COLUMN TROUBLESHOOTING
How Sensitive is My Method to changes. To test the sensitivity of your
Reversed-Phase HPLC method, change the mobile phase pH
Mobile Phase pH Changes?
Reproducibility Problems while keeping all other conditions
Not all compounds and not all methods
Caused by Changes in will be sensitive to mobile phase pH. constant and measure the affect on your
Mobile Phase pH Figure 1 shows a plot of retention time separation.
Day-to-day reproducibility problems, versus mobile phase pH for a basic, an
How Can I Avoid Reproducibility
and even column-to-column reproduc- acidic, and a neutral compound. The
Problems Caused by Changes in
ibility problems can be caused by retention of neutral compounds is little
Mobile Phase pH?
changes in mobile phase pH. Sometimes affected by mobile phase pH. However,
a change of as little as 0.1 pH units can the retention of basic compounds can 1st Try to develop separations that are
make the difference in an acceptable change dramatically in the pH range of not sensitive to changes in mobile phase
separation and an unacceptable one 5 to 10 while acidic compounds are pH. For example, if you are separating
(Figure 1). How can you know if your sensitive to pH between 2 and 5. An basic compounds use a mobile phase pH
separation is sensitive to mobile phase HPLC method may not be as robust as below 3 where retention will be little
pH and what can you do to protect you would like if you are operating at a affected by small changes in pH.
yourself from this problem? mobile phase pH where the retention of 2 nd Always make sure to buffer the
one or more analytes is sensitive to pH mobile phase properly. This includes
FIGURE 1 using a sufficient amount of buffer,
An Example of the Effect of Changes in FIGURE 2
usually 25 to 50 mM, and choosing the
Mobile Phase pH on Resolution The Effect of Mobile Phase pH on Basic,
Acidic, and Neutral Compounds right buffer for the pH that you have
chosen to use. See the table below for
40
information on buffer selection.
Base
(methylamphetamine) 3rd Always prepare the mobile phase
Rs = 3.0 the same way and always measure the
Retention Time (minutes)
available in today's market. But there's With today's increased pace of drug
more! ACE columns are now available discovery, fast and efficient methods
in seven bonded phases (C18, C8, C4, are the rule. Short, narrow-bore
CN, Phenyl, AQ, C18-HL) each columns are replacing the conventional
offering a unique, and in some cases, 4.6 x 250 mm versions. ACE HPLC
dramatically different selectivity. ACE columns are available in both 3 µm and 0 1 2 3 4 5 6 7 8
Time (min)
9 10 11 12 13
Further Information
Please contact your distributor to
recieve your copy of the latest Ace
catalogue, which contains additional
information on the full ACE HPLC
column range.
Advanced Chromatography Technologies (ACT), Aberdeen Science & Technology Park, Balgownie Technology Centre, Bridge of Don, Abe rdeen, AB22 8GW, UK.
Phone: +44 (0) 1224 704554 Fax: +44 (0) 1224 826622 E-mail: info@act-hplc.com Web Site: www.act-hplc.com