Вы находитесь на странице: 1из 1

10.

ERROR STATEMENTS

ERROR CAUSE ELIMINATION

E. d. Un. Number is too small (large negative) to be displayed change DP setting, channel constant setting

E. d. Ow. Number is too large to be displayed change DP setting, channel constant setting

increase table values, change input setting


E. t. Un. Number is outside the table range
(channel constant setting)

increase table values, change input setting


E. t. Ow. Number is outside the table range
(channel constant setting)

E. I. Un.
Input quantity is smaller than permitted input quan-
change input signal value or input (range) setting
tity range

E. I. Ow.
Input quantity is larger than permitted input quantity
change input signal value or input (range) setting
range

E. Hw. A part of the instrument does not work properly send the instrument for repair

E. EE
perform restoration of manufacture setting, upon repeated
Data in EEPROM corrupted
error statement send instrument for repair

E.. SET.
perform restoration of manufacture setting, upon repeated
Data in EEPROM outside the range
error statement send instrument for repair

E. CLR
upon repeated error statement send instrument for repair,
Memory was empty (presetting carried out)
possible failure in calibration

E.. OUT. Analogue output current loop disconnected check wire connection

82 | INSTRUCTIONS FOR USE OM 402UNI

Вам также может понравиться