements .........................................................................................5
Computer Requirements ...................................................................................6
Safety Information ...................................................................................................... 6
Operator Precautions ................................................................................................6
General Precautions..................................................................................................7
Safety Symbols.........................................................................................................7
System Design
The FS 3100 system includes the FS 3100 analyzer with injection valve and detector,
eight-channel pump, 90-position autosampler, and WinFLOW™ software. The user
orders the system with the desired analytical configuration. A computer and printer is
provided by the user. Other optional modules must be ordered separately.
Operating Principles
The FS 3100 combines several novel techniques for determining available and total
cyanide species including the following:
The sample is first treated with ligand exchange reagents (available or weak acid
dissociable (WAD) cyanide reagents A and B) to liberate cyanide ion (CN–) from weak
metal-cyano complexes. Under acidic conditions, CN– in the sample converts to HCN.
The HCN diffuses across a hydrophobic membrane into a basic acceptor solution where
it converts back to CN–. The cyanide ion is then detected amperometrically at the
flowcell. The analysis time for available cyanide is approximately two minutes.
Figure 1.1 depicts a general flow diagram for available cyanide utilizing sulfide
abatement reagents, as per ASTM D-6888-04.
Gas diffusion
module
To Carrier
waste
Acid reagent
Figure 1.1. General Flow Diagram For Available Cyanide as per ASTM D-6888-04 Using the
FS 3100
For total cyanide measurement, the sample passes through a UV digestor to break down
metal-cyano complexes. Under acidic conditions, CN– in the sample converts to HCN,
which diffuses across a hydrophobic membrane. The hydrogen cyanide gas absorbs into
a basic acceptor solution where it converts back to CN–. The cyanide ion is then
detected amperometrically at the flowcell. The analysis time for total cyanide is about
two minutes. Figure 1.2 depicts a general flow diagram for total cyanide using the
FS 3100.
NOTE: The FS 3100 methodology’s unique detection scheme differs from other
regulatory methods. Therefore, FS 3100 results may differ when
compared to results obtained by such methods.
Gas
diffusion
module
Base
To AMP reagent
waste detector
Sample
UV
digestor
To Carrier
waste
Air
TA 1
TA 2
Figure 1.2. General Flow Diagram for Total Cyanide Using the FS 3100
Pyridine
To Waste 570 nm
Chloramine-T
Sample
Air
Buffer
Figure 1.3. General Flow Diagram for Post-Distillation Total Cyanide Using the FS 3100
Features
• No distillation required for amperometric detection
• Available cyanide analysis in two minutes
• Total cyanide via UV digestion
• No pyridine or barbituric acid when measuring available cyanide
• No interference from thiocyanate
• WinFLOW Windows®-based software
• 90-position autosampler
• 2 ppb minimum level (ML)
• 0.5 ppb method detection limit (MDL)
Specifications
General Specifications
Dimensions
Weight
Analytical Configurations
• Available cyanide
• Total cyanide
• Post-distillation total cyanide, photometric
Options
• UV Digestor module
• Expanded Range™ photometric detector
• Photometric flowcell: PEEK®, 5- or 10-mm path length
Instrument Specifications
3090 Autosampler
• X-Y-Z autosampler
• 90-place maximum sample capacity
• Nine-place standards capacity
• 90-place sampler rack
• Dimensions: 25 cm H x 33 cm W x 33 cm D
9.8" H x 13" D x 13" D
• Space requirements: 76 cm H x 71 cm W x 71 cm D
30" H x 28" D x 28" D
• Weight: 7.9 kg (17.4 lbs)
• Input-output: one dedicated wash pump control relay; three programmable relays;
I/O port; two serial RS-232C ports
Mini 8 Pump
• Dimensions: 13 cm H x 17.5 cm W x 22 cm D
5" H x 7" W x 8.3" D
• Weight: 5.1 kg (12.1 lbs)
• Microprocessor-controlled DC motor
• RS-232 input for external control
• Variable speed: 0.4–45 rpm, adjustable in 0.1% steps or as flow rate
• Flow rate: 0.005–150 mL/minute
• Eight independently tensioned channels
• Eight rollers, 18/8 stainless steel, actively driven
• Differential pressure maximum: 1.0 bar (100 psi)
• Power requirements: 110–120 VAC, 50/60 Hz (2 x 500 mA Slo-Blo® fuses)
220–240 VAC, 50/60 Hz (2 x 250 mA Slo-Blo fuses)
• Power consumption: 30 watts
• Protection rating: IP 30
Valve
Amperometric Detector
• Amperometric at 0.00 V
• Flowcell: Silver working electrode; silver-silver chloride reference electrode;
stainless steel counter electrode
• Dimensions:
3.5" H x 18.25" W x 15.5" D (9 cm H x 46 cm W x 39 cm D)
Requirements
Space Requirements
• 66 cm L x 76 cm W x 41 cm H
26" L x 30" W x 16" H
Power Requirements
Computer Requirements
Safety Information
The OI Analytical FS 3100 meets the following International Certification when tested
in a typical configuration:
EN 1010
EN 55011:1994
IEC 801-2/CENELEC 60801-2/1000-1-2
IEC 801-3/CENELEC 50140-2/1000-4-3
IEC 801-4/1000-4-4
The FS 3100 has been designed and tested in accordance with recognized safety
standards and designed for use indoors. Using the instrument in a manner not specified
by the manufacturer can impair the instrument’s safety protection. Whenever the safety
protection of the FS 3100 has been compromised, disconnect the instrument from all
power sources and secure the instrument against unintended operation.
Operator Precautions
For operator safety, pay attention to WARNING and CAUTION statements throughout
this manual.
Warnings and precautions in this manual or on the instrument must be followed during
operation, service, and repair of the instrument. Failure to follow these warnings and
precautions violates the safety design standards and intended use of the instrument.
OI Analytical is not liable for the operator’s failure to comply with warnings and
precautions.
General Precautions
• Disconnect the AC power cord before removing covers.
• Replace or repair faulty or frayed insulation on power cords.
• Perform periodic leak checks on supply lines, fittings, and pneumatic plumbing.
• Arrange gas lines so they cannot become kinked, punctured, or otherwise damaged,
and do not interfere with foot traffic.
• Turn off the main power switch and disconnect the main power cord before using a
liquid solution to locate leaks.
• Wear safety glasses to prevent possible eye injury.
• Do not perform unauthorized modifications or substitute parts to the instrument that
are not OI Analytical original parts. Any unauthorized modifications or
substitutions voids the warranty.
• Verify all heated areas have cooled before handling or wear adequate hand
protection to prevent burns.
Safety Symbols
NOTE: Do not throw away the factory packaging. Keep it for possible future use.
This is one of the warranty conditions.
FS 3100 Front
Figure 2.1 depicts the front of the FS 3100, with callouts to the various elements.
Descriptions for these elements are listed below.
Temperature
Switch Bias Voltage
LCD
LED Indicators
Potentiometer Amperometric
Detector
Temperature
Verification Port
Gas Diffusion
Module
Six-port injection valve regulates the addition of samples and reagents through the
system.
Gas diffusion module is a manifold where samples and reagents mix. It holds the gas
diffusion membrane, which separates substances in a reagent stream based on differing
rates of diffusion across its material.
Potentiometer adjusts the amperometric detector’s bias potential between the reference
and working electrodes.
FS 3100 Back
Figure 2.2 depicts the back of the FS 3100, with callouts for the elements. Descriptions
for the elements are listed below.
Aux Detector
Connector
Valve
Connector
Pump
Connector
Power PC
Switch Connector
Power Receptacle Fuse Holder Sampler
Connector
Aux detector connector is a nine-pin female connector that joins the FS 3100 to an
optional additional detector using the appropriate cable (PN A001644), allowing
communication between the two.
Fuse holder contains fuses that protect the FS 3100 from a short-circuit condition.
Pump connector is a nine-pin male connector that joins the FS 3100 to the Mini 8
Pump using the appropriate cable (PN A001644).
Sampler connector is a 25-pin male connector that joins the FS 3100 to the
3090 Autosampler using the appropriate cable (PN 322319).
FIA Module The FIA Module (part number 322676H) is used to perform FIA Analytical Methods
and houses a six-port Injection Valve, a variable heater, and the applicable VersaChem
Multi-Test Manifold. Refer to Figure 2.3 for an example of a FIA Module.
Variable Heater
LEDs and Switch
VersaChem
Ports
Six-Port
Temperature Injection Valve
Verification Port
Heater
Inlet
Heater
Outlet
Figure 2.3. FIA Module
Heater Inlet – This port provides a connection to the heater assembly, which is
typically from the VersaChem outlet.
Heater Outlet – This port provides a connection to the heater assembly, which
typically connects to the flowcell or debubbler inlet, if applicable.
The six-port Injection Valve regulates the addition of samples and reagents into the
FS 3100 system. Figure 2.19 shows how the ports are labeled, while Table 2.5 indicates
each port’s purpose. Refer to Chapter 7, “Configuring the Six-Port Injection Valve (FIA
only)” on page 79, for more detailed information.
5
6 4
1 3
2
Figure 2.4. Six-Port Injection Valve
Port Connection
2 Carrier Line
3 To the VersaChem
5 Waste
6 Sample Probe
Platen
Control Panel
Platen
Extender
Release Latch
Control panel contains the pump display. It is used to set variables and to manually
start or stop the pump.
Platen extender extends the width of the platen and provides mounting for the pump
tube collar.
Pump head (not shown) contains the pump rollers that rotate and compress the pump
tubes, pushing liquid through the tubing.
Power Switch
Voltage
Selector
Fuse Holder
Power Receptacle
Fuse holder with voltage selector contains fuses that protect the pump from a short-
circuit condition. The voltage selector (not shown) sets the voltage setting to either
115 V or 230 V.
RS-232 port is a nine-pin female connector than joins the pump to the FS 3100 using
the appropriate cable (PN A001644).
Run/Stop
Mode Display
Direction arrows set the flow direction. During right-to-left operation, a negative sign
(–) appears on the display.
Max/Cal changes the pump speed between set speed and maximum speed, and is used
to calibrate flow rate or dispensing volume.
Mode selects the operational mode. Select PUMP mode for FS 3100 operation.
16-Channel Specifications
Front View
Control Platen
Panel
Platen
Extender
Release
Latch
Control Panel – Use this panel to set variables and manually start or stop the pump.
Platen – The platen attaches the pump tube on the pump head.
Platen Extender – Use the platen extender to extend the width of the platen and
provide mounting for the pump tube collar.
Pump Head – The pump head (not shown) contains the pump rollers that rotate and
compress the pump tubes, pushing liquid through the tubing.
Release Latch – This latch disengages the platen from the pump head.
Back View
Analog Interface
Port
RS-232 Port
Power Switch
Power Receptacle
Analog Interface Port – This port, labeled ANALOG, is not used with the FS 3100
system.
Fuse Holder with Voltage Selector – The fuse holder contains fuses that protect the
pump from short-circuiting. The Voltage Selector (not shown) sets the voltage setting
to either 115 V or 230 V.
Power Switch – This switch turns the power to the Precision Pump on or off.
RS-232 Port – This port nine-pin female connector, labeled RS-232 IN, connects the
pump to the Analysis Unit using the appropriate cable (part number A001644).
Z-drive Assembly
Autosampler Arm
Sample Vial Rack
Flowing rinse station washes the sample probe between samples. It includes tubing
that connects the rinse station to the rinse source and to the waste container.
Sample tray holds the standards vial rack and one sample vial rack. Ribs located on the
bottom of the sample tray hold the racks in place.
Sample vial rack holds sample vials. The autosampler accommodates 90 vials per
rack.
Standards vial rack (not shown) holds nine standard vials. The autosampler includes
ten standards vials, which are 30-mL conical centrifuge vials with caps.
Z-drive assembly includes a Y-axis slider block, guide plate, and sample probe. The
Z-drive assembly fits onto the autosampler arm.
Auxiliary-2nd Specifications
Photometric or
Amperometric • Dimensions: 15 cm H x 25 cm W x 38 cm D
Channel Analysis (6” H x 10” W x 15” D)
Unit for FIA • Weight: 6.8 kg (15 lb)
• Consists of the Auxiliary-Channel Analysis Unit Chassis, a variable heater, an
Injection Valve, a detector, and a flowcell
Temperature Six-Port
Switch Injection Valve Auxiliary-
LED Indicators Channel
Analysis Unit
Chassis
Temperature
Verification
Port
Flowcell
VersaChem Multi-
Test Manifold
Heater Heater
Inlet Outlet
VersaChem Six-Port
& Heater Injection Valve
Auxiliary-Channel Analysis Unit Chassis – The Auxiliary Chassis holds the FIA
Module (refer to “FIA Module” on page 25; part number 323887) and the Detector
Module (see “Detector Module” on page 29).
FIA Module Heater Inlet – This port provides a connection to the heater assembly, which is
typically from the VersaChem outlet.
Heater Outlet – This port provides a connection to the heater assembly, which
typically connects to the flowcell or debubbler inlet, if applicable.
LED Indicators – These lights describe the state of the heater. AT TEMP is lit when
the heater is at the temperature set by the Temperature Switch. NOT AT TEMP is lit
when either the heater is warming or the system is cooling.
Six-Port Injection Valve – The Injection Valve regulates the addition of samples and
reagents into the system; see “Heater Inlet” on page 26 for more detailed information.
Temperature Switch – This three-way switch varies the temperature among ambient,
37 °C, and 50 °C.
Detector Module Photometric Detector – The Photometric Detector measures the absorbance of colored
product in the analytical stream and provides a digital response proportional to the
absorbance; refer to “Photometric Detector” on page 45 and Chapter 8, “Photometric
Detector Module” on page 85, for further information.
Photometric Flowcell – The standard flowcell is a 5-mm light-pass flowcell; the 10-
mm light-pass flowcell is used for some methods to increase the sensitivity.
Back View
Auxiliary-
Channel
Analysis Unit
Power Chassis
Switch
Main Detector
Connection
Power
Receptacle
Fuse Holder
Auxiliary-Channel Analysis Unit Chassis – The Auxiliary Chassis holds the FIA
Module (refer to “FIA Module” on page 25; part number 322676H) and the Detector
Module (see “Detector Module” on page 29).
Fuse Holder – The fuse holder contains the fuses that protect the Analysis Unit from
short-circuiting.
Power Switch – This switch turns the power to the Analysis Unit on or off.
Analytical Cartridge
Each Analytical Cartridge includes a VersaChem Multi-Test ManifoldTM and the
appropriate Chemistry Kit, which includes the Tubing Kit, the fittings, a sample loop,
an optical filter, and the Analytical Methodology. Figure 2.25 illustrates the breakdown
of an Analytical Cartridge.
Analytical Cartridge
VersaChem
Chemistry Kit
Multi-Test
Manifold
The pre-configured cartridge design enables quick setup with reduced errors during
installation. In addition, the design facilitates quick conversion between tests with
reduced cross-contamination.
Chemistry Kit OI Analytical’s chemistry solution requires ordering specific Chemistry Kits. The
customer may then reconfigure the Precision Pump tubing to alternate among the
various chemistry applications. Contact OI Analytical for a current, complete list of
available kits.
VersaChem Multi- The VersaChem Multi-Test ManifoldTM provides an environment where the sample and
Test ManifoldTM reagents are combined. The VersaChem is a central unit for the sample, carrier, air,
reagents, and mixing coils.
1 A Carrier/Sample
2 B
3 C
4 D
E Mixing Coils
5
6 F
Outlet/Mixing Coil
Center Ports
Figure 2.14. VersaChem Multi-Test Manifold for FIA
Front
Z-Drive Assembly
Autosampler Arm
Autosampler Arm – The arm provides the probe motion in the X-Y plane
(horizontally).
Needle Manifold – Also referred to as the “Y-Axis Slider Block”, the needle manifold
holds the needle, needle sleeve, sample transfer line, and sleeve gas transfer line.
Rinse Station – The rinse station washes the sample probe between samples; tubing
connects the rinse station to the onboard peristaltic pump located on the back of the
Autosampler.
Sample Vial Racks – The racks contain 90 positions for 2- and 8-mL sample cups; the
Autosampler includes four 90-position sample racks, although optional 60-position
sample racks accommodate 4-mL and 12-mL sample cups.
Sample Tray – The sample tray can accommodate up to four sample racks; ribs located
on the bottom of the sample tray hold the sample vial racks in place.
Standards Vial Rack – The rack contains 12 positions for standards vials.
Z-Drive Assembly – This assembly includes a Y-axis slider block, guide plate, and
needle manifold; the Z-drive assembly fits onto the Autosampler arm.
Back
Power Peristaltic
Switch Pump
Adjustment
COM 2 Port
Screw
Power Onboard
Receptacle Peristaltic
Pump
USB Port
Auxiliary
COM 1 Port Port
Figure 2.16. 3360 Autosampler back view
COM 1 Port – This RS-232 Serial I/O port connects the Autosampler to the Analysis
Unit using the appropriate cable (part number 322319).
COM 2 Port – This RS-232 Serial I/O port connects the Autosampler to other external
devices, such as an autodilutor.
Onboard Peristaltic Pump – This two-channel pump moves the wash solution from
the wash source, through the wash station, and out to waste.
Peristaltic Pump Adjustment Screw – This screw fine-tunes the onboard peristaltic
pump’s flow rate.
Power Switch – This switch turns the power to the Autosampler on or off.
Detectors
The Detectors can accommodate various modules. They detect a wide range of analyte
concentrations with a single calibration using WinFLOWTM software. Please refer to
Chapter 8, “Detector(s)” on page 85, for detailed information about installing and
operating the detectors.
Amperometric
Bias Display Detector Module
Potentiometer
Red Wire
Black Wire
Bias Display – This LED display provides the detector bias potential, which is adjusted
by the potentiometer.
Black Wire – The black wire connects via a clip to the counter electrode (near where
the stainless steel fitting enters the flowcell body).
Potentiometer – The potentiometer adjusts the detector bias potential between the
reference and working electrodes over a range of -1.25 V to +1.25 V. The potentiometer
should be adjusted to a bias reading of 0.00.
Red Wire – The red wire connects the detector to the reference electrode plug.
White Wire – The white wire connects the detector to the working electrode plug.
Photometric
Flowcell
Detector Module
Outlet
Debubbler
To Waste
Debubbler
Sample In
Debubbler
Sample Out Debubbler
Figure 2.18. Photometric detection option
Debubbler – If applicable, the debubbler removes bubbles from the analytical stream.
Debubbler Sample Out – This port connects via Teflon® tubing from the Debubbler;
the sample without bubbles is sent to the flowcell.
Flowcell Outlet – This port (not shown) connects via the waste line to a waste
container.
Optional Components
Input
Output
Syringe Valve
Syringe
Safety Shield
Input – This port connects the syringe to tubing that routes to the autosampler.
Output – This port connects the syringe to tubing that routes to the waste bottle.
Safety Shield – The shield protects the autodilutor’s moving parts during operation.
Switching Valve – This valve (not shown) is the internal valve that switches between
the normal mode, which bypasses the syringe, and the dilute mode, which places the
syringe online.
Syringe – The syringe pulls precise, specified sample volumes and diluent to dilute the
sample in the desired manner.
Syringe Valve – This valve switches the syringe between input and output positions.
Back View
To Waste Port
To Autosampler
Power Receptacle
Fuse Holder
Power Switch
From Wash Port – This port connects tubing from the autodilutor to the wash bottle.
Fuse Holder – The fuse holder contains fuses that protect the autodilutor from short-
circuiting.
Power Receptacle – This receptacle provides electrical power via a power cord that
plugs into an AC outlet.
Sample In, To Probe – This port connects tubing from the autodilutor to the
autosampler probe.
Sample Out, To Pump – This port connects tubing from the autodilutor to the pump.
To Waste Port – This port connects tubing from the autodilutor to the waste bottle.
Inlet connects the UV Digestor to tubing that routes in from the FS 3100.
Outlet connects the UV Digestor to tubing that routes out to the FS 3100.
Fan provides cooling to the UV Digestor during operation. It starts when the power
switch is turned on.
Fuse holder contains fuses that protect the UV Digestor from a short-circuit condition.
Photometric Injection
Valve/Manifold Module Photometric Plug-In
Detector Module
Six-Port
Injection
Valve
Manifold
Debubbler
Inlet
Heater
Outlet
Debubbler
Heater outlet provides a connection from the heater assembly, which usually connects
to the debubbler inlet.
Six-port injection valve regulates the addition of samples and reagents through the
system.
Manifold provides the environment where sample and reagents are added and mixed to
generate a colored product
Before unpacking any FS 3100 components, review the dimensions required by the
FS 3100 listed in Chapter 1, “Dimensions” on page 3. In addition to this required area,
allow space for the operator-supplied computer and any associated peripherals. Space
for reagent bottles and waste containers is included in the FS 3100 dimensions.
The FS 3100 ships in multiple cartons. Any system accessories ship in separate
containers. After opening the shipping container, unpack the instrument and check the
items against the component list. If any damage appears, notify the carrier immediately.
Save all packing materials until verifying proper system operation.
Installing Fuses
Verify the correct fuses are installed by removing the fuse holder on the FS 3100 back.
a. Insert a small screwdriver into the notch between the fuse holder and the power
receptacle, depicted in Figure 3.1.
Aux Detector
Connector
Valve
Connector
Pump
Connector
Power PC Connector
Switch Power Receptacle Fuse Holder Sampler
Connector
b. Remove the fuse holder by prying up the edge with the screwdriver.
Line
Fuse PN
Voltage
3. If necessary, insert the two correct fuses into the fuse holder.
1. Install the autosampler cable by connecting it to the port labeled Sampler on the
back of the FS 3100.
The autosampler cable that connects the autosampler to the FS 3100 (PN 322319) is
included in the FS 3100 startup kit.
2. The other end of the cable connects to the COM 1 port on the back of the
3090 Autosampler.
3. Connect one end of the PC cable (PN A001644) to the port on the back of the
FS 3100 marked PC Connection.
4. Connect the other end of this cable to the serial port on the computer.
5. Install the pump cable (PN A001644) by connecting it to the port labeled Pump on
the back of the FS 3100.
6. Connect the other end of the cable to the RS-232 port on the back of the Mini 8
pump.
7. Install the power cord into the power receptacle on the FS 3100 back.
8. After verifying the instrument is turned off, plug the power cord into the electrical
power outlet.
Serial port
PC
PN A001644 PN A001644
PN 322319
Analytical Configuration
1. Remove the twist ties used to hold tubing in place during shipment.
2. Carefully pull the tubes away from the FS 3100, making sure not to detach any
connections.
3. Connect additional tubing from the valve and detector to the gas diffusion
module according to the detailed flow diagram in the analytical method.
The FS 3100 is configured at the factory to measure available cyanide when ordering
PN 322688 (System FS 3100, OIA 1677, Available CN) and PN 324405 (Cartridge,
OIA 1677, Available CN). This method does not use a UV digestion step. Refer to the
detailed flow diagram provided with Method OIA-1677 Available Cyanide by Ligand
Exchange and Flow Injection Analysis (PN 323452).
The FS 3100 is configured for total cyanide at the factory when ordering PN 322696
(System FS 3100, CN, Total, Gas Diffusion, Amperometric) and PN 325581 (Cartridge,
CN, Total). Refer to the detailed flow diagram provided with Method Total Cyanide by
UV Digestion and Flow Injection Analysis (PN 323153).
NOTE: If both available and total cyanide are measured routinely on most
samples, connect the UV Digestor in-line and switch on the lamp for total
cyanide or off for available cyanide analysis. A notation of the lamp status
should be made in the sample comments section in WinFLOW.
Injection Valve
The injection valve arrives from the factory with the tubing already installed. The valve
has six labeled ports as described in Table 3.2.
Port 6, sample in Sample line Pulls sample from the sample cups
in the autosampler.
Port 5, to waste Green/green pump tube Pulls sample through the sample
line; used to fill the sample loop.
Port 2, carrier in Carrier pump tube Pushes sample through the sample
loop; used to empty to sample loop
to the gas diffusion module.
Port 3, to gas Transmission tubing, 5" Connects the valve and gas diffusion
diffusion module module.
Port 1 and 4, Sample loop Loads and injects specific sample
sample loop volumes.
• Attach the sample pull-off pump tube specified in the analytical method to the
Luer-Lok® fitting at port 5. Use the appropriate Luer-Lok barbed fitting.
• Locate the carrier pump tubes specified in the flow diagram in the FS 3100
accessory kit. Both pump tubes attach with Luer-Lok fittings.
• The #3 transmission line attaches to the gas diffusion module. Both the sample line
and #3 line attach using a nut and ferrule.
The injection valve has two operational modes, load and inject. In the load position, the
sample loads into the sample loop while carrier flows to the gas diffusion module. In
the inject position, the sample enters the gas diffusion module with the carrier stream.
In the load position, depicted in Figure 3.3, the sample from the autosampler enters the
injection valve at port 6. It moves internally to port 1, through the sample loop to port 4,
and then internally to port 5, where it exits the valve to waste via the sample pull-off
pump tube. The carrier flows into the injection valve at port 2 and continues out port 3
to the gas diffusion module.
Pump
Sample
To pull-off
5 4 waste
To gas
diffusion
module
Sample in 6 3
1 2 Carrier
Front
In the inject position, depicted in Figure 3.4, the carrier stream, which enters the
injection valve at port 2, injects the sample in the sample loop into the gas diffusion
module. The stream moves internally to port 1, then pushes the sample through the
sample loop from port 1 to 4. It moves internally to port 3, where it exits the valve to
the gas diffusion module.
While the injection valve is in the inject position, the sample probe is in the wash
reservoir. Wash solution enters the valve at port 6, moves internally to port 5, and exits
the valve to waste via the sample pull-off pump tube.
Pump
Sample
To pull-off
waste
5 4
To gas
diffusion
module
Sample in 6 3
(wash from reservoir)
1 2 Carrier
Front
The sensitivity of the FS 3100 to cyanide is adjusted by changing the sample loop
volume. Larger injection volumes (200-μL sample loop) yield higher sensitivity and
allow lower detection limits. Smaller injection volumes allow higher cyanide
concentrations to be measured. To change the injection loop, unscrew the injection loop
from ports 1 and 4, then attach the new loop. Sample loop sizes specified in the
analytical method are located in the FS 3100 accessory kit.
Calculate sample loop volumes for 0.040" I.D. and 0.020" I.D. tubing using
Equation 3.1.
1. Locate the sample lines and “to waste” pump tubes in the autosampler accessory
kit.
2. Connect the sleeved end of the single channel sample line to the PEEK (or
stainless steel) sample probe.
3. Connect the end with the nut and ferrule to port 6 on the valve module for the
channel to be run.
Amperometric Detector
The amperometric detector arrives from the factory with the amperometric flowcell
mounted and installed and without the reference electrode installed. The reference
electrode is shipped in the startup kit.
The following procedure outlines how to install the amperometric detector reference
electrode.
5. Orient the flowcell so that the alignment dowel pins are to the back.
The working electrode should be to the left of the center of the flowcell (see
Figure 3.5 to identify flowcell components).
6. Push the flowcell onto the mounting pins located on the detector module.
7. Connect the red wire from the detector module to the reference electrode plug.
9. Connect the clip on the black wire to the counter electrode near where the
stainless steel fitting goes into the flowcell body.
10. Connect the flow tube from the base flow of the gas diffusion module to the inlet
on the left side of the flowcell body.
11. Connect the outlet flow tube with backpressure coil to the counter electrode
adapter fitting.
Screws (6)
Reference Plate
Electrode
Adapter Fitting
Reference Fitting
Electrode
Port
Counter Electrode
Alignment
Upper Cell Dowel Pins
Assembly
Working Threaded
Electrode Plate
Lower Cell
Assembly
Maintaining a Basic pH
The acid reagent concentration is higher than the base reagent concentration. Therefore,
mixing waste created by the system can create an acidic solution. This poses a potential
safety hazard since small cyanide volumes may also be present in the instrument
reagent waste stream.
WARNING: Failure to correctly install the waste containers can produce toxic
hydrogen cyanide gas.
NOTE: Dispose of all waste in a manner consistent with regulatory rules for your
laboratory.
The following procedure outlines how to treat the waste to maintain a basic pH.
Two 4-L plastic waste containers are provided in the FS 3100 accessory kit. Use
one for sample waste and the other for instrument reagent waste.
The following procedure outlines how to connect the sample waste container.
3. Connect the tubing from the injection valve to the sample waste container using
the Luer-Lok fitting.
The stopcock valve prevents fluid from inadvertently leaving the flowcell through the
reagent waste tubing connected between the flowcell and the reagent waste container.
The following procedure outlines how to install the stopcock valve.
CAUTION: Failure to open the stopcock and clip during instrument operation can
damage the gas diffusion membrane.
NOTE: The clip attached to the flowcell exit line can be removed if the stopcock is
installed.
NOTE: Allowing reference electrode filling solution to drain out of the flowcell
reduces the life of the electrode. Correctly using the stopcock valve
significantly increases the life of the reference electrode and reduces
routine flowcell maintenance.
1. Using the Luer-Lok fittings, connect the stopcock valve (PN A002586) to the top
of the reagent waste container, as depicted in Figure 3.6.
This stopcock valve is connected either directly to the waste container or with a
piece of tubing, as necessitated by the site configuration.
Flowcell
To
Reagent
Stopcock Valve
2. Attach the exit line out of the amperometric flowcell to the inlet side of the
stopcock.
Close the stopcock only after the instrument is shutdown to prevent the sodium
hydroxide acceptor reagent from backflowing out of the flowcell.
The following procedure outlines how to connect the reagent waste container.
3. Connect the waste line from the gas diffusion membrane module to the waste
container using the Leur-Lok fitting.
Installing WinFLOW
See the WinFLOW Operator’s Manual (located on the CD) for installation instructions.
3. Plug the power cord into the rear panel of the UV Digestor Module and an
appropriate wall socket.
4. Connect the inlet and outlet tubing according to the analytical method.
a. Connect one end of the auxiliary serial cable (part number A001644), included
with each additional Analysis Unit, to the serial port labeled TO AUX
DETECTOR on the back of the Main-Channel Analysis Unit.
b. Connect the other end of the auxiliary cable to the port labeled TO MAIN
DETECTOR on the back of the Auxiliary-Channel Analysis Unit.
5. Install the FS 3100 power cord into the power receptacle on the back of the
Analysis Unit(s).
6. Verify the Analysis Unit is powered off and connect the power cord to an electrical
outlet. Repeat this step for the Auxiliary-Channel Analysis Unit (if necessary), the
Autosampler, and the Precision Pump.
NOTE: The provided fluidic diagrams are examples of how to make fluidic
connections. Always refer to the Analytical Method for correct, detailed
fluidic connections.
1-Channel System Properly configure the VersaChem Multi-Test ManifoldTM and the Injection Valve
according to the procedures previously outlined in this manual and in the Analytical
Method. The setup should appear similar to Figure 3.8 (for FIA) or Figure 10.4 (for
SFA).
1 A Plumbed Debubbler
Empty Union
Detector
Deionized (DI) Water 6 F Filter
H To Waste
Sample Pull
Union Union Union
To Waste Backpressure
Flowcell Coil
Sample
Deionized (DI) Water
6-Way
6 5 Valve
Sample 1 4
Loop
2 3
Parallel 2-Channel If operating the FS 3100 in a parallel 2-channel platform, split the line connecting the
System for FIA Autosampler to the Analysis Units to accommodate operation in a tandem
configuration.
Use the following steps and Figure 3.9 to establish the appropriate connections for a
parallel 2-channel system for FIA.
Carrier 2 B
Color Reagent 3 C
Detector
Deionized (DI) Water 6 F Filter
H To Waste
Sample Pull
To Waste Union Union Union
Backpressure
Flowcell Coil
Series Sample In Line
Deionized (DI) Water
6-Way
6 5 Valve
Sample 1 4
loop
2 3
Legend
Parallel Sample In Line (from Autosampler)
Three-Port
Splitter Plumbed Debubbler
Empty Union
2 Backpressure
Carrier B coil
Exterior tubing,
0.8 mm I.D. Teflon®
OTCR (Open Tube Three-Port Splitter
Cadmium Reactor)
Color Reagent 3 C
Detector
Deionized (DI) Water 6 F Filter
H To Waste
Sample Pull
Union Union Union
To Waste Backpressure
Flowcell Coil
Figure 3.9. Fluidic connections for a 2-Channel System for FIA in parallel
1. Cut two lengths of the provided polyethylene tubing; ensure each cut is flat and
even.
2. Attach the line originating at the Autosampler to the top port of the three-way
splitter (part number A303-0114-00).
3. Attach the two lengths of tubing from step 1 to the left and right ports of the splitter.
4. Place a nut and a ferrule on the ends of the two split sample lines.
NOTE: Attach the nut to the sample line with the threaded portion facing the
corresponding end of the sample line. Attach the conical ferrule to the sample
line by orienting the wide end towards the conical indentation in the threaded
end of the nut.
5. Finger-tighten one of these lines onto port 6 of the Injection Valve on the Main-
Channel Analysis Unit.
NOTE: Connections between the sample lines and the three-way may require more
than physical adjunction. Cut segments of slightly larger diameter tubing for
use as sleeves for the connections, if needed. Adapt unused pump tubing for
this purpose.
6. Finger-tighten the other line onto port 6 of the Injection Valve on the Auxiliary-
Channel Analysis Unit.
7. Attach a sample pull line to port 5 of the Injection Valve on each Analysis Unit; this
line goes through the Precision Pump and out to waste.
NOTE: Adjust the timed events specified in the method to accommodate the multiple
Injection Valves and the additional sample load time, which varies with the
multi-channel configurations (serial or parallel).
NOTE: Simply arranging the fluidic and cable connections in a manner consistent
with multi-channel operation is insufficient for successful multi-channel
analysis. Verify the second Analysis Unit is an Auxiliary chassis and that the
WinFLOWTM software is configured for multi-channel data processing (using
the WinFLOW software manual) before operating.
Ensure the photometric plug-in detector module (PN A001821) contains a 570-nm
interference filter (PN A305-1570-00) in the filter holder, as depicted in Figure 3.10.
Filter Holder
Filter
Lamp
Photometric
Plug-in Detector
Module
Debubbler
2. Loosen the set screw in the filter holder with the P20 Allen wrench provided in the
FS 3100 accessory kit.
The wavelength is coded on the filter by three digits in the following manner:
“305-1xxx”. For example, a 570-nm filter is coded “305-1570.”
7. Return the filter holder in the lamp so the mirrored side faces the detector front.
This chapter provides information on installing and operating the Mini 8 Pump. Refer
to the IPC Pump Operator’s Manual included with the system for a full description of
its operation and maintenance.
Remove the packing checklist from the shipping container and check off items against
it. Leave accessories in the packing until ready to install them on the pump.
NOTE: Do not throw away the factory packaging. Keep it for possible future use.
CAUTION: Check the voltage selector setting on the back panel before turning on the
pump. Failure to set the voltage selector to the correct voltage can
permanently damage the pump.
3. Verify the voltage selector located on the pump back is set to the correct voltage.
The number visible in the window of the voltage selector on the fuse holder indicates
the pump’s nominal line voltage setting. See Figure 4.1 for the location of the voltage
selector and fuse holder.
Power Switch
Voltage
Selector
Fuse Holder
Power Receptacle
a. Insert the blade of a small screwdriver (size 0) into the slot above the power
receptacle and pry open the fuse holder. Pull the fuse holder straight out.
b. Remove the voltage selector from the fuse holder. The required voltage should
now be facing the window of the fuse holder
c. Rotate the voltage selector 180° and reinstall it in the fuse holder.
5. Plug the power cord into the power receptacles on the wall and pump back panel.
Pump Tube
Platen Extender Platen Tension Lever
Collar
Pump Tube
Collar
The pump uses platen extenders to extend the pump platen’s width. Pump tube collars
mount on the platen extenders. Always install platen extenders to provide the correct
pump tube tension.
Install platen extenders by pressing them into the grooves on each end of the platen. Be
sure the tubing groove in the extender matches the corresponding groove in the pump
platen.
The pump platens can be installed or removed individually without disturbing adjacent
channel operation.
Install a platen by positioning it above the matching guides on the pump base. Press
down until both sides lock into place. Be sure to observe the correct flow direction as
shown by the directional arrows on the platens. Install the platen so the arrow points
from the source on the right to the output on the left.
Remove a platen by pressing the release latches on both sides of the platen.
NOTE: The platens can install on the pump in either direction without damage.
However, always install platens with the arrow pointing in the correct flow
direction to avoid confusion when tracing flow through the system.
Two collars provide color coding for each pump tube. These collars serve the following
functions:
NOTE: Platen extenders must be installed on both sides of the platen. If not
already installed, install them before proceeding.
1. Remove a platen from the pump by pressing the release latches on both sides of
the platen.
3. Hook one of the collars into the channel on the platen bottom.
Be sure to observe the flow direction as shown by the directional arrows on the
platen top.
4. Stretch the pump tube slightly so the second collar fits into the channel on the
platen’s opposite side.
6. Verify the flow direction is correct from the input on the right to the output on the
left, and the directional arrow orients correctly.
Refer to the IPC Pump Operator’s Manual for a full description of independent pump
operation.
NOTE: New pump tubes require a break-in period for constant, reproducible flow
rates. Pump reagent water through new tubing for at least one to three
hours prior to running an application.
CAUTION: Siphoning of liquid can occur when disengaging the platens and releasing
tension on the pump tubes.
1. Verify the voltage selector located on the pump back is set to the correct voltage.
The voltage selector is part of the fuse holder. The number on the voltage selector
cover indicates the pump’s nominal line voltage setting. If necessary, convert the
pump to a different nominal operating voltage range (see “Installing the Mini 8
Pump” on page 48 in this chapter).
CAUTION: Failure to release tension shortens pump tube life. When the pump is idle
and the pump power is off, release tension on tubes not in use by pressing
down on the tension lever, or by pressing the release latches on either side
of the platen.
Overload situations occur mostly in the combined presence of large tubing inner
diameters and high speed.
This chapter provides information on installing and operating the 3090 Autosampler.
The 3090 Autosampler is designed for easy installation. Installation consists of three
parts: preparing for installation, assembling the autosampler, and connecting it to the
FS 3100. For the most part, the autosampler is installed without using tools. Remove
thumbscrews with tools if necessary, but only fingertighten when replacing them.
Several tasks are required to install the autosampler:
Inspect external packaging upon receipt for holes, tears, smashed corners, or any other
outward signs of damage from rough handling or abuse during shipment. Inspect all
items during unpacking and notify the carrier immediately of any damage.
NOTE: Do not throw away the factory packaging. Keep it for possible future use.
If the autosampler is shipped or removed from storage during cold weather, allow the
packaged equipment to attain room temperature before opening and exposing to warm,
humid air. Provide four to eight hours for this purpose.
Remove the packing checklist from the shipping container, and check off items against
it. Leave accessories in the packing until ready to install them on the autosampler.
Place the autosampler within 1.2 m (48") of a power outlet. Position the autosampler so
the power supply cord plug is easily accessible (is not blocked) and the plug can be
quickly disconnected if needed. The power supply socket is located on the back of the
autosampler below the power switch. Do not apply power to the power supply until
ready to operate the autosampler.
Attach the Z-drive assembly (PN 323158) to the autosampler arm to allow movement
and function of the sample probe. Figure 5.1 illustrates the Z-drive assembly
components while the following procedure outlines how to install the Z-drive assembly.
Sample Probe
Z-axis Drive
Assembly
Cable Sheath
Z-drive Assembly
Z-axis Slider
Sample Probe
Thumbscrews with
Bushings
Y-axis Lead
Screw Nut
Y-axis Slider
Block
1. Position the Z-drive assembly at the free end of the autosampler arm with the
Z-drive assembly pointing up.
2. Match the 6 x 3-mm grooves in the Y-axis slider block with the guide rails on the
autosampler arm.
3. Slide the block along the arm tube until the holes in the block align with the
matching holes in the Y-axis lead screw nut, as depicted in Figure 5.2.
Z-axis Drive
Cable Sheath
Sample Probe Assembly
Clamp
Z-drive Assembly
4. Secure the Y-axis slider block to the Y-axis lead screw nut using the 12-mm nylon
thumbscrews installed from the top (through the bushings).
CAUTION: Ensure the front white cable (labeled “A”) goes to the left side of the
Z-axis rotor (as viewed from the back of the autosampler. The rear white
cable (labeled “B”) goes to the right side of the Z-axis rotor (as viewed
from the back of the autosampler). Failure to connect the cables correctly
results in autosampler malfunction.
1. Loop the 1.5-mm O.D. PEEK Z-axis drive tubing around the bottom of the Z-axis
rotor groove, as depicted in Figure 5.3.
2. Attach the mounting blocks to the back of the autosampler chassis with the
stainless steel thumbscrews provided.
Mount the blocks with the holes to the far left- and far right-hand sides as viewed
from the back or the rotor will not function properly. Do not tighten the rotor clamp
at this time.
Z-axis Drive
Cable Sheath
Z-drive Assembly
Mounting Block Thumbscrews
(Two Each)
Figure 5.3. 3090 Autosampler, Back View, with Z-axis Rotor and Mounting Blocks
1. Install the clamp in the slot on the Z-axis slider, as depicted in Figure 5.1.
2. Install the sample probe through the slider block and push it through the clamp.
3. Move the Z-axis slider plus attached sample probe to the top of the Z-axis drive.
4. Leave approximately 105 mm (4.1") of the sample probe’s yellow support tube
extending above the top of the Z-axis slider (with the slider at the top of the Z-axis
drive).
6. Verify the probe tip clears the top of the rinse station when the autosampler is in
the home position above the rinse station.
The autosampler arm with the attached Z-axis drive can be manually moved to the
rinse station without damage to the autosampler.
7. Retain the sample transfer tubing with the spiral-wrap tie points at approximately
15 cm and 40 cm (6" and 16") above the top of the Z-axis drive, leaving an
untangled service loop of approximately 13–15 cm (5.1–6") above the probe.
The sample transfer tubing should still have slack remaining when the probe is at
the maximum downward limit.
CAUTION: Do not directly maneuver the sample probe to prevent damaging the
probe.
NOTE: To make fine adjustments to the sample probe X-Y targeting, loosen the
nylon screw on the probe guide plate Q/i- to ¼-turn and move the guide
plate up to 0.2 mm from the original location. Be sure to tighten the nylon
screw before operating the autosampler.
1. Adjust the Z-axis slider (with attached sample probe) so the slider is approximately
3 mm below the top of Z-axis drive.
2. Rotate the Z-axis rotor clockwise so the rotor stop pin is against the rotor stop.
Ensure the PEEK Z-axis drive tubing fully seats in the rotor clamp groove. Other-
wise, the PEEK tubing can slip, resulting in no movement of the Z-axis slider.
4. Manually rotate the Z-axis rotor back and forth several times and check for full
unhindered movement of the Z-axis slider.
5. With the Z-axis in the full-up position, hold the Z-axis slider and move the
sample probe tube up and down so that 3–6 mm (0.11–0.23") extends below the
sample probe guide plate.
Since the peristaltic pump inlet is at the top of the pump and the outlet is at the bottom,
the rinse water flows from the bottom to the top of the rinse station. Up-flow rinsing is
the most effective method for decontaminating the sample probe tube between samples.
Reversing the connections and the rinse water flow reduces the effectiveness of the
rinse station and can cause cross-contamination and unsatisfactory performance.
The waste rinse solution drains from the top of the rinse station by means of a pumped
drain, which is the standard arrangement for draining the rinse station. If using a
pumped drain is not desired, a gravity drain arrangement is satisfactory. This section
contains instructions for both the pumped drain and gravity drain arrangements.
The following procedure outlines how to connect the rinse station using the pumped
drain arrangement.
1. Connect the rinse water source to the onboard peristaltic pump by inserting the
3-mm I.D. Tygon® tubing onto the inlet at the top of the pump.
a. Use approximately 30 cm (12") of the 3-mm I.D. Tygon tubing provided for the
rinse solution uptake.
b. Insert one end of the 3-mm I.D. Tygon tubing onto the outlet at the bottom of
the pump.
Insert the tubing carefully because the peristaltic pump fitting grips the tubing
tightly. Applying too much force can break the fitting.
c. Insert the other end of the 3-mm I.D. Tygon tubing onto the rinse tube inlet at
the bottom of the rinse station.
3. Connect the rinse station to the second channel of the onboard peristaltic pump.
b. Insert the 3-mm I.D. Tygon tubing onto the top outlet of the rinse station.
c. Place the other end of the tubing onto the pump inlet (top of channel 2). Insert
the tubing carefully to avoid breaking the fitting.
4. Connect the second channel of the onboard peristaltic pump to the rinse solution
waste container.
a. Use up to 1.8 m (71") of the tubing provided for the pumped drain.
b. Insert the 3-mm I.D. Tygon tubing onto the peristaltic pump outlet.
Insert the tubing carefully because the rinse station fitting grips the tubing
tightly. Applying too much force can break the fitting.
c. Place the other end of the tubing into the waste container.
Ensure the tubing outlet is not immersed in the waste solution. Immersing of
the drain tube outlet may cause the waste solution to back up and overflow.
The following procedure outlines how to connect the rinse station using a gravity drain.
1. Connect the rinse solution source to the onboard peristaltic pump by inserting the
3-mm I.D. Tygon tubing onto the inlet at the top of the pump.
a. Use approximately 30 cm (12") of the 3-mm I.D. Tygon tubing provided for the
rinse solution uptake.
b. Insert one end of the 3-mm I.D. Tygon tubing onto the outlet at the bottom of
the pump.
c. Insert the other end of the 3-mm I.D. Tygon tubing onto the rinse station inlet at
the bottom of the rinse station.
a. Use up to 1.8 m (71") of the 5-mm I.D. Tygon tubing provided for the gravity
drain.
b. Ensure the waste container is at least 30–60 cm (12–24") lower than the rinse
station outlet.
c. Insert the 5-mm I.D. Tygon tubing onto the rinse station outlet (on top).
Insert the tubing carefully because the rinse station fitting grips the tubing
tightly. Applying too much force can break the fitting.
d. Place the other end of the tubing into the rinse solution waste container.
Ensure the tubing outlet is not immersed in the waste solution. Immersing the
drain tube outlet may cause the waste solution to back up and overflow.
WARNING: Before loading or unloading any sample vial racks on the sample tray,
park the autosampler arm and probe in the home position by cycling the
power off and on. The home position is the initial position when the unit
powers on. Never attempt to load, unload, or reposition a sample vial rack
or sample vial while the autosampler is operating.
NOTE: Keep at least one copy of the assembly instructions provided for each
rack.
NOTE: Correctly placed sample vial racks do not move more than ±2 mm in either
a left-right or forward-backward direction. Tilted sample vials indicate an
improperly placed rack, which must be corrected before operating the
autosampler.
The following procedure outlines how to assemble and place the sample vial racks.
1. Snap the racks together as shown in the instructions included with each rack.
2. Place the sample vial rack so the feet on the rack’s underside engage the locating
ribs on the sample tray’s surface.
WARNING: Use only the external desktop power supply provided with the unit or an
exact replacement.
1. Plug the power supply cord into the autosampler power connector.
2. Plug the power supply cord into the appropriate wall power outlet.
3. Plug the one end of the autosampler cable (PN A001736) to the COM 1 port on
the autosampler back, depicted in Figure 5.6.
Power Receptacle
COM 1 Port
The autosampler cable communications cable is included in the FS 3100 startup kit.
4. Plug the other end into the Sampler port on the back of the FS 3100, depicted in
Figure 5.7.
Verifying Installation
After installing the 3090 Autosampler, verify it is installed correctly. Verifying
installation consists of two parts:
1. Ensure the communications interface between the autosampler and the FS 3100
works.
2. Visually inspect the sample probe, peristaltic pump tubing, and rinse station
and tubing for leaks or signs of damage.
NOTE: Before testing the sample probe, ensure all autosampler components are
installed correctly. Securely tighten all thumbscrews and ensure the
communications cable from the FS 3100 to the COM 1 port on the
autosampler is properly connected.
The following procedure outlines how to test the sample probe operation.
1. Load the autosampler sample tray with an empty sample vial rack.
For information about placing the sample vial rack, see “Assembling and Placing
the Sample Vial Racks” on page 62 in this chapter.
NOTE: The LED power indicator is green. When the Z-drive assembly is in the
home position, the indicator is located behind it.
2. Turn on the autosampler and verify the LED power indicator is on.
3. Through WinFLOW, designate sample positions at the left rear, left front, right
rear, and right front of the sample rack.
5. Command the autosampler to move the sample probe to the designated sample
positions.
6. Check that the sample probe correctly accesses each position and the probe
descends into the center of each sample vial.
CAUTION: Ensure no air bubbles are visible in the rinse uptake tubing before running
samples with the autosampler.
NOTE: If flushing the rinse system during initial startup, first use a 2% nitric acid
solution as the rinse agent.
NOTE: The green LED indicator along the autosampler X-axis lights when the
power is on.
For more information about proper connections, see “Connecting the Rinse Sta-
tion” on page 58 in this chapter.
3. Adjust the peristaltic pump shoe using an Allen wrench on the adjustment screw
until achieving the desired rinse solution flow rate.
Figure 5.8 depicts the location of the peristaltic pump and the adjustment screw.
Two-Channel
Peristaltic
Pump
Peristaltic
Pump
Adjustment
Screw
4. Purge air from the rinse system by placing the rinse solution uptake tubing in the
rinse solution source and running the rinse solution through the rinse station.
The 3090 Autosampler runs until it reaches the end of the sampling sequence.
a. Remove the rinse solution uptake tubing from the rinse solution source.
b. Allow the peristaltic pump to run until all solution drains from the tube
attached to the rinse station outlet.
c. If not using reagent water for the rinse solution, flush the rinse system with
reagent water before shutting down the autosampler.
For more information, see “Flushing the Rinse Station and Flow Path” on
page 68 in this chapter.
Flushing the rinse system during initial startup of the autosampler removes any
contaminants that could cause interference during sample analysis. Flushing the rinse
system after using strong rinse agents prevents flow path degradation and failure, as
well as potentially inaccurate results.
The following procedure outlines how to flush the rinse station and flow path.
2. Run the rinse solution through the rinse station and flow path for 5–10 minutes.
3. After flushing the rinse system, proceed with the sampling sequence or drain the
rinse system as part of the shutdown procedure.
For information about running the sampling sequence, see “Operating the 3090
Autosampler” on page 66 in this chapter. For more information about draining the
rinse system, see the previous section, “Shutting Down the 3090 Autosampler” on
page 67 in this chapter.
Analytical Method
Each configuration includes an analytical method. Read the information carefully
before beginning any analysis for the first time. The analytical method provides the
following information:
• The flow diagram, which describes the component layout, all mixing coils, pump
tubes, and reagents.
• The analytical method, which describes the analysis principles, interferences, raw
materials required, reagent and calibrant preparation, specific operating
precautions, and notes.
Reagent Preparation
The following guidelines should be observed when preparing reagents for use in the
FS 3100 system.
• Degas by placing under a strong vacuum for five minutes. Magnetic stirring or
sonication aids in the degassing process. Vacuum filtration can be used to
satisfy both the filtering and degassing requirements.
• Purge with a stream of nitrogen gas (or other inert gas) through a glass frit for
approximately five minutes.
• After degassing, store the degassed reagents in a tightly sealed container to protect
them from reabsorbing atmospheric gases. For best results, store degassed reagents
under a slight vacuum when not in use.
Calibrant Preparation
Unless stated otherwise, all measurements should be made with Class A glassware to
achieve the greatest accuracy. Use Equation 6.1 to calculate the volume of stock (or
intermediate) calibrant to use.
(Equation 6.1) C1 × V1 = C2 × V2
Solving this equation for the volume of stock solution to be used (V1) obtains the
C2 × V2
following: V 1 = -----------------
-
C1
Since the desired concentration (C2), the final volume (V2), and the concentration of the
stock solution (C1) are all known for any given calibrant concentration in a defined
volume, the volume of stock solution to be used (V1) is easily calculated.
Waste
It is the laboratory’s responsibility to comply with all federal, state, and local
regulations governing waste management, particularly the hazardous waste
identification rules and land-disposal restrictions. In addition, it is the laboratory’s
responsibility to protect air, water, and land resources by minimizing and controlling all
releases from fume hoods and bench operations. Compliance is also required with any
sewage discharge permits and regulations.
WARNING: Failure to correctly install the waste containers can produce toxic
hydrogen cyanide gas.
CAUTION: Failure to open the stopcock during instrument operation may cause
damage to the gas diffusion membrane.
NOTE: Before beginning an analysis, read the entire operating protocol given in
the analytical method.
NOTE: The FS 3100 and the autosampler must be turned on before starting
WinFLOW. After turning on the FS 3100, an eight-second delay occurs
before it resets. Start the software after reset takes place.
1. Connect all reagent lines to the reagents as specified in the analytical method.
2. Ensure all waste lines are connected to the waste containers provided.
See Chapter 3, “Installing Sample and Reagent Waste Containers” on page 39 for
specific instructions.
3. Open the stopcock valve on the reagent waste container or release the clip on the
exit line from the amperometric flowcell.
7. Start WinFLOW.
Upon starting WinFLOW, the Select View dialog box displays, as depicted in
Figure 6.1.
The Specify filename for results dialog box appears, as depicted in Figure 6.4.
15. Go to the Pumps menu and select the appropriate pump to start.
NOTE: The pump can be run at a higher speed to lessen the time it takes for
startup solution to reach the gas diffusion module. Be sure to stabilize the
system at the speed the application runs. Tests are run at lower speed.
NOTE: The sample waste line from the valve module may exhibit some pulsing.
16. Verify all solutions are pumping and flow is smooth by pulling a tube out of the
startup solution for a few seconds, repositioning it in the solution, and watching the
flow of the bubble.
17. Allow the system to continue pumping startup solution while preparing the
method and sample table in WinFLOW.
Software Setup
Refer to the WinFLOW Operator’s Manual for setup instructions.
Sample Analysis
The following sections describe how sample analysis is conducted, from data collection
through low level cyanide testing.
NOTE: This section assumes reagent solution has been pumping for
approximately 10–15 minutes, and the amperometric flowcell has been
conditioned and stabilized with cyanide standards (see Chapter 7, “Cell
Conditioning” on page 87).
Data Collection
The following procedure provides a quick recap of data collection. Refer to the
WinFLOW Operator’s Manual for more information on starting data collection and
initiating baseline data collection.
1. Access the serial communications window from the Window menu in Win-
FLOW.
If the serial communications window does not display this value, use the ampero-
metric detector’s potentiometer, depicted in Figure 6.6, to adjust the value.
Chassis
Potentiometer
Amperometric
Detector
Gas Diffusion
Module Six-port Injection Amperometric
Valve Flowcell
3. Connect the red wire from the detector to the reference electrode plug.
5. Connect the clip on the black wire to the counter electrode near where the
stainless steel fitting goes into the flowcell body.
6. After connecting the three cell electrode wires, observe the signal percent value.
This value should change either up or down and start to return to 50%.
Observing no change after connecting the electrode wires indicates a bad connec-
tion in one of the electrode wires. Reconnect the wires and try again.
NOTE: Baseline stabilization may take five minutes or longer after introducing
reagents.
When deciding if the baseline meets these criteria, keep in mind that the real-time
scaling is dynamic, and the scale adjusts to fit the highest point on the screen at any one
time.
When monitoring the baseline, note the scale on the vertical axis. When the scale is
0–500 pA, it expands and a good baseline could give the appearance of a noisy
baseline.
Figure 6.7 shows a very good baseline. An absorbance reading of 500 pA is equivalent
to approximately 1 ppb, depending on the specific method. Before declaring the
baseline noisy, check the scale on the vertical axis. Once the SYNC peak appears, the
“noisy” looking baseline becomes a flat line.
If using the photometric detector option, establish a stable baseline on startup solution
then switch to reagents. Proceed with the instrument stabilization procedure outlined in
the specific analytical method. If the baseline does not stabilize on reagents within
10 minutes, see Chapter 8, “Troubleshooting” on page 95.
Stopping a Run
Refer to the WinFLOW Operator’s Manual for information on stopping a run.
CAUTION: Siphoning of liquid can occur when disengaging the platens and releasing
tension on the pump tubes.
1. Pump reagent water through the system for at least 10–15 minutes.
If using the photometric detector option, reconnect the startup solution and allow it
to run for 10–15 minutes.
3. After cleaning the system, stop the pump by pressing the rotation direction switch
to standby.
4. Release tension on the pump tubes by pressing down on the tension lever or by
pressing the release latches on both sides of the platen.
UV Digestor Operation
The following procedure outlines how to operate the UV Digestor module.
1. Verify the correct UV lamp and digestor coils are installed for the specific appli-
cation.
See Chapter 7, “The UV Digestor module uses two 312-nm, 8-watt UV lamps (PN
A001682) and two quartz digestor coils (PN 319798). The following procedure
describe how to replace these components.” on page 93 for instructions on chang-
ing these components.
4. Refer to the analytical method for specific instructions on instrument and baseline
stabilization.
Operating Notes
The following sections provide notes for consideration while operating the FS 3100
system.
To decrease the sensitivity of a test to detect higher concentrations, use one or more of
the following techniques:
• The peak shapes are consistent and smooth. The peaks should not be too abrupt, be
reasonably bell-shaped, and not have significant tailing.
• The sample zone dispersion should produce sample-to-sample interaction or
carryover of <1% in most cases.
• The baseline must have a low enough noise component so the minimum
quantitation concentration produces a peak that is at least twice the standard
Clean deionized water is the most critical element for successful low-level analysis.
Impure water is observed in the analysis as a high background, resulting in noise and an
inability to obtain the desired detection limits. Deionized water must meet ASTM
Type I specifications for purity. It is most easily obtained from laboratory purification
systems commercially available from Millipore, Barnstead, or others that deliver water
at approximately 18 mega-ohm resistivity. Supply these units with water pretreated
through distillation or an ion exchange filter system, rather than connecting them
directly to tap water. Another source of acceptable water is to pass ASTM Type II water
through both strongly acidic cation and strongly basic anion exchange resins.
A clean laboratory environment is also important. Proper air circulation and filtering
systems may be necessary in some cases.
Dedicated Glassware
Keeping dedicated glassware for reagents and standards for each analysis is another
good practice. Initially clean the glassware, followed by several rinses with ASTM
Type I water. After each use, rinse the glassware immediately and store completely
filled with ASTM Type I water until the next use. Any container used for the samples
themselves should also be rinsed thoroughly with acid and water.
Reagent Filtering
Most instrumentation systems benefit from filtering reagents used in the analysis. The
filtering process can be a source of contamination either from the filter media itself or
cross-contamination from other uses. If possible, keep separate filtering apparatus. One
inexpensive and efficient filtering method for smaller volumes uses large 60-cc
disposable syringes and readily available syringe filters. The syringe can easily be
cleaned for another use and the filters are disposable. In all cases, rinse the filter itself
with water before filtering reagents unless the filters in use are shown not to be
contaminated. Avoid using most paper filters.
Accuracy
Accuracy describes how closely to an accepted value an experimental value falls. In the
laboratory, accuracy is generally expressed as percent recovery (%R). Calculate %R
using Equation 6.2. Acceptable accuracy is generally 90–110% throughout most ranges
for most analyses.
Experimental value
(Equation 6.2) %R = ---------------------------------------------- × 100
Accepted value
Precision
Standard deviation
(Equation 6.3) %RSD = --------------------------------------------- × 100
Mean
Note that “mean” and “x” are equivalent terms for the average of a set of values. For
small sets of data, determine the standard deviation (S) using Equation 6.4.
∑ (x – x)
2
i
(Equation 6.4) S = =1
i----------------------------
N–1
Where:
Calculate Method Detection Limit (MDL) using the USEPA procedure (40 CFR part
136, Appendix B, “Definition and Procedure for the Determination of the Method
Detection Limit”).
Where:
The following test log provides a convenient format for recording this information. It
can be reproduced and used to log test parameters and results.
Date
Test
Operator
Run number
Analysis range (analyte
concentration)
Cycle duration time
(seconds)
Wavelength (nm)
High standard height
High standard
concentration
Middle standard height
Middle standard
concentration
Low standard height
Low standard
concentration
Initial serial
communications window
readings
Method name
Table name
Results file name
Reagents
Operational notes
Troubleshooting notes
Maintenance notes
This chapter discusses the routine and nonscheduled maintenance of the FS 3100.
NOTE: When analyzing certain types of mining samples, the gas diffusion
membrane may need to be changed weekly or even daily.
Exterior Maintenance
• Wipe exterior surfaces with a dampened soft cloth using mild detergent and water
to remove loose dust. Do not use abrasive cleaners.
• Wipe up any noncorrosive spills immediately with a soft cloth dampened with
water and wipe dry with a soft cloth. Neutralize corrosive spills immediately with
an appropriate compound, rinse thoroughly with water, and air dry. If liquids spill
into the instrument, immediately turn the power off and unplug it from the power
1. With the pump turned off, unscrew the three holding screws on the upper portion
of the gas diffusion module, depicted in Figure 7.1.
Holding Screw
Upper
Manifold
Membrane
Lower
Manifold
Overtightening may rupture the membrane, causing baseline instability and noise.
Potentiometer
A potentiometer is mounted on the amperometric detector just below the bias display.
This potentiometer adjusts the detector bias potential between the reference and
working electrodes over a range of –1.25 V to +1.25 V. Adjust the potentiometer until
achieving a bias reading of 0.00.
• % Full scale = 0.0, which means the current is below the lowest measurable
• % Full scale = 100, which means the current is above the highest measurable
Amperometric Flowcell
Figure 7.2 shows an exploded schematic of the amperometric flowcell.
Screws (6)
Reference Plate
Electrode
Adapter
Fitting
Reference Fitting
Electrode
Port
Counter
Electrode
Alignment
Upper Cell Dowel Pins
Assembly
Working Threaded
Electrode Plate
Lower Cell
Assembly
The silver working electrode requires little maintenance and should never need
replacing. However, periodic cleaning restores method sensitivity. The following
procedure outlines the cleaning procedures for the working electrode.
1. Loosen the screws holding the two halves of the flowcell together.
2. Using a small amount of toothpaste on a cotton swab, gently polish the electrode
surface until any discoloration of the silver is removed and the electrode is shiny.
Cell Conditioning
After cleaning the working electrode, run the flowcell for a period of time to increase
its sensitivity. When first using the flowcell, it produces a noisy, erratic baseline. To
correct this, inject 10-ppm cyanide standards repeatedly until the cell response rises to a
consistent level.
NOTE: The actual flowcell response varies from cell to cell but usually rises to
approximately 7–8 million picoamperes for a 10-ppm cyanide sample.
After about one hour, the cell’s baseline stabilizes.
The reference electrode uses a protected silver-polymer gel design. The silver wire
electrode is immersed and sealed in a polymer gel solution, and is separated from the
main flow-through channel in the flowcell by a low flow frit in the reference electrode.
Refer to Figure 7.2 and the following procedure for replacing the reference electrode.
2. Remove the protective storage boot from the replacement reference electrode.
• Release the tension on the pump tube platens if stopping the pump for longer than
15 minutes.
• Clean and lubricate the platens and rollers every week. Use silicone oil
(PN A001283) to clean.
• Check the pump tubes visually and physically for integrity. With the pump tubes off
the platens, gently stretch the ends. Replace the pump tube if it lacks elasticity. Also
replace flattened or darkened pump tubes.
• A new calibration is required after changing pump tubes due to the resulting
changes and variations in flow rate. Refer to the IPC Pump Operator’s Manual for
a full description of its operation and maintenance.
Clean the 3090 Autosampler both daily and weekly to prevent damage and extend its
life. Clean up spills and remove contaminants, such as abrasives, from the
autosampler’s moving parts. Also, chemically neutralizing spills may be necessary. The
following sections explain daily and weekly cleaning procedures.
Using the 3090 Autosampler often results in spills on autosampler components such as
the sample tray. Clean the autosampler daily by completing the following steps:
For information about shutting down the autosampler, see Chapter 5, “Shutting
Down the 3090 Autosampler” on page 67.
2. Wipe the sample tray, autosampler chassis, and autosampler arm using a towel
dampened with a laboratory grade cleaning agent.
CAUTION: Do not allow the cleaning agent to come into contact with the lead screws.
Also, never lubricate either of the two lead screws.
3. Repeat step 2, using a towel dampened with clear water to remove any remaining
contaminants.
4. Thoroughly dry the sample tray, autosampler chassis, and autosampler arm
using a dry towel.
Weekly Cleaning
Although daily cleaning removes spills and contaminants from most of the autosampler
components, the following procedure outlines how to clean the autosampler more
thoroughly once a week.
WARNING: Never lubricate the lead screws. The lead screw nuts are compounded
with a dry film lubricant. Oiling the lead screws causes gumming, galling,
and binding of the sample probe assembly.
For information about removing the sample tray, see “Replacing the Sample Tray”
on page 92.
3. Wipe loose particles off the Y-axis lead screw with a dry, lint-free cloth.
The Y-axis lead screw is a large metal screw located inside the autosampler arm
tubing, depicted in Figure 7.3.
Z-axis Drive
Sample Probe Assembly Cable Sheath
Clamp
Z-drive Assembly
Slider Block
Sample Probe Guide Plate
Guide Rail
4. Wipe the autosampler exterior and base until clean using a towel dampened with a
laboratory grade cleaning agent, followed by a towel dampened with clear water.
NOTE: Pay special attention to the slider block and guide rails along the tube of
the autosampler arm.
For information about replacing the tray, see “Replacing the Sample Tray” on
page 92.
2. Visually inspect the sample probe, peristaltic pump tubing, and rinse station and
its tubing for leaks or signs of deterioration.
For more information, see the appropriate section in this chapter on replacing vari-
ous components.
CAUTION: Replace the new tubing carefully. Damage may result if applying too
much force.
4. Carefully push the new tubing onto the mounting block fittings.
CAUTION: Applying too much force when removing the sample probe can damage
the Z-drive assembly.
For information about installing the sample probe, see Chapter 5, “Installing the
Sample Probe” on page 57.
2. Move the autosampler arm 20–30 cm (8–12") away from the home position by
gently pushing it.
Moving the autosampler arm ensures the sample probe is not damaged while
replacing the rinse station tubing.
Apply only a linear force when removing the tubing to prevent breaking the fit-
tings.
b. Remove the rinse station tube from the mounting block by lifting the tube
straight up.
5. Carefully push the rinse station tube into the mounting block and rotate it
clockwise ¼ turn.
Apply only a linear force when replacing the tubing to prevent the fittings from
breaking.
3. Move the autosampler arm 20–30 cm (8–12") away from the home position by
gently pushing it.
Moving the autosampler arm ensures the sample probe assembly is not damaged
while replacing the sample tray.
5. Raise the front edge of the damaged tray at least 2.5 cm (1").
If removing the sample tray is difficult, raise the front edge higher before sliding it
forward.
WARNING: Do not attempt to change the UV lamps or coils without first powering off
the module.
NOTE: The light emitted from the lamp ON indicator is not hazardous. The
material of the lamp ON indicator blocks UV spectrum light energy.
2. Unscrew the six screws that hold the top cover in place.
4. Remove the lamp from the sockets by carefully rotating it to release the prongs on
each end.
5. Pull the digestor coil and lamp slightly away from the unit and remove the lamp.
11. After installing the lamp and coil, check for leaks or restrictions in the fittings and
tubing.
12. Connect the appropriate cartridge tubing to the inlet and outlet fittings.
This chapter discusses the most common problems that can occur when using the
FS 3100, along with their most probable causes and corresponding corrective actions.
Before using this chapter as a guide, become thoroughly familiar with the operation and
maintenance information in the previous chapters. If a problem still exists after
reviewing the following charts or if a particular problem is not addressed, contact
OI Analytical Customer Support at (800) 366-1911 or (979) 690-1711 for assistance.
• Ensure the FS 3100 and the autosampler are plugged into an electrical outlet and
the power switches are turned on.
• Review the analytical methods and flow diagram(s) to verify operating parameters.
Electrical/
Chemical Hydraulic Software
Mechanical
Electrical/
Chemical Hydraulic Software
Mechanical
Change only one component or variable at a time and note any changes and results. In
this manner, the problem becomes more clearly defined.
2. Note the purchase date (if available) to verify if warranty service or repair applies.
Symptom Tables
Following are symptoms of some common problems, possible causes, and some
potential solutions.
Table 8.2. Main Unit Symptoms, Probable Causes, and Corrective Actions
FS 3100 does not COM port not working Verify functionality of the
respond to WinFLOW COM port.
Try another COM port.
Busy COM port Close all other programs.
Com port set incorrectly Set the COM port correctly
in the Configure WinFLOW
dialog box.
Serial cable not connected Reconnect both ends of the
securely cable.
Detector nonfunctional or Replace or reseat the detector
not seated fully in the FS in the FS 3100 chassis.
3100 chassis
SYNC peak does not SYNC ignore time set too Shorten the SYNC ignore
get marked long time.
Peak finding parameter too Decrease the requirements
stringent for rise and fall in the peak
finding parameter.
Poor reagents or standards Check peak heights of known
standards.
Remake standards and
reagents.
Sample loop not filling Adjust the valve inject time.
Replace pump tubes.
Valve not switching to Check the settings in the
inject Timed Events Editor screen.
Repair or replace the valve.
No reagent flow Verify reagent flow by
introducing a bubble and
watching it flow through the
tubing.
Refill reagent bottles.
Check for obstructions to
flow.
Table 8.2. Main Unit Symptoms, Probable Causes, and Corrective Actions (Continued)
Table 8.2. Main Unit Symptoms, Probable Causes, and Corrective Actions (Continued)
Injection Valve
Table 8.3 lists the symptoms, probable causes, and corrective actions for the injection
valve.
Table 8.3. Injection Valve Symptoms, Probable Causes, and Corrective Actions
Valve does not rotate Improper software setting Check the setting in the
within the method selected Timed Events Editor screen.
Injection valve loop Oil or grease on the Clean by injecting or
traps air bubbles injection loop walls aspirating soapy water or
methanol into the injection
loop.
Replace the injection loop if
necessary.
Poor sensitivity Oxidized or dirty silver Remove and polish the silver
working electrode working electrode.
Wrong injection valve loop Verify use of a 200-μL
size injection loop.
Injection loop not filling Adjust the settings in the
completely Timed Events screen.
Depleted reference Replace reference electrode.
electrode filling solution
Poor sensitivity– Clogged or coated gas Replace the gas diffusion
continued diffusion membrane membrane.
Incorrect electrode bias Set the correct bias voltage in
voltage the method (0.00 on the
detector potentiometer).
Improperly made reagents Check reagent preparation;
or calibrants remake reagents and
calibrants if necessary.
Bubble trapped in the Pinch the flowcell pull-
flowcell through tube and waste line
for five seconds, then release.
Worn pump tubes Replace the pump tubes.
Noisy baseline Ruptured gas diffusion Replace the gas diffusion
membrane membrane.
Depleted reference Replace the reference
electrode filling solution electrode.
Dissolved gases in reagents Degas all the reagents and
or carrier carrier.
Bubble trapped in the Refill the reference electrode
reference electrode reservoir to eliminate all
reservoir bubbles.
Baseline level (% Gas diffusion membrane Replace the gas diffusion
Full Scale) gradually saturated with water membrane.
rises to 100% over
several days/weeks
Table 8.4. Detector/Flowcell Symptoms, Probable Caues, and Corrective Actions (Continued)
Mini 8 Pump
Table 8.5 lists the symptoms, probable causes, and corrective actions for the Mini 8
Pump.
Table 8.5. Mini 8 Pump Symptoms, Probable Causes, and Corrective Actions
Pump does not power Unit not plugged in Plug in the power cord.
up
Blown fuse Replace the fuse in the back
of instrument.
Pump does not rotate No power to module See “Pump does not power
up.”
Carrier or reagent WinFLOW not set correctly Check the setting in the
flow ceases Timed Events screen.
Leakage in system Check the valve, cartridge,
and fittings for leaks.
Empty reagent containers Refill the reagent containers.
Blockage in flow lines Check fittings for blockages.
Table 8.5. Mini 8 Pump Symptoms, Probable Causes, and Corrective Actions (Continued)
Donor flow surges Worn air pump tube Replace the air pump tube.
Replace the air (orange/
white) pump tube with a
larger pump tube (black/
black or white/white).
Overtightened fittings Avoid using a wrench on the
fitting, which should be
fingertightened.
Replace the tubing on
overtightened fittings.
Dirty UV Digestor coil(s) Replace the UV digestor
coils.
3090 Autosampler
The following sections describe potential problems and solutions for the 3090
Autosampler.
A power system problem can cause a 3090 Autosampler malfunction. This may be
occurring if the green LED power indicator is off. If the autosampler is not functional,
check the power source.
1. Check the electrical outlet and see if the external power supply is plugged in.
2. Check the power switch of the external power supply and see if it is turned on.
2. Check the RS-232 cable to ensure it is plugged into the COM 1 port on the
autosampler and it is properly tightened.
3. Check the host computer to ensure the RS-232 cable is connected to the appropriate
COM port and it is properly tightened.
For more information about connecting the RS-232 cable, see Chapter 5, “Estab-
lishing External Connections” on page 62.
If the RS-232 cable is connected properly and the autosampler is still not
communicating with the PC, ensure WinFLOW is configured correctly.
2. Check the software configuration for the correct COM port selection and baud rate
(9600, N, 8, 1).
1. Ensure the Y-axis slider block and Z-drive assembly are installed.
If the Z-drive assembly is not installed, follow the instructions provided in Chapter
5, “Mounting the Z-drive Assembly” on page 55 to install it.
The home position flag is shown in Figure 8.1. If the flag is damaged, replace the
entire Y-axis slider block.
For information about mounting the Y-axis slider block on the autosampler arm, see
Chapter 5, “Mounting the Z-drive Assembly” on page 55.
Clamp
Z-drive assembly
Figure 8.1. Z-drive Assembly with Y-axis Block Home Position Flag
NOTE: If freeing the Z-drive assembly is not possible, replace it. See Chapter 5,
“Mounting the Z-drive Assembly” on page 55, for information. Order a
new Z-drive assembly from OI Analytical.
This chapter provides a list of replacement parts and support items for the FS 3100 and
its associated options. An asterisk indicates replacement parts that are considered
expendable (XPN). Replace expendable parts regularly, since they may become
deformed or broken. Keep a supply of expendable parts in stock.
Table 9.1 lists the replacement parts for the 3090 Autosampler.
Product Unit PN
Product Unit PN
Table 9.6 lists the replacement parts for the flowcell assembly and related components.
Figure 9.1 depicts the Amperometric Flowcell and its components.
Screws (6)
Reference Plate
Electrode
Adapter Fitting
Reference Fitting
Electrode
Port
Counter Electrode
Alignment
Upper Cell Dowel Pins
Assembly
Working Threaded
Electrode Plate
Lower Cell
Assembly
Table 9.6 lists the replacement parts for the amperometric flowcell.
Table 9.7 lists the replacement parts for the injection valve loop (PN A001626).
Product Unit PN
Product Unit PN
Product Unit PN
Table 9.16 lists the replacement test papers for the system.
Table 9.17 lists part numbers for tools useful in operating the system.
Product Unit PN
Table 9.18 lists the replacement tubing and tubing assemblies for the system.
Product Unit PN
Glossary
Interference filter The interference filter separates the light wavelength that
(photometer) enters the detector before the light passes through the
analytical stream.
Inverse chemistry A colorimetric analysis in which sample addition results in
color reduction (decreased absorbance).
ISE Ion-selective electrode.
Linearity The degree of proportionality between peak height and
analyte concentration.
Mixing coil A series of loops or turns on a cartridge that create mixing
(analytical action and allows a time delay needed for a chemical
cartridge) reaction.
Noise High frequency background fluctuations in voltage caused
by electronic, hydraulic, or chemical disturbances.
Peristaltic pump A pumping system that moves liquids by compressing
flexible tubing between a solid surface (platen) and a roller.
Since the flexible tubing contains the liquid, no contact
occurs with the pump mechanism.
Photodiode A diode that measures light energy intensity from the
change in the breakdown resistance.
Platen latch Located on the sides of the pump that compresses the
flexible tubing in the pump tube holders against the rollers
to create peristaltic pumping action.
Precision Repeatability of a test.
Predilution loop A hydraulic unit designed to dilute the original sample and
(analytical introduce it into the reagent stream.
cartridge)
Pump tube Calibrated plastic tubes installed in the pump platens and
stretched across the pump rollers. They deliver samples,
reagents, and air to the analytical cartridge. The flow rate of
the pump tubes depends on the tube’s internal diameter and
the pump speed. The color-coded shoulders on each tube
identify the internal diameter of the pump tubes.
Reagent A chemical substance or material that aids in or creates a
specific chemical reaction.
Recipient stream The reagent stream containing the analyte after dialysis
(dialyzer) (lower portion of the dialyzer block). Also see “Donor
Stream.”
Risetime Electronic filtering by a second-order Bessel filter. Risetime
in seconds is typically equivalent to two times the time-
constant in seconds.
T
Test 82
Test Performance Log 81
U
USEPA Method OIA-1677 1
UV Digestor Module 2, 35, 77
Digestor Coil 93
Maintenance
Changing the Digestor Coil 93
Changing the UV Lamp 93
Replacement Part Number 111
UV Lamp 93
W
Waste 70, 106
WinFLOW 1, 3, 71, 74
Installing 42
Replacement Part Number 111
Working Electrode 86